K Number
K170158
Manufacturer
Date Cleared
2017-06-01

(134 days)

Product Code
Regulation Number
880.5200
Panel
HO
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PowerGlide ST™ Midline Catheter is inserted into a patient's vascular system for short term use to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerGlide ST™ Midline Catheter is suitable for use with power injectors.

Device Description

Bard Access Systems, Inc.'s PowerGlide ST™ Midline Catheter is a sterile, single use device designed to provide access to the patient's vascular system. The device is intended for short term use (

AI/ML Overview

This document describes the premarket notification for a medical device, the PowerGlide ST™ Midline Catheter. It focuses on demonstrating substantial equivalence to a predicate device, not on proving clinical effectiveness through novel AI or diagnostic algorithms. Therefore, many of the requested criteria related to AI performance, ground truth establishment, expert readers, and MRMC studies are not applicable to this type of regulatory submission.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document lists performance tests and states that the "subject device met all predetermined acceptance criteria". However, the specific quantitative acceptance criteria values and the exact reported performance results for each test are not disclosed in this summary. It only indicates that the device passed.

Performance TestAcceptance Criteria (Not Explicitly Stated)Reported Performance
Dilator Outer DiameterWithin design specifications (Bard internal stds)Met criteria
Dilator Effective LengthWithin design specifications (BS EN ISO 11070:2014)Met criteria
Dilator Luer ConnectorComply with ISO 594-1 and 594-2 (6% luer taper)Met criteria
Catheter/Dilator Tip Adhesion Break ForceForce required to remove dilator (Bard internal stds)Met criteria
Dilator Assembly TensileDoes not fail in tension (BS EN ISO 11070:2014)Met criteria
Catheter/Dilator Assembly Burst PressureGreater than maximum use pressure (ISO 10555-1:2013)Met criteria

Biocompatibility Tests:

Biological TestAcceptance Criteria (Not Explicitly Stated)Reported Performance
CytotoxicityPer ISO 10993-5 (2009)Met criteria
SensitizationPer ISO 10993-10 (2010)Met criteria
Intracutaneous ReactivityPer ISO 10993-10 (2010)Met criteria
Acute Systemic ToxicityPer ISO 10993-11 (2006)Met criteria
PyrogenicityPer ISO 10993-11 (2006)Met criteria
HemocompatibilityPer ISO 10993-4 (2002, amended 2006)Met criteria
GenotoxicityPer ISO 10993-3 (2014), ISO/TR 10993-33 (2015)Met criteria
Sterilization ResidualsPer ISO 10993-7Met criteria

2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes used for the described physical and biological performance tests. It also does not discuss data provenance (e.g., country of origin, retrospective/prospective) as these are non-clinical, in-vitro/bench tests on device components, not clinical data sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable. The "ground truth" for this medical device submission is based on engineering specifications, physical/chemical properties, and biological safety standards (ISO standards). It does not involve expert interpretations of medical images or diagnoses that would require human expert consensus.

4. Adjudication method for the test set
This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic accuracy, where there's a need to resolve discrepancies between human readers or between AI and human reads. This document describes bench testing and biocompatibility assessments which do not involve this type of adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. This submission is for a physical medical device (catheter) and does not involve AI or human readers for diagnostic purposes. Therefore, an MRMC study and effects on human reader performance are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. This device is a physical medical catheter, not an algorithm.

7. The type of ground truth used
The "ground truth" for this device's acceptance is based on:

  • Engineering design specifications: The physical dimensions and properties of the dilator and catheter.
  • Industry Standards (ISO/BS EN ISO): Established international and national standards for medical devices, specifically for intravascular catheters, luer connectors, and biocompatibility.
  • Bard internal standards and procedures: Proprietary testing methods and acceptance criteria used by the manufacturer.

8. The sample size for the training set
This question is not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

9. How the ground truth for the training set was established
This question is not applicable for the same reason as point 8.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).