K Number
K170095

Validate with FDA (Live)

Date Cleared
2017-03-08

(56 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vial2Bag™ Direct Connect is indicated to serve as a connecting part between the IV bag and an external IV line. The integrated vial adapter makes it possible to reconstitute and admix drugs from a vial into the infusion solution.

Device Description

The proposed device, Vial2Bag Direct Connect, is intended for use in healthcare facilities or in home environment by the patient or care-giver to aid and support prescribed treatment and therapy.

The proposed device. Vial2Bag Direct Connect is available with 13mm Vial Adapter or 20mm Vial Adapter. The device consists of the Vial2Bag piercing spike and cover, the twist-off connector and an integrated Vial Adapter (13mm or 20mm) for access to the drug/solution vial.

Up to 3 Vial2Bag Direct Connect can be consecutively connected by inserting the spike of one device into the twist off connector of the previous device.

The device does not contain any medicinal substances, and can be used with standard drug vials.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Vial2Bag™ Direct Connect), not an AI/ML device. Therefore, the requested information regarding AI/ML device acceptance criteria and study details is not applicable to this document. The document describes modifications to an existing device and the performance testing conducted to ensure substantial equivalence to a predicate device.

However, I can extract information related to the device's performance testing and general characteristics from the provided text.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists performance tests conducted, but it does not explicitly define specific numerical acceptance criteria for each test. Instead, it states for all tests: "All testing met the required acceptance criteria." This implies that internal, pre-defined acceptance criteria were used and successfully achieved.

Performance Test NameReported Device Performance
Product Functionality According to IFUMet internal performance standards
Total Penetration ForceMet internal performance standards
Cap Detachment ForceMet internal performance standards
Vial Adapter Detachment Force from Drug VialMet internal performance standards
Breaking Torque TestMet internal performance standards
Spike Tip Ductility TestMet internal performance standards
Air Leakage TestMet internal performance standards
PackagingTested per ISO 11607-1; Met required acceptance criteria
Sterilization (SAL)Met minimum SAL 10^-6
Residuals of ETO and ECHMet requirements of ISO 10993-7:2008 for prolonged exposure
Bacterial Endotoxin (LAL method)Acceptable Endotoxin level 0.5 EU/ml or 20 EU/device; sensitivity of 0.005 EU/ml achieved
Cytotoxicity (ISO 10993-5)Successfully completed biocompatibility testing
Sensitization (ISO 10993-10)Successfully completed biocompatibility testing
ASTM Hemolysis (ASTM F756 and ISO 10993-4)Successfully completed biocompatibility testing
Intracutaneous Reactivity (ISO 10993-10)Successfully completed biocompatibility testing
Systemic Toxicity (Acute Systemic Injection, ISO 10993-11)Successfully completed biocompatibility testing
USP Rabbit Pyrogen (USP 151)Successfully completed biocompatibility testing

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes used for each performance test. It only states that "The modifications to the proposed device were evaluated within the Medimop design control system." The data provenance is internal to Medimop Medical Project Ltd., based in Israel. The studies are prospective in the sense that they were conducted specifically for this 510(k) submission to evaluate the modified device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not applicable. The device is a physical medical device, not an AI/ML product that requires expert-established ground truth for performance evaluation in the context of image interpretation or diagnosis. The "ground truth" for the performance tests would be defined by engineering specifications and relevant regulatory standards.

4. Adjudication Method for the Test Set:

Not applicable for a physical device's performance testing. The tests are typically objective measurements against predefined criteria.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable. This is a physical device, not an AI/ML diagnostic tool.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

Not applicable. This is a physical device, not an AI/ML algorithm.

7. Type of Ground Truth Used:

For the performance tests, the "ground truth" is a combination of:

  • Internal performance standards: Defined by the manufacturer (Medimop) for functionality, penetration force, detachment forces, torque, ductility, and air leakage.
  • International Standards: e.g., ISO 11607-1 for packaging, ISO 11135-1:2007 for sterilization, ISO 10993-7:2008 for ETO and ECH residuals, ISO 10993-1 and G95-1 for biocompatibility assessment, and specific ISO 10993 sub-standards, ASTM F756, and USP 151 for various biocompatibility tests.
  • Regulatory requirements: e.g., acceptable Endotoxin levels.

8. Sample Size for the Training Set:

Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable.

{0}------------------------------------------------

TMENT OF H

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 8, 2017

Medimop Medical Project Ltd. Ms. Ilanit Goldgraber Director, Regulatory Affairs 17 Hatidhar Street Ra'anana, 43665 ISRAEL

Re: K170095

Trade/Device Name: Vial2Bag™ Direct Connect Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: January 9, 2017 Received: January 11, 2017

Dear Ms. Ilanit Goldgraber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina Kiang -

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170095

Device Name Vial2Bag Direct Connect

Indications for Use (Describe)

The Vial2Bag™ Direct Connect is indicated to serve as a connecting part between the IV line. The integrated vial adapter makes it possible to reconstitute and admix drugs from a vial into the infusion solution.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY (K170095)

Company Name:

Medimop Medical Projects Ltd 17 Hatidhar St Ra'anana 4366519 Israel Phone: 972-9-7757122 Fax: 972-9-7485916

Contact Person:

Ilanit Goldgraber Director, Regulatory Affairs Phone: 972-9-7778666 Fax: 972-9-7485916 E-mail: ilanit.goldgraber@westpharma.com

US Agent:

Kevin Lentz Sr. Director, Regulatory Affairs West Pharmaceutical Services, Inc. 530 Herman O. West Drive Exton, PA 19341 Phone: (610) 594-4353 Email: kevin.lentz@westpharma.com

Preparation date:

3/7/2017

Classification:

Regulation Name:Intravascular Administration Set
Trade Name:Vial2Bag Direct Connect
Common/Usual Name:Vial2Bag DC
Product Code:LHI
Regulation No.:880.5440
Class:II
Panel Identification:General Hospital Panel
Predicate Devices:Vial2Bag Direct Connect, cleared by 510(k) K140730

Device Description

The proposed device, Vial2Bag Direct Connect, is intended for use in healthcare facilities or in home environment by the patient or care-giver to aid and support prescribed treatment and therapy.

The proposed device. Vial2Bag Direct Connect is available with 13mm Vial Adapter or 20mm Vial Adapter. The device consists of the Vial2Bag piercing spike and cover, the twist-off connector and an integrated Vial Adapter (13mm or 20mm) for access to the drug/solution vial.

{4}------------------------------------------------

Up to 3 Vial2Bag Direct Connect can be consecutively connected by inserting the spike of one device into the twist off connector of the previous device.

The device does not contain any medicinal substances, and can be used with standard drug vials.

Indications for use

The Vial2Bag™ Direct Connect is indicated to serve as a connecting part between the IV bag and an external IV line. The integrated vial adapter makes it possible to reconstitute and admix drugs from a vial into the infusion solution.

Technological Characteristics and Substantial Equivalence Discussion:

The proposed device, Vial2Bag Direct Connect, has the same indications for use and principle of operation as the predicate device. Vial2Bag Direct Connect. from Medical Projects, Ltd. cleared by 510(k) number K140730.

The following modifications have been made to the Vial2Bag Direct Connect originally cleared by 510(k) K140730:

  • Proposed device labeling (IFU) includes optional instructions for the connection of up to 3 Vial2Bag Direct Connect for consecutive connection by inserting the spike of one device into the twist off connector of the previous device.
  • Vial2Bag Direct Connect 20mm S (Short) Spike option reduces the vial piercing spike length . compared to the predicate device
  • . Vial2Bag Direct Connect 20mm C-Spike option contains larger diameter spike compared to the predicate device
  • Packaging changed from Ethyl Vinyl Acetate (EVA) / Tyvek pouch to a PETG blister with Tyvek seal cover stock.

The predicate and proposed devices are designed and manufactured by the same organization and share the same materials and function.

The predicate and proposed devices share the same principle of operation and fundamental technology when utilizing the Vial2Bag Direct Connect.

Areas forComparisonClaimed SubstantiallyEquivalent ProductVial2Bag Direct ConnectK140730Proposed DeviceVial2Bag Direct ConnectComparison
Indicationfor UseThe Vial2Bag™ Direct Connect isindicated to serve as a connectingpart between the IV bag and anexternal IV line. The integratedvial adapter is used to reconstituteand admix drugs from a vial intothe infusion solution.The Vial2Bag™ Direct Connect isindicated to serve as a connectingpart between the IV bag and anexternal IV line. The integratedvial adapter is used to reconstituteand admix drugs from a vial intothe infusion solution.Identical
Areas forComparisonClaimed SubstantiallyEquivalent ProductVial2Bag Direct ConnectK140730Proposed DeviceVial2Bag Direct ConnectComparison
SterilizationMethodEthylene OxideEthylene OxideIdentical
SterilityAssuranceLevelSAL 10-6SAL 10-6Identical
Single useYesYesIdentical
BodyMaterialPolycarbonatePolycarbonateIdentical
ExpirationDate5 years3 yearsModified
VialAdapterDiameter30.4mm to fit 20mm Vials18.5mm to fit 13mm Vials30.4mm to fit 20mm Vials18.5mm to fit 13mm VialsIdentical
LabelingPredicate device labeling (IFU)includes connection instructions toIV bagProposed device labeling (IFU)includes optional instructions forthe connection of up to 3 Vial2BagDirect Connect for consecutiveconnection by inserting the spikeof one device into the twist offconnector of the previous device.Modified
PiercingSpikePlastic - Single LumenVial2Bag Direct Connect 20mm S(Short) Spike reduces the vialpiercing spike length compared tothe predicate deviceVial2Bag Direct Connect 20mmC-Spike contains larger diameterspike compared to the predicatedeviceModified
"TightGrip"featureIncludedIncludedIdentical
PackagingMaterialTyvek / Ethyl Vinyl Acetate(EVA)PETG blister and Tyvek sealModified

{5}------------------------------------------------

{6}------------------------------------------------

Performance Testing:

The modifications to the proposed device were evaluated within the Medimop design control system. A risk assessment was performed to ensure that the proposed device modifications did not introduce any new potential risks. The following tests were performed on the proposed device, Vial2Bag Direct Connect, as a result of the risk assessment to ensure the all potential risks associated with the device modifications were mitigated to acceptable levels.

  • Product Functionality According to IFU ●
  • . Total Penetration Force
  • Cap Detachment Force
  • Vial Adapter Detachment Force from Drug Vial
  • Breaking Torque Test
  • Spike Tip Ductility Test ●
  • . Air Leakage Test
  • . Packaging
Performance Testing Summary
Test NameTesting Standard
Product Functionality Accordingto IFUTested to internal performance standards
Total PenetrationForceTested to internal performance standards
Detachment ForceTested to internal performance standards
Vial Adapter Detachment Forcefrom Drug VialTested to internal performance standards
Breaking Torque TestTested to internal performance standards
Spike Tip Ductility TestTested to internal performance standards
Air Leakage TestTested to internal performance standards
PackagingTested per ISO 11607-1

All testing met the required acceptance criteria.

Sterilization

  • The Vial2Bag Direct Connect is sterilized by Ethylene Oxide in a cycle validated according to ● standard ISO11135-1:2007 using the "overkill method". The sterilization subcontractor is Mediplast Israel Ltd.

{7}------------------------------------------------

  • The sterilization process is validated to a minimum SAL 106. Residuals of ETO and ECH were . tested after two sterilization cycle and met the requirements os ISO 10993-7:2008 for prolonged exposure.

Biocompatibility

  • Representative samples from every lot of product are tested for bacterial endotoxin by a validated ● Limulus Amoebocyte Lysate (LAL) method.
    • The acceptable Endotoxin level is 0.5 EU/ml (Endotoxin Units) or 20 EU/device. O
    • The sensitivity of the LAL assay is 0.005 EU/ml. o
  • A biocompatibility assessment was conducted according to ISO 10993-1 and the FDA Blue Book . Memorandum # G95-1 for this material following Double EtO. The Vial2Bag™ DC has been classified as follows:
    • O Category: Externally Communicating
    • Contact Duration: Prolonged o
    • O Device Body Contact: Blood Path Indirect
  • Based upon this assessment, the following biocompatibility testing has been successfully ● completed.
    • Cytotoxicity (Tested to ISO 10993-5) O
    • Sensitization (Tested to ISO 10993-10 O
    • ASTM Hemolysis (Tested to ASTM F756 and ISO 10993-4) O
    • Intracutaneous Reactivity (Tested to ISO 10993-10) o
    • Systemic Toxicity (Acute Systemic Injection) (Tested to ISO 10993-11) O
    • USP Rabbit Pyrogen (Material Mediated Pyrogenicity) (Tested to USP 151) O

Conclusion

The evaluation of the proposed device, Vial2Bag Direct Connect, does not raise any additional concerns regarding safety and effectiveness and is considered to be substantially equivalent to the predicate device, Vial2Bag Direct Connect, 510(k) K140730.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.