(161 days)
The LS-5™ ePTFE Suture is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery.
Each LSI SOLUTIONS® LS-5™ ePTFE sterile surgical suture with 2 ferrules is held in a customized tray, with a suture release feature and is intended for single patient use. The LS-5™ ePTFE Suture is designed to enable the rapid, easy and reliable loading of suture into compatible LSI SOLUTIONS® Suture Placement Devices. LS-5™ ePTFE Suture is a non-absorbable, monofilament expanded polytetrafluoroethylene (ePTFE) suture. A short length of modified surgical stainless steel tubing, called a ferrule, is attached to each end of the suture for interfacing with LSI SOLUTIONS® Suture Placement Devices. The LS-5™ ePTFE Suture also includes a detachable clear suture tube to keep the suture from tangling. The LS-5™ ePTFE Suture is undyed and contains no additives. No significant loss in tensile strength retention occurs in vivo. The LS-5TM ePTFE Suture is MR safe.
This document describes the FDA's 510(k) clearance for the LS-5™ ePTFE Suture, a Class II medical device. The clearance is based on demonstrating substantial equivalence to a predicate device, the W. L. Gore and Associates Inc. GORE-TEX™ Suture. The information provided outlines the performance testing conducted to support this equivalence.
Here's an analysis of the acceptance criteria and the study conducted:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the LS-5™ ePTFE Suture primarily rely on meeting the United States Pharmacopeia (USP) monograph for Nonabsorbable Sutures and demonstrating comparability to the predicate device, GORE-TEX™ Suture.
| Acceptance Criterion | Reported LS-5™ ePTFE Suture Performance |
|---|---|
| Mechanical Performance: | |
| USP <881> Sutures - Tensile Strength requirements | LSI has performed testing to demonstrate the LS-5™ ePTFE Suture is comparable to the predicate device for tensile strength and complies with USP requirements for Nonabsorbable Sutures Section <881> Sutures - Tensile Strength for 4-0 suture (as the predicate, CV-5 GORE-TEX® Suture, is promoted as comparable to USP size 4-0 surgical suture). |
| USP <871> Sutures – Needle Attachment requirements (for ferrule attachment) | Ferrule Attachment complies with the requirements for standard needle attachments as defined in Sutures – Needle Attachment <871>. |
| USP <861> Sutures - Diameter requirements | The LS-5™ ePTFE Suture is offered with a surgical suture diameter comparable to the predicate device, CV-5 GORE-TEX® Suture, which is designated to approximately correlate with a surgical suture diameter of USP size 2-0. While the internal microporous airspaces of ePTFE slightly exceed USP requirements, LSI has performed testing to demonstrate the LS-5™ ePTFE Suture is comparable to CV-5 GORE-TEX® Suture for surgical suture diameter. The labeling will include a tabular comparison of its diameter and tensile strength to the predicate and USP requirements for 2-0, 3-0, and 4-0. |
| Biocompatibility: | |
| Compliance with ISO 10993-1:2009 Standards | Biocompatibility evaluation conducted in accordance with FDA Guidance for Industry, Use of International Standard ISO 10993-1:2009. The LS-5™ ePTFE Suture is classified as an implant device, contacting blood, for a permanent duration, identical to the predicate device. |
| Biocompatibility Endpoints (Systemic Toxicity, Intracutaneous, Muscle Implantation, Cytotoxicity, Sensitization, Irritation, Pyrogen, Hemocompatibility) | Leveraged existing biocompatibility testing from the ePTFE suture material manufacturer for USP Systemic Toxicity, USP Intracutaneous Toxicity, and USP Muscle Implantation Studies. Additionally, LSI conducted Cytotoxicity, Sensitization, Irritation, Rabbit Pyrogen, and Hemocompatibility testing. A combination of testing and risk assessment was completed to ensure biocompatibility. |
| Other Performance: | |
| Compatibility with LSI Suture Placement Devices | LSI has performed testing to ensure that the LS-5™ ePTFE Suture is compatible with the LSI SOLUTIONS® Suture Placement Devices. |
| Bacterial Endotoxins Testing (BET) | BET performed to confirm the subject device does not contain gram-negative endotoxins. |
| Stability Testing | Stability testing performed to support the labeled 2-year shelf life. |
2. Sample Size for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes (number of sutures tested) for each individual mechanical or biocompatibility test. It indicates that testing was performed "to demonstrate" compliance and comparability.
The data provenance for the biocompatibility testing includes:
- Existing data: Leveraged from the manufacturer of the ePTFE suture material (likely prospective testing originally conducted by the material manufacturer).
- New data: Conducted by LSI Solutions (prospective testing).
The document does not specify the country of origin for the data, but given it's an FDA submission, the testing would generally align with international standards recognized by the FDA.
3. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable to the presented document. The ground truth for this device's performance is established by objective, standardized measurements against recognized industry standards (USP monographs for sutures) and direct comparison to a predicate device, not by expert consensus or interpretation of subjective data. Physical and chemical properties are measured directly.
4. Adjudication method for the test set
This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjective assessments (e.g., medical image interpretation) where multiple experts might disagree. In this case, performance is measured against objective standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There was no MRMC comparative effectiveness study performed. This device is a surgical suture, not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
There was no standalone (algorithm-only) performance study done. This is a physical medical device, not a software algorithm.
7. The type of ground truth used
The ground truth used for demonstrating the LS-5™ ePTFE Suture's performance is:
- USP Monograph Standards: Specific quantitative requirements for suture diameter, tensile strength, and needle attachment.
- Predicate Device Characteristics: Direct comparison of the LS-5™ ePTFE Suture's physical properties (diameter, tensile strength, material composition) to those of the legally marketed GORE-TEX™ Suture.
- International Biocompatibility Standards: Compliance with ISO 10993-1:2009 and related biological endpoints.
8. The sample size for the training set
This section is not applicable. This device is not an AI/ML model, so there is no concept of a "training set."
9. How the ground truth for the training set was established
This section is not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2017
LSI Solutions % Mr. Kevin Bentley Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Rd Victor, New York 14564
Re: K163639
Trade/Device Name: LS-5 ePTFE Suture Regulation Number: 21 CFR 878.5035 Regulation Name: Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture Regulatory Class: Class II Product Code: NBY Dated: May 5, 2017 Received: May 9, 2017
Dear Mr. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -ਟਤ For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name LS-5TM ePTFE Suture
Indications for Use (Describe)
The LS-5™ ePTFE Suture is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for LSASolutions, a company that provides innovative solutions for minimally invasive surgery. The logo is displayed in white against a blue background. The text "INNOVATIVE SOLUTIONS FOR MINIMALLY INVASIVE SURGERY" is written in a smaller font size below the company name.
510(k) SUMMARY
| Submitted By: | LSI SOLUTIONS®, Inc. |
|---|---|
| 7796 Victor-Mendon Road | |
| Victor, NY 14564 |
Contact Person: Jennifer Landry Regulatory Specialist Phone: (585) 760-6429 (585) 742-1039 Fax: Email: jlandry@lsisolutions.com
Date Prepared: May 17, 2017
LS-5TM ePTFE Suture Trade Name:
Common Name: Nonabsorbable Surgical Suture
Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture Classification Name: (21 CFR 878.5035, Product Code NBY)
Device Classification: Class II
Predicate Device:
The predicate device, W. L. Gore and Associates Inc. GORE-TEX™ Suture ePTFE Nonabsorbable Monofilament (hereinafter known as GORE-TEX® Suture), was selected to
{4}------------------------------------------------
demonstrate substantial equivalence to the LS-5™ ePTFE Suture. GORE-TEX® Suture was initially approved under Premarket Approval Number P820083 on December 30, 1985 and was reclassified to Class II on September 9, 1999 (reference 21 CFR Part 878, Docket No. 94P-0347). The LS-5™ ePTFE Suture is designed and manufactured to be comparable to CV-5 GORE-TEX® Suture, which is designated to approximately correlate with a surgical suture diameter of USP size 2-0. The LS-5™ ePTFE Suture is also similar to CV-5 GORE-TEX® Suture for tensile strength and needle attachment. W. L. Gore and Associates Inc. promotes the CV-5 GORE-TEX® Suture as being comparable to USP size 4-0 monofilament polypropylene surgical suture.
LSI SOLUTIONS® Suture Placement Devices and Accessories, which was cleared under 510(k) Premarket Notification K100593 on September 24, 2010, and LSI SOLUTIONS® Suture QUICK LOAD® Products, which was cleared under 510(k) Premarket Notification K031443 on June 18, 2003 are used as reference predicates representative of LSI SOLUTIONS® Suture Placement Devices and the technological differences between the predicate and subject device.
Device Description
Each LSI SOLUTIONS® LS-5™ ePTFE sterile surgical suture with 2 ferrules is held in a customized tray, with a suture release feature and is intended for single patient use. The LS-5™ ePTFE Suture is designed to enable the rapid, easy and reliable loading of suture into compatible LSI SOLUTIONS® Suture Placement Devices. LS-5™ ePTFE Suture is a non-absorbable, monofilament expanded polytetrafluoroethylene (ePTFE) suture. A short length of modified surgical stainless steel tubing, called a ferrule, is attached to each end of the suture for interfacing with LSI SOLUTIONS® Suture Placement Devices. The LS-5™ ePTFE Suture also includes a detachable clear suture tube to keep the suture from tangling. The LS-5™ ePTFE Suture is undyed and contains no additives. No significant loss in tensile strength retention occurs in vivo. The LS-5TM ePTFE Suture is MR safe.
Intended Use
The LS-5™ ePTFE Suture is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery.
{5}------------------------------------------------
The predicate device, GORE-TEX® Suture, is indicated for use in all types of soft tissue approximation, including use in cardiovascular surgery, but not for use in ophthalmic surgery, microsurgery, and neural tissue. It is recommended for use where reduced suture line bleeding during cardiovascular anastomotic procedures is desired.
According to the FDA Guidance for Industry. Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA, any surgery types, body sites, or patient populations where evidence demonstrates that the suture should not be used should be included in the contraindications of the labeling. The LS-5™ ePTFE Suture will include a contraindication that the subject device is not for use in ophthalmic surgery, microsurgery, and neural tissue. Therefore, the difference in indications do not impact the safety and effectiveness of the LS-5™ ePTFE Suture. Both the subject and predicate device are intended to be used in the approximation of all types of soft tissue approximation, including use in cardiovascular surgery, and are contraindicated for use in ophthalmic surgery, microsurgery, and neural tissue.
Technological Characteristics (comparison to Predicate Device)
The LS-5™ ePTFE Suture is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate device. Both the subject and predicate devices are non-absorbable, monofilament suture manufactured from ePTFE. The surgical suture of both the subject and predicate devices is undyed and contains no additives. The LS-5™ ePTFE Suture will be offered swaged on LSI's proprietary stainless steel ferrules for use with LSI's automated suturing technology. The ferrule attachments are the only technological difference between the LS-5™ ePTFE Suture and the predicate device. The ferrules are attached to both ends of the suture in the same manner as standard needle attachments and facilitate the use of the suture with the reference predicate, LSI SOLUTIONS® Suture Placement Devices. The ferrules have identical material characteristics as the ferrules used in the reference device, specifically LSI Non-Absorbable QUICK LOAD® Suture. Similar to standard needle attachments, once suturing is complete, the ferrules are removed from the suture.
The subject device is packaged and sterilized in an equivalent manner and has equivalent labeling claims to the predicate and reference predicate devices, including indications, contraindications,
{6}------------------------------------------------
warnings, cautions and precautions. The subject device is considered substantially equivalent and has the same technological characteristics to its predicate device through comparison in design, intended use, material composition, function, and range of sizes.
Performance Testing Summary
As recommended by the FDA Guidance for Industry, Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA, the LS-5™ ePTFE Suture was subject to the requirements of the United States Pharmacopeia (USP) monograph for Nonabsorbable Sutures. Performance testing includes:
- Sutures Tensile Strength <881> .
- Ferrule Attachment which complies with the requirements for standard needle attachments . as defined in Sutures – Needle Attachment <871>.
The LS-5™ ePTFE Suture is offered with a surgical suture diameter comparable to the predicate device, CV-5 GORE-TEX® Suture, which is designated to approximately correlate with a surgical suture diameter of USP size 2-0. Both the LS-5IM ePTFE Suture and predicate device are manufactured from ePTFE. The internal microporous airspaces of ePTFE that result from expanding pure polytetrafluoroethylene (PTFE) produce a less dense structure that slightly exceed USP requirements for Nonabsorbable Sutures Section <861> Sutures - Diameter. LSI has performed testing to demonstrate the LS-5™ ePTFE Suture is comparable to CV-5 GORE-TEX® Suture for surgical suture diameter.
The LS-5™ ePTFE Suture is also similar to CV-5 GORE-TEX® Suture for tensile strength and needle attachment, which is promoted as being comparable to USP size 4-0 surgical suture. LSI has performed testing to demonstrate the LS-5™ ePTFE Suture is comparable to the predicate device for tensile strength and needle attachment and complies with USP requirements for Nonabsorbable Sutures Section <881> Sutures - Tensile Strength and Section <871> Sutures -Needle Attachment for 4-0 suture. In accordance with the FDA Guidance for Industry, Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA, the labeling of the LS-5™ ePTFE Suture will include a tabular comparison of its diameter and tensile
{7}------------------------------------------------
strength to the predicate device and a tabular comparison of the USP requirements for USP size 2-0, 3-0 and 4-0.
Biocompatibility evaluation for the LS-5™ ePTFE Suture was conducted in accordance with the FDA Guidance for Industry, Use of International Standard ISO 10993-1:2009, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," to demonstrate that the LS-5™ ePTFE Suture is biocompatible and substantially equivalent to the predicate device. The LS-5™ ePTFE Suture is classified as an implant device, contacting blood, for a permanent duration, which is the identical type and duration of patient contact as the predicate device. LSI has leveraged the following biocompatibility testing from the manufacturer of the ePTFE suture material:
- USP Systemic Toxicity Study .
- USP Intracutaneous Toxicity Study .
- USP Muscle Implantation Study .
Additionally, LSI has conducted the following biocompatibility testing:
- Cytotoxicity .
- . Sensitization
- . Irritation
- Rabbit Pyrogen .
- Hemocompatibility .
The ePTFE suture material used in the LS-5™ ePTFE Suture is proven to be remarkably inert and the same polymer that the legally marketed predicate device is made of. A combination of testing and risk assessment was completed to ensure the biocompatibility of the LS-5™ ePTFE Suture.
LSI has also performed testing to ensure that the LS-5™ ePTFE Suture is compatible with the LSI SOLUTIONS® Suture Placement Devices, Bacterial Endotoxins Testing (BET) to confirm that the subject device does not contain gram-negative endotoxins and stability testing to support the labeled 2 year shelf life.
{8}------------------------------------------------
Clinical Testing
The technological characteristics, indications for use, material, manufacturing and sterilization processes are the same as the predicate and reference predicate devices, therefore, no clinical studies were deemed necessary to demonstrate the safety and effectiveness of the subject device.
Substantial Equivalence
Both the GORE-TEX® Suture and the LS-5™ ePTFE Suture are non-absorbable, monofilament ePTFE surgical suture that are indicated for the use in all types of soft tissue approximation, including use in cardiovascular surgery. The differences between the subject and predicate device introduce no new risks and have no negative impact to the safety and effectiveness of the surgical suture. The LS-5™ ePTFE Suture is substantially equivalent to the currently marketed W. L. Gore and Associates, Inc. GORE-TEX® Suture which was approved under Premarket Approval Number P820083 on December 30, 1985 and was reclassified to Class II on September 9, 1999 (reference 21 CFR Part 878, Docket No. 94P-0347).
§ 878.5035 Nonabsorbable expanded polytetrafluoroethylene surgical suture.
(a)
Identification. Nonabsorbable expanded polytetrafluoroethylene (ePTFE) surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from ePTFE and is intended for use in soft tissue approximation and ligation, including cardiovascular surgery. It may be undyed or dyed with an approved color additive and may be provided with or without an attached needle(s).(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.