AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tyvek® Pouches and Rolls with STERRAD® Chemical Indicator are intended to enclose medical devices that are to be terminally sterilized in the STERRAD Sterilization Systems and Cycles, in a single or double pouch configuration, and to indicate, by color change, that the pouch has been exposed to sterilant. After completion of the sterilization process, the pouch/roll is intended to maintain sterility of the enclosed device until used.

The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2014 [Type I]) that changes from red to the color indicated on the comparator bar included on the shelf pack (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD Sterilization Systems.

The pouches and rolls can be used in the STERRAD Sterilization Systems and Cycles shown in Table 1.

Device Description

The modified Tyvek Pouch/Roll with STERRAD Chemical Indicator is constructed from Tyvek plastic films, with the STERRAD Chemical Indicator printed onto the Tyvek film. The self-seal pouch permits sealing of the pouch without need of heat-sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the STERRAD Sterilization Systems.

The modified Tyvek Pouch/Roll with STERRAD Chemical Indicator is intended to enclose medical devices to be terminally sterilized in the STERRAD Sterilization Systems in a single or double pouch configuration. The medical devices are inserted into the pouch/roll, sealed, and then sterilized in the STERRAD Sterilization System.

The only difference between the modified device and the predicate device, is the increased post-processing shelf-life to 12 months.

The STERRAD Chemical Indicator offers an additional way to verify processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. STERRAD Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the hydrogen peroxide. The color of the Chemical Indicator changes from red to the color indicated by the comparator bar included on the shelf pack (or lighter) when exposed to hydrogen peroxide.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the "Tyvek® Pouch/Roll with STERRAD® Chemical Indicator with Extended Post-Processing Shelf-Life."

It's important to note that this document is a 510(k) summary, which is a premarket notification to the FDA. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed scientific publication of an efficacy study. Therefore, some of the requested information (like specific sample sizes for training sets, number of experts for ground truth, or MRMC studies) is typically not included in such summaries.


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criterion for the modification is related to the shelf-life of the product.

Acceptance CriteriaReported Device Performance
Original Device (Predicate K141693): Maintain sterility of the enclosed medical device for 30 days post-sterilization.Reported Performance: The predicate device maintained sterility for 30 days post-sterilization. (This is implied as it was the cleared shelf-life).
Modified Device (K163598): Maintain sterility of the enclosed medical device for 12 months post-sterilization across various STERRAD® Sterilization Systems and Cycles.Reported Performance: "12 Months Sterility Maintenance of the Tyvek Pouches/Rolls with STERRAD Chemical Indicator Post Processing using the STERRAD 100S Sterilization System" - Pass"12 Months Sterility Maintenance of the Tyvek Pouches/Rolls with STERRAD Chemical Indicator Post Processing in the STERRAD 100S Sterilization System using LONG Cycle" - Pass... (and similarly for other STERRAD® systems: NX Advanced, NX Standard, 100NX Standard, 100NX Express, 100NX Duo) - All Pass

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the numerical sample size (e.g., number of pouches/rolls tested for each condition) for the "Maintenance of Package Integrity" verification tests. It only lists the different STERRAD® systems and cycles that were tested to support the 12-month shelf-life.
  • Data Provenance: The data appears to be prospective verification testing conducted by the manufacturer, Advanced Sterilization Products. The country of origin for the data is not specified, but given the FDA submission, it's typically conducted under US or international standards accepted by the FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in a 510(k) summary. The "ground truth" here is the objective measurement of sterility maintenance, likely performed by a microbiology lab, rather than an expert interpretation of data points.

4. Adjudication Method for the Test Set

  • Not applicable as the "ground truth" is based on objective laboratory testing for sterility, not subjective expert assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a medical packaging with a chemical indicator, not an AI-powered diagnostic or assistive technology for human readers. There is no AI component or human reader involvement in interpreting its primary function (sterility maintenance).

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is not an algorithm or AI system. The "performance" being evaluated is the physical and chemical properties of the packaging and indicator.

7. The Type of Ground Truth Used

  • The ground truth for the "Maintenance of Package Integrity" tests is objective laboratory measurements of sterility over time, specifically for 12 months after processing in various STERRAD® sterilization systems. This would involve microbial ingress testing or similar methods to confirm the sealed package prevents contamination. The chemical indicator's functionality (color change) is also objectively evaluated.

8. The Sample Size for the Training Set

  • Not applicable. This product is a physical device being evaluated for its material properties and performance, not a machine learning model that requires a training set. The "testing" referred to in the document is verification testing.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for a machine learning model.

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or a stream.

July 31, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Advanced Sterilization Products Ms. Laurie Cartwright Associate Director, Regulatory Affairs 33 Technology Drive Irvine. CA 92618

Re: K163598

Trade/Device Name: Tyvek® Pouch/Roll with STERRAD® Chemical Indicator Regulation Number: 21 CFR 880.6850 /21 CFR 880.2800 Regulation Name: Sterilization Wrap/Sterilization Process Indicator Regulatory Class: Class II Product Code: KCT, JOJ Dated: July 05, 2017 Received: July 07, 2017

Dear Laurie Cartwright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mark S. Fellman -S

for

Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163598

Device Name

Tyvek® Pouch/Roll with STERRAD® Chemical Indicator

Indications for Use (Describe)

Tyvek® Pouches and Rolls with STERRAD® Chemical Indicator are intended to enclose medical devices that are to be terminally sterilized in the STERRAD Sterilization Systems and Cycles, in a single or double pouch configuration, and to indicate, by color change, that the pouch has been exposed to sterilant. After completion of the sterilization process, the pouch/roll is intended to maintain sterility of the enclosed device until used.

The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2014 [Type I]) that changes from red to the color indicated on the comparator bar included on the shelf pack (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD Sterilization Systems.

The pouches and rolls can be used in the STERRAD Sterilization Systems and Cycles shown in Table 1.

Table 1- Indications for Use
STERRAD® / CycleSingle-channel stainlesssteel lumensOther lumensLoad requirements
STERRAD® 100S≥1 mm ID x ≤125 mm long≥2 mm ID x ≤250 mm long≥3 mm ID x ≤400 mm longMetal and nonmetal≥6 mm ID x ≤310 mmlong≤10 lumens per load
STERRAD NX®with or withoutALLClear™ TechnologySTANDARD≥1 mm ID x ≤150 mm long≥2 mm ID x ≤400 mm longN/A≤10 lumens per load ≤10.7lbs
ADVANCED≥1 mm ID x ≤500 mm longN/A≤10 lumens per load ≤10.7lbs
STERRAD® 100NXwith or withoutALLClear™ TechnologySTANDARD≥0.7 mm ID x ≤500 mm longN/A≤10 lumens per load ≤21.4lbs
EXPRESSN/AN/A≤10.7 lbs on bottom shelfonly
DUON/AN/AAccessory devices (e.g.,cameras and light cords) only2 cameras maximum

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image features a logo with a stylized design. The logo consists of a triangle and the letters 'SP' integrated into a single, continuous line. The entire design is enclosed within a rounded, pill-shaped border, and there is a circled 'R' symbol in the upper right corner, indicating a registered trademark. The logo is presented in a light blue color.

510(k) Summary

Advanced Sterilization Products Tyvek® Pouch/Roll with STERRAD® Chemical Indicator with Extended Post-Processing Shelf-Life

General Information

Submitter Name:Advanced Sterilization ProductsDivision of Ethicon, Inc., a Johnson & Johnson company
Address:33 Technology DriveIrvine, CA 92618
Contact Person:Laurie CartwrightAssociate Director, Regulatory AffairsTelephone: (949) 789-3877Email: lcartwr@its.jnj.com
Date Prepared:July 26, 2017
510(k) Number:K163598
Device Name
Proprietary Name:Tyvek® Pouch/Roll with STERRAD® Chemical Indicator
Common Name:Sterilization Pouch/Roll with Chemical Indicator
Classification Name:Sterilization Wrap/Sterilization Process Indicator
Device Class:Class II
Product Code:KCTJOJ
CFR Section:21 CFR 880.685021 CFR 880.2800

Predicate Device

Tyvek Pouch/Roll with STERRAD Chemical Indicator cleared under K141693.

Device Description

The modified Tyvek Pouch/Roll with STERRAD Chemical Indicator is constructed from Tyvek plastic films, with the STERRAD Chemical Indicator printed onto the Tyvek film. The self-seal pouch permits sealing of the pouch without need of heat-sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the STERRAD Sterilization Systems.

The modified Tyvek Pouch/Roll with STERRAD Chemical Indicator is intended to enclose medical devices to be terminally sterilized in the STERRAD Sterilization Systems in a single or double pouch configuration. The medical devices are inserted into the pouch/roll, sealed, and then sterilized in the STERRAD Sterilization System.

The only difference between the modified device and the predicate device, is the increased post-processing shelf-life to 12 months.

The STERRAD Chemical Indicator offers an additional way to verify processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. STERRAD

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The letters are white and are set against a light blue background. There is a registered trademark symbol in the upper right corner of the logo.

Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the hydrogen peroxide. The color of the Chemical Indicator changes from red to the color indicated by the comparator bar included on the shelf pack (or lighter) when exposed to hydrogen peroxide.

The product codes for all Tyvek Pouches and Rolls with STERRAD Chemical Indicator are shown in the following table:

Product TypeProduct Codes
Tyvek Self-Seal Pouches with STERRAD ChemicalIndicator (123XX)12320, 12326, 12335, 12332, 12342, 12340, 12348,12356
Tyvek Rolls with STERRAD Chemical Indicator(124XX)12407, 12410, 12415, 12420, 12425, 12435, 12442,12450
Tyvek Heat-Seal Pouches with STERRADChemical Indicator (125XX)12521, 12526, 12532, 12541, 12548, 12543*, 12544*,12557*, 12558*, 12559*, 12560*
  • Indicates Product Codes exclusively for markets outside the USA.

Intended Use/Indications for Use

The intended use of the Tyvek Pouch/Roll with STERRAD Chemical Indicator, as described in the 21 CFR 880.6850/21 CFR 880.2800, has not changed as a result of the proposed modification. The Tyvek Pouch/Roll with STERRAD Chemical Indicator (predicate and modified device) is intended to enclose a medical device that is to be sterilized in the STERRAD Sterilization System, and to indicate that the pouch/roll has been exposed to hydrogen peroxide during the sterilization process.

The modified device indication for use is provided below:

Tyvek® Pouches and Rolls with STERRAD® Chemical Indicator are intended to be used to enclose medical devices that are to be terminally sterilized in the STERRAD Sterilization Systems and Cycles, in a single or double pouch configuration, and to indicate, by color change, that the pouch has been exposed to sterilant. After completion of the sterilization process, the pouch/roll is intended to maintain sterility of the enclosed(s) device until used.

The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2014 [Type I]) that changes from red to the color indicated on the comparator bar included on the shelf pack (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD Sterilization Systems.

The pouches and rolls can be used in the STERRAD Sterilization Systems and Cycles shown in Table 1.

STERRAD® / CycleSingle-channel stainless steel lumensOther lumensLoad requirements
STERRAD® 100S≥1 mm ID x ≤125 mm long≥2 mm ID x ≤250 mm long≥3 mm ID x ≤400 mm longMetal and nonmetal≥6 mm ID x≤310 mm long≤10 lumens per load
STERRAD NX® with orwithout ALLClear™TechnologySTANDARD≥1 mm ID x ≤150 mm long≥2 mm ID x ≤400 mm longN/A≤10 lumens per load≤10.7 lbs
ADVANCED≥1 mm ID x ≤500 mm longN/A≤10 lumens per load≤10.7 lbs
STERRAD® 100NXwith or withoutALLClear™ TechnologySTANDARD≥0.7 mm ID x ≤500 mm longN/A≤10 lumens per load≤21.4 lbs
EXPRESSN/AN/A≤10.7 lbs on bottomshelf only
DUON/AN/AAccessory devices (e.g.,cameras and light cords).only 2 camerasmaximum

Table 1- Indications for Use

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The letters are white and outlined in a light blue color. The logo is enclosed in a rounded rectangle, also in light blue. There is a small registered trademark symbol in the upper right corner of the logo.

Non-Clinical Data

Verification testing was conducted in support of the extended post-processing shelf-life to the Tyvek Pouch/Roll with STERRAD Chemical Indicator that is the subject of this submission; all testing yielded passing results. This testing is summarized in the following table.

VerificationTestingDescriptionResults(Pass/Fail)
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing using theSTERRAD 100S Sterilization SystemPass
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing in the STERRAD100S Sterilization System using LONG CyclePass
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing in the STERRADNX Sterilization System using ADVANCED CyclePass
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing in the STERRADNX Sterilization System using STANDARD CyclePass
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing in the STERRAD100NX Sterilization System using STANDARD CyclePass
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing in the STERRAD100NX Sterilization System using EXPRESS CyclePass
Maintenance ofPackageIntegrity12 Months Sterility Maintenance of the Tyvek Pouches/Rolls withSTERRAD Chemical Indicator Post Processing in the STERRAD100NX Sterilization System using DUO CyclePass

Summary of Performance Testing

Clinical Data

No clinical data was submitted in support of this Premarket Notification.

Summary

The Tyvek Pouch/Roll with STERRAD Chemical Indicator with Extended Post-Processing Shelf-Life and its predicate device, the Tyvek Pouch/Roll with STERRAD Chemical Indicator cleared under K141693, have the same intended use, technological characteristics, design, materials, and principles of operation. Results of maintenance of package integrity verification studies demonstrated that after completion of the sterilization process, the modified device maintains sterility of enclosed medical device(s) for a period of 12 months compared to period of 30 days for the predicate device. Furthermore, the modified device indication for use statement has been updated for clarity (including moving of the post processing shelflife dating to the description section of the Instruction for Use). Refer to the following tables for comparisons between the modified and the predicate device characteristics.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with the letters "ASP" in white against a light blue background. The "A" is represented by a triangle, and the "S" and "P" are stylized and connected. A small registered trademark symbol is visible in the upper right corner of the logo.

Predict Tyvek® Pouch/Roll withModified Device: Tyvek® Pouch/Roll with STERRAD®Chemical Indicator with Extended Post-Processing Shelf-Life
STERRAD® Chemical Indicator (K141693)(K163598)
Intended UseThe Tyvek Pouch/Roll with STERRAD Chemical Indicator isintended to enclose a medical device that is to be sterilized inthe STERRAD Sterilization System, and to indicate that thepouch/roll has been exposed to hydrogen peroxide during thesterilization process.Same
Indications forUseTyvek® Pouches and Rolls with STERRAD® ChemicalIndicator are intended to be used to enclose medical devicesthat are to be terminally sterilized in the STERRADSterilization Systems and Cycles, in a single or double pouchconfiguration, and to indicate, by color change, that the pouchhas been exposed to sterilant. After completion of thesterilization process, the pouch/roll maintains sterility untilthe pouch/roll is used; in the United States, this is for a periodof 30 days.The STERRAD Sterilization Systems and Cycles that can beused are shown in Table 1. The product codes for all TyvekPouches and Rolls with STERRAD Chemical Indicator arelisted in Table 2.The pouches and rolls are printed with a chemical indicatorbar which is a process indicator (ISO 11140-1:2005) thatchanges from red to yellow (or lighter) when exposed tohydrogen peroxide vapor during processing in the STERRADSterilization Systems.Tyvek® Pouches and Rolls with STERRAD® Chemical Indicatorare intended to be used to enclose medical devices that are to beterminally sterilized in the STERRAD Sterilization Systems andCycles, in a single or double pouch configuration, and to indicate,by color change, that the pouch has been exposed to sterilant.After completion of the sterilization process, the pouch/roll isintended to maintain sterility of the enclosed device(s) until used.The pouches and rolls are printed with a chemical indicator barwhich is a process indicator (ISO 11140-1:2014 [Type I]) thatchanges from red to the color indicated on the comparator barincluded on the shelf pack (or lighter) when exposed to hydrogenperoxide vapor during processing in the STERRAD SterilizationSystems.The pouches and rolls can be used in the STERRAD SterilizationSystems and Cycles shown in Table 1.

Comparison of Device Characteristics

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for ASP. The logo is a rounded rectangle with the letters "ASP" in white inside. The "A" is a triangle, and the "P" is a backwards "P". There is a registered trademark symbol in the upper right corner of the logo.

Predict Tyvek® Pouch/Roll with STERRAD® Chemical Indicator (K141693)Modified Device: Tyvek® Pouch/Roll with STERRAD® Chemical Indicator with Extended Post-Processing Shelf-Life (K163598)
STERRAD® / CycleSingle-channel stainless steel lumensOther lumensLoad requirementsSTERRAD® / CycleSingle-channel stainless steel lumensOther lumensLoad requirements
STERRAD® 100S≥1 mm ID x ≤125 mm long≥2 mm ID x ≤250 mm long≥3 mm ID x ≤400 mm longMetal and nonmetal≥6 mm ID x ≤310 mm long≤10 lumens per loadSTERRAD® 100S≥1 mm ID x ≤125 mm long≥2 mm ID x ≤250 mm long≥3 mm ID x ≤400 mm longMetal and nonmetal≥6 mm ID x ≤310 mm long≤10 lumens per load
STERRAD® 50≥1 mm ID x ≤125 mm long≥2 mm ID x ≤250 mm long≥3 mm ID x ≤400 mm longMetal and nonmetal≥6 mm ID x ≤310 mm long≤10 lumens per loadSTERRAD NX® with or without ALLClear™ TechnologySTANDARD≥1 mm ID x ≤150 mm long≥2 mm ID x ≤400 mm longADVANCED≥1 mm ID x ≤500 mm longN/A≤10 lumens per load≤10.7 lbsN/A≤10 lumens per load≤10.7 lbs
STERRAD® 200≥1 mm ID x ≤125 mm long≥2 mm ID x ≤250 mm long≥3 mm ID x ≤400 mm longTeflon® and polyethylene≥6 mm ID x ≤310 mm long≤12 lumens per load≤36.48 lbsSTERRAD® 100NX with or without ALLClear™ TechnologySTANDARD≥0.7 mm ID x ≤500 mm longEXPRESSN/ADUON/AN/A≤10 lumens per load≤21.4 lbsN/A≤10.7 lbs on bottom shelf onlyN/AAccessory devices (e.g., cameras and light cords) only 2 cameras maximum
STERRAD NX®Standard≥1 mm ID x ≤150 mm long≥2 mm ID x ≤400 mm longAdvanced≥1 mm ID x ≤500 mm longN/A≤10 lumens per load≤10.7 lbsN/A≤10 lumens per load≤10.7 lbs
STERRAD 100NX®Standard≥0.7 mm ID x ≤500 mm longEXPRESSN/ADUON/AN/A≤10 lumens per load≤21.4 lbsN/A≤10.7 lbs on bottom shelf onlyN/AAccessory devices (e.g., cameras and light cords) only 2 cameras maximum

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The "A" is represented by a triangle, and the "S" and "P" are connected with a single line. The logo is enclosed in a rounded rectangle with a light blue background. There is a registered trademark symbol in the upper right corner of the logo.

Predict Tyvek® Pouch/Roll withSTERRAD® Chemical Indicator (K141693)Modified Device: Tyvek® Pouch/Roll with STERRAD®Chemical Indicator with Extended Post-Processing Shelf-Life(K163598)
Design FeaturesTwo films which are heat sealed together on three sides forindividual pouches and on two sides for roll form pouches.A Chemical Indicator (CI) bar is printed onto the Tyvek sideof the pouch. For the self-seal pouch, the CI is printed ontoback surface of the Tyvek. For the heat-seal pouch and roll,the CI is printed onto the Tyvek, but is encased in the seam ofthe pouch between the Tyvek and transparent film.Same
Chemical IndicatorSTERRAD indicator inkSame
Chemical Indicator Device DesignThe color of the Chemical Indicator changes from red to color indicated by comparator bar (or lighter) when exposed to hydrogen peroxide vapor during the sterilization process in the STERRAD Sterilization Systems.Same
Post-Processing Sterility Shelf-Life30 days12 months

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The letters are white and are set against a light blue background. The logo is enclosed in a rounded rectangle shape. There is a registered trademark symbol in the upper right corner of the logo.

Predict Tyvek® Pouch/Roll withSTERRAD® Chemical Indicator (K141693)Modified Device: Tyvek® Pouch/Roll with STERRAD®Chemical Indicator with Extended Post-Processing Shelf-Life
BiocompatibilityIn vivo testing demonstrated that the pouch samples, afterSTERRAD processing, met the acceptance criteria per ISO10993-11 and 10993-10 and can be safely used.No new testing. Predicate Device testing applies to modifieddevice; no change to device materials
End Point / PostProcessing ColorStabilityTyvek Pouches/Rolls, after STERRAD processing,maintained the endpoint CI color reaction.No new testing. Predicate Device testing applies to modifieddevice; no change to device materials
MaterialCompatibilitySeal strength test, microbial barrier properties, burst test,and peel open test demonstrated acceptable materialcompatibility characteristics of the Tyvek Pouches/Rollsafter STERRAD processingNo new testing. Predicate Device testing applies to modifieddevice; no change to device materials
Sterilant PenetrationEfficacyThe sterilant penetrated through the "double pouch"configuration under the worst case half-cycle conditionsand the sterility assurance level (SAL) of 10-6 wasdemonstrated.No new testing. Predicate Device testing applies to modifieddevice; no change to device materials
Seal StrengthSeal strength of the processed Tyvek Pouches/Rolls ismaintained at the end of pre-processing shelf life.No new testing. Predicate Device testing applies to modifieddevice; no change to device materials
Chemical Indicator(CI) FunctionalityCI functionality of the processed Tyvek Pouches/Rolls ismaintained at the end of pre-processing shelf life.No new testing. Predicate Device testing applies to modifieddevice; no change to device materials
Maintenance ofPackageIntegrity30 days sterility maintenance post processing inSTERRAD Sterilizers12 months sterility maintenance post processing in STERRADSterilizers

Comparison of Biocompatibility and Performance Testing

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo with the letters ASP inside of a rounded rectangle. The letters are white, and the rectangle is a light blue color. There is a registered trademark symbol in the upper right corner of the rectangle.

In conclusion, the intended use, operating principle, performance specification, material and design of the modified device are same as the predicate device and the minor differences between the modified device and its predicate device do not raise any new questions of safety or effectiveness. Thus, the Tyvek Pouch/Roll with STERRAD Chemical Indicator with Extended Post-Processing Shelf-Life is substantially equivalent to the predicate device cleared in K141693.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).