K Number
K163544
Date Cleared
2017-01-12

(27 days)

Product Code
Regulation Number
862.1150
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lumipulse G Prolactin Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G Prolactin on the LUMIPULSE G System.

Device Description

Lumipulse G Prolactin Calibrator kit contains 2 bottles (1,5 mL each) of Lumipulse G Prolactin Calibrators 1 and 2. Calibrator 2 contains prolactin in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide.

Lumipulse® G Prolactin Calibrators CAL 230954, Liquid 1x2 concentrations Each calibrator kit contains one bottle each of Calibrators 1 – 2. The calibrator kit is packaged separately.

CAL 1 0 ng/mL Prolactin calibrator (1 x 1.5 mL)
CAL 2 400 ng/mL Prolactin calibrator (1 x 1.5 mL)

Contains prolactin in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide.

AI/ML Overview

The provided text describes the performance characteristics and studies for the Lumipulse® G Prolactin Calibrators, a medical device used for in vitro diagnostic calibration.

Here's an analysis of the acceptance criteria and the studies performed, based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

This device is a calibrator, not a diagnostic test with typical sensitivity/specificity metrics. The performance criteria focus on its ability to provide stable and accurate reference points for the Lumipulse G Prolactin assay.

Acceptance Criteria CategorySpecific Criteria / TestReported Device Performance
TraceabilityTraceable to 3rd International Standard, 1988 (code: 84/500) provided by NIBSC.Calibrators prepared gravimetrically, traceable to NIBSC 3rd International Standard (84/500).
Value Assignment (Calibrator 2)Acceptable rate mean ratio between secondary calibrator and Lumipulse G Prolactin Calibrator 2: 0.95 - 1.05.Achieved acceptable rate mean ratio (0.95 - 1.05).
Value Assignment (Target Range)Calibrator 1: 0 ng/mLCalibrator 2: 380 – 420 ng/mLCalibrator 1: 0 ng/mLCalibrator 2: 400 ng/mL (within target range).
Shelf-Life Stability (Long-term)12 months at 2-10°C, with maintained sensitivity, specificity (implied by assay performance), and reproducibility.Product passed 13-month test; 12 months shelf-life confirmed (stored at 2-10°C).
Transport Simulation (Temperature Stress - Tokyo)Visual assessment: 0/28 under stress conditions meeting specification.Sensitivity (S/N): ≥51 under stress conditions.Accuracy (vs. test values): within 100 ± 20% under stress conditions.Visual assessment: 0/28 under stress conditions, met specification.Sensitivity (S/N): 171 under stress conditions, met S/N≥51 specification.Accuracy: 99-109% under stress conditions, met 100 ± 20% specification.
Real-Time (Intended Storage) Stability (Malvern)Calibrators determined stable at intended storage conditions for up to 6 months (interim results). Will continue to support 12 months.Stable for up to 6 months at 2-10°C for all 3 lots (ongoing study).
Transport Simulation Stability (Malvern)%CV ≤ 10% for all transport conditions.Met %CV≤ 10% at study initiation (ongoing study).
Open-vial (In-Use) StabilityMean concentration of each panel from each time point within ±10% mean concentration at Study Initiation.%CV value of each time point ≤ 10%.Testing continues until data supports 12 months (ongoing study).

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly delineate a "test set" in the context of typical algorithm evaluation (e.g., retrospective vs. prospective patient data). Instead, the studies involve multiple lots of the calibrators themselves.

  • Sample Size for Calibrators:
    • Long-term stability (Tokyo): 3 lots of reagents (calibrators and samples).
    • Value Assignment: 10 replicates of Lumipulse G Prolactin Calibrators and 10 replicates of secondary calibrators for calculation and adjustment.
    • Real-Time (Intended Storage) Stability (Malvern): 3 lots of calibrators.
    • Transport Simulation (Malvern): Not specified number of "calibrators" but describes the conditions and testing.
    • Open-vial (In-Use) Stability: Not specified number of calibrator vials directly, but "panels" are used for evaluation.
  • Data Provenance:
    • Tokyo, Japan: Long-term stability and one transport simulation study were conducted here using the LUMIPULSE Forte (f) instrument (which uses the same cartridges and calibrators as the LUMIPULSE G1200 System).
    • Malvern, PA, USA: Real-time (intended storage) stability and a separate transport simulation study are ongoing at this location using the LUMIPULSE G1200 System.

Since this is a calibrator, the "data" primarily refers to the performance of the calibrator itself and its ability to maintain its specified values over time and under various conditions, rather than a patient dataset in the traditional sense. These are prospective studies of the calibrator product.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

N/A. This is a calibrator, not a diagnostic device that interprets clinical output requiring expert review of medical images or patient data to establish ground truth. The "ground truth" for the calibrator levels and traceability is established through reference to an international standard (NIBSC 3rd International Standard, 1988 (code: 84/500)) and gravimetric preparation.


4. Adjudication Method for the Test Set

N/A. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple human readers or experts provide independent assessments that then need to be reconciled for a definitive ground truth. This is not applicable for a device like a calibrator, where performance is measured against established analytical standards and physical properties.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

N/A. This is a calibrator for an in vitro diagnostic assay, not an AI-powered diagnostic device or an imaging system. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

N/A. This is an in vitro diagnostic calibrator, meaning it's a physical and chemical component used to calibrate an automated assay system. It doesn't involve "algorithms" or "human-in-the-loop performance" in the way an AI diagnostic device would.


7. The Type of Ground Truth Used

The ground truth for the calibrators is established through:

  • International Standard Traceability: The calibrators are traceable to the 3rd International Standard, 1988 (code: 84/500) provided by the National Institute for Biological Standards and Control (NIBSC).
  • Gravimetric Preparation: The calibrators are prepared gravimetrically, meaning their concentrations are determined by precise weighing of components to achieve the desired analyte levels.
  • Secondary Calibrator Rate-Matching: Calibrator 2 is rate-matched to a respective secondary calibrator.

This ground truth is based on established international standards and precise laboratory manufacturing and measurement techniques, rather than clinical outcomes or pathology.


8. The Sample Size for the Training Set

N/A. There is no "training set" in the context of artificial intelligence or machine learning for this device. The device is a calibrator, not an algorithm that learns from data.


9. How the Ground Truth for the Training Set Was Established

N/A. As stated above, this device does not involve a training set for an algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an emblem featuring a stylized caduceus-like symbol, with three human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.

January 12, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FUJIREBIO DIAGNOSTICS, INC DIANA DICKSON DIRECTOR, CLINICAL AND REGULATORY SCIENCE 201 GREAT VALLEY PARKWAY MALVERN PA 19355

Re: K163544

Trade/Device Name: Lumipulse® G Prolactin Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT Dated: December 15, 2016 Received: December 16, 2016

Dear Diana Dickson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For:

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163544

Device Name Lumipulse® G Prolactin

Indications for Use (Describe) Lumipulse G Prolactin Calibrators

Lumipulse & Prolactin Calibrators are for in vitro diagnostic use in the calibration of Lumipulse & Prolactin on the LUMIPULSE G System.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes. To the right of the graphic is the company name, "FUJIREBIO" in large, blue letters. Below the company name is the text "Diagnostics, Inc." in smaller, black letters.

Section 5 510(k) SUMMARY

Lumipulse® G Prolactin

January 6, 2017 Date:

  • 510(k) Number: K163544
  • Submitter: Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355
  • Contact Person: Stacey Dolan Manager, Requlatory Affairs Phone: (610) 240-3843 Fax: (610) 240-3803 Email: dolans@fdi.com
  • Lumipulse® G Prolactin Calibrators Device Name: Classification: Class II JIT 75, Clinical Chemistry 21 CFR § 862.1150
  • Predicate Device: K052982 ROCHE DIAGNOSTICS Cobas Elecsys Prolactin II Calset

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Device Description:

Lumipulse G Prolactin Calibrator kit contains 2 bottles (1,5 mL each) of Lumipulse G Prolactin Calibrators 1 and 2. Calibrator 2 contains prolactin in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide.

Lumipulse® G Prolactin Calibrators CAL 230954, Liquid 1x2 concentrations Each calibrator kit contains one bottle each of Calibrators 1 – 2. The calibrator kit is packaged separately.

  • CAL 1 0 ng/mL Prolactin calibrator (1 x 1.5 mL)
    CAL 2 400 ng/mL Prolactin calibrator (1 x 1.5 mL)
  • Contains prolactin in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for "FUJIREBIO Diagnostics, Inc." The logo features a stylized graphic on the left, consisting of blue and green shapes forming a partial circle with a white teardrop shape in the center. To the right of the graphic, the word "FUJIREBIO" is written in large, bold, blue letters. Below "FUJIREBIO", the words "Diagnostics, Inc." are written in a smaller, thinner font, also in blue.

Device Name:

Lumipulse G Prolactin Calibrators

Device Intended Use:

Lumipulse G Prolactin Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G Prolactin on the LUMIPULSE G System.

Device Indications for Use:

Same as Intended Use

Substantial Equivalence:

Comparison between the Lumipulse G Prolactin Calibrators and Roche Diagnostics Cobas Elecsys Prolactin II Calset - C-

Calibrator Set
Similarities and Differences
ParameterLumipulse GProlactin(Proposed Device)Cobas Elecsys Prolactin II Calset(Predicate Device)K052982
Intended UseLumipulse G Prolactin Calibrators arefor in vitro diagnostic use in thecalibration of Lumipulse G Prolactin onthe LUMIPULSE G System.Elecsys Prolactin II CalSet is usedfor calibrating the quantitativeElecsys Prolactin II assay on theElecsys immunoassay systems.
Instrument SystemLUMIPULSE G SystemElecsys and cobas eimmunoassay analyzers
AnalyteProlactinSame
Number of Calibrators2 calibratorsSame
Standardization/TraceabilityThe calibrators for use with LumipulseG Prolactin are prepared gravimetricallyand are traceable to the 3rd InternationalStandard, 1988 (code: 84/500) providedby the National Institute for BiologicalStandards and Control (NIBSC).The assay is standardized usingthe 3rd IRP WHO ReferenceStandard 84/500
Level2 levels0 ng/mL and 400 ng/mL2 levelsLow: approx. 2 µIU/mLHigh: approx. 2,000 µIU/mL
FormLiquid - ready to useSame
MatrixSodium chloride in Tris buffer withprotein stabilizer (bovine). Preservative:sodium azideBuffered equine serum

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and white shapes. To the right of the graphic is the text "FUJIREBIO" in a bold, blue sans-serif font. Below "FUJIREBIO" is the text "Diagnostics, Inc." in a smaller, black sans-serif font.

Performance Characteristics

Data were generated using the LUMIPULSE G1200 System.

    1. Analytical performance:
    • a. Precision/Reproducibility: Not applicable.
    • b. Linearity/assay reportable range: Not applicable.
    • c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The calibrators for use with Lumipulse G Prolactin are prepared gravimetrically and are traceable to the 3th International Standard, 1988 (code: 84/500) provided by the National Institute for Biological Standards and Control (NIBSC).

Value Assignment

Prolactin Antigen Grade is obtained from CLINIQA. The working stock solution is prepared by dissolving Prolactin antigen in 2-amino-2-hydroxymethyl-1,3propandiol-based buffer. The working stock solution is assigned a provision value based on the measurement using LUMIPULSE G Prolactin on the LUMIPULSE G System with secondary calibrators. Lumipulse G Prolactin Calibrators are made by calculating the amount of working stock solution and 2-amino-2-hydroxymethyl-1,3propandiol-based buffer required to achieve the desired 400 ng/mL calibrator value (Lumipulse G Prolactin Calibrator 2) and adding them together gravimetrically. The 0 ng/mL Calibrator (Lumipulse G Prolactin Calibrator 1) only contains the 2-amino-2hydroxymethyl-1,3-propandiol-based buffer. Lumipulse G Prolactin Calibrator 2 is rate-matched to the respective secondary calibrator based on the measurement in Lumipulse G Prolactin and adjusted if necessary. The measurement of Lumipulse G Prolactin Calibrators (10 replicates) and secondary calibrators (10 replicates) are performed on 3 runs using the LUMIPULSE G System. The mean ratio is the mean counts of the secondary calibrator to the respective Lumipulse G Prolactin Calibrator. The acceptable rate mean ratio is 0.95 - 1.05. This solution is stored at 2 to 10 °C.

CalibratorTarget Values (ng/mL)Target Range (ng/mL)
Cal 10 ng/mLN/A
Cal 2400 ng/mL380 – 420 ng/mL

Stability

Stability of the calibrators is determined by long-term stability study (Tokyo, Japan) using the LUMIPULSE Forte (f) instrument (only sold in Japan). Intended storage and transport stability study is still on-going (Malvern, PA). Based on accelerated stability study the sponsor determined that the calibrators have a shelf-life stability of 12 months

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a blue and green design with a white teardrop shape in the center. The company name is written in blue, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.

when stored at 2-10ºC. The sponsor's protocol and acceptance criteria was reviewed and found to be acceptable.

*Note: (The LUMIPULSE f System uses the same Lumipulse G Prolactin Immunoreaction Cartridges and Calibrators as the LUMIPULSE G1200 System)

1. Studies conducted at Fujirebio, Inc. (Tokyo, Japan)

  • 1.1 Long-term stability
    The long-term stability data were obtained on the LUMIPULSE G System by measuring the Prolactin calibrators and samples for 3 lots of reagents that were stored at 10°C for 0, 3, 7, and 13 months. Sensitivity, specificity and reproducibility were calculated.

The results of sensitivity, accuracy and reproducibility confirmed that in a 13-month test. standards were met and the product passed. Therefore, the shelf life of the Lumipulse G Prolactin Calibrators stored at 2-10°C for 12 months was confirmed.

1.2 Transport Simulation

Transportation Testing (Temperature Stress Testing) data were obtained by stressing reagents. calibrator (standard solution), substrate solution, and diluent under the following conditions: 37°C, 48 hours ± 2 hours → -20°C, 48 hours ± 2 hours → 30°C, 96 hours ± 4 hours-> storage conditions (4°C), 24 hours ± 2 hours. Sensitivity, accuracy, and visual inspection were performed.

Visual assessment of 5 items was 0/28 under control conditions and 0/28 under stress conditions, thus meeting the specification. Sensitivity was 162 under control conditions and 171 under stress conditions, in each case meeting the S/N≥51 specification. Accuracy (versus test values) was 95-101% under control conditions and 99-109% under stress conditions, in each case meeting the specification for "within 100 ± 20%." Therefore, the results of sensitivity, accuracy, and visual inspection all met specification.

2. Studies conducted at Fujirebio Diagnostics, Inc. (Malvern, PA)

2.1 Real Time (Intended Storage) Stability

Real-time (Intended Storage) stability is being evaluated as follows:

In the on-going real-time stability study, the Lumipulse G Prolactin calibrators are stored at 2-10 C. The Prolactin calibrators are tested in duplicate at specified intervals over the shelf life of the device up to the planned shelf life plus one month (13 months).

Real-time intended storage stability is calculated based on the percent difference for the panels at each time point when compared to the value received at Time Point 0 (study initiation).

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in blue text. The graphic is a combination of blue and green shapes, resembling a stylized water droplet or leaf. The text "FUJIREBIO" is in a bold, sans-serif font, with "Diagnostics, Inc." in a smaller font size below it.

Time Point 2 (Month 6) was the second time point for testing. The Lumipulse G Prolactin calibrators were determined to be stable at the intended storage conditions for up to 6 months for all 3 lots.

The testing will continue with this stability protocol until data to support a claim of 12 months is achieved.

2.2 Transport Simulation Stability Study

Transport simulation of the Lumipulse G Prolactin calibrators was examined in terms of temperature. For the temperature test, the calibrators were stored in the following condition: 37 C ± 2 C 24±2 hours → 25 C ± 2 C 24±2 hours → -20 C ± 10 C 24±2 hours → 6 C ± 4 C 24±2 hours. To determine the impact of temperature on the Lumipulse G prolactin Calibrators, a performance analysis was conducted.

After each stress condition, all Lumipulse G Prolactin ICs and Lumipulse G Prolactin Calibrators were moved to the next cycle temperature. After all cycles were complete, a visual inspection was performed and the Lumipulse G Prolactin calibrators were stored at the intended storage condition of 2-10°C. Stressed Lumipulse G Prolactin IC's and Calibrators were tested within 3 calendar days after the completion of all cycles. Calibrators were tested in duplicate, controls in singlicate and panels in triplicate.

At study initiation, the assay met the acceptance criteria of %CV≤ 10% for all transport conditions tested. The stability study is ongoing.

Shelf life

The shelf life for Lumipulse G Prolactin Calibrators is 12 months at 2-10°C, based on the longterm stability study conducted at Fujirebio Inc. and the Real Time Stability study conducted at Fujirebio Diagnostics, Inc..

Transport Conditions

Lumipulse G Prolactin Calibrators are shipped at 2-10°C.

  • 2.3 Open-vial (In-Use) Stability
    Open-vial (In-use) stability is being evaluated as follows:

In the on-going open-vial stability study (included in the real-time stability study), the Lumipulse G Prolactin calibrators are stored at 2-10 C and both calibrators were opened and then closed at the start of the study (Time Point). The Prolactin calibrators are tested in duplicate at specified intervals (6 months and 12 months) over the shelf life of the device up to the planned shelf life plus one month (13 months).

Open-vial (in-use) stability is calculated based on the percent difference for the panels at each time point when compared to the value received at Time Point 0 (study initiation).

The mean concentration of each panel from each time point must be within ±10% mean concentration at Study Initiation. The % CV value of each time point must be ≤ 10%.

{8}------------------------------------------------

Premarket Notification Lumipulse® G Prolactin

Image /page/8/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes that resemble a water droplet. To the right of the graphic is the company name, "FUJIREBIO," in bold, blue letters. Below the company name is the text "Diagnostics, Inc." in a smaller, black font.

The testing will continue with this stability protocol until data to support a claim of 12 months is achieved.

  • d. Detection limit: Not applicable.
  • e. Analytical specificity: Not applicable.
  • f. Assay cut-off: Not applicable.
    1. Comparison studies:
    • a. Method Comparison: Not applicable.
    • b. Method Comparison: Not applicable.
    • c. Matrix Comparison: Not applicable.

3. Clinical studies:

  • a. Clinical sensitivity: Not applicable.
  • b. Clinical specificity: Not applicable.
  • c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
  • Clinical cut-off: 4. Not applicable.
    1. Expected values/Reference range: Not applicable.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and green shapes, and the company name on the right. The company name is written in a bold, blue font, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.

Proposed Labelling

The labelling satisfies the requirements of 21 CFR Part 809.10.

Standard/Guidance Document Referenced:

  • ISO 17511:2003 Measurement of Quantities in Biological Samples Metrological ● Traceability of Values Assigned to Calibrator and Control Materials
  • CLSI EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents: Approved Guideline ●
  • Guidance for Industry and Food and Drug Administration Staff eCopy Program for Medical ● Device Submissions (December 3, 2015)
  • Guidance for Industry and Food and Drug Administration Staff - Refuse to Accept Policy for 510(k)s (August 4, 2015)
  • Guidance Document Guidance for Industry Abbreviated 510(k) Submissions for In Vitro ● Diagnostic Calibrators; Final (February 22, 1999)

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.