K Number
K163521
Device Name
Lumipulse G LH Calibrators
Date Cleared
2017-01-12

(28 days)

Product Code
Regulation Number
862.1150
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Lumipulse® GLH Calibrators are for in vitro diagnostic use in the calibration of Lumipulse GLH on the LUMIPULSE G System.
Device Description
The Lumipulse® G LH Calibrator kit contains 2 bottles (1.5mL each) of Lumipulse® G LH Calibrators 1 and 2. Calibrators 1 and 2 contain luteinizing hormone (LH) in 0.15M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide. Lumipulse® G LH Calibrators CAL 230916, Liquid 1x2 concentrations Each calibrator kit contains one bottle each of Calibrators 1 – 2. The calibrator kit is packaged separately. CAL 1 0 mlU/mL LH calibrator (1 x 1.5 mL) CAL 2 250 mIU/mL LH calibrator (1 x 1.5 mL) *Contains luteinizing hormone (LH) in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide The Lumipulse® G LH calibrator set is designed specifically for use on the LUMIPULSE® G System which has been previously cleared under k142895.
More Information

No
The device description and performance studies focus on the chemical composition, stability, and traceability of calibrators for an in vitro diagnostic system. There is no mention of AI or ML in the text.

No.
The "Intended Use" states that the device is "for in vitro diagnostic use in the calibration of Lumipulse GLH on the LUMIPULSE G System," which means it is used for testing within a lab setting, not directly for treating a patient.

No
Explanation: This device is a calibrator, which is used to calibrate diagnostic devices, but it is not a diagnostic device itself. The "Intended Use" states it is "for in vitro diagnostic use in the calibration of Lumipulse GLH on the LUMIPULSE G System".

No

The device description clearly states that the device is a kit containing liquid calibrators (bottles of solution) which are physical components, not software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The "Intended Use / Indications for Use" section explicitly states: "Lumipulse® GLH Calibrators are for in vitro diagnostic use in the calibration of Lumipulse GLH on the LUMIPULSE G System." This is the primary indicator of an IVD.
  • Device Description: The description details the composition of the calibrators, which are reagents used in a laboratory setting to calibrate an analytical instrument (the LUMIPULSE G System) for measuring a specific analyte (luteinizing hormone - LH) in biological samples. This aligns with the nature of IVD devices.
  • Intended User / Care Setting: The mention of "in vitro diagnostic use" further reinforces its classification as an IVD, as these devices are used outside of a living organism to diagnose or aid in the diagnosis of a condition.
  • Performance Studies: While the performance studies focus on the stability and traceability of the calibrators themselves, this is a crucial aspect of ensuring the accuracy and reliability of the overall diagnostic test performed using the LUMIPULSE G System. The studies are designed to demonstrate that the calibrators are suitable for their intended diagnostic purpose.
  • Predicate Device: The mention of a predicate device (K111023; Abbott ARCHITECT® LH Calibrators) which is also a calibrator for an LH assay, further supports the classification as an IVD. Predicate devices are typically other legally marketed IVDs.

In summary, the explicit statement of "in vitro diagnostic use" in the intended use, the nature of the device as a calibrator for a diagnostic system, and the context of its use in a laboratory setting all confirm that this device is an IVD.

N/A

Intended Use / Indications for Use

Lumipulse® GLH Calibrators are for in vitro diagnostic use in the calibration of Lumipulse GLH on the LUMIPULSE G System.

Product codes

JIT

Device Description

The Lumipulse® G LH Calibrator kit contains 2 bottles (1.5mL each) of Lumipulse® G LH Calibrators 1 and 2. Calibrators 1 and 2 contain luteinizing hormone (LH) in 0.15M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide.

Lumipulse® G LH Calibrators CAL 230916, Liquid 1x2 concentrations

Each calibrator kit contains one bottle each of Calibrators 1 – 2. The calibrator kit is packaged separately.

CAL 1 0 mlU/mL LH calibrator (1 x 1.5 mL) CAL 2 250 mIU/mL LH calibrator (1 x 1.5 mL) *Contains luteinizing hormone (LH) in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

  1. Analytical performance:
    a. Precision/Reproducibility: Not applicable.
    b. Linearity/assav reportable range: Not applicable.
    c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability:
The calibrators for use with Lumipulse G LH are prepared gravimetrically and are traceable to the 200 International Standard, 1988 (code: 80/552) provided by the National Institute for Biological Standards and Control (NIBSC).

Value Assignment:
LH Antigen Grade is obtained from CLINIQA. A working stock solution is prepared by dissolving LH antigen in 2-amino-2-hydroxymethyl-1,3-propandiol-based buffer. The working stock solution is assigned a provisional value based on the measurement (6 replicates) using Lumipulse G LH on the LUMIPULSE G system with secondary calibrators. Lumipulse G LH Calibrators are made by calculating the amount of working stock solution and 2-amino-2-hydroxymethyl-1,3-propandiol-based buffer required to achieve the desired 250 mlU/mL calibrator value (Lumipulse G LH Calibrator 2) and adding them together gravimetrically. The 0 mlU/mL calibrator (Lumipulse G LH Calibrator 1) only contains the 2-amino-2-hydroxymethyl-1,3propandiol-based buffer. Lumipulse G LH Calibrator 2 is rate-matched to the respective secondary calibrator based on the measurement in Lumipulse G LH and adjusted if necessary. The measurement of Lumipulse G LH calibrators (10 replicates) and secondary calibrators (10 replicates) are performed on three runs on the LUMIPULSE G system. The mean ratio is the mean counts of the secondary calibrator to the respective Lumipulse G LH calibrator. The acceptable rate mean ratio for the 3 runs is 0.95 – 1.05. This solution is stored at 2 to 10°C.

Stability:
Stability of the calibrators is determined by long-term stability study (Tokyo. Japan) using the LUMIPULSE Forte (f) instrument (only sold in Japan). Intended storage and transport stability study is still on-going (Malvern, PA). Based on real-time stability study the sponsor determined that the calibrators have a shelf-life stability of 12 months when stored at 2-10ºC. The sponsor's protocol and acceptance criteria was reviewed and found to be acceptable.
Note: (The LUMIPULSE f System uses the same Lumipulse G LH Immunoreaction Cartridges and Calibrators as the LUMIPULSE G1200 System).

Studies conducted at Fujirebio, Inc. (Tokyo, Japan):
Shelf-Life Stability:
Three lots of Lumipulse G LH Calibrators were tested on real-time stability with Lumipulse forte (f), a predecessor instrument of LUMIPULSE G1200. LUMIPULSE f and LUMIPULSE G1200 have a confirmed agreement rate.
Three lots of Lumipulse G LH Calibrators were stored at 10℃ for Month 0, 3, 7 and 13 and tested for sensitivity, accuracy and within-run reproducibility.
The results met the criteria for up to 13 months in the 3 lots of Lumipulse G LH Calibrators. Thus it was demonstrated that the LUMIPULSE G LH Calibrators were stable up to 13 months at 10℃ on the LUMIPULSE f System.
Regarding the sensitivity, 0 and 20 mIU/mL LH solutions were tested at each test point in replicates of 3 and the ratio of the amount of the luminescence ("20 mIU/mL" /"0 mIU/mL" LH) were calculated. The criterion was 20 or higher in the ratio.
Regarding the accuracy. 3 samples that ranged 5,850-162,217 mlU/mL were tested . in replicates of 3 at each time point and the ratios of the measured values to the preassigned values were confirmed. The criterion was that the ratios should be within 100 +/- 20% for each replicate.
Regarding within-run reproducibility, 4 samples that ranged 5.850-162.217 mlU/mL were tested in replicates of 6 at each time point. The criterion for the precision was 10% or lower in the CV for each sample.
The shelf life of Lumipulse G LH Calibrators is 12 months at 2 - 10℃.

Transport Conditions:
The transport simulation stability of Lumipulse G LH packed for overseas transport was examined in terms of temperature change.
Temperature change: The Lumipulse G LH Immunoreaction Cartridges, Lumipulse G LH Calibrators, LUMIPULSE G Substrate, LUMIPULSE G Wash Solution and LUMIPULSE G Specimen Diluent 1 were subjected to the following temperature stress: 37 °C 48 +/- 2 hours -> -20 °C 48 +/- 2 hours -> 30 °C 96 +/- 4 hours -> 4 °C 24 +/- 2 hours, according to the stress condition in the internal protocol. Then a visual inspection was carried out for the Lumipulse G LH Cartridges and it was confirmed that they cleared the criteria. Also a sensitivity test and accuracy test were conducted and compared with the intended condition (control). The sensitivity testing was conducted with 0 and 2 mlU/mL LH solutions and the criterion was the ratio of the amount of luminescence ("2 mlU/mL" / "0 mIU/mL" LH) to be 21 or higher. The accuracy testing was conducted with 3 control samples (concentration: 4.1-55.6 mlU/mL with LUMIPULSE G1200) and 4 serum samples (concentration: 11.1-294.0 mlU/mL with LUMIPULSE G1200) in replicates of 1 for control samples and 2 for the serum samples with LUMIPULSE G1200, and the criteria of each value after the stress were 100 +/- 20% of the reference value of each serum sample.
The results of the sensitivity test and accuracy met the criteria. Therefore, the temperature change described above did not affect the performance of Lumipulse G LH using the LUMIPULSE G1200 System.
Lumipulse G LH Calibrators are shipped at 2-10°C.

Studies conducted at Fujirebio Diagnostics, Inc. (Malvern, PA):
Real Time (Intended Storage and Open Use):
In the on-going real-time stability study, the Lumipulse G LH calibrators are stored at 2-10°C. The LH calibrators are tested on one Lumipulse G1200 instrument in duplicate at specified intervals over the shelf life of the device up to the planned shelf life plus one month (13 months).
Real-time intended storage stability is calculated based on the percent difference for the panels at each time point when compared to the value received at Time Point 0 (study initiation). The mean concentration of each panel from each time point must be within ±10% mean concentration at Study Initiation. The % CV value of each panel at each time point must be ≤ 10%. For all 3 lots, Time Point 2 (6 months) was the second time point for testing. The Lumipulse G LH Calibrators were determined to be stable at the intended storage conditions for up to 6 months for all 3 lots.
The testing will continue with this stability protocol until data to support a claim of 12 months is achieved.

Transport Simulation Stability Study:
Transport simulation of the Lumipulse G LH calibrators was examined in terms of temperature. For the temperature test, the calibrators were stored in the following condition: 37 °C 24±2 hours → 25 °C 24±2 hours → -20 °C 24±2 hours → 6 °C 24±2 hours. To determine the impact of temperature on the Lumipulse G LH Calibrators, a performance analysis was conducted.
After each stress condition, all Lumipulse G LH ICs and Lumipulse G LH Calibrators were moved to the next cycle temperature. After all cycles were complete, a visual inspection was performed and the Lumipulse G LH calibrators were stored at the intended storage condition of 2-10°C. Stressed Lumipulse G LH IC's and Calibrators were tested within 3 calendar days after the completion of all cycles. Calibrators were tested in duplicate, controls in replicates of 1 and panels in replicates of 3.
At study initiation, the assay met the acceptance criteria of %CV ≤10% for all transport conditions tested. This stability study is still ongoing.
Transport Conditions Lumipulse G LH Calibrators are shipped at 2-10°C.

d. Detection limit: Not applicable.
e. Analytical specificity: Not applicable
f. Assay cut-off: Not applicable.

  1. Comparison studies:
    a. Method Comparison with predicate device: Not applicable.
    b. Method Comparison Not applicable.
    c. Matrix Comparison Not applicable.

  2. Clinical studies:
    a. Clinical sensitivity: Not applicable
    b. Clinical specificity: Not applicable
    c. Other clinical supportive data (when a. and b. are not applicable): Not applicable

  3. Clinical cut-off: Not applicable.

  4. Expected values/Reference range: Not applicable.

Key Metrics

Not Found

Predicate Device(s)

K111023

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FUJIREBIO DIAGNOSTICS, INC DIANA DICKSON DIRECTOR, CLINICAL AND REGULATORY SCIENCE 201 GREAT VALLEY PARKWAY MALVERN PA 19355

January 12, 2017

Re: K163521

Trade/Device Name: Lumipulse® G LH Calibrators Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT Dated: December 14, 2016 Received: December 15, 2016

Dear Diana Dickson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

1

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For:

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K163521

Device Name

Lumipulse® GLH Calibrators

Indications for Use (Describe)

Lumipulse GLH Calibrators are for in vitro diagnostic use in the calibration of Lumipulse GLH on the LUMIPULSE G System.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes that resemble a water droplet. To the right of the graphic is the company name, "FUJIREBIO" in large, bold, blue letters. Below the company name is "Diagnostics, Inc." in smaller, black letters.

Section 5 510(k) SUMMARY

Lumipulse® G LH Calibrators

Date:January 6, 2016
------------------------
  • 510(k) Number: K163521
  • Submitter: Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355
  • Contact Person: Stacey Dolan Manager, Requlatory Affairs Phone: (610) 240-3843 Fax: (610) 240-3803 Email: dolans@fdi.com
Device Name:Lumipulse® G LH Calibrators
Common Name:Human Luteinizing Hormone
Classification:Class II
JIT
75, Clinical Chemistry
21 CFR § 862.1150
Predicate Device:K111023
Abbott Laboratories

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Abbott ARCHITECT® LH Calibrators

Device Description:

The Lumipulse® G LH Calibrator kit contains 2 bottles (1.5mL each) of Lumipulse® G LH Calibrators 1 and 2. Calibrators 1 and 2 contain luteinizing hormone (LH) in 0.15M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide.

Lumipulse® G LH Calibrators CAL 230916, Liquid 1x2 concentrations

Each calibrator kit contains one bottle each of Calibrators 1 – 2. The calibrator kit is packaged separately.

CAL 1 0 mlU/mL LH calibrator (1 x 1.5 mL) CAL 2 250 mIU/mL LH calibrator (1 x 1.5 mL) *Contains luteinizing hormone (LH) in 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). Preservative: sodium azide

4

Image /page/4/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic element on the left, followed by the company name in blue text. The graphic element is a combination of blue and green shapes, with a white teardrop shape in the center. The text "FUJIREBIO" is in a bold, sans-serif font, and "Diagnostics, Inc." is in a smaller font below it.

The Lumipulse® G LH calibrator set is designed specifically for use on the LUMIPULSE® G System which has been previously cleared under K142895.

Device Name:

Lumipulse G LH Calibrators

Device Intended Use:

Lumipulse G LH Calibrators are for in vitro diagnostic use in the calibration of Lumipulse G LH on the LUMIPULSE G System.

Device Indications for Use:

Same as Intended Use

Substantial Equivalence:

Comparison between the Lumipulse G LH Calibrators and Abbott ARCHITECT LH Calibrators

Similarities and Differences
ParameterLumipulse G
LH Calibrators
(Proposed Device)Abbott ARCHITECT LH
Calibrators (Predicate
Device)
K111023
Intended UseLumipulse G LH Calibrators are
for in vitro diagnostic use in the
calibration of Lumipulse G LH on
the LUMIPULSE G System.For the calibration of the
ARCHITECT i System
when used for the
quantitative determination
of human luteinizing
hormone (LH) in human
serum and plasma.
AnalyteLuteinizing hormonesame
Composition• Calibrators 1-2: Luteinizing
hormone (from human pituitary)
• Calibrators 1-2: Diluent: Tris
buffer with protein (bovine) and
chemical stabilizers
• Calibrators 1-2: Preservative:
Sodium azide• Calibrators B-F:
Luteinizing hormone
(from human pituitary)
• Calibrator A-F: Diluent:
Phosphate buffer with
protein stabilizers
(bovine)
• Calibrators A-F:
Preservatives: ProClin
300, ProClin 950
LH Calibrators• 2 levels
• 0 and 250 mlU/mL human
luteinizing hormone• 6 levels
• 0.00, 1.00, 3.50, 15.00,
50.00, 250.00 mlU/mL
human luteinizing
hormone
Standardization/TraceabilityCalibration of the LUMIPULSE G
LH is traceable to in-house
reference calibrators, whoseThe calibrators are
referenced to the World
Health Organization

5

Image /page/5/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in blue text. The graphic includes a blue and white abstract shape, with vertical green lines next to it. The text "FUJIREBIO" is in a bold, sans-serif font, with "Diagnostics, Inc." in a smaller font below it.

Similarities and Differences
ParameterLumipulse G
LH Calibrators
(Proposed Device)Abbott ARCHITECT LH
Calibrators (Predicate
Device)
K111023
values have been assigned to the
2nd International Standard, 1988
(code: 80/552) provided by the
National Institute for Biological
Standards and Control (NIBSC).(WHO) Luteinizing
Hormone (LH) Human,
Pituitary 2nd International
Standard 80/552.

Performance Characteristics

Data were generated using the LUMIPULSE G1200 System.

    1. Analytical performance:
    • a. Precision/Reproducibility: Not applicable.
    • b. Linearity/assav reportable range: Not applicable.
    • c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The calibrators for use with Lumipulse G LH are prepared gravimetrically and are traceable to the 200 International Standard, 1988 (code: 80/552) provided by the National Institute for Biological Standards and Control (NIBSC).

Value Assignment

LH Antigen Grade is obtained from CLINIQA. A working stock solution is prepared by dissolving LH antigen in 2-amino-2-hydroxymethyl-1,3-propandiol-based buffer. The working stock solution is assigned a provisional value based on the measurement (6 replicates) using Lumipulse G LH on the LUMIPULSE G system with secondary calibrators. Lumipulse G LH Calibrators are made by calculating the amount of working stock solution and 2-amino-2-hydroxymethyl-1,3-propandiol-based buffer required to achieve the desired 250 mlU/mL calibrator value (Lumipulse G LH Calibrator 2) and adding them together gravimetrically. The 0 mlU/mL calibrator (Lumipulse G LH Calibrator 1) only contains the 2-amino-2-hydroxymethyl-1,3propandiol-based buffer. Lumipulse G LH Calibrator 2 is rate-matched to the respective secondary calibrator based on the measurement in Lumipulse G LH and adjusted if necessary. The measurement of Lumipulse G LH calibrators (10 replicates) and secondary calibrators (10 replicates) are performed on three runs on the LUMIPULSE G system. The mean ratio is the mean counts of the secondary calibrator to the respective Lumipulse G LH calibrator. The acceptable rate mean ratio for the 3 runs is 0.95 – 1.05. This solution is stored at 2 to 10°C.

6

Image /page/6/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and green shapes that resemble a water droplet or a stylized flower. To the right of the graphic, the words "FUJIREBIO" are written in a bold, blue sans-serif font. Below "FUJIREBIO", the words "Diagnostics, Inc." are written in a smaller, black sans-serif font.

CalibratorTarget Values (mIU/mL)Target Ranges (mIU/mL)
Level 10N/A
Level 2250238 - 262

Stability

Stability of the calibrators is determined by long-term stability study (Tokyo. Japan) using the LUMIPULSE Forte (f) instrument (only sold in Japan). Intended storage and transport stability study is still on-going (Malvern, PA). Based on real-time stability study the sponsor determined that the calibrators have a shelf-life stability of 12 months when stored at 2-10ºC. The sponsor's protocol and acceptance criteria was reviewed and found to be acceptable.

*Note: (The LUMIPULSE f System uses the same Lumipulse G LH Immunoreaction Cartridges and Calibrators as the LUMIPULSE G1200 System).

See Section 14 - Sterilization and Shelf Life for more stability study related data and information.

1. Studies conducted at Fujirebio, Inc. (Tokyo, Japan)

Shelf-Life Stability

Three lots of Lumipulse G LH Calibrators were tested on real-time stability with Lumipulse forte (f), a predecessor instrument of LUMIPULSE G1200. LUMIPULSE f and LUMIPULSE G1200 have a confirmed agreement rate.

Three lots of Lumipulse G LH Calibrators were stored at 10℃ for Month 0, 3, 7 and 13 and tested for sensitivity, accuracy and within-run reproducibility.

The results met the criteria for up to 13 months in the 3 lots of Lumipulse G LH Calibrators. Thus it was demonstrated that the LUMIPULSE G LH Calibrators were stable up to 13 months at 10℃ on the LUMIPULSE f System.

  • . Regarding the sensitivity, 0 and 20 mIU/mL LH solutions were tested at each test point in replicates of 3 and the ratio of the amount of the luminescence ("20 mIU/mL" /"0 mIU/mL" LH) were calculated. The criterion was 20 or higher in the ratio.
  • Regarding the accuracy. 3 samples that ranged 5,850-162,217 mlU/mL were tested . in replicates of 3 at each time point and the ratios of the measured values to the preassigned values were confirmed. The criterion was that the ratios should be within 100 +/- 20% for each replicate.
  • . Regarding within-run reproducibility, 4 samples that ranged 5.850-162.217 mlU/mL were tested in replicates of 6 at each time point. The criterion for the precision was 10% or lower in the CV for each sample.

Shelf-Life

The shelf life of Lumipulse G LH Calibrators is 12 months at 2 - 10℃.

7

Image /page/7/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in blue text. The graphic appears to be a combination of shapes and lines, possibly representing water droplets or a stylized flower. The text "FUJIREBIO" is in a larger, bolder font, while "Diagnostics, Inc." is in a smaller font below.

Transport Conditions

The transport simulation stability of Lumipulse G LH packed for overseas transport was examined in terms of temperature change.

Temperature change: The Lumipulse G LH Immunoreaction Cartridges, Lumipulse G LH Calibrators, LUMIPULSE G Substrate, LUMIPULSE G Wash Solution and LUMIPULSE G Specimen Diluent 1 were subjected to the following temperature stress: 37 °C 48 +/- 2 hours -> -20 °C 48 +/- 2 hours -> 30 °C 96 +/- 4 hours -> 4 °C 24 +/- 2 hours, according to the stress condition in the internal protocol. Then a visual inspection was carried out for the Lumipulse G LH Cartridges and it was confirmed that they cleared the criteria. Also a sensitivity test and accuracy test were conducted and compared with the intended condition (control). The sensitivity testing was conducted with 0 and 2 mlU/mL LH solutions and the criterion was the ratio of the amount of luminescence ("2 mlU/mL" / "0 mIU/mL" LH) to be 21 or higher. The accuracy testing was conducted with 3 control samples (concentration: 4.1-55.6 mlU/mL with LUMIPULSE G1200) and 4 serum samples (concentration: 11.1-294.0 mlU/mL with LUMIPULSE G1200) in replicates of 1 for control samples and 2 for the serum samples with LUMIPULSE G1200, and the criteria of each value after the stress were 100 +/- 20% of the reference value of each serum sample.

The results of the sensitivity test and accuracy met the criteria. Therefore, the temperature change described above did not affect the performance of Lumipulse G LH using the LUMIPULSE G1200 System.

Lumipulse G LH Calibrators are shipped at 2-10°C.

    1. Studies conducted at Fujirebio Diagnostics, Inc. (Malvern, PA)

Real Time (Intended Storage and Open Use)

Real-time (Intended Storage and open use) stability is being evaluated as follows: In the on-going real-time stability study, the Lumipulse G LH calibrators are stored at 2-10°C. The LH calibrators are tested on one Lumipulse G1200 instrument in duplicate at specified intervals over the shelf life of the device up to the planned shelf life plus one month (13 months).

Real-time intended storage stability is calculated based on the percent difference for the panels at each time point when compared to the value received at Time Point 0 (study initiation). The mean concentration of each panel from each time point must be within ±10% mean concentration at Study Initiation. The % CV value of each panel at each time point must be ≤ 10%. For all 3 lots, Time Point 2 (6 months) was the second time point for testing. The Lumipulse G LH Calibrators were determined to be stable at the intended storage conditions for up to 6 months for all 3 lots.

The testing will continue with this stability protocol until data to support a claim of 12 months is achieved.

8

Image /page/8/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized symbol on the left and the company name on the right. The symbol is a combination of blue and green shapes, with a white teardrop shape in the center. The company name is written in blue, with "FUJIREBIO" in a larger font size than "Diagnostics, Inc."

Transport Simulation Stability Study

Transport simulation of the Lumipulse G LH calibrators was examined in terms of temperature. For the temperature test, the calibrators were stored in the following condition: 37 °C 24±2 hours → 25 °C 24±2 hours → -20 °C 24±2 hours → 6 °C 24±2 hours. To determine the impact of temperature on the Lumipulse G LH Calibrators, a performance analysis was conducted.

After each stress condition, all Lumipulse G LH ICs and Lumipulse G LH Calibrators were moved to the next cycle temperature. After all cycles were complete, a visual inspection was performed and the Lumipulse G LH calibrators were stored at the intended storage condition of 2-10°C. Stressed Lumipulse G LH IC's and Calibrators were tested within 3 calendar days after the completion of all cycles. Calibrators were tested in duplicate, controls in replicates of 1 and panels in replicates of 3.

At study initiation, the assay met the acceptance criteria of %CV ≤10% for all transport conditions tested. This stability study is still ongoing.

Transport Conditions Lumipulse G LH Calibrators are shipped at 2-10°C.

  • d. Detection limit: Not applicable.
  • e. Analytical specificity: Not applicable
  • f. Assay cut-off: Not applicable.
    1. Comparison studies:
    • a. Method Comparison with predicate device: Not applicable.
    • b. Method Comparison Not applicable.
    • c. Matrix Comparison Not applicable.
    1. Clinical studies:
    • a. Clinical sensitivity: Not applicable
    • b. Clinical specificity: Not applicable

9

Image /page/9/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a combination of blue and green shapes, with a white teardrop shape in the center. The text "FUJIREBIO" is in blue, and "Diagnostics, Inc." is in a smaller font size below the company name.

  • Other clinical supportive data (when a. and b. are not applicable): C. Not applicable
    1. Clinical cut-off: Not applicable.
    1. Expected values/Reference range: Not applicable.

Proposed Labeling

The labeling satisfies the requirements of 21 CFR Part 809.10.

Standard/Guidance Document Referenced:

  • . ISO 17511:2003 Measurement of Quantities in Biological Samples - Metrological Traceability of Values Assigned to Calibrator and Control Materials
  • CLSI EP25-A - Evaluation of Stability of In Vitro Diagnostic Reagents: Approved Guideline
  • . Guidance for Industry and Food and Drug Administration Staff - eCopy Program for Medical Device Submissions (December 3, 2015)
  • . Guidance for Industry and Food and Drug Administration Staff - Refuse to Accept Policy for 510(k)s (August 4, 2015)
  • Guidance Document - Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.