K Number
K163455
Device Name
SurgiCube
Date Cleared
2017-09-28

(293 days)

Product Code
Regulation Number
878.5070
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SurgiCube® is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m³ at the surgical site and instruments.

The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

Device Description

The SurgiCube® is an apparatus that provides a localized, optimally HEPA filtered, surgical environment to carry out minimally invasive surgery for ophthalmic procedures. It supplies air around the operating surface using uni directional air flow technique. Possible sources of contamination are sidelined. The patient is physically positioned outside the surgical area; just the operating area is in the field. The surgical team can move around the operating surface without interfering with the air. The uni directional flow together with correct surgical draping create a compact and manageable surgical area that will eliminate airborne contamination risks for patients. The SurgiCube® is equipped with a multiple mechanical filter system, including a High Efficiency Particle Air (HEPA) filter with an efficiency of filtering 99,995% of all particles of 0.3um.

AI/ML Overview

I am sorry, but the provided text describes a medical device, the SurgiCube, for creating a sterile surgical environment, and its regulatory clearance (510(k) submission) by the FDA. The information provided heavily pertains to engineering and performance specifications, and clinical validation for the device's ability to maintain a clean environment.

The request, however, asks for information typically associated with the evaluation of AI/ML-enabled medical devices, specifically:

  • Acceptance criteria and device performance in a table format, often related to metrics like sensitivity, specificity, AUC, etc.
  • Sample sizes for test sets and data provenance for an AI/ML model.
  • Number and qualifications of experts for ground truth establishment.
  • Adjudication methods.
  • Multi-reader multi-case (MRMC) comparative effectiveness studies with AI assistance.
  • Standalone algorithm performance.
  • Type of ground truth used (e.g., pathology, outcomes data).
  • Sample size for the training set and how its ground truth was established.

The provided document does NOT contain any of this information as the SurgiCube is not an AI/ML device. It is an air-handling apparatus designed to reduce microorganisms and particulate matter.

Therefore, I cannot fulfill your request based on the provided text.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2017

Surgicube International B.v. Ger Vijfvinkel CEO Seggelant-noord 4 Vierpolders, 3237 MG NL

Re: K163455

Trade/Device Name: Surgicube Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus For A Surgical Operating Room Regulatory Class: Class II Product Code: ORC Dated: August 24, 2017 Received: August 28, 2017

Dear Ger Vijfvinkel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, PhD Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163455

Device Name SurgiCube®

Indications for Use (Describe)

SurgiCube® is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-fittered to reduce the presence of particulate matter to reduce the microorganisms to a level of < S CFU per m3 at the surgical site and instruments.

The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for SurgiCube International. The logo features a blue square with a white curved line inside, next to the text "SurgiCube" in a blue sans-serif font. Below "SurgiCube" is the word "International" in a smaller, also blue, sans-serif font. The logo is clean and modern, with a focus on the company name.

510(k) Summary

Date the summary was prepared is September 25, 2017

The submitter of the 510(k) is:

  • Ger Vijfvinkel CEO SurgiCube International B.V. Seggelant-Noord 4 3237 MG Vierpolders The Netherlands Phone +31 181 760 660 Email g.vijfvinkel@surgicube.com

Device subject to this 510(k)

Trade name:SurgiCube®
Common name:Air handling apparatus for surgery
Classification nameAir-handling apparatus for a surgical operating room
Device classII
Regulation number21 CFR 878.5070
Product codeORC

Predicate device

The legally marketed device to which substantial equivalence

is claimed is: Operio K153498 Cleared 07/20/2016

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for "SurgiCube International". The logo consists of a blue square with a white square inside, and a white curved line inside the inner square. To the right of the square is the text "SurgiCube" in a blue sans-serif font, with the word "International" underneath in a smaller font.

Device description

The SurgiCube® is an apparatus that provides a localized, optimally HEPA filtered, surgical environment to carry out minimally invasive surgery for ophthalmic procedures. It supplies air around the operating surface using uni directional air flow technique. Possible sources of contamination are sidelined. The patient is physically positioned outside the surgical area; just the operating area is in the field. The surgical team can move around the operating surface without interfering with the air. The uni directional flow together with correct surgical draping create a compact and manageable surgical area that will eliminate airborne contamination risks for patients. The SurgiCube® is equipped with a multiple mechanical filter system, including a High Efficiency Particle Air (HEPA) filter with an efficiency of filtering 99,995% of all particles of 0.3um.

Indications for use

SurgiCube®is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m³ at the surgical site and instruments.

The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

Summary of the technological characteristics compared to the predicate device

SurgiCube® and the predicate device, Operio, have the same intended use and similar technological characteristics.

lt is similar in the following way:

  • The positioning to the surgical site of the delivery of the air flow is the same.
  • . The surgical indication, ophthalmic surgery is the same.
  • The cleaning efficiency of the HEPA filtered clean air over the surgical site is under 5 CFU for both the subject device (0.12 CFU/ m3) and the predicate device (0.4 CFU/ m3).
  • The air flow from the subject device is HEPA-filtered and has the same efficiency (99.995%) to reduce the presence of particulate matter and microorganisms at the surqical site.

lt differs in the following way:

  • The subject device is not a portable device; it is a stand-alone device. ●
  • . The subject device doesn't use a sterile shield in front of the air outlet, instead it uses a single use sterile drape attached onto the side columns.
  • . The reach of the clean air flow is up to 32" cm from the HEPA filter compared to 47" for the predicate device.
  • . The delivered air volume to the surgical site for the subject device is 353-777 CFM compared to 235 CFM for the predicate device.

The above stated differences do not introduce any new hazards and test results can confirm that the subject device is as safe and effective for use. In the hazard analysis possible risks have been mitigated that could be associated with the use of the subject device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for SurgiCube International. The logo consists of a blue square with a white outline on the left side, with a curved white line inside the square. To the right of the square is the text "SurgiCube" in a blue sans-serif font, with the word "International" in a smaller font below it.

Comparison table:

CharacteristicSubject devicePredicate device
Review PanelSurgiCube®General HospitalOperio K153498General Hospital
Indications for useSurgiCube® is a stand-alone devicethat creates a surgical operatingenvironment with a directed, non-turbulent flow of air to the surgical siteduring ophthalmic surgery and to thesterile instruments used duringsurgery. The air flow from the device isHEPA-filtered to reduce the presenceof particulate matter to reduce themicroorganisms to a level of < 5 CFUper m³ at the surgical site andinstruments.The surgical site of the patient isintended to be placed under the airflow which is directed downwards tothe surgical site and/or instrumentswithin 32" (80 cm) in length. Deviceeffectiveness may not be reliablydetectable at a distance of 32" (80 cm)from the air flow outlet, andeffectiveness depreciates beyond thisspecified area.Operio is a portable device for use in asurgical operating room that producesa directed, non-turbulent flow of air tothe surgical site during ophthalmicsurgery and to the sterile instrumentsused during surgery. The air flow fromthe device is HEPA-filtered to reducethe presence of particulate matter toreduce the microorganisms to a level of< 5 CFU per m³ at the surgical site andinstruments.The air flow is intended to be directedparallel to the surgical site and/orinstruments within: 20" (50 cm) inwidth, 47" (120 cm) in length and 15"(40 cm) in height. Device effectivenessmay not be reliably detectable at adistance of 47 inches (120 cm) from theair flow outlet, and effectivenessdepreciates beyond this specified area.
PhysicalDimension102-150 x 63 x 95 inches24 x 18 x 47-67 inches
Weight992 – 1763 lbs92 lbs
MaterialStainless steel, aluminium and safetyglassAluminium, powder coated steel, ABSplastic, stainless steel
ElectricalSpecifications1. Mains power input: 120 V (AC), 60Hz2. Power consumption: 350 VA1. Mains power input: 120 V (AC), 60 Hz2. Power consumption: 290 VA
Type of devicePrescription usePrescription use
MovabilitySurgiCube® is a stand-alone deviceOperio is a portable device
Air filtrationHEPA filtered air with at least 99.995%efficiency against 0.3 µm particlesHEPA filtered air with at least 99.995%efficiency against 0.3 µm particles
HEPA filter mediaWaterproof glass fibreNelior membrane media.Folded sheets of nelior.
Air inlet Pre-filterMediaSynthetic fibres and high efficiencywater resistant glass fibrePolyamide mesh
RegularmaintenanceHEPA filter to be replaced after 1 yearand pre-filters to be replaced after 6months.HEPA filter to be replaced after 2000hours/yearly.
Airflow capacitySC100: 353 CFMSC180: 636 CFMSC200: 706 CFMSC220: 777 CFMAirflow is 235 CFM
EfficiencyBacteria carrying particles at thesurgical site is < 0.12 CFU/m³Bacteria carrying particles at thesurgical site is 0.4 CFU/m3.
Airflow speed0.45 m/s0.4 m/s
Air volumedelivered to incisionin CFMSC100: 353 CFMSC180: 636 CFMSC200: 706 CFMSC220: 777 CFMAirflow over surgical site is 235 CFM
Measured averageparticulate densityat surgical incision18 particulates of 0.5 µm/f³ (666/m³)1.1 particulate of 0.5 µm/f³
Air flow positionAir flow is provided over the surgicalsite and instruments.Air flow is provided over the surgicalsite and instruments.
Sterile accessorySingle-use sterile drapeSingle-use sterile shield
Sterility Assurance(SAL)for steriledrape10⁻⁶ SAL10⁻⁶ SAL
Sterilization Methodfor sterile drapeETOGamma irradiation

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for SurgiCube International. The logo consists of a blue square with a white curved line inside, next to the text "SurgiCube" in a light blue font. Below "SurgiCube" is the word "International" in a smaller, light blue font.

Summary of Testing

SurgiCube® fulfils and has been tested against the electromagnetic compatibility requirements. An electromagnetic compatibility test has been carried out at an accredited test laboratory. The report indicates that the SurqiCube® complies to the FCC requlation and 47 CFR 15. The relevant clauses for electrical safety were tested to confirm that the SurgiCube® complies with the relevant electrical requirements of IEC 60601-1 to ensure that no hazardous situation can occur for the patient. The test results have been presented in the submission.

The following non-clinical tests has been conducted: Simulated use test, Particulate counting test, Colony Forming Units counting test, Air velocity test, Air leakage test, Smoke tests and Air cleaning efficiency in ambient area to demonstrate air flow and efficiency.

The bench tests demonstrate that SurgiCube® met all performance and acceptance criteria, and does not introduce considerations for safety and efficacy.

Summary of clinical testing

Clinical tests with 26 patients were carried out at ophthalmic procedures. Measurements where made near the surgical site and over the instruments where the results demonstrated a statistically significant reduction in CFU levels as well as particulates. Air viables and particles were measured in order to determine the cleanliness of the clean air area at the surgical site and the instrument site when the SurgiCube® was used for air filtering on patients.

The analysis of the data revealed that it is possible to state at a 95% confidence level that

  • The average number of air viables is with 0.12 CFU/m³ well below 5 CFU/m³ 1.
    1. The average number of particles with a size of ≥0.5 µm is with 666 per m³ well below 3520 per m3
    1. A significant improvement is achieved when comparing the values inside the Surgicube® with the values outside the Surgicube®:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for SurgiCube International. The logo consists of a blue square with a white curved line inside, next to the text "SurgiCube" in a light blue font. Below "SurgiCube" is the word "International" in a smaller, light blue font. The logo is clean and modern, with a focus on the company's name and international presence.

a. Approximately 470 times less particles of ≥0.5 µm per m³ (313 821 in background vs. 666 inside Surgicube) and approximately 725 times less particles of ≥5.0 µm per m³ (1952 in background vs. 2.69 in Surgicube)

b. More than 126 times less CFU per m³ (15.19 in background vs. 0.12 inside Surgicube)

Conclusion

Based on a comparison of the performance characteristics, the tests performed, clinical and nonclinical, it is concluded that the subject device SurgiCube® is substantially equivalent to the predicate device.

§ 878.5070 Air-handling apparatus for a surgical operating room.

(a)
Identification. Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.(b)
Classification. Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.