K Number
K163367
Date Cleared
2017-08-11

(254 days)

Product Code
Regulation Number
866.5920
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

GenetiSure Dx Postnatal Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) and copy-neutral loss of heterozygosity (cnLOH) in genomic DNA obtained from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. GenetiSure Dx Postnatal Assay is intended for the detection of CNVs and cnLOH associated with developmental delay, intellectual disability, congenital anomalies or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate. Interpretation of assay results is intended only by healthcare professionals, board certified in clinical cytogenetics. The assay is intended to be used on the SureScan Dx Microarray Scanner System and analyzed by CytoDx Software.

This device is not intended to be used for standalone diagnostic purposes, pre-implantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.

Device Description

The GenetiSure Dx Postnatal Assay is a clinical laboratory in vitro diagnostic assay for performing molecular karyotyping based on array comparative genomic hybridization (aCGH) and single nucleotide polymorphism (SNP) analysis from blood samples of post-natal patients who are suspected of having a genomic abnormality. This molecular karyotyping is a modified in situ hybridization technique that allows detection and mapping of DNA sequence copy difference(s) between two genomes in a single experiment. In molecular karyotyping analysis, two differentially labeled genomic DNAs (subject/test sample and a reference sample) are co-hybridized to complementary nucleic acid sequences synthesized in situ on a microarrav slide.

Locations of copy number variation (CNVs) and copy-neutral loss of heterozygosity (cnLOH) in the DNA segments of the subject sample genome are revealed by variable fluorescence intensity on the microarray.

The assay compares the patient sample against a sex-matched reference sample. Genomic DNA (gDNA) is extracted from the patient's whole blood and then is fluorescently labeled in parallel with the reference sample using two different fluorescent dyes. The two labeled samples are hybridized to complementary sequences (probes) that are printed on a CGH+SNP microarray.

After hybridization, the microarrays are washed and then scanned. The data from the microarray images are converted to numeric data. The relative abundance of the target sequences is computed based on the relative intensities of the fluorophores in the patient and reference samples hybridized to each of the probe sequences.

The numeric data is then processed using software specifically designed to report CNVs by chromosomal location. The reported CNVs are interpreted by a Board Certified Cytogeneticist, Molecular Geneticist, Molecular Pathologist, or similarly qualified clinician who has been trained to identify the clinically relevant CNVs, determine clinical significance, and report out these findings. cnLOH in patient samples is also reported to the clinician.

AI/ML Overview

The provided text describes the GenetiSure Dx Postnatal Assay, a qualitative assay for detecting copy number variations (CNV) and copy-neutral loss of heterozygosity (cnLOH) in genomic DNA from peripheral whole blood.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are not explicitly stated in a single, consolidated table with pass/fail values. Instead, they are described within the context of each analytical performance study (Reproducibility, Precision, Accuracy, Stability, Limit of Detection, Analytical Specificity). The "reported device performance" is essentially the results presented in various tables and text summaries throughout Section K, "Performance Characteristics."

Below is a synthesized table based on the provided text, outlining the implied acceptance criteria (where quantifiable and explicitly stated in relation to a target metric) and the reported performance:

CategoryAcceptance Criteria (Implied/Stated)Reported Device Performance
Reproducibility (Overall Pairwise Replicate Agreement - 50% Overlap)Not explicitly stated as a single numeric criteria for overall, but implied high agreement for various categories.Overall CN gains and losses: 85.0% (50% overlap), 82.3% (80% overlap)
Small CNVs (5-20 probes): N/A (grouped by kb/probe count)CN Gain: 85.7% (50% overlap), 84.4% (80% overlap)
Large CNVs (>20 probes): N/ACN Loss: 84.6% (50% overlap), 81.3% (80% overlap)
cnLOH regions: N/AcnLOH: 89.1% (50% overlap), 87.9% (80% overlap)
Reproducibility (Positive Percent Agreement - 50% Overlap)N/AAll CNVs: 89.3%
cnLOH: 92.7%
Precision (Between-Lot Reagent and Scanner - Pairwise Replicate Agreement - 50% Overlap)Not explicitly stated as a single numeric criteria for overall, but implied high agreement for various categories.Small CNVs (5-20 probes): 83.33%
Large CNVs (>20 probes): 98.39%Large CNVs (>20 probes): 98.39%
cnLOH regions: 80.80%cnLOH regions: 80.80%
Precision (DNA Extraction - Pairwise Replicate Agreement - 50% Overlap)Small CNVs (5-20 probes): 70%Small CNVs (5-20 probes): 82.17%
Large CNVs (>20 probes): 80%Large CNVs (>20 probes): 98.47%
cnLOH regions: N/AcnLOH regions: 81.15%
Accuracy (Confirmation Rate - 50% Overlap, Excluding Indeterminate CNVs)N/A (grouped by probe count and kb)All CNVs: 92.6% (Scheme b)
cnLOH: 85.8% (Scheme b)
Accuracy (Endpoint Agreement - ≤2 probes for CNVs, ≤50 probes for cnLOH)N/AAll CNVs (combined start/stop): 87.4%
cnLOH (combined start/stop): 91.6%
Whole Blood Stability (Confirmation Rate - 50% Overlap)Small CNVs (5-20 probes): 75%Small CNVs (5-20 probes): Met criteria
Large CNVs (>20 probes): 90%Large CNVs (>20 probes): Met criteria
Interfering Substances (Confirmation Rate - 50% Overlap)Small CNVs (5-20 probes): 75%Small CNVs (5-20 probes): Met criteria
Large CNVs (>20 probes): 90%Large CNVs (>20 probes): Met criteria
Cross-ContaminationNo suspected cross-contamination detected, supporting appropriateness of workflow.No suspected cross-contamination detected.
Limit of Detection (DNA Input)Data supports conservative LLOD at 375 ng and common ULOD at 1000 ng for CN and cnLOH. LLOD could be 250ng for CNVs only.Performance does not decline down to 375 ng; robust at 500 ng.
Limit of Detection (Mosaicism)Reliable detection of large CNV (>100 probes) down to 50% level.Large CNV (>100 probes) reliably detected at 50% or greater admixture.
Clinical Validity (PPA for CNVs only)N/A. The comparator here is the collection site's own interpretation.76.8% (95%CI=68.2%-83.6%)
Clinical Validity (NPA for CNVs only)N/A. The comparator here is the collection site's own interpretation.95.5% (95%CI=93.7%-96.8%)
Clinical Validity (PPA for all aberrations)N/A. The comparator here is the collection site's own interpretation.76.7% (95%CI=68.3%-83.5%)
Clinical Validity (NPA for all aberrations)N/A. The comparator here is the collection site's own interpretation.89.8% (95%CI=87.3%-91.8%)

2. Sample Size Used for the Test Set and the Data Provenance

  • Reproducibility:
    • Sample Size: 48 test samples.
    • Data Provenance: Cell-lines with known aberrations ("common syndromes," "analytically challenging regions," "claimed minimal resolution," "varying aberration size ranges," and "genomic coverage of aberrations"). Origin not explicitly stated, but cell lines are generally laboratory-derived. It's a prospective experimental study.
  • Precision (Between-Lot Reagent and Scanner):
    • Sample Size: 48 test samples.
    • Data Provenance: Cell-lines with known aberrations (copy number gains, losses, and cnLOH). Origin not explicitly stated. It's a prospective experimental study.
  • Precision (DNA Extraction):
    • Sample Size: 24 samples.
    • Data Provenance: Cell-lines (implied, as they were tested by each operator for DNA extraction and then run in the assay). Origin not explicitly stated. It's a prospective experimental study.
  • Accuracy:
    • Sample Size: 556 eligible samples total, comprising:
      • 451 aberrant genomic DNA (gDNA) samples from established commercial cell lines.
      • 76 archived clinical gDNA samples from whole blood specimens of anonymized patients.
      • 5 globally recognized syndrome reference panel gDNA samples.
      • 24 fresh blood-derived gDNA samples from phenotypically normal subjects.
    • Data Provenance:
      • Commercial cell lines, globally recognized reference panels (implies international or broadly recognized sources).
      • Archived clinical gDNA samples are retrospective, anonymized from whole blood specimens; country of origin not specified.
      • Fresh blood samples from phenotypically normal subjects are likely prospective, implying a controlled collection process, but specific origin not detailed.
  • Stability (Whole Blood):
    • Sample Size: 24 whole blood specimens (12 male, 12 female).
    • Data Provenance: Obtained from a blood bank; implies clinical samples. Origin not explicitly stated. It's a prospective experimental study.
  • Limit of Detection (DNA Input):
    • Sample Size: 24 gDNA samples.
    • Data Provenance: Obtained from Coriell Institute for Medical Research (a well-known biobank).
  • Limit of Detection (Mosaicism):
    • Sample Size: Aberrant cell line DNAs mixed with reference background DNA. Specific number of samples/admixtures not explicitly quantified but described as "admixtures."
    • Data Provenance: Cell-line DNAs. Origin not explicitly stated.
  • Analytical Specificity (Interfering Substances):
    • Sample Size: 24 whole blood samples (12 male, 12 female).
    • Data Provenance: Drawn from phenotypically normal subjects. Origin not explicitly stated.
  • Analytical Specificity (Cross Contamination):
    • Sample Size: 2 male and 2 female Coriell DNA samples.
    • Data Provenance: Coriell DNA samples.
  • Clinical Validity:
    • Sample Size: 800 samples from patients suspected of pathogenic aberrations (SPA samples) + 100 samples from phenotypically normal individuals = Total 900 samples.
    • Data Provenance:
      • SPA samples from three (3) regionally distinct clinical institutions (implies clinical, retrospective or newly collected for study, "collected from" rather than "archived"). Country of origin not explicitly stated.
      • Phenotypically normal samples used to assess aberrations in a normal population. Origin not explicitly stated for these.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Accuracy Study (Comparator step for Agilent aberrations): Ground truth was established by independent (non-Agilent) commercially available microarray-based assays, analytically validated for copy number detection, and if discrepancies arose, by qPCR. This suggests comparative technical methods rather than direct expert interpretation as the primary ground truth for individual aberration confirmation.
  • Clinical Validity Study:
    • Ground Truth Establishment: The aberrations detected by the GenetiSure Dx Postnatal Assay for all 900 samples were interpreted by one of four cytogeneticists.
    • Qualifications of Experts: The experts were cytogeneticists. No specific experience level (e.g., "10 years of experience") is provided, but it is implicitly understood that a "cytogeneticist" is a qualified professional in this field.

4. Adjudication Method for the Test Set

  • Reproducibility, Precision (Between-Lot Reagent and Scanner), DNA Extraction Precision, Whole Blood Stability, Interfering Substances, Cross-Contamination: The primary adjudication method for individual aberration calls was Pairwise Replicate Analysis (PRA), requiring at least a 50% overlap of chromosomal coordinates for confirmation between replicates. An 80% overlap criterion was also used for some analyses, described as "more stringent." For DNA extraction, results were based on 18 replicates per sample.
  • Accuracy Study:
    • Initial confirmation of Agilent aberrations was against microarray-based comparator platforms (requiring 50% overlap).
    • Discrepancy Resolution/Adjudication: If an Agilent CNV aberration could not be confirmed by the microarray comparator, an analytically validated method (qPCR) was employed. Additionally, 5% randomly selected "confirmed" CNV aberrations were also subjected to qPCR, and CNVs near the limit of resolution (from normal whole blood samples) were also confirmed by the third method. This represents a 2+1 (or 1+1+1) type adjudication involving multiple technical methods.
  • Clinical Validity Study:
    • Initial Interpretation: Interpretations (Benign, Likely Benign, VOUS, Likely Pathogenic, Pathogenic) were made by one of four cytogeneticists.
    • Confirmation: All reported Pathogenic and Likely Pathogenic CNVs were subject to confirmation by alternative methods:
      • Review of prior confirmation data from the sample collection site (e.g., qPCR, FISH, karyotype).
      • If no prior data, qPCR assays were performed for confirmation.
    • For cnLOH, no analytical confirmation was done, as clinical practice involves follow-up studies, which were beyond the study scope. This implies that for clinical validity related to cnLOH, the cytogeneticist's interpretation was the primary "ground truth" within the study's scope without further technical confirmation for the study itself.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance?

No MRMC study was performed, and this device (GenetiSure Dx Postnatal Assay) is not an AI-assisted diagnostic tool. It is a molecular diagnostic assay using microarray technology, and its software (CytoDx) performs feature extraction and aberration identification, but there is no mention of AI or machine learning algorithms, nor human reader improvement with AI assistance. The interpretation step is performed by human experts (cytogeneticists).

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done?

Yes, the analytical performance studies (Reproducibility, Precision, Accuracy, Limit of Detection, Analytical Specificity) essentially evaluate the standalone performance of the assay and its associated software (CytoDx Software) in detecting and identifying aberrations before a human cytogeneticist's interpretation. The "Confirmation Rate" and "Pairwise Replicate Agreement" metrics in these sections reflect the algorithmic performance in agreement with reference methods or across replicates.

The clinical validity study also evaluates the device's ability to identify aberrations that are subsequently interpreted by experts, but the core analytical performance measures the direct output of the assay system.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • Analytical Studies (Reproducibility, Precision, LOD, Analytical Specificity):
    • Known aberrations in cell lines/reference DNA: Many studies used samples with "known aberrations" from commercial cell lines or reference panels (e.g., Coriell Institute). The nature of "known" implies a pre-established characterization, likely through previous molecular testing, expert characterization, or consensus characterization of these widely used reference materials.
    • Microarray-based comparators and qPCR: For the Accuracy study, the ground truth for establishing confirmation rates was based on agreement with independent, commercially available, analytically validated microarray-based assays, and/or qPCR for discrepancy resolution.
  • Clinical Validity Study:
    • Clinical site historical array data: For comparison, the device's results were compared to "historical array data from the respective collection site," which were generated using internal laboratory methods.
    • Expert cytogeneticist interpretation (with confirmation): The final interpretation of pathogenicity (and thus the "ground truth" for classifying samples as pathogenic/non-pathogenic within the study) was established by one of four cytogeneticists, with Pathogenic and Likely Pathogenic CNVs subject to confirmation by alternative methods (prior site data or qPCR). For cnLOH, the cytogeneticist's interpretation served as the ground truth without further technical confirmation in the study.

8. The Sample Size for the Training Set

The provided documentation does not explicitly detail a "training set" size. This is a 510(k) submission for a diagnostic assay, and the studies described are primarily analytical validation and clinical validation studies, focused on demonstrating performance metrics (accuracy, precision, reproducibility) against defined standards or comparator methods. For this type of device, the "training" (development) of the assay and its algorithms would have occurred prior to these validation studies. Information about the data used during the development phase of the CytoDx software or the assay itself (which could be considered training data for algorithm development) is not included in this document. The samples described in the performance characteristics sections are test/validation sets.

9. How the Ground Truth for the Training Set Was Established

As no specific "training set" is described for this 510(k) submission, information on how its ground truth was established is not provided. It is presumed that any internal development or training data used to optimize the assay or software algorithms would have utilized similar methods for ground truth establishment as outlined in the analytical and clinical validation sections (e.g., reference materials, comparator methods).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a stacked formation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 11, 2017

Agilent Technologies, Inc. Bill Kurani, Ph.D. Director of Regulatory Affairs and Ouality Assurance 5301 Stevens Creek Blvd. Santa Clara, CA 95051

Re: K163367

Trade/Device Name: GenetiSure Dx Postnatal Assay Regulation Number: 21 CFR 866.5920 Regulation Name: Postnatal chromosomal copy number variation detection system Regulatory Class: Class II Product Code: PFX Dated: July 11, 2017 Received: July 12, 2017

Dear Bill Kurani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Reena Philip -S

Reena Philip, Ph.D. Director Division of Molecular Genetics and Pathology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163367

Device Name GenetiSure Dx Postnatal Assay

Indications for Use (Describe)

GenetiSure Dx Postnatal Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) and copy-neutral loss of heterozygosity (cnLOH) in genomic DNA obtained from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. GenetiSure Dx Postnatal Assay is intended for the detection of CNVs and cnLOH associated with developmental delay, intellectual disability, congenital anomalies or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate. Interpretation of assay results is intended only by healthcare professionals, board certified in clinical cytogenetics. The assay is intended to be used on the SureScan Dx Microarray Scanner System and analyzed by CytoDx Software.

This device is not intended to be used for standalone diagnostic purposes, pre-implantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5. 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).

A. 510(k) Number:

K163367

B. Purpose for Submission:

Clearance of new device

C. Submitter Information:

  • Submitter: Aqilent Technologies, Incorporated 5301 Stevens Creek Boulevard Santa Clara, CA 95051
    Establishment Registration No: 2916205
Contact:Lois NakayamaSr. Regulatory Affairs Specialist
Phone:+1 408 553 2715
E-mail:lois.nakayama@agilent.com

Contact: Bill Kurani Director, Regulatory Affairs and Quality Assurance Phone: +1 408 553 2007 E-mail: bill.kurani@agilent.com

Contact: Philip Klimbal Program Manager Phone: +1 858 373 6490 E-mail: phil.klimbal@agilent.com

Date Prepared: November 28, 2016

D. Name of Device and Classification

Name: GenetiSure Dx Postnatal Assay

Classification: Class II, 21 CFR § 866.5920 Postnatal chromosomal copy number variation detection system

Product Code: PFX

Panel: Immunology (82)

{4}------------------------------------------------

E. Predicate Device

Affymetrix CytoScan Dx Assay (K130313)

The Agilent GenetiSure Dx Postnatal Assay (the device) is substantially equivalent to the Affymetrix CytoScan Dx Assay (the predicate) as described in the premarket notification K130313.

F. Type of Test or Tests performed:

Chromosomal microarray

G. System Description:

1. Device Description

a) Overview

The GenetiSure Dx Postnatal Assay is a clinical laboratory in vitro diagnostic assay for performing molecular karyotyping based on array comparative genomic hybridization (aCGH) and single nucleotide polymorphism (SNP) analysis from blood samples of post-natal patients who are suspected of having a genomic abnormality. This molecular karyotyping is a modified in situ hybridization technique that allows detection and mapping of DNA sequence copy difference(s) between two genomes in a single experiment. In molecular karyotyping analysis, two differentially labeled genomic DNAs (subject/test sample and a reference sample) are co-hybridized to complementary nucleic acid sequences synthesized in situ on a microarrav slide.

Locations of copy number variation (CNVs) and copy-neutral loss of heterozygosity (cnLOH) in the DNA segments of the subject sample genome are revealed by variable fluorescence intensity on the microarray.

The assay compares the patient sample against a sex-matched reference sample. Genomic DNA (gDNA) is extracted from the patient's whole blood and then is fluorescently labeled in parallel with the reference sample using two different fluorescent dyes. The two labeled samples are hybridized to complementary sequences (probes) that are printed on a CGH+SNP microarray.

After hybridization, the microarrays are washed and then scanned. The data from the microarray images are converted to numeric data. The relative abundance of the target sequences is computed based on the relative intensities of the fluorophores in the patient and reference samples hybridized to each of the probe sequences.

The numeric data is then processed using software specifically designed to report CNVs by chromosomal location. The reported CNVs are interpreted by a Board Certified Cytogeneticist, Molecular Geneticist, Molecular Pathologist, or similarly qualified clinician who has been trained to identify the clinically relevant CNVs,

{5}------------------------------------------------

determine clinical significance, and report out these findings. cnLOH in patient samples is also reported to the clinician.

b) Components of the Product

  • i. Microarray
K1201A    GenetiSure Dx Postnatal Assay
         K1201-64500 GenetiSure Dx Postnatal Array
         K1201-64600 GenetiSure Dx Postnatal Gasket

Agilent will provide to customers 4x180K aCGH+SNP microarray slides, manufactured under QSR, as the microarray component of this kit. Each single-use microarray slide carries four identical microarrays.

The microarray contains approximately 107,000 probes optimized for CNV analysis, and approximately 59,000 bi-allelic SNP probes. Each probe is approximately 60 bases long.

The CNV probes are distributed across the entire genome with a higher density of probes in regions designated by the International Standards for Cytogenomic Arrays (ISCA) consortium to be of clinical interest. These regions include:

  • telomere and selected centromere regions, .
  • . microdeletion / duplication regions,
  • dosage sensitive regions, and ●
  • · regions associated with X-linked disorders.

Overall, 94% of the genome is targeted with at least 5 CNV probes per 400 kb. The ISCA regions are targeted with a median probe spacing of approximately 1 CNV probes per 3.5 kb.

The SNP probes are designed to known SNPs that overlap restriction digestion sites (Alu I/Rsa I). These probes allow for identification of cnLOH.

Overall, 91% of the genome is targeted with at least 100 SNP probes per 10 Mb.

{6}------------------------------------------------

ii. Gasket slides

The gasket slides are single-use, silicone gasket backing slides that hold the samples during hybridization to the microarrays.

iii. Reagent

K1201-64100GenetiSure Dx DNA Labeling Kit
K1201-64105GenetiSure Dx Labeling Kit, -20C Part
K1201-64110GenetiSure Dx Labeling Kit, RT Components
K1201-64200GenetiSure Dx Hybridization Kit
K1201-64300GenetiSure Dx Wash Buffer Set
K1201-64305GenetiSure Dx Wash Buffer 1, 4L
K1201-64310GenetiSure Dx Wash Buffer 2, 4L
K1201-64400GenetiSure Dx Cot-1 Human DNA

A complete list of laboratory equipment and reagents required is provided in the GenetiSure Dx Postnatal Assay Instructions for Use (IFU).

iv. Software

K1203-10000 CytoDx 1.0 Software

Agilent CytoDx Software performs feature extraction, CNV and cnLOH identification and reporting on the microarray TIF images generated by the SureScan Dx Microarray Scanner.

The software application has three functional components, described briefly below.

    1. Feature Extraction uses image files created by the scanner, verifies the quality of the scanned image, extracts the intensity information and calculates signal and background for each feature/probe on the array, applying appropriate normalization.
    1. CNV and Allele Identifier uses the data produced during the Feature Extraction step to evaluate the intensity of each probe, applying algorithms that identify aberrations, SNPs, CNVs and cnLOH compared to the reference sample.
    1. Chromosome Viewer uses the CNV Table and LOH Intervals Table produced during CNV and Allele Identification to overlay the chromosomal aberrations or LOH detected onto an image of the chromosomes to provide a graphical representation of the results.

{7}------------------------------------------------

c) Components Required but Not Included in the System

  • Special Instrument Requirements i.
    SureScan Dx Microarray Scanner G5761A

d) Specimen Processing

The Agilent GenetiSure Dx Postnatal Kit Instructions for Use (IFU) provides instructions to enable the user to process DNA obtained from blood specimens. The illustration and text below provide an overview.

Image /page/7/Figure/5 description: This image shows a flowchart of the microarray process. The process starts with DNA isolation, quantitation, and qualitative analysis, followed by sample fragmentation, and sample labeling. The process ends with microarray processing, which includes 24-hour hybridization, microarray washing, microarray scanning, and aberration reporting.

Image /page/7/Figure/6 description: The image is titled, "Figure 1-Agilent GenetiSure Dx Postnatal Kit Sample Processing Workflow from gDNA to Report Generation." The title describes the image as a figure that shows the sample processing workflow of the Agilent GenetiSure Dx Postnatal Kit. The workflow starts with gDNA and ends with report generation. The figure is labeled as Figure 1.

i. Specimen requirements

The GenetiSure Dx Postnatal Assay is for use with gDNA from whole blood specimens collected in tubes using EDTA as the anticoagulant. Blood specimens may be stored at 2-8°C for up to 7 days prior to DNA extraction. Two hundred (200) microliters of whole blood are used for DNA extraction, and 0.5 micrograms (500 ng) of DNA are required to perform the GenetiSure Dx Postnatal Assay.

ii. Workflow sequence, from sample to report

{8}------------------------------------------------

The Agilent GenetiSure Dx Postnatal Kit starts with purified gDNA isolated from EDTA whole blood samples using the Qiagen QIAmp DSP DNA Blood Mini Kit (part number 61104).

    1. gDNA is quantified using a double-stranded DNA-based fluorometric method. For each sample to be tested, 0.5 µg of the subject's DNA is processed in parallel with 0.5 uq of the sex-matched reference DNA included in the GenetiSure Dx DNA Labeling Kit.
    1. Both the gDNA of the subject and the sex-matched reference are restriction digested with Alu I and Rsa I restriction enzymes included in the Labeling Kit.
    1. After digestion, the samples are labeled in parallel with the reference sample using the fluorescent dyes provided in the Labeling Kit. The subject sample is labeled with cyanine 5 (Cy5) dye and the sex-matched reference sample is labeled with cyanine 3 (Cy3) dye.
    1. The two labeled samples are hybridized onto a single microarray of a 4x180k aCGH+SNP slide using the reagents in the GenetiSure Dx Hybridization Kit, GenetiSure Dx Gasket 4xArray Slides (part of the Postnatal Assay) and Hybridization Chamber Kit. Prepared slides are hybridized for 24 hours at 67℃ in a light eliminating Hybridization Oven rotating at 20 rpm.
    1. After hybridization, the microarrays are washed using the GenetiSure Dx Wash Buffer Set and transferred into a SureScan Microarray Scanner slide holder.
    1. The microarray slides are then scanned in the SureScan Dx Microarray Scanner.
    1. The scanner generated image is then processed using the Agilent CytoDx Software, and the image data are converted to numeric data using the Feature Extraction module of the software. The relative abundance of the target sequences is computed by the Analytics module, based on the relative intensities of the fluorophores in the patient and reference samples hybridized to each of the probe sequences. CNVs and cnLOH are reported by chromosomal location.
    1. The reported CNVs and cnLOH are interpreted by a Board Certified Cytogeneticist, Molecular Geneticist, Molecular Pathologist, or similarly qualified clinician who has been trained to identify the clinically relevant CNVs, determine clinical significance, and report out these findings.

iii. Quality Control (QC)

Internal control probes on each array are used to calculate array QC metrics and assess the quality of data. As external controls, Agilent Male and Agilent Female reference DNA (provided with the Labeling Kit) are sexmatched, processed alongside, and co-hybridized with each test sample. These reference DNAs are used for data normalization and aberration detection on a per test sample basis, and to aid in troubleshooting, if necessary.

The following QC checks are required to assure reliable results:

{9}------------------------------------------------

    1. Sample input: only samples with sufficient amount of gDNA obtained by DNA extraction/purification procedures proceed to labeling: minimum of 500 ng is required.
    1. Labeling/In-Process QC: only samples passing DNA yield and specific activity measurements proceed to array hybridization. The required amount of fluorescently labeled DNA obtained after labeling/purification procedures is 8-15 ug, and the specific activity, i.e. the amount of dye (Cy3 or Cy5) incorporated into DNA after labeling and purification is 20-45 pmol Cv3 dye/ug of DNA or 20-40 pmol Cy5 dye/ug of DNA.
    1. Array QC metrics: The software uses the signal from probes on the microarray to perform a series of data verifications that detect laboratory processing anomalies. These include automated grid finding, probe-toprobe noise, signal-to-noise ratios and SNP call rates. Only arrays passing the QC metrics proceed to analysis. If the assay fails any of the array QC metrics, the software will generate a report for review of the QC metrics, but "sign-off" will not be allowed for the report.

2. Description of Test Report

The aberrations identified in a patient sample by the CytoDx algorithms can be viewed from the Triage View screen of the CytoDx software. The final Cyto Report is generated when the Lab Director signs off on the sample results, and lists the aberrations.

Clinical interpretation is performed by a Board Certified Cytogeneticist, Molecular Geneticist, Molecular Pathologist, or similarly qualified clinician who has been trained to identify the clinically relevant CNVs and cnLOH intervals, determine clinical significance, and report out these findings. Clinical interpretation of the aberration results takes place during classification of the aberrations (e.g., as Pathogenic, Likely pathogenic, VOUS, Likely benign, or Benign) in the CytoDx software. To select the appropriate classification, the clinician can rely on prior knowledge, comparisons to tracks or to other samples, and references to gene ontology database that describe the functions and disease associations for genes impacted by the aberrations. The classification assignments appear in the Cyto Report in the Classifications section.

Details of the report are available in the IFU.

H. Intended Use:

1. Intended Use:

GenetiSure Dx Postnatal Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) and copy- neutral loss of heterozygosity (cnLOH) in genomic DNA obtained from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. GenetiSure Dx Postnatal Assay is intended for the detection of CNVs and cnLOH associated with developmental delay, intellectual disability, congenital anomalies, or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate.

{10}------------------------------------------------

Interpretation of assay results is intended to be performed only by healthcare professionals, board certified in clinical cytogenetics or molecular genetics. The assay is intended to be used on the SureScan Dx Microarray Scanner System and analyzed by CytoDx Software.

This device is not intended to be used for standalone diagnostic purposes, preimplantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.

2. Indications for Use:

Same as above.

3. Special Condition for Use Statement(s):

For prescription use only.

{11}------------------------------------------------

I. Standard/Guidance Document Referenced (if applicable):

Table 1 – Documents referenced

NoStandard DevelopingOrganizationStandard DesignationNumber and DateTitle of Standard
1ISO14971 Second edition2007-03-01;14971:2012Medical devices - Application of riskmanagement to medical devices
2ISOISO 23640:2011In vitro diagnostic medical devices --Evaluation of stability of in vitro diagnosticreagents
3AAMI ANSI ISO15223-1:2012Medical devices - Symbols to be used withmedical devices labels, labeling, andinformation to be supplied - Part 1: Generalrequirements
4AAMI ANSI IEC62304:2006Medical device software - Software life cycleprocesses
5AAMI ANSI IEC62366-1 Edition 1.02015-02Medical devices - Part 1: Application ofusability engineering to medical devices
6CLSIEP07-A2 - 05/21/2007Interference Testing in Clinical Chemistry;Approved Guideline - Second Edition
7CLSIEP12-A2 - 01/30/2014User Protocol for Evaluation of QualitativeTest Performance
8CLSIEP25-A - 01/15/2013Evaluation of Stability of In Vitro DiagnosticReagents; Approved Guideline
9CLSIMM13-A - 12/01/2005Collection, Transport, Preparation, andStorage of Specimens for Molecular Methods;Approved Guideline
10CLSIMM21: 1st Edition – Aug2015Genomic Copy Number Microarrays forConstitutional Genetic and OncologyApplications

J. Substantial Equivalence Discussion

The predicate device for the GenetiSure Dx Postnatal Assay is the Affymetrix CytoScan Dx Assay, cleared in January, 2014 under K130313.

Comparisons between the GenetiSure Dx Postnatal Assay and its predicate device are presented in the following tables.

{12}------------------------------------------------

Table 2-Similarities between assay and predicate

GenetiSure Dx Postnatal AssayAffymetrix CytoScan Dx Assay
(Device)(Predicate)K130313
Indications forUseGenetiSure Dx Postnatal Assay is aqualitative assay intended for thepostnatal detection of copy numbervariations (CNV) and copy-neutral loss ofheterozygosity (cnLOH) in genomic DNAobtained from peripheral whole blood inpatients referred for chromosomal testingbased on clinical presentation. GenetiSureDx Postnatal Assay is intended for thedetection of CNVs and cnLOH associatedwith developmental delay, intellectualdisability, congenital anomalies, ordysmorphic features. Assay results areintended to be used in conjunction withother clinical and diagnostic findings,consistent with professional standards ofpractice, including confirmation byalternative methods, parental evaluation,clinical genetic evaluation, andcounseling, as appropriate. Interpretationof assay results is intended to beperformed only by healthcareprofessionals, board certified in clinicalcytogenetics or molecular genetics. Theassay is intended to be used on theSureScan Dx Microarray Scanner Systemand analyzed by CytoDx Software.This device is not intended to be used forstandalone diagnostic purposes, pre-implantation or prenatal testing orscreening, population screening, or for thedetection of, or screening for, acquired orsomatic genetic aberrations.CytoScan® Dx Assay is a qualitativeassay intended for the postnataldetection of copy number variations(CNV) in genomic DNA obtained fromperipheral whole blood in patientsreferred for chromosomal testing basedon clinical presentation. CytoScan® DxAssay is intended for the detection ofCNVs associated with developmentaldelay, intellectual disability, congenitalanomalies, or dysmorphic features.Assay results are intended to be used inconjunction with other clinical anddiagnostic findings, consistent withprofessional standards of practice.including confirmation by alternativemethods, parental evaluation, clinicalgenetic evaluation, and counseling, asappropriate. Interpretation of assayresults is intended to be performed onlyby healthcare professionals, boardcertified in clinical cytogenetics ormolecular genetics. The assay isintended to be used on the GeneChip®System 3000Dx and analyzed byChromosome Analysis Suite DxSoftware (ChAS Dx Software).This device is not intended to be usedfor standalone diagnostic purposes,pre-implantation or prenatal testing orscreening, population screening, or forthe detection of, or screening for,acquired or somatic genetic aberrations.
SpecialConditionsFor prescription useSame.
Sample TypePeripheral whole bloodSame.
TechnologyMicroarray for comparative genomichybridizationSame.
SoftwareAssay-specific software is used to performfeature extraction, CNV and cnLOHidentification and reporting on themicroarray images.Same.
Assay stepsStarts with purified genomic DNA (gDNA)and ends with microarray intensity data.Same.
Quality ControlsIn-process QC checks, external controlsand array QC metrics are used to monitorand assess the quality of results.Same.
ReportThe device reports the copy numberchange (gain, loss) and loss ofheterozygosity aberrations, andposition/location of the aberrant segmentacross the queried genome.Same.
LimitationsThis device is not intended to be used forstandalone diagnostic purposes, pre-implantation or prenatal testing orscreening, population screening, or for thedetection of, or screening for, acquired orsomatic genetic aberrations.Same.
ClinicalValidationCompare test results with availablediagnosis for sample.Same.

{13}------------------------------------------------

Table 3-Differences between assay and predicate

GenetiSure Dx Postnatal Assay(Device)Affymetrix CytoScan Dx Assay(Predicate)
Array Format60-mer probesFour microarrays on a single 1"x 3" glassslide25-mer oligosIndividual microarrays housed in aGeneChip cartridge
Method of ArrayManufactureOn slide (in-situ) synthesis of probes usingink-jet printingOn-wafer synthesis of probes usingphotolithography
DNAFragmentation/LabelingFragmented DNA is directly labelled withfluorescent dye (Cy3 and Cy5) beforehybridization.Fragmented DNA is PCR amplifiedand then labelled with biotin beforehybridization.
Data is produced in two intensity channelswhich are then compared to generate aLogRatioSingle channel data is produced whichis later compared to an in silicoreference to produce a LogRatio.
HybridizationCohybridization of labeled sample andreference for direct on-array comparisonHybridization of single labeled samplewhich is compared to an in silicoreference.
Washing /Staining ofMicroarraysManual washing process in accordancewith instruction for the validated diagnosticassayAutomated processing with FS450Dxfluidics station for both washing andstaining steps
EquipmentSureScan Dx Microarray ScannerGeneChip System 3000Dx Scanner

K. Performance Characteristics

1. Analytical Performance

a) Reproducibility

The aim of the Reproducibility Study was to demonstrate that GenetiSure Dx Postnatal Assay achieves acceptable, reproducible results when performed at multiple laboratory sites by multiple operators over multiple days. Replicates of forty-eight (48) test samples were processed by two separate operators, at each of

{14}------------------------------------------------

three individual clinical laboratories, in three (3) one-week intervals for a total of 864 data points.

The forty-eight (48) test samples were selected from cell-lines with a wide range of known aberrations (copy number gains, losses, and copy neutral loss of heterozygosity (cnLOH)). The aberrations met the following criteria: common syndromes ('known syndromic regions'), analytically challenging regions, claimed minimal resolution, varying aberration size ranges, and genomic coverage of aberrations. Multiple samples had multiple aberrations spanning multiple criteria. Test sample selection criteria encompasses aberrations expected to be found in normal whole blood samples.

All individual aberrations reported within each processed test sample, regardless of expected pathogenicity, were compared to their respective replicates (18 replicates for each aberration, operator by site by week) by pairwise replicate analysis (PRA), requiring at least 50% overlap of chromosomal coordinates for confirmation. Positive agreement was assessed separately for small copy number variants (CNVs, 5-20 probes contained with the aberration), larger CNVs (>20 probes), or cnLOH regions. The results demonstrate that the pre-defined acceptance criteria were met for each category with a pairwise replicate agreement of 80.22%, 95.83%, and 89.08%, respectively. Using a more stringent 80% overlap criteria for pairwise replicate agreement, acceptance criteria were also met.

Data are further refined by size, probe number, aberration type, and study variable (e.g. operator, site, test sample). Alternative metrics of positive percent agreement, call rate, and breakpoint accuracy/endpoint deviation are presented.

To provide additional insight into the reproducibility of the test as a function of reported aberration size (in kb), the data were categorized into more refined size bins (see Table 4). The results demonstrated that when comparing all replicates of all test samples across all days, sites, and operators, using a 50% aberration overlap criteria, the overall pairwise replicate agreement across all sizes of CN gains and losses was 85.0%. Pairwise replicate agreement across the various kb bins ranged from 75.9% to 100%. For copy number gains, the overall agreement was 85.7%; for losses, the overall agreement was 84.6%. For cnLOH, the overall pairwise replicate agreement was 89.1%. Applying a more stringent 80% overlap criteria produced overall agreements of 82.3% for CN gains and losses combined, 84.4% for gains, 81.3% for losses, and 87.9% for cnLOH.

When assessing specifically the agreement between replicates for positive aberration calls by PPA analysis, the agreement was 89.3% for all copy number calls and 92.7% for cnLOH using the 50% overlap criteria.

Call rate averaged 78.1% for CNVs, and 74.9% for cnLOH.

{15}------------------------------------------------

Table 4-Reproducibility of Aberrations Categorized by Size (in kb) and Type

AberrationTypeAberrationRange (kb)# AberrationsCall Rate (%)PairwiseReplicateAgreement (%)PPA (%)
50%80%50%80%
CN Gain10 - 50551.282.582.582.982.9
50 - 100368.796.396.397.397.3
100 - 2001350.579.979.880.180.0
200 - 5002682.786.384.691.589.4
500 - 1000979.780.278.987.686.0
1000 - 2000772.190.782.892.481.6
2000 - 50001165.175.975.979.779.7
5000 +1393.298.498.499.199.1
Total8773.885.784.489.988.2
CN Loss10 - 501451.676.876.177.676.2
50 - 1002100.0100.089.5100.089.5
100 - 2002381.482.378.188.182.9
200 - 5003182.681.875.586.378.7
500 - 10005572.881.276.285.278.5
1000 - 20003083.386.485.991.591.0
2000 - 50001888.987.485.189.987.3
5000 +20100.0100.0100.0100.0100.0
Total19380.184.681.389.084.8
All CNVs(Gain &Loss)Total28078.185.082.389.385.8
cnLOH5000 - 100002150.677.176.877.476.8
10000 - 200001191.599.096.499.496.7
20000 +13100.0100.098.4100.098.4
Total4574.989.187.992.791.1

When results were binned by the number of probes in an aberration, rather than size in kb, using the 50% overlap criteria, the overall pairwise replicate agreement was similar to the above (see Table 5): 86.2% for combined CN gains and losses

{16}------------------------------------------------

(ranging from 70.6% to 100%), 86.1% for gains alone, 86.3% for losses alone, and 89.1% for cnLOH.

Using the 80% overlap criteria, overall agreements were 84.6% for CN gains and losses combined, 85.3% for gains, 84.2% for losses, and 88.4% for cnLOH. PPA for the 50% overlap criteria was 90.9% and 92.7% for CNVs and cnLOH, respectively. Call rate averaged 78.1% for CNVs, and 74.9% for cnLOH.

{17}------------------------------------------------

Table 5-Reproducibility of Aberrations Categorized by Probe Number and Type
AberrationTypeAberrationRange (#Probes)# AberrationsCall Rate (%)Pairwise ReplicateAgreement (%)PPA (%)
Overlap50%80%
CN Gain5 - 71138.076.576.569.069.0
7 - 101554.170.669.472.870.6
10 - 152387.989.488.394.092.7
15 - 201166.582.479.986.883.1
20 - 30989.697.597.597.997.9
30 - 100370.393.093.095.095.0
100 - 500372.390.290.293.293.2
500 +12100.0100.0100.0100.0100.0
Total8773.886.185.390.589.4
CN Loss5 - 73661.176.675.680.979.1
7 - 103965.577.675.482.879.6
10 - 154281.985.581.090.585.0
15 - 201896.995.994.097.695.7
20 - 301687.589.185.791.687.9
30 - 1001092.293.292.996.396.0
100 - 50017100.0100.0100.0100.0100.0
500 +15100.0100.0100.0100.0100.0
Total19380.186.384.291.188.5
AII CNV(Gain &Loss)Total28078.186.284.690.988.8
cnLOH100 to 2002554.880.380.382.082.0
200 to 50013100.0100.097.7100.097.7
> 5007100.0100.0100.0100.0100.0
Total4574.989.188.492.791.8

{18}------------------------------------------------

CONCLUSIONS: More refined categorization of aberrations by size and probe number, and further analyses of these confirmations using PPA and call rate calculations, support the reproducibility conclusions established for small (5-20 probes), large (>20 probes), and cnLOH categories by pairwise confirmation. In general, pairwise replicate agreement, PPA, and call rate increase with aberration size and probe number, although the aberration numbers within each bin varies.

b) Precision

i. Between-Lot Reagent and Scanner Precision

The aim of the Between-Lot Reagent and Scanner Precision Study was to demonstrate that GenetiSure Dx Postnatal Assay achieves acceptable, precise results when performed using multiple reagent manufacturing lots and when analyzed on multiple scanner instruments. Forty-eight (48) test samples containing a range of chromosomal aberrations (copy number gains, losses, and copy-neutral loss of heterozygosity (cnLOH)) were processed by multiple operators, using combinations of three (3) reagent lots and three (3) scanner instruments across three (3) processing weeks at a single site for a total of 432 data points.

Individual aberrations called within each processed test sample were compared to their respective replicates (9 replicates for each aberration, representing 3x3 reagent-lot/scanner combinations) by pairwise replicate analysis (PRA), requiring at least 50% overlap of chromosomal coordinates for confirmation. Agreement was assessed separately for small copy number variants (CNVs, 5-20 probes contained within the aberration), larger CNVs (>20 probes), or cnLOH regions. The results demonstrate that the pre-defined acceptance criteria were met for each category with a pairwise replicate agreement of 83.33%, 98.39%, and 80.80%, respectively. Results were similar when using a more stringent 80% overlap criteria for pairwise replicate agreement. In addition, no substantial differences were observed when the pairwise replicate agreement was assessed separately for inter-lot vs. intra-lot replicate pairs, or for inter-scanner vs. intra-scanner replicate pairs.

Data were further refined by size, probe number, aberration type, and study variable (e.g. reagent lot, scanner, processing week). Alternative metrics of positive percent agreement, call rate, and breakpoint accuracy/endpoint deviation are presented.

To provide additional insight into the precision of the test as a function of reported aberration size (in kb), the data were categorized into more refined size bins (see Table 6). The results demonstrated that when comparing all replicates of all test samples across all lots, scanners, and weeks, using a 50% aberration overlap criteria, the overall pairwise replicate agreement across all sizes of copy number qains and losses was 89.0%. Pairwise

{19}------------------------------------------------

replicate agreement across the various kb bins ranged from 76.9% to 100%. For copy number gains, the overall agreement was 85.4%; for copy number losses, the overall agreement was 91.3%. For cnLOH, the overall pairwise replicate agreement was 80.8%. Applying a more stringent 80% overlap criteria produced overall agreements of 87.2% for all CNVs (gains or losses) combined, 84.0% for CN gains, 89.2% for CN losses, and 76.0% for cnLOH.

When assessing specifically the agreement between replicates for positive aberration calls by PPA analysis, the agreement was 93.0% for all copy number calls and 87.3% for cnLOH using the 50% overlap criteria.

Call rate averaged 83.0% for CNVs, and 75.6% for cnLOH.

Table 6-Reproducibility of Aberrations Categorized by Size (in kb) and Type

Pairwise ReplicateAgreement (%)PPA (%)
AberrationTypeAberrationRange (kb)#AberrationsCall Rate(%)Overlap50%80%
50%80%
20-50659.376.976.980.580.5
50 - 100690.885.275.591.881.1
100 - 2001660.378.877.482.580.2
200 - 5003062.281.980.385.182.4
CN Gain500 - 1000661.077.876.978.877.3
1000 - 2000458.384.784.786.986.9
2000 - 50001893.988.688.693.493.4
5000 +17100.0100.0100.0100.0100.0
Total10375.085.484.089.988.1
10-502070.680.378.886.083.9
50 - 100494.590.390.394.994.9
100 - 2003382.891.989.593.890.9
200 - 5002481.586.886.591.991.5
CN Loss500 - 10003192.289.583.593.386.8
1000 - 20001591.995.093.797.295.8
2000 - 50006100.0100.096.3100.096.3
5000 +3399.799.399.099.799.3
Total16688.091.389.294.692.2
All CNVsTotal26983.089.087.293.090.8

{20}------------------------------------------------

Pairwise ReplicateAgreement (%)PPA (%)
AberrationTypeAberrationRange (kb)#AberrationsCall Rate(%)Overlap50%80%
cnLOH5000 - 100002366.750%80%79.575.9
10000 - 200001881.572.770.392.681.4
20000 +595.688.078.995.995.9
Total4675.692.292.287.381.0

When results were binned by the number of probes in an aberration, rather than size in kb, using the 50% overlap criteria, the overall pairwise replicate agreement was similar to the above (see Table 7): 89.6% for combined CN gains and losses (ranging from 70.6% to 100%), 85.6% for CN gains alone, 92.1% for CN losses alone, and 80.8% for cnLOH.

Using the 80% overlap criteria, overall agreements were 88.4% for CN gains and losses combined, 84.6% for CN gains, 90.8% for CN losses, and 76.4% for cnLOH. PPA for the 50% overlap criteria was 93.7% and 87.3% for CNVs and cnLOH, respectively. Call rate averaged 83.0% for CNVs, and 75.6% for cnLOH.

Table 7-Reproducibility of Aberrations Categorized by Probe Number and Type

AberrationTypeAberrationRange (#Probes)#AberrationsCall Rate(%)Pairwise ReplicateAgreement (%)PPA (%)
Overlap
5-72251.474.774.775.575.5
7-101051.170.669.271.268.5
10-152076.787.883.891.786.4
15 - 201480.282.182.188.988.9
CN Gain20 - 30683.384.383.388.987.8
30 - 100774.693.793.795.795.7
100 - 50010100.0100.0100.0100.0100.0
500 +14100.0100.0100.0100.0100.0
Total10375.085.684.690.288.9
CN Loss5-73972.687.686.891.590.4
7-102881.081.979.188.885.3

{21}------------------------------------------------

AberrationTypeAberrationRange (#Probes)#AberrationsCall Rate(%)Pairwise ReplicateAgreement (%)PPA (%)
Overlap
10-152287.989.887.694.191.7
15 - 201289.894.091.096.693.3
20 - 3017100.0100.0100.0100.0100.0
30 - 1001399.298.396.499.197.2
100 - 50015100.0100.0100.0100.0100.0
500 +20100.0100.0100.0100.0100.0
Total16688.092.190.895.594.0
All CNVsTotal26983.089.688.493.792.2
100 to 2003369.475.172.282.077.9
200 to 5001189.994.484.896.986.2
cnLOH500 +2100.0100.0100.0100.0100.0
Total4675.680.876.487.381.5

CONCLUSIONS: More refined categorization of aberrations by size and probe number, and further analyses of these confirmations using PPA and call rate calculations, support the conclusions regarding precision established for CNV (5-20 probes and >20 probes) and cnLOH categories by pairwise confirmation.

DNA Extraction Precision ii.

The extraction precision study was performed to assess the aberration calling concordance of the GenetiSure Dx Postnatal Assay across repeated DNA extractions from the same blood sample performed by different operators in multiple runs, to determine repeatability and precision.

A panel of twenty-four (24) samples was tested by each operator. The gDNA from the samples was extracted using the same lot of the Qiagen kit, at one site, in duplicate, by 3 operators, on 3 separate days, for a total of 432 extractions (3 operators x 3 days x 2 duplicates x 24 samples). Each of the three operators labelled a set of 48 extracted samples (24 different samples x 2 extractions) per week for 3 weeks for a total of 144 test results run per operator over the course of the study. The extracted gDNA replicates were tested in the GenetiSure Dx Postnatal Assay in 3 weeks, each corresponding to a specific day of extraction.

Primary analysis was performed using pairwise comparison of aberration results on each of the 18 replicates (3 operators x 3 days x 2 duplicates) for

{22}------------------------------------------------

each sample. An aberration was considered confirmed if at least 50% of the region of aberration overlapped between the replicates being compared.

The results of the 50% overlap analysis demonstrated that results obtained from multiple extractions of the same sample were highly concordant, regardless of operator and day upon which the samples were extracted. The individual Pairwise Replicate Agreement % values stratified by week or operator were similar to each other and to the overall averages shown in Table 8 (82.17% for copy number variants (CNVs) called by 5-20 probes, 98.47% for CNVs called by >20 probes, and 81.15% for copy-neutral loss of heterozygosity (cnLOH)), which further supports that similar assay performance can be expected from different extractions, personnel, days, and samples.

{23}------------------------------------------------

CategoryAberrationTypeUniqueAberrationsPairwiseConfirmedNumber ofComparisonsPercentConfirmedAcceptanceCriteria
5 to 20Gain364326539180.24NA
5 to 20Loss465817695383.66NA
5 to 20Gain orLoss82101431234482.1770
>20Gain294244438796.74NA
>20Loss3349324932100.00NA
>20Gain orLoss629176931998.4780
--cnLOH161959241481.15NA

Table 8 - Pairwise Replicate Agreement Based on 50% Overlap of Base Pairs

2. Accuracy

The analytical accuracy of the GenetiSure Dx Postnatal Assay was investigated by evaluating samples obtained from multiple sources and carrying a wide variety of chromosomal aberrations. Of 626 samples tested in total, 556 eligible samples constituted a comprehensive panel for accuracy evaluation. The diversified sample panel consisted of 451 aberrant genomic DNA (gDNA) samples derived from established commercial cell lines, 76 archived clinical gDNA samples isolated from whole blood specimens of anonymized patients, 5 globally recognized syndrome reference panel gDNA samples, and 24 fresh blood-derived gDNA samples extracted from whole blood of phenotypically normal subjects.

The samples were selected to maximize the variation across the genome with consideration for gain and loss segments of various sizes/number of probes, chromosomal representation, CNV regions in genic and non-genic regions, and in telomeric and centromeric regions. A total of 2187 CNV regions and 292 cnLOH regions covered 91% of the genome. These aberrations were more prevalent in nontelomeric/non-centromeric regions than in telomeric/centromeric (1337 regions vs 1130 regions). A total of 23% (508 out of 2187 regions) of the CNVs had high (>45%) GC content.

For all the 556 samples evaluated, 4.1 (Mean) ± 2.1 (SD) CNVs were detected on average per sample (Median=4), of which 2.3 ± 1.4 were small CNVs with 5-20 probes and 1.8 ± 1.6 were large CNVs with >20 probes (Median=2 for both size categories). More descriptive statistics for these 556 samples can be found in 9 (All Samples). These statistics remained almost unchanged if all the aberrant samples were considered together (532 samples consisting of both clinical and cell line samples). When considering the 24 normal blood-derived samples only, the average number of CNVs per sample was approximately halved: 2.0 (Mean) ± 1.5 (SD), Median=2. The majority of

{24}------------------------------------------------

these detected CNVs were (5-20 probes), with more being CN Loss than CN Gain. Large CNVs (>20 probes) detected in normal blood samples were relatively rare (Mean =0.25, Median=0 per sample).

Table 9: Aberration Counts Per Sample

All Samples Included in Evaluation (N= 556)
DistributionCN GainCN LossCNV 5-20probesCNV >20probesCNV-allcnLOH
Max2798282833
90th Percentile344361
Median122240
10th Percentile011020
Min000000
Mean1.62.52.31.84.10.5
Std Dev1.71.61.41.62.13.0
Normal Whole Blood Samples Only (N= 24)
DistributionCN GainCN LossCNV 5-20probesCNV >20probesCNV-allcnLOH
Max355251
90th Percentile234140.5
Median011.5020
10th Percentile000000
Min000000
Mean0.671.31.80.252.00.08
Std Dev0.871.31.40.531.50.28

{25}------------------------------------------------

Due to the diversified sample panel composition and lack of an applicable universal comparator, independent (non-Agilent) commercially available microarray based assays analytically validated for copy number detection were employed to assist accuracy assessment of CNV aberrations and resolve discrepancies. The samples were tested through the GenetiSure Dx Postnatal Assay using standard procedures in a designated Agilent laboratory.

A target Agilent aberration was deemed "confirmed" if a minimum percent overlap was found with comparator aberration call(s) of the same type (gain. Ioss, or cnLOH). For the data presented below, a 50% overlap was required, with the Agilent aberrations being compared independently to the other platforms. All eligible Agilent aberrations were assessed, one at a time (Figure 2).

Image /page/25/Figure/2 description: The image shows a diagram that explains the concept of '% Overlap'. The diagram includes labels such as 'Agilent Comparator', 'Target Agilent Aberration', 'Ia', 'Ib', and 'U'. The formula for calculating '% Overlap' is given as (Ia+Ib)/U.

Figure 2: Method for comparison of Agilent detected aberrations with those of comparator platform(s).

If an Agilent CNV aberration could not be "confirmed" by the microarray-based comparator, another analytically validated method (qPCR) was employed to adjudicate the results. To avoid bias in the assessment, an additional 5% randomly selected "confirmed" CNV aberrations (separately selected for CNVs>20 probes and CNVs with 5-20 probes) were also included in this discrepancy resolution testing. Other CNVs directly subject to a third method confirmation included CNVs near the limit of resolution selected from normal whole blood samples.

The results are summarized and stratified by copy number state, size or probe number, and genomic region. The results shown in Table 9 are presented as stratified by probe number, either including indeterminate CNVs as unconfirmed (scheme a) or excluding indeterminate CNVs (scheme b). In Table 10, the results are presented using the same schemes for treatment of indeterminate CNVs, with the results stratified by length in kb.

{26}------------------------------------------------

Scheme a: Including Indeterminate CNVs as "Not Confirmed"

TYPEAberrationRange (#of Probes)SampleSize(N)*#ConfirmedConfirmation Rate (95%CI)FPR (95% CI)
Gain5-7483879.2% (65.7%, 88.3%)20.8% (11.7%, 34.3%)
7-101018584.2% (75.8%, 90.0%)15.8% (10.0%, 24.2%)
10-1519715277.2% (70.8%, 82.5%)22.8% (17.5%, 29.2%)
15-201018382.2% (73.6%, 88.4%)17.8% (11.6%, 26.4%)
20-5014810470.3% (62.5%, 77.0%)29.7% (23.0%, 37.5%)
50-500826478.0% (67.9%, 85.6%)22.0% (14.4%, 32.1%)
500 +16916698.2% (94.9%, 99.4%)1.8% (0.6%, 5.1%)
Total84669281.8% (79.1%, 84.3%)18.2% (15.7%, 20.9%)
Loss5-721619690.7% (86.1%, 93.9%)9.3% (6.1%, 13.9%)
7-1020216581.7% (75.8%, 86.4%)18.3% (13.6%, 24.2%)
10-1525721684.0% (79.1%, 88.0%)16.0% (12.0%, 20.9%)
15-201259072.0% (63.6%, 79.1%)28.0% (20.9%, 36.4%)
20-501309573.1% (64.9%, 80.0%)26.9% (20.0%, 35.1%)
50-50022521796.4% (93.1%, 98.2%)3.6% (1.8%, 6.9%)
500 +18618096.8% (93.1%, 98.5%)3.2% (1.5%, 6.9%)
Total1341115986.4% (84.5%, 88.2%)13.6% (11.8%, 15.5%)
All CNVs2187185184.6% (83.1%, 86.1%)15.4% (13.9%, 16.9%)
cnLOH100-2001329471.2% (63.0%, 78.2%)28.8% (21.8%, 37.0%)
200-5001029694.1% (87.8%, 97.3%)5.9% (2.7%, 12.2%)
500 +5858100.0% (93.8%, 100.0%)0.0% (0.0%, 6.2%)
Total29224884.9% (80.4%, 88.6%)15.1% (11.4%, 19.6%)

{27}------------------------------------------------

  • The number of aberrations analyzed in each range bin, including indeterminate CNVs as "not confirmed".

{28}------------------------------------------------

Scheme b: Excluding Indeterminate CNVs

TYPEAberrationRange (#of Probes)SampleSize(N)*#ConfirmedConfirmation Rate (95%CI)FPR (95% CI)
Gain5-7433888.4% (75.5%, 94.9%)11.6% (5.1%, 24.5%)
7-10918593.4% (86.4%, 96.9%)6.6% (3.1%, 13.6%)
10-1517515286.9% (81.1%, 91.1%)13.1% (8.9%, 18.9%)
15-20918391.2% (83.6%, 95.5%)8.8% (4.5%, 16.4%)
20-5012410483.9% (76.4%, 89.3%)16.1% (10.7%, 23.6%)
50-500726488.9% (79.6%, 94.3%)11.1% (5.7%, 20.4%)
500 +16916698.2% (94.9%, 99.4%)1.8% (0.6%, 5.1%)
Total76569290.5% (88.2%, 92.3%)9.5% (7.7%, 11.8%)
Loss5-719719699.5% (97.2%, 99.9%)0.5% (0.1%, 2.8%)
7-1018316590.2% (85.0%, 93.7%)9.8% (6.3%, 15.0%)
10-1523121693.5% (89.6%, 96.0%)6.5% (4.0%, 10.4%)
15-201029088.2% (80.6%, 93.1%)11.8% (6.9%, 19.4%)
20-501149583.3% (75.4%, 89.1%)16.7% (10.9%, 24.6%)
50-50022221797.7% (94.8%, 99.0%)2.3% (1.0%, 5.2%)
500 +18418097.8% (94.5%, 99.2%)2.2% (0.8%, 5.5%)
Total1233115994.0% (92.5%, 95.2%)6.0% (4.8%, 7.5%)
All CNVs1998185192.6% (91.4%, 93.7%)7.4% (6.3%, 8.6%)
cnLOH100-2001329471.2% (63.0%, 78.2%)28.8% (21.8%, 37.0%)
200-500999697.0% (91.5%, 99.0%)3.0% (1.0%, 8.5%)
500 +5858100.0% (93.8%, 100.0%)0.0% (0.0%, 6.2%)
Total28924885.8% (81.3%, 89.4%)14.2% (10.6%, 18.7%)

{29}------------------------------------------------

  • The number of aberrations analyzed in each range bin, excluding indeterminate CNVs

Table 10: GenetiSure Dx Postnatal Assay Accuracy for all aberration regions stratified by aberration size (kb) when compared to the comparator method (Schemes a & b)

Scheme a: Including Indeterminate CNVs as "Not Confirmed"

TYPEAberrationRange (kb)SampleSize(N)*#ConfirmedConfirmation Rate (95%CI)FPR (95% CI)
Gain20-100693753.6% (42.0%, 64.9%)46.4% (35.1%, 58.0%)
100-200947579.8% (70.6%, 86.7%)20.2% (13.3%, 29.4%)
200-30013612088.2% (81.7%, 92.6%)11.8% (7.4%, 18.3%)
300-5001239274.8% (66.5%, 81.6%)25.2% (18.4%, 33.5%)
500-1,000907077.8% (68.2%, 85.1%)22.2% (14.9%, 31.8%)
1000-10,00016813681.0% (74.3%, 86.2%)19.0% (13.8%, 25.7%)
10,000 +16616297.6% (94.0%, 99.1%)2.4% (0.9%, 6.0%)
Total84669281.8% (79.1%, 84.3%)18.2% (15.7%, 20.9%)
Loss10-100885967.0% (56.7%, 76.0%)33.0% (24.0%, 43.3%)
100-20020718087.0% (81.7%, 90.9%)13.0% (9.1%, 18.3%)
200-30012911488.4% (81.7%, 92.8%)11.6% (7.2%, 18.3%)
300-50011610388.8% (81.8%, 93.3%)11.2% (6.7%, 18.2%)
500-1,00020916478.5% (72.4%, 83.5%)21.5% (16.5%, 27.6%)

{30}------------------------------------------------

1000-10,00039835188.2% (84.6%, 91.0%)11.8% (9.0%, 15.4%)
10,000 +19418896.9% (93.4%, 98.6%)3.1% (1.4%, 6.6%)
Total1341115986.4% (84.5%, 88.2%)13.6% (11.8%, 15.5%)
All CNVs2187185184.6% (83.1%, 86.1%)15.4% (13.9%, 16.9%)
cnLOH5,000-10,000936165.6% (55.5%, 74.5%)34.4% (25.5%, 44.5%)
10,000-20,000948489.4% (81.5%, 94.1%)10.6% (5.9%, 18.5%)
20,000 +10510398.1% (93.3%, 99.5%)1.9% (0.5%, 6.7%)
Total29224884.9% (80.4%, 88.6%)15.1% (11.4%, 19.6%)
  • The number of aberrations analyzed in each range bin, including indeterminate CNVs as "not confirmed".

Scheme b: Excluding Indeterminate CNVs

ТҮРЕAberrationRange (kb)SampleSize(N)*#ConfirmedConfirmation Rate (95%CI)FPR (95% CI)
Gain20-100533769.8% (56.5%, 80.5%)30.2% (19.5%, 43.5%)
100-200877586.2% (77.4%, 91.9%)13.8% (8.1%, 22.6%)
200-30012312097.6% (93.1%, 99.2%)2.4% (0.8%, 6.9%)
300-5001079286.0% (78.2%, 91.3%)14.0% (8.7%, 21.8%)
500-1,000777090.9% (82.4%, 95.5%)9.1% (4.5%, 17.6%)
1000-10,00015313688.9% (82.9%, 92.9%)11.1% (7.1%, 17.1%)
10,000 +16516298.2% (94.8%, 99.4%)1.8% (0.6%, 5.2%)

{31}------------------------------------------------

Total76569290.5% (88.2%, 92.3%)9.5% (7.7%, 11.8%)
Loss10-100765977.6% (67.1%, 85.5%)22.4% (14.5%, 32.9%)
100-20018718096.3% (92.5%, 98.2%)3.7% (1.8%, 7.5%)
200-30011511499.1% (95.2%, 99.8%)0.9% (0.2%, 4.8%)
300-50010510398.1% (93.3%, 99.5%)1.9% (0.5%, 6.7%)
500-1,00017916491.6% (86.6%, 94.9%)8.4% (5.1%, 13.4%)
1000-10,00037935192.6% (89.5%, 94.8%)7.4% (5.2%, 10.5%)
10,000 +19218897.9% (94.8%, 99.2%)2.1% (0.8%, 5.2%)
Total1233115994.0% (92.5%, 95.2%)6.0% (4.8%, 7.5%)
All CNVs1998185192.6% (91.4%, 93.7%)7.4% (6.3%, 8.6%)
cnLOH5,000-10,000936165.6% (55.5%, 74.5%)34.4% (25.5%, 44.5%)
10,000-20,000928491.3% (83.8%, 95.5%)8.7% (4.5%, 16.2%)
20,000 +10410399.0% (94.8%, 99.8%)1.0% (0.2%, 5.2%)
Total28924885.8% (81.3%, 89.4%)14.2% (10.6%, 18.7%)
  • The number of aberrations analyzed in each range bin, excluding indeterminate CNVs

Analysis was also conducted using alternative methods of comparing the aberrations to the comparator calls. Results using those methods were similar to those presented (data not shown). When calculated using one of the alternative methods, which included a predefined 65% minimum overlap criterion and a composite of aberrations from two platforms as a comparator, and a definitive scheme where by "indeterminate" calls were excluded, the aberration confirmation rates (% agreement with comparator) were 93.5% for larger CNVs with >20 probes, 92.5% for smaller CNVs with 5-20 probes, and 90.1% for all cnLOH aberrations.

{32}------------------------------------------------

The endpoint agreements were also analyzed. Agreements for one comparator are shown in Table 11. Analysis by other methods were found to have similar results.

Table 11: Endpoint Agreement for comparator platform 1; Binning by Number of Probes; Start/Stop Breakpoints Combined; Endpoint Agreement Criteria: ≤2 probes for CNVs (Gain/Loss), ≤50 probes for cnLOH

TYPEAberration Range(# of Probes)Breakpoints, NBreakpoint Agreement, NBreakpoint Agreement, % (95% CI)
Gain5-7484797.9% (89.1%, 99.6%)
7-1012611692.1% (86.0%, 95.6%)
10-1525822386.4% (81.7%, 90.1%)
15-201108880.0% (71.6%, 86.4%)
20-5012811589.8% (83.4%, 94.0%)
50-50012410483.9% (76.4%, 89.3%)
500 +29625887.2% (82.9%, 90.5%)
Total109095187.2% (85.1%, 89.1%)
Loss5-723623499.2% (97.0%, 99.8%)
7-1012411794.4% (88.8%, 97.2%)
10-151168976.7% (68.3%, 83.5%)
15-20846982.1% (72.6%, 88.9%)
20-5013811079.7% (72.2%, 85.6%)
50-50042038291.0% (87.8%, 93.3%)
500 +35028380.9% (76.4%, 84.6%)
Total1468128487.5% (85.7%, 89.1%)
All CNVs2558223587.4% (86.0%, 88.6%)
cnLOH100-20018817693.6% (89.2%, 96.3%)
200-50019417690.7% (85.8%, 94.1%)
500 +11610489.7% (82.8%, 94.0%)
Total49845691.6% (88.8%, 93.7%)

{33}------------------------------------------------

In summary, the analytical performance of the GenetiSure Dx Postnatal Assay has been determined to be highly accurate and appropriate for its intended use as a source of reliable information contributing to a clinical diagnosis in subjects with potential genetic aberrations.

{34}------------------------------------------------

3. Stability

a) Array and Reagent Stability

This study was designed to determine the stability of eight components in the context of the assay workflow.

Stability testing consisted of four different arms of the study:

  • . Shelf Life: testing after storage of assay components under recommended conditions at defined intervals after manufacturing.
  • . Multi-use testing: testing after components of the assay that are stored at -15°C to -25°C are subject to different numbers of freeze/thaw cycles
  • . Transport: Similar to Shelf life testing, but assay components were subject to a transportation simulation prior to storage at recommended conditions.
  • . In-use testing: testing after different components of the assay are stored for various times after preparation or testing where reaction intermediates are stored for various times.

Results of the study to date indicate that all tested conditions pass all metrics under all conditions tested, supporting a shelf life of the assay components of 12 months. In addition, multi-use and in-use testing support the tested conditions of the assay:

Up to:

  • 8x freeze/thaw for GenetiSure Dx DNA Labeling Kit components;
  • . 16x freeze/thaw or 120 davs' storage at -15°C to -25°C for reconstituted 10x Blocking Agent;
  • 16x freeze/thaw for Cot-1 DNA;
  • 60 days' storage of an open microarray package under specified ● conditions, and
  • 30 days' storage of the digested gDNA or labeled DNA reaction ● intermediates at -15°C to -25°C.

b) Whole Blood Stability

To determine the stability of whole blood specimens prior to gDNA isolation, 24 whole blood specimens, 12 male and 12 female, were obtained from a blood bank and gDNA was isolated from the specimens at 1, 3, 7, and 10 days after initial collection.

A list of aberrations for each sample extracted on Days 3, 7 and 10 were reported and compared with the 'Day 1' list for the same sample. An aberration from the 'Day 1' sample was considered confirmed in the stored samples if the test result identifies a region of aberration that overlaps the 'Day 1' region by at least 50%. Blood was considered stable when stored for a given time when 75% of small CNV (gain/loss) aberrations, called by 5-20 copy number probes, and 90% of the larger CNV (qain/loss)

{35}------------------------------------------------

aberrations, called by >20 probes, were confirmed. An additional analysis was also performed using an 80% overlap criteria.

The results of both the 50% and 80% overlap analysis methods demonstrated that whole blood specimens may be stored for up to 10 days at 2-8ºC prior to gDNA isolation. Samples stored for this period of time and processed with the GenetiSure Dx Postnatal Assay produced acceptable results.

4. Limit of Detection

a) DNA Input

To determine the analytical sensitivity, or the Limit of Detection (LOD) of the GenetiSure Dx Postnatal Assay, this study was conducted to evaluate the minimum and maximum amounts of genomic DNA (gDNA) acceptable as the assay input to detect copy number variations (CNVs) and copy neutral loss of heterozygosity (cnLOH) accurately. Twenty-four (24) gDNA samples with known chromosomal aberrations were obtained from Coriell Institute for Medical Research (Coriell). These DNA samples were tested in the assay using two (2) lots of reagents across a range of varied DNA input levels from 0.125 µg (125 ng) to 1 µg (1000 ng), with 0.5 µg (500 ng) as the recommended input quantity (standard).

The study assessed the impact of various gDNA input on aberration calling and determined the upper and lower limits of detection (ULOD and LLOD) of the assay by comparing the percentage of aberrations confirmed at each non-standard DNA input level against pre-defined acceptance criteria. Data from this study support the use of 500 ng as the recommended input amount. The study data, and supplemental data generated under similar study conditions, demonstrate that performance does not decline down to 375 ng. The data support a conservative LLOD at 375 ng and a common ULOD at 1000 ng for both copy number and cnLOH aberrations. For copy number aberrations only, the LLOD could be further reduced to 250 ng.

The assay performs robustly at the recommended input amount of 500 ng and is stable at considerably lower and higher amounts.

b) Mosaicism

To determine the sensitivity (LoD) for the detection of mosaic cytogenetic abnormalities, aberrant cell line DNAs containing known copy number changes were mixed with a reference background DNA in different percentages to mimic various levels of mosaicism. The admixtures were analyzed for the presence of the copy number changes detectable in the pure aberrant cell line DNAs. Secondary analysis addressed the rate of

{36}------------------------------------------------

false positive calls present in admixtures as compared to the pure reference sample.

Large copy number aberrations could be reliably detected when present in a 50% or greater admixture. Some aberrations were correctly identified at lower than a 50% level, but the sensitivity of detection was reduced. Results were similar for both gains and losses. Smaller aberrations could not be reliability detected in any of the admixtures. Despite the reduced sensitivity, the specificity of aberrations called remained largely unaffected across all levels of mosaicism evaluated and both size ranges.

The results demonstrate that the GenetiSure Dx Postnatal Assay has the sensitivity to detect large mosaic copy number gains and losses (>100 probes) down to the 50% level.

5. Analytical Specificity

a) Interfering Substances

To determine the effects of interfering substances on the results of the GenetiSure Dx Postnatal Assay, the study evaluated the impact of hemoglobin, conjugated bilirubin, unconjugated bilirubin and triglycerides (triolein) spiked into whole blood prior to gDNA isolation.

Blood drawn from twelve (12) phenotypically normal males and twelve (12) phenotypically normal females was used in the testing.

The list of aberrations for each sample containing a given interferent was reported and compared with the 'non-adulterated control' list for the same sample. An aberration was considered confirmed if the test result identified a region of aberration that overlaps between the sample and control by at least 50%. The test was considered robust to a given interferent when 75% of the CNVs (gains/losses) in the 5-20 probe category and 90% of the CNVs (gains/losses) in the >20 probe category in the 'non-adulterated control' were confirmed.

An additional analysis was also performed at 80% overlap between regions. The results of both the 50% and 80% overlap analysis methods demonstrated that the test results are not altered by the presence of excessive hemoglobin, triglycerides, or bilirubin (conjugated or unconjugated) in the patient whole blood specimen.

b) Cross Contamination

The GenetiSure Dx Postnatal Assay consists of a glass slide composed of four (4) independent microarrays that are sealed by a gasket slide during sample hybridization. Cross-contamination can arise during sample processing, especially during hybridization set-up, when samples are

{37}------------------------------------------------

loaded onto each of the 4 adjacent arrays or during the overnight incubation due to gasket leakage. The presence of contamination can result in corrupt and inaccurate patient data.

This study was designed to determine if cross contamination occurs during the routine GenetiSure Dx Postnatal Assay workflow and, if so, what the impact on data would be. For this study, two (2) male and two (2) female Coriell DNA samples, each with distinctive sets of known chromosomal aberrations, were tested across multiple microarray slides under conditions that would either allow or prevent detection of cross contamination between the adjacent arrays on the slides. Four (4) microarray slides served as the "non-contaminated condition" with four replicates of the same sample placed on each of the four arrays of the slides. Six (6) slides served as a test for "potential cross-contamination" that could occur between adjacent arrays within a single slide during the hybridization set-up or overnight incubation. For these slides, the sample replicates were alternated on the slide with sample replicates from a different sample. The copy number variation (CNV) and copy-neutral loss of heterozygosity (cnLOH) aberration results from the "potential cross-contamination" microarray slides were compared to the aberration results from the "non-contaminated condition" microarray slides, using a 50% overlap criteria, to determine if detectable cross contamination had occurred on the test slides. Additionally, gasketrelated cross-contamination was evaluated by use of three (3) different lots of gasket slides.

The 50% overlap analysis demonstrated that results obtained were highly concordant between the aberrations detected in both the "noncontaminated condition" and the "potential cross-contamination condition", easily meeting the established acceptance criteria, which were established based on the tested condition not impacting the detection of aberrations in a given sample. No suspected cross contamination was detected. This supports the appropriateness of the sample handling workflow, the accuracy of the data collected, and the integrity of the gasket slide materials for use with the GenetiSure Dx Postnatal Assay.

6. Clinical Validity

Eight hundred (800) samples from patients suspected of having pathogenic aberrations (SPA samples) were processed utilizing the GenetiSure Dx Postnatal Assay. The samples had been collected from three (3) regionally distinct clinical institutions that offered postnatal array testing for the detection of chromosomal abnormalities. One hundred (100) samples from phenotypically normal individuals were also processed using the GenetiSure Dx Postnatal Assay and were used to assess the aberrations that might be expected to be found in a normal (non-patient) population.

{38}------------------------------------------------

The aberrations detected in each sample, for all nine hundred (900) samples, were interpreted by one of four cytogeneticists as Benign, Likely Benign, Variant Of Unknown Significance (VOUS), Likely Pathogenic, or Pathogenic.

The test results, per sample, were compared to historical array data from the respective collection site, which were generated using the methods established at each laboratory.

All reported Pathogenic and Likely Pathogenic copy number variants (CNVs), gains and losses, were subject to confirmation by alternative methods:

  • (i) Confirmation data from the sample collection site was reviewed. If prior confirmation data, from the collection sites, was available (e.g., quantitative polymerase chain reaction (qPCR), fluorescence in situ hybridization (FISH) or karyotype), it was considered adequate, and no additional confirmation test was performed.
  • (ii) Remaining aberrations were assessed by qPCR assays for confirmation per pre-determined quidelines. Briefly, at least 1 gPCR CNV confirmatory assay was required within each target genomic region (copy number aberration interval). For larger regions (>400 kb), more than 1 qPCR assay was preferred, ideally with qPCR assays distributed across the target region, if possible. The actual number of qPCR assays selected was in part determined by the availability of predesigned assays in a given target region and the quantity of DNA available for analysis. Aberrations without any assays available in the region or without sufficient DNA were excluded from qPCR confirmation.

In routine clinical practice, reported Pathogenic and Likely Pathogenic copyneutral loss of heterozygosity (cnLOH) aberrations are not subject to analytical confirmation. Clinical follow up (such as parental, methylation and/or sequencing studies to understand the mechanism or impacts of the cnLOH) is conducted. Such follow up was beyond the scope of this study, and hence was not performed. The lack of clinical follow up on these aberrations does not impact the data interpretation for this study or the outcome of the study.

a) Study Results

The diagnostic yield for the GenetiSure Dx Postnatal Assay, when considering only copy number aberrations, was 15%. This increased to 20% when cnLOH aberrations were also considered.

{39}------------------------------------------------

CollectionSiteNumberofSamplesCollectionSite:Number ofPathogenicCallsCollection Site:Diagnostic YieldGenetiSureDx PostnatalAssay:Number ofPathogenicCallsGenetiSure DxPostnatal Assay:Diagnostic Yield
Copy Number Aberrations Only
Site 12572911% (8.0%, 15.7%)3915% (11.3%, 20.1%)
Site 23133511% (8.2%, 15.2%)3311% (7.6%, 14.4%)
Site 32304821% (16.1%, 26.6%)4520% (15.0%, 25.2%)
TOTAL80011214% (11.8%, 16.6%)11715% (12.3%, 17.2%)
All Aberrations (Copy Number and cnLOH)
Site 12572911% (8.0%, 15.7%)4819% (14.4%, 23.9%)
Site 23133912% (9.2%, 16.6%)6019% (15.2%, 23.9%)
Site 32304821% (16.1%, 26.6%)5122% (17.3%, 28.0%)
TOTAL80011615% (12.2%, 17.1%)15920% (17.3,22.8%)

Table 12: Diagnostic Yield by Collection Site (95% CI)

When considering only copy number aberrations, diagnostic yields were comparable between the collection sites and the GenetiSure Dx Postnatal Assay; however, differences were seen among the three collection sites, with Site 3 having a higher diagnostic yield than either Site 1 or Site 2. As the results from the GenetiSure Dx Postnatal Assay show the same trend across collection sites, this difference is likely due to diversity in the population of patients referred to the three sites, rather than a difference in the interpretation of aberrations, as the same set of cytogeneticists assessed the GenetiSure Dx Postnatal Assay aberrations, irrespective of the collection site.

The non-pathogenic category is encompassed by samples with interpretations of VOUS, Likely Benign, Benign, or those with no aberrations reported. Results of the PPA and NPA analysis are presented considering only copy number aberrations (Table 13) or considering both

{40}------------------------------------------------

copy number and cnLOH aberrations (Table 14). For the copy number aberration only analysis, samples with pathogenic cnLOH aberrations were considered as non-pathogenic, unless they also included a pathogenic copy number aberration.

Collection Site Aberration Interpretation
Pathogenic InterpretationNon-Pathogenic Interpretation
GenetiSure Dx PostnatalAssay InterpretationPathogenicLikelyPathogenicVOUSLikelyBenign1Normal2Total
PathogenicInterpretationPathogenic561490382
LikelyPathogenic124110835
Non-PathogenicInterpretationVOUS583503280
Normal267801509603
Total79331351552800
PPA386/112 = 76.8% (95%CI=68.2%-83.6%)
NPA4657/688 = 95.5% (95%CI=93.7%-96.8%)

Table 13: Comparison of Sample Classification between the Collection Site and the GenetiSure Dx Postnatal Assay, Considering only Copy Number Aberrations

"One Site 2 sample was presented with the interpretation on Likely Benign.

2 Samples from the GenetiSure Dx Postnatal Assay or Site 1 with either only Benign or Likely Benign aberrations, or samples without aberrations are classified as "Normal". Site 3 and Site 2 provided sample classification of "Normal".

Positive Percent Agreement (PPA): Percent [(Agilent GenetiSure Dx Postnatal Assay = Pathogenic & Collection site classification = Pathogenic)/(Collection site classification = Pathogenic)]

'Negative Percent Agreement (NPA): Percent [(Agilent GenetiSure Dx Postnatal Assay = Non-pathogenic & Collection site classification = Non-Pathogenic)/(Collection = Non-Pathogenic)]

When considering only copy number aberrations in the sample classification, PPA was 76.8% and NPA was 95.5%. In total, twenty-six (26) samples which were determined to have Pathogenic or Likely Pathogenic copy number aberrations by the collection sites were reported as non-pathogenic by the GenetiSure Dx Postnatal Assay. Most of these aberrations were either detected by GenetiSure Dx Postnatal Assay, but interpreted differently by the cytogeneticist, or below the detection limit of the Assay.

Table 14: Comparison of Sample Classification between the Collection Site and the GenetiSure Dx Postnatal Assay, Considering Copy Number and cnLOH Aberrations

Collection Site Aberration Interpretation
Pathogenic InterpretationNon-Pathogenic Interpretation

{41}------------------------------------------------

GenetiSure Dx PostnatalAssay InterpretationPathogenicLikelyPathogenicVOUSLikelyBenign1Normal2Total
PathogenicInterpretationPathogenic561490382
LikelyPathogenic1452303577
Non-PathogenicInterpretationVOUS51059046120
Normal275741434521
Total82341651518800
PPA389/116 = 76.7% (95%CI=68.3%-83.5%)
NPA4614/684 = 89.8% (95%CI=87.3%-91.8%)

1 One Site 2 sample was presented with the interpretation on Likely Benign.

2 Samples from the GenetiSure Dx Postnatal Assay or Site 1 with either only Benign or Likely Benign aberrations, or samples without aberrations are classified as "Normal". Site 3 and Site 2 provided sample classification of "Normal".

Positive Percent Agreement (PPA): Percent [(Agilent GenetiSure Dx Postnatal Assay = Pathogenic & Collection site classification = Pathogenic)//(Collection = Pathogenic)]

"Negative Percent Agreement (NPA): Percent [(Agilent GenetiSure Dx Postnatal Assay = Non-pathogenic & Collection site classification = Non-Pathogenic)/(Collection site classification = Non-Pathogenic)]

When considering all aberrations, PPA remained similar at 76.7%, and NPA dropped to 89.8%, which is consistent with the higher diagnostic vield for the GenetiSure Dx Postnatal Assay when considering all aberrations. Twenty-seven (27) samples were called as Pathogenic at the collection sites and non-pathogenic by the GenetiSure Dx Postnatal Assay.

A list of observed syndromes was compiled based on the pathogenic and likely pathogenic calls detected from the 800 SPA samples (Table 15). In total, 36 distinct syndromes were identified which encompassed 73 cases from the clinical study sample set.

Syndrome AssociationNumber ofCases
10q26 Deletion Syndrome1
13q Deletion Syndrome1
15q11.2 Deletion Syndrome5
15q13.3 Microdeletion Syndrome1
Syndrome AssociationNumber of Cases
15q25 Deletion Syndrome1
16p11.2 Microdeletion1
16p11.2 Microduplication2
16p12.1 Deletion Syndrome1
16p13.11 Microdeletion2
16p13.11 Microduplication neurocognitive disorder susceptibility locus3
1q21.1 Deletion Syndrome2
1q21.1 Duplication Syndrome2
22q11.2 Duplication Syndrome3
2q37 Deletion Syndrome4
3q29 Deletion Syndrome1
7q11.23 Duplication Syndrome1
8p23.1 Microdeletion/CDH syndrome1
8p23.1 Microduplication1
Angelman/Prader-Willi Syndrome6
Charcot-Marie-Tooth Neuropathy, Type 1a1
DiGeorge Syndrome7
Distal 22q11.2 Deletion Syndrome1
Downs Syndrome/Trisomy 216
Ichthyosis, X-Linked/ STS Deficiency1
Isodicentric Chromosome 15 Syndrome1
Jacobsen/ 11q Deletion1
Klinefelter Syndrome4
Mental Retardation-Hypotonic Facies Syndrome, X-Linked/ Smith-Fineman-Myers1
Neuropathy, Hereditary, With Liability to Pressure Palsies; HNPP1
Sotos Syndrome-1/ 5q35 Deletion Syndrome1
Triple X Syndrome2
Trisomy 9 mosaicism1
Turner Syndrome1
Syndrome AssociationNumber of Cases
Williams-Beuren Syndrome2
Xq26.3 Duplication Syndrome1
Other2
Total73

Table 15: List of Syndromes in the Clinical Study Sample Set

{42}------------------------------------------------

{43}------------------------------------------------

As per the SPA samples, data obtained for the phenotypically normal samples were aggregated at the sample level. Results of the sample level analysis are shown in the table below.

{44}------------------------------------------------

Table 15-Phenotypically Normal Individual Sample Level Summary

InterpretationNumber of Samples
Normal (all Likely Benign, Benign, or no aberrations identified)78
VOUS14
Likely Pathogenic5
Pathogenic3
Total100

In total, eight (8) of the 100 samples had aberrations classified as Likely Pathogenic or Pathogenic, the details of those aberrations are listed in Table 17

Table 17-Pathogenic and Likely Pathogenic Aberrations Detected in Phenotypically Normal Samples

Sample IDTypeSizeChr BandInterpretation
CLIS-CMC-0033-FGain100 kb22q13.33Pathogenic
CLIS-CMC-0216-FLoss18.2 kb16q24.3Pathogenic
CLIS-CMC-0253-MGain646 kbXp21.2-p21.1L. Pathogenic
CLIS-CMC-0259-FcnLOH20 Mb2q14.3-q22.3L. Pathogenic
CLIS-CMC-0281-FLoss13.6 kb15q11.2L. Pathogenic
CLIS-CMC-0372-McnLOH11 Mb6q23.3-q25.1L. Pathogenic
CLIS-CMC-0497-MGain59Mb+93MbXp + XqPathogenic
CLIS-CMC-0507-FGain1.5 Mb16p13.11L. Pathogenic

Of the eight (8) samples with Pathogenic or Likely Pathogenic aberrations reported, two (2) of those were cnLOH aberrations reported as Likely Pathogenic. The other six (6) samples contained copy number changes, five (5) of which were confirmed by qPCR. One (1) of the copy number changes, a 100 kb gain on Chromosome 22, was not confirmed by qPCR. Of note, one of the samples was identified as containing an additional X

{45}------------------------------------------------

chromosome (Karyotype 47, XXY), which is commonly associated with Klinfelter Syndrome (https://ghr.nlm.nih.gov/condition/klinefelter-syndrome).

PPA and NPA analysis revealed strong correlation with previous clinical data, especially for copy number changes. Addition of cnLOH aberrations in the comparison did not significantly impact the PPA, but the reduced NPA is consistent with the fact that few cnLOH were reported as pathogenic by the collection sites.

§ 866.5920 Postnatal chromosomal copy number variation detection system.

(a)
Identification. A postnatal chromosomal copy number variation detection system is a qualitative assay intended for the detection of copy number variations (CNVs) in genomic DNA obtained from whole blood in patients referred for chromosomal testing based on clinical presentation. It is intended for the detection of CNVs associated with developmental delay, intellectual disability, congenital anomalies, or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate. Interpretation of assay results is intended to be performed by qualified healthcare professionals such as clinical cytogeneticists or molecular geneticists. This device is not intended to be used for standalone diagnostic purposes, pre-implantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following information:
(i) A detailed description of all components in the test system that includes:
(A) A description of the assay components, array composition and layout, all required reagents, instrumentation, and equipment, including illustrations or photographs of non-standard equipment or methods;
(B) A description of the design of the array in terms of chromosomal coverage and probe density for different regions;
(C) An identification of the number of probes and size of the CNVs reported at the lower range of the assay;
(D) Detailed documentation of the device software, including standalone software applications and hardware-based devices that incorporate software;
(E) Methodology and protocols for detecting copy number and visualizing results;
(F) A description of the result outputs along with sample reports, and a description of any links to external databases provided by the device to the user or accessed by the device;
(G) Specifications for the methods to be used in specimen collection, extraction (including DNA criteria for DNA quality and quantity to perform the assay), and storage; and
(H) A description of appropriate internal and external controls that are recommended or provided. The description must identify those control elements that are incorporated into the testing procedure.
(ii) Information that demonstrates the performance characteristics of the system, including:
(A) Device reproducibility data generated, at a minimum, using three sites, with two operators at each site, for three non-consecutive days using at least three instruments. A well-characterized panel of samples that provide a wide range of CNVs (
i.e., gains, losses, adequate size coverage across the range of sizes claimed by the device, adequate chromosomal coverage, challenging regions in the genome, CNVs reported at the lower range of the assay, interstitial, subtelomeric, and pericentromeric rearrangements, aneuploidy, unbalanced translocations, mosaicism, and known syndromic regions) must be used. The results must be itemized for all CNVs detected in each sample across all replicates and summarized in a tabular format stratified by size range and range of probe numbers for gains and losses separately and calculated for overall. The results must be analyzed using pairwise replicate agreement, and summarized as overall pairwise replicate agreement as well as pairwise replicate agreement conditional on replicates having a positive copy number state call (gains or losses), call rate, CNV size variation, and endpoint agreement;(B) Device accuracy data using cell lines and clinical samples representing a variety of CNVs and syndromes. In this analytical study, accuracy must be determined for every CNV detected in a particular sample. The accuracy data provided must include the copy number state determination and endpoint accuracy. The accuracy samples must cover different genomic variations across the genome (
i.e., gains, losses, adequate CNV size coverage across the range of sizes claimed by the device, adequate chromosomal coverage, challenging regions in the genome, CNVs reported at the lower range of the assay, interstitial, subtelomeric, and pericentromeric rearrangements, aneuploidy, unbalanced translocations, mosaicism, and known syndromic regions). CNVs identified by the device must be compared to comparator method(s). Agreement between the CNVs detected by the array and the comparator must be summarized in a tabular format that includes the positive percent agreement and false positive rate stratified by size range and range of probe numbers for gains and losses separately and calculated for overall;(C) Assay performance data for CNVs reported at the lower range of the assay for both gains and losses;
(D) Device analytical sensitivity data, including DNA input and limit of detection for mosaicism, if applicable;
(E) Device analytical specificity data, including interference, carryover, and cross-contamination data;
(F) Device stability data, including real-time stability under various storage times, temperatures, and freeze-thaw conditions;
(G) Specimen matrix comparison data if more than one specimen type or anticoagulant can be tested with the device;
(H) Data that demonstrates the clinical validity, including diagnostic yield, of the device using a minimum of 800 retrospective clinical samples that were collected prospectively and obtained from three or more clinical laboratories, with results interpretation equally divided between two or more qualified healthcare professionals (
e.g., cytogeneticists). Patients must be representative of the intended use population and not limited to common syndromes. Diagnostic yield data must be summarized in tabular format and stratified by the comparison methodologies. Data must also be summarized comparing interpretation of results, with description of reasons for variability in calls between the device and the standard of care methods. Data to support the accuracy of calls for known syndromes must be included; and(I) Data that demonstrates device results when a minimum of 100 apparently healthy, phenotypically normal individuals are tested and interpreted by one or more cytogeneticists blinded to the patient status.
(iii) Identification of risk mitigation elements used by the device, including a description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing.
(2) The labeling required under § 809.10 of this chapter must include:
(i) A warning statement that the device is not intended to be used for standalone diagnostic purposes, pre-implantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations;
(ii) Limitations regarding the assay's performance with respect to validated CNVs reported at the lower range of the assay, stratified by size range and range of probe numbers for gains and losses separately; and limitations regarding problematic (hypervariable) regions, loss of heterozygosity, mosaicism, and inability to detect balanced translocations, as appropriate;
(iii) A warning statement that interpretation of assay results is intended to be performed by qualified healthcare professionals such as clinical cytogeneticists or molecular geneticists; and,
(iv) A description of the performance studies performed in accordance with paragraph (b)(1)(ii) of this section and a summary of the results.