(94 days)
The Medtronic Transportation Cassettes are intended for use in healthcare facilities to organize, enclose, sterlize, transport, and store medical devices and other instrumentation between surgical uses. The Medtronic Transportation/Sterilization Cassettes are not intent own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.
The Medtronic Transportation/Sterilization Cassettes include cases, trays, lids, caddies, modules, or brackets fabricated from a variety of materials commonly used to enclose, protect, and organize Medtronic orthopedic or neurological non-sterile devices, which meet national or international standards. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. The Medtronic Transportation / Sterilization Cassettes design consists of multiple components designed to be integrated into a single unit, which protects the interior components during transportation, sterilization, and storage. These components consist of: an outer base, internal individual trays with or without lids, caddies and an outer locking lid. The components if the Medtronic Transportation/Sterilization Cassettes are fabricated from intrinsically stable metals and thermoplastic polymers. All of the components of the Medtronic Transportation/Sterilization Cassettes are perforated with an evenly distributed hole pattern, and are designed to be used for steam sterilization. Since the Medtronic Transportation/Sterilization Cassettes are perforated, an FDA-cleared wrap must be used for sterilization and to maintain the sterility of the contents. The Medtronic Transportation/Sterilization Cassettes are designed to be used with standard autoclaves used in hospitals and healthcare facilities. As such, the Medtronic Transportation/Sterilization Cassettes are effective for containing devices during sterilization and have been designed to withstand repeated steam sterilization cycles.
I am sorry, but the provided text does not contain the detailed information necessary to complete the requested table and study description. The document is a 510(k) summary for a medical device (Medtronic Transportation/Sterilization Cassettes) that states no changes were made to the device itself, only to the labeling to add manual cleaning instructions.
Therefore, the document explicitly states that the original biocompatibility and verification/validation data still applies, and it does not provide new, detailed acceptance criteria and performance data for the device's primary function of sterilization and transport. It mentions a cleaning validation study was performed for the new manual cleaning instructions, but provides no details on its methodology, sample size, or specific acceptance criteria and results beyond stating it was "effective."
Without specific acceptance criteria and reported device performance (e.g., sterilization efficacy rates, barrier properties, mechanical durability under specific tests), and details on the study design (sample sizes, ground truth establishment, expert qualifications, adjudication methods, etc.), I cannot fulfill your request.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, with flowing lines above them, possibly representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 23, 2017
Medtronic Sofamor Danek, USA, Inc. Julie Bassett Regulatory Affairs Program Manager 1800 Pyramid Place Memphis. Tennessee 38132
Re: K163279
Trade/Device Name: Medtronic Transportation/Sterilization Cassettes Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 18, 2016 Received: November 21, 2016
Dear Julie Bassett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K163279
Device Name
Medtronic Transportation/Sterilization Cassettes
Indications for Use (Describe)
The Medtronic Transportation Cassettes are intended for use in healthcare facilities to organize, enclose, sterlize, transport, and store medical devices and other instrumentation between surgical uses. The Medtronic Transportation/Sterilization Cassettes are not intent own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.
Sterilization validations for the worst case Medtronic Transportation Cassettes (22.75 x 11.26 x 5.5 inches) included implants and common surgical instruments such as rasps, drivers, trials, handles, inserters, probes, drills, etc. The validated total weight was 28.4lbs. The validated worst case loading configurations of the Medtronic Transportation/Sterlization Cassettes included the following worst case lumen dimensions:
■ 363 x 1.575mm
■ 247.5 x 4.1mm
Sterilization parameters
| Cycle | Temperature | Exposure time | Minimum dry time |
|---|---|---|---|
| Gravity displacement | 250°F (121°C) | 30 Minutes | 30 Minutes |
| Gravity displacement | 270°F (132°C) | 15 Minutes | 30 Minutes |
| Gravity displacement | 275°F (135°C) | 10 Minutes | 30 Minutes |
| Dynamic-air-removal(4 pre-conditioning pulses) | 270°F (132°C) | 4 Minutes | 30 Minutes |
| Dynamic-air-removal(4 pre-conditioning pulses) | 275°F (135°C) | 3 Minutes | 30 Minutes |
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
Medtronic Sofamor Danek Medtronic Transportation/Sterilization Cassettes January 2017
| I. | Company: | Medtronic Sofamor Danek1800 Pyramid PlaceMemphis, Tennessee 38132(800) 876-3133 | |
|---|---|---|---|
| Contact: | Julie BassettRegulatory Affairs Program ManagerDirect Telephone: (901) 399-3248Fax: (901) 346-9738 | ||
| Date Prepared: | January 25, 2017 | ||
| II. | Device: | ||
| Name of Device: | Medtronic Transportation/Sterilization Cassettes | ||
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes &Other Accessories (21 CFR 880.6850) | ||
| Classification: | Class II | ||
| Product Code: | KCT | ||
| Predicate Device: | K152241 – MedtronicTransportation/Sterilization Cassettes(SE 1/20/2016)The predicate has not been subject of a designrelated recall |
{4}------------------------------------------------
Description: III.
The Medtronic Transportation/Sterilization Cassettes include cases, trays, lids, caddies, modules, or brackets fabricated from a variety of materials commonly used to enclose, protect, and organize Medtronic orthopedic or neurological non-sterile devices, which meet national or international standards. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. The Medtronic Transportation / Sterilization Cassettes design consists of multiple components designed to be integrated into a single unit, which protects the interior components during transportation, sterilization, and storage. These components consist of: an outer base, internal individual trays with or without lids, caddies and an outer locking lid. The components if the Medtronic Transportation/Sterilization Cassettes are fabricated from intrinsically stable metals and thermoplastic polymers. All of the components of the Medtronic Transportation/Sterilization Cassettes are perforated with an evenly distributed hole pattern, and are designed to be used for steam sterilization. Since the Medtronic Transportation/Sterilization Cassettes are perforated, an FDA-cleared wrap must be used for sterilization and to maintain the sterility of the contents. The Medtronic Transportation/Sterilization Cassettes are designed to be used with standard autoclaves used in hospitals and healthcare facilities. As such, the Medtronic Transportation/Sterilization Cassettes are effective for containing devices during sterilization and have been designed to withstand repeated steam sterilization cycles.
No changes are being made to the Medtronic Transportation/Sterilization Cassettes from the cassettes cleared in K152241, SE 1/20/2016. The purpose of this 510(k) is to update the labeling, specifically the information for use (IFU). The IFU is being updated to add manual cleaning instructions and to centralize similar information within the IFU and avoid potential confusion.
{5}------------------------------------------------
Device List:
1850060 Case - Triple Generic Outer Base (22.74 x 11.260 x 5.040 inches) 1850064 Lid - Generic Outer Lid (22.75 x 11.260 x 0.470 inches) 7022101L Tray Lid (21 x 10.13 x 0.075 inches) P1850061 Tray 1 (20.75 x 9.79 x 1.32 inches) P1850062 Tray 2 (21 x 10.13 x 1.69 inches) P1850063 Tray 3 (21 x 10.13 x 1.38 inches) 7059532 Large Caddy (9.47 x 6.37 x 1.3 inches) 7059532L Large Lid (5.85 x 4.725 x 0.095 inches) P9213018 Small Caddy (2 x 1.5 x 1.025 inches) P9213018 Small Lid (2 x 1.29 x 0.095 inches)
IV. Indications For Use:
The Medtronic Transportation/Sterilization Cassettes are intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.
Sterilization validations for the worst case Medtronic Transportation/Sterilization Cassette (22.75 x 11.26 x 5.5 inches) included implants and common surgical instruments such as rasps, drivers, trials, handles, inserters, probes, drills, etc. The validated total weight was 28.4lbs. The validated worst case loading configurations of the Medtronic Transportation/Sterilization Cassette included the following worst case lumen dimensions:
- 363 x 1.575mm - 247.5 x 4.1mm
{6}------------------------------------------------
| Sterilization parameters | |||
|---|---|---|---|
| Cycle | Temperature | Exposure time | Minimum dry time |
| Gravity displacement | 250°F (121°C) | 30 Minutes | 30 Minutes |
| Gravity displacement | 270°F (132°C) | 15 Minutes | 30 Minutes |
| Gravity displacement | 275°F (135°C) | 10 Minutes | 30 Minutes |
| Dynamic-air-removal(4 pre-conditioning pulses) | 270°F (132°C) | 4 Minutes | 30 Minutes |
| Dynamic-air-removal(4 pre-conditioning pulses) | 275°F (135°C) | 3 Minutes | 30 Minutes |
Comparison of Technological Characteristics: V.
Since there have been no changes to the Medtronic Transportation/Sterilization Cassettes, the technological characteristics for the cassettes remain unchanged from those characteristics cleared in K152241, SE 1/20/2016. In addition, modifications to the device labeling do not impact the technical characteristics of the Medtronic Transportation/Sterilization Cassettes. As shown in Table 1: Substantial Equivalence - Comparison of Technological Characteristics for the Predicate and Subject Medtronic Transportation/Sterilization Cassettes and Table 2: Substantial Equivalence - Comparison of Performance Data for the Predicate and Subject Medtronic Transportation/Sterilization Cassettes., the Medtronic Transportation/Sterilization Cassettes in this 510(k) submission are identical to the Medtronic Transportation / Sterilization Cassettes cleared in K152241. SE 1/20/2016.
Table 1: Substantial Equivalence – Comparison of Technological Characteristics for the Predicate and Subject Medtronic Transportation/Sterilization Cassettes
| Feature | Predicate Device | Subject Device |
|---|---|---|
| Trade Name | Medtronic Transportation/Sterilization Cassettes | Identical |
| Fundamental Scientific Technology | Sterilization Cassette | Identical |
{7}------------------------------------------------
| Feature | Predicate Device | Subject Device |
|---|---|---|
| Intended Use | The MedtronicTransportation/Sterilization Cassettesare intended for use in healthcarefacilities to organize, enclose, sterilize,transport, and store medical devices andother instrumentation between surgicaland other medical uses. The MedtronicTransportation/Sterilization Cassettesare not intended on their own tomaintain sterility; it is intended to beused in conjunction with a legallymarketed, validated, FDA-clearedsterilization wrap. | Identical |
| Product Code | KCT | Identical |
| MaterialComposition | Thermoplastic polymers, aluminum,and stainless steel | Identical |
| Design | A base, a lid with a locking latch,and individual inserts | Identical |
| Dimensions | The greatest challenge dimensionwas assessed to be:22.75 x 11.26 x 5.51 inchesThe inserts are offered in differentsizes | Identical |
| Configuration | Perforated bases, lids, and inserts | Identical |
| Air Permeance | Yes | Identical |
| Percent Perforation | Evenly distributed hole pattern. | Identical |
| SterilizationMethod | Pre-Vacuum andGravity Displacement | Identical |
| Reusable | Yes | Identical |
Table 2: Substantial Equivalence – Comparison of Performance Data for the Predicate and
Subject Medtronic Transportation/Sterilization Cassettes
| Feature | Predicate Device | Subject Device |
|---|---|---|
| Sterilant Penetration | Yes | Identical |
| Microbial Barrier Properties | To be used with an FDAapproved sterilization wrap | Identical |
| Material Compatibility withSterilization Method | Materials are compatiblewith sterilization method | Identical |
| Toxicological Properties | Materials arebiocompatible | Identical |
{8}------------------------------------------------
Performance Data: VI.
Biocompatibility:
No changes are being made to the Medtronic Transportation/Sterilization Cassettes – only the labeling is being updated; therefore, the original biocompatibility data still applies.
{9}------------------------------------------------
Verification/Validation Testing:
No changes are being made to the Medtronic Transportation/Sterilization Cassettes only the labeling is being updated; therefore, the original verification/validation data still applies.
However, due to the addition of the manual cleaning instructions for the Medtronic Transportation/Sterilization Cassettes, a cleaning validation study was performed. The cleaning validation study demonstrated that the manual cleaning parameters were effective and will adequately clean Medtronic Transportation/Sterilization Cassettes.
Conclusion: VII.
In conclusion, the subject Medtronic Transportation/Sterilization Cassettes are substantially equivalent to predicate device Medtronic Transportation/Sterilization Cassettes cleared in K152241. Based on the intended use, technological characteristics, and performance data, the subject Medtronic Transportation / Sterilization Cassettes are substantially equivalent and are as safe and as effective as the legally marketed predicate device.
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).