(151 days)
Not Found
No
The document explicitly states "Not Found" for mentions of AI, DNN, or ML, and the core technology described is Pulse Decomposition Analysis (PDA), a scientific method, not an AI/ML algorithm.
No.
The device is described as a "cardiovascular monitoring device" that measures vital signs but does not provide any treatment or therapy.
No
The device measures physiological parameters (blood pressure, heart rate, respiration rate) but does not provide a diagnosis or aid in diagnosing a disease or condition. It is a monitoring device.
No
The device description explicitly states it is a "cardiovascular monitoring device that non-invasively measures continuous blood pressure, heart rate and respiration rate via a finger cuff". This indicates the presence of hardware components (the finger cuff and the monitoring device itself) in addition to any software used for data processing and display.
Based on the provided information, the CareTaker device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs analyze samples taken from the body (in vitro - "in glass"). This includes things like blood, urine, tissue, etc.
- The CareTaker device measures physiological parameters on the body (in vivo - "in the living"). It uses a finger cuff to non-invasively measure blood pressure, heart rate, and respiration rate directly from the patient.
The description clearly states it's a "cardiovascular monitoring device that non-invasively measures continuous blood pressure, heart rate and respiration rate via a finger cuff". This is a direct measurement from the patient's body, not an analysis of a sample taken from the body.
N/A
Intended Use / Indications for Use
CareTaker is intended to noninvasively and continuously measure a patient's blood pressure ("BP"), heart rate ("HR"), ould respiration rate ("RSP"), which are derived from the pulse pressure waveform using the scientific method of Pulse Decomposition Analysis ("PDA") for use on adult patients at rest. CareTaker is calibrated using a manual method or any A MI 81060 compliant BP device, or is automatically calibration mode. All parameters derived by Care Taker are reported to an integrated LCD screen and optionally to a remote display monitor (RDDS) via standard radio transmission protocols. The device is intended for use by clinicians or other properly trained medical personnel and does not provide any physiological alarm functions.
Product codes
DXN
Device Description
CareTaker is a cardiovascular monitoring device that non-invasively measures continuous blood pressure, heart rate and respiration rate via a finger cuff based on the scientific method of Pulse Decomposition Analysis ("PDA"). CareTaker is calibrated using any manual or AAMI 81060 compliant blood pressure device or is automatically calibrated using its self-calibration mode.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
finger
Indicated Patient Age Range
adult patients
Intended User / Care Setting
clinicians or other properly trained medical personnel
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
CareTaker (K151499), Nexfin HD (K072049), HealthPatch MD (K152139)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
April 21, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
CareTaker Medical, LLC Jeff Pompeo President & CEO 3042 Berkmar Drive, Suite A Charlottesville, Virginia 22901
Re: K163255
Trade/Device Name: CareTaker4 Physiological Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 15, 2017 Received: March 16, 2017
Dear Jeff Pompeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
1
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mada Jellm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K163255
Device Name
CareTaker4 Physiological Monitor
Indications for Use (Describe)
CareTaker is intended to noninvasively and continuously measure a patient's blood pressure ("BP"), heart rate ("HR"), ould respiration rate ("RSP"), which are derived from the pulse pressure waveform using the scientific method of Pulse Decomposition Analysis ("PDA") for use on adult patients at rest. CareTaker is calibrated using a manual method or any A MI 81060 compliant BP device, or is automatically calibration mode. All parameters derived by Care Taker are reported to an integrated LCD screen and optionally to a remote display monitor (RDDS) via standard radio transmission protocols. The device is intended for use by clinicians or other properly trained medical personnel and does not provide any physiological alarm functions.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the The to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
3
510(k) Summary
510(k) Summary of Safety and Effectiveness
Preparation Date: November 9, 2016 1)
2) Submitted by:
CareTaker Medical, LLC 3042 Berkmar Drive, Suite A Charlottesville, Virginia 22901-1455 User Fee Organization Number 397095 Owner/Operator #: TBD
Contact Person/Prepared by: 3)
Jeff Pompeo President & CEO CareTaker Medical 3042 Berkmar Drive, Suite A Charlottesville, Virginia 22901-1455 Phone: 434-409-1945 Email: Jeff@caretakermedical.net
4) Device Identification:
Trade Name: CareTaker Wireless Vital Signs Monitor Common Name: CareTaker4 Physiological Monitor Classification: 21 CFR 870.1130, Product Code: DXN -System, Measurement, Blood-Pressure, Non-Invasive Device Class: II
- Predicate Devices: CareTaker (K151499), Nexfin HD (K072049), HealthPatch MD 5) (K152139)
- Device Description: CareTaker is a cardiovascular monitoring device that non-6) invasively measures continuous blood pressure, heart rate and respiration rate via a finger cuff based on the scientific method of Pulse Decomposition Analysis ("PDA"). CareTaker is calibrated using any manual or AAMI 81060 compliant blood pressure device or is automatically calibrated using its self-calibration mode.
Intended Use: CareTaker is intended to noninvasively and continuously measure a 7) patient's blood pressure ("BP"), heart rate ("HR") and respiration rate ("RSP"), which are derived from the pulse pressure waveform using the scientific method of Pulse Decomposition Analysis ("PDA") for use on adult patients at rest. CareTaker is calibrated
4
using a manual method or any AAMI 81060 compliant BP device, or is automatically calibrated using its self-calibration mode. All parameters derived by CareTaker are reported to an integrated LCD screen and optionally to a remote display monitor ("RDDS") via standard radio transmission protocols. The device is intended for use by clinicians or other properly trained medical personnel and does not provide any physiological alarm functions.
8) Comparison to Predicates:
Provision of continuous non-invasive blood pressure and heart rate
Both the predicate CareTaker and the present CareTaker system use the Pulse Decomposition Analysis method for the tracking of blood pressure. Both devices measure blood pressure within AAMI SP-10 guidelines. Both devices require calibration and are intended for use on adult patients in clinical settings administered by trained medical staff. Both systems are identical with regard to electronic circuitry, pneumatic layout and components and algorithmic pulse analysis. While the CareTaker4 has an onboard display, the predicate CareTaker does not. While the CareTaker4 is capable of self-calibration, the predicate CareTaker is not. While the CareTaker4 is capable of determining respiration rates, the predicate CareTaker is not. Both systems operate at a coupling pressure significantly lower than normal diastole with a lower risk of occluding blood flow to the monitored digit, making the technology is safer.
Provision of calibrated blood pressure readings
The CareTaker system and the predicate Nexfin_HD device provide calibrated blood pressure readings. While the Nexfin HD uses the Penaz/Physiocal method to obtain blood pressure calibration, the CareTaker uses the principle of oscillometry. Both devices measure blood pressure within AAMI SP-10 guidelines. Both devices obtain the required arterial pulse pressure signal from a digit. However, because they use different technologies for the determination of blood pressure, their coupling to the monitored digit is very different. Since the CareTaker system operates at a significantly lower coupling pressure with a lower risk of occluding blood flow to the monitored digit, the technology is safer. The ability of the CareTaker system to operate with lower, constant coupling pressure also has significant consequences regarding portability and versatility of the technology compared to that of the predicate Nexfin device.
Provision of respiration rate readings
With regard to respiration rate detection, both the CareTaker4 and the predicate HealthPatch MD analyze a pulsatile, heartbeat-related signal. In the case of the CareTaker4 it is the derivative arterial pressure pulse, in the case of the respiration detection predicate device it is an EKG signal. In both systems signal amplitude and inter-beat interval-related changes are used to detect and determine the respiration rate.
5
9) Conclusions:
Provision of continuous non-invasive blood pressure and heart rate
The CareTaker4 device is substantially equivalent to the predicate CareTaker device in regard to providing continuous blood pressure and heartrate readings.
Provision of calibrated blood pressure readings
The CareTaker4 device is substantially equivalent to the NexFin HD in terms of providing calibrated blood pressure readings from a finger cuff, although CareTaker4 uses a different methodology (PDA). Like the Nexfin HD, CareTaker4 has clinically validated evidence attesting adherence to SP10 standards and followed quality management system rigor to comply with IEC 60601 safety standards.
Provision of respiration rate readings
The CareTaker4 device is substantially equivalent to the HealthPatch MD in terms of providing a respiration rate derived from a pulse sensor, although CareTaker4 uses a differentiated arterial pulse signal while the predicate device uses an EKG signal. The CareTaker4 has clinically validated evidence attesting performance equivalent to or better than the predicate device.