K Number
K163090

Validate with FDA (Live)

Date Cleared
2017-05-23

(200 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:

· Transurethral resection of prostate (TURP) for benign prostatic hyperplasia

  • · Transurethral incision of the prostate (TUIP) or bladder neck
  • · Transurethral resection of bladder tumors (TURBT)
  • · Cystodiathermy

· Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)

Device Description

The Plasma Edge system is a manual surgical device, consisting of a single-use electrode with cable range, an active and passive working element reusable and an adaptator to an HF generator compatible.

The electrodes consist of an active tip, two wires threaded through ceramic tubes to connect the active tip to the body of the loop, allowing the HF energy to reach the active tip.

It has to be used with continuous flow irrigation of saline solution (NaCl 0,9%) that reaches the operative site through a resectoscope. The HF energy delivered from the generator to the electrode ionizes the gas of the saline solution, creating a plasma for the cutting, coagulation and vaporization of tissues.

AI/ML Overview

The provided text describes the Plasma EDGE System and its performance testing for a 510(k) premarket notification. However, it does not contain specific acceptance criteria with numerical targets or detailed results that demonstrate the device meets those criteria from a clinical study. Instead, it focuses on demonstrating substantial equivalence to predicate devices through various non-clinical tests and a comparison of technological characteristics.

Therefore, many of the requested elements for a study proving device performance against acceptance criteria cannot be fully extracted.

Here's an analysis of the provided text in relation to your request:


1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, accuracy, or other quantitative measures typically found in clinical performance studies of AI/diagnostic devices. The "Performance testing" section instead lists various types of engineering and non-clinical evaluations that were conducted.

If we interpret "acceptance criteria" broadly to mean successful completion of required testing, the table would look like this (though actual numerical results for "reported device performance" are missing for most items):

Acceptance Criteria (Implied)Reported Device Performance
Biocompatibility in accordance with ISO-10993 standardsTesting for Cytotoxicity, Chemical Analysis, Biological Safety conducted and passed.
Electrical safety according to AAMI/ANSI ES 60601-1Tested by an independent laboratory and passed.
Electromagnetic compatibility (EMC) according to IEC 60601-1-2Tested by an independent laboratory and passed.
Cleaning and sterilization validation for reusable componentsValidated by an independent laboratory (Steam Sterilization per AAMI TIR N°12, ISO 17664, ISO 17665; Cleaning/Disinfection per AAMI TIR N°30).
Shelf life validation (3 years for electrodes)Validated (sterilization method: ethylene oxide).
Bench top validation testing (End of life, breakdown, compatibility, ex vivo tissues)Tested to demonstrate product safety and efficiency (details of "efficiency" missing).
Risk analysis according to ISO-14971Carried out in accordance with established in-house acceptance criteria based on ISO-14971:2007.

The document states, "Comparative performance testing demonstrate that the device performed as well as, or better than, the predicate device," but does not provide the specific comparative results or metrics.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily describes non-clinical performance and benchtop testing. It does not mention a "test set" in the context of clinical data, nor does it specify sample sizes for any clinical or human-in-the-loop evaluation, or data provenance. The "ex vivo tissues" mentioned in benchtop testing do not constitute a clinical "test set."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The document does not describe a clinical study involving human judgment or interpretation that would require expert-established ground truth. The evaluations are technical and benchtop in nature.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as there is no clinical "test set" or human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an electrosurgical unit and accessories, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to AI assistance for human readers would not be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical device, not an algorithm. Its performance is inherent in its physical and electrical characteristics when used by a surgeon. The device itself operates "standalone" in the sense that its electrosurgical function is performed directly, but it is always "human-in-the-loop" as it requires a surgeon for operation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. The "ground truth" for the tests described are engineering specifications, standardized test methods (e.g., ISO, AAMI, IEC standards), and physical/biological properties (e.g., cytotoxicity, electrical safety, sterilization efficacy).

8. The sample size for the training set

Not applicable. The Plasma EDGE System is a hardware device; it does not involve machine learning or AI models that require a "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 23, 2017

Chirurgie Innovation % Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313

Re: K163090 Trade/Device Name: Plasma EDGE System Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FAS, GEI, FDC Dated: May 15, 2017 Received: May 16, 2017

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Charles Viviano -S

For

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163090

Device Name

Plasma EDGE System

Indications for Use (Describe)

The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:

· Transurethral resection of prostate (TURP) for benign prostatic hyperplasia

  • · Transurethral incision of the prostate (TUIP) or bladder neck
  • · Transurethral resection of bladder tumors (TURBT)
  • · Cystodiathermy

· Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

{3}------------------------------------------------

K163090 Page 1 of 7

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge SystemTraditional 510(k) application
Rev. 4Page 1 / 7

SECTION 5:

510(K) SUMMARY

SubmitterChirurgie Innovation27 Place Victor SchoelcherF-91300 Massy FRANCE
ContactsGuillaume NouryCEOregulatory@orange.frPhone : +33 6 16 16 63 30
Date07/29/2016
Trade NamePlasma EDGE System
Common nameBipolar electrodes
Classification NameElectrosurgical cutting and coagulation device and accessoriesResectoscope, working element
ClassII
Product CodeClassification product code : FASSubsequent product code : GEI / FDC
CFR section876.4300 / 878.4400 / 876.1500
Device panelRegulation Medical Specialty : Gastroenterology / Urology510k Review Panel : General & Plastic Surgery
Legally marketedpredicate devicesK994166: Axipolar Resectoscope Electrode - manufactured byGyrus Medical Ltd.K120567: Gyrus ACMI PK® Button Electrode - manufactured byGyrus Medical Ltd.

Product Description

The Plasma Edge system is a manual surgical device, consisting of a single-use electrode with cable range, an active and passive working element reusable and an adaptator to an HF generator compatible.

The electrodes consist of an active tip, two wires threaded through ceramic tubes to connect the active tip to the body of the loop, allowing the HF energy to reach the active tip.

{4}------------------------------------------------

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge System
Traditional 510(k) application
Rev. 4Page 2 / 7

It has to be used with continuous flow irrigation of saline solution (NaCl 0,9%) that reaches the operative site through a resectoscope. The HF energy delivered from the generator to the electrode ionizes the gas of the saline solution, creating a plasma for the cutting, coagulation and vaporization of tissues.

Electrodes range:

The list of electrodes is divided in 2 ranges: SIDE LOAD and FRONT LOAD.

Each range got is specific way to be assembly into the working element. The only physical difference between both ranges is the electrode cable connexion.

End user can be used to both versions so we decided to offer the choice between them.

Working element range:

The working elements are compatible with the 3 main existing optics brands: STORZ, ACMI and OLYMPUS.

The working elements will be divided in 2 families: side load connection and front load connection.

Adaptor:

This will allow to connect the Plasma EDGE electrode range to a bipolar HF generator system

Indications for use

The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:

  • Transurethral resection of prostate (TURP) for benign prostatic hyperplasia
  • Transurethral incision of the prostate (TUIP) or bladder neck
  • Transurethral resection of bladder tumors (TURBT)

{5}------------------------------------------------

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge SystemTraditional 510(k) application
Rev. 4Page 3 / 7
  • Cystodiathermy
    • Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)

Performance testing

Risk analysis:

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO-14971:2007

Performance tests were carried out to ensure that the system functions as intended and meets design specifications. The following performance tests and usability studies were conducted:

Biocompatibility testing

The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, `Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995. The testing included the following tests: • Biological Safety – toxicology: AAMI ANSI ISO 10993-1:2009; Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process · Cytotoxicity: AAMI ANSI ISO 10993-5:2009; Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity • Chemical Analysis: AAMI ANSI ISO 10993-12:2012; Biological evaluation of medical devices - Part 12: Sample Preparation and Reference Materials

Electrical safety

Electrical Safety was tested by an independent laboratory according to AAMI/ANSI ES 60601-1:2005 and C1:2009 and A2:2010; Medical Electrical Equipment - Part 1.1 General requirements for safety and essential performance.

Electromagnetic compatibility (EMC)

{6}------------------------------------------------

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge System
Traditional 510(k) application
Rev. 4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              Rev. 4Page 4 / 7

Electromagnetic compatibility with the existing HF generator was tested by an independent laboratory according to IEC 60601-1 -2 : 2007

Cleaning and sterilization validation for the reusable working element range:

Steam sterilization validation was tested by an independent laboratory following AAMI TIR N°12, ISO 17664 and ISO 17665

Cleaning and disinfection validation was tested by an independent laboratory following AAMI TIR Nº30

Bench top validation testing

End of life simulation report, breakdown simulation report, working element compatibility report and test on ex vivo tissues have been tested on the Plasma Edge System to demonstrate the product safety and the efficiency.

Summary of Sterilization and Shelf Life Discussion

The electrodes are delivered in a sterile state and are intended for single patient use only. The validated sterilization method used is ethylene oxide. The product has a shelf life of three (3) years.

List of standards

StandardsDescription
ISO 10993-1Biological evaluation of medical devices - Part 1: Evaluation and testingwithin a risk management process
ISO 10993-5Biological evaluation of medical devices -- Part 5: Tests for In Vitrocytotoxicity
ISO 10993-7Biological evaluation of medical devices - part 7: ethylene oxidesterilization residuals [including: technical corrigendum 1 (2009)]
ISO 10993-10Biological evaluation of medical devices - Part 10: Tests for irritationand skin sensitization

{7}------------------------------------------------

K163090
Page 5 of 7

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge System
Traditional 510(k) application
Rev. 4Page 5 / 7
StandardsDescription
ISO 10993-11Biological evaluation of medical devices - Part 11: Tests for systemictoxicity
ISO 11135Sterilization of health-care products - ethylene oxide - requirements forthe development, validation and routine control of a sterilization processfor medical devices
ISO 11607-1Packaging for terminally sterilized medical devices - Part 1:Requirements for materials, sterile barrier systems and packagingsystems
ISO 11607-2Packaging for terminally sterilized medical devices - Part 2: Validationrequirements for forming, sealing and assembly processes
ISO 17664ISO 17665-1AAMI TIRN°12Sterilization of medical devices - Information to be provided by themanufacturer for the processing of resterilizable medical devicesSterilization of health care products -- Moist heat -- Part 1:Requirements for the development, validation and routine control of asterilization process for medical devices
AAMI TIRN°30Cleaning and disinfection validation: following FDA guidance"Reprocessing medical device in health care settings: validationmethods and labeling guidance for industry and food administrationstaff issued March 17, 2015"
IEC 60601-1Medical electrical equipment - Part 1: General requirements for basicsafety and essential performance
IEC 60601-1-2General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances- Requirements and tests
IEC 60601-1-6Medical electrical equipment - Part 1-6: General requirements for basicsafety and essential performance - Collateral standard: Usability
IEC 60601-2-2Medical electrical equipment - Part 2-2: Particular requirements for thebasic safety and essential performance of high frequency surgicalequipment and high frequency surgical accessories

{8}------------------------------------------------

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge SystemTraditional 510(k) application
Rev. 4Page 6 / 7
StandardsDescription
IEC 60601-2-18Medical electrical equipment - Part 2-18: Particular requirements for thebasic safety and essential performance of endoscopic equipment
IEC 62366Medical devices – Application of usability engineering to medical devices
ISO 14971Medical devices - application of risk management to medical devices.
ISO 15223-1Medical devices - symbols to be used with medical device labels,labeling, and information to be supplied - part 1 : general requirements

Comparison of Technological characteristics:

ELECTRODEProposed devicePredicate devicePredicate device
Plasma Edge (cutting loop electrode / vaporization electrode)AxipolarResectoscopeElectrode(K994166)Gyrus ACMI PK®Button Electrode(K120567)
Indications for use- Transurethral resection of prostate (TURP)- Transurethral incision of the prostate (TUIP) or bladder neck- Transurethral resection of bladder tumours (TURBT)- Cystodiathermy-Transurethral electrovaporization of the prostate- Transurethral resection of prostate (TURP)- Transurethral incision of the prostate (TUIP) or bladder neck- Transurethral resection of bladder tumours (TURBT)- Cystodiathermy-- Transurethral electrovaporization of the prostate
SterilizationEtoEtoEto

{9}------------------------------------------------

K163090 Page 7 of 7

Section 5: 510(k) Summary
ApplicantChirurgie Innovation
Device :Plasma Edge SystemTraditional 510(k) application
Rev. 4Page 7 / 7
ELECTRODEProposed devicePredicate devicePredicate device
Plasma Edge (cutting loop electrode / vaporization electrode)AxipolarResectoscopeElectrode(K994166)Gyrus ACMI PK®Button Electrode(K120567)
DisposableYesYes
Energy typeHigh frequencyHigh frequency
ModeBipolarBipolar
User interfaceFootswitchFootswitch
Use only inConductiveMediaThe electrode is to be activated only when immersed in a conductive media such as standard saline solutionThe electrode is to be activated only when immersed in a conductive media such as standard saline solution
ElectrodemanipulationWorking elementWorking element
WorkingelementcompatibilityStorzOlympusACMIStorzOlympusACMI
Sterilization ofworkingelementAutoclaveAutoclave

Conclusion:

There is no difference between the Plasma EDGE resection system and the predicate devices in terms of intended use, principle of operation, and the technology used for device performance. In other words, there is no difference technically, clinically and biologically.

The Plasma EDGE resection system was subjected to verification testing to confirm device performance. There is no new technology and no difference that would raise new or different questions of safety or efficiency. Comparative performance testing demonstrate that the device performed as well as, or better than, the predicate device.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).