K Number
K162884
Manufacturer
Date Cleared
2017-01-12

(90 days)

Product Code
Regulation Number
882.4300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DSG™ Zavation Screw System is indicated for use with the Zavation Spinal System during pedicle screw insertion to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the pedicle screw and may indicate contact of the tip with soft tissues and possible vertebral cortex perforation. The DSG™ Zavation Screw System is indicated for use in both open and percutaneous (MIS) surgical approaches to the spine, with options of direct insertion of the screw in bone or after a step of preparation of the pilot hole with sensor equipped instruments.

Device Description

The DSG™ Zavation Screw System is a modification to the cleared DSG™ Threaded Drill System and consists of the DSG™ Electronic T-handle, Ratcheting Handle, DSG™ Pin (active stylet), and the previously cleared Zavation Spinal System (K153404). These components are purchased and shipped as a complete system from Zavation, with the DSG™ Threaded Drill System components and Zavation Spinal System components individually packaged. The complete system is provided with the modified instructions for use of the DSG™ Zavation Screw System.

All of the patient-contacting materials are categorized per FDA's guidance on ISO 10993-1 as externally communicating materials that are in contact with the body for a limited duration, and are unchanged from the prior clearance. Certain components of the device are single-use while others are re-usable; certain components are provided sterile while others are sterilized by the end user.

The device is intended for use by surgeons in a professional healthcare environment, and utilizes sensing technology to detect the impediately surrounding tissues while inserting pedicle screws either through a previously drilled pilot hole or directly into bone. The surgeon can either drill and/or tap the screw hole prior to inserting the pedicle screw, or can use the system to directly insert the screw into the bone without a pilot hole. As the screw is manually advanced into the bone, the distal sensor measures the electrical impedance of the immediately surrounding tissues. The device produces real-time visual and audible signals to indicate changes in impedance associated with possible vertebral perforation.

AI/ML Overview

The provided text describes the 510(k) summary for the SpineGuard DSG™ Zavation Screw System. It outlines the device, its intended use, and comparative information with a predicate device. However, it does not detail specific acceptance criteria with numerical targets or a comprehensive study plan with the level of detail requested for AI/device performance.

Based on the information provided, here's what can be extracted and what is missing:

1. A table of acceptance criteria and the reported device performance:

The document describes "performance testing" but does not define explicit acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy, or a specific range of values for mechanical properties). Instead, it states that "All tests were passed, demonstrating equivalent performance according to device specifications and thus supporting substantial equivalence." The tests are:

TestTest Method SummaryReported Device Performance
Cadaver TestingCadaveric usability testing was performed to demonstrate the usability and placement accuracy of the device.Pass
Mechanical TestingMechanical testing was performed to demonstrate the performance and integrity of the system in implanting pedicle screws without a pilot hole.Pass
BiocompatibilityPerformed in accordance with ISO-10993Pass
Sterilization ValidationEtO sterilization cycle designed and validated per NF EN ISO 11737-2Pass
Electrical SafetyPerformed in accordance with IEC 60601-1Pass

Missing Information: Specific quantitative acceptance criteria (e.g., "placement accuracy within X mm" for cadaver testing, or specific thresholds for mechanical integrity). The "Pass" result indicates that the device met internal specifications, but these specifications are not detailed in the provided text.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not specified for any of the performance tests. For "Cadaver Testing," the sample size (number of cadavers, or number of pedicle screws inserted) is not mentioned. For "Mechanical Testing," the number of units tested is also not specified.
  • Data Provenance: Not specified. It's unclear if the cadaver testing was performed in the US or another country. The document notes the sponsor is in France.
  • Retrospective or Prospective: Not explicitly stated, though cadaver testing would typically be considered prospective for the device evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. For cadaver testing, it would likely involve surgeons, but their number and experience are not mentioned.

4. Adjudication method for the test set:

  • Adjudication Method: Not specified. If multiple experts were involved (which is not stated), the method for resolving discrepancies (e.g., 2+1, 3+1, none) is not described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is not an AI/ML-driven diagnostic or image analysis tool for "human readers." It's a surgical guidance system providing real-time feedback (visual and audible alerts) based on impedance measurements during pedicle screw insertion. Therefore, an MRMC study related to human readers improving with AI assistance is not applicable to this device. The "feedback to the surgeon" is a direct function for intraoperative guidance, not a tool for interpreting images or data that human "readers" would then review.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device's function is inherently "human-in-the-loop" as it provides feedback to the surgeon. It's not a standalone diagnostic algorithm. The "performance data" describes the device's accuracy and integrity when used by a human. So, a standalone algorithm performance without human involvement is not applicable in the AI sense. Its "standalone" performance would be about the accuracy of its impedance detection, which is implicitly covered by the "Pass" results in the cadaver and mechanical testing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Cadaver Testing: The "placement accuracy" implies that the true anatomical position and any perforations would be verified (e.g., radiography, CT scan, or direct visual inspection post-dissection), which would serve as the ground truth. However, the specific method for establishing this ground truth is not detailed.
  • Mechanical Testing: Ground truth would be based on engineering specifications and measurements (e.g., force, torque, displacement thresholds).

8. The sample size for the training set:

This device does not appear to be an AI/ML device that requires a distinct "training set" in the context of machine learning. Its operation is based on pre-programmed impedance thresholds for tissue differentiation. Therefore, this question is not applicable in the context of the provided information.

9. How the ground truth for the training set was established:

As there's no mention of a "training set" in the context of an AI/ML algorithm, this question is also not applicable. The device's impedance thresholds would likely be established through prior research and experimentation on tissue types, rather than a "training set" with established ground truth labels in the machine learning sense.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is rendered in black, and the text is also in black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 12, 2017

SpineGuard, S.A. % John Smith, MD, JD Partner Hogan Lovells U.S. LLP 555 Thirteenth Street NW Washington, DC 20004

Re: K162884

Trade/Device Name: SpineGuard DSG Zavation Screw System Regulation Number: 21 CFR 882.4300 Regulation Name: Manual Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBG Dated: October 14, 2016 Received: October 14, 2016

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Hoffmann -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162884

Device Name SpineGuard DSG Zavation Screw System

Indications for Use (Describe)

The DSG™ Zavation Screw System is indicated for use with Zavation pedicle screws during pedicle screw insertion to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the pedicle screw and may indicate contact of the tip with soft tissues and possible vertebral cortex perforation. The DSG™ Zavation Screw System is indicated for use in both open and percutaneous (MIS) surgical approaches to the spine, with options of direct insertion of the screw in bone or after a step of preparation of the pilot hole with sensor equipped instruments.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

SpineGuard, S.A.'s DSG™ Zavation Screw System

Sponsor

SpineGuard, S.A. 10 cours Louis Lumière Vincennes, France 94300 Phone: +33 1 45 18 45 19 Facsimile: +33 1 45 18 45 20 Contact Person: Stephane Bette Date Prepared: December 19, 2016

Name of Device: DSG™ Zavation Screw System

Common or Usual Name: Manual cranial drill and accessories

Classification Name: 21 C.F.R. § 882.4300 (Manual cranial drills, burrs, trephines, and their accessories)

Classification Panel: Neurology

Regulatory Class: Class II

Product Code: HBG

Predicate and Reference Devices

SpineGuard, S.A. DSG™ Threaded Drill System (K152747, Predicate)

Zavation Spinal System (K153404, Reference)

Device Description

The DSG™ Zavation Screw System is a modification to the cleared DSG™ Threaded Drill System and consists of the DSG™ Electronic T-handle, Ratcheting Handle, DSG™ Pin (active stylet), and the previously cleared Zavation Spinal System (K153404). These components are purchased and shipped as a complete system from Zavation, with the DSG™ Threaded Drill System components and Zavation Spinal System components individually packaged. The complete system is provided with the modified instructions for use of the DSG™ Zavation Screw System.

All of the patient-contacting materials are categorized per FDA's guidance on ISO 10993-1 as externally communicating materials that are in contact with the body for a limited duration, and are unchanged from the prior clearance. Certain components of the device are single-use while

{4}------------------------------------------------

others are re-usable; certain components are provided sterile while others are sterilized by the end user.

The device is intended for use by surgeons in a professional healthcare environment, and utilizes sensing technology to detect the impediately surrounding tissues while inserting pedicle screws either through a previously drilled pilot hole or directly into bone. The surgeon can either drill and/or tap the screw hole prior to inserting the pedicle screw, or can use the system to directly insert the screw into the bone without a pilot hole. As the screw is manually advanced into the bone, the distal sensor measures the electrical impedance of the immediately surrounding tissues. The device produces real-time visual and audible signals to indicate changes in impedance associated with possible vertebral perforation.

Intended Use / Indications for Use

The DSG™ Zavation Screw System is indicated for use with the Zavation Spinal System during pedicle screw insertion to provide feedback to the surgeon via visual and audible alerts that indicate a change in impedance at the tip of the pedicle screw and may indicate contact of the tip with soft tissues and possible vertebral cortex perforation. The DSG™ Zavation Screw System is indicated for use in both open and percutaneous (MIS) surgical approaches to the spine, with options of direct insertion of the screw in bone or after a step of preparation of the pilot hole with sensor equipped instruments.

Summary of Technological Characteristics

The subject and predicate devices function based on the same technological principles, including the sensing of tissue impedance in tissues surrounding the targeted treatment area. At a high level, both devices are based on the following same technological elements:

  • Both devices contain the DSG™ Electronic T-handle, Ratcheting Handle, and DSG™ Pin ● (active stylet). These components are unmodified from the predicate clearance (K152747).
    The following technological differences exist between the subject and predicate devices:

  • The subject device does not include the DSG™ Threaded Drill shaft or DSG™ Sleeve ● components, in order to facilitate use with the cleared Zavation Spinal System.

DSG™ Zavation Screw System(subject device)DSG™ Threaded Drill System(K152747)
Intended Use/ Indicationsfor UseThe DSG™ Zavation Screw System isindicated for use with the ZavationSpinal System during pedicle screwinsertion to provide feedback to thesurgeon via visual and audible alertsthat indicate a change in impedanceat the tip of the pedicle screw andmay indicate contact of the tip withsoft tissues and possible vertebralcortex perforation. The DSG™Zavation Screw System is indicatedThe DSG™ Threaded Drill System isintended to be used for the preparationof pedicle screw holes. The DSG™Threaded Drill System is indicated foruse during pedicle screw pilot holedrilling to provide feedback to thesurgeon via visual and audible alerts thatindicate a change in impedance at the tipof the probe and may indicate contact ofthe tip with soft tissues and possiblevertebral cortex perforation. The DSG™
DSG™ Zavation Screw System(subject device)DSG™ Threaded Drill System(K152747)
for use in both open andpercutaneous (MIS) surgicalapproaches to the spine, with optionsof direct insertion of the screw in boneor after a step of preparation of thepilot hole with sensor equippedinstruments.Threaded Drill System is indicated foruse in both open and percutaneous(MIS) surgical approaches to the spine.The DSG™ Threaded Drill System isalso indicated for use with fluoroscopicguidance in percutaneous (MIS) surgicalapproaches to the spine.
The DSG™ Threaded Drill System alsois specifically indicated for use inintraoperative electromyographic("EMG") surveillance to assist in thelocation and evaluation of spinal nervesduring surgery of the spine, byadministration of low voltage electricalenergy to tissues and nerves at theoperative site, and EMG monitoring ofmuscle groups associated with thosenerves.
Handle ShapeT-Handle with RatchetT-Handle with Ratchet
ComponentsDSG™ Electronic T-handle,Ratcheting Handle, and DSG™ Pin(active stylet)DSG™ Electronic T-handle, RatchetingHandle, and DSG™ Pin (active stylet),DSG™ Threaded Drill shaft and DSG™Sleeve
SafetyFeaturesDevice cannot be turned off untilbattery exhausted. Prevents reuse ofdevice.Device cannot be turned off until batteryexhausted. Prevents reuse of device.
Power SourceLithium-Ion BatteryLithium-Ion Battery
SterilitySterile/Non-sterileSterile/Non-sterile
Single Use orReusableSingle-use (DSG™ Electronic T-handle and DSG™ Pin (active stylet))Re-usable (Ratcheting Handle)Single-use (DSG™ Electronic T-handleand DSG™ Pin (active stylet))Re-usable (Ratcheting Handle, DSG™Threaded Drill shaft and DSG™ Sleeve)
Distal ShaftShapeZ-Direct Screw Driver shaft (K153404;reference device) and DSG™ PinStraight (cannulated) and DSG™ Pin
DimensionsDSG™ Pin: 373mm x 1.65mmDSG™ Pin: 373mm x 1.65mm; 292mmx 1.65mmThreaded Drill: 1.7mm inner diameter(hole) cannulated metal shaft withflanges (threads) and cutting edges;Thread outer diameters of 4.0, 5.5 and8.0mm
DSG™ Zavation Screw System ( subject device )DSG™ Threaded Drill System (K152747)
Circuit BoardCapacitors, Resistors and Diodes -Firmware (programmable chip) on circuit boardCapacitors, Resistors and Diodes -Firmware (programmable chip) on circuit board

A table comparing the key features of the subject and predicate devices is provided below.

{5}------------------------------------------------

{6}------------------------------------------------

Performance Data

Performance testing for the subject device consisted of simulated insertion testing and cadaver testing to establish that the DSG™ Zavation Screw System is as safe and as effective as the predicate device to which it is a modification. All tests were passed, demonstrating equivalent performance according to device specifications and thus supporting substantial equivalence.

No animal or clinical tests were performed in support of this premarket notification.

Data supporting the labeling change that is the subject of this 510(k) notice is outlined below:

TestTest Method SummaryResults
Cadaver TestingCadaveric usability testing was performed to demonstrate the usability and placement accuracy of the device.Pass
Mechanical TestingMechanical testing was performed to demonstrate the performance and integrity of the system in implanting pedicle screws without a pilot hole.Pass
BiocompatibilityPerformed in accordance with ISO-10993Pass
Sterilization ValidationEtO sterilization cycle designed and validated per NF EN ISO 11737-2Pass
Electrical SafetyPerformed in accordance with IEC 60601-1Pass

Like its predicate, the SpineGuard DSG™ Zavation Screw System complies with the following recognized standards:

  • ASTM F138-13, Standard specification for wrought 18 chromium-14 nickel-2.5 molybdenum . stainless steel bar and wire for surgical implants (2013).
  • . ASTM F899-12b, Standard specification for wrought stainless steel for surgical instruments (2012).
  • . IEC 60601-1 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance, incl. Amendment 1 (2007).
  • . ISO 10993-5, Biological Evaluation of Medical Devices: Tests for in vitro cytotoxicity (2010).
  • . ISO 10993-7, Biological evaluation of medical devices: Ethylene Oxide sterilization residuals (2008).
  • ISO 10993-10, Biological Evaluation of Medical Devices: Tests for irritation and sensitization ● (2010).
  • . ISO 10993-11, Biological Evaluation of Medical Devices: Tests for systemic toxicity (2009).

{7}------------------------------------------------

  • . ISO 11607-1. Packaging for terminally sterilized medical devices: Requirements for materials. sterile barrier systems and packaging systems (2006).
  • . ISO 11607-2, Packaging for terminally sterilized medical devices: Validation requirements for forming, sealing and assembly processes (2006).
  • . ISO 11737-1, Sterilization of medical devices - Microbiological methods: Determination of a population of microorganisms on products (2006).
  • . ISO 11737-2, Sterilization of medical devices - Microbiological methods: Tests of sterility performed in the definition, validation and maintenance of a sterilization process (2010).
  • . ASTM D4169, Standard practice for performance testing of shipping containers and systems (2014).
  • . ASTM F1980, Standard guide for accelerated aging of sterile medical device packages (2011).
  • ISO 11135-1, Sterilization of health care products Ethylene oxide: Requirements for the . development, validation and routine control of a sterilization process for medical devices (2007).

Conclusions

The DSG™ Zavation Screw System is as safe and effective as the predicate DSG™ Threaded Drill System (K152747). The subject device has the same intended use, and very similar indications for use, technological characteristics and principles of operation, as its predicate device. The minor technological differences between the DSG™ Zavation Screw System and its predicate device raise no new issues of safety or effectiveness. In addition, performance data demonstrate that the DSG™ Zavation Screw System is as safe and effective as the predicate device. Thus, the DSG™ Zavation Screw System is substantially equivalent.

§ 882.4300 Manual cranial drills, burrs, trephines, and their accessories

(a)
Identification. Manual cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments that are used without a power source on a patient's skull.(b)
Classification. Class II (performance standards).