K Number
K162818
Date Cleared
2017-01-09

(95 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nittile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) is a non- sterile disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs (tested with breakthrough times) as per attached table.

Device Description

Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Druzs (Blue)

AI/ML Overview

This document is a 510(k) premarket notification for a medical glove, not an AI/ML device. Therefore, the requested information regarding acceptance criteria, study details, ground truth, and sample sizes for AI/ML performance is not applicable.

The document discusses the device's substantial equivalence to a predicate device and its indications for use, specifically for protecting against contamination and certain chemotherapy drugs.

Here's the relevant information provided in the document:

1. A table of acceptance criteria and the reported device performance

The document does not specify "acceptance criteria" in the context of device performance metrics often associated with AI/ML. Instead, it provides breakthrough times for the glove's resistance to various chemotherapy drugs. This can be interpreted as the device's performance regarding chemical impermeability.

Chemotherapy Drug (Concentration)Reported Breakthrough Time
Carmustine (BCNU) (3.3 mg/ml)10.1 minutes
Cisplatin (1.0 mg/ml)> 240 minutes
Cyclophosphamide (Cytoxan) (20> 240 minutes
mg/ml)
Dacarbazine (DTIC) (10.0 mg/ml)> 240 minutes
Doxorubicin Hydrochloride (2.0> 240 minutes
mg/ml)
Etoposide (Toposar) (20.0 mg/ml)> 240 minutes
Fluorouracil (50.0 mg/ml)> 240 minutes
Methotrexate (25 mg/ml)> 240 minutes
Mitomycin C (0.5 mg/ml)> 240 minutes
Paclitaxel (Taxol) (6.0 mg/ml)> 240 minutes
Thiotepa (10.0 mg/ml)20.3 minutes
Vincristine Sulfate (1.0 mg/ml)> 240 minutes

2-9. Other Requested Information (Sample size, data provenance, experts, adjudication, MRMC, standalone performance, ground truth type, training set details)

This information is not applicable to this document as it pertains to a traditional medical device (examination glove) and not an AI/ML powered device. The document does not describe any studies involving algorithms, data sets, human readers, or expert consensus for AI/ML performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 9, 2017

Hartalega Sdn. Bhd. Nurul Kong Ouality Assurance Senior Manager No. 7 Kawasan Perusahaan Suria Bestari Jaya, 45600 MY

Re: K162818

Trade/Device Name: Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZC Dated: December 9, 2016 Received: December 13, 2016

Dear Nurul Kong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K162818

Device Name

Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Druzs (Blue)

Indications for Use (Describe)

The Nittile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) is a non- sterile disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs (tested with breakthrough times) as per attached table.

The list of Chemotherapy Drugs tested (with breakthrough times) is as per below:

Carmustine (BCNU) (3.3 mg/ml)10.1 minutes
Cisplatin (1.0 mg/ml)> 240 minutes
Cyclophosphamide (Cytoxan) (20 mg/ml)> 240 minutes
Dacarbazine (DTIC) (10.0 mg/ml)> 240 minutes
Doxorubicin Hydrochloride (2.0 mg/ml)> 240 minutes
Etoposide (Toposar) (20.0 mg/ml)> 240 minutes
Fluorouracil (50.0 mg/ml)> 240 minutes
Methotrexate (25 mg/ml)> 240 minutes
Mitomycin C (0.5 mg/ml)> 240 minutes
Paclitaxel (Taxol) (6.0 mg/ml)> 240 minutes
Thiotepa (10.0 mg/ml)20.3 minutes
Vincristine Sulfate (1.0 mg/ml)> 240 minutes
  • Please note that the following drugs have extremely low permeation times of 10.1 minutes with Carmustine (3.3 mg/ml) and 20.3 minutes with Thiotepa (10.0 mg/ml).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.