(25 days)
The Siemens Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations. *
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. This system is designed for whole body oncology, neurology and cardiology examinations. The Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
The software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The Biograph Horizon PET/CT systems and Biograph mCT Flow PET/CT systems that are the subject of this Premarket Notification are identical in design, material, functionality, technology and energy source to the commercially available Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems.
The provided document describes a 510(k) Premarket Notification for Siemens Biograph Horizon PET/CT and Biograph mCT/mCT Flow PET/CT Systems. The purpose of this submission is not to introduce a new device, but to support an additional indication for use for low dose lung cancer screening in high-risk populations. The key information for this request revolves around demonstrating that the CT component of these combined PET/CT systems is substantially equivalent to standalone SOMATOM CT systems (K142955) that already have this specific indication.
Therefore, the "proof" is based on demonstrating equivalence to an already cleared device, rather than a de novo clinical study proving effectiveness or a multi-reader, multi-case study in the context of AI assistance. The document explicitly states: "No additional performance testing beyond what was cleared in the original Premarket Notifications is required." This means that the original performance tests for the standalone CT systems are being leveraged.
Below is an attempt to structure the information based on your request, noting that some of your specific categories (like AI assistance, ground truth for training sets, number of experts for ground truth, and adjudication methods for a novel device) are not directly applicable to this specific 510(k) submission due to its nature as an "additional indication for use" based on substantial equivalence.
Acceptance Criteria and Device Performance (Based on Substantial Equivalence to Predicate)
The acceptance criteria for the new indication for use (low dose lung cancer screening) are based on demonstrating that the performance of the CT component within the Biograph Horizon PET/CT and Biograph mCT/mCT Flow PET/CT Systems is identical to that of the previously cleared standalone SOMATOM CT systems (K142955) for this specific application.
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category (for CT Component) | Specific Criterion (Performance Parameter Comparison) | Reported Device Performance (Subject Device vs. Predicate) |
|---|---|---|
| Imaging Performance | CT number accuracy | "no differences exist in the parameters" |
| CT number uniformity | "no differences exist in the parameters" | |
| Spatial resolution (MTF, max in-plane resolution) | "no differences exist in the parameters" | |
| Slice thickness/sensitivity profile (min slice width) | "no differences exist in the parameters" | |
| Noise properties (NPS and image Noise (std dev)) | "no differences exist in the parameters" | |
| Contrast to Noise Ratio | "no differences exist in the parameters" | |
| Operational Performance | Maximum scan speed | "no differences exist in the parameters" |
| Minimum reconstructed slice interval | "no differences exist in the parameters" | |
| Overall Equivalence | Design, material, functionality, technology, energy source | "identical" (to commercially available predicate systems) |
| Intended Use | "no difference in the intended use of the device" | |
| Radiation Safety Standards | Compliance with 21 CFR 1020.30, 1020.33, 1040.10, 1040.11 | |
| Electrical/Mechanical Safety Standards | Compliance with IEC 60601-1 series and AAMI / ANSI ES60601-1 | |
| Software Lifecycle Standards | Compliance with IEC 62304 | |
| Risk Management | Compliance with ISO 14971 |
Conclusion from performance testing: "The test results demonstrate that the subject devices perform the same as the standalone SOMATOM CT systems. Since the standalone systems have been identified as suitable for lung cancer screening (K142955), the use of the CT component of the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems is likewise suitable for lung cancer screening."
Study Details Proving Device Meets Acceptance Criteria
This submission relies on demonstrating substantial equivalence to existing, legally marketed predicate devices, specifically the standalone SOMATOM CT systems (K142955) that were previously cleared for low dose lung cancer screening. Therefore, the "study" is primarily a comparative analysis of technical parameters and design specifications, rather than a new clinical trial.
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable in the context of patient data for this specific 510(k). The "test set" here refers to the technical parameters of the devices.
- Data Provenance: The comparison is made between the technical specifications and existing performance validation (from previous 510(k) clearances) of the Biograph Horizon/mCT/mCT Flow PET/CT systems and the SOMATOM CT systems (K142955). This is an engineering and regulatory comparison of device characteristics and previously validated performance, not a new retrospective or prospective clinical data study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not directly applicable. The "ground truth" for this submission is whether the technical parameters of the CT component in the PET/CT system are demonstrably identical to those of the predicate standalone CT system. This is established through engineering specifications, previous performance reports, and adherence to recognized standards, rather than expert consensus on a scan dataset for a new device. The original clearance for the predicate CT system would have involved relevant expert input.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. This is not a study involving human reader interpretation of images that would require an adjudication method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done. If so, what was the effect size of how much human readers improve with AI versus without AI assistance:
- No MRMC study was conducted for this 510(k) submission. The device (PET/CT system) is an imaging equipment; it is neither an AI algorithm nor an AI-assisted diagnostic tool in the sense of the question. The request is to add an indication for a human-interpreted CT scan for screening.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The CT component itself is a standalone imaging modality within the PET/CT system. The "performance" assessment was of these physics/engineering parameters to ensure they are identical to the predicate CT system. There is no AI algorithm being evaluated for standalone performance in this submission.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For this specific submission, the "ground truth" is the proven performance and technical specifications of the predicate device (SOMATOM CT systems K142955) regarding its suitability for low dose lung cancer screening. The comparison aims to establish that the subject device's CT component matches these established characteristics.
- The basis for the predicate device's clearance for lung cancer screening likely involved referencing clinical literature, such as the National Lung Screening Trial (NLST), and guidelines from professional medical societies, which would implicitly rely on outcomes data and expert consensus for defining suitable screening parameters and populations.
8. The sample size for the training set:
- Not applicable. This submission is for an imaging device, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established:
- Not applicable (as above).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff entwined with snakes, often used as a symbol of medicine and healthcare. The caduceus is depicted with three intertwined snakes, forming a stylized representation of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 25, 2016
Siemens Medical Solutions USA, Inc. % M. Alaine Medio, RAC PET and PCS Regulatory Projects Manager 810 Innovation Drive KNOXVILLE TN 37932
Re: K162736
Trade/Device Name: Biograph Horizon PET/CT Systems Biograph mCT PET/CT and Biograph mCT Flow PET/CT Systems Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, JAK Dated: September 29, 2016 Received: September 30, 2016
Dear Mr. Medio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162736
Device Name
Biograph Horizon PET/CT Systems Biograph mCT PET/CT and Biograph mCT Flow PET/CT Systems
Indications for Use (Describe)
The Siemens Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing. diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations *
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Siemens Biograph Horizon PET/CT Systems and Biograph mCT and mCT Flow PET/CT Systems 510(k) Premarket Notification
s required by 21 CFR Part 807.92
Identification of the Submitter
| Submitter: | M. Alaine Medio, RACPET and PCS Regulatory Projects ManagerSiemens Medical Solutions USA, Inc.Molecular Imaging810 Innovation DriveKnoxville, TN 37932 |
|---|---|
| Manufacturer: | Siemens Medical Solutions USA, Inc.Molecular Imaging2501 North Barrington RoadHoffman Estates, IL 60192 |
| Telephone Number: | (865)218-2703 |
| Fax Number: | (865)218-3019 |
| Date of Submission: | September 29th, 2016 |
ldentification of the product
| Device Proprietary Name: | Biograph Horizon PET/CT Systems | Biograph mCT PET/CT andBiograph mCT Flow PET/CT Systems |
|---|---|---|
| Common Name | Positron Emission Tomography (PET) SystemComputed Tomography (CT) System | |
| Classification Name: | Emission Computed Tomography System per 21 CFR 892.1200Computed Tomography X-Ray System per 21 CFR 892.1750 | |
| Product Code: | 90 KPS and 90 JAK | |
| Classification Panel: | Radiology | |
| Device Class: | Class II |
{5}------------------------------------------------
Marketed Devices to which Equivalence is claimed
Predicate:
| Device Proprietary Name: | Biograph Horizon PET/CT Systems |
|---|---|
| Manufacturer: | Siemens Medical Solutions USA, Inc. |
| Product Code: | 90 KPS and JAK |
| Device Class: | Class II |
| 510(k) Number: | K152880 |
Reference Devices:
| Device Proprietary Name: | Biograph mCT PET/CT andBiograph mCT Flow PET/CTSystems | SOMATOM CT Scanners |
|---|---|---|
| 510(k) : | K151486 | K142955 |
Device Description:
The Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. This system is designed for whole body oncology, neurology and cardiology examinations. The Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
The software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The Biograph Horizon PET/CT systems and Biograph mCT Flow PET/CT systems that are the subject of this Premarket Notification are identical in design, material, functionality, technology and energy source to the commercially available Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems.
Intended Use:
The Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems are radioloqical imaqing systems that are a combination of a positron emission tomography (PET) camera system for nuclear medicine images, and a computed tomography (CT) camera
{6}------------------------------------------------
system for x-ray images. The nuclear medicine images and the x-ray images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (i.e., distribution of radiopharmaceuticals). The PET and CT portions of the system may be used independently or in combination. The PET and CT images may be transferred to other systems for radiation therapy planning or additional processing.
This submission was made to support an additional indication for use related to lung cancer screeninq within the subset of the overall intended use of the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems:
This CT system can be used for low dose lung cancer screening in high risk populations. *
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
This revised indication for use is cleared for the standalone SOMATOM CT systems (K142955) that are incorporated into the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems.
In accordance with guidance document "Guidance for Industry. General/Specific Intended Use" issued November 4, 1998, and presentation "FDA/MITA Meeting on LDCT Lung Cancer Screening", the modified more specific indication for use does not alter the intended use for the legally marketed predicate devices with a general indication for use, nor does it alter the intended diagnostic effect in comparison to the predicate devices.
Based on the analysis conducted within the original 510(k) for the SOMATOM CT systems (K142955) and the integration of the CT subsystem, without altering the technological characteristics, into the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems, the more specific use of low dose lung cancer screening is a subset of the general use rather than a new intended use. Further, the updated specific indication is substantially equivalent to the inclusion of this specific indication within the predicate device.
Revised Indications for Use:
The Siemens Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowinq for single modality CT and / or PET diagnostic imaging.
{7}------------------------------------------------
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations. *
- As defined by professional medical societies. Please refer to clinical literature. including the results of the National Lung Screeninq Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Performance Testinq / Safety and Effectiveness:
The Biograph Horizon PET/CT systems and Biograph mCT Flow PET/CT systems that are the subject of this Premarket Notification are identical in design, material, functionality, technology and energy source to the commercially available Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems. No additional performance testing beyond what was cleared in the original Premarket Notifications is required.
CT Testing in accordance with the requirements of the following FDA Performance Standards for lonizing Radiation Emitting Products and Light emitting products for CT equipment with a laser localizer was performed and documented in a Dosimetry and Imaging Performance Report. Further, each CT subsystem is tested and passes the Applicable Performance Standards prior to shipment:
- 21 CFR 1020.30 (a) ●
- 21 CFR 1020.30 (b)(36)(iii)-(v)
- 21 CFR 1020.30 (b)(58)-(62) ●
- 21 CFR 1020.30 (h)(3)(vi)-(viii) ●
- 21 CFR 1020.33
- 21 CFR 1040.10
- . 21 CFR 1040.11
Applicability
Technique factors CT, Scan, Scan Time, Tomogram, Dose Information to be provided for users Computed Tomography (CT) equipment Laser Products Specific purpose laser products
The Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems were designed in accordance with Design Controls and in accordance with the following FDA recognized standards:
- . Recognition Number 19-1: IEC 60601-1-2 Edition 3: 2007-03
- Recognition Number 19-4: AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 ●
- Recognition Number 5-89: IEC 60601-1-6 Edition 3.1 2013-10 ●
- Recognition Number 12-256: IEC 60601-2-44 Edition 3.1 2012-09
- Recognition Number 12-269: IEC 60601-1-3 Edition 2.1 2013-04
- Recognition Number 5-40: ISO 14971 Second Edition 2007-03 ●
- Recognition Number 13-79: IEC 62304 Edition 1.1 2015-06
- Recognition Number 12-265: NEMA NU 2-2012 ●
- Recognition Number 12-300: NEMA PS 3.1 - 3.20 (2016)
{8}------------------------------------------------
Siemens Biograph Horizon PET/CT Systems and Biograph mCT and mCT Flow PET/CT Systems 510(k) Premarket Notification
- Recognition Number 12-225: NEMA XR 25 ●
- . Recognition Number 12- 270: NEMA 61223-3-5 First Edition 2004
- Recognition Number 12-226: NEMA 61223-2-6 Second Edition: 2006 .
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beqinning early in the design cycle and continuinq throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
Technical parameters used to test general CT use are applicable for lung cancer screening. The technical parameters of the CT subsystem for the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems and the commercially available SOMATOM standalone CT systems devices with the additional lung cancer screening indication (K142955) were compared, and no differences exist in the parameters. Included in this evaluation, are the following parameters:
- CT number accuracy .
- . CT number uniformity
- Spatial resolution (MTF, maximum in-plane resolution) ●
- . Slice thickness/sensitivity profile (minimum slice width)
- Noise properties (NPS and image Noise (standard deviation))
- Contrast to Noise Ratio ●
- . Maximum scan speed
- Minimum reconstructed slice interval .
The test results demonstrate that the subject devices perform the same as the standalone SOMATOM CT systems. Since the standalone systems have been identified as suitable for lung cancer screening (K142955), the use of the CT component of the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems is likewise suitable for lung cancer screening.
Statement regarding Substantial Equivalence:
The Biograph Horizon PET/CT Systems and Biograph mCT and mCT Flow PET/CT systems are commercially available and no changes to technological, functional or clinical features have been made. There is no difference in the intended use of the device. The CT systems being referenced as predicate devices (K14955) are integrated into the PET/CT systems as a subsystem of the device and no differences in functionality have been made compared to the commercially available systems (with the exception of the tilt functionality). All parameters associated with low dose lung cancer screening are the same between the devices subject to this application and the predicate devices referenced.
{9}------------------------------------------------
Siemens Biograph Horizon PET/CT Systems and Biograph mCT and mCT Flow PET/CT Systems 510(k) Premarket Notification
Based on the information provided in the Premarket Notification, Siemens Medical Solutions USA, Inc. considers the Biograph Horizon PET/CT systems and Biograph mCT and mCT Flow PET/CT systems with the added low dose lung cancer screening indication to be substantially equivalent to the currently commercially available predicate device.
§ 892.1200 Emission computed tomography system.
(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.