K Number
K162722
Device Name
HyperSense
Date Cleared
2017-04-20

(203 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HyperSense is a software feature intended for use on GE MR 1.5T and 3.0T Systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction that allows users to reduce scan times or increase scan resolution. HyperSense can be used for non-contrast enhanced imaging of the head, neck, spine, extremities, pelvis, and abdomen.

Device Description

HyperSense is a software feature used on GE 1.5T and 3.0T MR systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction, enabling faster imaging without the penalties commonly found with conventional parallel imaging. HyperSense is intended to be used with volumetric acquisitions, and can be combined with other methods of acceleration (ARC) for achieving high signal to noise ratio with shorter acquisition times. HyperSense can deliver higher spatial resolution images or reduced scan times.

AI/ML Overview

HyperSense Device Acceptance Criteria and Study Details

The provided document describes the GE HyperSense, a software feature for MR systems that uses sparse data sampling and iterative reconstruction to reduce scan times or increase resolution.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Measured Metric)Reported Device Performance (HyperSense with predicate)
Image quality (legibility of morphological features)Comparable imaging performance
Signal to Noise Ratio (SNR)Evaluated (results not explicitly stated, but "passing results" generally imply meeting predetermined criteria)
Spatial ResolutionEvaluated (results not explicitly stated, but "passing results" generally imply meeting predetermined criteria)

Note: The document states that "the non-clinical testing was completed with passing results per the pass/fail criteria defined in the test cases" for SNR and spatial resolution. For clinical testing, it states "The clinical results demonstrated that HyperSense maintains comparable imaging performance results as its predicate devices."

2. Sample Size for Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated for either phantom or clinical studies.
  • Data Provenance: Not explicitly stated for the clinical study. It mentions "sample clinical images are included in this submission," suggesting the data was likely from relevant GE MR systems. The study is referred to as "A clinical study has been performed," implying a prospective nature, though not explicitly stated.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not explicitly stated.
  • Qualifications of Experts: Not explicitly stated.

4. Adjudication Method

  • Adjudication Method: Not explicitly stated.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, an MRMC comparative effectiveness study is not mentioned. The clinical study focused on evaluating the impact of HyperSense on image quality compared to its predicate devices, not on human reader improvement with AI assistance.

6. Standalone Performance Study

  • Standalone Study: Yes, a standalone performance study in the form of phantom testing was conducted to evaluate the impact of HyperSense on Signal to Noise Ratio (SNR) and spatial resolution. The "clinical study" also appears to be a standalone assessment of the device's image quality. The device is a "software only feature" that enhances image acquisition and reconstruction, implying its direct impact on the image itself, which is a standalone performance by the algorithm.

7. Type of Ground Truth Used

  • Phantom Testing: Objective measurements of SNR and spatial resolution.
  • Clinical Study: "Legibility of morphological features," which implies a qualitative assessment by experts (implicitly radiologists or medical professionals experienced with MRI interpretations). This would typically involve expert consensus or assessment against established diagnostic criteria. The document also mentions a "peer reviewed journal article describing a study of the HyperSense technique as supporting evidence," which would also likely rely on similar ground truth methodologies.

8. Sample Size for Training Set

  • Sample Size for Training Set: Not mentioned. HyperSense is an acceleration technique and iterative reconstruction method. While it's a software feature, the process of developing and tuning such algorithms often involves various datasets for training and validation, but these details are not provided in the summary.

9. How Ground Truth for Training Set Was Established

  • How Ground Truth for Training Set was Established: Not mentioned. Given the nature of an iterative reconstruction algorithm, "ground truth" for training would typically involve high-quality, fully sampled MR images that the algorithm aims to replicate or improve upon from undersampled data. The process would likely involve various forms of quantitative image metrics and expert review to optimize the reconstruction parameters.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 20, 2017

Ge Medical Systems, LLC Glen Sabin Regulatory Affairs Director 3200 Grandview Blvd Waukesha. Wisconsin 53188

Re: K162722

Trade/Device Name: HyperSense Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: March 15, 2017 Received: March 16, 2017

Dear Glen Sabin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael O'Hara

For

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162722

Device Name HyperSense

Indications for Use (Describe)

HyperSense is a software feature intended for use on GE MR 1.5T and 3.0T Systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction that allows users to reduce scan times or increase scan resolution. HyperSense can be used for non-contrast enhanced imaging of the head, neck, spine, extremities, pelvis, and abdomen.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue. The background is white.

In accordance with 21 CFR 807.92 the following summary of information is provided:
Date:19 April 2017
Primary Contact Person:Glen SabinRegulatory Affairs Director, MRGE Healthcare, (GE Medical Systems, LLC)Phone: 262-521-6848Fax: 414-908-9585
Secondary Contact Person:James McMahonSenior Director, Regulatory Affairs - MRGE Healthcare, (GE Medical Systems, LLC)Phone: 508-382-2858
Device Trade Name:HyperSense
Common/Usual Name:Magnetic Resonance Diagnostic Device
Classification Names:892.1000
Product Code:LNH
Predicate Devices:Predicate Device: ARC (K142085)
Device Description:HyperSense is a software feature used on GE 1.5T and3.0T MR systems.HyperSense is an acceleration technique based onsparse data sampling and iterative reconstruction,enabling faster imaging without the penaltiescommonly found with conventional parallel imaging.HyperSense is intended to be used with volumetricacquisitions, and can be combined with other methodsof acceleration (ARC) for achieving high signal to noiseratio with shorter acquisition times. HyperSense candeliver higher spatial resolution images or reducedscan times.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white water droplet shapes around the letters. The logo is simple and recognizable.

Indications for Use:HyperSense is a software feature intended for use onGE MR 1.5T and 3.0T Systems. HyperSense is anacceleration technique based on sparse data samplingand iterative reconstruction that allows users to reducescan times or increase scan resolution. HyperSensecan be used for non-contrast enhanced imaging of thehead, neck, spine, extremities, pelvis, and abdomen.
Comparison of IntendedUseBoth HyperSense and the predicate device aremagnetic resonance imaging devices intended fordiagnostic use.Both indications for use statements are functional innature, and do not list specific diseases or conditions.HyperSense and the predicate device are indicated forthe same patient population, and for the same clinicalsetting. Both indications for use statements include arange of intended anatomies.Therefore, GE Healthcare believes that HyperSense hasthe same intended use as the predicate device inaccordance with the FDA's guidance document "The510(k) Program: Evaluating Substantial Equivalence inPremarket Notifications [510(k)]", dated 28 July 2014.
Comparison ofTechnologicalCharacteristics:The most notable technological difference betweenHyperSense and the predicate device is that HyperSenseadopts compressed sensing acquisition method basedon sparse data compressibility allowing scan timereduction while maintaining appropriate image quality.Per bench and clinical data collected to validate theindications for use, this technological difference doesnot raise any different questions of safety andeffectiveness. Both devices must address questions ofwhether they provide an adequate level of imagequality appropriate for diagnostic use. The performancedata described in this submission include results ofboth bench testing and clinical testing that show theimage quality performance of HyperSense compared tothe predicate device

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are also some water droplet-like shapes surrounding the circle. The logo is simple and recognizable, representing the company's brand identity.

Performance DataSummary of Non-Clinical Tests:
HyperSense is a software only feature and complies with the following voluntary standards:AAMI/ANSI 62304 AAMI/ANSI ES60601-1 IEC 60601-2-33
In addition, HyperSense complies with NEMA PS3.1-3.18 for DICOM conformance.
HyperSense complies with the same applicable standards as the software used in the predicate device.
As with the predicate device, the following quality assurance measures were applied to the development of HyperSense:Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation)
Phantom testing has been conducted to evaluate the impact of HyperSense on Signal to Noise Ratio (SNR), and spatial resolution.
The non-clinical testing was completed with passing results per the pass/fail criteria defined in the test cases.
Summary of Clinical Tests:
A clinical study has been performed to evaluate the impact of HyperSense on image quality, as measured by the legibility of morphological features. Additionally, a peer reviewed journal article describing a study of the HyperSense technique has been provided in this submission as supporting evidence. The clinical results demonstrated that HyperSense maintains comparable imaging performance results as its predicate devices (K142085). Sample clinical images are included in this submission.
Conclusion:GE Healthcare believes that the HyperSense softwarefeature has substantially the same intended use as thepredicate ARC. This 510(k) submission includesinformation on the technological characteristics of theHyperSense feature, as well as performance datademonstrating that HyperSense is as safe and effectiveas the predicate, and does not raise different questionsof safety and effectiveness.In conclusion, GE Healthcare believes that theHyperSense feature is substantially equivalent to thepredicate device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are white swirls around the letters, creating a dynamic and recognizable design. The logo is simple, clean, and easily identifiable.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.