(90 days)
No reference devices were used in this submission.
No
The document describes a standard automated blood coagulation analyzer and associated reagents, focusing on analytical performance studies. There is no mention of AI, ML, or related technologies in the intended use, device description, or performance studies.
No
This device is an in vitro diagnostic analyzer used to measure coagulation factors in blood samples, which aids in diagnosis rather than providing direct therapy.
Yes
The device, Sysmex CS-2100i, is clearly stated as an "in vitro diagnostic use" device for analyzing various blood coagulation parameters, which are used to determine specific conditions or levels of substances in patient samples to aid in diagnosis.
No
The device description explicitly states it is an "automated blood coagulation instrument" and mentions hardware components like an "Information Processing Unit (IPU) screen" and the ability to print on "external printers or transmitted to a host computer." The 510(k) is for reagent applications used with this hardware.
Yes, this device is an IVD (In Vitro Diagnostic).
The "Intended Use / Indications for Use" section explicitly states: "The Sysmex CS-2100i is a fully automated blood coagulation analyzer intended for in vitro diagnostic use..."
Furthermore, the descriptions for the associated reagents also state "In vitro diagnostic reagent".
N/A
Intended Use / Indications for Use
For Sysmex® CS-2100i:
The Sysmex CS-2100i is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory.
For determination of:
- Prothrombin Time (PT) seconds and PT INR with Dade® Innovin® .
- . Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
- . Fibrinogen (Fbg) with Dade® Thrombin Reagent
- . Coagulation Factor V with Dade® Innovin®
- . Coagulation Factor VII with Dade® Innovin®
- Protein C with Protein C Reagent
- . Antithrombin (AT) with INNOVANCE® Antithrombin
- Protein C with Berichrom® Protein C
- D-dimer with INNOVANCE® D-Dimer
The performance of this device has not been established in neonate and pediatric patient populations.
For Coagulation Factor V Deficient Plasma:
Intended Use
In vitro diagnostic reagent for the determination of the activity of coagulation factor V in human plasma.
For Coagulation Factor VII Deficient Plasma:
Intended Use
In vitro diagnostic reagents for the determination of the activity of coagulation factor II (prothrombin), coagulation factor VII and coagulation factor X in human plasma by coagulometric methods.
For Protein C Reagent: Intended Use Protein C Reagent is a coagulation test for the quantitative determination of protein C activity in human plasma.
For Berichrom® Protein C:
Intended Use
For the quantitative determination of functionally active protein C using a chromogenic substrate as an aid in the diagnosis of inherited and acquired deficiencies.
Product codes (comma separated list FDA assigned to the subject device)
JPA, GGP, GJT
Device Description
The Sysmex® CS-2100i is an automated blood coagulation instrument which can analyze samples using clotting, chromogenic and immunoassay methods. Analysis results are displayed on the Information Processing Unit (IPU) screen. They can be printed on external printers or transmitted to a host computer. Sold separately from the instrument are the associated
- Reagents
- Controls
- Calibrators
- Consumable materials
The subject of this 510(k) notification are reagent applications which perform the coagulation tests Coagulation Factor V with Dade® Innovin®, Coagulation Factor VII with Dade® Innovin®, Protein C with Protein C Reagent and Protein C with Berichrom® Protein C.
The analysis principles used on the instrument are reflected by the reagent application testing provided in this 510(k) notification and is described in the below table.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
The performance of this device has not been established in neonate and pediatric patient populations.
Intended User / Care Setting
The intended Environment of Use is a clinical central/hospital laboratory.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Method comparison
Method comparison studies designed according to EP09-A3 CLSI Guideline 'Measurement Procedure Comparison and Bias Estimation Using Patient Samples' were conducted at four external sites in the United States, all sites using the same protocol. Samples were measured on both the predicate device (Sysmex® CA-1500) as well as the new device (Sysmex® CS-2100i), in random order to eliminate any inherent bias. Results were compared by Passing-Bablok regression analysis as well as Bland-Altman plots. Results from each application met the predetermined acceptance criteria.
Reproducibility Studies
Twenty-day precision studies were performed at two external sites in Germany and one external site in the United States. Testing followed the scheme of two runs per day, with two replicates per run, at each of the three sites according to CLSI EP05-A2 'Evaluation of Precision Performance of Quantitative Measurement Methods'. The order of the analysis of parameter, samples and quality control samples for each run and day varied to avoid an inherent bias to the study. One calibration curve of each calibrated application was used in the study. Within Run, Between Run, Between Day, and Total (within site) were calculated.
Detection Capability Results
Detection capability studies were measured for the calibrated assays on the Sysmex® CS-2100i: Coagulation Factor V with Dade® Innovin®, Coagulation Factor VII with Dade® Innovin®, Protein C with Protein C Reagent, and Protein C with Berichrom® Protein C. Studies were conducted following the CLSI document EP17-A2 'Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures'. Data for all tested reagents met the predetermined acceptance criteria and support the lower limit of the clinically reportable range claim.
Linearity & Measuring Range
Linearity studies were performed for the calibrated assays on the Sysmex® CS-2100i: Coagulation Factor V with Dade® Innovin®, Coagulation Factor VII with Dade® Innovin®, Protein C with Protein C Reagent, and Protein C with Berichrom® Protein C. All reagents met the predetermined acceptance criteria and support the clinically reportable range claim. Studies were conducted as described in CLSI EP6-A 'Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach'.
Reference Interval
Reference interval studies were conducted at three clinical study sites in the United States following the guidance of CLSI document EP28-A3c 'Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory'. The study population did not include neonate and pediatric sample populations.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Method Comparison:
Passing-Bablok Regression
- Factor V with Dade® Innovin®: N=589, Sample range=7.2 - 145.1 (% of norm), r=0.983, Slope=0.990 (0.977, 1.003), Intercept=1.861 (1.025, 2.784)
- Factor VII with Dade® Innovin®: N=492, Sample range=6.2 - 148.8 (% of norm), r=0.991, Slope=1.039 (1.028, 1.051), Intercept=0.636 (-0.135, 1.104)
- Protein C with Protein C Reagent: N=584, Sample range=10.3 - 129.3 (% of norm), r=0.977, Slope=0.966 (0.952, 0.981), Intercept=-0.372 (-1.207, 0.470)
- Protein C with Berichrom® Protein C: N=507, Sample range=11.3 - 134.7 (% of norm), r=0.994, Slope=0.942 (0.934, 0.951), Intercept=-0.292 (-1.037, 0.399)
Reproducibility Summarized Table (%CV)
Within Run:
- Coagulation Factor V with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site=4.07 – 7.40, 2nd Site=2.07 - 3.34, 3rd Site=2.70 - 6.09, Sites Combined=3.16 - 5.01
- Coagulation Factor VII with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site=1.74 - 2.38, 2nd Site=1.71 – 2.87, 3rd Site=1.99 - 2.59, Sites Combined=1.85 - 2.63
- Protein C with Protein C Reagent (10.1 - 131.0% of norm): 1st Site=2.26 - 3.75, 2nd Site=2.08 - 2.68, 3rd Site=2.43 - 7.02, Sites Combined=2.46 - 4.78
- Protein C with Berichrom® Protein C (10.0 - 138.0% of norm): 1st Site=1.02 - 5.34, 2nd Site=1.00 - 4.56, 3rd Site=1.20 - 5.26, Sites Combined=1.15 - 4.63
Between Run:
- Coagulation Factor V with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site=0.00 - 2.80, 2nd Site=0.00 - 3.87, 3rd Site=0.00 - 3.36, Sites Combined=1.76 - 2.55
- Coagulation Factor VII with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site=0.65 - 1.41, 2nd Site=0.00 - 1.35, 3rd Site=0.00 - 2.32, Sites Combined=0.00 - 1.72
- Protein C with Protein C Reagent (10.1 - 131.0% of norm): 1st Site=0.00 - 1.65, 2nd Site=0.85 - 2.70, 3rd Site=0.00 - 2.50, Sites Combined=0.56 - 1.95
- Protein C with Berichrom® Protein C (10.0 - 138.0% of norm): 1st Site=0.54 - 2.51, 2nd Site=0.32 - 2.23, 3rd Site=0.00 - 1.20, Sites Combined=0.84 - 1.66
Between Day:
- Coagulation Factor V with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site=0.00 - 1.55, 2nd Site=0.00 - 2.56, 3rd Site=0.00 - 2.81, Sites Combined=0.00 - 1.64
- Coagulation Factor VII with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site=0.00 - 1.02, 2nd Site=1.48 – 2.68, 3rd Site=0.00 - 1.92, Sites Combined=0.89 - 1.67
- Protein C with Protein C Reagent (10.1 - 131.0% of norm): 1st Site=0.00 - 2.41, 2nd Site=0.00 - 1.20, 3rd Site=0.00 - 2.60, Sites Combined=0.00 - 2.17
- Protein C with Berichrom® Protein C (10.0 - 138.0% of norm): 1st Site=0.00 - 0.57, 2nd Site=0.00 - 1.02, 3rd Site=0.00 - 2.36, Sites Combined=0.00 - 1.40
Total CV (Within Site and Sites Combined):
- Coagulation Factor V with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site Total CV=4.64 - 7.40, 2nd Site Total CV=3.37 - 5.12, 3rd Site Total CV=3.85 - 6.95, Sites Combined Total CV=4.78 - 10.43
- Coagulation Factor VII with Dade® Innovin® (6.0 - 149.0% of norm): 1st Site Total CV=2.01 - 2.81, 2nd Site Total CV=2.58 – 3.56, 3rd Site Total CV=2.23 - 3.31, Sites Combined Total CV=2.57 - 3.59
- Protein C with Protein C Reagent (10.1 - 131.0% of norm): 1st Site Total CV=2.78 - 4.55, 2nd Site Total CV=2.36 - 3.83, 3rd Site Total CV=3.47 - 7.49, Sites Combined Total CV=3.47 - 5.85
- Protein C with Berichrom® Protein C (10.0 - 138.0% of norm): 1st Site Total CV=1.40 - 5.91, 2nd Site Total CV=1.40 - 5.18, 3rd Site Total CV=1.70 - 5.26, Sites Combined Total CV=2.19 - 5.69
Detection Capability Results:
Limit of Quantitation Studies
- Coagulation Factor V with Dade® Innovin®: Lower Limit of Clinically Reportable Range=6.0 (% of norm), Measured Limit of Quantitation=4.80 (% of norm), Maximum Total Error=2.04 (% of norm)
- Coagulation Factor VII with Dade® Innovin®: Lower Limit of Clinically Reportable Range=6.0 (% of norm), Measured Limit of Quantitation=3.39 (% of norm), Maximum Total Error=0.21 (% of norm)
- Protein C with Protein C Reagent: Lower Limit of Clinically Reportable Range=10.1 (% of norm), Measured Limit of Quantitation=9.35 (% of norm), Maximum Total Error=3.62 (% of norm)
- Protein C with Berichrom® Protein C: Lower Limit of Clinically Reportable Range=10.0 (% of norm), Measured Limit of Quantitation=8.32 (% of norm), Maximum Total Error=2.07 (% of norm)
Linearity & Measuring Range:
- Coagulation Factor V with Dade® Innovin®: Measured Linear Range=3.4 – 180.7% of norm, Clinically Reportable Range=6.0 – 149.0% of norm
- Coagulation Factor VII with Dade® Innovin®: Measured Linear Range=4.3 – 179.5% of norm, Clinically Reportable Range=6.0 – 149.0% of norm
- Protein C with Protein C Reagent: Measured Linear Range=7.0 – 187.7% of norm, Clinically Reportable Range=10.1 – 131.0% of norm
- Protein C with Berichrom® Protein C: Measured Linear Range=7.1 – 181.3% of norm, Clinically Reportable Range=10.0 – 138.0% of norm
Reference Interval Summary Table:
- Coagulation Factor V with Dade® Innovin®: N=193, Sysmex® CS-2100i Reference Interval=79.8% (5th Percentile of norm)
- Coagulation Factor VII with Dade® Innovin®: N=193, Sysmex® CS-2100i Reference Interval=65.6% (5th Percentile of norm)
- Protein C with Protein C Reagent: N=193, Sysmex® CS-2100i Reference Interval=75.6% (5th Percentile of norm)
- Protein C with Berichrom® Protein C: N=191, Sysmex® CS-2100i Reference Interval=84.9% (5th Percentile of norm)
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Sysmex® Automated Blood Coagulation Analyzer CA-1500 (K011235), Factor V with Dade® Innovin® K993299, Factor VII with Dade® Innovin® K993299, Protein C with Protein C Reagent K000649, Protein C with Berichrom® Protein C K001645
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
No reference devices were used in this submission.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 864.5425 Multipurpose system for in vitro coagulation studies.
(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
0
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of a caduceus, a symbol often associated with healthcare. The caduceus is composed of three intertwined snakes and a staff.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 26, 2016
Siemens Healthcare Diagnostics Products GmbH Ms. Donna Noeh Regulatory Affairs Manager Emil-von-Behring Strasse 76 35041 Marburg, Germany
Re: K162688 Trade/Device Name: Sysmex® CS-2100i Coagulation Factor V Deficient Plasma Coagulation Factor II. VII and X Deficient Plasmas Protein C Reagent Berichrom® Protein C Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: JPA, GGP, GJT Dated: September 26, 2016 Received: September 27, 2016
Dear Ms. Noeh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Leonthena R. Carrington -S
Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K162688
Device Name
Sysmex® CS-2100i, Coagulation Factor V Deficient Plasma, Coagulation Factor II, VII, and X Deficient Plasmas, Protein C Reagent, Berichrom® Protein C
Indications for Use (Describe)
For Sysmex® CS-2100i:
The Sysmex CS-2100i is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory.
For determination of:
- Prothrombin Time (PT) seconds and PT INR with Dade® Innovin® .
- . Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
- . Fibrinogen (Fbg) with Dade® Thrombin Reagent
- . Coagulation Factor V with Dade® Innovin®
- . Coagulation Factor VII with Dade® Innovin®
- . Protein C with Protein C Reagent
- . Antithrombin (AT) with INNOVANCE® Antithrombin
- Protein C with Berichrom® Protein C
- D-dimer with INNOVANCE® D-Dimer
The performance of this device has not been established in neonate and pediatric patient populations.
For Coagulation Factor V Deficient Plasma:
Intended Use
In vitro diagnostic reagent for the determination of the activity of coagulation factor V in human plasma.
For Coagulation Factor VII Deficient Plasma:
Intended Use
In vitro diagnostic reagents for the determination of the activity of coagulation factor II (prothrombin), coagulation factor VII and coagulation factor X in human plasma by coagulometric methods.
For Protein C Reagent: Intended Use Protein C Reagent is a coagulation test for the quantitative determination of protein C activity in human plasma.
For Berichrom® Protein C:
Intended Use
For the quantitative determination of functionally active protein C using a chromogenic substrate as an aid in the diagnosis of inherited and acquired deficiencies.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
3
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92 and follows the FDA guidance 'The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]', issued July 28, 2014.
Submitter 1
Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg, Germany
Contact Person: | Donna Noeh |
---|---|
Email: | donna.noeh@siemens.com |
Phone: | + (49) 6421 39 5107 |
Facsimile: | + (49) 6421 39 4977 |
Date Prepared: | November 18, 2016 |
2 Device
3
Name of Device: | Sysmex® Automated Blood Coagulation Analyzer CS-2100 |
---|---|
Common or Usual Name: | Sysmex® CS-2100i; CS-2100i |
Classification Name: | Multipurpose system for in vitro coagulation studies |
(21 CFR 864.5425) | |
Regulatory Class: | II |
Product Code: | JPA |
510(k) Review Panel | Hematology |
Predicate Device | |
Name of Device: | Sysmex® Automated Blood Coagulation Analyzer CA-1500 |
(K011235) | |
Common or Usual Name: | Sysmex® CA-1500 |
Classification Name: | Multipurpose system for in vitro coagulation studies |
(21 CFR 864.5425) | |
Regulatory Class: | II |
Product Code: | JPA |
510(k) Review Panel | Hematology |
The predicate has not been subject to a design-related recall for any of the applications associated with this premarket notification. No reference devices were used in this submission.
5
| Reagent Name | K number | Regulation
Number | Regulatory
Class | Product Code | Panel |
|-------------------------------------------|----------|----------------------|---------------------|--------------|------------|
| Factor V with
Dade®
Innovin® | K993299 | 864.7290 | 2 | GJT | Hematology |
| Factor VII with
Dade®
Innovin® | K993299 | 864.7290 | 2 | GJT | Hematology |
| Protein C with
Protein C
Reagent | K000649 | 864.7290 | 2 | GGP | Hematology |
| Protein C with
Berichrom®
Protein C | K001645 | 864.7290 | 2 | GGP | Hematology |
Reagent Applications that are the subject of this 510(k) notification:
Device Description / Test Principle 4
The Sysmex® CS-2100i is an automated blood coagulation instrument which can analyze samples using clotting, chromogenic and immunoassay methods. Analysis results are displayed on the Information Processing Unit (IPU) screen. They can be printed on external printers or transmitted to a host computer. Sold separately from the instrument are the associated
- Reagents
- Controls
- Calibrators ●
- Consumable materials ●
The subject of this 510(k) notification are reagent applications which perform the coagulation tests Coagulation Factor V with Dade® Innovin®, Coagulation Factor VII with Dade® Innovin®, Protein C with Protein C Reagent and Protein C with Berichrom® Protein C.
The analysis principles used on the instrument are reflected by the reagent application testing provided in this 510(k) notification and is described in the below table.
Table of Sysmex® CS-2100i Analysis Principles | ||
---|---|---|
Reagent | Application | Methodology |
Dade® Innovin® with | ||
Coagulation Factor V | ||
Deficient Plasma | Coagulation Factor V with | |
Dade® Innovin® | Clotting | |
(extrinsic pathway) |
6
| Dade® Innovin® with
Coagulation Factor VII
Deficient Plasma | Coagulation Factor VII with
Dade® Innovin® | Clotting
(extrinsic pathway) |
|-------------------------------------------------------------------|-----------------------------------------------|---------------------------------|
| Protein C Reagent | Protein C with
Protein C Reagent | Clotting |
| Berichrom® Protein C | Protein C with
Berichrom® Protein C | Chromogenic |
The intended Environment of Use is a clinical central/hospital laboratory.
7
Instrument (main unit)
Front View of the Sysmex CS-2100i (main unit) Figure 1:
Image /page/7/Figure/3 description: This image is a line drawing of a piece of equipment with multiple labeled parts. The parts are labeled with the numbers 1 through 9. The equipment has a boxy shape with multiple compartments and openings.
(1) Power connector: Connected to the power cable.
- (2) Power Switch: Turns the power ON and OFF.
- (3) Light shield lid: Open this cover to set reagents and perform maintenance.
- (4) Alarm indicator LED: This displays alarms affecting the instrument.
- (5) Cuvette hopper: Holds cuvettes and automatically supplies them to the instrument.
- (6) Start button: This is the same as the Start button on the IPU screen.
(7) Mechanical stop switch: Press this switch to immediately stop the instrument's mechanical movement.
(8) Cuvette trash tray: Used cuvettes are discarded here.
(9) Sampler: The sampler automatically transports samples that are set in the sample rack to the aspiration position.
8
Figure 2: Informational Processing Unit (IPU) Sysmex CS-2100i
Image /page/8/Figure/2 description: The image shows a computer setup with four labeled parts. The labels point to the monitor (1), the computer tower (2), the keyboard (3), and the mouse (4). The monitor is placed on top of the computer tower, and the keyboard is in front of the tower. The mouse is located to the right of the keyboard.
- (1) Touch panel display: Displays the IPU screen. It can also be used as a touch panel.
- (2) IPU Main Unit: This is the Main Unit of IPU.
- (3) Keyboard: Used to operate the IPU together with the touch panel.
- (4) Mouse: Used to operate the IPU together with the touch panel.
The instrument is capable of measuring in the following analysis modes:
(1) Normal mode: Samples for all the analyses including re-analyses are taken into the instrument at the same time and analyzed. Automatic re-analysis can also be performed.
(2) Micro-sample mode: The sample volume from samples set in the sampler or STAT holder is taken into the instrument for each analysis through dispensing sample probe and analyzed. This analysis mode can also be performed with less sample volume than normal mode; however, automatic re-analysis cannot be performed.
9
5 Intended Use / Indications for Use
The Sysmex® CS-2100i is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory.
For determination of:
- Prothrombin Time (PT) seconds and PT INR with Dade® Innovin®
- Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
- Fibrinogen (Fbg) with Dade® Thrombin Reagent
- Coagulation Factor V with Dade® Innovin® ●
- Coagulation Factor VII with Dade® Innovin®
- Protein C with Protein C Reagent ●
- Antithrombin (AT) with INNOVANCE® Antithrombin ●
- Protein C with Berichrom® Protein C
- D-dimer with INNOVANCE® D-Dimer .
The performance of this device has not been established in neonate and pediatric patient populations.
Coagulation Factor V with Dade® Innovin®
In vitro diagnostic reagent for the determination of the activity of factor V in human plasma.
Coagulation Factor VII with Dade® Innovin® Plasma
In vitro diagnostic reagent for the determination of the activity of coagulation factor VII (prothrombin), coagulation factor VII and coagulation factor X in human plasma by coagulometric methods.
Protein C with Protein C Reagent
Protein C Reagent is a coagulation test for the quantitative determination of protein C activity in human plasma.
Protein C with Berichrom® Protein C
For the quantitative determination of functionally active protein C using a chromogenic substrate as an aid in the diagnosis of inherited and acquired deficiencies.
Comparison of Technological Characteristics with the Predicate Device 6
Both the subject and predicate instruments employ the same technological characteristics in that they automatically analyze various clotting tests using reagents, calibrators and controls previously cleared for automated coagulation analyzers. The reagents perform at least equally well on both the subject and predicate instruments. At a high level, the devices have the following same technological elements:
10
Similarities between CS-2100i and CA-1500 | ||
---|---|---|
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
Regulatory Classification | JPA, Class II | |
System, Multipurpose for in | ||
vitro coagulation studies | Same | |
Intended Use Statement | The Sysmex® CS-2100i is a fully | |
automated blood coagulation | ||
analyzer intended for in vitro | ||
diagnostic use using plasma collected | ||
from venous blood samples in 3.2% | ||
sodium citrate tubes to analyze | ||
clotting, chromogenic and | ||
immunoassay methods in the clinical | ||
laboratory. | ||
For determination of: | ||
• Prothrombin Time (PT) seconds | ||
and PT INR with Dade® | ||
Innovin® | ||
• Activated Partial Thromboplastin | ||
Time (APTT) with Dade® | ||
Actin® FSL | ||
• Fibrinogen (Fbg) with Dade® | ||
Thrombin Reagent | ||
• Coagulation Factor V with | ||
Dade® Innovin® | ||
• Coagulation Factor VII with | ||
Dade® Innovin® | ||
• Protein C with Protein C Reagent | ||
• Antithrombin (AT) with | ||
INNOVANCE® Antithrombin | ||
• Protein C with Berichrom® | ||
Protein C | ||
• D-dimer with INNOVANCE® D- | ||
Dimer | ||
The performance of this device has | ||
not been established in neonate and | ||
pediatric patient populations. | The intended use of the Sysmex® | |
CA-1500 is as a fully automated, | ||
computerized blood plasma | ||
coagulation analyzer for in vitro | ||
diagnostic use in clinical | ||
laboratories. | ||
The instrument uses citrated | ||
human plasma to perform the | ||
following parameters and | ||
calculated parameters: | ||
Clotting Analysis Parameters: | ||
Prothrombin Time (PT); Activated | ||
Partial Thromboplastin Time | ||
(APTT); Fibrinogen (Clauss); | ||
Batroxobin Time; Extrinsic Factors | ||
(II, V, VII, X); Intrinsic Factors | ||
(VIII, IX, XI, XII); Protein C. | ||
Chromogenic Analysis | ||
Parameters: Antithrombin III; | ||
Factor VIII; Plasminogen; | ||
Heparin; Protein C; α2- | ||
Antiplasmin. | ||
Immunologic Analysis Parameters: | ||
D-dimer. | ||
Calculated Parameters: PT Ratio; | ||
PT INR; PT %; Derived | ||
Fibrinogen; Factor Assays % | ||
Activity. | ||
Sample Type | Human plasma | |
3.2% sodium citrate | Same | |
Similarities between CS-2100i and CA-1500 | ||
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
Application type | Clotting Applications: | Same |
Prothrombin Time (PT) with | ||
Dade® Innovin®; | ||
Activated Partial Thromboplastin | ||
Time (APTT) with | ||
Dade® Actin® FSL; | ||
Fibrinogen (Clauss) with | ||
Dade® Thrombin Reagent; | ||
Coagulation Factor V with | ||
Dade® Innovin® | ||
Coagulation Factor VII with | ||
Dade® Innovin® | ||
Protein C with Protein C Reagent | ||
Chromogenic Application: | ||
Antithrombin with | ||
INNOVANCE® Antithrombin; | ||
Protein C with | ||
Berichrom® Protein C | Same | |
Immuno-Chemical Application: | ||
D-dimer with | ||
INNOVANCE® D-Dimer | Same | |
Calculated Application: | ||
PT INR with Dade® Innovin® | Same | |
Clinical Reportable Range | Fibrinogen with Dade® Thrombin | |
Reagent 50 to 860 mg/dL; | Same | |
Antithrombin with INNOVANCE® | ||
Antithrombin 9.0 to 128% of norm; | ||
D-dimer with INNOVANCE® D- | ||
dimer 0.19 to 35.2 mg/L FEU; | ||
Coagulation Factor V with Dade® | ||
Innovin® 6.0 to 149.0% of norm; | ||
Coagulation Factor VII with Dade® | ||
Innovin® 6.0 to 149.0% of norm; | ||
Protein C with Protein C Reagent | ||
10.1 to 131.0% of norm | ||
Specimen Processing | Automatic Pipetting and Dilution | Same |
Similarities between CS-2100i and CA-1500 | ||
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
Random Access | Yes | Same |
Liquid Level Sensing | Yes - reagent and sample | Same |
Bar Code Reader | Sample and reagent | Same |
STAT Testing | Yes | Same |
Sampling Capabilities | Normal and Micro Mode | Same |
Sample Volumes (Plasma) | PT with Dade® Innovin® (50 µL) | |
APTT with Dade® Actin® FSL | ||
(50 µL) | ||
Fibrinogen with Dade® Thrombin | ||
Reagent (10 µL) | ||
Coagulation Factor V with Dade® | ||
Innovin® (5 µL) | ||
Coagulation Factor VII with Dade® | ||
Innovin® (5 µL) | ||
Protein C with Protein C Reagent | ||
(5 µL) | ||
Protein C with Berichrom® | ||
Protein C (5 µL) | Same | |
Similarities between CS-2100i and CA-1500 | ||
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
Sample Volumes in Micro Mode | ||
(Plasma) | PT with Dade® Innovin® (50 µL) | |
APTT with Dade® Actin® FSL | ||
(50 µL) | ||
Fibrinogen with Dade® Thrombin | ||
Reagent (10 µL) | ||
Coagulation Factor V with Dade® | ||
Innovin® (5 µL) | ||
Coagulation Factor VII with Dade® | ||
Innovin® (5 µL) | ||
Protein C with Protein C Reagent | ||
(5 µL) | ||
Protein C with Berichrom® | ||
Protein C (15 µL) | Same | |
Rinse & Buffer Solutions | ||
On-board | ||
External | CA-CLEAN I | |
CA-CLEAN II | ||
Dade® Owren's Buffer | ||
Water | Same | |
Light Source | ||
Chromogenic | ||
Immuno-chemical | Halogen Lamp | |
Halogen Lamp | Same | |
Same | ||
Probes | 1 Sample probe; | |
1 Reagent probe | Same | |
Wavelengths used in Analysis | Antithrombin with | |
INNOVANCE® Antithrombin | ||
(405 nm) | ||
Coagulation Factor V with | ||
Dade® Innovin® | ||
(Default = 660 nm; | ||
sub-wavelength = none) | ||
Coagulation Factor VII with Dade® | ||
Innovin® | ||
(Default = 660 nm; Sub-wavelength= | ||
none) | ||
Protein C with Protein C Reagent | ||
(Default = 660 nm; Sub-wavelength= | Same | |
Similarities between CS-2100i and CA-1500 | ||
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
none) | ||
Protein C with | ||
Berichrom® Protein C | ||
(Default = 405 nm; Sub-wavelength= | ||
none) | ||
Temperature Control | Sample incubation well: | |
37 °C ± 1.0 °C | Same | |
Differences between CS-2100i and CA-1500 | ||
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
Clinical Reportable Range | Protein C with Berichrom® Protein C | |
10.0 to 138% of norm | 5.0 to 138% of norm | |
Operating Principle | Clotting: | |
Transmitted Light Detection | ||
(Absorbance) at 340, 405, 575, 660 or | ||
800 nm. Wavelengths 340, 405 and | ||
575 are technically available but not | ||
validated in combination with the | ||
intended applications. | Scattered Light Detection at 660 | |
nm | ||
Chromogenic: | ||
Transmitted Light Detection | ||
(Absorbance) at 340, 405, 575, 660, | ||
800 nm. Wavelengths 340, 575, 660, | ||
and 800 are technically available but not | ||
validated in combination with the | ||
intended applications. | Transmitted Light Detection | |
(Absorbance) at 405, 575, 800 nm | ||
Immunochemical: | ||
Transmitted Light Detection | ||
(Absorbance) at 340, 405, 575, 660 or | ||
800 nm. Wavelengths 340, 405, 575, | ||
and 800 are technically available but not | ||
validated in combination with the | ||
intended applications. | Transmitted Light Detection | |
(Absorbance) at 405, 575, or 800 | ||
nm | ||
Wavelengths* used in Analysis | ||
*The default wavelength is | ||
normally used to generate the | ||
reported value of the measurement. | ||
The sub-wavelength is run in | PT (seconds) Dade® Innovin® | |
(Default = 660 nm; Sub- Wavelength= | ||
800 nm) | PT (seconds) with Dade® | |
Innovin® Default = 660 nm; Sub- | ||
Wavelength= none) | ||
parallel. If a light intensity error | ||
occurs by using the default | ||
wavelength the value from the sub- | ||
wavelength is used automatically. | PT (INR) with Dade® Innovin® | |
(Default = 660 nm; Sub-Wavelength= | ||
800 nm) | PT (INR) with Dade® Innovin® | |
(Default = 660 nm; Sub- | ||
Wavelength= none) | ||
APTT with Dade® Actin® FSL | ||
Activated PTT Reagent (Default = 660 | ||
nm; Sub-Wavelength= 800 nm) | APTT with Dade® Actin® FSL | |
Activated PTT Reagent (Default = | ||
660 nm; Sub-Wavelength= none) | ||
Differences between CS-2100i and CA-1500 | ||
Analyzer Component | Proposed Device | |
Sysmex® CS-2100i | Predicate Device | |
Sysmex® CA-1500 | ||
Light Source | ||
Clotting | Halogen Lamp | Light Emitting Diode |
Cap Piercing | Cap Piercer only | Both Cap Piercer model and Non- |
Cap Piercer models are available | ||
Temperature Control | Detector: 37 ± 0.5 °C | |
Reagent probe: 37.5 ± 0.5 °C | Detector: 37 ± 1.0 °C | |
Reagent probe: 37 ± 1.0 °C | ||
Reagent Cooling | 10 ± 2 °C, when ambient | |
temperature is 20 – 28 °C. | ||
During operation 4 – 15 °C, when | ||
ambient temperature is 15 – 30 °C | 15 ± 2 °C, when ambient | |
temperature is 15 - 30 °C | ||
Pipetting Capabilities | Reagent probe: | |
20 – 200 μL | ||
Sample probe: | ||
4 – 270 μL | Reagent probe: | |
3 – 200 μL | ||
Sample probe: | ||
5 – 450 μL | ||
Sample Volumes (Plasma) | Antithrombin with | |
INNOVANCE® Antithrombin | ||
(14 μL) | ||
D-dimer with INNOVANCE® | ||
D-Dimer (15 μL) | Antithrombin with | |
INNOVANCE® Antithrombin | ||
(10 μL) | ||
D-dimer with INNOVANCE® | ||
D-Dimer (13 μL) |
Similarities between CS-2100i and CA-1500
11
12
13
14
15
Differences between Sysmex® CS-2100i and Sysmex® CA-1500
16
The above described differences do not raise new questions as to safety and effectiveness of the new device.
17
7 Performance Data
The following performance data were provided in support of the substantial equivalence determination.
7.1 Method comparison
Method comparison studies designed according to EP09-A3 CLSI Guideline 'Measurement Procedure Comparison and Bias Estimation Using Patient Samples' were conducted at four external sites in the United States, all sites using the same protocol.
Samples were measured on both the predicate device (Sysmex® CA-1500) as well as the new device (Sysmex® CS-2100i), in random order to eliminate any inherent bias. Results were compared by Passing-Bablok regression analysis as well as Bland-Altman plots. Results from each application met the predetermined acceptance criteria. The following summary of Passing-Bablok regression from the multi-center study shows that the proposed and predicate devices provide equivalent results when used in a clinical setting.
| Application | N | Sample range
(% of norm) | r | Slope
(95% CI) | Intercept
(95% CI) |
|-------------------------------------|-----|-----------------------------|-------|-------------------------|---------------------------|
| Factor V with Dade® Innovin® | 589 | 7.2 - 145.1 | 0.983 | 0.990
(0.977, 1.003) | 1.861
(1.025, 2.784) |
| Factor VII with Dade® Innovin® | 492 | 6.2 - 148.8 | 0.991 | 1.039
(1.028, 1.051) | 0.636
(-0.135, 1.104) |
| Protein C with Protein C Reagent | 584 | 10.3 - 129.3 | 0.977 | 0.966
(0.952, 0.981) | -0.372
(-1.207, 0.470) |
| Protein C with Berichrom® Protein C | 507 | 11.3 - 134.7 | 0.994 | 0.942
(0.934, 0.951) | -0.292
(-1.037, 0.399) |
7.2 Reproducibility Studies
Twenty-day precision studies were performed at two external sites in Germany and one external site in the United States. Testing followed the scheme of two runs per day, with two replicates per run, at each of the three sites according to CLSI EP05-A2 'Evaluation of Precision Performance of Quantitative Measurement Methods'. The order of the analysis of parameter, samples and quality control samples for each run and day varied to avoid an inherent bias to the study. One calibration curve of each calibrated application was used in the study. Within Run, Between Run, Between Day, and Total (within site) were calculated. The data is summarized in the following tables.
18
Sysmex® CS-2100i: Reproducibility Summary Table, Within Run | ||||
---|---|---|---|---|
Application | ||||
(measuring interval) | 1st Site | |||
Within Run | ||||
(%CV) | 2nd Site | |||
Within Run | ||||
(%CV) | 3rd Site | |||
Within Run | ||||
(%CV) | Sites Combined | |||
(%CV) | ||||
Coagulation Factor V with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 4.07 – 7.40 | 2.07 - 3.34 | 2.70 - 6.09 | 3.16 - 5.01 |
Coagulation Factor VII with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 1.74 - 2.38 | 1.71 – 2.87 | 1.99 - 2.59 | 1.85 - 2.63 |
Protein C with | ||||
Protein C Reagent | ||||
(10.1 - 131.0% of norm) | 2.26 - 3.75 | 2.08 - 2.68 | 2.43 - 7.02 | 2.46 - 4.78 |
Protein C with | ||||
Berichrom® Protein C | ||||
(10.0 - 138.0% of norm) | 1.02 - 5.34 | 1.00 - 4.56 | 1.20 - 5.26 | 1.15 - 4.63 |
19
Sysmex® CS-2100i: Reproducibility Summary Table, Between Run | ||||
---|---|---|---|---|
Application | ||||
(measuring interval) | 1st Site | |||
Between | ||||
Run | ||||
(%CV) | 2nd Site | |||
Between | ||||
Run | ||||
(%CV) | 3rd Site | |||
Between | ||||
Run | ||||
(%CV) | Sites Combined | |||
(%CV) | ||||
Coagulation Factor V with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 0.00 - 2.80 | 0.00 - 3.87 | 0.00 - 3.36 | 1.76 - 2.55 |
Coagulation Factor VII with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 0.65 - 1.41 | 0.00 - 1.35 | 0.00 - 2.32 | 0.00 - 1.72 |
Protein C with | ||||
Protein C Reagent | ||||
(10.1 - 131.0% of norm) | 0.00 - 1.65 | 0.85 - 2.70 | 0.00 - 2.50 | 0.56 - 1.95 |
Protein C with | ||||
Berichrom® Protein C | ||||
(10.0 - 138.0% of norm) | 0.54 - 2.51 | 0.32 - 2.23 | 0.00 - 1.20 | 0.84 - 1.66 |
Sysmex® CS-2100i: Reproducibility Summary Table, Between Day | ||||
---|---|---|---|---|
Application | ||||
(measuring interval) | 1st Site | |||
Between | ||||
Day | ||||
(%CV) | 2nd Site | |||
Between | ||||
Day | ||||
(%CV) | 3rd Site | |||
Between | ||||
Day | ||||
(%CV) | Sites Combined | |||
(%CV) | ||||
Coagulation Factor V with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 0.00 - 1.55 | 0.00 - 2.56 | 0.00 - 2.81 | 0.00 - 1.64 |
Coagulation Factor VII with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 0.00 - 1.02 | 1.48 – 2.68 | 0.00 - 1.92 | 0.89 - 1.67 |
Protein C with | ||||
Protein C Reagent | ||||
(10.1 - 131.0% of norm) | 0.00 - 2.41 | 0.00 - 1.20 | 0.00 - 2.60 | 0.00 - 2.17 |
Protein C with | ||||
Berichrom® Protein C | ||||
(10.0 - 138.0% of norm) | 0.00 - 0.57 | 0.00 - 1.02 | 0.00 - 2.36 | 0.00 - 1.40 |
20
CS-2100i: Reproducibility Summary Table, Total CV (Within Site and Sites Combined) | ||||
---|---|---|---|---|
Application | ||||
(measuring interval) | 1st Site | |||
Total CV | ||||
Within Site | ||||
(%CV) | 2nd Site | |||
Total CV | ||||
Within Site | ||||
(%CV) | 3rd Site | |||
Total CV | ||||
Within Site | ||||
(%CV) | Total CV | |||
Sites Combined | ||||
(%CV) | ||||
Coagulation Factor V with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 4.64 - 7.40 | 3.37 - 5.12 | 3.85 - 6.95 | 4.78 - 10.43 |
Coagulation Factor VII with | ||||
Dade® Innovin® | ||||
(6.0 - 149.0% of norm) | 2.01 - 2.81 | 2.58 – 3.56 | 2.23 - 3.31 | 2.57 - 3.59 |
Protein C with | ||||
Protein C Reagent | ||||
(10.1 - 131.0% of norm) | 2.78 - 4.55 | 2.36 - 3.83 | 3.47 - 7.49 | 3.47 - 5.85 |
Protein C with | ||||
Berichrom® Protein C | ||||
(10.0 - 138.0% of norm) | 1.40 - 5.91 | 1.40 - 5.18 | 1.70 - 5.26 | 2.19 - 5.69 |
21
Detection Capability Results 7.3
Detection capability studies were measured for the calibrated assays on the Sysmex® CS-2100i: Coagulation Factor V with Dade® Innovin®, Coagulation Factor VII with Dade® Innovin®, Protein C with Protein C Reagent, and Protein C with Berichrom® Protein C. Studies were conducted following the CLSI document EP17-A2 'Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures'. Data for all tested reagents met the predetermined acceptance criteria and support the lower limit of the clinically reportable range claim.
Sysmex® CS-2100i: Summary of Limit of Quantitation Studies | |||||
---|---|---|---|---|---|
Application | Lower Limit of | ||||
Clinically | |||||
Reportable | |||||
Range | |||||
(% of norm) | Measured Limit | ||||
of Quantitation | |||||
based on | |||||
predicate | |||||
(% of norm) | Maximum | ||||
Total Error | |||||
(% of norm) | |||||
Coagulation Factor V with Dade® | |||||
Innovin® | 6.0 | 4.80 | 2.04 | ||
Coagulation Factor VII with | |||||
Dade® Innovin® | 6.0 | 3.39 | 0.21 | ||
Protein C with | |||||
Protein C Reagent | 10.1 | 9.35 | 3.62 | ||
Protein C with | |||||
Berichrom® Protein C | 10.0 | 8.32 | 2.07 |
22
Linearity & Measuring Range 7.4
Linearity studies were performed for the calibrated assays on the Sysmex® CS-2100i: Coagulation Factor V with Dade® Innovin®, Coagulation Factor VII with Dade® Innovin®, Protein C with Protein C Reagent, and Protein C with Berichrom® Protein C. All reagents met the predetermined acceptance criteria and support the clinically reportable range claim. Studies were conducted as described in CLSI EP6-A 'Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach'.
Application | Measured Linear Range | Clinically Reportable Range |
---|---|---|
Coagulation Factor V with | ||
Dade® Innovin® | 3.4 – 180.7% of norm | 6.0 – 149.0% of norm |
Coagulation Factor VII with | ||
Dade® Innovin® | 4.3 – 179.5% of norm | 6.0 – 149.0% of norm |
Protein C with | ||
Protein C Reagent | 7.0 – 187.7% of norm | 10.1 – 131.0% of norm |
Protein C with | ||
Berichrom® Protein C | 7.1 – 181.3% of norm | 10.0 – 138.0% of norm |
7.5 Reference Interval
Reference interval studies were conducted at three clinical study sites in the United States following the guidance of CLSI document EP28-A3c 'Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory'. The summary is provided below. The study population did not include neonate and pediatric sample populations.
Sysmex® CS-2100i: Reference Interval Summary Table | ||
---|---|---|
Application | N | Sysmex® CS-2100i Reference Interval |
(5th Percentile of norm) | ||
Coagulation Factor V with | ||
Dade® Innovin® | 193 | 79.8% |
Coagulation Factor VII with | ||
Dade® Innovin® | 193 | 65.6% |
Protein C with | ||
Protein C Reagent | 193 | 75.6% |
Protein C with | ||
Berichrom® Protein C | 191 | 84.9% |
23
8 Conclusions
Because the predicate device was cleared based in part on the results of clinical studies, and because clinical settings are required for a well-validated device, clinical testing was required to support substantial equivalence.
The non-clinical data support the safety of the device.
The clinical data demonstrate that the Sysmex® CS-2100i performs comparably to the predicate device that is currently marketed for the same intended use.
The data submitted for this premarket notification demonstrates that the device raises no new concerns as to safety and effectiveness when compared to the predicate device, and is substantially equivalent to the predicate device.