(270 days)
VIDAS TOXO IgM Test System (K923166)
Not Found
No
The device description details a standard immunoassay process and mentions results being determined automatically by the Elecsys software by comparing signals to a cutoff value, which is a typical automated analysis method, not indicative of AI/ML. There are no mentions of AI, ML, or related concepts in the provided text.
No.
This device is an in vitro diagnostic immunoassay used for qualitative detection of antibodies, aiding in diagnosis. It does not provide any therapeutic intervention or treatment.
Yes
The "Intended Use / Indications for Use" section states that the assay "may be used as an aid in the diagnosis of an acute or recent Toxoplasma gondii infection in suspected patients and pregnant women." This directly indicates its role in diagnosis.
No
The device description clearly details a physical immunoassay process involving reagents, microparticles, and electrochemiluminescence detection, which are hardware components. While software is mentioned for result determination, it is integral to a larger hardware-based system.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The "Intended Use / Indications for Use" section explicitly states that the Elecsys Toxo IgM immunoassay is "for the in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma." The phrase "in vitro" is a key indicator of an IVD, meaning it's used to test samples outside of the body.
- Purpose: The assay is used as an "aid in the diagnosis of an acute or recent Toxoplasma gondii infection," which is a diagnostic purpose.
- Sample Type: It analyzes "human serum and plasma," which are biological samples.
- Device Description: The description details a laboratory-based immunoassay process involving reagents, microparticles, and detection methods, all typical of IVD devices.
- Quality Control: The mention of "PreciControl Toxo IgM is used for quality control" further supports its use in a clinical laboratory setting for diagnostic testing.
N/A
Intended Use / Indications for Use
The Elecsys Toxo IgM immunoassay is for the in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma. This assay may be used as an aid in the diagnosis of an acute or recent Toxoplasma gondii infection in suspected patients and pregnant women. Patient testing must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immuno-assay analyzers. This assay has not been cleared by the FDA for blood/plasma donor screening.
PreciControl Toxo IgM is used for quality control of the Elecsys Toxo IgM immunoassay on the Elecsys and cobas e immunoassay analyzers.
Product codes (comma separated list FDA assigned to the subject device)
LGD, JJX
Device Description
The Elecsys Toxo IgM is a two-step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. In the first incubation 10 uL of sample are automatically prediluted 1:20 with Elecsys Diluent Universal and T. gondii-specific recombinant antigen labeled with a ruthenium complex is added. Anti-Toxo IgM antibodies present in the sample react with the ruthenium-labeled T. gondii-specific recombinant antigen. Then biotinylated monoclonal human-IgM- specific antibodies and streptavidin-coated microparticles are added. The complex becomes bound to the solid phase via interaction of biotin and streptayidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results are determined automatically by the Elecsys software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by Toxo IgM calibration.
The reagent working solutions include:
- rackpack (kit placed on instrument)
- Streptavidin coated microparticles, -
- Reagent 1 (Toxoplasma-Ag~ Ru(bpy)2+3) and -
- Reagent 2 (Anti human IgM Ab~biotin). -
- Included with the kit: -
- . Calibrator 1 (negative calibrator) Human serum non-reactive for anti Toxo IgM
- . Calibrator 2 (positive calibrator) Human serum reactive for anti Toxo IgM
PreciControl Toxo IgM is a ready for use control serum based on human serum. The controls are used for monitoring the accuracy of the Elecsys Toxo IgM immunoassay.
The PreciControl Toxo IgM includes:
- PC Toxo IgM 1 (Human serum negative for Toxo IgM antibodies, preservatives) .
- PC Toxo IgM 2 (Human serum positive for Toxo IgM antibodies, preservatives) .
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
The patient's age ranged from 2 to 86.
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical Performance Evaluation
- Precision: Determined at one site using Elecsys reagents, human sera and controls, 6 times daily for 10 days (n = 60). Results showed repeatability, between-run, and intermediate precision with CVs ranging from 1.5% to 5.2%.
- Reproducibility: Determined at two sites using Elecsys reagents and human sera, 6 times daily for 10 days at each site (n = 120). Results showed repeatability, between-run, intermediate precision, and between-site precision, with total precision CVs ranging from 4.9% to 13.6%.
- Cross-reactivity: 218 potentially cross-reacting samples were tested. Most showed 0% cross-reactivity. Some notable cross-reactivity percentages were AMA (6.7%), EBV (10%), HBV (4.8%), Malaria (8%), and Rubella (10%).
- Matrix equivalency study: Evaluated serum/gel separation tubes, lithium heparin plasma, K2-EDTA plasma, and sodium citrate plasma. A minimum of 49 sample pairs were drawn into serum and plasma collection tubes. All results met acceptance criteria.
- Limitations - interference: The assay is unaffected by Bilirubin (
§ 866.3780
Toxoplasma gondii serological reagents.(a)
Identification. Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toToxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyToxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoanToxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines connecting them. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2017
ROCHE DIAGNOSTICS KELLI TÜRNER REGULATORY AFFAIRS PRINCIPAL 9115 HAGUE RD. INDIANAPOLIS IN 46250
Re: K162678
Trade/Device Name: Elecsys Toxo IgM. Elecsys Toxo IgM PreciControl Regulation Number: 21 CFR 866.3780 Regulation Name: Toxoplasma gondii serological reagents Regulatory Class: II Product Code: LGD, JJX Dated: May 19, 2017 Received: May 22, 2017
Dear Ms. Turner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ribhi Shawar -S For
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K162678
Device Name Elecsys Toxo IgM Immunoassay
Indications for Use (Describe)
Elecsys Toxo IgM Immunoassay
The Elecsys Toxo IgM immunoassay is for the in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma. This assay may be used as an aid in the diagnosis of an acute or recent Toxoplasma gondii infection in suspected patients and pregnant women. Patient testing must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay.
The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immuno-assay analyzers. This assay has not been cleared by the FDA for blood/plasma donor screening.
PreciControl Toxo IgM
PreciControl Toxo IgM is used for quality control of the Elecsys Toxo IgM immunoassay on the Elecsys and cobas e immunoassay analyzers.
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Elecsys Toxo IgM 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
Submitter Name | Roche Diagnostics |
---|---|
Address | 9115 Hague Road |
Indianapolis, IN 46250 | |
Contact | Kelli Turner |
Phone: (317) 521-4515 | |
FAX: (317) 521-3080 | |
Email: kelli.turner@roche.com | |
Date Prepared | June 21, 2017 |
Proprietary Name | 1) Elecsys Toxo IgM |
- PreciControl Toxo IgM |
| Common Name | 1) Toxo IgM - PreciControlToxo IgM |
| Classification Name | 1) Toxoplasma gondii serological reagents - Single (specified) analyte controls (assayed and unassayed) |
| Product Codes | 1) LGD, 866.3780 - JJX, 862.1660 |
| Predicate Device | 1) VIDAS TOXO IgM Test System (K923166) |
| Establishment Registration | Roche Diagnostics GmbH in Mannheim, Germany, is 9610126
Roche Diagnostics GmbH in Penzberg, Germany, is 9610529
Roche Diagnostics in the United States is 1823260 |
4
1. DEVICE DESCRIPTION
The Elecsys Toxo IgM is a two-step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection. In the first incubation10 uL of sample are automatically prediluted 1:20 with Elecsys Diluent Universal and T. gondii-specific recombinant antigen labeled with a ruthenium complex is added. Anti-Toxo IgM antibodies present in the sample react with the ruthenium-labeled T. gondii-specific recombinant antigen. Then biotinylated monoclonal human-IgM- specific antibodies and streptavidin-coated microparticles are added. The complex becomes bound to the solid phase via interaction of biotin and streptayidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results are determined automatically by the Elecsys software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by Toxo IgM calibration.
Reagents 1.1.
The reagent working solutions include:
- rackpack (kit placed on instrument)
- Streptavidin coated microparticles, -
- Reagent 1 (Toxoplasma-Ag~ Ru(bpy)2+3) and -
- Reagent 2 (Anti human IgM Ab~biotin). -
- Included with the kit: -
- . Calibrator 1 (negative calibrator) Human serum non-reactive for anti Toxo IgM
- . Calibrator 2 (positive calibrator) Human serum reactive for anti Toxo IgM
5
1.2. Control
PreciControl Toxo IgM is a ready for use control serum based on human serum. The controls are used for monitoring the accuracy of the Elecsys Toxo IgM immunoassay.
The PreciControl Toxo IgM includes:
- PC Toxo IgM 1 (Human serum negative for Toxo IgM antibodies, preservatives) .
- PC Toxo IgM 2 (Human serum positive for Toxo IgM antibodies, preservatives) .
2. INDICATIONS FOR USE
The Elecsys Toxo IgM immunoasay is for the in vitro qualitative detection of IgM antibodies to Toxoplasma gondii in human serum and plasma. This assay may be used as an aid in the diagnosis of an acute or recent Toxoplasma gondii infection in suspected patients and pregnant women. Patient testing must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers. This assay has not been cleared by the FDA for blood/plasma donor screening.
Comparison with the predicate device
The Elecsys Toxo IgM immunoasay was compared to a commercially marketed kit VIDAS TOXO IgM Test System (K923166) by Biomerieux. Both kits have the same intended use and use the same methodology. Below are tables comparing the reagents and the procedures of the two devices.
Similarities | ||
---|---|---|
Item | Device | Predicate |
Intended Use | The Elecsys Toxo IgM | |
immunoassay is for the in | ||
vitro qualitative detection of | ||
IgM antibodies to | ||
Toxoplasma gondii in human | ||
serum and plasma. This | ||
assay may be used as an aid | ||
in the diagnosis of an acute | ||
or recent Toxoplasma gondii | The VIDAS® TOXO IgM | |
(TXM) assay is intended | ||
for use with a VJDAS® | ||
(Vitek ImmunoDiagnostic | ||
Assay System) instrument | ||
as an automated enzyme- | ||
linked fluorescent | ||
immunoassay (ELFA) for | ||
the presumptive qualitative | ||
Similarities | ||
Item | Device | Predicate |
infection in suspected | ||
patients and pregnant | ||
women. Patient testing must | ||
be performed in conjunction | ||
with an anti- Toxoplasma | ||
gondii IgG antibody assay. | ||
The | ||
electrochemiluminescence | ||
immunoassay “ECLIA” is | ||
intended for use on Elecsys | ||
and cobas e immuno-assay | ||
analyzers. This assay has not | ||
been cleared by the FDA for | ||
blood/plasma donor | ||
screening. | ||
PreciControl Toxo IgM is | ||
used for quality control of | ||
the Elecsys Toxo IgM | ||
immunoassay on the Elecsys | ||
and cobas e immunoassay | ||
analyzers. | detection of anti- | |
Toxoplasma gondii IgM | ||
antibodies in human serum, | ||
as an aid in the diagnosis of | ||
acute, recent, or reactivated | ||
Toxoplasma gondii | ||
infection. This assay must | ||
be performed in | ||
conjunction with an anti- | ||
Toxoplasma gondii IgG | ||
antibody assay. VIDAS | ||
TXM assay performance | ||
has not been established for | ||
prenatal screening or | ||
newborn testing. This assay | ||
has not been cleared by the | ||
FDA for blood/plasma | ||
donor screening. | ||
Calibrator | Included with the kit | Same |
Assay type | Sandwich ELISA | Same |
Instrument Platform | Fully automated system | Same |
Differences | ||
Item | Device | Predicate |
Assay Protocol | Electrochemiluminescence | |
Immunoassay (ECLIA) | Enzyme-linked fluorescent | |
immunoassay (ELFA) | ||
Sample type | Serum, lithium heparin | |
plasma, potassium EDTA | ||
plasma and sodium citrate | ||
plasma | Serum | |
Calibrators | Two | One |
6
NON-CLINICAL PERFORMANCE EVALUATION 3.
3.1. Precision
Representative performance data on the analyzers are given below. Results obtained in individual laboratories may differ.
7
Precision was determined at one site using Elecsys reagents, human sera and controls in a modified protocol (EP5-A) of the CLSI (Clinical and Laboratory Standards Institute): 6 times daily for 10 days (n = 60). The following results were obtained:
| | | | Repeatability | | Between-run | | Intermediate
precision | |
|-----------------|----|-------------|---------------|--------|-------------|--------|---------------------------|--------|
| Sample | N | Mean [s/co] | SD [s/co] | CV (%) | SD [s/co] | CV (%) | SD [s/co] | CV (%) |
| PCd) Toxo IgM 1 | 60 | 0.146 | 0.004 | 2.5 | 0.003 | 1.9 | 0.005 | 3.2 |
| PC Toxo IgM 2 | 60 | 1.61 | 0.037 | 2.3 | 0.045 | 2.8 | 0.059 | 3.6 |
| Serum Pool 1 | 60 | 0.147 | 0.003 | 2.3 | 0.005 | 3.1 | 0.006 | 3.9 |
| Serum Pool 2 | 60 | 2.06 | 0.075 | 3.7 | 0.031 | 1.5 | 0.082 | 4.0 |
| Serum Pool 3 | 60 | 3.49 | 0.173 | 4.9 | 0.060 | 1.7 | 0.183 | 5.2 |
d) PC = PreciControl
Note: intermediate precision is within lab precision data.
Reproducibility 3.2.
Representative performance data on the analyzers are given below. Results obtained in individual laboratories may differ.
Reproducibility was determined at two sites using Elecsys reagents and human sera in a modified protocol (EP5-A) of the CLSI (Clinical and Laboratory Standards Institute): 6 times daily for 10 days at each site (n = 120). The following results were obtained:
Repeatability | Between-run | Intermediate precision | ||||||||
---|---|---|---|---|---|---|---|---|---|---|
Sample | N | Mean [s/co] | SD [s/co] | CV (%) | SD [s/co] | CV (%) | SD [s/co] | CV (%) | ||
Serum Pool 1 | 120 | 0.161 | 0.003 | 1.9 | 0.006 | 3.5 | 0.006 | 4.0 | ||
Serum Pool 2 | 120 | 2.01 | 0.065 | 3.3 | 0.03 | 1.5 | 0.072 | 3.6 | ||
Serum Pool 3 | 120 | 3.37 | 0.135 | 4.0 | 0.073 | 2.2 | 0.153 | 4.6 | ||
Between site | Total precision | |||||||||
Sample | N | Mean [s/co] | SD [s/co] | CV (%) | SD [s/co] | CV (%) | ||||
Serum Pool 1 | 120 | 0.161 | 0.021 | 13.0 | 0.022 | 13.6 | ||||
Serum Pool 2 | 120 | 2.01 | 0.067 | 3.3 | 0.098 | 4.9 | ||||
Serum Pool 3 | 120 | 3.37 | 0.174 | 5.2 | 0.232 | 6.9 |
8
3.3. Cross-reactivity
A total of 218 potentially cross-reacting samples were collected and tested with the Elecsys Toxo IgM | |
---|---|
assay. The results are shown in the table below. |
| Antibody or Disease
State | # Tested | Elecsys Toxo IgM
Positive | % Cross-reactivity |
|------------------------------|----------|------------------------------|--------------------|
| AMA | 15 | 1 | 6.7 |
| ANA | 26 | 0 | 0 |
| Chlamydia | 5 | 0 | 0 |
| CMV | 13 | 0 | 0 |
| EBV | 10 | 1 | 10 |
| Gonorrhea | 5 | 0 | 0 |
| HAV | 10 | 0 | 0 |
| HBV* | 21 | 0* | 4.8 |
| HCV | 12 | 0 | 0 |
| HIV | 13 | 0 | 0 |
| HSV | 9 | 0 | 0 |
| Influenza | 16 | 0 | 0 |
| Malaria* | 25 | 1* | 8 |
| Parvo B19 | 11 | 0 | 0 |
| Rubella* | 10 | 0* | 10 |
| Syphilis | 5 | 0 | 0 |
| TPHA | 4 | 0 | 0 |
| VZV | 8 | 0 | 0 |
*contains 1 HBV sample repeatedly equivocal by Elecsys and negative with reference, 1 Malaria sample repeatedly equivocal by Elecsys and positive with reference, 1 Rubella sample reactive in Elecsys and initially equivocal with reference (no repetition possible due to low sample volume). Equivocal samples are counted as positive.
9
3.4. Matrix equivalency study
Studies were conducted to evaluate the suitability of the following 5 types of blood collectin tubes: serum/gel separation tubes, lithium heparin plasma. K2-EDTA plasma and sodium citrate plasma. A minimum of 49 sample pairs were drawn into serum and plasma collection tubes. The recovery of the analyte in the presence of the anti coagulant was compared to serum. All results met the acceptance criteria.
Limitations - interference 3.5.
The assay is unaffected by:
Compound | Concentration |
---|---|
Bilirubin | 0.5 COI recovery ± 20 %. |
Samples should not be taken from patients receiving therapy with high biotin doses (i.e. >5 mg/day) until at least 8 hours following the last biotin administration.
No interference was observed from rheumatoid factors up to a concentration of 3720 IU/mL. The high-dose hook effect does not lead to false-negative results in the Elecsys Toxo IgM assay. In vitro tests were performed on 18 commonly used pharmaceuticals and in addition on spiramycine, sulfadiazine, folinic acid and pyrimethamine. No interference with the assay was found.
As with many u-capture assays an interference with unspecific IgM is observed. Increasing amounts of unspecific IgM may lead to a decrease in the recovery of positive samples with the Elecsys Toxo IgM assay.
In rare cases, interference due to extremely high titers of antibodies to immunological components, streptavidin or ruthenium can occur. These effects are minimized by suitable test design.
For diagnostic purposes, the results should always be assessed in conjunction with the patient's medical history, clinical examination and other findings.
*COI= cut off index
10
3.6. Hook Effect
For verification of this claim and due to the fact that spiking of the analyte is not possible, three high positive serum samples (single donors), with a COI value of 30.5, 31.9 and 64.5 COI, (based on the measured Elecsys Toxo IgM result) were used for this experiment. It was diluted with native human negative serum in decreasing gradients. All dilution steps were measured in three-fold determination. The median of the signals for each dilution is calculated and, in relation to the dilution factor shown for each dilution in a diagram.
Acceptance criterion: For the Elecsys Toxo IgM assay it is claimed that High-dose Hook effect does not lead to false-negative results
Assay Cut-off 3.7.
The cut-off for the Elecsys Toxo IgM immunoassay was established by the use of sample collectives characterized with commercially available assays.
The classification of samples, based on the cut-off establishment, verification, and validation, is as follows:
- Sample results 1.0 COI = Reactive sample .
CLINICAL PERFORMANCE EVALUATION 4.
Summary of clinical performance
To characterize the performance of the Elecsys Toxo IgM immunoassay, relative to an FDA cleared method, multicenter studies in US and EU were performed. In total, 601 samples were tested as detailed below.
11
4.1. Study 1
A total of 440 prospective specimens were obtained from US reference laboratory. All specimens represented samples for whom anti Toxo testing had been ordered per clinical routine. The patient population consisted of 162 males and 278 females. There were 158 pregnant women among the females. The patient's age ranged from 2 to 86. The performance of the Elecsys Toxo IgM immunoassay, relative to an FDA cleared method is presented in the table below.
Reference Toxo IgM Method | |||||
---|---|---|---|---|---|
Elecsys | |||||
Toxo IgM | Positive | Equivocal | Negative | Total | |
Reactive | 2 | 0 | 2 | 4 | |
Equivocal | 0 | 0 | 4 | 4 | |
Non-reactive | 0 | 1 | 431 | 432 | |
Total | 2 | 1 | 437 | 440 |
US prospectively collected overall patient population
| Agreement Classification | Numerator/
Denominator | Percent
Agreement | 95 % Confidence
Interval |
|--------------------------|---------------------------|----------------------|-----------------------------|
| Positive agreement | 2/3 | 66.7 | 9.43 - 99.2 |
| Negative agreement | 431/437 | 98.6 | 97.0 - 99.5 |
Equivocals are counted as discrepant results against the Elecsys.
4.2. Study 2
A total of 60 retrospective samples were obtained from a US commercial sample vendor. The patient population consisted of 4 males and 56 females. There were 50 pregnant women among the females. All specimens represented samples from whom anti-Toxo testing had been ordered per clinical routine. The performance of the Elecsys Toxo IgM immunoassay, relative to an FDA cleared method is presented in the table below.
US retrospectively collected overall patient population
Reference Toxo IgM Method | |
---|---|
--------------------------- | -- |
12
| Elecsys
Toxo IgM | Positive | Equivocal | Negative | Total | |
---|---|---|---|---|---|
Reactive | 0 | 0 | 0 | 0 | |
Equivocal | 0 | 0 | 1 | 1 | |
Non-reactive | 0 | 0 | 59 | 59 | |
Total | 0 | 0 | 60 | 60 |
| Agreement Classification | Numerator/
Denominator | Percent
Agreement | 95 % Confidence
Interval |
|--------------------------|---------------------------|----------------------|-----------------------------|
| Positive agreement | 0/0 | NA | NA |
| Negative agreement | 59/60 | 98.3 | 91.1 - 100 |
4.3. Study 3
One hundred one (101) samples collected prospectively were obtained from the general population in Europe under suspicion of Toxoplasmosis consisting of 5 males and 96 females with an age range of 22 to 69. The agreement results are shown below.
EU prospectively collected overall patient population
Reference Toxo IgM Method | |||||
---|---|---|---|---|---|
Elecsys | |||||
Toxo IgM | Positive | Equivocal | Negative | Total | |
Reactive | 81 | 1 | 5 | 87 | |
Equivocal | 0 | 1 | 1 | 2 | |
Non-reactive | 4 | 1 | 7 | 12 | |
Total | 85 | 3 | 13 | 101 |
| Agreement Classification | Numerator/
Denominator | Percent
Agreement | 95 % Confidence
Interval |
|--------------------------|---------------------------|----------------------|-----------------------------|
| Positive agreement | 81/86 | 94.2 | 87.0 - 98.1 |
| Negative agreement | 7/14 | 50.0 | 23.0 - 77.0 |
13
Equivocals are counted as discrepant results against the Elecsys. In case of unanimously equivocal samples, samples will not be considered for agreement calculation.
Of the 101 prospectively collected European specimens described above, 87 were from pregnant women. The agreement results are shown below.
Reference Toxo IgM Method | |||||
---|---|---|---|---|---|
Elecsys | |||||
Toxo IgM | Positive | Equivocal | Negative | Total | |
Reactive | 68 | 0 | 5 | 73 | |
Equivocal | 0 | 1 | 1 | 2 | |
Non-reactive | 4 | 1 | 7 | 12 | |
Total | 72 | 2 | 13 | 87 |
EU prospectively collected pregnant women
| Agreement Classification | Numerator/
Denominator | Percent
Agreement | 95 % Confidence
Interval |
|--------------------------|---------------------------|----------------------|-----------------------------|
| Positive agreement | 68/73 | 93.2 | 84.7 - 97.7 |
| Negative agreement | 7/13 | 53.9 | 25.1 - 80.8 |
Equivocals are counted as discrepant results against the Elecsys. In case of unanimously equivocal samples, samples will not be considered for agreement calculation.
CONCLUSIONS 5.
The information provided in this Premarket Notification [510(k)] shows that the device Elecsys Toxo IgM is safe and effective and will support a substantial equivalence determination to the predicate device, VIDAS TOXO IgM Test System (K923166).