K Number
K162641

Validate with FDA (Live)

Date Cleared
2017-06-19

(270 days)

Product Code
Regulation Number
888.3310
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The R3™ Constrained Liner Acetabular System is an uncemented prosthesis intended to replace a hip joint. The Constrained Liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained Liner is intended for single use only.

Device Description

Subject of this Traditional premarket notification is the addition of new liner sizes to the Smith & Nephew R3" Constrained Liner Acetabular System. The R3 Constrained Liner Acetabular System currently consists of the R3 Constrained Liner offered in sizes of 52mm to 66/70mm cleared via premarket notification K111635. The proposed devices are additional sizes to the R3 Constrained Liners that were previously cleared. The proposed devices will be offered in sizes of 48mm, 50mm, 72-74mm, and 76-80mm, and will be marketed as part of the R3 Constrained Liner Acetabular System.

The new R3 Constrained Liner sizes are line additions to the primary predicate R3 Constrained Liners cleared via premarket notification K111635. The subject devices are similar to the primary predicate R3 Constrained Liners in that they are a multipiece acetabular component consisting of a bipolar bearing that articulates with a captured outer polyethylene liner, locking ring, and outer poly support ring. The subject devices and the primary predicate R3 Constrained Liners utilize a tripolar design allowing motion at two interfaces: the femoral head with the bipolar bearing and the bipolar bearing with the outer liner. The subject devices are identical in design to the cleared R3 Constrained Liners, with the exception that they are smaller or larger in overall size. The new R3 Constrained Liner sizes are also similar to the following additional predicates: R3 Constrained Liners (K122139), REFLECTION Constrained Liners (K021803, K033442), R3 XLPE Liners (K113848), and DePuy Self-Centering Hip Prostheses (K033273). The subject devices are also similar to reference predicate Osteonics Constrained Acetabular Inserts (P960047). The subject devices utilize seven components manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE), forged Titanium Alloy (Ti-6Al-4V), and Cobalt Chrome (CoCr).

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device, specifically the Smith & Nephew, Inc. R3TM Constrained Liner. This type of submission is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device, rather than proving its safety and effectiveness from scratch through clinical studies. Therefore, much of the information requested in a typical AI/ML medical device study (e.g., sample size for training/test sets, expert qualifications, MRMC studies, standalone performance) is not applicable or not provided in this context.

However, I can extract information related to the acceptance criteria and the type of study conducted to demonstrate equivalence for this specific device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (What was measured)Reported Device Performance (How the device performed)
Attachment loads: Push-in, Pull-out, Shuck-out, Push-outMet (Demonstrated no new issues)
Dislocation resistanceMet (Demonstrated no new issues)
Range of Motion (ROM)Met (Evaluated, demonstrated no new issues)
Pyrogen limit specification (≤ 20 EU/device)Met (Using Limulus Amebocyte Lysate (LAL) test)

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not explicitly stated as this was mechanical and LAL testing, not a clinical study on a patient population. The document refers to "the subject devices" and "the new R3 Constrained Liner sizes," implying a representative number of physical devices were tested.
  • Data Provenance: The testing was conducted internally by the manufacturer, Smith & Nephew, Inc. The document does not specify a country of origin for the data beyond that it's for a US regulatory submission. The studies were prospective in the sense that they were designed and executed to evaluate the new device sizes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable. The "ground truth" for mechanical and LAL testing is the established engineering specifications and pyrogen limits, not expert human assessment of images or clinical outcomes.

4. Adjudication method for the test set:

  • Not applicable. This device underwent mechanical and LAL testing, not human adjudication of results.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/ML device, and no MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI/ML device. The "standalone" performance here refers to the device's mechanical and biological properties independent of human interaction during testing.

7. The type of ground truth used:

  • Mechanical Testing: Engineering specifications and performance standards for hip prostheses.
  • Biological Testing: Established pyrogen limits (≤ 20 EU/device).

8. The sample size for the training set:

  • Not applicable. As a mechanical medical device, there is no "training set" in the context of AI/ML. The device's design is based on established biomechanical principles and materials science.

9. How the ground truth for the training set was established:

  • Not applicable. There is no training set. The "ground truth" for the device's design and manufacturing is derived from engineering design principles, materials science, and compliance with relevant industry standards for orthopedic implants.

Summary:

This 510(k) submission demonstrates substantial equivalence by showing that the new R3 Constrained Liner sizes have equivalent technological characteristics and performance to already legally marketed predicate devices. The "study" here involves mechanical testing (attachment loads, dislocation resistance, range of motion) and biological testing (Limulus Amebocyte Lysate (LAL) test for pyrogens). The primary goal was to ensure that the new, larger/smaller sizes of the liner did not introduce any new issues related to safety or effectiveness compared to the previously cleared sizes and other predicates. Clinical data was explicitly stated as "not needed to support substantial equivalence."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three overlapping profiles, suggesting a sense of community or collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the emblem.

June 19, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Smith & Nephew, Inc. Dongeun Kim Regulatory Affairs Specialist I 7135 Goodlett Farms Parkway Cordova, Tennessee 38016

Re: K162641 Trade/Device Name: Smith & Nephew, Inc. R3TM Constrained Liner Regulation Number: 21 CFR 888.3310 Regulation Name: Hip Joint Metal/Polymer Constrained Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWZ Dated: May 18, 2017 Received: May 19, 2017

Dear Dongeun Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162641

Device Name Smith & Nephew, Inc. R3TM Constrained Liner

Indications for Use (Describe)

The R3™ Constrained Liner Acetabular System is an uncemented prosthesis intended to replace a hip joint. The Constrained Liner is intended for primary or revision patients at hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained Liner is intended for single use only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Smith & Nephew. The logo consists of an orange flower-like symbol on the left, followed by the text "We are smith&nephew". The text "We are" is in gray, while "smith&nephew" is in orange, matching the color of the flower symbol. The logo is simple and clean, with a focus on the company name.

Submitted by:Smith & Nephew, Inc.Orthopaedic Division1450 Brooks RoadMemphis, Tennessee 38116
Date of Summary:June 19, 2017
Contact Person and Address:Dongeun KimRegulatory Affairs Specialist IT 901-399-1151F 901-566-7120
Name of Device:Smith & Nephew, Inc. R3TM Constrained Liner
Common Name:Acetabular Liner, Constrained
Device Classification Name andReference:21 CFR 888.3310 Hip Joint Metal/Polymer ConstrainedCemented or Uncemented Prosthesis
Device Class:Class II
Panel Code:Orthopaedics/87
Product Code:KWZ

Device Description

Subject of this Traditional premarket notification is the addition of new liner sizes to the Smith & Nephew R3" Constrained Liner Acetabular System. The R3 Constrained Liner Acetabular System currently consists of the R3 Constrained Liner offered in sizes of 52mm to 66/70mm cleared via premarket notification K111635. The proposed devices are additional sizes to the R3 Constrained Liners that were previously cleared. The proposed devices will be offered in sizes of 48mm, 50mm, 72-74mm, and 76-80mm, and will be marketed as part of the R3 Constrained Liner Acetabular System.

The new R3 Constrained Liner sizes are line additions to the primary predicate R3 Constrained Liners cleared via premarket notification K111635. The subject devices are similar to the primary predicate R3 Constrained Liners in that they are a multipiece acetabular component consisting of a bipolar bearing that articulates with a captured outer polyethylene liner, locking ring, and outer poly support ring. The subject devices and the primary predicate R3 Constrained Liners utilize a tripolar design allowing motion at two interfaces: the femoral head with the bipolar bearing and the bipolar bearing with the outer liner. The subject devices are identical in design to the cleared R3 Constrained Liners, with the exception that they are smaller or larger in overall size. The new R3 Constrained Liner sizes are also similar to the following additional predicates: R3 Constrained Liners (K122139), REFLECTION Constrained Liners (K021803, K033442), R3 XLPE Liners (K113848), and DePuy Self-Centering Hip Prostheses (K033273). The subject devices are also similar to reference predicate Osteonics Constrained Acetabular Inserts (P960047). The subject devices utilize seven components manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE), forged Titanium Alloy (Ti-6Al-4V), and Cobalt Chrome (CoCr).

{4}------------------------------------------------

Intended Use

The R3™ Constrained Liner Acetabular System is an uncemented prosthesis intended to replace a hip joint. The Constrained Liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation. bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained Liner is intended for single use only.

Technological Characteristics

Mechanical testing has been conducted for the subject devices to address the attachment loads (push-in, pull-out, shuck-out, and push-out) and dislocation resistance. The range of motion (ROM) has also been evaluated. A review of the testing has demonstrated that there are no new issues related to safety or effectiveness of the subject devices. Clinical data was not needed to support substantial equivalence.

The Limulus Amebocyte Lysate (LAL) test was used to determine the subject devices meets the specified pyrogen limit specification of less than or equal to 20 EU/device.

Substantial Equivalence Information

The intended use, indications for use, and technological characteristics of the subject R3 Constrained Liners are substantially equivalent to the predicate devices. A comparison of the subject devices to the predicate devices are provided in Table 5.1 below.

{5}------------------------------------------------

Design FeaturesSubject R3ConstrainedLiner510(k) NumberManufacturerIndications for Use/Intended UseR3 ConstrainedLiner (PrimaryPredicate)R3 ConstrainedLiner (AdditionalPredicate)REFLECTIONConstrainedLiner (AdditionalPredicate)R3 XLPE Liner(AdditionalPredicate)DePuy Self-CenteringHip Prosthesis(Additional Predicate)OsteonicsConstrainedAcetabularInsert(ReferencePredicate)
The R3Constrained LinerAcetabularSystem is anuncementedprosthesisintended toreplace a hip joint.The ConstrainedLiner is intendedfor primary orrevision patientsat high risk for hipdislocation due toa history of priordislocation, boneloss, soft tissuelaxity,neuromusculardisease, or intra-operativeinstability and forwhom all otheroptions toconstrainedacetabularcomponents haveK111635Smith & Nephew,Inc.The R3Constrained LinerAcetabularSystem is acemented oruncementedprosthesisintended toreplace a hip joint.The ConstrainedLiner is intendedfor primary orrevision patientsat high risk for hipdislocation due toa history of priordislocation, boneloss, soft tissuelaxity,neuromusculardisease, or intra-operativeinstability and forwhom all otheroptions toconstrainedacetabularThe R3Constrained LinerAcetabularSystem is acemented oruncementedprosthesisintended toreplace a hip joint.The ConstrainedLiner is intendedfor primary orrevision patientsat high risk for hipdislocation due toa history of priordislocation, boneloss, soft tissuelaxity,neuromusculardisease, or intra-operativeinstability and forwhom all otheroptions toconstrainedacetabularK122139Smith & Nephew,Inc.TheREFLECTIONConstrained LinerAcetabularSystem is acemented oruncementedprosthesisintended toreplace a hip joint.The ConstrainedLiner is intendedfor primary orrevision patientsat high risk for hipdislocation due toa history of priordislocation, boneloss, soft tissuelaxity,neuromusculardisease, or intra'operativeinstability and forwhom all otheroptions toconstrainedK021803,K033442Smith & Nephew,Inc.Hip componentsare indicated forindividualsundergoingprimary andrevision surgerywhere othertreatments ordevices havefailed inrehabilitating hipsdamaged as aresult of trauma ornoninflammatorydegenerative jointdisease (NIDJD)or any of itscompositediagnoses ofosteoarthritis,avascularnecrosis, traumaticarthritis, slippedcapital epiphysis,fused hip, fractureof the pelvis, anddiastrophic variantK113848Smith & Nephew,Inc.Indications for theUltimaTM Unipolarheads and the Self-CenteringTM HipProstheses previouslycleared for Hemi-HipArthroplasty remainunchanged. Additions tothe indications for useinclude:Self-CenteringTM HipProstheses and unipolarfemoral heads are alsointended to be used fortotal hip arthroplasty toprovide increased patientmobility and reduce painby replacing thedamaged hip jointarticulation in patientswhere there is evidenceof sufficient sound boneto seat and support thecomponents, when usedin conjunction with aUHMWPE bearingsurface having an insideK033273DePuy Orthopaedics,Inc.The OsteonicsConstrainedAcetabular Insertis indicated as acomponent of atotal hipprosthesis inprimary orrevision patientsat high risk of hipdislocation, boneloss, joint or softtissue laxity,neuromusculardisease orintraoperativeinstability.P960047HowmedicaOsteonics Corp.

ਣ のない 이후 പ് d Liner to the Constrainc rs ണ്ട ട് ത ਾਪੂ: rison of t .. ﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ hle

5-3

{6}------------------------------------------------

K162641
Size Offering(Outer Diameter)MaterialsSterilization
Constrained Lineris intended forsingle use only.The R3components havebeen considered.48mm, 50mm, 72-74mm, and 76-80mmUHMWPE, Ti-6Al-4V, CoCrEthylene Oxide
Constrained Lineris intended forsingle use only.The R3components havebeen considered.52mm to66/70mmUHMWPE, Ti-6Al-4V, CoCrEthylene Oxide
Constrained Lineris intended forsingle use only.The R3components havebeen considered.52mm to66/70mmUHMWPE, Ti-6Al-4V, CoCrEthylene Oxide
acetabularcomponents havebeen considered.46/48mm-70/76mmUHMWPE, Ti-6Al-4V, CoCrEthylene Oxide
Hip componentsare also indicatedfor inflammatorydegenerative jointdisease includingrheumatoidarthritis, arthritissecondary to avariety of diseasesand anomalies,and congenitaldysplasia;treatments ofnonunion, femoralneck fracture, andtrochantericfractures of theproximal femurwith headinvolvement thatare unmanageableusing othertechniques;endoprosthesis,femoralosteotomy, orGirdlestoneresection; fracture-dislocation of thehip; and correctionof deformity.22mm-36mmXLPEEthylene Oxide
diameter correspondingto outside diameter ofthe metallic cup that isutilized. Use in total hipreplacement is indicatedin the following additionalconditions:1. A severely painfuland/or disabled joint fromosteoarthritis, traumaticarthritis, rheumatoidarthritis, or congenital hipdysplasia2. Failed previous hipsurgery including jointreconstruction, internalfixation, arthrodesis,hemiarthroplasty, or totalhip replacement.39mm-57mmUHMWPE
50mm-74mmUHMWPE, Ti-6Al-4V, CoCr

Page 4 of 5

5-4

{7}------------------------------------------------

Conclusion

This Traditional 510(k) Premarket Notification is being submitted to request clearance for the new R3 Constrained Liner sizes of 48mm, 50mm, 72mm-74mm, and 76mm-80mm. Based on the similarities to the predicate devices and a review of the mechanical testing performed, the subject devices are substantially equivalent to the above predicate devices.

§ 888.3310 Hip joint metal/polymer constrained cemented or uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”