K Number
K162622
Date Cleared
2016-10-18

(28 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

Device Description

The modified Fujifilm Endoscope Models EC-600HL and EC-600LS are comprised of three main sections: an operation section, an insertion portion, and an umbilicus. The operation section controls the angulation (up/down/left/right) of the distal end of the endoscope. The insertion contains class fiber bundles, several channels and a complementary metal-oxide-semiconductor (CMOS) image sensor. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor. The endoscope also contains several channels to deliver air/water and provide suction, as well as a forceps channel. The forceps channel is used to introduce endoscope accessories such as biopsy forceps during the procedure. The umbilicus section consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source.

AI/ML Overview

The provided text is a 510(k) summary for the Fujifilm Endoscope Models EC-600HL and EC-600LS. This document describes modifications to existing devices and asserts substantial equivalence to predicate devices, rather than presenting a study for a novel device to meet specific acceptance criteria for a new clinical indication. Therefore, much of the requested information (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, MRMC study results, standalone performance) is not applicable or not provided in this regulatory submission for device modifications.

Here's the available information based on your request:

Acceptance Criteria and Reported Device Performance

The submission states that performance testing was conducted for the modified devices to "ensure that the modified device performs equivalently to the predicate EC-600HL and EC-600LS." The acceptance criterion for each test was that "the device met the pre-defined acceptance criteria for the test," implying equivalence to the predicate device's performance specifications.

Acceptance Criterion (Type of Test)Reported Device Performance
Field of viewMet pre-defined acceptance criteria
Bending capabilityMet pre-defined acceptance criteria
Rate of air supplyMet pre-defined acceptance criteria
Rate of water supplyMet pre-defined acceptance criteria
Rate of suctionMet pre-defined acceptance criteria
Working lengthMet pre-defined acceptance criteria
Diameter of forceps channelMet pre-defined acceptance criteria
Viewing directionMet pre-defined acceptance criteria
Maximum insertion portion widthMet pre-defined acceptance criteria
ResolutionMet pre-defined acceptance criteria
LG output (Light Guide output)Met pre-defined acceptance criteria

(Note: The document only states that the device 'met pre-defined acceptance criteria' for each test. It does not provide the specific numerical criteria or the measured performance values. The acceptance criteria are implicitly based on the performance characteristics of the predicate device.)

Other Requested Information:

  1. Sample size used for the test set and the data provenance: Not explicitly stated. The performance data seems to be based on engineering and bench testing, not clinical studies with human subjects or retrospective data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable for this type of engineering performance testing. Ground truth for these tests would likely be established through calibrated measurement equipment and established engineering standards.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for this type of engineering performance testing.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This submission is for modifications to an endoscope, not an AI-powered diagnostic device.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an algorithm-only device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the performance characteristics listed, the ground truth would typically be established through physical measurements, calibration, and engineering specifications, rather than clinical outcomes or expert consensus.
  7. The sample size for the training set: Not applicable. This is not a machine learning or AI device that requires a training set in the conventional sense.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming a wing-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 18, 2016

Fujifilm Medical Systems USA, Inc. Shraddha Morc Specialist, Regulatory Affairs and Quality Assurance 10 High Point Drive Wayne, New Jerscy 07470

Re: K162622

Trade/Device Name: Fujifilm Endoscope Models EC-600HL and EC-600LS Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: FDF Dated: September 19, 2016 Received: September 20, 2016

Dear Shraddha Morc:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labcling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not mislcading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Shraddha Morc

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDcvices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sinccrely,

For Division

Douglas Silverstein -S 2016.10.18 13:53:51 -04'00'

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162622

Device Name

Fujifilm Endoscope Models EC-600HL and EC-600LS

Indications for Use (Describe)

This device is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Fujifilm's EC-600HL and EC-600LS

Submitter's Information

Fujifilm Medical Systems U.S.A., Inc. Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293

Contact Person:

Shraddha More Specialist, Regulatory Affairs and Quality Assurance Telephone: (973) 686-2627 Ext. 522627 Facsimile: (973) 633-8818 E-Mail: smore@fujifilm.com

Date Prepared: January 27, 2017

ldentification of the Subject Device:

Proprietary/Trade Name:Fujifilm Endoscope Models EC-600HL and EC-600LS
Common Name:Video Endoscope
Device Class:Class II
Review Panel:Gastroenterology/Urology

Classification Information:

Colonoscope and Accessories (Flexible/Rigid), 21 C.F.R. § 876.1500 Product Code: FDF

Primary Predicate

Fujifilm Endoscope Models EC-600HL and EC-600LS (K143732)

Purpose of the Special 510(k) notice

The EC-600HL and EC-600LS are modifications of predicate EC-600HL and EC-600LS respectively

Intended Use

This device is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine.

{4}------------------------------------------------

Device Description

The modified Fujifilm Endoscope Models EC-600HL and EC-600LS are comprised of three main sections: an operation section, an insertion portion, and an umbilicus. The Table 5-1 describes the insertion and non-insertion portions of the devices. The operation section controls the angulation (up/down/left/right) of the distal end of the endoscope. The insertion contains class fiber bundles, several channels and a complementary metal-oxide-semiconductor (CMOS) image sensor. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor. The endoscope also contains several channels to deliver air/water and provide suction, as well as a forceps channel. The forceps channel is used to introduce endoscope accessories such as biopsy forceps during the procedure. The umbilicus section consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The Figure 10-1 below shows main components of the endoscopes.

The subject device is used in combination with FUJIFILM's video processor, light source and peripheral devices such as water tank, endoscope accessories, monitor, printer, electrosurgical instruments, foot switch, and cart. All of these were previously cleared in K143732.

Image /page/4/Figure/4 description: The image shows a diagram of a system with various components and labels. The diagram includes pipes, valves, and other mechanical parts, all interconnected. The labels provide information about the function and purpose of each component within the system. The diagram appears to be technical in nature, possibly related to engineering or mechanics.

Figure 10-1: Components of the Endoscope EC-600HL and EC-600LS

{5}------------------------------------------------

Technological Characteristics

Table 10-1: Comparison between technological characteristics of EC-600HL and its predicate
Predicate Device ModelEC-600HLProposed Device ModelEC-600HL
Device NameFujifilm Video EndoscopeEC-600HLFujifilm Video EndoscopeEC-600HL
510(k) NumberK143732Pending
Indications for Use(IFU)This device is intended for the visualizationof the lower digestive tract, specifically forthe observation, diagnosis, andendoscopic treatment of the rectum andlarge intestine.Same as K143732
Viewing directionForward/ 0 degreeSame as K143732
Observation range2-100mmSame as K143732
Field of View170 degreesSame as K143732
DistortioncharacteristicsOrthogonal ProjectionSame as K143732
Image sensorsCMOSSame as K143732
Distal end diameter12.8mmSame as K143732
Flexible portiondiameter12.8mmSame as K143732
Maximum insertiondiameter14.3mmSame as K143732
Up180 degreesSame as K143732
BendingDown180 degreesSame as K143732
capabilityLeft160 degreesSame as K143732
Right160 degreesSame as K143732
Forceps channeldiameter4.2mmSame as K143732
Working length1690mmSame as K143732
Total length1990mmSame as K143732
WJ FunctioncSame as K143732
Location of WJ inletOn the light guide connectorSame as K143732
Video ProcessorEPX-4440HD- Light Source:XL-4450- Processor:VP-4440HDSame as K143732
PeripheralsWater Tank WT-2Water Tank, WT-4Same as K143732
Endoscopic Accessory(i.e. Forceps)Same as K143732
MonitorSame as K143732
PrinterSame as K143732
NoneSame as K143732
Electrosurgical InstrumentsSame as K143732
Foot SwitchSame as K143732
CartSame as K143732
Predicate Device ModelEC-600HLProposed Device ModelEC-600HL
AccessoriesCleaning Brush WB11002FW2Same as K143732
Channel Cleaning BrushWB5021FW2Same as K143732
Forceps Valve(FOV-DV7)Same as K143732
Ventilation Adapter(AD-7)Same as K143732
Cleaning Adapter Kit (CA-510/A)Image: Cleaning Adapter KitSame as K143732
J Tube (JT-500)Image: J TubeSame as K143732
Air/Water button(AW-500)Minor material and design modifications
Suction button (SB-500)Minor material and design modifications
Optional itemsWater Jet Inlet capSame as K143732
Air leak tester LT-7FSame as K143732

{6}------------------------------------------------

{7}------------------------------------------------

Predicate Device ModelEC-600LSProposed Device ModelEC-600LS
Device NameFujifilm Video EndoscopeFujifilm Video Endoscope
EC-600LSEC-600LS
510(k) NumberK143732Pending
Indications for Use(IFU)This device is intended for the visualizationof the lower digestive tract, specifically forthe observation, diagnosis, and endoscopictreatment of the rectum and large intestine.Same as K143732
Viewing directionForward/ 0 degreeSame as K143732
Observation range2-100mmSame as K143732
Field of View170 degreesSame as K143732
DistortioncharacteristicsOrthogonal ProjectionSame as K143732
Image sensorsCMOSSame as K143732
Distal end diameter11.5mmSame as K143732
Flexible portiondiameter11.5mmSame as K143732
Maximum insertiondiameter13.1mmSame as K143732
BendingcapabilityUp180 degreesSame as K143732
Down180 degreesSame as K143732
Left160 degreesSame as K143732
Right160 degreesSame as K143732
Forceps channeldiameter3.8mmSame as K143732
Working length1690mmSame as K143732
Total length1990mmSame as K143732
WJ FunctionYesSame as K143732
Location of WJ inletOn the light guide connectorSame as K143732
Video ProcessorEPX-4440HDLight Source: XL-4450Processor:VP-4440HDSame as K143732
Water Tank WT-2Water Tank, WT-4Same as K143732
Endoscopic Accessory(i.e. Forceps)Same as K143732
PeripheralsMonitorSame as K143732
PrinterSame as K143732
NoneSame as K143732
Electrosurgical InstrumentsSame as K143732
Foot SwitchSame as K143732
CartSame as K143732

Table 10-2: Comparison between technological characteristics of EC-600LS and its predicate

{8}------------------------------------------------

Predicate Device ModelEC-600HLProposed Device ModelEC-600HL
AccessoriesCleaning brush WB11002FW2Same as K143732
Cleaning brush WB11002FW2Same as K143732
Channel Cleaning BrushWB5021FW2Same as K143732
Cleaning Adapter Kit (CA-510/A)Image: [Cleaning Adapter Kit]Same as K143732
Forceps Valve(FOV-DV7)Ventilation Adapter(AD-7)Same as K143732
J Tube (JT-500)Image: [J Tube]Same as K143732
Optional itemsAir/Water button(AW-500)Minor material and design modifications
Suction button (SB-500)Minor material and design modifications
Water Jet Inlet capAir leak tester LT-7FSame as K143732

Principles of Operation

The principles of operation of the FUJIFILM Endoscope Models EC-600HL and EC-600LS are identical to those of the predicate device, FUJIFILM 600 Series Endoscope EC-600HL and EC-600LS (K132210). All three devices utilize a retrograde approach for visualization of the areas of the lower digestive tracts.

{9}------------------------------------------------

Summary of the Design Modifications

The proposed devices have been modified from their corresponding predicate devices in their materials. A new epoxy resin will be used in modified devices instead of the current epoxy resin during their manufacturing. Both epoxy resins will be used for the repair of predicate as well as subject devices.

This material change does not affect the intended use or fundamental scientific technology of the proposed endoscopes; therefore the intended use and fundamental scientific technology of the proposed endoscopes remain the same as the legally marketed EC-600HL and EC-600LS (K143732). See Table 10-1 and Table 10-2 for a detailed similarities and differences comparison between the technological characteristics of proposed device and the legally marketed device.

Performance Data

Fujifilm conducted the following performance testing of the modified EC-600HL and EC-600LS to ensure that the modified device performs equivalently to the predicate EC-600HL and EC-600LS:

  • Field of view .
  • Bending capability .
  • Rate of air supply .
  • Rate of water supply .
  • . Rate of suction
  • . Working length
  • . Diameter of forceps channel
  • . Viewing direction
  • Maximum insertion portion width .
  • Resolution ●
  • . LG output

In all cases, the device met the pre-defined acceptance criteria for the test.

Substantial Equivalence

EC-600HL and EC-600LS has the same intended use and similar indications, principles of operation, and technological characteristics as their predicate devices EC-600HL and EC-600LS. The minor differences in the modified devices technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the EC-600HL and EC-600LS are as safe and effective as their predicate devices EC-600HL and EC-600LS respectively. Thus, the modified EC-600HL and EC-600LS are substantially equivalent to their respective predicate devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.