K Number
K162547
Device Name
Viveve System
Manufacturer
Date Cleared
2016-10-06

(24 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.

Device Description

The Viveve System utilizes monopolar radiofrequency (RF) energy to selectively heat a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue. The generator delivers energy to the treatment tip to create an electric field under the treatment tip (electrode). The mechanism of action is the application of RF energy to the tissue causing coagulation and/or hemostasis.

The Viveve System consists of four (4) primary components:

  • An RF Generator to provide the heating energy. The Generator incorporates the Cooling Module to supply the coolant which provides the cooling energy.
  • A hand piece that couples the cooling and heating energy to the tissue through the treatment tip (below).
  • A footswitch that allows the user to turn the RF Energy on or off.
  • Sterile Disposable Treatment Tip.

Accessories include:

  • Coupling Fluid
  • Cryogen
  • Return Cable
  • Return Pad
  • Power Cord
AI/ML Overview

This document is a 510(k) premarket notification for the Viveve System, an electrosurgical device. It primarily focuses on demonstrating substantial equivalence to a predicate device due to design modifications rather than presenting a study to prove new acceptance criteria or performance claims.

Here's an analysis of the provided text in relation to your questions:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly provide a table of acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy) for a diagnostic output. Instead, it discusses demonstrating conformity to safety and performance standards for an electrosurgical device.

The "reported device performance" is essentially that the device continues to meet these standards and design parameters after modifications. It doesn't present new performance data for a diagnostic task.

Here’s a summary based on the provided text, reinterpreting "acceptance criteria" as conformity to relevant standards and "reported device performance" as confirmation of that conformity:

Acceptance Criteria Category (Reinterpreted)Reported Device Performance
Safety and Essential Performance StandardsMet acceptance criteria, confirming continued conformance to IEC 60601-1:2005 (2006/07-3rd edition) and IEC 60601-2-2:2009.
Electromagnetic Compatibility (EMC) StandardsMet acceptance criteria, confirming continued conformance to IEC 60601-1-2:2007.
Design Parameters (for Proposed Modifications)Design verification bench testing and software validation ensured the proposed modifications perform within design parameters and under proper environmental conditions. Results met acceptance criteria.
Biocompatibility StandardsBiocompatibility testing for the Treatment Tip and Coupling Fluid (parts with patient contact) was performed according to ISO 10993-1:2009. The results met acceptance criteria.
Risk Analysis StandardsRisk analysis was completed in accordance with criteria based on ISO 14972:2007.
Sterilization (Method and Parameters)Remains the same as the predicate device.
Shelf Life (Treatment Tips) & Packaging ValidationTreatment Tips shelf life was tested and verified to extend to three years. Packaging validation testing was completed.
System Functionality, Safety, or Effectiveness (Overall)The modifications had "no effect on system functionality, safety, or effectivity," and the "results of performance testing demonstrate that the modifications to the Viveve System do not affect the safety, efficacy, or performance of the generator." The device continues to be "substantially equivalent" to the predicate.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily describes non-clinical performance data (bench testing, software validation, biocompatibility) rather than a clinical test set with patient data. Therefore, the concept of "sample size" in the context of patients, "country of origin," or "retrospective/prospective" study design is not applicable here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided as the submission relies on non-clinical performance data (engineering tests, standard compliance) rather than expert-adjudicated clinical data to establish ground truth for a diagnostic task.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as the document describes non-clinical engineering and safety testing, not a clinical study involving human judgment and adjudication for a test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is an electrosurgical cutting and coagulation device, not an AI-powered diagnostic tool, so this type of study is irrelevant to its purpose.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The Viveve System is an electrosurgical device, not an algorithm or AI product. Its performance is evaluated through its physical and electrical characteristics as determined by engineering tests, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical performance data, the "ground truth" essentially comes from:

  • Industry and regulatory standards: (e.g., IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 10993-1, ISO 14972).
  • Design specifications and performance requirements: The device's components and system are tested against their defined technical specifications and operational parameters.
  • Predicate device characteristics: The modified device is compared to the previously cleared predicate device to ensure substantial equivalence.

8. The sample size for the training set

This information is not applicable. The Viveve System is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping faces.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 6, 2016

Viveve Inc. Mr. Jim Talbot Vice President Quality and Regulatory Affairs 150 Commercial Street Sunnyvale, California 94086

Re: K162547

Trade/Device Name: Viveve System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 9, 2016 Received: September 12, 2016

Dear Mr. Talbot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162547

Device Name Viveve System

Indications for Use (Describe)

The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 7

510(k) SUMMARY

REGULATORY AUTHORITY 7.1

Safe Medical Devices Act of 1990, 21 CFR 807.92

7.2 APPLICANT INFORMATION

Applicant:Viveve Inc.150 Comercial StreetSunnyvale, CA 94086 USA
Contact:Jim TalbotVice President, Quality Assurance and Regulatory Affairsjtalbot@viveve.comPhone: (408) 470-3206Fax: (408) 530-1919

Date Prepared: September 9, 2016

7.3 SUBJECT DEVICE INFORMATION

Trade Name:Viveve System
Common Name:Electrosurgical System
Product Code:GEI
Classification Name:Electrosurgical Cutting and Coagulation Device and
Accessories (21 CFR 878.4400)
Device Panel:General Surgery/Restorative Devices
Device Classification:Class II

PREDICATE DEVICE 7.4

  • TivaMed Cooled RF System (K082962) .

7.5 DEVICE DESCRIPTION

The Viveve System utilizes monopolar radiofrequency (RF) energy to selectively heat a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue. The generator delivers energy to the treatment tip to create an electric field under the treatment tip (electrode). The mechanism of action is the application of RF energy to the tissue causing coagulation and/or hemostasis.

{4}------------------------------------------------

The Viveve System consists of four (4) primary components:

  • . An RF Generator to provide the heating energy. The Generator incorporates the Cooling Module to supply the coolant which provides the cooling energy.
  • A hand piece that couples the cooling and heating energy to the tissue through ● the treatment tip (below).
  • A footswitch that allows the user to turn the RF Energy on or off. ●
  • Sterile Disposable Treatment Tip. ●

Accessories include:

  • Coupling Fluid
  • Cryogen ●
  • Return Cable ●
  • . Return Pad
  • Power Cord

7.6 INDICATION FOR USE

The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.

7.7 TECHNICAL CHARACTERISTICS

The Viveve System is an electrosurgical device that delivers radiofrequency (RF) energy to selectively heat a given area of tissue beneath the surface, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue. The application of RF energy causes the tissue to coagulate and/or become hemostatic.

7.8 SUBSTANTIAL EQUIVALANCE

The Viveve System is substantially equivalent to the predicate device listed. The principle of operation between the predicate device and the subject device remain the same, however the subject device is modified with various design changes to the Generator, Hand Piece, and Treatment Tip. These changes have been tested and verified to have no effect on system functionality, safety, or effectivity. The function of the subject device and indication for use have not changed, therefore, the subject device is substantially equivalent to the predicate device listed.

ItemViveve Systemand Accessories(Subject Device)TivaMed Cooled RF System(now Viveve System)(Predicate Device)
510(k) NumberSubject 510(k)K082962
Legal ManufacturerViveve, Inc.TivaMed, Inc. (now Viveve, Inc.)
ContractManufacturerStellartech Research CorporationStellartech Research Corporation
ItemViveve Systemand Accessories(Subject Device)TivaMed Cooled RF System(now Viveve System)(Predicate Device)
Indication for UseThe Viveve System is indicated for usein general surgical procedures forelectrocoagulation and hemostasis.The TivaMed Cooled RF System isindicated for use in general surgicalprocedures for electrocoagulation andhemostasis.
FDA ClassificationClass IIClass II
CFR/Product code21 CFR 878.4400/GEI21 CFR 878.4400/GEI
Invasiveness oftreatmentNon-invasive. Device applied to thesurface.Non-invasive. Device applied to thesurface.
Principles ofoperationsRadiofrequency (RF) energy selectivelyheats a given volume of tissue beneaththe surface, while cryogen is delivered tothe inside of the Treatment Tip to coolthe surface tissue. The Treatment Tip isplaced on the surface of the skin and theinternal tissues are heated while thesurface tissue is protected. (Reversethermal gradient)Radiofrequency (RF) energy selectivelyheats a given volume of tissue beneaththe surface, while cryogen is delivered tothe inside of the Treatment Tip to coolthe surface tissue. The Treatment Tip isplaced on the surface of the skin and theinternal tissues are heated while thesurface tissue is protected. (Reversethermal gradient)
EnergyRFRF
Treatment TypeMonopolarMonopolar
Input power100-240 VAC @ 50/60 Hz100-240 VAC @ 50/60 Hz
Maximum power(generator)240 Watts240 Watts
Operating Frequency6 MHz6 MHz
Voltage waveform6.0 MHz continuous sinusoidalwaveforms6.0 MHz continuous sinusoidalwaveforms
Electrode probeMonopolarMonopolar
Impedance range25 - 120 Ω25 - 120 Ω
PackagingTyvek PouchTyvek Pouch
SterilityETOETO
Cooling solutionCryogenCryogen

Table 7.8.1: Summary Comparison of Technical Characteristics

{5}------------------------------------------------

7.9 PERFORMANCE DATA

The modifications made to the Viveve System were found to not affect safety or performance through design verification bench testing, which confirmed the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:

  • Safety and essential performance testing in accordance with IEC 60601-1:2005 (2006/07-3rd edition) and IEC 60601-2-2:2009
  • Electromagnetic compatibility in accordance with IEC 60601-1-2:2007 ●

{6}------------------------------------------------

  • . Design verification testing to ensure the proposed modifications perform within design parameters under the proper environmental conditions.
  • Software validation was conducted to ensure the proposed modification . perform with the design parameters.

7.9.1 Non-Clinical Performance Data

Design verification and biocompatibility testing was performed to ensure the subject device Viveve System functions according to the intended use. The results met the acceptance criteria. Risk analysis was also completed in accordance with criteria based on ISO 14972:2007. Electromagnetic compatibility and electrical safety testing was completed in compliance with IEC 60601-1 and IEC 60601-1-2. Biocompatibility testing was conducted for the Treatment Tip and Coupling Fluid according to ISO 10993-1:2009. Other components of the Viveve System do not have direct contact with patient tissue, therefore biocompatibility testing was not required. Sterilization for the subject device remains the same as the predicate device sterilization method and parameters. However, shelf life for the Treatment Tips was tested and verified to extend to three years. Packaging validation testing was also completed against the subject device along with the shelf life testing.

7.9.2 Clinical Performance Data

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

CONCLUSION 7.10

The indications for use and the technical characteristics of the Viveve System has not changed from the predicate device. The results of performance testing demonstrate that the modifications to the Viveve System do not affect the safety, efficacy, or performance of the generator, therefore, the subject device is substantially equivalent to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.