(285 days)
The COSMED K5 Wearable Metabolic Technology is a device designed to measure cardio-respiratory and metabolic functions during stress testing, rehabilitation, sports medicine and other related activities, performed in professional healthcare facilities only. The system is not intended for Home Use.
The main measurements reported by the system are Oxygen Uptake production, Ventilation, Heart Rate and Energy Expenditure.
K5 is intended to use with adults and children over the age of 14 years.
It is to be used by physicians or by trained personnel on a physician responsibility.
The system and its accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals. It must not be intended as a monitoring device, nor as a sole means for determining a patient's diagnosis but for the purpose of assisting the clinician in the diagnosis of cardio-pulmonary diseases
The COSMED K5 Wearable Metabolic Technology is a portable unit, designed for the measurement of resting and exercise metabolism in the laboratory, utilizing international measurement quidelines published by relevant scientific societies. The K5 consists of the following main parts:
- Main portable unit
- Optional Bluetooth receiver
- PC software
- Face mask with head cap
- Turbine flowmeter with optoelectronic reader and wind cover
- Harness
- Heart rate monitor belt
- 02 sensor
- Battery charger + batteries
- AC/DC adapter
- USB cable
- Carrying case
- Optional pressure regulator for calibration
The K5 system is intended to be used the laboratory, by either direct connection or by telemetry. The system offers both 'breath-by-breath' and 'mixing chamber' operational modes.
Here's an analysis of the acceptance criteria and study information for the COSMED K5 Wearable Metabolic Technology, based on the provided FDA 510(k) summary:
Note: The provided document is an FDA 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study report. Therefore, some of the requested information, particularly regarding specific study details like training set provenance, expert qualifications, and detailed statistical methodologies for clinical trials, is not fully elaborated in this type of document.
1. Table of Acceptance Criteria and Reported Device Performance
The document mentions "Accuracy validation (Vt, RF, VO2, VCO2)" as a non-clinical test. The table on page 7 provides specific measurement accuracies for ventilation, VO2, and VCO2, which serve as the acceptance criteria and reported performance.
| Parameter | Acceptance Criteria (Subject Device) | Reported Device Performance (Subject Device) | Predicate Device Accuracy |
|---|---|---|---|
| Ventilation | <2% or 50 ml/min | <2% or 50 ml/min | <2% or 50 ml/min |
| Oxygen Uptake (VO2) | <3% | <3% | <3% or 50 ml/min |
| CO2 Production (VCO2) | <3% | <3% | <3% or 50 ml/min |
2. Sample Size Used for the Test Set and Data Provenance
The document refers to "non-clinical testing" for "Accuracy validation (Vt, RF, VO2, VCO2)." However, it does not specify the sample size used for these tests, nor does it provide details on the data provenance (e.g., country of origin, retrospective or prospective nature of data). These non-clinical tests are typically performed in a controlled laboratory environment using simulated physiological conditions or human volunteers, but the specifics are not detailed here.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the 510(k) summary. The accuracy validation tests for metabolic devices often rely on comparison to a gold standard measurement method or calibrated gas mixtures, rather than expert consensus on patient data. Therefore, the concept of "experts establishing ground truth for the test set" in the context of this device's performance validation as presented is not applicable in the way it might be for image-based diagnostic AI.
4. Adjudication Method for the Test Set
This information is not provided and is generally not applicable to the type of non-clinical accuracy validation tests described for this device. Adjudication methods are typically used in clinical studies involving interpretation by multiple observers where there might be disagreement in assessing an outcome, which is not the case for direct physiological measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or at least not reported in this 510(k) summary. This type of study is usually conducted for diagnostic imaging devices where human readers interpret images with and without AI assistance to determine the AI's impact on diagnostic accuracy. The COSMED K5 is a measurement device for physiological variables, not an interpretative diagnostic imaging tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The performance data presented in the table above (Ventilation, VO2, VCO2 accuracy) represents the standalone performance of the device's measurement algorithms. The device directly measures and calculates these physiological parameters. There isn't an "algorithm only" vs. "human-in-the-loop" distinction in the same way there would be for an AI-powered diagnostic tool. The device is designed to provide direct physiological measurements, and the accuracy validation confirms its ability to do so.
7. The Type of Ground Truth Used
For the accuracy validation tests (Vt, RF, VO2, VCO2), the ground truth would typically be established by:
- Highly calibrated reference systems or gas analyzers: For VO2 and VCO2 measurements, this would involve comparing the device's readings against known gas concentrations or a highly accurate reference metabolic cart.
- Volumetric calibration for ventilation (Vt, RF): Using precision syringes or flow calibrators to provide known volumes and rates of airflow.
The document does not explicitly state the specific "gold standard" methods used, but for this type of device, it would be based on physical standards and calibrated instruments.
8. The Sample Size for the Training Set
The 510(k) summary does not provide information on a "training set" in the context of machine learning. This device measures physiological variables directly and calculates parameters based on known physiological equations and sensor readings. It's not an AI/ML device that requires a large dataset for training a predictive model in the conventional sense. The "training" of such a device would involve calibration and optimization of its internal algorithms and sensor performance based on engineering principles and validation data, not a "training set" of patient data for an AI model.
9. How the Ground Truth for the Training Set Was Established
As there is no "training set" for an AI/ML model described for this device, the question of how its ground truth was established is not applicable in the AI sense. The development and calibration of the K5's measurement capabilities would rely on rigorous engineering, physical principles, and laboratory testing using established standards and known physiological conditions.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of three human profiles facing right, stacked on top of each other. The profiles are stylized with flowing lines. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2017
Cosmed S.r.l. % Roger Gray VP Quality and Regulatory Donawa Lifescience Consulting Srl Piazza Albania 10 Rome, 00153 IT
Re: K162515
Trade/Device Name: COSMED K5 Wearable Metabolic Technology Regulation Number: 21 CFR 868.1880 Regulation Name: Pulmonary-Function Data Calculator Regulatory Class: Class II Product Code: BZC Dated: May 17, 2017 Received: May 22, 2017
Dear Roger Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Lori A. Wiggins -S6
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K162515
Device Name
COSMED K5 Wearable Metabolic Technology
Indications for Use (Describe)
The COSMED K5 Wearable Metabolic Technology is a device designed to measure cardio-respiratory and metabolic functions during stress testing, rehabilitation, sports medicine and other related activities, performed in professional healthcare facilities only. The system is not intended for Home Use.
The main measurements reported by the system are Oxygen Uptake production, Ventilation, Heart Rate and Energy Expenditure.
K5 is intended to use with adults and children over the age of 14 years.
It is to be used by physicians or by trained personnel on a physician responsibility.
The system and its accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals. It must not be intended as a monitoring device, nor as a sole means for determining a patient's diagnosis but for the purpose of assisting the clinician in the diagnosis of cardio-pulmonary diseases
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for COSMED - The Metabolic Company. On the left side of the logo is a blue griffin inside of a blue circle. To the right of the griffin is the word "COSMED" in large blue font, with the words "The Metabolic Company" in a smaller font below it.
510(k) Summary
| Device Name: | COSMED K5 Wearable Metabolic Technology |
|---|---|
| Type of 510(k) submission: | Traditional |
| Date of submission: | 12 June 2017 |
| Manufacturer: | Cosmed SrlVi dei Piani di Monte Savello 37Albano Laziale00041 RomeItaly |
| Phone: | +39 06 931 5492 |
| Fax: | +39 06 931 4580 |
| FDA Establishment Reg. Number: | 8021084 |
| 510(k) Owner and Submitter: | Cosmed SrlVi dei Piani di Monte Savello 37Albano Laziale00041 RomeItaly |
| Owner/Operator Reg. Number: | 8021084 |
| 510(k) Application Correspondent: | Mr Roger GrayVP Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania 1000153 RomeItaly |
| Phone: | +39 06 578 2665 |
| Fax: | +39 06 574 3786 |
| Email: | rgray@donawa.com |
| FDA Product Code: | BZC |
| FDA Regulation Number: | 21 CFR 868.1880 |
| FDA Classification Name: | Calculator, Pulmonary Function Data |
| Classification Panel: | Anesthesiology |
| Common Name: | Pulmonary Function Data Calculator |
| FDA Classification: | Class II |
| Submission Type: | 510(k) |
Indications for Use:
The COSMED K5 Wearable Metabolic Technology is a device designed to measure cardio-respiratory and metabolic functions during stress testing, rehabilitation, sports medicine and other related activities, performed in professional healthcare facilities only. The system is not intended for Home Use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for COSMED, The Metabolic Company. On the left is a blue circle containing a blue griffin. To the right of the circle is the word "COSMED" in large blue letters. Below the word "COSMED" are the words "The Metabolic Company" in smaller letters.
The main measurements reported by the system are Oxygen Uptake, Carbon Dioxide production, Ventilation, Heart Rate and Energy Expenditure.
K5 is intended to use with adults and children over the age of 14 years.
It is to be used by physicians or by trained personnel on a physician responsibility.
The system and its accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals. It must not be intended as a monitoring device, nor as a sole means for determining a patient's diagnosis but for the purpose of assisting the clinician in the diagnosis of cardiopulmonary diseases.
Device Description:
The COSMED K5 Wearable Metabolic Technology is a portable unit, designed for the measurement of resting and exercise metabolism in the laboratory, utilizing international measurement quidelines published by relevant scientific societies. The K5 consists of the following main parts:
- Main portable unit ●
- Optional Bluetooth receiver
- PC software ●
- Face mask with head cap
- Turbine flowmeter with optoelectronic reader and wind cover
- Harness ●
- Heart rate monitor belt
- 02 sensor
- . Battery charger + batteries
- AC/DC adapter ●
- USB cable
- Carrying case ●
- Optional pressure regulator for calibration
The K5 system is intended to be used the laboratory, by either direct connection or by telemetry. The system offers both 'breath-by-breath' and 'mixing chamber' operational modes.
Performance data:
Measurement of exercise metabolism consists of collecting during physical activities (e.g. running, cycling, rowing, etc.) a number of physiological variables (including breathing flow, exhaled CO2 fraction, heart rate) and environmental signals (ambient temperature, pressure and humidity) in order to calculate the following main parameters:
- VO2 (ml/min) Oxygen Uptake, known also as Oxygen Consumption ●
- . VCO2 (ml/min) Carbon Dioxide production
- VE (I/min) Minute Ventilation
- HR (1/min) Heart Rate ●
- . RQ (---) Respiratory Quotient
Measurement of resting metabolism consists of measuring at rest the following main parameters:
- VO2 (ml/min) Oxygen Uptake, known also as Oxygen Consumption ●
- VCO2 (ml/min) Carbon Dioxide production .
- RQ (---) Respiratory Quotient ●
- EE (Kcal/day) Energy Expenditure ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for COSMED, The Metabolic Company. On the left is a blue circle containing a blue griffin. To the right of the circle is the company name, COSMED, in large blue letters. Below the company name is the tagline, "The Metabolic Company", in smaller letters.
Non-clinical testing:
Subject device non-clinical testing confirms that the device meets its specification, including the requirements of relevant standards:
- Electrical safety in accordance with IEC 60601-1:2005
- EMC in accordance with IEC 60601-1-2:2007
- Accuracy validation (Vt, RF, VO2, VCO2) .
- . Coexistence testing (FCC Part 15)
- Protection provided by enclosure (IEC 60529:2013) ●
In relation to biocompatibility, all of the K5 components that are either intended to contact the patient or may come into contact with the patient during use have been previously assessed for their biocompatibility, as appropriate for the nature of their body contact and contact duration, by FDA within already-cleared 510(k) submissions.
The results of the above testing, together with previous FDA clearances, assist in the demonstration of substantial equivalence of the subject device with the predicate device.
Substantial equivalence
The predicate device selected for comparison with the Cosmed K5 Wearable Metabolic Technology is:
| Predicate Device: | Jaeger Oxycon Mobile |
|---|---|
| Sponsor: | SensorMedics, Inc. |
| 510(k) Number: | K023120 |
| Clearance Date: | 12 November 2003 |
| FDA Product Code: | BZC |
| Classification Name: | Calculator, Pulmonary Function Data |
| Regulation No: | 21 CFR 868.1880 |
| Class: | II |
Predicate device comparison table:
Table 1 provides evidence of substantial equivalence of the subject device with the selected predicate device.
| Table 1: Predicate device comparison table | |||||
|---|---|---|---|---|---|
| Feature | Subject device | Predicate device | Similarity | ||
| Device name | K5 Wearable Metabolic Technology | Jaeger Oxycon Mobile | N/A | ||
| Device Manufacturer | Cosmed | SensorMedics | N/A | ||
| 510(k) Reference | This submission | K023120 | N/A | ||
| FDA Product Code | BZC | BZC | Same | ||
| FDA ClassificationName | Calculator, Pulmonary Function Data | Calculator, Pulmonary Function Data | Same | ||
| FDA RegulationNumber | 868.1880 | 868.1880 | Same | ||
| Device description | A portable cardiopulmonary functiontesting system offering breath-by-breath and sample mixing options,with wireless telemetry | A portable cardiopulmonary functiontesting system offering breath-by-breath sampling and wirelesstelemetry | Similar |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for COSMED, The Metabolic Company. On the left is a blue griffin inside of a blue circle. To the right of the griffin is the word "COSMED" in large blue letters, with the words "The Metabolic Company" in smaller letters underneath.
| Table 1: Predicate device comparison table | |||
|---|---|---|---|
| Feature | Subject device | Predicate device | Similarity |
| Indications for use | The COSMED K5 WearableMetabolic Technology is a devicedesigned to measure cardio-respiratory and metabolic functionsduring stress testing, rehabilitation,sports medicine and other relatedactivities, performed in professionalhealthcare facilities only. The systemis not intended for Home Use.The main measurements reported bythe system are Oxygen Uptake,Carbon Dioxide production,Ventilation, Heart Rate and EnergyExpenditure.K5 is intended to use with adults andchildren over the age of 14 years.It is to be used by physicians or bytrained personnel on a physicianresponsibility.The system and its accessories areindicated for the acquisition, analysis,formatting, display, printing andstorage of certain physiologic signals.It must not be intended as amonitoring device, nor as a solemeans for determining a patient'sdiagnosis but for the purpose ofassisting the clinician in the diagnosisof cardio-pulmonary diseases. | The Jaeger OXYCON MOBILEpulmonary function mobile testsystem is a device which monitors thecardio-respiratory functions duringstress testing, rehabilitation, sportsmedicine and other related activities.The OXYCON MOBILE system allowsthe use of telemetry for the monitoringof metabolic parameters, TheOXYCON MOBILE system isintended to use with adults andchildren over the age of 14 years. | Substantiallyequivalent |
| Device description | A portable cardiopulmonary functiontesting system offering breath-by-breath and sample mixing options,with wireless telemetry | A portable cardiopulmonary functiontesting system offering breath-by-breath sampling and wirelesstelemetry | Similar |
| User Population | Adults and children over the age of 14years | Adults and children over the age of 14years | Same |
| Use environment | Professional healthcare facilities | Indoor and outdoor use | Different |
| Device measurements | Cardiorespiratory functions duringstress testing, rehabilitation, sportmedicine and other related activities | Cardiorespiratory functions duringstress testing, rehabilitation, sportmedicine and other related activities | Same |
| Measurement mode(s) | Breath by breath and mixing chamber | Breath by breath only | Different |
| Output parameters | Ergospirometric key parameters suchas ventilation, VO2, VCO2, anaerobicthreshold, RQ, HR, VE/VO2,VE/VCO2. | Ergospirometric key parameters suchas ventilation, VO2, VCO2, anaerobicthreshold, RQ, HR, VE/VO2,VE/VCO2. | Same |
| Major separate systemcomponents | K5 unit, flowmeter, masks, receiverUSB stick (to PC) | Oxycon unit (2 pieces), flowmeter,masks, receiver unit (to PC) | Different |
| Flowmeter technology | Bi-directional turbine | Bi-directional turbine | Same |
| O2 sensor technology | Electrochemical | Electrochemical | Same |
| CO2 sensor technology | Infrared | Thermal conductivity | Different |
| Barometric pressuresensor technology | Piezo-resistive | Piezo-resistive | Same |
| Telemetry | Bluetooth, bi-directional | 2401-2495 MHz, bi-directional | Different |
| Table 1: Predicate device comparison table | |||
| Feature | Subject device | Predicate device | Similarity |
| User interface | LCD touchscreen plus PC keyboardand mouse | PC Keyboard and mouse | Different |
| PC Software | Windows based application | Windows based application | Same |
| Biocompatibility | All patient contact components arebiocompatible and used in alreadylegally marketed devices with thesame intended use | All patient contact components arebiocompatible and used in alreadylegally marketed devices with thesame intended use | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Anatomical sites | Patient's mouth, through face mask | Patient's mouth, through face mask | Same |
| Energy used | External or internal power supply | External or internal power supply | Same |
| Energy delivered | No energy delivered to patient | No energy delivered to patient | Same |
| Standards: Safety | IEC 60601-1: class II / InternalElectric Power Source - type BF | IEC 60601-1: class II / InternalElectric Power Source - type BF | Same |
| Battery | Lithium-Ion, rechargeable | Lithium-Ion, rechargeable | Same |
| Dimensions | 1 unit 174 x 111 x 64 mm | 2 units 126 x 96 x 41mm | Different |
| Weight | 900g (31.7 oz) | 950g (33.4 oz) | Similar |
| Measurement Range | Ventilation: 0-300 l/minVO2: 0.1-7 l/minVCO2: 0.1-7 l/min | Ventilation: 0-300 l/minVO2: 0-7 l/minVCO2: 0-7 l/min | Different |
| Measurement Accuracy | Ventilation: <2% or 50 ml/minVO2: <3%VCO2: <3% | Ventilation: <2% or 50 ml/minVO2: <3% or 50 ml/minVCO2: <3% or 50 ml/min | Different |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for COSMED, The Metabolic Company. On the left is a blue circle with a white griffin inside. To the right of the circle is the company name, COSMED, in large blue letters, with the tagline "The Metabolic Company" in smaller letters below.
The subject device and the predicate device have many identical or similar properties or features. The differences that exist and are identified in the above table include:
- Measurement mode(s)
- Major separate system components
- CO2 sensor technology ●
- Use environment
- . Telemetry
- User interface ●
- Dimensions
- Measurement range
- . Measurement accuracy
None of the identified differences introduce new aspects of safety or effectiveness.
12.4 Conclusion
Based on the information contained within this submission, it is concluded that the COSMED K5 Wearable Metabolic Technology is substantially equivalent to the identified predicate device already in interstate commerce within the USA.
§ 868.1880 Pulmonary-function data calculator.
(a)
Identification. A pulmonary-function data calculator is a device used to calculate pulmonary-function values based on actual physical data obtained during pulmonary-function testing.(b)
Classification. Class II (performance standards).