K Number
K162509

Validate with FDA (Live)

Date Cleared
2017-03-10

(183 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Infrared Non-Contact Thermometers, ACT8100, is non-sterile, reusable clinical thermometers intended for the intermittent determination of human body temperature measured at the center of the forehead without contact. The device is intended for use on people of all ages, except pre-term babies or very small (small for gestational age). The device is intended for use in clinical and home use environments.

Device Description

The Infrared Non-Contact Thermometer, ACT8100 is a hand-held device powered by batteries and designed to measure human body temperature without contacting patients. This thermometer can switch modes between forehead and ambient temperatures. The ambient temperature mode is DIRECT MODE and the forehead temperature mode is ADJUSTED MODE. The forehead temperature of the ADJUSTED MODE is calculated by converting the measured temperature of the DIRECT MODE to an oral equivalent temperature without contacting the patient's forehead. Infrared Non-Contact Thermometer, ACT8100 uses a thermopile integrated with a thermistor, a thermistor mounted on the head of the thermometer for ambient temperature readings, and an optoelectronic mechanism that focuses the infrared energy emitted from the user s forehead for the detection of noncontact use and calibration of temperature reading.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Infrared Non-Contact Thermometer, Model ACT8100, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard & Requirement)Reported Device Performance (ACT8100)
Laboratory Accuracy (ASTM E1965-98:2009 & ISO 80601-2-56:2009)
35.0 °C ~ 42.0 °C± 0.2 °C (for 35.0~42.0°C)
Below 34.9 °C or Above 42.1 °C± 0.3 °C (for below 34.9°C or above 42.1°C)
Repeatability (EN12470-5:2003, subclause 6.3.4)Under 0.3℃
Measuring Range32.0 °C ~ 43.9 °C (89.6 °F ~ 111.0 °F)

Note: The document states that the ACT8100's accuracy and measuring range "All meet ASTM E 1965-98:2009". The specific criteria for ASTM E1965-98:2009 for the full range are a bit more detailed, but the provided table summarizes the reported performance against the specified ranges.

2. Sample Size Used for the Test Set and Data Provenance

The document states:

  • "The three groups of subjects being tested were: (1) Infants of under 1 year old, (2) Children between 1 to 5 years old, (3) Adults"
  • However, the specific number of subjects (sample size) within each group or the total sample size for the clinical study is not provided.
  • Data Provenance: The country of origin for the data is not explicitly stated in the provided text. The submission is from TAIWAN (Easywell Biomedicals, Inc.), so it's likely the study was conducted there, but this is not confirmed. The study was prospective in nature, as it involved taking measurements from subjects using the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • The document does not specify the number of experts used to establish the ground truth or their qualifications.
  • The ground truth was established by a "reference thermometer, Actherm Digital Clinical Thermometer ACT 2130 (K031905), either at the rectal, oral, or axillary site depending on their age group." While this is a reference device, it doesn't involve "experts" in the sense of human readers adjudicating images; rather, it's a direct measurement.

4. Adjudication Method for the Test Set

  • An adjudication method (like 2+1, 3+1) is not applicable here as the ground truth was established by direct temperature measurements from a reference thermometer, not by human interpretation that requires consensus.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not explicitly mentioned or performed in the context of human readers improving with AI vs. without AI assistance. This device is a standalone non-contact thermometer, not an AI-assisted diagnostic tool that humans would interpret.
  • The study did compare the ACT8100's measurements to a predicate device (NTF 3000) and a reference standard (ACT 2130), demonstrating similar clinical biases and repeatability, but this is not an MRMC study.

6. Standalone Performance (Algorithm Only)

  • Yes, a standalone performance study was done. The entire clinical evaluation detailed in Section X ("Clinical Testing") is a standalone performance assessment of the ACT8100. It directly measures the device's accuracy and repeatability against a reference standard and compares it to a predicate device without human interpretation of the device's output.

7. Type of Ground Truth Used

  • The type of ground truth used was measurements from a reference standard device, specifically the "Actherm Digital Clinical Thermometer ACT 2130 (K031905), either at the rectal, oral, or axillary site depending on their age group." This is clinical measurement-based ground truth.

8. Sample Size for the Training Set

  • The document does not provide information on a "training set" or its sample size. This type of non-AI, hardware-based medical device typically does not involve machine learning training sets in the same way an AI-powered diagnostic algorithm would. The device's calibration and algorithm for converting infrared readings to oral equivalent temperatures would have been developed during its R&D phase, but a distinct "training set sample size" with associated ground truth for a retrospective ML model is not applicable here.

9. How the Ground Truth for the Training Set Was Established

  • As there's no mention of a distinct "training set" in the context of machine learning, the question of how its ground truth was established is not applicable based on the provided information. The device's internal algorithms would have been developed and validated through engineering and laboratory testing, likely using controlled temperature sources and potentially clinical data, but not typically referred to as a "training set" in this context.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of a human figure, represented by three overlapping profiles facing to the right, with a stylized wave or ribbon element below them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 10, 2017

Easywell Biomedicals, Inc. Arthur Chen Quality Assurance Manager Suite 2, 2F, No.9, Jhanye 1st Rd, Hsinchu Science Park Hsinchu, 30078 TAIWAN

Re: K162509

Trade/Device Name: Infrared Non-contact Thermometer, Model ACT8100 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 26, 2017 Received: February 15, 2017

Dear Arthur Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". There is also a "-5" below the name. The signature is a series of loops and curves.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162509

Device Name

Infrared Non-Contact Thermometer, Model ACT8100

Indications for Use (Describe)

The Infrared Non-Contact Thermometers, ACT8100, is non-sterile, reusable clinical thermometers intended for the intermittent determination of human body temperature measured at the center of the forehead without contact. The device is intended for use on people of all ages, except pre-term babies or very small (small for gestational age). The device is intended for use in clinical and home use environments.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Easywell Biomedicals. The logo features the word "EASYWELL" in blue, with a green checkmark-like symbol incorporated into the "W". Below the word "EASYWELL" is the word "Biomedicals" in a smaller font. A curved line of green dots appears below the word "Biomedicals", adding a visual element to the logo.

K162509 510(k) Summary

I. Submitter

EASYWELL BIOMEDICALS, INC.

Suite 2, 2F, No.9, Jhanye 1st Rd., Hsinchu Science Park, 30078 Hsinchu, TAIWAN

Phone:+886-3-666-9596

FAX: +886-3-666-9697

Contact Person:Richard Hsieh

Date: January 25, 2017

II. Device

Trade Name:Infrared Non-Contact Thermometer, Model ACT8100

Common Name: Infrared Thermometer

Regulation Number: 21 CFR 880. 2910

Regulation Name:Clinical Electronic Thermometer

Regulatory Class : II

Product Code:FLL

III. Predicate Device

K134043 : BRAUN No Touch + Forehead Thermometer; ModelNTF3000US (US Model)

Reference Device

K031905: Actherm Digital Clinical Thermometer ACT 2130

IV. Device Description

The Infrared Non-Contact Thermometer, ACT8100 is a hand-held device powered by batteries and designed to measure human body temperature without contacting patients. This thermometer can switch modes between forehead and ambient temperatures. The ambient temperature mode is DIRECT MODE and the forehead temperature mode is ADJUSTED MODE. The forehead temperature of the ADJUSTED MODE is calculated by converting the measured temperature of the DIRECT MODE to an oral equivalent temperature without contacting the patient's forehead. Infrared Non-Contact Thermometer, ACT8100 uses a thermopile integrated with a thermistor, a thermistor mounted on the head of the thermometer for ambient temperature readings, and an optoelectronic mechanism that focuses the infrared energy emitted from the user s forehead for the detection of noncontact use and calibration of temperature reading.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Easywell Biomedicals. The logo has the word "EASYWELL" in blue, with the "V" in "EASYWELL" being a lighter shade of blue. Below the word "EASYWELL" is the word "Biomedicals" in a smaller font. Below the word "Biomedicals" is a curved line of blue dots.

Physical and Performance characteristics:

Dimension163×44×26mm approximately
WeightApproximately 75g without battery; 88g with batteries
MeasuringRange32.0 °C ~ 43.9 °C (89.6 °F ~ 111.0 °F)
Accuracy± 0.2 °C (for 35.0~42.0°C); ± 0.3°C (for below 34.9°C or above42.1°C)

V. Indication for Use

The Infrared Non-Contact Thermometers, ACT8100, is non-sterile, reusable clinical thermometers intended for the intermittent determination of human body temperature measured at the center of the forehead without contact. The device is intended for use on people of all ages, except pre-term babies or very small (small for gestational age). The device is intended for use in clinical and home use environments.

VI. Conformance Standards

The Infrared Non-Contact Thermometers conform to the following standards:

    1. ISO 80601-2-56:2009 Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
    1. ASTM E1965-98:2009 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature.
    1. IEC 60601-1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard Electromagnetic compatibility -Requirements and tests
    1. IEC 60601-1-11:2010 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
    1. ISO 14971:2007 Medical Devices: Application of Risk Management to Medical Devices.
    1. Software: FDA Guidance for the Content of Premarket Submissions for Software contained in Medical Device
    1. ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process.
    1. ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity.
    1. ISO 10993-10: 2010 Biological evaluation of medical devices. Tests for irritation and delayedtype hypersensitivity

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Easywell Biomedicals. The logo is in blue and features the word "EASYWELL" in a stylized font. Below the word is the word "Biomedicals" in a smaller font. Underneath the words is a curved line made of blue-green dots.

VII. Similarities/Differences between the subject and predicate device:
Subject Device(s)Predicate Device
ELEMENT OFCOMPARISONACT8100No touch + foreheadthermometer(NTF3000US)Comparison
ClassificationThermometerClinical Electronic,Class II DeviceThermometerClinical Electronic,Class II DeviceThe deviceclassifications areidentical
Intended UseA non-sterile, reusableclinical thermometerintended for theintermittentdetermination of humanbody temperaturemeasured at the centerof the forehead withoutcontact. The device isintended for use onpeople of all ages,except pre-term babiesor very small (small forgestational age). Thedevice is intended foruse in clinical and homeA non-sterile, reusableclinical thermometerintended for theintermittentdetermination of thehuman's bodytemperature in a touchand no touch on thecenter of the foreheadas the measurement siteon people of all ages.The intended usestatements aresimilar
SensorInfrared: converts auser's foreheadtemperature using theinfrared energy emittedin the area around theuser's forehead to anoral equivalenttemperature whenplaced up to 2-8cmaway.Infrared: converts auser'sforeheadtemperature using theinfrared energy emittedin the area around theuser's forehead to anoral equivalenttemperature whenplaced in contact withthe subject's foreheador up to 2 inches away.Technologicalcharacteristics aresimilar
ELEMENT OFCOMPARISONSubject Device(s)Predicate DeviceComparison
PowerRequirementACT8100:1.5V AAA size Alkaline battery *21.5V AA size battery *2Battery size different, but the power requirements are similar (voltage outputs are the same)
Dimensions:163×44×26 mmDimensions:Length:148Head: 44×26 mmBody: 37×24 mmHead dimension is similar, the body length is different.
FunctionalityMaterials:All skin contacting materials have been tested in accordance with ISO 10993-1、-5、-10.Materials:All skin contacting materials have been tested in accordance with ISO 10993-1、-5、-10.Material characteristics all meet biological requirements
Measuring ranges:$32.0 °C ~ 43.9 °C$Measuring ranges:$34.4 °C ~ 42.2 °C$All meet ASTM E 1965-98:2009
Accuracy:$35.0 °C ~ 42.0 °C: ± 0.2 °C$$<35.0 °C$ or $>42.0 °C: ± 0.3 °C$Accuracy:$35.0 °C ~ 42.0 °C: ± 0.2 °C$Out of above range: ± 0.3 °CAll meet ASTM E 1965-98:2009
Forehead measuring distance:2~8 cmForehead measuring distance:0~5 cmSimilar

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Easywell Biomedicals. The logo features the word "EASYWELL" in a stylized font, with the "V" in "WELL" replaced by a green checkmark. Below the word "EASYWELL" is the word "Biomedicals" in a smaller, sans-serif font. A curved line of green dots appears below the word "Biomedicals".

VIII. Comparison of Technological Characteristics with the Predicate Device:

The technological characteristics of ACT8100 and predicate device (NTF3000US) are similar. Both convert user's forehead temperature using the infrared energy emitted in the area around the user 's forehead to an oral equivalent temperature.

The differences between ACT 8100 and predicate device (NTF3000US) do not raise different questions of safety or effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Easywell Biomedicals. The logo is in blue and green, with the word "EASYWELL" in large, bold letters. Below the word "EASYWELL" is the word "Biomedicals" in smaller letters. A curved line of blue and green dots appears below the word "Biomedicals".

DifferenceDifference DescriptionReasons why not raising new questions
Battery sizeACT 8100 using AAA size,the predicated device(NTF3000US) using AA sizeThe output voltage is the same, and thedifferences are verified not to raise newquestions regarding safety and effectivenessrequirements within standard IEC60601-1electrical safety requirements and IEC 60601-1-2 EMC test.
DimensionsThe design of shape is notvery similar.The differences are verified not to raise newquestions regarding safety and effectivenessrequirements within standard IEC60601-1electrical safety requirements & ASTM E1965-98:2009
AmbientmeasuringrangesACT 8100 (32.0°C ~ 43.9°C)is a little wider than The NoTouch + ForeheadThermometer (34.4°C ~42.2°C).The differences are verified not to raise newquestions regarding safety and effectivenessrequirements within standard IEC60601-1electrical safety requirements & ASTM E1965-98:2009
Foreheadmeasuringdistance(sensor)The predicated device(NTF3000US) designed forboth contact and no-contact.ACT 8100 is designed foronly non-contact.The differences are verified not to raise newquestions regarding safety and effectivenessrequirements within standard IEC60601-1electrical safety requirements & ASTM E1965-98:2009

IX. Non-Clinical Safety and Performance Testing

Non-clinical performance tests used to demonstrate the substantial equivalence of ACT 8100 to the predicate device. The device conforms to the requirements of laboratory accuracy, general electrical safety, electromagnetic compatibility and biocompatibility. Thus, the device has the same performance characteristics as the predicate device.

Non-Clinical PerformanceTestingStandard/Guidance Applied and Conformed
Laboratory accuracyASTM E1965-98:2009ISO 80601-2-56:2009
General Electrical SafetyIEC 60601-1:2012IEC 60601-1-11:2010
ElectromagneticCompatibilityIEC 60601-1-2:2014
BiocompatibilityISO 10993-1;ISO 10993-5; ISO 10993-10
SoftwareIEC 62304
UsabilityIEC 62366
Risk ManagementISO 14971

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Easywell Biomedicals. The logo is blue and features the word "EASYWELL" in a stylized font. Below the word "EASYWELL" is the word "Biomedicals" in a smaller font. Underneath the word "Biomedicals" is a curved line of blue-green dots.

X. Clinical Testing

A clinical study was performed to determine the clinical accuracy and to provide comparison with the predicate device. The three groups of subjects being tested were: (1) Infants of under 1 year old, (2) Children between 1 to 5 years old, (3) Adults

Each group of subjects underwent a series of temperature measurement at their forehead and back of ear using ACT 8100 and BRUAN Infrared no touch + forehead thermometer NTF 3000. This series of measurement was intended to indicate and ultimately compare the repeatability of ACT 8100 and NTF 3000. The repeatability of both devices is found to be under 0.3℃. which is within the acceptable range as required by EN12470-5:2003, subclause 6.3.4. In addition, the repeatability of the both devices were found to be similar, further indicating that the ACT 8100 has reasonable accuracy and reliability. The three subject groups underwent a series of temperature measurement using the reference thermometer, Actherm Digital Clinical Thermometer ACT 2130 (K031905), either at the rectal, oral, or axillary site depending on their age group. These temperature measurements were then each compared to the temperature measurements taken by ACT 8100 and that of NTF 3000. The clinical biases of these measurements taken by ACT 8100 were found to be similar to that of NTF 3000, indicating that ACT 8100 produces effective and accurate temperature measurements.

XI. Conclusion

Infrared Non-Contact Thermometers, ACT8100 has similar intended use, principles of operation, and similar technological characteristics as the predicate device identified. Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. Results from the clinical study demonstrate substantial equivalence of accuracy and reliability of ACT 8100 and its similarity with NTF 3000. Thus, the Infrared Non-Contact Thermometer is substantially equivalent to the predicate device.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.