K Number
K162451

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-11-28

(88 days)

Product Code
Regulation Number
866.3309
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana® HSV 1+2/VZV Assay is an in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection and differentiation of herpes simplex virus type 1, herpes simplex virus type 2, and varicella-zoster virus DNA isolated and purified from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection. The Solana® HSV 1+2/VZV Assay is intended to aid in the diagnosis of herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus active cutaneous or mucocutaneous infections. Negative results do not preclude herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus infections and should not be used as the sole basis for diagnosis, treatment or other management decisions. The Solana® HSV 1+2/VZV Assay is intended for use only with the Solana® instrument.

Warning: The Solana® HSV 1 + 2/VZV Assay is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS). The Solana® HSV 1 + 2/VZV Assay is not intended for use in prenatal screening.

Device Description

The Solana® HSV 1+2/VZV Assay amplifies and detects viral DNA isolated from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection.

The assay consists of two (2) major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to HSV-1, HSV-2 and/or VZV using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

Patient specimen is transferred to a Process Tube, subjected to heat treatment at 95°C for 5 minutes and mixed and vortexed. The processed sample is transferred to a Reaction Tube and mixed. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of specific target sequence. In Solana, the target sequences are amplified by HSV-1, HSV-2 and/or VZV specific primers and detected by HSV-1, HSV-2 and/or VZV specific fluorescence probes included in the Reaction Tube. A competitive process control (PRC) is included in the Process Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to HSV-1, HSV-2, VZV or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana will then report the test results to the user on its display screen, and the results can be printed via an attached printer.

AI/ML Overview
{
  "1. A table of acceptance criteria and the reported device performance": {
    "HSV-1 Cutaneous Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 85.1% to 100%)",
        "Specificity": "97.8% (95% CI: 95.0% to 99.1%)"
      }
    },
    "HSV-2 Cutaneous Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "92.3% (95% CI: 75.9% to 97.9%)",
        "Specificity": "94.4% (95% CI: 90.8% to 96.6%)"
      }
    },
    "VZV Cutaneous Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 85.1% to 100%)",
        "Specificity": "96.5% (95% CI: 93.0% to 98.3%)"
      }
    },
    "HSV-1 Mucocutaneous Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 96.7% to 100%)",
        "Specificity": "96.4% (95% CI: 94.0% to 97.8%)"
      }
    },
    "HSV-2 Mucocutaneous Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "99.1% (95% CI: 95.0% to 99.8%)",
        "Specificity": "97.2% (95% CI: 95.4% to 98.3%)"
      }
    },
    "VZV Mucocutaneous Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 51.0% to 100%)",
        "Specificity": "98.6% (95% CI: 96.9% to 99.4%)"
      }
    },
    "HSV-1 Uncategorized Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 86.7% to 100%)",
        "Specificity": "97.6% (95% CI: 93.1% to 99.2%)"
      }
    },
    "HSV-2 Uncategorized Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 82.4% to 100%)",
        "Specificity": "96.6% (95% CI: 92.3% to 98.6%)"
      }
    },
    "VZV Uncategorized Lesions": {
      "Acceptance Criteria": {
        "Sensitivity": "≥ 85%",
        "Specificity": "≥ 95%"
      },
      "Reported Device Performance": {
        "Sensitivity": "100% (95% CI: 81.6% to 100%)",
        "Specificity": "94.1% (95% CI: 87.8% to 97.3%)"
      }
    }
  },
  "2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)": {
    "Total Sample Size": "1062 fresh lesion specimens (combined Cutaneous, Mucocutaneous, and Uncategorized)",
    "Data Type": "Prospective study",
    "Country of Origin": "United States (three sites across the United States)",
    "Specific Sample Sizes by Lesion Type for Clinical Performance": {
      "Cutaneous Lesions (HSV-1)": "249 specimens",
      "Cutaneous Lesions (HSV-2)": "275 specimens",
      "Cutaneous Lesions (VZV)": "222 specimens",
      "Mucocutaneous Lesions (HSV-1)": "501 specimens",
      "Mucocutaneous Lesions (HSV-2)": "610 specimens",
      "Mucocutaneous Lesions (VZV)": "372 specimens",
      "Uncategorized Lesions (HSV-1)": "148 specimens",
      "Uncategorized Lesions (HSV-2)": "166 specimens",
      "Uncategorized Lesions (VZV)": "119 specimens"
    }
  },
  "3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)": "Not specified. The ground truth was established by comparator methods (ELVIS cell culture system for HSV-1/2 and H&V mixed cells with DFA for VZV), which are FDA-cleared laboratory methods, rather than expert consensus on images or other forms of data typically requiring expert review.",
  "4. Adjudication method (e.g. 2+1, 3+1, none)  for the test set": "For discrepant results between the Solana assay and the primary comparator method, an additional RT-PCR assay was used for secondary adjudication. The document notes how many of the discrepant cases were resolved by this additional RT-PCR assay (e.g., 'Three (3) of the five (5) positives was positive by an additional RT-PCR assay' for HSV-1 cutaneous lesions). This indicates a form of adjudication involving a third, independent test.",
  "5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance": "Not applicable. This is a molecular diagnostic assay, not an imaging device that uses AI for interpretation by human readers. The clinical study compares the device's performance against established laboratory methods.",
  "6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done": "Yes, a standalone study was conducted. The Solana® HSV 1+2/VZV Assay is an in vitro diagnostic test that processes specimens and provides results using on-board method-specific algorithms on the Solana® instrument. The reported performance characteristics (clinical sensitivity and specificity) are for the device operating as a standalone system against the comparator methods.",
  "7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)": "Laboratory comparator methods: ELVIS cell culture system for HSV-1 and HSV-2, and H&V mixed cells with DFA cell culture systems for VZV. Discrepancies were resolved using an additional RT-PCR assay.",
  "8. The sample size for the training set": "Not explicitly stated for a distinct training set. The descriptions focus on the performance evaluation of the final device. For analytical performance (e.g., LOD, cross-reactivity, interference), contrived samples and specific strains were used for testing, but these are not referred to as a 'training set' in the context of machine learning. Clinical studies used fresh lesion specimens as described in section 2.",
  "9. How the ground truth for the training set was established": "Not applicable, as a distinct training set (in the machine learning sense) with established ground truth is not described for this diagnostic device. Analytical sensitivities (LOD) were determined using quantified viral cultures (TCID50/mL), which would serve as a 'truth' for analytical performance testing."
}

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 28, 2016

Ouidel Corporation Ronald H. Lollar Senior Director, Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, OH 45701

Re: K162451

Trade/Device Name: Solana® HSV 1+2/VZV Assay Regulation Number: 21 CFR 866.3309 Regulation Name: Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel Regulatory Class: II Product Code: PGI Dated: August 31, 2016 Received: September 1, 2016

Dear Mr. Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Stephen J. Lovell -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162451

Device Name Solana® HSV 1+2/VZV Assay

Indications for Use (Describe)

The Solana® HSV 1+2/VZV Assay is an in vitro diagnostic test, using isothermal amplification technology (helicasedependent amplification, HDA), for the qualitative detection and differentiation of herpes simplex virus type 1, herpes simplex virus type 2, and varicella-zoster virus DNA isolated and purified from cutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection. The Solana® HSV 1+2/VZV Assay is intended to aid in the diagnosis of herpes virus 1. herpes simplex virus 2 and varicella-zoster virus active cutaneous infections. Negative results do not preclude herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus infections and should not be used as the sole basis for diagnosis, treatment or other management decisions. The Solana® HSV 1+2/VZV Assay is intended for use only with the Solana® instrument.

Warning: The Solana® HSV 1+2/VZV Assay is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS). The Solana® HSV 1+2/VZV Assay is not intended for use in prenatal screening.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

{4}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 - Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

August 31, 2016

A. 510(k) Number:

K162451

B. Purpose for Submission:

To obtain substantial equivalence for the Solana® HSV 1+2/VZV Assay when performed on the Solana® instrument

C. Measurand:

HSV-1: US7: glycoprotein I

HSV-2: noncoding region between UL47: VP13/14 and UL48: VP16

VZV: ORF6: DNA-helicase primase

D. Type of Test:

{5}------------------------------------------------

Helicase-dependent amplification (HDA)

E. Applicant:

Quidel Corporation

F. Proprietary and Established Names:

Solana® HSV 1+2/VZV Assay

G. Regulatory Information:

Table 1.Regulatory Information
Product CodeClassificationRegulation SectionPanel
PGI — HerpesVirus (VZV, HSV-1, HSV-2) DNADetection Assayfor CutaneousandMucocutaneousLesion SamplesClass II(SpecialControls)21 CFR 866.3309 – Herpes virusnucleic acid-based cutaneousand mucocutaneous lesion panelMicrobiology(83)

H. Intended Use:

1. Intended Use(s):

The Solana® HSV 1+2/VZV Assay is an in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection and differentiation of herpes simplex virus type 1, herpes simplex virus type 2, and varicella-zoster virus DNA isolated and purified from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection. The Solana® HSV 1+2/VZV Assay is intended to aid in the diagnosis of herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus active cutaneous or mucocutaneous infections. Negative results do not preclude herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus infections and should not be used as the sole basis for diagnosis, treatment or other management decisions.

{6}------------------------------------------------

The Solana® HSV 1+2/VZV Assay is intended for use only with the Solana® instrument.

Warning: The Solana® HSV 1 + 2/VZV Assay is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS). The Solana® HSV 1 + 2/VZV Assay is not intended for use in prenatal screening.

2. Indication(s) for Use:

Same as intended Use

    1. Special conditions for use statement(s):
    • For in vitro diagnostic use only
    • For prescription use only
    1. Special instrument requirements:

Solana® instrument

l. Device Description:

The Solana® HSV 1+2/VZV Assay amplifies and detects viral DNA isolated from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection.

The assay consists of two (2) major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to HSV-1, HSV-2 and/or VZV using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

Patient specimen is transferred to a Process Tube, subjected to heat treatment at 95°C for 5 minutes and mixed and vortexed. The processed sample is transferred to a Reaction Tube and mixed. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of specific target sequence. In Solana, the target sequences are amplified by HSV-1, HSV-2 and/or VZV specific primers and detected by HSV-1, HSV-2 and/or VZV specific fluorescence probes included in the Reaction Tube. A competitive process control (PRC) is included in the Process Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by specific primers and detected by a PRC specific fluorescence probe.

{7}------------------------------------------------

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to HSV-1, HSV-2, VZV or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana will then report the test results to the user on its display screen, and the results can be printed via an attached printer.

Materials Provided:

Cat. #M302

48 Tests per Kit

Table 2.Kit Components
ComponentQuantityStorage
Process Buffer Tubes48 tubes/kit 1.6 mL2°C to 8°C
Reaction Tubes48 tubes/kit2°C to 8°C

Materials required but not provided:

  • . External controls for HSV-1, HSV-2, or VZV (e.g. Quidel Molecular HSV/VZ Control Set, which contains positive and negative controls, serves as an external processing and extraction control)
  • Sterile DNase-free filter-blocked or positive displacement micropipettor tips ●
  • Micropipettor ●
  • Stopwatch or timer ●
  • Scissors or a blade ●
  • Heat block capable of 95° C ± 2° C temperature ●
  • Solana workflow tray and transfer rack ●
  • Solana Instrument
  • Thermometer

Substantial Equivalence Information: J.

    1. Predicate device name(s):
      Lyra® Direct HSV 1+2/VZV Assay
    1. Predicate 510(k) number(s): K133448

{8}------------------------------------------------

3. Comparison with predicate:

Table 3. Similarities
ItemSubject DeviceSolana® HSV 1+2/VZV AssaySubject DeviceLyra® Direct HSV 1 + 2/VZVAssay K133448
Intended UseThe Solana® HSV 1+2/VZV Assay is an in vitro diagnostic test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection and differentiation of herpes simplex virus type 1, herpes simplex virus type 2, and varicella-zoster virus DNA isolated and purified from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection. The Solana® HSV 1+2/VZV Assay is intended to aid in the diagnosis of herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus active cutaneous or mucocutaneous infections. Negative results do not preclude herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus infections and should not be used as the sole basis for diagnosis, treatment or other management decisions. The Solana® HSV 1+2/VZV Assay is intended for use only with the Solana® instrument.Warning: The Solana® HSV 1 + 2/VZV Assay is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections ofThe Lyra® Direct HSV 1 + 2/VZV Assay is an in vitro multiplex Real-Time PCR test for qualitative detection and differentiation of herpes simplex virus type 1, herpes simplex virus type 2, and varicella-zoster virus DNA isolated and purified from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or varicella-zoster infection. The Lyra® Direct HSV 1 + 2/VZV Assay is intended to aid in the diagnosis of herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus active cutaneous or mucocutaneous infections. Negative results do not preclude herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus infections and should not be used as the sole basis for diagnosis, treatment or other management decisions. The Lyra® Direct HSV 1 + 2/VZV Assay is not intended for use with cerebral spinal fluid. The
Table 3.      Similarities
ItemSubject DeviceSolana® HSV 1+2/VZV AssaySubject DeviceLyra® Direct HSV 1 + 2/VZVAssay K133448
the central nervous system. TheSolana® HSV 1 + 2/VZV Assay is notintended for use in prenatalscreening.Lyra® Direct HSV 1 + 2/VZVAssay is not intended for usein prenatal screening. Thedevice is not intended forpoint-of-care use.
DetectionTechniquesMultiplex assay using differentreporter dyes for each targetMultiplex assay usingdifferent reporter dyes foreach target
Identification ofHSV-1, HSV-2, andVZVYesYes
Assay ResultsQualitativeQualitative
Sample Typescutaneous or mucocutaneous lesionswab specimens obtained fromsymptomatic patientscutaneous or mucocutaneouslesion swab specimensobtained from symptomaticpatients
Extraction MethodsNot requiredNot required

{9}------------------------------------------------

Table 4. Differences
ItemSubject DeviceSolana® HSV 1+2/VZV AssaySubject DeviceLyra® Direct HSV 1 + 2/VZVAssay K133448
Viral TargetHSV-1: US7: glycoprotein IHSV-1: glycoprotein G
HSV-2: noncoding region betweenUL47: VP13/14 and UL48: VP16HSV-2: glycoprotein G
VZV: ORF6: DNA-helicase primaseVZV: ORF6: DNA-helicaseprimase

{10}------------------------------------------------

Table 4. Differences
ItemSubject DeviceSolana® HSV 1+2/VZV AssaySubject DeviceLyra® Direct HSV 1 + 2/VZVAssay K133448
Assay MethodologyHelicase-Dependent Amplification(HDA) detecting the presence orabsence of viral DNA in clinicalspecimensPCR-based system fordetecting the presence orabsence of viral DNA inclinical specimens
Amplification andDetectioninstrumentsSolana® instrumentLife TechnologiesQuantStudio™ Dx, theApplied Biosystems® 7500Fast Dx, or the CepheidSmartCycler® II System

K. Standard/Guidance Document Referenced (if applicable):

Not applicable

L. Test Principle:

Patient specimen is transferred to a Process Tube, subjected to heat treatment at 95°C for 5 minutes and mixed and vortexed. The processed sample is transferred to a Reaction Tube and mixed. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of specific target sequence. In Solana, the target sequences are amplified by HSV-1, HSV-2 and/or VZV specific primers and detected by HSV-1, HSV-2 and/or VZV specific fluorescence probes included in the Reaction Tube. A competitive process control (PRC) is included in the Process Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to HSV-1, HSV-2, VZV or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board

{11}------------------------------------------------

method-specific algorithms. Solana will then report the test results to the user on its display screen, and the results can be printed via an attached printer.

M. Performance Characteristics:

1. Analytical performance:

a. Precision/Reproducibility:

Reproducibility

The reproducibility of the Solana HSV 1+2/VZV Assay was evaluated at three laboratory sites. A reproducibility panel containing 30 contrived samples, manufactured as high negative samples (n=3; 1/18x or 1/27x LOD (C₂o – ၆၈ concentration)) for HSV-1 MacIntyre, HSV-2 G, and VZV Ellen, low positive samples (n=3; near the assay limit of detection) for HSV-1, HSV-2 and VZV, moderate positive samples (n=3; 3x LOD) for HSV-1, HSV-2 and VZV and negative samples (n=3) was used for the study. The samples were randomized and blind-coded within each panel, and the operator tested one (1) panel, together with three (3) positive and three (3) negative external controls, in three runs. The panels were run by two operators at each testing site for five (5) non-consecutive days.

Table 5. Reproducibility Summary
SITEOverall Percent Agreement
HSV-1 MacIntyreSite #1Site #2Site #395% Confidence Interval
Rate of Detection% AgreementRate of Detection% AgreementRate of Detection% AgreementRate of Detection% Agreement
High Negative(3.50× 101 TCID50/mL)11/3036.75/3016.714/3046.730/9033.324.5 to 43.0
Low Positive(6.30 × 102 TCID50/mL)30/3010030/3010030/3010090/9010095.9 to 100
Moderate Positive(1.89 × 103 TCID50/mL)30/3010030/3010030/3010090/9010095.9 to 100
Negative0/301000/301000/301000/9010095.9 to 100
Positive Control30/3010030/3010030/3010090/9010095.9 to 100
Negative Control0/301000/301000/301000/9010095.9 to 100

{12}------------------------------------------------

SITEOverall PercentAgreement
HSV-2 GSite #1Site #2Site #3Overall PercentAgreement95%ConfidenceInterval
Rate ofDetection%AgreementRate ofDetection%AgreementRate ofDetection%AgreementRate ofDetection%Agreement
High Negative(3.71 × 102 TCID50/mL)12/30408/3026.77/3023.327/9030.021.5 to 40.
Low Positive(6.67 × 103 TCID50/mL)30/3010030/3010030/3010090/9010095.9 to 100
Moderate Positive(2.00 × 104 TCID50/mL)30/3010030/3010030/3010090/9010095.9 to 10
Negative0/301000/301000/301000/9010095.9 to 100
Positive Control30/3010030/3010030/3010090/9010095.9 to 100
Negative Control0/301000/301000/301000/9010095.9 to 100
SITEOverall PercentAgreement
VZV EllenSite #1Site #2Site #3Overall PercentAgreement95%ConfidenceInterval
Rate ofDetection%AgreementRate ofDetection%AgreementRate ofDetection%AgreementRate ofDetection%Agreement
High Negative(5.50 × 10-3 TCID50/mL)11/3036.72/306.78/3026.721/9023.315.8 to 33.
Low Positive(1.49 × 10-1 TCID50/mL)30/3010030/3010030/3010090/9010095.9 to 100
Moderate Positive(4.46 × 10-1 TCID50/mL)30/3010030/3010030/3010090/9010095.9 to 100
Negative0/301000/301000/301000/9010095.9 to 100
Positive Control30/3010030/3010030/3010090/9010095.9to 100

The results suggest that there are no significant differences between different users using different instruments at different sites on different days.

  • b. Linearity/assay reportable range:
    Not applicable – This assay is qualitative.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Traceability:

Not applicable. This assay is qualitative.

Specimen Stability:

{13}------------------------------------------------

A study was performed to demonstrate specimen stability in various transport media types: M4, M4RT, M5, M6, and UTM when stored at room temperature (30 °C) for up to 48 hours, ≤ -20 °C, or 2 to 8 °C for up to 7 days.

HSV-1, HSV-2, and VZV virus stocks (HSV-1 McIntyre, HSV-2 G, and VZV Ellen) were diluted to their respective 2x LOD concentrations in each in of five (5) transport medium pooled negative matrix (Remel M4, Remel M6, Remel M6, Remel M4RT or UTM transport medium).

The transport media systems containing the contrived samples were stored T at three different conditions: Condition 1 = room temperature (30 ± 2 °C) for 50 hours, Condition 2 = 2 to 8 °C for 8 days, and Condition 3 = -20 °C for 8 days.

For Condition 1, the specimens were stored at room temperature (30 ± 2 °C) and tested at 0, 24, 48, and 50 hours post-storage. For Condition 2, the specimens were stored at 2-8 °C and tested at Day 0, Day 1, Day 3, Day 4, Day 7, and Day 8 poststorage. For Condition 3, the specimens were stored at -20 °C and tested at Day 0, Day 1, Day 3, Day 4, Day 7, and Day 8 post-storage.

Based on this study, HSV-1, HSV-2, and VZV samples at 2x LOD were stable when stored in M4, M4-RT, M5, M6, and UTM negative matrix for up to 50 hours at 30 °C ± 2 °C. HSV-1, HSV-2, and VZV samples were stable when stored in M4, M4-RT, M5, M6, and UTM negative matrix for up to 8 days at 2° to 8 °C. HSV-1, HSV-2, and VZV samples were stable when stored in M4, M4-RT, M5, M6, and UTM negative matrix for up to 8 days at 2° to 8 °C.

Controls:

. External controls for HSV-1, HSV-2, or VZV (e.g. Quidel Molecular HSV/VZV Control Set, which contains positive and negative controls, serves as an external processing and extraction control) were run on the Solana HSV 1+2/VZV Assay each day of testing.

The assay controls are described as follows:

  • a. The process control is used to monitor sample processing, to detect HDA inhibitory specimens and to confirm the integrity of assay reagents and Solana instrument functionality. The process control is included in the Process Buffer tube.

{14}------------------------------------------------

  • The external positive control may be treated as a patient specimen. The b. control should be sampled and tested as if it were a specimen and processed as described in the Assay Procedure. The external positive control is intended to monitor substantial reagent and instrument failure.
  • c. The external negative control may be treated as a patient specimen. The control should be sampled and tested as if it were a specimen and processed as described in the Assay Procedure. The external negative control is used to detect reagent or environmental contamination (or carry-over) by HSV 1+2/VZV DNA or amplicon.

d. Detection limit:

The analytical sensitivity (limit of detection or LOD) of the Solana HSV 1+2/VZV Assay was determined using quantified (TCID50/mL) cultures of two (2) HSV-1 strains, two (2) HSV-2 strains, and two (2) VZV strains, serially diluted in negative matrix. Each dilution was run as 20 replicates in the Solana HSV 1+2/VZV assay. Analytical sensitivity (LOD) is defined as the lowest concentration at which at least 95% of all replicates tested positive. The demonstrated LOD for each strain tested is shown below:

Table 6.LOD Values
VirusTCID50/mL
HSV-1 MacIntyre6.30 × 102
HSV-1 3165.47 × 104
HSV-2 G6.67 × 103
HSV-2 COMP1.62 × 105
VZV Ellen1.49 × 10-1
VZV 99391.65 × 102
  • e. Analytical specificity:

Cross Reactivity:

A study was performed to evaluate the performance of the Solana HSV 1+2/VZV Assay in the presence of sixty-four (64) organisms that might be found in lesion specimens. Each potentially cross-reactive microorganism was tested at clinically relevant levels of viruses and bacteria: ≥10° CFU/mL; viruses: ≥ 10° copies (cp), viral particles (vp) or TCID50/mL.

{15}------------------------------------------------

Table 7.Potentially Cross-reactive Organisms
OrganismTestConcentrationOrganismTestConcentration
Acholeplasma laidlawi7.10E+06 CFU/mLKlebsiella pneumoniae1.61E+06 CFU/mL
Acinetobacter calcoaceticus9.27E+06 CFU/mLLactobacillus acidophilus2.49E+06 CFU/mL
Adenovirus 71.58E+05 TCID50/mLLegionella pneumophila1.76E+06 CFU/mL
Bacteroides fragilis1.19E+06 CFU/mLMeasles virus1.95E+05 TCID50/mL
Bordetella bronchiseptica1.97E+06 CFU/mLMobiluncus mulieris2.54E+06 CFU/mL
Bordetella pertussis7.21E+06 CFU/mLMoraxella cartarrhalis1.26E+06 CFU/mL
Candida albicans2.00E+06 CFU/mLMumps virus5.89E+05 TCID50/mL
Candida glabrata3.93E+06 CFU/mLMycoplasma hominis1.30E+06 CFU/mL
Chlamydia trachomatis3.00E+06 CFU/mLMycoplasma hyorhinis6.60E+06 CFU/mL
Chlamydophila pneumoniae1.25E+06 IFU/mLMycoplasma orale3.08E+06 CFU/mL
Clostridium perfringens1.06E+06 CFU/mLMycoplasma pneumoniae3.16E+06 CFU/mL
Coronavirus OC438.51E+05 TCID50/mLMycoplasma salivarium1.67E+06 CFU/mL
Corynebacterium diphtheriae1.51E+06 CFU/mLNeisseria gonorrhoeae1.23E+06 CFU/mL
Coxsackievirus B43.16E+05 TCID50/mLParainfluenza Type 13.97E+05 TCID50/mL
Cytomegalovirus Towne VR-9772.14E+05 TCID50/mLParainfluenza Type 23.15E+05 TCID50/mL
Echovirus 112.14E+05 TCID50/mLParainfluenza Type 32.56E+05 TCID50/mL
Enterococcus faecalis3.45E+06 CFU/mLParainfluenza Type 41.37E+05 TCID50/mL
Enterovirus 701.78E+05 TCID50/mLProteus mirabilis1.19E+06 CFU/mL
Epstein Barr Virus1.34E+05 vp/mLPseudomonas aeruginosa1.32E+06 CFU/mL
Table 7. Potentially Cross-reactive Organisms
OrganismTest ConcentrationOrganismTest Concentration
Escherichia coli8.42E+06 CFU/mLRSV A Long1.95E+05 TCID50/mL
Gardnerella vaginalis1.20E+06 CFU/mLRSV B Washington3.43E+05 TCID50/mL
Haemophilis influenza type A5.33E+06 CFU/mLRubella Virus2.09E+05 TCID50/mL
HBV synthetic DNA6.80E+05 cp/mLSalmonella enteriditis5.40E+06 CFU/mL
HCV synthetic RNA1.96E+05 cp/mLSalmonella typhimurium1.01E+06 CFU/mL
HHV-63.30E+05 TCID50/mLStaphylococcus aureus1.02E+06 CFU/mL
HHV-71.15E+05 TCID50/mLStaphylococcus saprophyticus2.00E+06 CFU/mL
HHV-81.26E+05 TCID50/mLStreptococcus agalactiae2.20E+06 CFU/mL
HIV purified RNA1.60E+05 cp/mLStreptococcus pneumoniae2.18E+06 CFU/mL
hMPV A13.66E+05 TCID50/mLStreptococcus pyogenes1.29E+06 CFU/mL
HPV4.30E+05 cp/uLToxoplasma gondii1.06E+06 tachyzoites/mL
Influenza A/Mexico/4108/20092.88E+05 vp/mLTrichomonas vaginalis1.00E+06 trophozoites/mL
Influenza B Hong Kong VR-7911.91E+05 TCID50/mLUreaplasma uralyticum1.23E+06 CFU/mL

{16}------------------------------------------------

No cross-reactivity was observed with the sixty-four (64) microorganisms tested with the Solana HSV 1+2/VZV Assay. Interference:

The performance of Solana HSV 1+2/VZV Assay was evaluated with potentially interfering substances that may be present in lesion specimens. A panel composed of twenty-six (26) substances was tested in the absence or presence of HSV-1, HSV-2, or VZV (Maclntyre, G, Ellen strains, respectively) at 2X LOD in the Solana HSV 1+2/VZV Assay. There was no evidence of interference caused by the substances tested at the concentrations shown below.

{17}------------------------------------------------

Table 8.Potential Interfering Substances and Concentrations Tested
SubstanceTestConcentrationSubstanceTestConcentration
Abreva7%Female Urine7%
Acetamidophenol10 mg/mLKY Jelly7%
Acyclovir7 mg/mLLanacane3.50%
Albumin3.3 mg/mLLeukocytes2.5x105 cells/mL
Blood/EDTA0.63%Listerine7%
Carmex7%Male Urine7%
Casein7 mg/mLMiconazole 17%
Chlorpheniramine5 mg/mLMiconazole 37%
Colgate7%Mucin60 µg/mL
Cornstarch2.5 mg/mLPreparation H7%
Dextromethorphan5 mg/mLReleev7%
Douche7%Seminal Fluid2%
Feces0.22%Tioconazole 17%

Analytical Reactivity (Inclusivity):

The inclusivity of the Solana HSV 1+2/VZV Assay was further evaluated by functional testing of viral strains in addition to those strains used in the LOD study. The clinical panel consisted of two (2) strains of HSV-1, three (3) strains of HSV-2, and five (5) strains of VZV at concentrations near the level of detection (LOD) of the assay.

Table 9. Inclusivity Strains
StrainTCID50/mLInclusive (Yes or No)
HSV-1 Isolate #11.26 × $10^3$Yes
HSV-1 Isolate #31.26 × $10^3$Yes
HSV-2 Strain MS1.33 × $10^4$Yes
HSV-2 Isolate #251.33 × $10^4$Yes

{18}------------------------------------------------

Table 9. Inclusivity Strains
StrainTCID50/mLInclusive (Yes or No)
HSV-2 Isolate #321.33 × 104Yes
VZV Strain 828.05 × 100Yes
VZV Strain 1302.41 × 101Yes
VZV Strain 2758.05 × 100Yes
VZV Strain B2.41 × 101Yes
VZV Strain D8.05 × 100Yes
  • f. Assay cut-off:
    Not applicable.

2. Comparison studies:

  • a. Method comparison with predicate device:
    Not applicable

  • b. Matrix comparison:
    Not applicable

    1. Clinical studies:
    • Clinical Sensitivity: a.

Performance characteristics of the Solana " HSV 1+2/VZV Assay were established during a prospective study between February and April 2016. One thousand sixtytwo (1062) fresh lesion specimens, collected for herpes simplex/varicella-zoster identification, have been included in this study at three sites across the United States. A single specimen was collected per patient. The specimens were processed and tested with Solana " HSV 1+2/VZV Assay on the Solana "instrument at the sites. Testing of the comparator methods (culture with DFA) were performed at one central location.

{19}------------------------------------------------

Combined Data

Cutaneous Lesions

Two-hundred seventy-five (275) active cutaneous lesion specimens were cultured for HSV-1, using the FDA-cleared ELVIS cell culture system and were also tested with the subject device for HSV-1 viral DNA. Samples which gave HSV-2 positive results from the ELVIS method were excluded from the HSV-1 performance calculation because of the inability of the ELVIS method to distinguish an HSV-1 positive sample when HSV-2 was detected first (n=26). The table below details the HSV-1 results for the remaining two-hundred forty-nine (249) specimens.

Table 10. HSV-1 Results
Comparator: ELVIS® HSV ID and D3 Typing Test
Solana HSV 1 + 2/VZV AssayPositiveNegativeTotal
Positive225*27
Negative0222222
Total22227249
95% CI
Sensitivity22/22100%85.1% to 100%
Specificity222/22797.8%95.0% to 99.1%
  • Three (3) of the five (5) positives was positive by an additional RT-PCR assay.

Two-hundred seventy-five (275) active cutaneous lesion specimens were cultured for HSV-2, using the FDA-cleared ELVIS cell culture system and were also tested with the subject device for HSV-2 viral DNA. The table below details the HSV-2 results for the two-hundred seventy-five (275).

Table 11.HSV-2 Results
Comparator: ELVIS® HSV ID and D³ Typing Test
Solana HSV 1 + 2/VZVAssayPositiveNegativeTotal
Positive2414*28
Negative2**235237
Total26249275
95% CI
Sensitivity24/2692.3%75.9% to 97.9%
Specificity235/24994.4%90.8% to 96.6%
  • Thirteen (13) of the fourteen (14) positives were positive by an additional RT-PCR assay.

** Two (2) of the two (2) negatives were positive by an additional RT-PCR assay.

{20}------------------------------------------------

Two-hundred and seventy-five (275) active cutaneous lesion specimens were cultured for VZV using the H&V mixed cells with DFA cell culture systems and were also tested with the subject device for VZV viral DNA. Due the presence of either HSV-1 or HSV-2, fifty-one (51) specimens have been excluded from analysis. Two (2) specimens were contaminated or had toxic cultures. These fifty-three (53) specimens have been excluded from analysis. The table below details the VZV results for the remaining two-hundred and twenty-two (222) specimens.

Table 12.VZV Results
Comparator: DSFA and Culture with DFA
Solana HSV 1 + 2/VZV AssayPositiveNegativeTotal
Positive227*29
Negative0193193
Total22200222
95% CI
Sensitivity22/22100%85.1% to 100%
Specificity193/20096.5%93.0% to 98.3%
  • Six (6) of the seven (7) positives were positive by an additional RT-PCR assay.

Mucocutaneous Lesions

Six-hundred twenty-one (621) active mucocutaneous lesion specimens were cultured for HSV-1, using the FDA-cleared ELVIS cell culture system and were also tested with the subject device for HSV-1 viral DNA. Seven (7) specimens were contaminated in the ELVIS cell culture. Two (2) specimens were invalid in Solana HSV 1 + 2/VZV Assay. Two (2) specimens were contaminated and invalid. Samples which gave HSV-2 positive results from the ELVIS method were excluded from the HSV-1 performance calculation because of the inability of the ELVIS method to distinguish an HSV-1 positive sample when HSV-2 was detected first (n=109). Thus, one hundred and twenty (120) specimens have been excluded from further analysis. The table below details the HSV-1 results for the remaining five-hundred and one (501) specimens.

{21}------------------------------------------------

Table 13.HSV-1 Results
Comparator: ELVIS® HSV ID and D3 Typing Test
Solana HSV 1 + 2/VZV AssayPositiveNegativeTotal
Positive11314*127
Negative0374374
Total113388501
95% CI
Sensitivity113/113100%96.7% to 100%
Specificity374/38896.4%94.0% to 97.8%
  • Six (6) of the fourteen (14) positives were positive by an additional RT-PCR assay.

Six-hundred twenty-one (621) active mucocutaneous lesion specimens were cultured for HSV-2, using the FDA-cleared ELVIS cell culture system and were also tested with the subject device for HSV-2 viral DNA. Seven (7) specimens were contaminated in the ELVIS cell culture. Two (2) specimens were invalid in Solana HSV 1 + 2/VZV Assay. Two (2) specimens were contaminated and invalid. These eleven (11) specimens have been excluded from further analysis. The table below details the HSV-2 results for the remaining six-hundred ten (610) specimens.

Table 14.HSV-2 Results
Comparator: ELVIS® HSV ID and D³ Typing Test
Solana HSV 1 + 2/VZV AssayPositiveNegativeTotal
Positive10814*122
Negative1**487488
Total109501610
95% CI
Sensitivity108/10999.1%95.0% to 99.8%
Specificity487/50197.2%95.4% to 98.3%
  • Eleven (11) of the fourteen (14) positives were positive by an additional RT-PCR assay.

** One (1) of one (1) negative was positive by an additional RT-PCR assay.

Six-hundred twenty one (621) active mucocutaneous lesion specimens were cultured for VZV using the H&V mixed cells with DFA cell culture systems and were also tested with the subject device for VZV viral DNA. Due the presence of either HSV-1 or HSV-2, two hundred thirty-six (236) specimens have been excluded from analysis. Nine (9) specimens were contaminated in culture, and four (4) specimens were invalid in Solana HSV 1 + 2/VZV Assay. These two hundred forty-nine (249) specimens have been excluded from analysis. The

{22}------------------------------------------------

table below details the VZV results for the remaining three hundred seventy-two (372) specimens.

Table 15. VZV Results
Comparator: DSFA and Culture with DFA
Solana HSV 1 + 2/VZV AssayPositiveNegativeTotal
Positive45*9
Negative0363363
Total4368372
95% CI
Sensitivity4/4100%51.0% to 100%
Specificity363/36898.6%96.9% to 99.4%
  • One (1) of the five (5) positives was positive by an additional RT-PCR assay.

Note: The data presented for the detection of VZV is consistent with limited presence of VZV in mucocutaneous lesions. The use of mucocutaneous lesions has no discernible impact on the performance characteristics of Solana HSV 1 + 2/VZV Assay.

Uncategorized Lesions

One hundred sixty-six (166) active lesion specimens (not categorized as cutaneous or mucocutaneous) were cultured for HSV-1, using the FDA-cleared ELVIS cell culture system and were also tested with the subject device for HSV-1 viral DNA. Samples which gave HSV-2 positive results from the ELVIS method were excluded from the HSV-1 performance calculation because of the inability of the ELVIS method to distinguish an HSV-1 positive sample when HSV-2 was detected first (n=18). The table below details the HSV-1 results for the remaining one hundred and forty-eight (148) specimens.

Table 16.HSV-1 Results
Comparator: ELVIS ® HSV ID and D3 Typing Test
Solana HSV 1 + 2/VZVAssayPositiveNegativeTotal
Positive253*28
Negative0120120
Total25123148
95% CI
Sensitivity22/22100%86.7% to 100%
Specificity120/12397.6%93.1% to 99.2%
  • Three (3) of the three (3) positives was positive by an additional RT-PCR assay.

{23}------------------------------------------------

One hundred sixty-six (166) active lesion specimens (not categorized as cutaneous or mucocutaneous) were cultured for HSV-2, using the FDA-cleared ELVIS cell culture system and were also tested with the subject device for HSV-2 viral DNA. The table below details the HSV-2 results for the one hundred sixtysix (166).

Table 17.HSV-2 Results
Comparator: ELVIS® HSV ID and D3 Typing Test
Solana HSV 1 + 2/VZVAssayPositiveNegativeTotal
Positive185*23
Negative0143143
Total18148166
95% CI
Sensitivity18/18100%82.4% to 100%
Specificity143/14896.6%92.3% to 98.6%
  • Five (5) of the five (5) positives were positive by an additional RT-PCR assay.

One hundred sixty-six (166) active lesion specimens (not categorized as cutaneous or mucocutaneous) were cultured for VZV using the H&V mixed cells with DFA cell culture systems and were also tested with the subject device for VZV viral DNA. Due the presence of either HSV-1 or HSV-2, forty-six (46) specimens have been excluded from analysis. One (1) specimen was contaminated in culture. These forty-seven (47) specimens have been excluded from analysis. The table below details the VZV results for the remaining one hundred nineteen (119) specimens.

Table 18.VZV Results
Comparator: DSFA and Culture with DFA
Solana HSV 1 + 2/VZV AssayPositiveNegativeTotal
Positive176*23
Negative09696
Total17102119
95% CI
Sensitivity17/17100%81.6% to 100%
Specificity96/10294.1%87.8% to 97.3%
  • Five (5) of the six (6) positives were positive by an additional RT-PCR assay.

{24}------------------------------------------------

  • b. Clinical specificity:
    See Section 3a.

  • Other clinical supportive data (when a. and b. are not applicable): ﻥ
    Not applicable

    1. Clinical cut-off:
      Not applicable

5. Expected values:

The expected values of the Solana HSV 1+2/VZV Assay were established during a prospective study conducted between February and May 2016. One thousand sixty-two (1062) specimens have been included in this study at three (3) sites across the United States. A single specimen was collected per patient. The specimens were processed and tested with Solana HSV 1+2/VZV Assay on the Solana instrument at the sites.

The expected value of HSV-1, HSV-2, and VZV with the Solana HSV 1+2/VZV Assay has been calculated for the combined sites based on the category of specimen (cutaneous, mucocutaneous, or uncategorized lesion source) and the age of the patient.

Table 19. Combined Study – Expected Values (Cutaneous) (N=273)
HSV-1HSV-2VZV
AgeTotal #Total PositivePrevalenceTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
≤ 5 years9111.1%90N/A9444.4%
6 to 21 years431023.3%4349.3%430N/A
22 to 59 years180168.9%1802614.4%1801910.6%
≥ 60 years430N/A43818.6%43614.0%
Table 20.Expected Values (Cutaneous) (N=273)
HSV- 1HSV-2VZV
Total #TotalPositivePrevalenceTotal #TotalPositivePrevalenceTotal #TotalPositivePrevalence
Genital - penis103109.7%1031817.5%10311.0%
skin lesion1701710.0%1701911.2%1702816.5%

{25}------------------------------------------------

Table 21. Combined Study – Expected Values (Mucocutaneous) (N=617)*
HSV-1HSV-2VZV
AgeTotal #Total PositivePrevalenceTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
≤ 5 years21419.0%210N/A2114.8%
6 to 21 years1584125.9%1582918.4%1580N/A
22 to 59 years3857419.2%3858923.1%38582.1%
≥ 60 years531120.8%5359.4%5311.9%
  • Four (4) specimens were invalid and removed from analysis
Table 22. Expected Values (Mucocutaneous) (N=619)*
HSV-1HSV-2VZV
Total#TotalPositivePrevalenceTotal #TotalPositivePrevalenceTotal #TotalPositivePrevalence
Anorectal26726.9%26726.9%2613.8%
genital – vaginal/cervical4498017.8%44911224.9%44951.1%
Nares23521.7%2314.3%23313.0%
Ocular7228.6%70N/A7114.3%
Oral lesion1123632.1%11232.7%1120N/A
  • Four (4) specimens were invalid and removed from analysis
Table 23. Combined Study – Expected Values (Uncategorized Lesion Source) (N=166)
HSV-1HSV-2VZV
AgeTotal #Total PositivePrevalenceTotal #Total PositivePrevalenceTotal #Total PositivePrevalence
≤ 5 years1119.1%110N/A110N/A
6 to 21 years281035.7%280N/A2827.1%
22 to 59 years891516.9%891820.2%891415.7%
≥ 60 years3825.3%38513.2%38821.1%

N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Instrument: Solana™ Instrument

O. System Descriptions:

  1. Modes of Operation:

{26}------------------------------------------------

The Solana instrument heats each reaction tube to 64°C. If present, the target sequence is amplified by HSV-1, HSV-2 and/or VZV specific primers and detected by HSV-1, HSV-2 and/or VZV specific fluorescence probes included in the Reaction Tube. Each probe has a florescent dye of specific wavelength. The target probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target probes carry a ribonucleic acid. Upon annealing to the respective amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. The Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes No

P. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and the special controls.

§ 866.3309 Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel.

Link to an amendment published at 90 FR 55982, Dec. 4, 2025. (a)
Identification. A herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel is a qualitative in vitro diagnostic device intended for the simultaneous detection and differentiation of different herpes viruses in cutaneous and mucocutaneous lesion samples from symptomatic patients suspected of Herpetic infections. Negative results do not preclude infection and should not be used as the sole basis for treatment or other patient management decisions. The assay is not intended for use in cerebrospinal fluid samples.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer design and selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation of a clinical study using lesion samples in which Herpes Simplex Virus 1, Herpes Simplex Virus 2, or Varicella Zoster Virus DNA detection was requested. The study must compare the device performance to an appropriate well established reference method.
(4) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(5) The device labeling must include a limitation statement that reads: “The device is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS).”
(6) Premarket notification submissions must include quality assurance protocols and a detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(7) The risk management activities performed as part of the manufacturer's 21 CFR 820.30 design controls must document an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.