K Number
K162417
Date Cleared
2016-09-28

(29 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACUSON Freestyle™ Utrasound System is intended for diagnostic imaging or fluid flow analysis of the human body including: Fetal. Abdominal, Intraoperative Neurological, Pediatric, Small Organ. Neonatal Cephalic. Cardiac, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal Superficial.

Device Description

The ACUSON Freestyle™ Ultrasound System is an ultrasound imaging system, which operates with linear, curvilinear, array transducers. The transducers may be used in either a wireless or wired configuration through a cable connected to the system.

AI/ML Overview

The provided text is a 510(k) summary for the ACUSON Freestyle™ Ultrasound System, detailing its substantial equivalence to previously cleared predicate devices. It focuses on the device's technical characteristics, intended use, and compliance with various safety standards.

However, the document explicitly states that "clinical data is not required for substantial equivalence." This means it does not contain information about specific acceptance criteria related to device performance in a clinical setting, nor does it describe a study that proves the device meets such criteria.

Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria and performance studies, including sample size, data provenance, ground truth establishment, or multi-reader multi-case studies, because this type of clinical performance data is not present in the provided text.

The document mainly focuses on comparing the new device's technical specifications and intended uses to those of predicate devices to demonstrate substantial equivalence, rather than providing performance metrics from clinical trials.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2016

SIEMENS MEDICAL SOLUTIONS, INC. % Mr. MARK JOB THIRD PARTY REVIEWER REGULATORY TECHNOLOGY SERVICES LLC 1394 25TH STREET, NW BUFFALO MN 55313

Re: K162417

Trade/Device Name: Acuson Freestyle Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 29, 2016 Received: August 30, 2016

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

ACUSON FreestyleTM Ultrasound System

Indications for Use (Describe)

The ACUSON Freestyle™ Utrasound System is intended for diagnostic imaging or fluid flow analysis of the human body including: Fetal. Abdominal, Intraoperative Neurological, Pediatric, Small Organ. Neonatal Cephalic. Cardiac, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal Superficial.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Number (if known):

ACUSON FreestyleTM Diagnostic Ultrasound System Device Name:

Diagnostic Imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation
ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify:(Note 2)Other:HarmonicImaging
Opthalmic
FetalPPP
AbdominalPPP
Intraoperative (Note 1)PPP
Intraoperative NeurologicalPPP
PediatricPPP
Small Organ (Note 3)PPP
Neonatal CephalicPPP
Adult Cephalic
CardiacPPP
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VesselPPP
Laparoscopic
Musculo-skeletalConventionalPPP
Musculo-skeletal SuperficialPPP
Other

N=new indication, P = Previously Cleared, Blank: Not Claimed

Additional Comments:

Note 1: For example Cardiac Note 2: B Mode and PWD mode, or Cardiac Doppler and PW Mode Note 3: Breast, Testes, Thyroid, Penis

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{4}------------------------------------------------

510(k) Number (if known):

L8-3 Linear Transducer Device Name:

Intended Use:

Diagnostic Imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify:Note 2)Other:HarmonicImaging
Opthalmic
FetalPPP
AbdominalPPP
Intraoperative (Note 1)PPP
Intraoperative NeurologicalPPP
PediatricPPP
Small Organ (Note 3)PPP
Neonatal CephalicPPP
Adult Cephalic
CardiacPPP
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VesselPPP
Laparoscopic
Musculo-skeletalConventionalPPP
Musculo-skeletal SuperficialPPP
Other

N=new indication, P = Previously Cleared, Blank: Not Claimed

Additional Comments: Note 1: For example Cardiac Note 2: B Mode and PWD mode, or color Doppler and PW Mode Note 3: Breast, Testes, Thyroid, Penis

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{5}------------------------------------------------

510(k) Number (if known):

Device Name:

L12-5 Linear Transducer

Intended Use:

Diagnostic Imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
ABMPWDCWDColorAmplitudeDoppler DopplerColorVelocityImagingCombined(Specify:Note 2)Other: HarmonicImaging
Opthalmic
FetalPPP
AbdominalPPP
Intraoperative (Note 1)PPP
Intraoperative NeurologicalPPP
PediatricPPP
Small Organ (Note 3)PPP
Neonatal CephalicPPP
Adult Cephalic
CardiacPPP
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VesselPPP
Laparoscopic
Musculo-skeletal ConventionalPPP
Musculo-skeletal SuperficialPPP
Other

N=new indication, P = Previously Cleared, Blank: Not Claimed

Additional Comments:

Note 1: For example Cardiac Note 2: B Mode and PWD mode, or color Doppler and PW Mode Note 3: Breast, Testes, Thyroid, Penis

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{6}------------------------------------------------

510(k) Number (if known):

Device Name:

C5-2 Curvilinear Transducer

Intended Use:

Diagnostic Imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
ABMPWDCWDColorDopplerAmplitudeDopplerColorVelocityImagingCombined(Specify:Note 2)Other:HarmonicImaging
Opthalmic
FetalPPP
AbdominalPPP
Intraoperative (Note 1)PPP
Intraoperative NeurologicalPPP
PediatricPPP
Small Organ (Note 3)PPP
Neonatal CephalicPPP
Adult Cephalic
CardiacPPP
Transesophageal
Transrectal
Transvaginal
Transurethral
Intravascular
Peripheral VesselPPP
Laparoscopic
Musculo-skeletalConventionalPPP
Musculo-skeletal SuperficialPPP
Other

N=new indication, P = Previously Cleared, Blank: Not Claimed

Additional Comments:

Note 1: For example Cardiac Note 2: B Mode and PWD mode, or color Doppler and PW Mode Note 3: Breast, Testes, Thyroid, Penis

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{7}------------------------------------------------

510(k) Summary

Sponsor:Siemens Medical Solutions, Inc.Ultrasound Division5168 Campus DrivePlymouth Meeting, PA 19462
---------------------------------------------------------------------------------------------------------------------

Contact Person: Primary Contact: Kevin Kong, RAC Telephone: (650) 969-9112

Secondary Contact:

Lawrence Engle Telephone: (610) 834-1220 Ext 102

  • Submission Date: July 8, 2016
    ACUSON Freestyle™ Ultrasound System Device Name:

  • Common Name: Diagnostic Ultrasound System

  • Classification: Regulatory Class: ll

Review Category: Tier II Classification Panel: 90, Radiology

Ultrasonic Pulsed Doppler Imaging System

  • 21 CFR # 892.1550 .
  • . Product Code IYN

Ultrasonic Pulsed Echo Imaging System

  • . 21 CFR # 892.1560
  • Product Code IYO ●

Image /page/7/Picture/18 description: The image shows the word "SIEMENS" in all capital letters. The font is a sans-serif typeface, and the letters are a teal color. The background is white, which makes the text stand out.

{8}------------------------------------------------

Diagnostic Ultrasound Transducer

  • 21 CFR # 892.1570 .
  • . Product Code ITX

Legally Marketed Predicate Devices

The ACUSON Freestyle™ Ultrasound System in this 510k is a modification to Penrith Elettra Diagnostic Ultrasound System (Primary predicate), previously cleared in K100598. The secondary predicate device is the ACUSON S2000 as cleared on K152369.

Device Description:

The ACUSON Freestyle™ Ultrasound System is an ultrasound imaging system, which operates with linear, curvilinear, array transducers. The transducers may be used in either a wireless or wired configuration through a cable connected to the system.

Intended Use

The ACUSON Freestyle™ Ultrasound System is intended for diagnostic imaging or fluid flow analysis of the human body including: Fetal, Abdominal, Intraoperative, Intraoperative Neurological, Pediatric, Small Organ, Neonatal Cephalic, Cardiac, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal Superficial.

Image /page/8/Picture/11 description: The image shows the word "SIEMENS" in all capital letters. The text is a teal color. The font is sans-serif and appears to be bolded.

{9}------------------------------------------------

Substantial Equivalence

The submission device is a modification to Elettra Ultrasound System previously cleared in K100598 with regard use and technological characteristics and is also substantially equivalent to the ACUSON S2000 as cleared in K152369.

Freestyle™ Ultrasound System Substantial Equivalence Table

DescriptionACUSON S2000(Predicate K152369)Penrith Elettra(PredicateK100598)ACUSONFreestyle™ (Thissubmission)
Indications for Use(Clinical Applications)The ultrasound imaging systems are intended for the following applications: Fetal,Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac,Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, SuperficialMusculoskeletal, and Peripheral Vascular applications.The system also provides the ability to measure anatomical structures {fetal,abdominal, intraoperative, pediatric, small organ, neonatal cephalic, adult cephalic,cardiac, trans-esophageal, transrectal, transvaginal, peripheral vessel, musculo-skeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} andcalculation packages that provide information to the clinician that may be usedadjunctively with other medical data obtained by a physician for clinical diagnosispurposes.The Arterial Health Package (AHP) software provides the physician with thecapability to measure Intima Media Thickness and the option to referencenormative tables that have been validated and published in peer-reviewed studies.The information is intended to provide the physician with an easily understood toolfor communicating with patients regarding state of their cardiovascular system.This feature should be utilized according to the "ASE Consensus Statement; Useof Carotid Ultrasound to Identify Subclinical Vascular Disease and EvaluateCardiovascular Disease Risk: A Consensus Statement from the AmericanAssociation of Echocardiography; Carotid Intima-Media Thickness Task Force,DiagnosticImaging or fluidflow analysis ofthe human bodyas follows: Fetal,Abdominal,IntraoperativeIntraoperativeNeurological,Pediatric, SmallOrgan, NeonatalCephalic, Cardiac,Peripheral Vessel,Musculo-skeletal(Conventional),Musculo-skeletal(Superficial)Same as K100598

Image /page/9/Picture/6 description: The image contains the word "SIEMENS" in a bold, teal-colored font. The letters are all capitalized and evenly spaced. The word appears to be a logo or brand name.

{10}------------------------------------------------

DescriptionACUSON S2000(Predicate K152369)Penrith Elettra(PredicateK100598)ACUSONFreestyle™ (Thissubmission)
Endorsed by the Society for Vascular Imaging. The ACUSON Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as wellas visualization of other devices in the heart of adult and pediatric patients.
Product Code(s)System: IYO, IYNTransducer: ITXCatheter: OBJ-System: IYO, IYN-Transducers: ITX-System: IYO, IYN-Transducers: ITX
Transducer Types-Linear Array-Phased Array-Curved Array-Pencil-TEE-Linear Array-Curvilinear Array-Linear Array-Curvilinear Array
DescriptionACUSON S2000(Predicate K152369)Penrith Elettra(PredicateK100598)ACUSONFreestyle™ (Thissubmission)
Modes of Operation-B Mode-B Mode-B Mode
-M Mode-Pulsed Wave Doppler (PWD)-Continuous Wave Doppler (CWD)-Color Doppler-Amplitude Doppler-Combined (BMDC)-Color DopplerMode-AmplitudeDoppler Mode-Color DopplerMode-AmplitudeDoppler Mode
Functionality withSiemens ARTISSystemN/AN/A
Imaging
Multi-Hertz multiplefrequency imaging
Acoustic outputdisplay standardcompliance
Measurements(Distance, Area,Elliptical)
Trapezoidal (Wide)ImagingN/A
DescriptionACUSON S2000(Predicate K152369)Penrith Elettra(PredicateK100598)ACUSONFreestyle™ (Thissubmission)
WirelessTransducer
Wireless UltrasoundImage TransmissionN/AB, Color,AmplitudeB, Color,Amplitude
Meets FCC Part 15Subpart B:UnintentionalRadiatorsN/AYesYes
Meets FCC Part 15Subpart C: IntentionalRadiatorsN/AYesYes
Meets FCC Part 15Subpart F : UltraWideband OperationN/AYesYes
Meets FCC Part 95(WMTS)N/ANoNo
Frequency Range RFTransmitterN/A7.5-8.5 GHz7.5-8.5 GHz
Lithium Ion BatteryOperationN/AYesYes
Connectivity
Wireless NetworkConnectivityN/A
External AntennaN/AN/A
DICOM Compatibility

Image /page/10/Picture/3 description: The image shows the word "SIEMENS" in a teal color. The letters are in a bold, sans-serif font. The background is white, which makes the text stand out.

{11}------------------------------------------------

SIEMENS

{12}------------------------------------------------

Image /page/12/Picture/3 description: The image shows the word "SIEMENS" in a teal color. The letters are all capitalized and in a bold font. The word is centered horizontally and vertically in the image. The background is white.

{13}------------------------------------------------

Non-clinical Tests submitted, referenced, or relied on for determination of substantial equivalence

The devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfectiveness as well as thermal, electrical, electromagnetic and mechanical safety and to conform with applicable medical devices safety standards. The systems comply with the following standards:

  • · AAMI/ANSI 60601-1: Medical Electrical Equipment Part 1 General Requirements for Basic Safety and Essential Performance
  • · IEC 60601-1-2: Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance: Electromagnetic Compatibility
  • · IEC 60601-2-37: Medical Electrical Equipment Part 2-37 For Basic Safety and Essential Performance of Ultrasonic Medical Diagnostic Monitoring Equipment
  • IEC 62366: Medical Devices Application of Usability .
  • IEC 62304: Medical Device Software Software Life Cycle Process .
  • NEMA UD-2: Acoustic Output Measurement Standard for Diagnostic Ultrasound ●
  • NEMA UD-3: Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
  • · ISO 10993-1: Biological Evaluation of Medical Devices

Summary discussion of the clinical tests submitted, referenced, or relied on for determination of substantial equivalence

The ACUSON Freestyle™ Ultrasound System is a class II device, and uses the same technology and operating principle as the predicate devices; clinical data is not required for substantial equivalence.

Conclusion

As shown by the substantial equivalence table above, the ACUSON Freestyle™ is substantially equivalent as the predicate devices; Penrith Elettra Ultrasound System (K100598) and the ACUSON S2000 (K152369).

SIEMENS

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.