(81 days)
Puritan Fecal Opti-Swab Collection and Transport System is intended for use in the collection and transport of clinical fecal and rectal swab specimens to preserve the viability of enteric bacteria during transport from the collection site to the testing laboratory for bacteriological examination and culture.
Puritan Fecal Opti-Swab Collection and Transport System is comprised of a sterile peel pouch containing a sterile polyester flock swab applicator for collecting clinical fecal and rectal specimens and a polypropylene vial containing 2 ml of Fecal Opti-Swab medium.
Fecal Opti-Swab medium is a nonnutritive balanced salt solution containing phosphates to provide buffering capability, chloride salts to provide essential ions that help maintain osmotic balance. Agar is a solidifying agent that increases the viscosity of the medium. Sodium thioglycolate and L-cysteine provides an oxygen-reduced environment. It is recommended for maintaining the viability of enteric bacteria during the transport to the laboratory.
The document describes the Puritan Fecal Opti-Swab Collection and Transport System, a device for collecting and transporting fecal and rectal swab specimens to preserve the viability of enteric bacteria. The study presented is a substantial equivalence comparison with a predicate device, not a comparative effectiveness study with human readers or a standalone algorithm performance study.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly list "acceptance criteria" in a typical quantitative sense (e.g., target sensitivity/specificity). Instead, it states that the Puritan Fecal Opti-Swab Collection and Transport System was compared to a predicate device, and "No notable differences in recovery were detected" and that the performance "is not adversely affected by fecal material." It also reports on pH stability, cytotoxicity, and sterilization.
| Feature/Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Microbial Recovery (vs. Predicate Device) | No notable differences in recovery compared to predicate device. | "No notable differences in recovery were detected between the Puritan Fecal Opti-Swab Collection and Transport System and the predicate device." |
| Microbial Recovery (in Fecal Matrix) | Not adversely affected by fecal material. | "It was determined that clinical fecal material does not notably alter the recovery of the challenge organisms from the Puritan Fecal Opti-Swab Collection and Transport System." |
| pH Stability | pH of 6.90-7.50 up to expiry date. | "All samples tested were found to maintain pH within the specified range [6.90-7.50] for up to 15 months after manufacturing date." |
| Cytotoxicity | No evidence of cytotoxicity. | "No evidence of cytotoxicity was detected." |
| Sterilization | Sterilized according to ANSI/AAMI/ISO 11137:2006 guidelines. | "Puritan Fecal Opti-Swab Collection and Transport System is sterilized by gamma irradiation and validated following ANSI/AAMI/ISO 11137:2006, Sterilization of health care products Radiation guidelines." |
| Shelf Life | 15 months. | The device demonstrated pH stability and microbial recovery up to the expiry date (15 months). |
| Specimen Storage for C. difficile | Up to 48 hours at 2 – 8 °C and up to 24 hours at 20 – 25 °C. | Stated as the intended performance characteristic for both test and predicate device. Testing confirmed viability to meet this (implied by recovery testing). |
| Specimen Storage for Other Target Organisms | Up to 72 hours at 2 – 8 °C and up to 48 hours at 20 – 25 °C. | Stated as the intended performance characteristic for both test and predicate device. Testing confirmed viability to meet this (implied by recovery testing). |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document does not explicitly state the specific number of samples (e.g., number of swabs, fecal samples, or bacterial cultures) used for each recovery and stability test. It mentions "known inoculum of ATCC type culture and clinically significant microorganisms" and "three representative challenge organisms," but not specific quantities.
- Data Provenance: The study appears to be prospective in nature, as it involves preparing inocula and conducting experiments in a controlled lab setting to compare the Puritan device with a predicate. The specific country of origin for the data is not mentioned, but the sponsor is Puritan Medical Products LLC from Guilford, ME, USA.
The study compares the device to the predicate following CLSI M40-A2 guidelines, which are standard for microbiological transport systems.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- This information is not provided as the study focuses on the performance of a transport medium for microbial viability, not on diagnostic interpretation by human experts. The "ground truth" here is the known quantity and viability of bacterial cultures established through standard microbiology laboratory techniques.
4. Adjudication Method for the Test Set:
- An adjudication method is typically used when human interpretation is involved and discrepancies arise. As this study is a laboratory-based evaluation of microbial viability in a transport medium, an adjudication method for a test set is not applicable and therefore not described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, an MRMC comparative effectiveness study was not done. This type of study assesses the effect of a device (like AI software) on human reader performance. The Puritan Fecal Opti-Swab is a physical collection and transport system, not an AI-powered diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is not relevant to this device or study.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, a standalone algorithm performance study was not done. This device is a manual collection and transport system, not an algorithm or software.
7. The Type of Ground Truth Used:
- The ground truth used was based on known bacterial ATCC type cultures and clinically significant microorganisms with known concentrations (inoculum). This is a laboratory-established ground truth, typically validated by quantitative plating methods to determine viable colony-forming units (CFUs). For the fecal matrix experiment, "clinical fecal material" was used, and presumably, spiked with known organisms to assess recovery.
8. The Sample Size for the Training Set:
- This device is a physical product (collection and transport system) and not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The stability and recovery tests are analogous to validation or performance testing for a traditional device.
9. How the Ground Truth for the Training Set Was Established:
- As there is no training set for an algorithm, this question is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 4, 2016
PURITAN MEDICAL PRODUCTS LLC MEHDI KARAMCHI VICE PRESIDENT OF SCIENTIFIC AFFAIRS 31 SCHOOL STREET P.O. BOX 149 GUILFORD ME 04443
Re: K162284
Trade/Device Name: Puritan Fecal Opti-swab Collection And Transport System Regulation Number: 21 CFR 866,2390 Regulation Name: Transport culture medium Regulatory Class: I Product Code: JSM, LIO, JTW, JTX Dated: August 11, 2016 Received: August 15, 2016
Dear Mr. Karamchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ribhi Shawar -S
For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162284
Device Name
Puritan Fecal Opti-Swab Collection and Transport System
Indications for Use (Describe)
Puritan Fecal Opti-Swab Collection and Transport System is intended for use in the collection and transport of clinical fecal and rectal swab specimens to preserve the viability of enteric bacteria during transport from the collection site to the testing laboratory for bacteriological examination and culture.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Puritan® Fecal Opti-Swab™ Collection and Transport System
5.1 Sponsor
| 5.1 | Sponsor | |
|---|---|---|
| Puritan Medical Products LLC | ||
| 31 School St., Guilford, ME | ||
| 04443 | ||
| Contact: | Mehdi Karamchi | |
| Telephone Number: | 207-876-3311 | |
| Fax Number: | 207-876-2680 | |
| Date: | August 11, 2016 | |
| 5.2 | Device Name | |
| Classification Name: | Transport Culture Medium | |
| Common Name: | Specimen Collection and Transport System | |
| Proprietary Name: | Puritan® Fecal Opti-Swab™ Collection and Transport System | |
| 5.3 | Regulatory Information | |
| A. Regulatory Section: | 21 CFR 866.2390 | |
| B. Classification: | Class I | |
| C. Product Code: | LIO, JSM; JTW; JTX | |
| D. Panel: | Microbiology (83) |
5.4 Predicate Device
Copan Fecal Swab Collection, Transport and Preservation System manufactured by Copan Diagnostics Inc. of Italy
510K Number: K142094
{4}------------------------------------------------
5.5 Device Description
Puritan Fecal Opti-Swab Collection and Transport System is comprised of a sterile peel pouch containing a sterile polyester flock swab applicator for collecting clinical fecal and rectal specimens and a polypropylene vial containing 2 ml of Fecal Opti-Swab medium.
Fecal Opti-Swab medium is a nonnutritive balanced salt solution containing phosphates to provide buffering capability, chloride salts to provide essential ions that help maintain osmotic balance. Agar is a solidifying agent that increases the viscosity of the medium. Sodium thioglycolate and L-cysteine provides an oxygen-reduced environment. It is recommended for maintaining the viability of enteric bacteria during the transport to the laboratory.
5.6 Intended Use
Puritan Fecal Opti-Swab Collection and Transport System is intended for use in the collection and transport of clinical fecal and rectal swab specimens to preserve the viability of enteric bacteria during transport from the collection site to the testing laboratory for bacteriological examination and culture.
5.7 Indication(s) For Use
Puritan Fecal Opti-Swab Collection and Transport System is intended for use in the collection and transport of clinical fecal and rectal swab specimens to preserve the viability of enteric bacteria during transport from the collection site to the testing laboratory for bacteriological examination and culture.
5.8. Substantial Equivalence statement
Puritan Fecal Opti-Swab Collection and Transport System is similar in design, manufacturing and intended usage to the predicate device. Both the Puritan and the predicate device are single-use devices intended for collection and transport of clinical fecal and rectal swab specimens to preserve the viability of enteric bacteria during transport from the collection site to the testing laboratory for bacteriological examination and culture.
{5}------------------------------------------------
| Puritan and Predicate Device Comparison | ||
|---|---|---|
| Item | Test Device | Predicate |
| Intended USE | Puritan Fecal Opti-SwabCollection and Transport Systemis a sterile ready-to- use systemintended for collection andtransport of clinical fecal andrectal swab specimens topreserve the viability of entericbacteria during transport fromthe collection site to the testinglaboratory for bacteriologicalexamination and culture. | The Copan FecalSwabCollection, Transport andPreservation System isintended for the collectionof rectal swab and fecalspecimens and to preservethe viability of entericpathogenic bacteria duringtransport from thecollection site to the testinglaboratory. In thelaboratory, FecalSwabspecimens are processedusing standard clinicallaboratory operatingprocedures for culture. |
| Single-use Device | Yes | Same |
| Medium Formulation | Chloride saltsSodium saltsPhosphate bufferL-CysteineAgarWater | Chloride saltsSodium saltsPhosphate bufferL-CysteineAgarWater (differentformulation) |
| pH | 6.90-7.50 | Same |
| Storage Temperature | 2-25°C (refrigerated and roomtemperature) | 5-25°C (refrigerated androom temperature) |
| Container | Polypropylene conical bottomvial | Same |
| Product Configuration | Medium in vial & capSystem including Medium and | Same |
| swab in peel pouch option. | ||
| Swab Shaft | Plastic | Same |
| Swab Tip | Flock Swab | Same |
| Shelf Life | 15 months | Same |
| Specimen Storage | For C. difficile, up to 48 hours at2 – 8 °C and up to 24 hours at 20– 25 °C. For other targetorganisms organisms: up to 72hours at 2 – 8 °C and up to 48hours at 20 – 25 °C. | Same |
| Medium Volume | 2 mL | Same |
| Organisms Tested | Campylobacter jejuniClostridium difficileEnterococcus faecalisEscherichia coliSalmonella typhimuriumShigella sonneiVibrio parahaemolyticusYersinia enterocolitica | Same |
{6}------------------------------------------------
5.9 Recovery Testing
To determine the ability of the Puritan Fecal Opti-Swab Collection and Transport System to maintain viability of different strains of enteric bacteria, known inoculum of ATCC type culture and clinically significant microorganisms were prepared in a phosphate-buffered saline matrix and inoculated into the Puritan Fecal Opti-Swab Collection and Transport medium and compared to the predicate device following Clinical and Laboratory Standards Institute (CLSI) M40-A2 guidelines. No notable differences in recovery were detected between the Puritan Fecal Opti-Swab Collection and Transport System and the predicate device.
Recovery was also evaluated using three representative challenge organisms prepared in a fecal matrix to demonstrate that the performance of the Puritan Fecal Opti-Swab Collection and Transport medium is not adversely affected by fecal material. It was determined that clinical fecal material does not notably alter the recovery of the challenge organisms from the Puritan Fecal Opti-Swab Collection and Transport System.
{7}------------------------------------------------
5.10 Stability Testing
Stability tests were performed on Puritan Fecal Opti-Swab Collection and Transport System to verify the ability of the aged products to maintain microbial recovery up to the expiry date.
5.11 pH Stability
The pH of the test device was measured at predetermined time intervals for up to 15 months after manufacturing date. The test was performed using calibrated pH meter with random samples from three different lots of Puritan Fecal Opti-Swab Collection and Transport System. All samples tested were found to maintain pH within the specified range.
Cytotoxicity 5.12
Cytotoxicity testing was conducted to evaluate the glue, shaft and the flock tipped swabs for potential cytotoxicity effect following ISO Elution Method-1X MEM Extract. No evidence of cytotoxicity was detected.
5.13 Sterilization
Puritan Fecal Opti-Swab Collection and Transport System is sterilized by gamma irradiation and validated following ANSI/AAMI/ISO 11137:2006, Sterilization of health care products Radiation guidelines.
5.14 Conclusion
The conclusions drawn from the nonclinical tests demonstrate the device is as safe and effective as the legally marketed device identified above.
§ 866.2390 Transport culture medium.
(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).