K Number
K162161

Validate with FDA (Live)

Date Cleared
2016-12-16

(136 days)

Product Code
Regulation Number
882.5880
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

Device Description

This submission is based upon the established Stimwave product family offering additional surgical kitting options (Spare Lead Kit and Sterile Revision Kit), updates to the surgical kits (Receiver Kit and Trial Lead Kit) to include new accessory components (Needle, and RF Stylet), a new Wearable Antenna Assembly (WAA) model utilizing a Low Energy (LE) Bluetooth module, a USB battery charger, and upgrades to the WaveCrest Application.

The Stimwave Technologies Incorporated (Stimwave) Freedom Spinal Cord SCS System (System) is used for spinal column stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or lower limbs including unilateral pain. The therapy utilizes pulsed electrical current to create an energy field that acts on nerves near the spinal column. The System is comprised of an implantable stimulator (Freedom-8A/4A Stimulator), receiver component (Receiver), and an externally worn transmitter (Wearable Antenna Assembly (WAA)) to power the device. The System is implanted only following a successful trial period with the Freedom-8A/4A Trial Lead.

AI/ML Overview

The Stimwave Freedom Spinal Cord Stimulator (SCS) System (K162161) sought clearance by demonstrating substantial equivalence to its predicate devices (K160600, K150517, and K141399). Therefore, the acceptance criteria and performance are primarily based on non-clinical testing verifying its components meet design requirements and applicable voluntary standards. There was no clinical testing required or performed for this submission as the device's indications for use and underlying technology were considered equivalent to the already legally marketed predicate devices with established safety and efficacy.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding performance statistics in a pass/fail format typical for classification. Instead, it describes various performance tests and states that the device "complies" or "passed" the criteria of the respective standards. The acceptance criteria essentially align with the requirements of the listed voluntary standards and the device's design specifications for safety and functionality.

Acceptance Criteria CategoryStandard / Test TypeReported Device Performance
Material BiocompatibilityISO 10993-1:2009, Blue Book Memorandum G95-1- No negative impacts from materials for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity. - Materials (Pellethane 55D, Pt-Ir) have extensive record of chronic and carcinogenetic safety. - WAA is "noncontacting device" and outside the scope of ISO 10993-1:2009 for body contact. - Conclusion: Meets biological safety and compatibility requirements.
Protection from Temperature ChangeAAMI ANSI ISO 14708-3:2008- Freedom-8A/4A Stimulator: Functional, safe rating, passed post visual inspection, and passed temperature change testing. - WAA: Met passing criteria of visual and functional inspections, no physical damage, fully operational.
Atmospheric Pressure ChangeAAMI ANSI ISO 14708-3:2008- Freedom-8A/4A Stimulator: Functional following post testing functionality inspection, passed atmospheric pressure change testing. - WAA: Met passing criteria of visual and functional inspections, no physical damage, fully operational.
External Defibrillation ExposureAAMI ANSI ISO 14708-3:2008- Freedom-8A/4A Stimulator and Receiver: Verified as functional after exposure. - Conclusion: Complies with testing requirements.
Thermal ShockApplicable Standard (not explicitly named, implies AAMI ANSI ISO 14708-3:2008 context)- Freedom-8A/4A Stimulator: "No irreversible damage" and fully functional, no physical anomalies present. - Conclusion: Complies with thermal shock design requirements and applicable standard.
Leakage CurrentApplicable Standard (not explicitly named, implies AAMI ANSI ISO 14708-3:2008 context)- Freedom-8A/4A Stimulator: Produced zero leakage current on all tested paths for all tested samples. - Conclusion: Complies with leakage design requirements and applicable standard.
Stylet Insertion/Withdrawal ForceDesign Specifications- Stylet within Stimulator: Required less than 2.5N insertion/withdrawal force. - Receiver within Stimulator: Required less than 2.2N insertion/withdrawal force. - Visual inspection confirmed no damage to stimulator samples. - Conclusion: Complies with design specifications.
Mechanical TestingDesign Requirements- Freedom-8A/4A Stimulator: Passed all criteria of tensile and torsion testing, showing no visible damage to stimulator body or functional damage to components. - Conclusion: Complies with all stimulator mechanical design requirements.
WAA Robustness (Push, Drop, Impact, Mold Stress Relief)IEC 60601-1- Determined to be robust to withstand expected damage. Met passing criteria of visual and functional inspections, showed no physical damage, fully operational. - Conclusion: Satisifes design requirements and applicable standard.
WAA Identification, Marking, DocumentsIEC 60601-1- Analysis of labeling determined compliance with requirements. All requirements and markings clearly identified and viewable. - Conclusion: Complies with the requirements of the standard.
WAA Means of Protection, Creepage Distances, Air ClearancesIEC 60601-1- Analysis of design determined system satisfies requirements. - Conclusion: Satisfies requirements of the applicable standard.
Ingress of WaterIEC 60529- WAA: Met passing criteria of visual and functional inspections, showed no physical damage, fully operational. - Conclusion: Satisfies design requirements and applicable standard.
Particulate MatterIEC 60529- WAA: Met passing criteria of visual and functional inspections, showed no physical damage, fully operational. - Conclusion: Satisfies design requirements and applicable standard.
Electromagnetic Compatibility (EMC)IEC 60601-1-2- WAA: Met all acceptance criteria for emissions, low-frequency magnetic fields, immunity, electrostatic discharge, radiated RF electromagnetic fields, electrical fast transients and magnetic fields. Operated within all test limits, showed no physical damage, and fully operational. - Conclusion: Satisfies the standard.
Software VerificationDesign Requirements- Passed all verification tests outlined. - Design requirements for Software Verification met.
Overall ComplianceNational and International Standards, Design Inputs/Specifications- Complies with applicable standards for neurostimulators, electrical safety, electromagnetic interference and compatibility, packaging, and sterilization. - Output meets design inputs and specifications. - Passed all testing. - Conclusion: Substantially equivalent to legally marketed predicate devices, and differences do not raise questions of safety and effectiveness.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily refers to "tested samples" for hardware components and "analysis of labeling" or "analysis of design" for other aspects. Specific sample sizes for each test are not provided. The data provenance is non-clinical, meaning it's from laboratory and engineering testing, not patient data. No country of origin is specified for the testing data itself, but the submission is from a US company (Fort Lauderdale, Florida). It is by nature retrospective since it refers to testing already completed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the submission focuses on non-clinical performance data and engineering verification. The "ground truth" here is adherence to specified voluntary standards (e.g., AAMI ANSI ISO 14708-3:2008, IEC 60601-1, IEC 60529, IEC 60601-1-2) and internal design requirements, rather than expert clinical consensus or patient-derived ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept (adjudication for ground truth establishment) is relevant for studies involving human interpretation or clinical data analysis where disagreements need resolution. The tests described are objective engineering and laboratory tests against predefined criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case comparative effectiveness study was not done. This device is a spinal cord stimulator, not an AI-assisted diagnostic or imaging system. The submission explicitly states: "There was no clinical testing required to support the medical device, as the indications for use are equivalent to the legally marketed predicate devices." and "There was no clinical testing performed on this device since performance testing demonstrated similar performance as the legally marketed predicate devices."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a medical implant system; it does not involve an algorithm performing a standalone diagnostic or interpretative task. The "software" mentioned (WaveCrest Application) is for controlling the stimulator parameters, not for standalone diagnostic performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in this context is the predefined requirements from voluntary international standards (e.g., ISO 10993 for biocompatibility, IEC 60601 series for electrical safety, AAMI ANSI ISO 14708-3 for active implantable medical devices) and the device's internal engineering design specifications. The compliance with these standards and specifications serves as the basis for demonstrating safety and effectiveness for substantial equivalence.

8. The sample size for the training set

Not applicable. This is a medical device, not a machine learning or AI model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As there is no AI/ML component requiring a training set, the concept of establishing ground truth for a training set does not apply.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, with three human profiles forming the staff and a cloth draped around the base.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 16, 2016

Stimwave Technologies Incorporated Elizabeth Greene Vice President of Quality Assurance and Regulatory Affairs 901 East Las Olas Boulevard, Suite 201 Fort Lauderdale, Florida 33301

Re: K162161

Trade/Device Name: Freedom Spinal Cord Stimulation (SCS) System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZB Dated: November 16, 2016 Received: November 18, 2016

Dear Elizabeth Greene:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,
William J. Heetderks -A
Digitally signed by William J. Heetderks -ADN: c=US, o=U.S. Government, ou=HHS, ou=NIH,ou=People,0.9.2342.19200300.100.1.1=0010149848,cn=William J. Heetderks -ADate: 2016.12.16 14:42:53 -05'00'

for

Carlos L. Peña, PhD, MS
Director
Division of Neurologicaland Physical Medicine Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K162161

Device Name: Freedom Spinal Cord Stimulator (SCS) System

Indications For Use:

The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the words "stim wave" is the word "TECHNOLOGIES" in dark blue. To the right of the word "wave" are three curved lines in dark blue that resemble radio waves.

510(k) Summary

for

Freedom Spinal Cord Stimulator (SCS) System

1. Submission Sponsor

Stimwave Technologies Incorporated
901 East Las Olas Boulevard
Suite 201
Fort Lauderdale
Florida 33301
USA
Phone:800.965.5134
Fax:800.965.5134
Contact:Elizabeth Greene, Vice President of Quality Assurance and Regulatory Affairs

2. Date Prepared

September 7, 2016

3. Device Identification

Trade/Proprietary Name:Freedom Spinal Cord Stimulator (SCS) System
Common/Usual Name:Spinal Cord Stimulator
Classification Name:Stimulator, Spinal-Cord, Implanted (Pain Relief)
Classification Regulation:882.5880
Product Code:GZB
Device Class:Class II
Classification Panel:Neurology

4. Legally Marketed Predicate Device(s)

Stimwave Freedom SCS System FR8A/FR4A, LBRD-915-2A (K160600) Stimwave Freedom SCS System FR8A/FR4A, LBRD-915-2A (K150517) Stimwave Freedom SCS System FRE4-A001, FRT4-A001, WAA-A012 (K141399)

5. Device Description

This submission is based upon the established Stimwave product family offering additional surgical kitting options (Spare Lead Kit and Sterile Revision Kit), updates to the surgical kits (Receiver Kit and Trial Lead Kit) to include new accessory components (Needle, and RF Stylet), a new Wearable Antenna Assembly (WAA) model utilizing a Low Energy (LE) Bluetooth module, a USB battery charger, and upgrades to the WaveCrest Application.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in large, blue, lowercase letters. The word "wave" is in large, green, lowercase letters. Below the words is the word "TECHNOLOGIES" in smaller, blue, uppercase letters inside of a blue oval. To the right of the words is a blue graphic of concentric arcs.

The Stimwave Technologies Incorporated (Stimwave) Freedom Spinal Cord SCS System (System) is used for spinal column stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or lower limbs including unilateral pain. The therapy utilizes pulsed electrical current to create an energy field that acts on nerves near the spinal column. The System is comprised of an implantable stimulator (Freedom-8A/4A Stimulator), receiver component (Receiver), and an externally worn transmitter (Wearable Antenna Assembly (WAA)) to power the device. The System is implanted only following a successful trial period with the Freedom-8A/4A Trial Lead.

A polyurethane (Pellethane 55D) casing with an embedded receiver, Freedom-8A Stimulator. flexible circuit board and electrodes (Platinum Iridium 90:10) that is Freedom-4A placed in the patient's evidural space. The Freedom-8A Stimulator Stimulator has eight (8) electrodes, and the Freedom-4A Stimulator has four (4) electrodes. Identical to K150517 and K160600. Receiver A copper and PEEK cable with dual couplers; placed within the center lumen of the Freedom-8A or Freedom-4A Stimulator with the distal end combination of Receiver and Stimulator being placed under the skin. Two (2) Receivers are provided with each kit. Identical to K160600, but quantity is increased from one to two. A stainless steel wire with a polypropylene handle that is inserted Stylet(s) into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Receiver Kit, one straight and one bent, each with diameter of 0.30 mm. Similar to K150517 and K160600, but with a larger diameter. Needle A 13-gauge stainless steel needle that acts as a conduit for passage of the Stimulator into the epidural space. Similar to K141399, but provided with a larger gauge. Guidewire A stainless steel, rigid, solid core guidewire used to create a hollow pathway in the evidural space for the Stimulator to pass through easily. Identical to K141399, K150517, and K160600. Suture Sleeve Cap A polyurethane (Pellethane 55D) cap that is placed over the proximal end of the Stimulator. The Suture Sleeve Cap is attached to the Freedom-8A/4A Stimulator and can be sutured to tissue to reduce the possibility of device migration. Identical to K141399, K150517, and K160600.

Freedom-8A and Freedom-4A Stimulator (Receiver Kit)

Wearable Antenna Assembly (WAA Kit)

WAAThe WAA housing includes the following components:
A. Microwave Field Stimulator (MFS) – A printed circuit board (PCB) that generates 915 MHz RF power with embedded waveform parameter settings and switches for changing parameter settings as needed by the user. Similar to K150517 with design updates for USB charging and PCB layout;
Wearable Antenna Assembly (WAA Kit)
B. Switch Membrane – A elastomeric silicon rubber pad that
corresponds to switches on the MFS that allows the user to turn
the device on/off or increase or decrease power amplitude as well
as interpret device power status (On, Off, Charging, Transmitting,
and Bluetooth® Connection). Identical to the component used for
K152178 (StimQ PNS System);
C. Battery Assembly – A battery and wire assembly for charging and
the MFS for power delivery. New component for WAA.
Transmitting (Tx) Antenna Assembly – An antenna and coaxial cable
assembly that is attached to the WAA that is used to transmit
microwave energy to the implanted Stimulator. Similar design and
functionality as K152178, with design updates for indication.
Charger Kit
Battery ChargerAn off-the-shelf battery charger that uses a power adapter and USB to
micro-USB cable to recharge the encased lithium ion battery of the
WAA. New accessory component used to recharge the WAA.
Freedom-8A and Freedom-4A Spare Lead (Spare Lead Kit)
Freedom-8ASpare Lead,Freedom-4ASpare LeadA polyurethane (Pellethane 55D) casing with an embedded receiver,flexible circuit board and electrodes (Platinum Iridium 90:10) that isplaced in the patient's epidural space. The Freedom-8A Spare Leadhas eight (8) electrodes, and the Freedom-4A Spare Lead has four (4)electrodes. Identical in design and functionality to the Freedom-8Aand Freedom-4A Stimulator.
RF StyletA copper and PEEK cable with dual couplers; placed within thecenter lumen of the Freedom-8A or Freedom-4A Spare Lead with thedistal end combination of RF Stylet and Spare Lead being placedunder the skin. Two (2) RF Stylets are provided with each kit.Identical in design and functionality to the Receiver.
Stylet(s)A stainless steel wire with a polypropylene handle that is insertedinto the open central lumen of the stimulator to provide rigidityduring implantation. Two (2) stylets are provided in the Spare LeadKit, one straight and one bent, each with diameter of 0.30 mm.Identical in design and functionality to the stylets in the Receiver Kit.
NeedleA 13-gauge stainless steel needle that acts as a conduit for passage ofthe Spare Lead into the epidural space. Identical in design andfunctionality to the needle in the Receiver Kit.
GuidewireA stainless steel, rigid, solid core guidewire used to create a hollowpathway in the epidural space for the Spare Lead to pass througheasily. Identical in design and functionality to the guidewire in theReceiver Kit.
Freedom-8ATrial Lead,Freedom-4ATrial LeadA polyurethane (Pellethane 55D) casing with an embedded receiver, flexible circuit board and electrodes (Platinum Iridium 90:10) that is placed percutaneously in the patient's epidural space. The Freedom-8A Trial Lead has eight (8) electrodes, and the Freedom-4A Trial Lead has four (4) electrodes. Identical in design and functionality to the Freedom-8A and Freedom-4A Stimulator.
RF StyletA copper and PEEK cable with dual couplers; placed within the center lumen of the Freedom-8A or Freedom-4A Trial Lead with the distal end combination of RF Stylet and Trial Lead being placed under the skin. Two (2) RF Stylets are provided with each kit. Identical in design and functionality to the Receiver.
Stylet(s)A stainless steel wire with a polypropylene handle that is inserted into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Trial Lead Kit, one straight and one bent with a diameter of 0.30 mm. Identical in design and functionality to the stylets in the Receiver Kit.
NeedleA 13-gauge stainless steel needle that acts as a conduit for passage of the Trial Lead into the epidural space. Identical in design and functionality to the needle in the Receiver Kit.
GuidewireA stainless steel, rigid, solid core guidewire used to create a hollow pathway in the epidural space for the Trial Lead to pass through more easily. Identical in design and functionality to the guidewire in the Receiver Kit.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a large, purple, sans-serif font. Below and to the right of "stim" is the word "wave" in a green, sans-serif font. Below "stim wave" is the word "TECHNOLOGIES" in a smaller, blue, sans-serif font inside of a rounded rectangle. To the right of the words is a graphic of three concentric arcs in a blue-purple color.

Wearable Antenna Assembly (WAA Kit)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a blue sans-serif font, and the word "wave" is in a green sans-serif font. Below the words "stim wave" is the word "TECHNOLOGIES" in a smaller blue sans-serif font with a blue line above and below the word. To the right of the words is a graphic of blue concentric semi-circles.

Suture Sleeve Cap A polyurethane (Pellethane 55D) cap that is placed over the proximal end of the Stimulator. The Suture Sleeve Cap is attached to the Freedom-8A/4A Spare Lead and can be sutured to tissue to reduce the possibility of device migration. Identical in design and functionality to the Suture Sleeve Cap in the Receiver Kit.

Freedom-8A and Freedom-4A Trial Lead (Trial Lead Kit)

Sterile Revision Kit

Stylet(s)A stainless steel wire with a polypropylene handle that is inserted into the open central lumen of the stimulator to provide rigidity during implantation. Two (2) stylets are provided in the Sterile Revision Kit, one straight and one bent with a diameter of 0.30 mm. Identical in design and functionality to the stylets in the Receiver Kit.
NeedleA 13-gauge stainless steel needle that acts as a conduit for passage of the Stimulator/Lead into the epidural space. Identical in design and functionality to the needle in the Receiver Kit.
GuidewireA stainless steel, rigid, solid core guidewire used to create a hollow pathway in the epidural space for the Stimulator/Lead to pass through easily. Identical in design and functionality to the guidewire in the Receiver Kit.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below the words is the word "TECHNOLOGIES" in blue. To the right of the words are three curved lines that look like sound waves in blue.

6. Indication for Use Statement

The Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The Freedom-8A Trial Lead Kit is only used in conjunction with the Freedom-8A Stimulator Receiver Kit, and the Freedom-4A Trial Lead Kit is used for either the Receiver Kit Freedom-4A Stimulator or the Receiver Kit Freedom-8A Stimulator. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

7. Substantial Equivalence Discussion

The following table compares the Stimwave Freedom SCS System to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

ComparatorStimwaveFreedom SCSSystem(K162161)StimwaveFreedom SCSSystem(K160600)StimwaveFreedom SCSSystem(K150517)StimwaveFreedom SCSSystem(K141399)
Product CodeGZBSame as K162161Same as K162161Same as K162161
Regulation No.882.5880Same as K162161Same as K162161Same as K162161
Regulation NameStimulator, Spinal-Cord,Implanted (Pain Relief)Same as K162161Same as K162161Same as K162161
Intended UseStimulation of spinal cord forchronic, intractable pain oftrunk and lower limbsSame as K162161Same as K162161Same as K162161
Mode of ActionRF wireless transmission ofenergy to producestimulation at Stimulatorelectrodes. WAA sends apulsed RF signal on a carrierfrequency of 915MHz to theStimulatorSame as K162161Same as K162161Same as K162161
Implant SiteEpidural space, L5 to T5Same as K162161Same as K162161Same as K162161
EnvironmentalUseHospital, HomeSame as K162161Same as K162161Same as K162161
Intended ClinicianOrthopedic, Neurosurgeon,AnesthesiologistSame as K162161Same as K162161Same as K162161
Intended UserLaypersonSame as K162161Same as K162161Same as K162161
ElectrodeMaterialPlatinum-iridium 90:10Same as K162161Same as K162161Same as K162161
Stimulator BodyMaterialPolyurethane 2363-55DSame as K162161Same as K162161Same as K162161
Cable FeaturesMulti-lumen TubeSame as K162161Same as K162161Same as K162161
Stimulator Length45 centimetersSame as K162161Same as K162161Same as K162161
Diameter1.35 millimetersSame as K162161Same as K162161Same as K162161
Electrode ArrayLength24.0 millimeters52.0 millimetersSame as K162161Same as K162161Same as K162161
ComparatorStimwaveFreedom SCSSystem(K162161)StimwaveFreedom SCSSystem(K160600)StimwaveFreedom SCSSystem(K150517)StimwaveFreedom SCSSystem(K141399)
No. of Electrodes4 or 8Same as K162161Same as K1621614
Electrode Length3.0 millimetersSame as K162161Same as K162161Same as K162161
Electrode Spacing4.0 millimetersSame as FreedomSame as K162161Same as K162161
Electrode Surface Area12.72 mm²Same as K162161Same as K162161Same as K162161
Method of IntroductionPercutaneous and AnchorIncisionSame as K162161Same as K162161Same as K162161
Tissue ContactYesSame as K162161Same as K162161Same as K162161
SterilizationEthylene Oxide (EO)Same as K162161Same as K162161Same as K162161
LabelingLabeled as Sterile, SingleUse, Prescription DeviceSame as K162161Same as K162161Same as K162161
PackageBacker card and two sterilepouchesSame as K162161Same as K162161BlisterTray/Tyvek Lid
Pulse Frequency5 to 1500 HertzSame as K162161Same as K1621612 to 1500 Hertz
Pulse Width50 to 500 microsecondsSame as K162161Same as K162161Same as K162161
Current/Voltage RegulatedCurrentSame as K162161Same as K162161Same as K162161
Output Voltage (300 Ω)0 to 4.1 VSame as K162161Same as K1621610 to 6.3 V
Output Voltage (500 Ω)0 to 6.4 VSame as K162161Same as K1621610 to 7.2 V
Output Voltage (800 Ω)0 to 7.5 VSame as K162161Same as K1621610 to 8.0 V
Output Current (300 Ω)0 to 13.5 mASame as K162161Same as K1621610 to 21 mA
Output Current (500 Ω)0 to 12.8 mASame as K162161Same as K1621610 to 15 mA
Output Current (800 Ω)0 to 9.4 mASame as K162161Same as K1621610 to 10 mA
WaveformCharge Balanced (delayed)Biphasic asymmetricalSame as K162161Same as K162161Same as K162161
PolarityProgrammable(Anode, Cathode, or Off)Same as K162161Same as K162161Fixed
Pulse ShapeDecaying ExponentialSame as K162161Same as K162161Same as K162161
Avg. Current Density (300 Ω)105.0 mA/cm²Same as K162161Same as K162161111.6 mA/cm²
Avg. Current Density (500 Ω)95.1 mA/cm²Same as K162161Same as K16216196.7 mA/cm²
Avg. Current Density (800 Ω)69.0 mA/cm²Same as K162161Same as K16216177.0 mA/cm²
Max. Phase Charge* (300 Ω)6.8 µC/pulseSame as K162161Same as K16216110.5 µC/pulse
Max. Phase Charge* (500 Ω)6.4 µC/pulseSame as K162161Same as K1621617.2 µC/pulse
Max. Phase Charge* (800 Ω)4.7 µC/pulseSame as K162161Same as K1621615.0 µC/pulse
Max. Charge Density* (300 Ω)53.1 µC/cm²Same as K162161Same as K16216182.5 µC/cm²
Max. Charge Density* (500 Ω)50.3 µC/cm²Same as K162161Same as K16216156.6 µC/cm²
Max. Charge Density* (800 Ω)36.9 µC/cm²Same as K162161Same as K16216139.3 µC/cm²
ComparatorStimwaveFreedom SCSSystem(K162161)StimwaveFreedom SCSSystem(K160600)StimwaveFreedom SCSSystem(K150517)StimwaveFreedom SCSSystem(K141399)
Max. CurrentDensity* (300 Ω)106.1 mA/cm²Same as K162161Same as K162161165.1 mA/cm²
Max. CurrentDensity* (500 Ω)100.6 mA/cm²Same as K162161Same as K162161113.2 mA/cm²
Max. CurrentDensity* (800 Ω)73.9 mA/cm²Same as K162161Same as K16216178.6 mA/cm²
Net Charge0 μCSame as K162161Same as K162161Same as K162161
Avg. Phase Power(300 Ω)0.053 W/phaseSame as K162161Same as K1621610.060 W/phase
Avg. Phase Power(500 Ω)0.073 W/phaseSame as K162161Same as K1621610.076 W/phase
Avg. Phase Power(800 Ω)0.062 W/phaseSame as K162161Same as K1621610.060 W/phase
Avg. Phase PowerDensity (300 Ω)0.42 W/cm²/phaseSame as K162161Same as K1621610.48 W/cm²/phase
Avg. Phase PowerDensity (500 Ω)0.58 W/cm²/phaseSame as K162161Same as K1621610.59 W/cm²/phase
Avg. Phase PowerDensity (800 Ω)0.48 W/cm²/phaseSame as K162161Same as K1621610.60 W/cm²/phase
Pulse DeliveryModeContinuousSame as K162161Same as K162161Same as K162161
ON/OFF TimesNo CyclingSame as K162161Same as K162161Same as K162161
Current PathOptionsBipolarSame as K162161Same as K162161Same as K162161
Power DeliveryEmbedded receiver andcoupled receiver in lumen ofStimulator bodySame as K162161Coupled receiver,built intoStimulator bodyCoupled receiver,built intoStimulator body
TransmitFrequency915 MHzSame as K162161Same as K162161Same as K162161
MaterialPlatinum-iridium 90:10,Polyurethane 2363-55DSame as K162161Same as K162161Same as K162161
SterileYes - ethylene oxideSame as K162161Same as K162161Same as K162161
ContractSterilizerSteris Isomedix ServicesLife ScienceOutsourcingLife ScienceOutsourcingOscor, Inc.
Single-UseYesSame as K162161Same as K162161Same as K162161
Shelf Life2 yearSame as K162161Same as K1621611 year
Complies withISO 10993-1YesSame as K162161Same as K162161Same as K162161
Safety TestingPassedYesSame as K162161Same as K162161Same as K162161
MR ConditionalNo, MR UnsafeSame as K162161Same as K162161Yes
AccessoriesReceiver/RF Stylet,Stylet(s), Guidewire,Needle, Suture Sleeve CapReceiver,Stylet(s),Guidewire,Needle, SutureSleeve CapStylet(s),Guidewire,Needle, SutureSleeve CapStylet(s),Guidewire,Needle, SutureSleeve Cap
ChargerUSB ChargerWireless ChargerWireless ChargerWireless Charger
WearableAntennaAssemblyAluminum transmitter andseparate, connected AntennaABS plastictransmittercontained in awearable athleticpouch beltABS plastictransmittercontained in awearable athleticpouch beltThermoplasticpolyurethanebody worn device
ComparatorStimwaveFreedom SCSSystem(K162161)StimwaveFreedom SCSSystem(K160600)StimwaveFreedom SCSSystem(K150517)StimwaveFreedom SCSSystem(K141399)
Software Level ofConcernModerateSame as K162161Same as K162161Same as K162161
iPad ApplicationWaveCrest™ 2.2WaveCrest™WaveCrest™WaveCrest™
KitsReceiver Kit, Spare Lead Kit,Trial Lead Kit, Sterile RevisionKit, WAA Kit, Charger KitReceiver Kit,Trial Lead Kit,WAA KitReceiver Kit,Trial Lead Kit,WAA KitReceiver Kit,Trial Lead Kit,WAA Kit

Table 5A. Comparison of Characteristics

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in dark blue and is inside of a dark blue rounded rectangle. To the right of the word "wave" are three dark blue curved lines that resemble radio waves.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a dark blue font, and the word "wave" is in a green font. To the right of the word "wave" is a graphic of three curved lines that resemble radio waves. Underneath the word "stimwave" is the word "TECHNOLOGIES" in a smaller, dark blue font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in a blue sans-serif font on the top left. Below that, the word "wave" is in a green sans-serif font. Below that, the word "TECHNOLOGIES" is in a blue sans-serif font with a blue line above and below it. To the right of the word "wave" are three concentric semi-circles in a blue color.

(*) asterisk denotes that formulas were used for the calculations.

8. Biocompatibility Data

The materials of the Freedom-8A/4A Stimulator (K162161) in direct contact with tissue remain unchanged from the Freedom SCS System (K141399, K150517, and K160600) and thus, the biocompatibility tests conducted on representative subassemblies of the Freedom SCS System (Freedom-4, K141399) directly apply to the Freedom SCS System (K150517, K160600, and K162161). The materials, construction and intended use of the Freedom SCS System is comparable to the predicate device, and have a long history of safety with respect to biocompatibility. The biological safety of the Freedom-8A/4A Stimulator (same as the Freedom-4 Stimulator) was evaluated in accordance to ISO 10993-1:2009 and guidance document Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing of the Freedom-8A/4A Stimulator (same as the Freedom-4 Stimulator) for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity demonstrated no negative impacts from the materials that are used in the Freedom SCS System. The Freedom-8A/4A Stimulator materials in direct tissue contact include Pellethane 55D (Stimulator) and Pt-Ir (90:10) (Stimulator only), both having an extensive record (previously cleared and approved) of chronic and carcinogenetic safety. The Receiver/RF Stylet is never in direct or indirect contact with tissue. The WAA is intended to be on top of a thin shirt or article of clothing around the midsection of the patient. The User Manual provided to the patient describes that the WAA should always be worn on top of a layer of clothing. The WAA does not come into contact with the patient's skin. The categorization by nature of body contact of the WAA is thus "noncontacting device", and not included in the scope of ISO 10993-1:2009. The Freedom SCS System meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.

9. Non-Clinical Performance Data

The Freedom SCS System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The Freedom SCS System complies with all design requirements and applicable voluntary standards.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. To the right of the word "wave" are three curved lines that resemble radio waves. Underneath the words "stim wave" is the word "TECHNOLOGIES" in blue.

Stimwave Technologies Incorporated Traditional 510(k) Premarket Submission Freedom Spinal Cord Stimulator System

AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the Freedom-8A/4A Stimulator was functional. receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008. For atmospheric pressure change, the Freedom-8A/4A Stimulator were functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008. This testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the packaging and the Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.

For testing external defibrillation exposure, the Freedom-8A/4A Stimulator and Receiver were verified as functional after exposure to external defibrillation. Thus, the Freedom SCS System complies with testing as specified by AAMI ANSI ISO 14708-3:2008. The Receiver/RF Stylet testing presented for K162161 is leveraged from K160600 and is directly applicable for demonstration of device safety and efficacy as Receiver/RF Stylet remains the same.

Following the thermal shock testing, the Freedom-8A/4A Stimulator was found to have "no irreversible damage" and fully functional as specified by the manufacturer, and to have no physical anomalies present at the time of inspection. Thus, the Freedom-8A/4A Stimulator comply with the thermal shock design requirements and the applicable standard. This testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.

For leakage current testing, the Freedom-8A/4A Stimulator was produced zero leakage current on all tested paths for all tested samples. Thus, the Freedom-8A/4A Stimulator comply with the leakage design requirements and the applicable standard. This testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.

For testing the insertion and withdrawal of the stylet within the Stimulator, the stylet was found to require less than 2.5N of insertion or withdrawal force for all tested stylets in all tested stimulator samples. For testing the insertion and withdrawal of the Receiver within the Stimulator, the Receiver/RF Stylet was found to require less than 2.2N of insertion or withdrawal force for all tested stylets in all tested stimulator samples. Visual inspection confirmed no damage was present in any stimulator samples. Thus, the Freedom-8A/4A Stimulator and Receiver/RF Stylet comply with design specifications for stylet insertion and withdrawal force. The Stimulator testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in large, bold, dark blue font, and the word "wave" is in large, bold, green font. Below the words "stim wave" is the word "TECHNOLOGIES" in a smaller, dark blue font, inside of a dark blue rectangle with rounded corners. To the right of the words is a graphic of three curved lines, which are also dark blue.

Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet testing presented for K162161 is leveraged from K160600 and is directly applicable for demonstration of device safety and efficacy as Receiver/RF Stylet remains the same.

For mechanical testing, the Freedom-8A/4A Stimulator passed all criteria of the test, showing no visible damage to the stimulator body or functional damage to the components. Mechanical testing included tensile testing and torsion testing. Thus, the Freedom-8A/4A Stimulator comply with all stimulator mechanical design requirements. This testing presented for K162161 is leveraged from K150517 and is directly applicable for demonstration of device safety and efficacy as the Freedom-8A/4A Stimulator remains the same. The Receiver/RF Stylet is a passive component that does not impact the outcome of the leveraged tests for safety and effectiveness.

IEC 60601-1 - The WAA was tested for compliance with IEC 60601-1. For testing the WAA for protection from temperature change, including shipping and storage temperature ranges, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA of the Freedom SCS System satisfies the outlined protection from temperature change design requirements and the applicable standard, IEC 60601-1. For atmospheric pressure change testing, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA of the Freedom SCS System satisfies the outlined atmospheric pressure change design requirements and the applicable standard, IEC 60601-1. For the push, drop, impact and mold stress relief testing of the WAA, it was determined through testing that the WAA is robust to withstand expected damage in accordance with general safety standards. The WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom-8A/4A SCS System satisfies the outlined push, drop, impact, and mold stress relief design requirements and the applicable standard, IEC 60601-1. For the identification, marking and documents of the WAA it was determined through an analysis of the labeling that the WAA complies with the requirements of the standard. All requirements and markings are clearly identified and viewable either from the external case of the product or from within the accompanying documents. For the means of protection, creepage distances, and air clearances of the WAA it was determined through an analysis of the design that the system satisfies the requirements of the applicable standard, IEC 60601-1.

IEC 60529 - The WAA was tested for compliance with IEC 60529. For testing the ingress of water, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the outlined Ingress of Water design requirements and the applicable standard IEC 60529. For particulate matter testing, the WAA met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a large, blue, sans-serif font on the top left. To the right of "stim" is the word "wave" in a green, sans-serif font. Below "stim wave" is the word "TECHNOLOGIES" in a smaller, blue, sans-serif font inside of a blue rounded rectangle. To the right of the words is a graphic of three concentric semi-circles in blue.

outlined Particulate Matter design requirements and the applicable standard, IEC 60529.

IEC 60601-1-2 - The WAA was tested for compliance with IEC 60601-1-2. For testing the WAA for electromagnetic compatibility, the unit met all acceptance criteria for emissions, low-frequency magnetic fields, immunity, electrostatic discharge, radiated RF electromagnetic fields, electrical fast transients and magnetic fields. The WAA operated within all test limits and showed no physical damage and was fully operational. Thus, the WAA for the Freedom SCS System satisfies the IEC 60601-1-2 standard.

The Freedom SCS System complies with the applicable standards for neurostimulators, electrical safety, electromagnetic interference and compatibility, packaging, and sterilization. The software of the Freedom SCS System passed all verification tests outlined and the design requirements for Software Verification have been met. The device passed all the testing in accordance with national and international standards.

Following performance testing, it has been determined that the Freedom SCS System is substantially equivalent to legally marketed predicate devices for the therapy for chronic, intractable pain of the trunk and/or lower limbs, including unilateral pain.

Due to the similarities between the legally marketed predicate devices (K141399, K150517, and K160600), and the Freedom SCS System (K162161), Stimwave Technologies Incorporated has leveraged applicable performance testing in addition to completed a number of tests that demonstrates substantial equivalence to the legally marketed predicate devices. The Freedom SCS System meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety confirms that the output meets the design inputs and specifications. The Freedom SCS System passed all testing stated above as shown by the acceptable results obtained.

10. Clinical Performance Data

There was no clinical testing required to support the medical device, as the indications for use are equivalent to the legally marketed predicate devices. These types of devices, including the legally marketed predicate devices. have been on the market for many vears with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to legally marketed predicate devices when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. The Freedom SCS System has the same intended use as the legally marketed predicates devices and is implanted percutaneous into the epidural space ranging from T5 to L5. Performance tested verified that the Freedom SCS

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a large, purple, sans-serif font. The word "wave" is in a large, green, sans-serif font. The word "TECHNOLOGIES" is in a smaller, purple, sans-serif font and is inside of a rounded rectangle. To the right of the words is a series of three curved lines that are purple and resemble radio waves.

Stimwave Technologies Incorporated Traditional 510(k) Premarket Submission Freedom Spinal Cord Stimulator System

System complies with all applicable voluntary standards such as IEC 60601-1, AAMI ANSI ISO 14708-3, and IEC 60529. The Freedom SCS System also meets the design requirements where no applicable standard could be used. This included Receiver/RF Stylet performance testing, stimulator body durability testing, programmable parameters, as well as power and performance of the WAA. There were no recognized performance standards for this device. There was no clinical testing performed on this device since performance testing demonstrated similar performance as the legally marketed predicate devices. and materials for the implanted stimulator are the same as the legally marketed predicate devices.

It has been shown in this 510(k) submission that the difference between the Freedom SCS System and the legally marketed predicate devices do not raise any questions regarding its safety and effectiveness as compared to legally marketed predicate devices. Freedom SCS System, as designed and manufactured, is determined to be substantially equivalent to the referenced legally marketed predicate devices.

§ 882.5880 Implanted spinal cord stimulator for pain relief.

(a)
Identification. An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.(b)
Classification. Class II (performance standards).