K Number
K141399

Validate with FDA (Live)

Date Cleared
2014-10-10

(135 days)

Product Code
Regulation Number
882.5880
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stimwave Technologies Incorporated Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral pain. The FRT4-A001 device is solely used for trial stimulation (no longer than 30 days) of the permanent FRE4-A001 device.

Device Description

The Stimwave Technologies Incorporated (Stimwave) Freedom Spinal Cord SCS System (System) is used for spinal column neural stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or lower limbs including unilateral or bilateral pain. The therapy utilizes pulsed electrical current to create an electrical energy field that acts on nerves near the dorsal column of the spine. The System is comprised of an implantable stimulator (Freedom-4 Stimulator) and an externally worn transmitter (Wearable Antenna Assembly (WAA)) to power the device. The System is implanted only following a successful trial period with the Trial Freedom-4 Stimulator.

AI/ML Overview

The document describes the Stimwave Technologies Incorporated Freedom Spinal Cord Stimulator (SCS) System and provides information to support its substantial equivalence to legally marketed predicate devices.

Here's an analysis of the acceptance criteria and supporting studies, formatted as requested:

Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with specific numerical targets and corresponding reported device performance values in a side-by-side format. Instead, it describes compliance with various national and international standards and design requirements through a series of non-clinical tests. The "reported device performance" is generally stated as having "passed," "met the criteria," or being "lower/higher than the allowable limit."

However, a comparison table (Table 5A) is provided that compares the technological characteristics of the Stimwave Freedom SCS System with three predicate devices. This table implicitly defines "acceptance" as being similar to or within the range of these predicate devices for many parameters.

Implicit Acceptance Criteria and Reported Performance (derived from Table 5A and non-clinical testing sections):

Acceptance Criteria CategoryImplicit Acceptance Criteria / StandardReported Device Performance
General Device CharacteristicsSame intended use, implant site, environmental use, intended clinician/user, electrode material, stimulator body material, method of introduction, tissue contact, sterilization, labeling, package.All "Same as Freedom" across predicate devices.
Physical Dimensions
- Stimulator LengthComparable to predicate devices (30-110 cm)45 centimeters (within range)
- DiameterComparable to predicate devices (1.3-1.37 mm)1.35 millimeters (within range)
- Electrode Array LengthComparable to predicate devices (24.0 mm)24.0 millimeters (same)
- No. of ElectrodesComparable to predicate devices (4 or 8)4 (within range)
- Electrode LengthComparable to predicate devices (3.0 mm)3.0 millimeters (same)
- Electrode SpacingComparable to predicate devices (4.0 mm)4.0 millimeters (same)
- Electrode Surface AreaComparable to predicate devices (12.25 mm² - ~13 mm²)12.72 mm² (comparable)
Electrical Performance ParametersComparable to predicate devices within ranges/values.See Table 5A for detailed comparison across various parameters like Pulse Frequency, Pulse Width, Current/Voltage Regulation, Output Voltage/Current, Waveform, Pulse Shape, Current/Charge/Power Densities, Net Charge, Output Power. Generally comparable or within acceptable ranges.
Material BiocompatibilityCompliance with ISO 10993-1:2009 and Blue Book Memorandum G95-1 for an implant device in contact with tissue/bone for cytotoxicity, sensitization, irritation, intracutaneous reactivity, acute systemic toxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity."No negative impacts" from materials, meets biological safety and compatibility requirements.
Temperature Change (Storage)Functional, safe rating following post visual inspection. Compliance with AAMI ANSI ISO 14708-3:2008.Stimulator was functional, received a safe rating, passed change of temperature testing. External unit met passing criteria.
Atmospheric Pressure ChangeFunctional following post testing functionality inspection. Compliance with AAMI ANSI ISO 14708-3:2008.Stimulator was functional, passed atmospheric pressure change testing. External unit met passing criteria.
MRI Safety - RF Induced HeatingMaximum temperature increase lower than allowable limit for 1.5T and 3T MRI. Compliance with ASTM F2182-11a.Produced a maximum temperature increase lower than the allowable limit for 1.5T and 3T MRI, thus passed.
MRI Safety - Image ArtifactsMaximum measured artifact size lower than allowable limit for 3T MRI. Compliance with ASTM F2119-07.Produced a maximum measured artifact size lower than the allowable limit for 3T MRI, thus passed. No detrimental image artifacts.
MRI Safety - Induced DisplacementSignificantly lower deflection value than allowable test limit. Compliance with ASTM F2052-06.Produced a significantly lower deflection value than the allowable test limit. Passed ASTM acceptance criteria for deflection angle.
MRI Safety - Induced TorqueLowest score (no torque) according to qualitative scale. Compliance with ASTM F2213-06.Produced the lowest score, no torque. Will not present additional risk.
Electrical Safety - Leakage CurrentZero leakage current on all tested paths for all samples.Produced zero leakage current.
Stimulator Insertion/WithdrawalLess than 2.5N of insertion or withdrawal force. No damage.Required less than 2.5N; no damage.
Stimulator Body FlexNo visible damage to stimulator body or components.Passed all criteria, no visible damage.
Destructive Pull TestMaximum elongation of 0.5% (well below 5% limit).Maximum elongation 0.5% (well below test limit of 5%).
External Defibrillation ExposureVerified to function as outlined.Verified to function as outlined.
External Unit RobustnessWithstand expected damage (push, drop, impact, mold stress relief). Compliance with IEC 60601-1.Met passing criteria, no physical damage, fully operational.
Electromagnetic Compatibility (EMC)Met all acceptance criteria for emissions, magnetic fields, immunity (electrostatic discharge, radiated RF fields, electrical fast transients/bursts). Compliance with IEC 60601-1-2.Met all acceptance criteria and operated within test limits.
Ingress of Water/Particulate MatterMet passing criteria of visual and functional inspections. Compliance with IEC 60529.Met passing criteria, no physical damage, fully operational.
Software VerificationPassed all verification tests outlined and design requirements met.Passed all verification tests.

Study Details

The document refers to a "submission" rather than a single specific study. The evidence for meeting acceptance criteria is predominantly based on non-clinical performance testing and a comparison to predicate devices.

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a separate "test set" in the context of device performance, but rather individual components were tested according to various standards. For example, "all tested samples" for leakage current and "all tested stylets in all tested stimulator samples" for insertion/withdrawal. The exact number of units/components tested for each non-clinical test is not specified.
    • Data Provenance: The testing was conducted by Stimwave Technologies Incorporated to verify compliance with national and international standards (AAMI ANSI ISO, ASTM, IEC). The data is prospective in the sense that the testing was performed on the Stimwave Freedom SCS System components specifically for this submission. The "data provenance" regarding the location of testing is not specified, but it's presumed to be within the manufacturer's testing facilities or accredited laboratories.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable in the typical sense for this submission. The "ground truth" for the non-clinical tests is established by the specific requirements and acceptance criteria of the referenced national and international standards (e.g., ISO, ASTM, IEC). These standards are developed and recognized by expert bodies in their respective fields. No adjudication by specific medical experts for interpreting test results is described; rather, results are objectively compared against predefined numerical or qualitative criteria from the standards.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • "None" in the context of human expert adjudication as this was non-clinical engineering and materials testing against established standards. The review process for the submission at the FDA would involve regulatory experts assessing the manufacturer's presented data against requirements.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This document pertains to a Spinal Cord Stimulator, a medical device that delivers electrical impulses for pain relief, not an imaging or diagnostic AI device that would involve "human readers" or AI assistance in interpretation. The nature of this device does not fit the MRMC study paradigm.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This question is not applicable. The device is a physical medical implant system, not a standalone algorithm. Its performance is inherent to its design and function, not an AI algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for demonstrating substantial equivalence against legally marketed predicate devices primarily relies on:
      • Standardized Test Methods and Acceptance Criteria: As set by numerous international and national standards (e.g., ISO 10993, AAMI ANSI ISO 14708-3, ASTM F2182-11a, IEC 60601-1) for biocompatibility, safety, and performance.
      • Direct Comparison to Predicate Devices: Explicitly measuring and comparing physical and electrical parameters against established values of legally marketed devices (Table 5A).
      • Design Requirements: Manufacturer-defined design specifications and requirements that the device must meet, often derived from industry best practices and safety considerations.
    • There is no mention of pathology or outcomes data as "ground truth" for this specific 510(k) submission, as it focuses on non-clinical substantial equivalence.
  7. The sample size for the training set:

    • This is not applicable. This is not an AI/ML device requiring a training set. The term "training set" is usually associated with machine learning models.
  8. How the ground truth for the training set was established:

    • This is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines connecting them, resembling a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 10, 2014

Elizabeth Greene Regulatory Affairs Manager Stimwave Technologies Incorporated 9375 East Shea Boulevard Suite 147 Scottsdale, Arizona 85260

Re: K141399

Trade/Device Name: Freedom Spinal Cord Stimulator (SCS) System Regulation Number: 21 CFR 882.5880 Regulation Name: Stimulator, Spinal-Cord, Implanted (Pain Relief) Regulatory Class: Class II Product Code: GZB Dated: September 5, 2014 Received: September 8, 2014

Dear Ms. Greene:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141399

Device Name

Freedom Spinal Cord Stimulator (SCS) System

Indications for Use (Describe)

The Stimwave Technologies Incorporated Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral pain. The FRT4-A001 device is solely used for trial stimulation (no longer than 30 days) of the permanent FRE4-A001 device.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below the words is the word "TECHNOLOGIES" in blue, inside of a blue rounded rectangle. To the right of the words is a graphic of concentric circles.

510(k) Summary

for

Freedom Spinal Cord Stimulator (SCS) System

1. Submission Sponsor

Stimwave Technologies Incorporated 420 Lincoln Road Suite 365 Miami Beach Florida 33139 USA Phone: 800.965.5134 Fax: 800.965.5134 Contact: Elizabeth Greene, Regulatory Affairs Manager

2. Date Prepared

May 27, 2014

3. Device Identification

Trade/Proprietary Name:Freedom Spinal Cord Stimulator (SCS) System
Common/Usual Name:Spinal Cord Stimulator
Classification Name:Stimulator, Spinal-Cord, Implanted (Pain Relief)
Classification Regulation:882.5880
Product Code:GZB
Device Class:Class II
Classification Panel:Neurology

4. Legally Marketed Predicate Device(s)

Medtronic Mattrix 3271/3272 Neuromodulation System (K934065) Medtronic Xtrel, Model Number 3425 Receiver (K883780) ANS Renew Neurostimulation System Transmitter, Model 2508, Receiver Model 3408, Antennae Models 1220 and 1230, Lead Models 3143, 3146, 3153, 3156, 3183 and 3186, Extension Models 3382, 3383, 3341, 3342 and 3343 (K000852)

5. Device Description

The Stimwave Technologies Incorporated (Stimwave) Freedom Spinal Cord SCS System (System) is used for spinal column neural stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or lower limbs including unilateral or bilateral pain. The therapy utilizes pulsed electrical current to create an electrical energy field that

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the two words is the word "TECHNOLOGIES" in dark blue, inside of a rounded rectangle. To the right of the words are a series of concentric arcs in dark blue.

acts on nerves near the dorsal column of the spine. The System is comprised of an implantable stimulator (Freedom-4 Stimulator) and an externally worn transmitter (Wearable Antenna Assembly (WAA)) to power the device. The System is implanted only following a successful trial period with the Trial Freedom-4 Stimulator.

Freedom-4 Stimulator (Receiver Kit)

Freedom-4A polyurethane (Pellethane 55D) casing with an embedded receiver, flexible circuit
Stimulatorboard and electrodes (Platinum Iridium 90:10) that is placed percutaneously in the
patient's epidural space. The Stimulator has four (4) electrodes.
Stylet(s)A stainless steel wire with a nylon hub that is inserted into the open central lumen
of the stimulator to provide rigidity during implantation. Four (4) stylets are
provided in the Receiver Kit, two straight and two bent with diameters of 0.25 mm
or 0.30 mm.
TuohyA 14-gauge stainless steel needle that is inserted into the epidural space of the
Needlepatient and acts as a conduit to guide the introducer sheath.
IntroducerA PEBAX sheath that slides over the Tuohy needle before insertion, and acts as a
Sheathconduit for passage of the stimulator into the epidural space. The introducer sheath
is RF transparent, allowing the transmission of wireless power to the Stimulator for
intraoperative stimulation testing.
GuidewireA stainless steel, rigid, solid core guidewire used to create a hollow pathway in the
epidural space for the Stimulator to pass through more easily.
SutureA polyurethane (Pellethane 55D) cap that is placed over the proximal end of the
SleeveStimulator. The Sleeve Cap is attached to the Freedom-4 Stimulator and can be
Capsutured to tissue to reduce the possibility of device migration.

Wearable Antenna Assembly (WAA Kit)

WAAThe WAA device includes the following components sealed within the unit:
i. Transmitting (Tx) Antenna - Used to transmit microwave energy to theimplanted Stimulator;
ii. Microwave Field Stimulator (MFS) – A printed circuit board (PCB) thatgenerates 915 MHz RF power with embedded waveform parameter settings;
iii. User Interface (UI) Board – A PCB that contains the circuitry to allow the userto select clinician-programmed parameter combination options or increase ordecrease power amplitude;
iv. Battery Charger Board and Coil – A PCB and inductive charging coil thatreceives energy from an external battery-charging unit. The battery chargerboard communicates with the MFS to transfer charge into the 3.7V battery byfacilitating power transfer and warns the system when battery power is low;
v. Battery – A lithium ion rechargeable battery.
WirelessBatteryChargingPadA Qi-compliant charging pad that uses inductive charging technology to rechargethe encased lithium ion battery of the WAA.
AdjustableStrapAn elastic nylon strap that attaches to the two ends of the WAA. This strap can beadjusted to various lengths to accommodate the varying patient population.

Trial Freedom-4 Stimulator (Trial Kit)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is underlined. To the right of the words is a blue graphic of concentric arcs.

TrialFreedom-4StimulatorA polyurethane (Pellethane 55D) casing with an embedded receiver, flexible circuit board and electrodes (Platinum Iridium 90:10) that is placed percutaneously in the patient's epidural space. The Stimulator has four (4) electrodes.
Stylet(s)A stainless steel wire with a nylon hub that is inserted into the open central lumen of the stimulator to provide rigidity during implantation. Four (4) stylets are provided in the Receiver Kit, two straight and two bent with diameters of 0.25 mm or 0.30 mm.
TuohyNeedleA 14-gauge stainless steel needle that is inserted into the epidural space of the patient and acts as a conduit to guide the introducer sheath.
IntroducerSheathA PEBAX sheath that slides over the Tuohy needle before insertion, and acts as a conduit for passage of the stimulator into the epidural space. The introducer sheath is RF transparent, allowing the transmission of wireless power to the Stimulator for intraoperative stimulation.
GuidewireA stainless steel, rigid, solid core guidewire used to create a hollow pathway in the epidural space for the Stimulator to pass through more easily.

6. Indication for Use Statement

The Stimwave Technologies Incorporated Freedom Spinal Cord Stimulator (SCS) System is intended as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or lower limbs, including unilateral or bilateral pain. The FRT4-A001 device is for trial stimulation of the FRE4-A001 device for permanent implantation.

7. Substantial Equivalence Discussion

The following table compares the Stimwave Freedom SCS System to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence. Stimwave physically measured Medtronic and ANS predicate devices to obtain the values listed in Table 5A.

ComparatorStimwaveFreedom SCSSystemMedtronicMattrix 3271/3272(K934065)Medtronic Xtrel,Model Number3425 (K883780)ANS Renew(K000852)
Product CodeGZBGZB and GZFGZBGZB
Regulation No.882.5880882.5880882.5880882.5880
Regulation NameStimulator,Spinal-Cord,Implanted (PainRelief)Stimulator, Spinal-Cord, Implanted(Pain Relief)Stimulator, Spinal-Cord, Implanted(Pain Relief)Stimulator,Spinal-Cord,Implanted (PainRelief)
Intended UseStimulation ofspinal cord forchronic,intractable pain oftrunk and lowerlimbsSame as FreedomSame as FreedomSame as Freedom
ComparatorStimwaveFreedom SCSSystemMedtronicMattrix 3271/3272(K934065)Medtronic Xtrel,Model Number3425 (K883780)ANS Renew(K000852)
Implant SiteEpidural space,L5 to T5Same as FreedomSame as FreedomSame as Freedom
Environmental UseHospital, HomeSame as FreedomSame as FreedomSame as Freedom
Intended ClinicianOrthopedic,Neurosurgeon,AnesthesiologistSame as FreedomSame as FreedomSame as Freedom
Intended UserLaypersonSame as FreedomSame as FreedomSame as Freedom
Electrode MaterialPlatinum-iridium90:10Same as FreedomSame as FreedomSame as Freedom
Stimulator BodyMaterialPolyurethane2363-55DSame as FreedomSame as FreedomSame as Freedom
Cable FeaturesMultilumen TubeCoiled WiresCoiled WiresBraided Wire
Stimulator Length45 centimeters30 to 110centimeters30 to 110centimeters30 centimeters,and 60centimeters
Diameter1.35 millimeters1.3 millimeters1.3 millimeters1.37 millimeters
Electrode ArrayLength24.0 millimetersSame as FreedomSame as FreedomSame as Freedom
No. of Electrodes4Same as FreedomSame as Freedom4 or 8
Electrode Length3.0 millimetersSame as FreedomSame as FreedomSame as Freedom
Electrode Spacing4.0 millimetersSame as FreedomSame as FreedomSame as Freedom
Electrode SurfaceArea12.72 mm²12.25 mm²12.25 mm²"Approximately13 mm²"
Method ofIntroductionPercutaneousSame as FreedomSame as FreedomSame as Freedom
Tissue ContactYesSame as FreedomSame as FreedomSame as Freedom
SterilizationEthylene Oxide(EO)Same as FreedomSame as FreedomSame as Freedom
LabelingLabeled as Sterile,Single Use,PrescriptionDeviceSame as FreedomSame as FreedomSame as Freedom
PackageBlisterTray/Tyvek LidSame as FreedomSame as FreedomSame as Freedom
Pulse Frequency2 to 1500 Hertz5 to 240 Hertz5 to 1400 Hertz10 to 1500 Hertz
Pulse Width50 to 500microseconds50 to 500microseconds50 to 1000microseconds50 to 500microseconds
Current/VoltageRegulatedCurrentVoltageVoltageCurrent
Output Voltage(300 Ω)0 to 6.3 V0 to 7 V0 to 5.4 V0 to 5.7 V
Output Voltage(500 Ω)0 to 7.2 V0 to 10.8 V0 to 7.1 V0 to 7.6 V
Output Voltage(800 Ω)0 to 8.0 V0 to 11.6 V0 to 8.4 V0 to 9.6 V
Output Current(300 Ω)0 to 21 mA0 to 23.3 mA0 to 18.0 mA0 to 19.0 mA
Output Current(500 Ω)0 to 15 mA0 to 21.6 mA0 to 14.2 mA0 to 15.2 mA
ComparatorStimwaveFreedom SCSSystemMedtronicMattrix 3271/3272(K934065)Medtronic Xtrel,Model Number3425 (K883780)ANS Renew(K000852)
Output Current(800 Ω)0 to 10 mA0 to 14.5 mA0 to 10.5 mA0 to 12.0 mA
WaveformCharge Balanced(delayed)BiphasicasymmetricalCharge BalancedBiphasicasymmetricalCharge BalancedBiphasicasymmetricalCharge Balanced(delayed)Biphasicasymmetrical
Pulse ShapeDecayingExponentialDecayingExponentialDecayingExponentialDecayingExponential
Average CurrentDensity (300 Ω)111.6 mA/cm²175.0 mA/cm²125.8 mA/cm²117.7 mA/cm²
Average CurrentDensity (500 Ω)96.7 mA/cm²151.7 mA/cm²101.7 mA/cm²103.1 mA/cm²
Average CurrentDensity (800 Ω)77.0 mA/cm²106.7 mA/cm²75.8 mA/cm²86.2 mA/cm²
Maximum PhaseCharge* (300 Ω)10.5 µC/pulse11.7 µC/pulse18.0 µC/pulse9.5 µC/pulse
Maximum PhaseCharge* (500 Ω)7.2 µC/pulse10.8 µC/pulse14.2 µC/pulse7.6 µC/pulse
Maximum PhaseCharge* (800 Ω)5.0 µC/pulse7.3 µC/pulse10.5 µC/pulse6.0 µC/pulse
Maximum ChargeDensity* (300 Ω)82.5 µC/cm²97.2 µC/cm²150.0 µC/cm²73.1 µC/cm²
Maximum ChargeDensity* (500 Ω)56.6 µC/cm²90.0 µC/cm²118.3 µC/cm²58.5 µC/cm²
Maximum ChargeDensity* (800 Ω)39.3 µC/cm²60.4 µC/cm²87.5 µC/cm²46.2 µC/cm²
Maximum CurrentDensity* (300 Ω)165.1 mA/cm²194.4 mA/cm²150.0 mA/cm²146.2 mA/cm²
Maximum CurrentDensity* (500 Ω)113.2 mA/cm²180.0 mA/cm²118.3 mA/cm²116.9 mA/cm²
Maximum CurrentDensity* (800 Ω)78.6 mA/cm²120.8 mA/cm²87.5 mA/cm²92.3 mA/cm²
Net Charge0 µC0 µC0 µC0 µC
Average PhasePower (300 Ω)0.060 W/phase0.132 W/phase0.068 W/phase0.070 W/phase
Average PhasePower (500 Ω)0.076 W/phase0.166 W/phase0.074 W/phase0.090 W/phase
Average PhasePower (800 Ω)0.060 W/phase0.131 W/phase0.066 W/phase0.100 W/phase
Average PhasePower Density(300 Ω)0.48 W/cm²/phase1.10 W/cm²/phase0.57 W/cm²/phase0.54 W/cm²/phase
Average PhasePower Density(500 Ω)0.59 W/cm²/phase1.38 W/cm²/phase0.62 W/cm²/phase0.69 W/cm²/phase
Average PhasePower Density(800 Ω)0.60 W/cm²/phase1.09 W/cm²/phase0.55 W/cm²/phase0.77 W/cm²/phase
Pulse DeliveryModeContinuousContinuousContinuousContinuous
ComparatorStimwaveFreedom SCSSystemMedtronicMattrix 3271/3272(K934065)Medtronic Xtrel,Model Number3425 (K883780)ANS Renew(K000852)
ON/OFF TimesNo CyclingON/OFF CyclingOptionON/OFF CyclingOptionNo Cycling
Current PathOptionsBipolarBipolarBipolarBipolar
Power DeliveryCoupled receiver,built intoStimulator bodyCoupled receiver,Radio FrequencytransmissionCoupling receiver,Radio FrequencytransmissionCoupled receiver,hardwired withconnector
TransmitFrequency915 MHz2 MHz1.60 MHz2 MHz
MaterialPlatinum-iridium90:10,Polyurethane2363-55DSame as FreedomSame as FreedomSame as Freedom
SterileYes - ethyleneoxideSame as FreedomSame as FreedomSame as Freedom
Single-UseYesYesYesYes
Shelf Life1 year1 year1 year2 years
Complies with ISO10993-1YesYesYesYes
Safety TestingPassedYesYesYesYes

Table 5A. Comparison of Characteristics

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. Below that, the word "TECHNOLOGIES" is in blue and in a rounded rectangle. To the right of the words is a blue image of concentric circles.

{7}------------------------------------------------

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. To the right of the word "wave" are three curved lines that resemble radio waves. Underneath the words "stim wave" is the word "TECHNOLOGIES" in blue.

(*) asterisk denotes that formulas were used for the calculations.

8. Biocompatibility Data

The Freedom SCS System biocompatibility tests were conducted on representative subassemblies of the product. The materials, construction and intended use of the Freedom SCS System is comparable to the predicate device, and have a long history of safety with respect to biocompatibility. The biological safety of the Freedom-4 Stimulator was evaluated in accordance to ISO 10993-1:2009 and guidance document Blue Book Memorandum G95-1 Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing. Under these, for the stated indications for use, the device was classified as a (C), implant device in contact with tissue/bone. The results for the biocompatible testing for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, implantation (4, 8, and 13 weeks), and subchronic toxicity demonstrated no negative impacts from the materials that are used in the Freedom SCS System. The Freedom-4 Stimulator materials in direct tissue contact include Pellethane 55D and Pt-Ir (90:10), both having an extensive record (previously cleared and approved) of subchronic toxicity, chronic, and carcinogenetic safety. The WAA is intended to be on top of a thin shirt or article of clothing around the midsection of the patient. The User Manual provided to the patient describes that the WAA should always be worn on top of a layer of clothing. The WAA does not come into contact with the patient's skin. The categorization by nature of body contact of the WAA is thus "non-contacting device", and not included in the scope of ISO 10993-1:2009. The Freedom SCS System meets biological safety and compatibility requirements of ISO 10993-1:2009 and Blue Book Memorandum G95-1.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the words is the word "TECHNOLOGIES" in a smaller font, inside of a dark blue rounded rectangle. To the right of the words is a series of concentric arcs in dark blue.

9. Non-Clinical Performance Data

The Freedom SCS System was tested to verify that the performance meets the system design requirements as well as all applicable voluntary standards. The Freedom SCS System complies with all design requirements and applicable voluntary standards.

AAMI ANSI ISO 14708-3:2008 - For protection from temperature change including shipping and storage temperature ranges, the Freedom-4 Stimulator was functional, receiving a safe rating following post visual inspection and passed the change of temperature testing performed as specified by AAMI ANSI ISO 14708-3:2008.

For atmospheric pressure change, the Freedom-4 Stimulator was functional following post testing functionality inspection and passed atmospheric pressure change testing as specified by AAMI ANSI ISO 14708-3:2008.

For magnetic resonance imaging (MRI) radio frequency (RF) induced heating as related to specific absorbance rate (SAR), the Freedom-4 Stimulator produced a maximum temperature increase lower than the allowable limit for the 1.5T and 3T MRI procedure and thus passed the 1.5T and 3T testing.

ASTM F2182-11a - In accordance with F2182-11a - American Society for Testing and Materials (ASTM) International Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging, the Freedom-4 Stimulator showed that its presence would not cause injury to the patient with the implant during an MRI procedure. The Freedom-4 Stimulator is a passive implant that is not powered while the external unit is not transmitting to it. For MRI exposure. RF image artifact testing, the Freedom-4 Stimulator produced a maximum measured artifact size lower than the allowable limit for the 3T MRI procedure and thus passed the testing.

ASTM F2119-07 - In accordance with F2119-07 - American Society for Testing and Materials (ASTM) International Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants, the Freedom-4 Stimulator showed that it does not produce image artifacts in 3T MRI procedures that are detrimental to the outcome of the procedure. For MRI exposure, induced displacement, the Freedom-4 Stimulator produced a significantly lower deflection value than the allowable test limit.

ASTM F2052-06 - In accordance with F2052-06 - The American Society for Testing and Materials (ASTM) International Designation Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, the Freedom-4 Stimulator showed that it does not harm the patient due to its displacement by forces induced by MRI exposure. The Freedom-4 Stimulator passes the ASTM acceptance criteria for deflection angle in a 3T MRI system and will not present an additional risk or hazard to a patient in the 3T or less MRI environment. For MRI exposure, induced torque, the Freedom-4 Stimulator produced the lowest score, no torque, according to the qualitative scale outlined by the applicable

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in blue, and the word "wave" is in green. The word "TECHNOLOGIES" is in blue and is located under the word "wave". To the right of the words is a blue graphic that looks like sound waves.

ASTM standard.

ASTM F2213-06 - In accordance with F2213-06 - American Society for Testing and Materials (ASTM) International Standard Test Method for Measurement of Magnetically Induced Torque on Passive Implants in the Magnetic Resonance Environment, the Freedom-4 Stimulator must show that it does not harm the patient due to its torque by forces induced by MRI exposure. The Freedom-4 Stimulator will not present an additional risk or hazard to a patient in the 3T MRI environment or less with regard to torque. The stimulation waveforms following post-exposure showed no component damage had occurred in any stimulator. During testing, no component damage was observed in any waveform. Based on these test results, the Freedom-4 Stimulator will be fully functional following standard 1.5T and 3T MR procedures and its performance or functionality is not affected by such exposures. Following the thermal shock testing, the Freedom-4 Stimulator was found to have "no irreversible damage" and fully functional as specified by the manufacturer, and to have no physical anomalies present at the time of inspection. Thus, the Freedom-4 Stimulator complies with the thermal shock design requirements and the applicable standard. For leakage current testing, the Freedom-4 Stimulator produced zero leakage current on all tested paths for all tested samples. Thus, the Freedom-4 Stimulator complies with the leakage design requirements and the applicable standard. For testing the insertion and withdrawal of the stimulator within the stylet, the Freedom-4 Stimulator was found to require less than 2.5N of insertion or withdrawal force for all tested stylets in all tested stimulator samples. Visual inspection confirmed no damage was present in any stimulator samples.

Thus, the Freedom-4 Stimulator complies with all design. For stimulator body flex testing, the Freedom-4 Stimulator passed all criteria of the test, showing no visible damage to the stimulator body or the components. Thus, the Freedom-4 Stimulator complies with all stimulator body flex test design requirements. For transition stimulator body to the flex board, the Freedom-4 Stimulator passed all criteria of the test with no visible damage to the segment. For the destructive pull test, the Freedom-4 Stimulator was verified to function as outlined by the criteria of the test and the maximum elongation found in all samples was 0.5%, well below the test limit of 5%. For testing external defibrillation exposure, the Freedom-4 Stimulator was verified to function as outlined by the criteria of the testing protocol.

IEC 60601-1 - For testing the external unit for protection from temperature change, including shipping and storage temperature ranges, the external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the external unit of the Freedom SCS System satisfies the outlined protection from temperature change design requirements and the applicable standard, IEC 60601-1. For atmospheric pressure change testing, the external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the external unit of the Freedom SCS System satisfies the outlined atmospheric pressure change design requirements and the applicable standard, IEC

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in a large, dark blue font on the left side of the image. To the right of "stim" is the word "wave" in a large, green font. Below the words "stim wave" is the word "TECHNOLOGIES" in a smaller, dark blue font. To the right of the words is a graphic of concentric arcs in a light blue color.

60601-1. For the push, drop, impact and mold stress relief testing of the external unit, it was determined through testing that the external unit is robust to withstand expected damage in accordance with general safety standards. The external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the outlined push, drop, impact, and mold stress relief design requirements and the applicable standard, IEC 60601-1. For the identification, marking and documents of the external unit it was determined through an analysis of the labeling that the external unit complies with the requirements of the standard. All requirements and markings are clearly identified and viewable either from the external case of the product or from within the accompanying documents. For the means of protection, creepage distances, and air clearances of the external unit it was determined through an analysis of the design that the system satisfies the requirements of the applicable standard. IEC 60601-1.

IEC 60601-1-2 - For testing the external unit for electromagnetic compatibility, the unit met all acceptance criteria for emissions, low-frequency magnetic fields, immunity, electrostatic discharge, radiated RF electromagnetic fields, electrical fast transients and bursts, and magnetic fields. The external unit operated within all test limits and showed no physical damage and was fully operational. Thus, the external unit for the Freedom SCS System satisfies the IEC 60601-1-2 standard.

IEC 60529 - For testing the ingress of water, the external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the outlined Ingress of Water design requirements and the applicable standard IEC 60529. For particulate matter testing, the external unit met the passing criteria of both of the visual and functional inspections following the testing. It showed no physical damage and was fully operational. Thus, the WAA component of the Freedom SCS System satisfies the outlined Particulate Matter design requirements and the applicable standard, IEC 60529. The software of the Freedom SCS System passed all verification tests outlined and the design requirements for Software Verification have been met.

The Freedom SCS System complies with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.

Following performance testing, it has been determined that the Freedom SCS System is substantially equivalent to legally marketed predicate devices for the therapy for chronic, intractable pain of the trunk and/or lower limbs, including unilateral pain.

Stimwave Technologies Incorporated completed a number of tests for the Freedom SCS System that demonstrates substantial equivalence to the legally marketed predicate devices. The Freedom SCS System meets all the requirements for overall design,

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Stimwave Technologies. The word "stim" is in dark blue, and the word "wave" is in green. Below the words is the word "TECHNOLOGIES" in a smaller font, inside of a dark blue rounded rectangle. To the right of the words is a series of concentric arcs in dark blue.

sterilization, biocompatibility, and electrical safety confirms that the output meets the design inputs and specifications. The Freedom SCS System passed all testing stated above as shown by the acceptable results obtained.

10. Clinical Performance Data

There was no clinical testing required to support the medical device as the indications for use is equivalent to the legally marketed predicate device. These types of devices, including the legally marketed predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to legally marketed predicate devices when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. The Freedom SCS System has the same intended use as the legally marketed predicates devices and is implanted percutaneous into the epidural space ranging from T5 to L5. Performance tested verified that the Freedom SCS System complies with all applicable voluntary standards such as IEC 60601-1. AAMI ANSI ISO 14708-3, IEC 60529 and MRI safety tests. The Freedom SCS System also meets the design requirements where no applicable standard could be used. This included stimulator body durability testing, programmable parameters, as well as power and performance of the external unit. There were no recognized performance standards for this device. There was no clinical testing performed on this device since performance testing demonstrated similar performance as the legally marketed predicate devices, and materials for the implanted stimulator are the same as the legally marketed predicate devices.

It has been shown in this 510(k) submission that the difference between the Freedom SCS System and the legally marketed predicate devices do not raise any questions regarding its safety and effectiveness as compared to legally marketed predicate devices. Freedom SCS System, as designed and manufactured, is determined to be substantially equivalent to the referenced legally marketed predicate devices.

§ 882.5880 Implanted spinal cord stimulator for pain relief.

(a)
Identification. An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.(b)
Classification. Class II (performance standards).