(189 days)
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
The Device is equipped with three replaceable treatment heads; two treatment heads are electrode heads for micro current output to stimulate facial skin (Bigger one, and smaller one), the other treatment head is LED lamp holder.
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
For two micro current treatment heads of instrument, they can be exchanged for different skin part to use. The bigger one is for face tissue on cheeks and the smaller one is for facial tissue below eye region. It should be used about 10 minutes/2 to 3 times per week. The red is for red-light caring (about 3 minutes/2 to 3 times per week.)The product is supplied with an internal lithium rechargeable battery; It can be charged by 5V voltage of battery charger connected by standard USB cable. During battery charging, the LED lights start flash circling until charging is completed.
There are two kinds of electrode heads, one bigger electrode head is for cheek stimulation; and smaller electrode head is for stimulation on under eyes area. There are no electronic component inside of either heads. It is only for conduct stimulation current.
This document is a 510(k) Pre-Market Notification for the "Micro-current Wrinkle Reduction Facial Service, Model EP-400". It details the device's technical specifications and compares it to predicate devices to establish substantial equivalence. However, it does not contain a study proving the device meets specific acceptance criteria in terms of clinical performance or effectiveness beyond safety and technical standards.
The document discusses tests for safety and technical performance, but not clinical efficacy in reducing wrinkles or providing facial stimulation results. The "Test Summary" section lists various engineering and safety standards used for evaluation, which are common for device clearance but do not address clinical outcomes.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria for clinical performance (e.g., wrinkle reduction percentage, subjective improvement scores) or reported device performance against such metrics. Instead, the acceptance is based on compliance with various engineering, safety, and compatibility standards.
| Acceptance Criteria (Type) | Reported Device Performance (Compliance) |
|---|---|
| Electrical safety (IEC 60601-1, IEC 60601-2-10) | Compliant |
| Electromagnetic compatibility (IEC 60601-1-2) | Compliant |
| Photobiological safety (IEC 62471) | Compliant |
| Software verification and validation (FDA Guidance) | Compliant |
| Waveform output specifications (IEC 60601-2-10, Guidance for Powered Muscle Stimulator) | Compliant (waveform test report to verify output specifications) |
| Biocompatibility (ISO 10993-5, ISO 10993-10) | Compliant |
2. Sample size used for the test set and the data provenance
The document describes bench testing for engineering and safety standards, not a clinical test set with human subjects. Therefore, the concept of "sample size used for the test set" and "data provenance" (country of origin, retrospective/prospective) in a clinical context is not applicable here. The testing was performed on the device itself in a lab setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable as there is no clinical test set with a "ground truth" established by human experts for clinical outcomes. The "ground truth" for the engineering and safety tests would be the specifications outlined in the respective standards, verified by testing specialists in a lab.
4. Adjudication method for the test set
Not applicable as there is no clinical test set requiring adjudication of findings.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical therapy/cosmetic device, not an AI-powered diagnostic or assistive technology for human readers/clinicians, and no MRMC study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is geared towards AI/software performance. The document describes a physical medical device. While it mentions "Software verification and validation test" according to FDA guidance, this refers to the internal software controlling the device's functions, not an AI algorithm performing a standalone clinical task. Therefore, "standalone (algorithm only without human-in-the-loop performance)" is not applicable in the context of clinical efficacy for this device.
7. The type of ground truth used
For the safety and performance tests conducted, the "ground truth" would be the technical specifications, requirements, and acceptable ranges defined by the cited IEC and ISO standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 62471, ISO 10993-5, ISO 10993-10, and FDA Guidance for Pre Market Submissions and for Software Contained in Medical Devices, Guidance for Powered Muscle Stimulator).
8. The sample size for the training set
Not applicable. This is not a machine learning or AI algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for an AI algorithm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2017
Li-Tek Electronic Technology Corporation % Jet Li Regulation manager Guangzhou LETA Testing Technology Co., Ltd 6F, No.1 TianTai Road, Science City, LuoGang District Guangzhou, CN Guangdong
Re: K162106
Trade/Device Name: Micro-current Wrinkle Reduction Facial Service, Model EP-400 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: December 31, 2016 Received: January 9, 2017
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162106
Device Name
Micro-current Wrinkle Reduction Facial service, Model EP-400
Indications for Use (Describe)
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K162106
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
510k preparation date: 2017-01-25
1. Submitter's Information
510(k) Owner's Name: Li-Tek Electronics Technologies Establishment Registration Number: Address: No.8~13,the industrial park of Jinshagang Shixia village, Dalangtown, Dongguan city, Guangdong, China
Phone: 0769-83117755 Fax: 0769-83117759 Contact Person: Barry Yuan (Quality Director) E-mail: quality5@li-tek.com
Application Correspondent:
Company: Guangzhou LETA Testing Technology Co., Ltd. Address: 6F, No.1 TianTai road Science City, LuoGang District, GuangZhou City, China Contact Person: Mr. Jet Li Tile: Regulation Manager Tel: +86-20-22325619 Email: med-jl@foxmail.com
2. Subject Device Information
Type of 510(k) submission: Traditional Common Name: Laser surqical instrument for use in general and plastic surgery and in dermatology, Transcutaneous electrical nerve stimulator for pain relief.
Trade Name: Micro-current Wrinkle Reduction Facial service, model EP-400
Classification Name: Light Based Over The Counter Wrinkle Reduction; Stimulator,
{4}------------------------------------------------
Transcutaneous Electrical, Aesthetic Purposes Review Panel: General & Plastic Surgery, Neurology
Product Code: OHS, NFO
Requlation Number: 878.4810, 882.5890
Regulation Class: 2
3. Predicate Device Information
| Sponsor | BiosonicTechnologies,LIC. | EVERYWAYMEDICALINSTRUMENTCO., LTD. | Carol ColeCompany | Nutra Luxe MD,LLC |
|---|---|---|---|---|
| DeviceName | Beautiful ImageModel 900 FacialToning Device | MT-200 FacialMENS | NOFACE® Plus | Nutra Light Red |
| 510(k)Number | K130045 | K142794 | K103472 | K141308 |
| ProductCode | NFO | NFO | NFO | OHS |
| RegulationNumber | 882.5890 | 882.5890 | 882.5890 | 878.4810 |
| RegulationClass | 2 | 2 | 2 | 2 |
4. Device Description
The Device is equipped with three replaceable treatment heads; two treatment heads are electrode heads for micro current output to stimulate facial skin (Bigger one, and smaller one), the other treatment head is LED lamp holder.
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
For two micro current treatment heads of instrument, they can be exchanged for different skin part to use. The bigger one is for face tissue on cheeks and the smaller one is for facial tissue below eye region. It should be used about 10 minutes/2 to 3 times per week. The red is for red-light caring (about 3 minutes/2 to 3 times per week.)The product is supplied with an internal lithium rechargeable battery; It can be charged by 5V voltage of battery charger connected by standard USB cable. During battery charging, the LED lights start flash circling until charging is completed.
{5}------------------------------------------------
There are two kinds of electrode heads, one bigger electrode head is for cheek stimulation; and smaller electrode head is for stimulation on under eyes area. There are no electronic component inside of either heads. It is only for conduct stimulation current.
5. Intended Use / Indications for Use
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
6. Test Summary
EP-400 has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety test according to IEC 60601-1 and IEC 60601-2-10 standards
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Photobiological safety of lamps and lamp systems according to IEC62471
- � Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
- � Waveform test report to verify the output specifications of the device according to IEC 60601-2-10 and Guidance for Powered Muscle Stimulator.
7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of EP-400 is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety and effectiveness.
| ElementsofComparison | SubjectDevice | Predicate Device | Predicate Device | Predicate Device | PredicateDevice | Remark | |
|---|---|---|---|---|---|---|---|
| Basic Unit Characteristics | |||||||
| Device Nameand Model | Micro-currentWrinkleReductionFacial Service,model EP-400 | BeautifulImageModel 900FacialToning Device | MT-200MENS | NOFACE® Plus | Nutra Light Red | -- | |
| 510(K)Number | K162106 | K130045 | K142794 | K103472 | K141308 | -- | |
| Product Code | NFO, OHS | NFO | NFO | NFO | OHS | -- | |
| RegulationNumber | 882.5890878.4810 | 882.5890 | 882.5890 | 882.5890 | 878.4810 | -- | |
| ElementsofComparison | SubjectDevice | Predicate Device | Predicate Device | Predicate Device | PredicateDevice | Remark | |
| Intended Use | The device isintended forfacialstimulation byelectrodeheads forcosmetic use.The device isalso intendedfor thetreatment ofperiorbitalwrinkles withred LightEmitting Diode(LED) head. Itis for over-the-counter use. | BiosonicTechnologiesModel 900 FacialToning Device isintended for facialstimulation and isindicated forprescriptioncosmetic use. Theanatomical site forapplication of theModel 900 is theface. | TheEverywayFacialMENS,model: MT-200 isintended for facialstimulationandindicated for over-the-countercosmetic use. | TheNOFACE®Plus Facial ToningDevice is intendedforfacialstimulation and isindicated for over-the-countercosmetic use. (21CFR 801 SubpartC). The anatomicalsite for applicationof the NOFACE®Plus is the face. | The Nutra LightRed is a non-invasive LEDlight device isintended/indicated for over- the -counter use forthe treatment ofperiorbitalwrinkles, andrhytides. | SENote 1 | |
| Apply parts | For microcurrentstimulation:Face;For Red light:periorbital | Face | Face | Face | periorbital | SE | |
| PowerSources | 1200mAhlithium battery | One 6V battery | 9-Volt battery | 4 rechargeablebatteries | Internal NI-MHrechargeablebattery | SENote 1 | |
| Method of LineCurrentIsolation | Battery SupplyN/A | N/A | Type BF | N/A | N/A | SENote 1 | |
| For Micro current facial stimulation function | |||||||
| Number ofModes forMicro currentstimulation | 1 | 1 | 3 (8Hz, 9Hz, 10Hz) | 1 | N/A | SENote 1 | |
| Number ofChannels forMicro currentstimulation | 1 | 1 | 1 | 1 | N/A | SE | |
| Elements ofComparison | SubjectDevice | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Remark | |
| -Synchronousor Alternating | Alternating | N/A | Alternating | Alternating | N/A | SE Note 1 | |
| Regulated Currentor RegulatedVoltage | RegulatedVoltage | Both | 0-3mA (load 4kΩ) | Regulated Voltage | N/A | SE Note 1 | |
| Software/Firmware/Microprocessor control | Yes | Yes | Yes | Yes | Yes | SE | |
| AutomaticOverload Trip | Yes | Yes | Yes | Not required due tocircuit design | N/A | SE | |
| Automatic No-load Trip | Yes. | Yes | Yes | Yes | N/A | SE | |
| AutomaticShut Off | Yes. | Yes | Yes | Yes | N/A | SE | |
| PatientOverrideControl | Yes | Yes | N/A | Yes | N/A | SE | |
| On/OffStatus | Yes | Yes | Yes | Yes | N/A | SE | |
| IndicatororDisplay | LowBattery | Yes | Yes | Yes | Yes | N/A | SE |
| Voltage/CurrentLevel | Yes | Yes | Yes | Yes | N/A | SE | |
| Timer Range | 10 minutes | None | 20, 40 minutes andContinuous | 21 minutes | N/A | SE Note 1 | |
| HousingMaterials andConstruction | Console: ABSplastic | Thermoplastic | ABS | Thermo Plastic | medical gradebiocompatibilityplastics viainjection molding | SE Note 1 | |
| Output Specification | |||||||
| ElementsofComparison | SubjectDevice | Predicate Device | Predicate Device | Predicate Device | PredicateDevice | Remark | |
| Waveform | Pulsed Biphasic | Biphasic | Biphasic | Pulsed Monoilhasic | N/A | SENote 2 | |
| Shape | Rectangular | Rectangular | Rectangular | Modulated Square | N/A | SENote 2 | |
| MaximumOutputVoltage(+/-10%) | 1.23V @ 500Ω3.64V @ 2kΩ10.9V @ 10kΩ | 0.347V @ 500Ω1.242V @ 2kΩ5.780V @ 10kΩ | 1.78V @500Ω6.64V @2KΩ16.2V @10KΩ | 137mV @ 500Ω769mV @ 2kΩ3.82V @ 10kΩ | N/A | SENote 2 | |
| MaximumCurrentDensity | 2.46mA @500Ω1.82mA @ 2kΩ1.09mA @10kΩ | 0.647mA @ 500Ω0.625mA@ 2kΩ0.584mA@ 10kΩ | 3.56mA @500Ω3.32mA @2KΩ1.62mA @10KΩ | 274μΑ @ 500Ω387μA @ 2kΩ382μΑ @ 10kΩ | N/A | SENote 2 | |
| Frequencyrange | 59.3Hz | 0.62 1 - 308.6Hz | 8Hz, 9Hz, 10Hz | 8.40 Hz | N/A | SENote 2 | |
| Pulsewidthrange | 4ms | 3.24-1610ms | N/A | 119ms | N/A | SENote 2 | |
| Pulseduration | 4ms | 3.24-1610ms | ~49-63ms | N/A | N/A | SENote 2 | |
| Net Charge | 9.04μC@ 500Ω | 0 uC @5000 | N/A | N/A | N/A | SENote 2 | |
| MaximumPhaseCharge | 16.4µC@ 500Ω | 190µC@ 500Ω | 157.12uC(8Hz/500Ω) | 23.06µC@ 500Ω | N/A | SENote 2 | |
| MaximumCurrentDensity | Bigger electrodehead:0.044 mA/cm²@500ΩSmallerelectrode head:0.42 mA/cm²@500Ω | 1.486mA/cm²@500Ω | 0.26mA/cm²(8Hz/500Ω) | 0.419mA/cm²@500Ω | N/A | SENote 2 | |
| MaximumPowerDensity | Biggerelectrode head0.014 mW/cm²@ 500ΩSmallerelectrode head0.139 mW/cm²@ 500Ω | 366 µW/cm² @500Ω | 1.33mW/cm²(8Hz/10kΩ) | 3.22 µW/cm² @500Ω | - N/A | SENote 2 | |
| ON time | Constant | 10-30s | Constant | Constant | N/A | SE | |
| OFF time | None | 1-6s | None | None | N/A | SE | |
| Elements ofComparison | SubjectDevice | Predicate Device | Predicate Device | Predicate Device | PredicateDevice | Remark | |
| Contraction andRelaxation time | Adjustable, due to different modes. | Adjustable, due to different modes. | -- | -- | N/A | SE | |
| For LED red light irradiation function | |||||||
| LEDwavelength | 650+/-5nm | -- | -- | -- | 650 +/- 5nm | SE | |
| LED PowerDensity | 80mW/cm² | -- | -- | -- | 80mW/cm² | SE | |
| Additional Features | |||||||
| Environmentfor operating | Temperature: 5 ~ 40° C | + 50 to 1040 F(+10 to +40° C) | -- | -- | -- | SE | |
| Environmentfor storage | Temperature: - 25 ~70° CHumidity: 10 ~90% RH | - 29 to + 1670 F (- 34 to +76° C)0 to 95% - non-condensing | -- | -- | -- | SE | |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5, ISO10993-10 and IEC 62471:2006 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 , ISO10993-10 and IEC 62471:2006 requirements. | SE | |
| ElectricalSafety | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | SE | |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Comparison in Detail(s):
Note 1:
Although the Power Sources, Method of Line Current Isolation, Synchronous or Alternating, Regulated Current or Regulated Voltage, Timer Range, Console weight, Housing Materials and
{10}------------------------------------------------
Construction are a little different from the predicate devices, they are all compliant with requirements of IEC 60601-1, IEC 60601-1-2 and Guidance for Powered Muscle Stimulator. So the differences of the function specifications do not raise any safety or effectiveness issue.
Note 2:
Although the Waveform. Shape, maximum output voltage, maximum current density, frequency range, pulse width range, pulse duration, net charge, maximum phase charge, maximum current density, maximum power density, on time, off time, contraction time of subject device are a little different from the predicate devices, they are all compliant with the requirements of IEC 60601-1, IEC 60601-2-10, and Guidance for Powered Muscle Stimulator. So the differences of function specification do not raise any safety or effectiveness issue.
Conclusion:
The subject device "Micro-current Wrinkle Reduction Facial Service, model EP-400" is Substantial Equivalence to the predicate devices.
§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.
(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).