(227 days)
The GMK® knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- · Severely painful and/or disabled joint as a result of arthritis, theumatoid arthritis or polyarthritis
- · Avascular necrosis of femoral condyle.
- · Post traumatic loss of joint configuration.
- · Primary implantation failure.
Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the femoral components. In case a GMK Revision tibial tray is used, an extension stem must be implanted.
The GMK Sphere Tibial Insert Flex is a line extension to the GMK Sphere Total Knee System and is comprised of the following products:
- Tibial Insert Fixed Flex: Left and Right, Sizes 1-6, 10-20mm (including intermediate ● sizes 11mm and 13mm) UHMWPE (ISO 5834-2)Type 1, Fixation Screw: Ti6Al5V (ISO 5232-3)
The purpose of this submission is to modify the surgical technique provided to surgeons to show that the GMK Sphere Tibial Insert Flex titanium screws' use as optional, instead of required. This technique modification does not alter the intended use or outcomes.
This document describes a 510(k) premarket notification for the GMK® Sphere Tibial Insert Flex. The purpose of this submission is to modify the surgical technique for the device, making the use of titanium screws optional instead of required, and to introduce additional thickness sizes.
Here's an analysis of the provided text in relation to acceptance criteria and supporting studies:
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present a table of acceptance criteria with corresponding reported device performance metrics in the way one might expect for a diagnostic or AI-driven device. Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence to predicate devices and the lack of new safety and effectiveness questions arising from the proposed changes. The performance is assessed through various tests and analyses.
Acceptance Criterion (Implied) Reported Device Performance Biocompatibility "Biocompatibility testing conducted on the predicate devices for the same material supports the biological safety of the GMK Sphere Tibial Insert Flex. Additional testing was deemed unnecessary." Mechanical Performance (with optional screw use) "Based on the risk analysis, static analysis combined with previous static and dynamic/fatigue test on the clipping system, the GMK Sphere Tibial Insert Flex remains safe and consistent for the intended use though the fixation screw is not used." Sterilization Validation One of the analyses conducted or leveraged was "Sterilization Validation". No specific performance outcome is detailed, but the inclusion implies it met established standards. Shelf Life One of the analyses conducted or leveraged was "Shelf Life". No specific performance outcome is detailed, but the inclusion implies it met established standards. The shelf life is stated as 5 years in the comparison table (Page 5). Pyrogenicity (for applicable components, if claimed non-pyrogenic) "Limulus Amebocyte Lysate (LAL) testing was evaluated to establish the device meets pyrogen limit specifications." (Medacta does not intend to claim "non-pyrogenic" for Femoral Distal Augmentations, implying pyrogenicity was evaluated where relevant but not necessarily claimed as part of the primary device's properties). Safety and Effectiveness (overall, for modified features) "The fundamental scientific technology of the modified device has not changed relative to the predicate devices. The safety and effectiveness of the subject device is adequately supported by the substantial equivalence information, materials information, and analysis data provided..." -
Sample Size Used for the Test Set and Data Provenance:
The document describes engineering and laboratory tests rather than clinical studies with human subjects or a "test set" of patient data in the typical sense for an AI/diagnostic device. The tests mentioned include:
- Static and Dynamic/Fatigue Test on Clipping System (K090988): This test was leveraged from a previous submission (K090988). The sample size (e.g., number of components tested) for this mechanical testing is not specified in this document. The provenance is internal laboratory testing.
- Sterilization Validation, Shelf Life, LAL testing: These are standard laboratory tests conducted on device materials or products. The sample sizes for these tests are not specified, but would typically involve a statistically appropriate number of units as per relevant international standards (e.g., ISO, ASTM). The provenance is internal laboratory testing.
There is no mention of country of origin of data or whether it was retrospective or prospective, as it pertains to device testing, not patient data analysis.
-
Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not applicable to this submission. This document pertains to a medical device modification and its mechanical and material properties, not an AI/diagnostic device that requires expert-established ground truth from a test set of patient data. The "ground truth" here is established through engineering principles, material science, and regulatory standards.
-
Adjudication Method for the Test Set:
This is not applicable as there is no "test set" of patient cases requiring adjudication by experts. The determination of device safety and effectiveness is based on engineering analyses and adherence to material and manufacturing standards.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
This is not applicable. This document describes a traditional medical device (knee prosthesis) modification and does not involve AI assistance or human reader performance.
-
Standalone (Algorithm Only) Performance:
This is not applicable. The device is a physical knee prosthesis, not an algorithm.
-
Type of Ground Truth Used:
The "ground truth" in this context is established by:
- Engineering principles and performance standards: For static and dynamic/fatigue testing, the "ground truth" is that the device must withstand certain forces and cycles without failure, as determined by established engineering specifications and benchmarks for similar devices.
- Material properties and biocompatibility standards: For materials, sterilization, and LAL testing, the "ground truth" is adherence to ISO standards (e.g., ISO 5834-2 for UHMWPE, ISO 5832-3/4 for metals) and regulatory limits for pyrogens, ensuring biological safety.
- Substantial equivalence: The ultimate "ground truth" for this 510(k) submission is that the modified device is "as safe and effective as the predicate devices" based on the provided data and analyses.
-
Sample Size for the Training Set:
This is not applicable. There is no "training set" in the context of this traditional medical device submission. The device's design is based on established engineering principles and prior device designs, not machine learning training.
-
How the Ground Truth for the Training Set Was Established:
This is not applicable as there is no training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6. 2017
Medacta International SA % Elizabeth Rose Manager, Regulatory Affairs Mapi Usa, Inc 2343 Alexandria Drive Suite 100 Lexington, Kentucky 40504
Re: K162035
Trade/Device Name: GMK Sphere Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: February 9, 2017 Received: February 10, 2017
Dear Mrs. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K162035
Device Name GMK® Sphere
Indications for Use (Describe)
The GMK® knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- · Severely painful and/or disabled joint as a result of arthritis, theumatoid arthritis or polyarthritis
- · Avascular necrosis of femoral condyle.
- · Post traumatic loss of joint configuration.
- · Primary implantation failure.
Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the femoral components. In case a GMK Revision tibial tray is used, an extension stem must be implanted.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Medacta International. The logo features a blue geometric shape resembling two triangles, followed by the word "Medacta" in blue. A yellow line is underneath the word "Medacta", and the word "International" is in gray below the yellow line. A Swiss flag is on the right side of the word "International".
I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
| Contact Person: | Stefano Baj, Regulatory Affairs Manager |
|---|---|
| Date Prepared: | July 21, 2016 |
| Date Revised: | February 9, 2017 |
II. Device
| Device Proprietary Name: | GMK® Sphere |
|---|---|
| Common or Usual Name: | Tibial Inserts |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi constrained cemented prosthesis. |
| Regulation Number: | 21 CFR 888.3560 |
| Product Code: | JWH |
| Device Classification | 2 |
III. Predicate Device
Substantial equivalence is claimed to the following devices:
GMK Sphere - K121416, Medacta International SA GMK Sphere Extension – K140826, Medacta International SA
IV. Device Description
The GMK Sphere Tibial Insert Flex is a line extension to the GMK Sphere Total Knee System and is comprised of the following products:
- Tibial Insert Fixed Flex: Left and Right, Sizes 1-6, 10-20mm (including intermediate ● sizes 11mm and 13mm) UHMWPE (ISO 5834-2)Type 1, Fixation Screw: Ti6Al5V (ISO 5232-3)
{4}------------------------------------------------
The purpose of this submission is to modify the surgical technique provided to surgeons to show that the GMK Sphere Tibial Insert Flex titanium screws' use as optional, instead of required. This technique modification does not alter the intended use or outcomes.
The following components of the GMK Sphere have been cleared previously under K121416 and K140826 predicate device:
- Femoral Component Left and Right, Sizes 1-7 Co-Cr-Mo (ISO 5832-4 Third Edition . 2014-09-15 Implants for Surgery - Metallic Materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy)
- . Femoral Component Left and Right, Sizes 1+ to 6+ (intermediate sizes) Co-Cr-Mo (ISO 5832-4 Third Edition 2014-09-15 Implants for Surgery - Metallic Materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy)
- . Tibial tray fixed cemented Left and Right, 4 intermediate sizes Co-Cr-Mo (ISO 5832-4 Third Edition 2014-09-15 Implants for Surgery - Metallic Materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy)
- . Tibial Insert Fixed Flex, Left and Right, Sizes 1-6, 10mm -20mm UHMWPE (ISO 5834-2:2011 Implants for Surgery - Ultra-High-Molecular-Weight Polyethylene - Part 2: Moulded Forms) Type 1, Ti6Al5V (ISO 5832-3:1996 Implants for Surgery - Metallic Materials - Part 3: Wrought Titanium 6-Aluminium 4-Vanadium Alloy)
- . Tibial Insert Fixed Flex, Left and Right, Sizes 1-6, 11mm and 13mm UHMWPE (ISO 5834-2:2011 Implants for Surgery - Ultra-High-Molecular-Weight Polyethylene - Part 2: Moulded Forms) Type 1, Ti6Al4V (ISO 5832-3:1996 Implants for Surgery - Metallic Materials - Part 3: Wrought Titanium 6-Aluminium 4-Vanadium Alloy
- Instrumentation
The following components of the GMK Sphere have been cleared previously under the Medacta GMK Total Knee System, K121416:
- Resurfacing patella Sizes 1-4 (K090988 and K113571) ●
- Tibial tray fixed cemented Left and Right, Sizes 1-6 (K090988)
- Primary extension stem Ø 11mm/ L 65mm (K090988) and L 30mm (K133630) ●
V. Indications for Use
The GMK Knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, ● rheumatoid arthritis or polyarthritis
- . Avascular necrosis of femoral condyle.
- Post traumatic loss of joint configuration. ●
- Primary implantation failure. ●
{5}------------------------------------------------
Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision tibial tray is used, an extension stem must be implanted.
VI. Comparison of Technological Characteristics
The GMK Sphere Tibial Insert Flex and the predicate devices share the following characteristics:
- Indications for Use .
- Materials
- Packaging ●
- Device Usage ●
- Shelf Life ●
- Sterilization Method
The GMK Sphere Tibial Insert Flex is technologically different from the predicate devices as follows:
- Thickness sizes available .
- Optional use of fixation screw in the Surgical Technique
Biocompatibility testing conducted on the predicate devices for the same material supports the biological safety of the GMK Sphere Tibial Insert Flex. Additional testing was deemed unnecessary.
A comparison of the subject and predicate devices is provided in the table below.
| GMK Sphere Tibial InsertFlex | GMK Sphere Extension(K140826) | GMK Sphere(K121416) | |
|---|---|---|---|
| Subject Device | |||
| Material ofConstruction | UHMWPE (ISO 5834-2) withTitanium alloy Screw (ISO 5832-3) | Same | Same |
| SterilizationMethod | Ethylene Oxide | Same | Same |
| Device Usage | Single Use | Same | Same |
| Sizes | 6 sizes (1 to 6) | Same | Same |
| Thickness | 7 levels of thickness(10, 11, 12, 13, 14, 17, 20mm) | Same | 5 levels of thickness(10, 12, 14, 17, 20mm) |
| Shelf Life | 5 years | Same | Same |
Technological comparison
{6}------------------------------------------------
Discussion
As seen above, the difference between the subject and predicate devices is that the subject device has additional thicknesses. The main purpose of the submission is to update the surgical technique to make the fixation screw included in the packaging of the GMK Sphere Tibial Insert Flex an optional feature for additional fixation, instead of being required for fixation.
The fundamental scientific technology of the modified device has not changed relative to the predicate devices. The safety and effectiveness of the subject device is adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
Performance Data VII.
A review of the mechanical data on the subject and predicate devices indicates that the modified use of the subject device, usage of fixation screw of GMK Sphere Tibial Insert Flex as optional instead of required, is equivalent to devices currently cleared for use and does not alter the intended surgical outcomes. The purpose of this submission is to update the surgical technique to show that the GMK Sphere Tibial Insert Flex does not require the fixation screws included in the packaging and that they are optional devices for additional fixation. The modification was evaluated by risk analysis to identify any new risks associated with the change. Based on the risk analysis, static analysis combined with previous static and dynamic/fatigue test on the clipping system, the GMK Sphere Tibial Insert Flex remains safe and consistent for the intended use though the fixation screw is not used.
The following analyses were conducted or were leveraged to support the substantial equivalence of the subject devices:
- Static and Dynamic/Fatigue Test on Clipping System (K090988) ●
- Sterilization Validation
- Shelf Life ●
- Limulus Amebocyte Lysate (LAL) testing was evaluated to establish the device meets ● pyrogen limit specifications.
Medacta does not intend to make a claim of "non-pyrogenic" on the device labeling for the Femoral Distal Augmentations.
VIII. Conclusion
The information provided above supports that the GMK Sphere Tibial Insert Flex is as safe and effective as the predicate devices. The main purpose of the submission is to update the surgical technique to make the fixation screw included in the packaging of the GMK Sphere Tibial Insert Flex an optional feature for additional fixation, instead of being required for fixation. The change
{7}------------------------------------------------
to the surgical technique and addition of thicknesses available of the subject device as compared to the predicate devices, do not raise any new questions of safety and effectiveness. Therefore, it is concluded that the GMK Sphere Tibial Insert Flex is substantially equivalent to the predicate devices.
§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.
(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.