K Number
K162001

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-08-31

(42 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 16 Channel Shoulder Coil is to be used in conjunction with Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.

Device Description

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are designed for use with Magnetic Resonance Imaging (MRI) systems. The coil is designed to work in unison with the Body Coil of the MRI system, which will transmit the radio frequency (RF) signals, so that the coil may receive the resultant RF signal from the excited nuclei.

The proposed 1.5T and 3T 16Ch GE Shoulder Coils are identical to each other in construction with the exception that they are tuned to their respective frequencies and additional cable trap baluns are included on the 3T coil due to the shorter wavelengths associated with 3T. The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are designed as receive-only coils for high resolution diagnostic imaging of shoulder. The coils provide unilateral images (Left and Right) of the anatomy of interest.

AI/ML Overview

The provided text is a 510(k) premarket notification for magnetic resonance imaging (MRI) coils (1.5T and 3.0T 16CH GE Shoulder Coils). While it describes the device, its intended use, and its similarities to a predicate device to establish substantial equivalence, it does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI-powered diagnostic device.

This document is for a physical medical device (an MRI coil), and its "performance" is evaluated through non-clinical bench testing (SNR, image uniformity) and a clinical demonstration focusing on image quality for interpretation by a physician. It is not a document for an AI/ML-based diagnostic device where "acceptance criteria" would typically refer to metrics like sensitivity, specificity, or AUC, and "ground truth" would be established by expert consensus or biopsy.

Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving the device (in the context of an AI diagnostic tool) meets those criteria, as the provided text describes a physical MRI coil and its FDA clearance process.

Here's why each point you requested is not applicable or not explicitly detailed in the provided text for an AI/ML context:

  1. Table of acceptance criteria and reported device performance: The document states "Non-clinical verification and validation test results demonstrate that the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils... Meets the acceptance criteria and is adequate for its intended use." However, it does not provide a specific table of quantitative acceptance criteria for diagnostic performance (like sensitivity/specificity thresholds) or the numerical results against these criteria. It mentions tests for SNR and image uniformity and that a clinical radiologist reviewed images to confirm adequate quality, but no numerical thresholds or results are given.
  2. Sample size for the test set and data provenance: The document mentions "clinical demonstrations" and "Clinical Images review by Radiologist" but does not specify the sample size of these clinical demonstrations (number of patients/cases) or the data provenance (country, retrospective/prospective).
  3. Number of experts and qualifications for ground truth: "a clinical radiologist" is mentioned as reviewing the DICOM images for adequacy. The number of radiologists is not specified (e.g., singular "a radiologist" could imply one, but is not definitive for a study), nor are their specific qualifications (e.g., years of experience). For an AI device, this would typically involve multiple, highly qualified experts.
  4. Adjudication method: Not applicable/not mentioned. Ground truth for an AI device involves multiple readers and an adjudication process. Here, it's a "clinical radiologist" reviewing images for overall adequacy, not to establish a precise diagnostic ground truth for an AI algorithm.
  5. MRMC comparative effectiveness study: Not done. This is not an AI-assisted device.
  6. Standalone performance: The performance mentioned (SNR, uniformity) is inherent to the coil's function, not a diagnostic decision made by an algorithm.
  7. Type of ground truth used: For image quality, it was radiologist review of image adequacy ("confirm the image quality is adequate"). This is different from establishing a diagnostic ground truth (e.g., presence/absence of disease confirmed by pathology) for an AI algorithm.
  8. Sample size for the training set: Not applicable. This is not an AI/ML device that requires training data.
  9. How ground truth for training set was established: Not applicable.

In summary, the provided document details the regulatory pathway for a conventional MRI coil, not an AI-powered diagnostic device. The concepts of "acceptance criteria" and "study" in the context of AI diagnostic performance metrics are not applicable to this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol composed of three curved lines, representing the department's mission of protecting the health of all Americans and providing essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 31, 2016

INVIVO CORPORATION C/O MARK JOB REGULATORY TECHNOIOGY SERVICES LLC 1394 25TH STREET, NW BUFFALO MN 55313

Re: K162001

Trade/Device Name: 1.5T and 3.0T 16CH GE Shoulder Coils Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: August 18, 2016 Received: August 19, 2016

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara

For

Robert Ochs. PhD Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162001

Device Name

1.5T and 3.0T 16CH GE Shoulder Coils

Indications for Use (Describe)

The 16 Channel Shoulder Coil is to be used in conjunction with Magnetic Resonance Scanners to produce diagnostic images of the shoulder that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitted by:Invivo Corporation3545 SW 47th AveGainesville, FL 32608
Establishment Name:Invivo Corporation
EstablishmentRegistration Number:1056069
Contact Person:Ken RevennaughTitle: Director of Quality and RegulatoryInvivo Corporation3545 SW 47th AveGainesville, FL 32608Phone: (352) 384-8590E-mail: ken.revennaugh@philips.com
Date Prepared:August 15, 2016
Trade Name:1.5T and 3.0T 16 Ch GE Shoulder Coils
Common Name:Magnetic resonance diagnostic device
Classification Name:Coil, Magnetic Resonance, Specialty
ClassificationRegulation Number:892.1000
Classification:Class II
Classification Panel:Radiology
Product code:MOS

{4}------------------------------------------------

Device Description:

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are designed for use with Magnetic Resonance Imaging (MRI) systems. The coil is designed to work in unison with the Body Coil of the MRI system, which will transmit the radio frequency (RF) signals, so that the coil may receive the resultant RF signal from the excited nuclei.

The proposed 1.5T and 3T 16Ch GE Shoulder Coils are identical to each other in construction with the exception that they are tuned to their respective frequencies and additional cable trap baluns are included on the 3T coil due to the shorter wavelengths associated with 3T. The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are designed as receive-only coils for high resolution diagnostic imaging of shoulder. The coils provide unilateral images (Left and Right) of the anatomy of interest.

Variations between the proposed 1.5T and 3T 16Ch GE Shoulder Coils and the predicate devices HRS-63-8 and HRS-127-8 Shoulder Array Coils include MR system compatibility and a 16-channel connection to the MRI system. Mechanical design changes include modifications to the connector to permit the 16-channel connection and the specified MR system compatibility. In addition, modifications to the housing from three-part adjustable to two-part adjustable design improves workflow as well as meets customer requirements.

Proposed changes to the device

  • . Coil housing material changed from Lexan 950 to Lexan 925A for ease of availability with no impact to device safety and effectiveness
  • o Coil pad coating changed from Guardian MPGX to Polyscan coating for ease of availability with no impact to device safety or effectiveness.
  • Channel count was increased from 8 channels to 16 channels to meet the customer/market needs with no impact on device to safety and effectiveness.
  • The overall coil dimensions were changed from 9.1" x 6.2" x 11.7" to 11" x 11" x 12.25" to accommodate more channels with no impact to device safety and effectiveness.
  • The housing design was changed from a three-part adjustable housing to a two-part o adjustable housing to improve ease of use and workflow with no impact to device safety and effectiveness.
  • o The GE OEM connector was updated from an ODU Bendix connector to an ODU P port connector to accommodate specified MRI Scanners with no impact to device safety and effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Invivo. The logo consists of a circle with the number 1 inside, followed by the word "Invivo" in a stylized font. The color of the logo is purple. There is a trademark symbol to the right of the word Invivo.

Photograph:

Image /page/5/Picture/2 description: The image shows a CAD model of a medical device or equipment. The device has a gray rectangular base with a clear plastic component on top. The text at the bottom of the image says "side iso view w pods", which suggests that the image is a side isometric view of the device with pods.

{6}------------------------------------------------

System Compatibility:

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are designed to be used with GE 1.5T and 3.0T MRI Systems respectively. The Models 1.5T and 3.0T 16 Ch GE Shoulder Coils are 16-channel receive-only coils. Models 1.5T and 3.0T 16 Ch GE Shoulder Coils include identical coils with slight variations between system connectors to allow compatibility with respective MRI Scanners.

The Model 1.5T 16 Ch GE Shoulder Coil will operate with the GE 1.5T MRI Scanners such as GE Signa Voyager MRI Scanner and the Model 3.0T 16 Ch GE Shoulder Coil will operate with GE 3.0T MRI Scanners such as GE Signa Pioneer MRI Scanner.

Patient Safety

Description of design that assures that the coil will remain decoupled:

The coil incorporates a redundant decoupling strategy. Back-to-back fast recovery PIN diodes or Dual diodes are placed in series with an inductor across a capacitor. The inductor is tuned to shift the resonant frequency of the coil. During system transmitted pulses, a D.C. voltage is supplied to each channel of the array by the MRI system. This voltage drives the PIN diodes or Dual diodes they conduct, thus engaging the decoupling circuit. The coil resonant frequency then shifts away from the system resonant frequency. The coil therefore becomes a very poor receiver of RF energy at the system operating frequency during the transmit pulses.

In the event that the coil is placed in the magnet and the operator neglects to plug in the coil, the PIN diodes or Dual diodes function as a passive decoupling system. In this case, the decoupling takes place without the need for the voltage supplied by the MRI system.

Description of electrical isolation of the patient from the surface coil electrical conductors:

The electrical conductors are etched on FR4 printed circuit board. The circuit boards are encased in plastic housing. The plastic housing wall and/or foam is provided with a minimum of .10" thickness. At no time it is possible for the coil conductors to touch the patient.

The cable exits the coil superior to the patient's head to connect to the MRI system connector. The cables are of such a length to reach the MRI system coil part by running the cables away from patient. The cable is designed such that the cable is difficult to make the loop shape during use. The cables have insulation jackets and additional pad that increase separation of the cable if it did somehow become looped.

Indications for Use:

The 16 Channel Shoulder Coil is to be used in conjunction with a Magnetic Resonance Scanner to produce diagnostic images of the shoulder that can be interpreted by a trained physician.

{7}------------------------------------------------

Predicate Device Information:

Predicate Device NamePredicate 510(k)Submission Reference
HRS-63-8 and HRS-127-8 Shoulder Array CoilsK053017

Safety and Effectiveness:

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils labeling contain instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via Invivo Risk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via the product development process, verification and validation testing and safety features provided by the MRI Scanners.

Technological Characteristics:

The technological characteristics of the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are exactly the same as the predicate devices HRS-63-8 and HRS-127-8 Shoulder Array Coils.

The technological characteristic of a receive-only radio frequency (RF) coil is that the coil receives radio frequency signals from the tissue of interest. This technological characteristic of the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils has not changed as compared to the technological characteristic of the predicate devices HRS-63-8 and HRS-127-8 Shoulder Array Coils.

Based on the information provided above, the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are considered substantially equivalent to the predicate devices HRS-63-8 and HRS-127-8 Shoulder Array Coils (K053017) in terms of technological characteristics.

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils utilize the same technological characteristic as the predicate devices. Both have similar design, intended use, patient safety features, housing materials, overall dimensions, and operating principals; and substantially equivalent in their performance and effectiveness for their intended uses.

{8}------------------------------------------------

Following is the comparison of technological characteristics between the Predicate and the proposed devices:

Table-01Comparison of the Technological Characteristics of the Predicate Device and the Proposed Device
Device FeaturesPredicate DeviceK053017HRS-63-8 and HRS-127-8 Shoulder Array CoilsProposed Device1.5T and 3.0T 16 ChGE Shoulder CoilsComments
Indications for UseThe coil is to be used inconjunction with aMagnetic ResonanceScanner to producediagnostic images of theshoulder that can beinterpreted by a trainedphysicianThe 16 Channelshoulder coil is to beused in conjunctionwith a MagneticResonance Scanner toproduce diagnosticimages of the shoulderthat can be interpretedby a trained physicianSimilar, changedfrom a coil tospecific 16 ChannelShoulder Coil.No Change in theIntended Use of theCoil
Coil DimensionsLength x Width xHeight9.1 x 6.2 x 11.70 inches11 x 11 x 12.25 inchesThe change indimensions is toaccommodate morechannels. No impactto device safety andeffectiveness.
Coil Frequency63.87 MHz for 1.5T(or 127.73 MHz for 3.0T)IdenticalNo Change
Coil DesignReceive-only Phased ArrayIdenticalNo Change
Magnetic FieldOrientation (B0)HorizontalIdenticalNo Change
Coil Geometry/Housing DesignThree-part adjustablehousingTwo-part adjustablehousingThe change inhousing design wasto improve ease ofuse and workflow.No impact to devicesafety andeffectiveness.
Housing Thickness.10 inch insulationminimumIdenticalNo Change
Decoupling MethodLC Tank CircuitIdenticalNo Change

{9}------------------------------------------------

Table-01Comparison of the Technological Characteristics of the Predicate Device and the Proposed Device
Device FeaturesPredicate DeviceK053017HRS-63-8 and HRS-127-8 Shoulder Array CoilsProposed Device1.5T and 3.0T 16 ChGE Shoulder CoilsComments
Number ofChannels/Preamplifiers8 channels / 8preamplifiers16 channels / 16preamplifiersThe number ofchannels wasincreased to meet thecustomer and marketneeds. No impact todevice safety andeffectiveness.
Housing MaterialLexan 950Lexan 925ABoth are similarmaterials and havebeen used in currentproducts. No impactto device safety andeffectiveness.
Base PadUrethane Foam withGuardian MPGX coatingUrethane Foam withPolyscan CoatingBoth are similarmaterials and havebeen used in currentproducts. No impactto device safety andeffectiveness.
System Connector /CompatibilityODU Bendix Connectorfor GE 1.5T/3.0T ScannerODU P Port Connectorfor GE 1.5T/3.0TScannerTo be compatiblewith the specifiedMRI Scanners. Noimpact to devicesafety &effectiveness.

Non-Clinical testing and Performance Information:

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils comply with the following international and FDA-recognized consensus standards:

  • IEC 60601-1, 3rd edition ●
  • IEC 60601-2-33, 3rd edition ●
  • ISO 14971, 2nd edition ●
  • ISO 10993-1 ●
  • Device specific draft guidance document, entitled "Submission of Premarket ● Notifications for Magnetic Resonance Diagnostic Devices - July 14, 2015"

{10}------------------------------------------------

Bench performance testing includes both non-clinical testing and clinical demonstrations of the representative anatomic region (shoulder imaging) using FDA-cleared GE 1.5T and 3.0T scanners. The non-clinical testing characterizes the Signal-to-Noise Ratio (SNR), Image Uniformity and special purpose of the RF receive-only coil. The clinical demonstration DICOM images were reviewed by a clinical radiologist to confirm the image quality is adequate.

Non-clinical verification and validation test results demonstrate that the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils

  • Comply with the device specifications, identified international and FDA-recognized . consensus standards, and device specific guidance document entitled "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices"
  • Meets the acceptance criteria and is adequate for its intended use. .

Therefore, the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are as safe and effective as the predicate devices (K053017) in terms of safety and effectiveness.

Summary of Clinical Data:

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils do not require clinical study since substantial equivalence to the predicate device was demonstrated with the following attributes:

  • Design features:
  • Indication for use;
  • Technological characteristic;
  • Non-clinical performance testing, including Clinical Images review by Radiologist, and ●
  • Device Safety and effectiveness

Conclusion:

The proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils are substantially equivalent to the predicate devices HRS-63-8 and HRS-127-8 Shoulder Array Coils (K053017). The substantial equivalency is based on design features, principles of operation, technological characteristic, indications for use, magnet system compatibility, clinical image review, and safety and effectiveness of the proposed devices. Additionally, the non-clinical performance (verification and validation) tests, which complied with the device specifications, and requirements specified in the international and FDA-recognized consensus standards demonstrated that the proposed 1.5T and 3.0T 16 Ch GE Shoulder Coils meet the acceptance criteria and are adequate for the established intended use. The electrical isolation methods, decoupling method and housing material are similar for the predicate and proposed devices.

The results of both non-clinical and device performance tests demonstrate that the proposed device is as safe, as effective, and performs as well as the predicate device.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.