(115 days)
The UltraShape System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) for lipolysis (breakdown of fat) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference and for use on the flanks.
The UltraShape System is comprised of multiple components, including the control unit and two ultrasonic transducers. The UltraShape System selectively targets subcutaneous adipose tissue via focused ultrasound for the purpose of non-invasive aesthetic body contouring. The transducers are electro-mechanical devices that convert an electrical signal into mechanical (acoustical) energy. The operating parameters of the UltraShape System achieve selective disruption of adipose tissue without damaging neighboring tissues such as blood vessels, nerves, or muscle. The UltraShape System has two treatment modes available, the Single Focus Deep Mode and the Super Mode.
The Synergy UltraShape System, a focused ultrasound stimulator for aesthetic use, was evaluated for safety and effectiveness in non-invasive lipolysis of the flanks. The device delivers focused ultrasound energy to disrupt subcutaneous adipose tissue (SAT) for fat breakdown.
Here's a breakdown of the acceptance criteria and the study proving it meets those criteria:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Primary Endpoint) | Reported Device Performance | Met? |
|---|---|---|
| Fat reduction in flanks evaluated by visual assessment of randomized photographs of before (baseline) versus after (16 weeks following last treatment), as well as treated versus control flanks, by 2 blinded reviewers. | Primary: Met endpoint; blinded reviewers identified correctly (by agreement between the blinded reviewers) both the pre-/post-treatment photographs as well as the treated/un-treated flanks in 80% (36 of 45) subjects. (Sensitivity analysis: If one subject with poor photo quality, for whom fat reduction was observed, was imputed as a failure, the success rate would be 78%, which would not meet the 80% threshold. However, this subject was excluded from the primary endpoint assessment due to photograph quality, and the presented result is based on 45 subjects.) | Yes |
| Secondary Effectiveness Endpoints: | Secondary: | |
| * Fat thickness reduction on the treated flank compared to the control flank, measured by ultrasound. | * Fat thickness reduction results following UltraShape treatment were statistically significantly greater compared to controls as measured by ultrasound. | Yes |
| * Fat thickness reduction on the treated flank compared to the control flank, measured by caliper. | * Fat thickness reduction results following UltraShape treatment were greater compared to controls as measured by caliper (reaching statistical significance in absolute fat thickness reduction at 16 week follow up). | Yes |
| * Investigator satisfaction assessment. | * Investigators were satisfied with results following UltraShape treatment for 70% of the subjects at 4 week follow up, 53% of the subjects at 8 week follow up, and 43% of the subjects at 16 week follow up. | N/A |
| * Subject satisfaction self-assessment. | * At 4, 8 and 16 weeks following end of treatments, 43%, 47%, and 41% of the subjects were satisfied or very satisfied. | N/A |
| * Subjects' comfort level assessed immediately after each treatment. | * Subjects reported no to minimal pain for each of the three treatments. | N/A |
| Safety Results: No reported adverse events and anticipated immediate responses (mild erythema, edema) resolved completely within days without intervention. | No adverse events were reported after 141 treatment sessions. The only immediate responses recorded were mild and resolved completely within days without any intervention. | Yes |
2. Sample Size for the Test Set and Data Provenance
- Sample Size: 48 subjects were enrolled and treated in the study, with 46 subjects completing the study. The primary endpoint assessment was based on 45 subjects due to the poor quality of one post-treatment photograph.
- Data Provenance: The study was a prospective, single-arm, self-controlled, multicenter clinical study conducted at 3 U.S. sites. The data is thus prospective and from the U.S.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: 2 blinded reviewers.
- Qualifications: The document does not explicitly state the specific qualifications (e.g., radiologist, years of experience) of the blinded reviewers. It only specifies that they were "blinded reviewers."
4. Adjudication Method for the Test Set
- The primary endpoint was met "by agreement between the blinded reviewers." This implies a consensus-based adjudication, specifically that both reviewers had to agree on the correct identification of pre/post-treatment and treated/untreated flanks.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly stated as being performed to evaluate human reader improvement with or without AI assistance. This study focused on the device's efficacy in achieving fat reduction through visual assessment and physical measurements, not the performance of human readers assisted by AI.
6. Standalone (Algorithm Only) Performance
- This study did not focus on standalone algorithm performance. The UltraShape System is a device for physical treatment and relies on visual and measurement assessments by human observers (investigators, subjects, blinded reviewers) and physical measurement tools (ultrasound, caliper) to determine its effectiveness. While the device has software components for control and guidance, the performance evaluation is for the device's physical effect on the body, not a diagnostic algorithm.
7. Type of Ground Truth Used
- The ground truth for the primary endpoint was established by expert consensus (agreement between 2 blinded reviewers) on photographic assessments.
- For secondary endpoints, objective measurements (ultrasound, caliper) were used to quantify fat thickness reduction, and subjective assessments (investigator and subject satisfaction, comfort level) were collected.
8. Sample Size for the Training Set
- The document refers to "several separate prospective studies" for clinical evaluation and states that nonclinical performance testing was "previously conducted" and "In vivo testing in an animal model was performed." However, it does not explicitly state a separate "training set" for the UltraShape System's clinical evaluation in the context of this 510(k) submission. The clinical data presented here is the definitive study for the device's substantial equivalence to the predicate. The UltraShape System itself is a physical device, and its "training" might refer to development phases or prior versions, but not a dataset for an AI algorithm in the way it's typically understood. The presented clinical study is the test set for this specific submission's claims.
9. How the Ground Truth for the Training Set Was Established
- As noted above, a distinct "training set" for the clinical effectiveness of this specific device version isn't detailed in the same manner as an AI algorithm's training data. The device's development and prior clearances (K141708, K133238, K160896) would have involved extensive R&D and testing, where engineers and clinicians established desired performance characteristics. The nonclinical testing (biocompatibility, beam profile, acoustic power, thermal measurements, software verification, electrical safety, in vivo animal testing) described in the document served to validate the device's core functionality and safety prior to humanitarian clinical trials. The "ground truth" for these aspects would be derived from laboratory measurements, adherence to standards, and observation of physical effects.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 7, 2016
Syneron Medical Ltd. % Ms. Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K161952
Trade/Device Name: UltraShape System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System for Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: October 3, 2016 Received: October 3, 2016
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
UltraShape System
Indications for Use (Describe)
The UltraShape System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) for lipolysis (breakdown of fat) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference and for use on the flanks.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
Syneron Medical Ltd.'s UltraShape System
- Submitted by: Syneron Medical Ltd. P.O.B. 550 Industrial Zone Tavor Building Yokneam Illit, 20692 Israel Contact Person: Ruthie Amir, MD, Global Vice President of Clinical, Regulatory and Education Syneron Medical Ltd. P.O.B. 550 Industrial Zone Tavor Building Yokneam Illit, 20692 Israel Phone: 972-73-244-2200 Facsimile: 972-73-244-2202
Submission Correspondent:
Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103 Phone: 267 675 4600 Facsimile: 267 675 4601
Date Prepared: November 3, 2016
Name of Device
Syneron UltraShape System
Common or Usual Name
Focused Ultrasound Stimulator System for Aesthetic Use
Classification
Focused Ultrasound Stimulator System for Aesthetic Use
21 CFR 878.4590, Class II, product code OHV
Predicate Device
Syneron UltraShape (K141708)
Intended Use / Indications for Use
The UltraShape System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) for lipolysis (breakdown of fat) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference and for use on the flanks.
{4}------------------------------------------------
Device Description
The UltraShape System is comprised of multiple components, including the control unit and two ultrasonic transducers. The UltraShape System selectively targets subcutaneous adipose tissue via focused ultrasound for the purpose of non-invasive aesthetic body contouring. The transducers are electro-mechanical devices that convert an electrical signal into mechanical (acoustical) energy. The operating parameters of the UltraShape System achieve selective disruption of adipose tissue without damaging neighboring tissues such as blood vessels, nerves, or muscle. The UltraShape System has two treatment modes available, the Single Focus Deep Mode and the Super Mode.
Technological Characteristics
The UltraShape System has very similar technological characteristics compared to its predicate. The UltraShape and the predicate are both comprised of the system console, including the computer, and two ultrasonic transducers deliver the focused ultrasound energy beam to the targeted treatment area, and real-time optical and acoustic feedback (optional) on the treatment is provided via the tracking and guidance system. With both the UltraShape and its predicate, the transducers' functionality is based on the piezoelectric effect implemented with the ceramic element.
In addition, the UltraShape System has the same treatment parameters as the previously cleared UltraShape device, including the same frequency, burst duration, and user interface. In addition, the power intensity levels were previously cleared for the UltraShape. The focal distribution of the energy beams delivered to the treatment area is consistent between the UltraShape and its predicate, including the focal depth, diameter, and length. Software changes were also made to the updated version of the UltraShape. Notably, the UltraShape System has been evaluated in clinical, in vivo, and bench testing, and none of the changes raise new types of safety or effectiveness questions compared to the predicate.
Therefore, the UltraShape System has similar technological characteristics compared to its predicate.
Performance Data
Nonclinical Performance Testing. The following nonclinical performance testing was previously conducted to support the substantial equivalence of the UltraShape System to the predicate device, consistent with FDA's "Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" (2011). In all instances, the UltraShape System functioned as intended.
- . Biocompatibility testing in accordance with ISO 10993 for skin irritation, sensitization, cytotoxicity testing supported the biocompatibility of the patient-contacting components of the device.
- Beam profile testing demonstrated that the acoustic energy is delivered and concentrated in the desired target location.
{5}------------------------------------------------
- . Acoustic power testing demonstrated that the acoustic power of the transducers is highly predictable with low variability.
- . In vitro acoustic and thermal measurements demonstrated the safety of non-targeted tissues both proximal and distal to the targeted region.
- . Software verification and validation was performed, and demonstrated that the software performs as intended.
- . Electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2) and electromagnetic immunity testing was conducted and results were passing.
- In vivo testing in an animal model was performed which demonstrated the treatment effects of the UltraShape, and supported its safety and efficacy profile for the intended use.
Clinical Data: In addition, clinical evaluation of the intended population was performed in several separate prospective studies, including a single-arm, prospective, selfcontrolled study to evaluate the safety and effectiveness of the UltraShape device for noninvasive lipolysis of the flanks. A total of 48 subjects were enrolled and treated in the study (of which 46 subjects completed the study) at 3 U.S. sites. The study included females (83%) and males (17%) across a range of ages, races, ethnicities, and skin types. The mean age was 45 years and the majority of the subjects were Caucasian. The baseline mean weight was 68±9 kg and mean BMI was 25.28±2.08 kg/m².
Each subject received 3 biweekly treatments on one randomized flank, while the other flank remained un-treated throughout the study. Subject follow up was conducted at 4 weeks, 8 weeks, and 16 weeks after the last treatment. The study included visual assessment of the flanks by 2 blinded reviewers as well as fat thickness measurements of the post treatment flanks compared to baseline measurements.
Study results demonstrated that the UltraShape treatment at the flanks was accompanied with no or minimal discomfort, consistent with results observed for the prior UltraShape clearances. Throughout the study, no adverse events were reported, and the anticipated immediate responses after treatment (e.g., mild erythema, mild edema) were consistent with those observed with other similar devices, such as the previously cleared UltraShape systems (K133238, K160896) and Cynosure's SculpSure (K150230). The responses were mild and resolved completely within days without any intervention.
The primary endpoint was met, where 80% of treated versus control flanks and pre versus post treatment photographs were correctly identified by blinded reviewers. One subject was not included in the primary endpoint blinded review assessment because of the poor quality of the post-treatment photograph for this subject. Although results for this subject showed fat reduction at the treated flank by caliper (-1.0 mm ( 2.38%)) and by ultrasound (-0.55 mm (-2.40%)), if this subject was imputed as a failure in a sensitivity analysis of the primary endpoint, the study success rate would be 78%, which would not meet the primary endpoint threshold of 80%.
In addition, secondary effectiveness endpoints demonstrated that fat thickness reduction on the treated flank was statistically significantly greater compared to the control flank at each
{6}------------------------------------------------
follow up visit as measured by ultrasound. Fat thickness reduction results following UltraShape treatment were greater compared to controls as measured by caliper (reaching statistical significance in absolute fat thickness reduction at 16 week follow up). At 4 week follow up, investigator satisfaction measured 70%, and subsequent results reflected lower satisfaction with increasing time from treatment (53% at 8 weeks follow up, 43% at 16 weeks follow up). In terms of subject satisfaction, at 4, 8 and 16 weeks following end of treatments, 43%, 47%, and 41% subjects were satisfied or very satisfied. The study design and results are further summarized in the table below.
| Study Design | Prospective, single-arm, self-controlled, multicenter clinical study |
|---|---|
| Sample size | 48 subjects at 3 sites were enrolled and treated in the study |
| PrincipalEligibilityCriteria | Fat thickness of at least 1.5 cm in the treated area as measured by calibrated caliper. BMI interval: $22 \le BMI \le 30$ (normal to overweight, but not obese). General good health confirmed by medical history and skin examination of the treated area. |
| Follow upintervals | 3 treatment visits and follow up visits at 4 weeks, 8 weeks, and 16 weeks. |
| Endpoints | Primary: Fat reduction in flanks was evaluated by visual assessment of randomized photographs of before (baseline) versus after (16 weeks following last treatment), as well as treated versus control flanks, by 2 blinded reviewers. The primary endpoint was achieved when at least 80% of treated versus control flanks and pre- versus post-treatment photographs were correctly identified.Secondary: Fat thickness reduction on the treated flank compared to the control flank, measured by ultrasound. Fat thickness reduction on the treated flank compared to the control flank, measured by caliper. Investigator satisfaction assessment. Subject satisfaction self-assessment. Subjects' comfort level assessed immediately after each treatment. |
| EffectivenessResults | Primary: Met endpoint; blinded reviewers identified correctly (by agreement between the blinded reviewers) both the pre-/post-treatment photographs as well as the treated/un-treated flanks in 80% (36 of 45) subjects.Secondary: Fat thickness reduction results following UltraShape treatment were statistically significantly greater compared to controls as measured by ultrasound. Fat thickness reduction results following UltraShape treatment were greater compared to controls as measured by caliper (reaching statistical significance in absolute fat thickness reduction at 16 week follow up). Investigators were satisfied with results following UltraShape treatment for 70% of the subjects at 4 week follow up, 53% of the subjects at 8 week follow up, and 43% of the subjects at 16 week follow up. At 4, 8 and 16 weeks following end of treatments, 43%, 47%, and 41% of the subjects were satisfied or very satisfied. Subjects reported no to minimal pain for each of the three treatments. |
| SafetyResults | No adverse events were reported after 141 treatment sessions were conducted. The only immediate responses recorded were mild and resolved completely |
| within days without any intervention. |
{7}------------------------------------------------
Therefore, clinical evaluation of the UltraShape device demonstrated the favorable performance and safety profile of the device for lipolysis of the flanks. Results thus further support substantial equivalence of the device compared to the predicate.
Substantial Equivalence
The UltraShape has the same intended use and similar indications for use, technological characteristics and principles of operation as its predicate device. The technological differences between the UltraShape and its predicate mainly consist of minor improvements to the device to facilitate use. Nonclinical and clinical studies of the UltraShape have demonstrated the safety and effectiveness profile of the UltraShape in the intended population. Thus, the UltraShape is substantially equivalent to the predicate device.
Conclusion
Syneron's UltraShape System is a Focused Ultrasound Stimulator System for Aesthetic Use Class II device that has been evaluated in nonclinical testing in accordance with FDA's Special Controls Guidance Document. Testing demonstrated that the device performs as intended. The UltraShape device is substantially equivalent to its predicate device (K141708).
§ 878.4590 Focused ultrasound stimulator system for aesthetic use.
(a)
Identification. A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.