(59 days)
The Trojan XOXO Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
The Trojan™ XOXO™ Intense Ribbed Natural Rubber Latex Male Condom with Silicone and Aloe Vera OE Lubricant [Trojan™ XOXO™ Intense Ribbed condom] is a natural rubber latex sheath, which completely covers the penis with a fitted membrane. A silicone base lubricant containing Aloe Vera OE is applied directly to the condom. The condom is a bulbous reservoir end shape and relies on ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male condom) and ISO 4074:2002 Natural Rubber Latex Condoms - Requirements and test methods. The condom has ribs on the shaft and bulb. The Trojan™ XOXO™ Intense Ribbed condom has a nominal length of 185 ± 10 mm, a nominal thickness 0.09 ± .01mm, nominal flat-width of 54 ± 2 mm measured 30 mm from the open end and 65±2 mm bulb flat width. The Trojan™ XOXO™ Intense Ribbed condom is individually packaged into a foil container. Multiple individually foiled condoms may be packaged in a carton.
1. Table of Acceptance Criteria and Reported Device Performance
| Test Performed | Acceptance Criteria (Specification) | Reported Device Performance |
|---|---|---|
| Biocompatibility: Cytotoxicity ISO Elution Method | Adherence to ISO 10993-5 (2009) | Passed |
| Biocompatibility: Acute Systemic Toxicity | Adherence to ISO 10993-11 (2006) | Passed |
| Biocompatibility: Primary Rabbit Skin Irritation | Adherence to ISO 10993-10 (2010) | Passed |
| Biocompatibility: Rabbit Vaginal Irritation | Adherence to ISO 10993-10 (2010) | Passed |
| Biocompatibility: Rabbit Penile Irritation | Adherence to ISO 10993-10 (2010) | Passed |
| Biocompatibility: Guinea Pig Maximum Sensitization | Adherence to ISO 10993-10 (2010) | Passed |
| Physical Testing | Met specifications of ASTM D3492-08 and ISO 4074 | Met specifications for all three tested lots |
| Shelf Life (Accelerated Stability Study) | Provide a 3-year (36 month) shelf-life | Found to have a 3-year (36 month) shelf-life |
| Shelf Life (Real-time Stability Study) | Verify 5-year (60 month) expiration dating (ongoing) | Will be verified according to 21 CFR 801.435 (ongoing) |
2. Sample size used for the test set and the data provenance
- Biocompatibility Testing: The specific sample sizes for each animal study (rabbit skin, vaginal, penile irritation, guinea pig sensitization) are not explicitly stated in the provided document, but implied by the ISO standards. The tests are prospective as they are performed on the new device according to established protocols.
- Physical Testing: "Three (3) lots" of the Trojan™ XOXO™ Intense Ribbed condom were tested. The data provenance is internal to the manufacturer's testing process. The document does not specify the country of origin of the data, but the testing conforms to international (ISO) and US (ASTM) standards.
- Shelf-Life Testing: The accelerated stability study was conducted on a sample of the product. The real-time stability study is ongoing. The specific sample sizes are not detailed in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device (condom) and the assessment relies on standardized physical and biological performance tests, not on expert clinical interpretation of images or data. The "ground truth" is defined by the passing criteria of the specified international and national standards (ISO and ASTM).
4. Adjudication method for the test set
Not applicable. The tests are objective and based on measurable parameters defined by the ISO and ASTM standards. There is no subjective interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic or interpretive device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithmic device.
7. The type of ground truth used
The ground truth for the device's performance is established by regulatory and industry standards. Specifically:
- Biocompatibility: ISO 10993 series standards.
- Physical Testing: ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms) and ISO 4074: Natural Rubber Latex Condoms Requirements and Test Methods.
- Shelf-Life: 21 CFR 801.435.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As noted above, there is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2016
Church & Dwight Co., Inc. Lisa Burns Manager, Regulatory Affairs 500 Charles Ewing Blvd Ewing, New Jersey 08628
Re: K161915
Trade/Device Name: Trojan™ XOXOTM Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: HIS Dated: July 11, 2016 Received: July 12, 2016
Dear Lisa Burns:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
For Division
Douglas Silverstein -S 2016.09.09 16:17:20 -04'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K161915
Device Name
Trojan(TM) XOXO(TM) Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant
Indications for Use (Describe)
The Trojan XOXO Intense Ribbed Male Natural Rubber Latex Condom with Silicone and Aloe Vera OE Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
II. 510(k) Summary K161915
| Submitter Name: | Church & Dwight Co., Inc. |
|---|---|
| Submitter Address: | 500 Charles Ewing BoulevardEwing, NJ 08628 |
| Contact Person: | Lisa BurnsRegulatory Affairs ManagerChurch & Dwight Co., Inc.469 North Harrison StreetPrinceton, NJ 08543Tel: (609) 806.1997Fax: (609) 403.7411 |
| Date Prepared: | September 9, 2016 |
| Device Trade Name: | Trojan™ XOXO™ Intense Ribbed Male Natural RubberLatex Condom with Silicone and Aloe Vera OE Lubricant |
| Device Common Name: | Natural Rubber Latex condom with Lubricant |
| Product Code: | HIS - Condom (21 CFR § 884.5300) |
| Classification: | Class II |
| Predicate Device: | Trojan™ Jaguar Male Natural Rubber Latex Condom withWarming and Tingling Lubricant |
Device Description:
The Trojan™ XOXO™ Intense Ribbed Natural Rubber Latex Male Condom with Silicone and Aloe Vera OE Lubricant [Trojan™ XOXO™ Intense Ribbed condom] is a natural rubber latex sheath, which completely covers the penis with a fitted membrane. A silicone base lubricant containing Aloe Vera OE is applied directly to the condom. The condom is a bulbous reservoir end shape and relies on ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male condom) and ISO 4074:2002 Natural Rubber Latex Condoms - Requirements and test methods. The condom has ribs on the shaft and bulb. The Trojan™ XOXO™ Intense Ribbed condom has a nominal length of 185 ± 10 mm, a nominal thickness 0.09 ± .01mm, nominal flat-width of 54 ± 2 mm measured 30 mm from the open end and 65±2 mm bulb flat width.
{4}------------------------------------------------
The Trojan™ XOXO™ Intense Ribbed condom is individually packaged into a foil container. Multiple individually foiled condoms may be packaged in a carton.
Indications for Use:
The Trojan XOXO Intense Ribbed Natural Rubber Latex Male Condom with Silicone and Aloe Vera OE Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Technological Characteristics:
The Trojan™ XOXO™ Intense Ribbed condom has a bulbous reservoir end shape designed with 11 rows of continuous ribs on its shank and 9 rows of continuous ribs on its bulb. The Trojan™ XOXO™ Intense Ribbed condom is made from a pre-vulcanized natural rubber latex. The purpose and technological function of the pre-vulcanized natural rubber latex in the latex condom formulation is the same as the current natural rubber latex use in the predicate device. The Trojan™ XOXO™ Intense Ribbed condom has a silicone base lubricant with Aloe Vera OE.
Substantial Equivalence Discussion:
| Characteristic | NEW DEVICE:K161915 | PREDICATE DEVICE:K131887 |
|---|---|---|
| Indications for Use | For contraception and forprophylactic purposes (tohelp prevent pregnancyand the transmission ofsexually transmittedinfections). | Same |
| Device Composition | Pre-vulcanized NaturalRubber Latex andstabilizing additives | Natural Rubber Latex,vulcanizing additives andstabilizing additives |
| Lubricant | Silicone with Aloe | Glycol with sensate |
| Shape | Bulbous reservoir endwith ribs | Same |
| Nominal Length | 185 ± 10mm | Same |
| Nominal Flat Width | 54 ± 2mm | Same |
| Bulb Flat Width | 65 mm | Same |
| Thickness | 0.09 ± .01mm | 0.061 ± .01mm |
The indication for use for the Trojan™ XOXO™ Intense Ribbed condom and predicate device are the same; therefore, they have the same intended use.
{5}------------------------------------------------
Although there are differences in the thickness, lubricant, and latex formulations, these differences do not raise different questions of safety and effectiveness as compared to the predicate device.
Biocompatibility:
Biocompatibility testing was performed to evaluate the biocompatibility of the Trojan™ XOXO™ Intense Ribbed condom in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1(2009). The device passed each biocompatibility test identified below:
| Test Performed | ISO Standard |
|---|---|
| Cytotoxicity ISO Elution Method | ISO 10993-5 (2009) |
| Acute Systemic Toxicity | ISO 10993-11 (2006) |
| Primary Rabbit Skin Irritation | ISO 10993-10 (2010) |
| Rabbit Vaginal Irritation | ISO 10993-10 (2010) |
| Rabbit Penile Irritation | ISO 10993-10 (2010) |
| Guinea Pig Maximum Sensitization | ISO 10993-10 (2010) |
Physical Testing:
Three (3) lots of Trojan™ XOXO™ Intense Ribbed condom were tested and met specifications of ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms) and ISO 4074: Natural Rubber Latex Condoms Requirements and Test Methods.
Shelf Life:
The Trojan™ XOXO™ Intense Ribbed condom has a three year (36 month) shelflife based on the results of an accelerated stability study protocol created with reference to 21 CFR 801.435. Five year (60 month) expiration dating will be verified with real-time stability according to 21 CFR 801.435.
Conclusion
The Trojan™ XOXO™ Intense Ribbed condom has the same intended use as the predicate device. Based on performance data and biocompatibility review and testing and safety data, Trojan™ XOXO™ Intense Ribbed condom is substantially equivalent to the predicate device.
§ 884.5300 Condom.
(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.