K Number
K161800
Device Name
OnSite-IV
Manufacturer
Date Cleared
2016-12-06

(159 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OnSite-IV™ is a single use sterile device for the preparation and / or transfer of drugs from a standard pharmaceutical vial into an elastomeric pump or IV bag.

Device Description

The OnSite-IV™ device is a sterile, single use device that, through the use of an integrated vial spike, facilitates creation of a sterile fluid path to transfer liquid from the syringe body to a standard pharmaceutical vial with a 20 mm cap containing a drug, in order to mix or reconstitute the drug and aspirate and transfer the prepared drug back into the syringe body for delivery into an IV bag or elastomeric pump. The device is intended to be used by health care professionals (HCPs) such as physicians, nurses and pharmacists in a clinical setting.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (OnSite-IV™), not an AI/ML device. As such, the information typically requested regarding acceptance criteria and studies for AI/ML performance (like sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, and training set information) is not applicable here.

This document describes the equivalence of a physical medical device (Intravascular Administration Set) to existing predicate devices based on its design features, intended use, and non-clinical performance testing.

Here's a summary of the non-clinical testing performed and the general acceptance criteria inferred, as best as can be derived from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria values for each test, but rather confirms that the device "operates as specified" and "functions as intended." The "reported device performance" is a confirmation of passing these tests.

Acceptance Criteria CategorySpecific Test (Standard)Reported Device Performance
SterilitySterilization Validation (ANSI/AAMI/ISO 11137-2:2013)Achieves 10^-6 sterility assurance level (SAL)
BiocompatibilityCytotoxicity (ISO 10993-5)Passed
Maximization Sensitization (ISO 10993-10)Passed
Intracutaneous Reactivity Irritation (ISO 10993-10)Passed
Acute Systemic Toxicity (ISO 10993-11)Passed
Hemolysis, Extract Method (ASTM F 756)Passed
Hemolysis, Direct Method (ASTM F 756)Passed
Materials Mediated Pyrogen Test (ISO 10993-11, USP )Passed
Physical Performance (ISO 7886-1:1993 for syringes)Limits for Acidity or AlkalinityPassed
Limits for Extractable MetalsPassed
Tolerance on Graduated CapacityPassed
Dead SpacePassed
Liquid Leakage at Syringe Piston under CompressionPassed
Air Leakage at Syringe Piston under CompressionPassed
Air Leakage past Syringe Piston during AspirationPassed
Fiducial LinePassed
Fit of Piston in BarrelPassed
Additional Performance TestingLuer Connection Performance (ISO 594-1, ISO 594-2)Passed
Particulate Matter (USP )Passed
Performance of the Bacterial Retentive Filter (ASTM F2101)Passed
Internal Requirements (after sterilization, transport, aging)Visual InspectionAll features verified to operate as specified (passed)
Device Retained Volume After UseAll features verified to operate as specified (passed)
Drug MixingAll features verified to operate as specified (passed)
Force to Operate ValveAll features verified to operate as specified (passed)
Graduated Marking DurabilityAll features verified to operate as specified (passed)

2. Sample size used for the test set and the data provenance
Not applicable as this is a physical device, not an AI/ML system evaluated on a dataset. The testing refers to physical samples of the device. The document does not specify the number of devices or components tested for each non-clinical performance test.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth and expert review are concepts for diagnostic performance evaluation of AI/ML systems. This document describes physical and biological performance tests.

4. Adjudication method for the test set
Not applicable. Adjudication methods are relevant for human interpretation disagreement in diagnostic performance studies.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device, and no human reader studies are mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. The "ground truth" for this device's performance is adherence to established international standards (ISO, ASTM, USP) and internal specifications for physical and biological properties.

8. The sample size for the training set
Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established
Not applicable.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.