K Number
K161792

Validate with FDA (Live)

Date Cleared
2016-09-20

(82 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging.

The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Device Description

The PINPOINT Endoscopic Fluorescence Imaging System (PINPOINT, PINPOINT System) is comprised of an endoscopic video processor/ illuminator (VPI) which is capable of providing visible and near-infrared illumination to a surgical laparoscope, surgical laparoscopes optimized for visible (VIS) and near-infrared (NIR) illumination and imaging, a camera head that is also optimized for visible and near-infrared imaging, and a flexible light quide cable. The following laparoscope models are included as part of the PINPOINT System: SC9100, SC9104, SC9130, SC9131, SC9134, SC9144, SC9504, SC9534 and SC9544. These are the major components of the PINPOINT System.

During surgical procedures, PINPOINT may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the illuminator and the resulting reflected light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

To provide NIR fluorescence imaging, PINPOINT is used with the imaging agent, indocvanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the VPI, and the fluorescence response is then imaged with the camera, processed and displayed on an HD video monitor.

When used with a VIS/NIR laparoscope, PINPOINT can operate to provide illumination and imaging of both visible light and ICG fluorescence images simultaneously.

AI/ML Overview

The provided text does not contain detailed information about specific acceptance criteria or a dedicated study proving the device meets these criteria in the format requested. The document is a 510(k) summary for the PINPOINT Endoscopic Fluorescence Imaging System, which primarily focuses on demonstrating substantial equivalence to a predicate device.

However, based on the non-clinical performance testing section, we can infer some general acceptance criteria related to safety and performance through adherence to standards.

Here's an attempt to extract and present the information, acknowledging the limitations of the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria Category (Inferred)Specific Standard/Requirement Met (Reported Device Performance)
Basic Safety and Essential PerformanceIEC 60601-1:2012 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance conformance demonstrated.
Electromagnetic Compatibility (EMC)IEC 60601-1-2:2007 Medical Electrical Equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests conformance demonstrated.
Safety of Endoscopic EquipmentIEC 60601-2-18:2009 Medical Electrical Equipment – Part 2: Particular requirements for the basic safety and essential performance of endoscopic equipment conformance demonstrated.
Laser SafetyIEC 60825:2007 Safety of laser products – Part 1: Equipment classification and requirements conformance assessed by UL, classified as a Class 3R laser device with internal maximum Class 4 laser radiation.
In Vivo Fluorescence Imaging CapabilityAnimal testing using a porcine model validated the in vivo fluorescence imaging capability of the PINPOINT System.
Substantial EquivalenceDemonstrated substantial equivalence to the predicate device (PINPOINT Endoscopic Fluorescence Imaging System K150956) in terms of safety, effectiveness, and performance based on identical indications for use, technological characteristics, and principle of operation.

Regarding the study proving the device meets acceptance criteria:

The document describes "Non-Clinical Performance Testing" which includes adherence to various IEC standards and animal testing. This is the "study" demonstrating performance against safety and basic functional criteria. However, it's not a typical clinical study with patient outcomes or a reader study.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size:
    • For the in vivo fluorescence imaging capability, the "test set" was a porcine model. The exact number of animals is not specified.
    • For the other performance tests (IEC standards), these are typically performed on hardware and software and do not involve a "test set" in the sense of patient data.
  • Data Provenance:
    • The animal testing (porcine model) appears to be prospective and conducted by NOVADAQ.
    • The standards conformance testing was conducted by NOVADAQ and accredited third-party laboratories.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the document. The animal model study used to validate in vivo fluorescence imaging would likely have involved veterinarians or surgical experts, but their number and qualifications are not detailed. For the standards testing, "ground truth" is typically defined by the standard itself rather than expert consensus on a test set.

4. Adjudication Method for the Test Set

  • This information is not provided. Given the nature of the tests described (animal model and standards conformance), a formal adjudication method like 2+1 or 3+1 is unlikely to have been used, as these are typically reserved for image interpretation studies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study is mentioned. This device is an imaging system, not an AI-powered diagnostic tool, so such a study would not be directly applicable in the sense of "human readers improve with AI." The document focuses on the system's ability to provide visual assessment of vessels, blood flow, and bile ducts using near-infrared imaging alongside visible light.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This device is an imaging system used in real-time by surgeons; it is not an algorithm that performs a standalone diagnostic function. Therefore, a standalone algorithm performance study is not applicable and not mentioned. The device's "performance" is its ability to acquire and display images.

7. The type of ground truth used

  • For the in vivo fluorescence imaging validation, the ground truth was likely physiological observation within the porcine model by trained personnel, confirming the system's ability to visualize expected fluorescence in living tissue after ICG administration.
  • For the standards conformance tests, the "ground truth" is the requirements outlined in the respective IEC and UL standards.

8. The sample size for the training set

  • This device is an imaging system, not a machine learning algorithm that requires a "training set" in the conventional sense. Therefore, information about a training set sample size is not applicable and not provided.

9. How the ground truth for the training set was established

  • As there is no mention of a training set, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 20, 2016

Novadaq Technologies Inc. Jen Pendlebury Director of Regulatory Affairs 8329 Eastlake Drive. Unit 101 Burnaby, British Columbia Canada V5A 4W2

Re: K161792

Trade/Device Name: Pinpoint Endoscopic Fluorescence Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, IZI Dated: August 25, 2016 Received: August 26, 2016

Dear Jen Pendlebury:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Christopher J. Ronk -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161792

Device Name

PINPOINT Endoscopic Fluorescence Imaging System

Indications for Use (Describe)

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging.

The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 6 - 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92.

Trade Name:PINPOINT Endoscopic Fluorescence Imaging System
Device Model Number:PC9000
Common Name:Endoscope Video Imaging System
Classification:21 CFR § 876.1500
Classification Name:Laparoscope, General & Plastic Surgery; Angiographic X-raySystem
Product Code:GCJ; IZI
Classification:Class II
Manufacturer:Novadaq Technologies Inc.8329 Eastlake Drive, Unit 101Burnaby, British ColumbiaCanada V5A 4W2
Contact Name:Jen PendleburyDirector of Regulatory AffairsTel: (905) 629-3822 ext. 205Fax: (905) 249-0656
Date 510(k) Summary Prepared:June 29, 2016
Predicate Devices:PINPOINT Endoscopic Fluorescence Imaging System (K150956)(Novadaq Technologies Inc.)

Device Description:

The PINPOINT Endoscopic Fluorescence Imaging System (PINPOINT, PINPOINT System) is comprised of an endoscopic video processor/ illuminator (VPI) which is capable of providing visible and near-infrared illumination to a surgical laparoscope, surgical laparoscopes optimized for visible (VIS) and near-infrared (NIR) illumination and imaging, a camera head that is also optimized for visible and near-infrared imaging, and a flexible light quide cable. The following laparoscope models are included as part of the PINPOINT System: SC9100, SC9104, SC9130, SC9131, SC9134, SC9144, SC9504, SC9534 and SC9544. These are the major components of the PINPOINT System.

During surgical procedures, PINPOINT may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the illuminator and the resulting reflected light is imaged by the

{4}------------------------------------------------

camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

To provide NIR fluorescence imaging, PINPOINT is used with the imaging agent, indocvanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the VPI, and the fluorescence response is then imaged with the camera, processed and displayed on an HD video monitor.

When used with a VIS/NIR laparoscope, PINPOINT can operate to provide illumination and imaging of both visible light and ICG fluorescence images simultaneously.

Proposed Indications for Use of the PINPOINT System:

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging.

The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Summary of Technological Characteristics of the PINPOINT System and Predicate Device

Based on the technological characteristics and fundamental scientific premise, the PINPOINT System (PC900) has been determined to be substantially equivalent to the predicate device, the PINPOINT Endoscopic Fluorescence Imaging System (K150956). Both systems use the same technology and mode of imaging to provide real-time endoscopic visible and NIR fluorescence imaging during minimally invasive surgical procedures. These systems utilize the same mode of imaging - visible and near infrared fluorescence imaging, with ICG as the imaging agent, used in the hospital operating room. All of these systems have the same integral components - a light source console and a light cable for outputting light, a camera control unit for processing NIR and VIS light images, a coupler attached to the laparoscope and a camera head, and a laparoscope for VIS and NIR light illumination and imaging. The proposed PINPOINT laparoscopes are substantially equivalent to the existing PINPOINT laparoscopes (K150956) with minor modification to provide users with an alternative laparoscope.

Non-Clinical Performance Testing of the PINPOINT System

The PINPOINT System was designed and developed by Novadaq Technologies Inc. (NOVADAQ), in accordance with the applicable requirements and standards to establish performance and safety of the device. Device safety and performance were verified by tests conducted by NOVADAQ and accredited third party laboratories.

The PINPOINT System was tested in accordance with IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2007 Medical Electrical Equipment - Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility - Requirements and tests conformance testing was also conducted on the PINPOINT System and test results showed that PINPOINT conforms

{5}------------------------------------------------

to the applicable requirements. PINPOINT was also tested in accordance with IEC 60601-2-18:2009 Medical Electrical Equipment – Part 2: Particular reguirements for the basic safety and essential performance of endoscopic equipment, and has demonstrated conformance to the standard. Conformance of PINPOINT with IEC 60825:2007 Safety of laser products – Part 1: Equipment classification and requirements was assessed by Underwriters Laboratories Inc. (UL) and showed that PINPOINT is a Class 3R laser device with internal maximum Class 4 laser radiation.

Animal testing using a porcine model validated the in vivo fluorescence imaging capability of the PINPOINT System.

Conclusions

It has been demonstrated in this Special 510(k) submission that the proposed modification to the PINPOINT System is substantially equivalent to the predicate device in terms of safety, effectiveness and performance. This determination is based on the proposed and predicate devices having the same indications for use, technological characteristics and principle of operation. The device modification to the PINPOINT Endoscopic Fluorescence Imaging System outlined in this summary raises no issues related to its safety and effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.