(307 days)
No
The device description and performance studies focus on electrochemical measurement and standard software conversion of electrical signals to glucose concentration, with no mention of AI/ML algorithms for data analysis or interpretation.
No.
The device is for monitoring glucose levels (a diagnostic aid), not for directly treating a condition.
Yes
The device quantifies glucose levels in blood to monitor the effectiveness of diabetes control, which is a diagnostic function, even though it's explicitly stated to not be used for initial diagnosis or screening.
No
The device description explicitly states that the system consists of a physical "iGlucose meter" which is an electronic device, along with test strips and control solutions. While the meter utilizes software to convert electrical signals to glucose concentrations, it is fundamentally a hardware device with integrated software, not a software-only medical device.
Yes, this device is an IVD (In Vitro Diagnostic).
The intended use explicitly states: "The iGlucose Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use)..." This directly identifies it as an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The iGlucose Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The iGlucose Monitoring System is intended to be used by a single person and should not be shared.
The iGlucose Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iGlucose Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use.
The iGlucose Blood Glucose Test Strips are for use with the iGlucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Product codes (comma separated list FDA assigned to the subject device)
NBW, CGA
Device Description
The iGlucose Blood Glucose Monitoring System consists of the iGlucose meter, the iGlucose test strips and the Rightest Glucose Control Solutions (GC550 level 2 and 4 (previously cleared under K092052). Bionime is seeking FDA's clearance to market and commercially distribute in the United States the iGlucose Meter and its corresponding iGlucose Test Strips.
The iGlucose meter is an electronic device that utilizes electrical characteristics technology. When used with its corresponding iGlucose Test Strips, it quantitatively measures glucose in a small drop of fresh capillary whole blood (minimum 0.75 µL). The blood drop is placed on the test strip where it interacts with reagents in the presence of glucose oxidase to produce an electrical current proportional to the amount of glucose in the sample. The oxidase electrochemical sensor of the meter measures the current and using meter software converts it to the corresponding glucose concentration. The glucose result is then displayed on the meter within 5 seconds.
Other system components include:
- Rightest lancing device labeled for "Single Patient Use only" as well as its . disposable sterile lancets.
- FMK; Lancet, Blood, Class I, 21.CFR 878.4800, General & Plastic Surgery
- . Meter charger and cable
- . A clear cap for the lancing device
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
fingertips
Indicated Patient Age Range
Not for neonatal use. Intended for people with diabetes.
Intended User / Care Setting
self-testing outside the body (in vitro diagnostic use) by people with diabetes at home
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
a. Precision Evaluation Study
Study Type: Repeatability and Intermediate Precision Evaluation
Sample Size: 1500 tests for repeatability (5 venous blood samples, 10 replicates, 3 test strip lots, 10 meters); 900 tests for intermediate precision (3 glucose control solutions, 10 replicates, 3 test strip lots, 10 meters).
Key Results:
Repeatability:
Sample P-01: Mean 44.4 mg/dL, SD 1.6 mg/dL, CV 3.6%
Sample P-02: Mean 97.6 mg/dL, SD 2.0 mg/dL, CV 2.1%
Sample P-03: Mean 132.5 mg/dL, SD 2.3 mg/dL, CV 1.8%
Sample P-04: Mean 213.2 mg/dL, SD 3.5 mg/dL, CV 1.6%
Sample P-05: Mean 364.0 mg/dL, SD 5.4 mg/dL, CV 1.5%
Intermediate Precision:
Glucose levels CS-L: Mean 42.7 mg/dL, SD 1.2 mg/dL, CV 2.8%
Glucose levels CS-N: Mean 96.7 mg/dL, SD 2.0 mg/dL, CV 2.0%
Glucose levels CS-H: Mean 274.3 mg/dL, SD 4.6 mg/dL, CV 1.7%
The results demonstrate that the performance of the iGlucose meter falls within acceptance criteria that are equivalent to those produced by the predicate device.
b. Linearity Evaluation Study
Study Type: Linearity testing
Sample Size: 3 lots of iGlucose blood glucose test strips. Venous blood samples spiked to 15 levels of glucose concentrations.
Key Results: Linear regression analysis resulted in a mean slope ranging from 0.9808 to 0.9952 and an R2 value of 0.9996 to 0.9997. The results support the claim that the iGlucose Blood Glucose Monitoring System is linear between 20 and 600 mg/dL.
c. Interferences - Endogenous/Exogenous Substances
Study Type: Interference testing
Sample Size: 19 potential interfering substances evaluated using 3 lots of test strips at 2 glucose concentrations of venous blood samples, 5 replicates each.
Key Results: iGlucose meter measurements of samples containing ascorbic acid (> 5 mg/dL), cholesterol (> 600 mg/dL), and uric acid (> 9 mg/dL) resulted in biases greater than ±10% at normal glucose concentrations. Results for the other 16 substances were within acceptance criteria.
d. Interferences - Hematocrit
Study Type: Hematocrit level evaluation
Sample Size: Samples prepared ranging in HCT from 20 to 60% at 6 glucose concentration levels of venous blood samples, 5 replicates each.
Key Results: The measurements of the iGlucose meter were within acceptance criteria for the entire HCT range of 20-60%.
e. Sample Volume
Study Type: Sample volume evaluation
Sample Size: 3 lots of blood glucose test strips. Venous blood samples tested at 3 glucose concentration levels and at 9 sample volumes ranging from 0.60 to 3.0 µL.
Key Results: When sample volumes were less than 0.70μL, the iGlucose meter displayed an error message. At sample volumes greater than 0.75μL, all iGlucose meter test results were within the acceptance criteria of ±10% compared to reference measurements. Supports a minimum sample volume of 0.75 µL.
f. Other Bench Performance studies
Study Type: Traceability, Accurate meter display of Hi and Lo, Temperature and Humidity Effects, Altitude Effects, Cleaning and Disinfecting, Virucide Efficacy, Test Strip stability.
Key Results: All tests passed acceptance criteria.
- Traceability: Calibration traced to NIST, SRM 917c glucose standard.
- Accurate meter display of Hi and Lo: All test results passed and complied with acceptance criteria.
- Temperature and Humidity Effects: Results met acceptance criteria and support claims of 50 ~104 °F (10 ~ 40°C) for operating temperature and 10 ~ 90% for RH.
- Altitude Effects: Results met acceptance criteria and support operation at altitudes up to 10,000 ft.
- Cleaning and Disinfecting: Robustness studies demonstrated no change in performance or external materials after cleaning and disinfecting with CaviWipes Disinfecting Towelettes, simulating 5 years of multiple single-patient use.
- Virucide Efficacy: 2 minutes contact time with CaviWipe Disinfecting Towelettes effectively inactivated HBV.
- Test Strip stability: Protocol and acceptance criteria support claimed stability durations when stored at 4-30 ℃ and at a relative humidity of up to 90%.
g. Method Comparison - Lay-User Trial
Study Type: Lay-User Trial (clinical performance)
Sample Size: 153 subjects.
Key Results:
iGlucose vs. YSI for glucose concentration
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a family of three figures, with the heads of the figures forming the top part of the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Mav 2, 2017
BIONIME CORPORATION % NINA PELED, PH.D., M.B.A., REGULATORY CONSULTANT MEDICAL DEVICE REGULATORY CONSULTANTS 627 16TH AVE. MENLO PARK, CA 94025
Re: K161790
Trade/Device Name: iGlucose Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA Dated: March 23, 2017 Received: March 29, 2017
Dear Dr. Peled:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of
1
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K161790
Device Name
iGlucose Blood Glucose Monitoring System
Indications for Use (Describe)
The iGlucose Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The iGlucose Monitoring System is intended to be used by a single person and should not be shared.
The iGlucose Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iGlucose Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use.
The iGlucose Blood Glucose Test Strips are for use with the iGlucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
SECTION 10: EXECUTIVE 510(k) SUMMARY
Assigned 510(k) Number: K161790
1. Submitter Information
Company: | Bionime Corporation |
---|---|
No.100, Sec. 2, Daqing St., South Dist., | |
Taichung City 40242, Taiwan (R.O.C.) | |
Contact Person: | Nina Peled |
Regulatory Consultant to the Bionime Corporation | |
627 16th Ave. | |
Menlo Park, CA 94025 | |
npeled@yahoo.com | |
650 454 0322 | |
2. Date Prepared: | September 9, 2016 |
-
- Device Name: iGlucose Blood Glucose Monitoring System
-
- Product Code, Class, Regulation Number and Panel
Product Code | Class | Regulation Number | Panel |
---|---|---|---|
NBW; System, Test, Blood | |||
Glucose, OTC | Class II | 21.CFR 862.1345 | Clinical chemistry |
CGA; Glucose Oxidase, Glucose | Class II | 21.CFR 862.1345 | Clinical chemistry |
-
- Predicate Device: GE Blood Glucose Monitoring System 333 (listed in FDA's database as Ge333 Blood Glucose Monitoring System), K143387
6. Device Description and Principle of Operation:
The iGlucose Blood Glucose Monitoring System consists of the iGlucose meter, the iGlucose test strips and the Rightest Glucose Control Solutions (GC550 level 2 and 4 (previously cleared under K092052). Bionime is seeking FDA's clearance to market and commercially distribute in the United States the iGlucose Meter and its corresponding iGlucose Test Strips.
The iGlucose meter is an electronic device that utilizes electrical characteristics technology. When used with its corresponding iGlucose Test Strips, it quantitatively measures glucose in a small drop of fresh capillary whole blood (minimum 0.75 µL). The blood drop is placed on the test strip where it interacts with reagents in the presence of
4
glucose oxidase to produce an electrical current proportional to the amount of glucose in the sample. The oxidase electrochemical sensor of the meter measures the current and using meter software converts it to the corresponding glucose concentration. The glucose result is then displayed on the meter within 5 seconds.
Other system components include:
- Rightest lancing device labeled for "Single Patient Use only" as well as its . disposable sterile lancets.
- FMK; Lancet, Blood, Class I, 21.CFR 878.4800, General & Plastic Surgery
- . Meter charger and cable
- . A clear cap for the lancing device
7. Indications for Use:
The iGlucose Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The iGlucose Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.
The iGlucose Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iGlucose Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use.
The iGlucose Blood Glucose Test Strips are for use with the iGlucose Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
8. Special Conditions for Use Statement
- . The glucose meter and lancing device are for single patient use. Do not use on multiple patients. Do not share meter with anyone including other family members.
- . Do not use the lancing device for assisted blood draws by healthcare providers or at healthcare provision sites and do not share it with anyone else, even a family member.
- . The iGlucose Blood Glucose Monitoring System is not for use on neonates and should not be used in the critically ill patients, patients in shock, dehydrated patients or hyperosmolar patients.
- Do not reuse; each Test Strip is for single use only. ●
- . Do not test samples other than fresh capillary whole blood obtained from the fingertip.
5
- . Inaccurate test results may be obtained at altitudes greater than 10,000 ft above sea level.
- . Do not perform the blood glucose test at temperatures below 10°C (50°F) or above 40°C (104°F), nor below 10% or above 90% relative humidity.
- Hematocrits below 20% may cause higher results. Hematocrits above 60% may ● cause lower results.
- . High concentrations of Uric acid >9 mg/dL, Cholesterol >600 mg/dL, and Ascorbic acid (Vitamin C) >5 mg/dL may interfere with the glucose test causing inaccurate test results.
- iGlucose Blood Glucose Test Strips are designed for use with capillary whole blood samples. Do not use serum or plasma samples.
- . The iGlucose Blood Glucose Monitoring System should not be used to screen for or diagnose diabetes mellitus.
- For single-patient use only.
9. Comparison to the Predicate Device
The iGlucose Blood Glucose Monitoring System is a modified version of a cleared device, the GE Blood Glucose Monitoring System 333, K143387. The modifications to the cleared device include replacing the Bluetooth data telecommunication with a cell enabled transmission (GSM) as well as some minor customer convenience upgrades.
The comparison table below outlines the similarities and differences between the subject device, the iGlucose Blood Glucose Monitoring System, and the GE Blood Glucose Monitoring System 333 serving as the predicate device:
Similarities | |
---|---|
Fundamental Scientific Technology of the Device | Oxidase Electrochemical Sensor. The structure, function and performance of the blood glucose monitoring module of the meter and the structure, function and performance of its accompanying test strips is unchanged between the subject device and the predicate device. |
Analyte | Glucose |
Measuring Range | 20-600 mg/dL |
Sample | Fresh capillary whole blood |
Sample Volume | 0.75 µL minimum |
Strip Reagents | 1.Glucose Oxidase (GOD) 14.8% |
2.Potassium ferricyanide 39.5% |
6
Similarities | |
---|---|
3.Non-reactive ingredients 45.7% | |
HCT Range | 20-60% |
Interferences | Ascorbic acid > 5 mg/dL |
Cholesterol > 600 mg/dL | |
Unit of Measurement | mg/dL |
Open Vial Shelf Life | 3 months |
Test Time | 5 seconds |
Coding | Auto coding |
Operating Temperature Range | 50 ~104 °F (10 ~ 40°C) |
Operating Relative Humidity Range | 10 ~ 90% |
Operating Altitude | Up to 10,000 ft |
Test Strip Storage Conditions | 39 ~86 °F (4 ~ 30°C), 10 ~ 90% relative humidity |
Meter Storage Conditions | 14 ~140 °F (-10 ~ 60°C) |
Memory Capacity | 500 blood glucose test results with date and time |
Control Solution | 2 levels (Level 2 and 4, GC550) |
Performance | Both devices passed same acceptance criteria for bench |
performance: precision, linearity and interferences as well as clinical | |
performance: a lay-user trial. |
7
Differences | ||
---|---|---|
iGlucose Blood Glucose Monitoring System | GE Blood Glucose Monitoring System 333, K143387 | |
Indications for Use | The iGlucose Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The iGlucose Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. | |
The iGlucose Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iGlucose Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes or for neonatal use. | ||
The iGlucose Blood Glucose Test Strips are for use with the iGlucose Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. | The GE Blood Glucose Monitoring System 333 is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The GE Blood Glucose Monitoring System 333 is intended to be used by a single person and should not be shared. | |
The GE Blood Glucose Monitoring System 333 is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GE Blood Glucose Monitoring System 333 should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). | ||
The GE Blood Glucose Test Strips 333 are for use with the GE Blood Glucose Meter 333 to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. | ||
Power Supply | Rechargeable battery (3.7V, 1,000mAh Li-Ion) | Two 1.5V (AAA) batteries |
Meter Dimensions | $104 mm \times 49.8 mm \times 16.5 mm$ | $85 mm \times 57 mm \times 22.5 mm$ |
Meter Weight | $85.0 \pm 5 g$ with battery | $81.0 \pm 5 g$ with batteries |
Differences | ||
iGlucose Blood Glucose Monitoring | ||
System | GE Blood Glucose Monitoring System | |
333, K143387 | ||
Battery | ||
Life | About 500 tests per battery charge | |
without use of data transmission. | ||
Number of uses is reduced and varies | ||
when data is transmitted. | About 800 tests per battery life | |
LCD | ||
Display | ||
Area | 36.6 mm × 49 mm | 39 mm × 39.5 mm |
Data | ||
Transmission | GSM | Bluetooth 4.0 (Low energy) |
Power | ||
Saving | Automatically turns to standby mode. | Turns to standby by pressing the main |
button for 3 seconds. | ||
Backlight | Yes | No |
Error | ||
Messages | Spelled out on meter display | A numerical code is displayed |
Interferen | ||
ces | Uric acid > 9 mg/dL | Uric acid > 10 mg/dL |
8
The indications for use statement for the two devices is the same apart from limiting the iGlucose Blood Glucose Monitoring System to testing capillary blood samples from the fingertip only and thus eliminating the risk of lag in blood glucose levels of samples obtained from alternative body sites. Such lag may be present when going through a rapid change in blood glucose concentration. The predicate device lists testing of samples from the fingertips, forearm or palm.
There is also no change to the fundamental scientific technology of the device. The structure, function and performance of the blood glucose monitoring module in the two devices and the structure, function and performance of their accompanying test strips is the same.
10. Standards /Guidance Documents Referenced
- CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline- Third Edition(2014)
- CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline -Second Edition (2010)
9
- CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: ● A Statistical Approach; Approved Guideline (2014)
- IEC 61010-1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements (2001)
11. Performance Characteristics
Bench studies including evaluation of analytical performance as well as clinical performance studies were conducted and passed acceptance criteria. Software verification testing was carried out to ensure all meter functions and displayed error messages perform as intended.
a. Precision Evaluation Study
Following CLSI EP05-A3, repeatability testing was conducted over 10 days using 5 venous blood samples with glucose concentrations spanning the measuring range. Each measured in 10 replicates with 3 test strips lots and 10 meters. Intermediate precision testing was conducted over 10 days using 3 glucose control solutions each measured in 10 replicates with 3 test strips lots and 10 meters. A total of 1500 tests were performed for the repeatability evaluation and 900 for the intermediate precision evaluation. iGlucose meter results of venous blood samples were compared to those obtained by the YSI 2300 analyzer as the reference method.
10
Sample | P-01 | P-02 | P-03 | P-04 | P-05 |
---|---|---|---|---|---|
(1) Total test | |||||
numbers (n) | 300 | 300 | 300 | 300 | 300 |
(2) Mean mg/dL | |||||
(mmol/L) | 44.4 (2.5) | 97.6 (5.4) | 132.5 (7.4) | 213.2 (11.8) | 364.0 (20.2) |
(3) SD mg/dL | |||||
(mmol/L) | 1.6 (0.09) | 2.0 | |||
(0.11) | 2.3 (0.13) | 3.5 (0.19) | 5.4 (0.30) | ||
(4) CV (%) | 3.6% | 2.1% | 1.8% | 1.6% | 1.5% |
(i) Repeatability evaluation
(ii) Intermediate Precision evaluation
Glucose levels | CS-L | CS-N | CS-H |
---|---|---|---|
(1) Total test | |||
numbers (n) | 300 | 300 | 300 |
(2) Mean mg/dL | |||
(mmol/L) | 42.7 (2.4) | 96.7 (5.4) | 274.3 (15.2) |
(3) SD mg/dL | |||
(mmol/L) | 1.2 (0.07) | 2.0 (0.11) | 4.6 (0.26) |
(4) CV (%) | 2.8% | 2.0% | 1.7% |
The results of both repeatability and intermediate precision testing demonstrate that the performance of the iGlucose meter falls within acceptance criteria that are equivalent to those produced by the predicate device.
b. Linearity Evaluation Study
Following CLSI EP6-A, linearity testing of the measuring range was conducted using 3 lots of iGlucose blood glucose test strips. Venous blood samples were spiked to 15 levels of glucose concentrations ranging from 0 to 640 mg/dL and the iGlucose meter test results were compared to those obtained from the YSI 2300 analyzer as the reference method.
A linear regression analysis of the glucose measurements resulted in a mean slope ranging from 0.9808 to 0.9952 and an R2 value of 0.9996 to 0.9997. The results support the claim that the iGlucose Blood Glucose Monitoring System is linear between 20 and 600 mg/dL, a claim equivalent to that made for the predicate device.
11
c. Interferences - Endogenous/Exogenous Substances
Interference testing followed the CLSI EP7-A2. A total of 19 potential interfering substances were evaluated using 3 lots of test strips at 2 glucose concentrations of venous blood samples, 5 replicates each.
Study results indicate that iGlucose meter measurements of samples containing ascorbic acid (> 5 mg/dL), cholesterol > 600 mg/dL), and uric acid (> 9 mg/dL) resulted in biases greater than ±10% at normal glucose concentrations. These iGlucose meter limitations were equivalent to those found for the cleared predicate device. The results of the 16 remaining potential interfering substances were within acceptance criteria for both normal and high glucose concentrations.
d. Interferences - Hematocrit
Hematocrit levels were evaluated using 3 lots of blood glucose test strips. Samples were prepared ranging in HCT from 20 to 60% at 6 glucose concentration levels of venous blood samples, 5 replicates each. The iGlucose results were compared to the YSI 2300 analyzer as the reference method.
Study results indicate that the measurements of the iGlucose meter were within acceptance criteria for the entire HCT range of 20-60%, a claim equivalent to the one made for the cleared predicate device.
e. Sample Volume
A study was conducted to evaluate sample volumes using 3 lots of blood glucose test strips. Venous blood samples were tested at 3 glucose concentration levels and at 9 sample volumes ranging from 0.60 to 3.0 µL.
Study results demonstrated that when sample volumes were less than 0.70μL, the iGlucose meter displayed an error message. At sample volumes greater than 0.75μL, all iGlucose meter test results were within the acceptance criteria of ±10% compared to reference measurements. The results support a minimum sample volume of 0.75 µL for the iGlucose meter, a claim equivalent to that made for the predicate cleared device.
Other Bench Performance studies f.
Additional bench studies were conducted and all passed acceptance criteria including:
- . Traceability. The calibration of the iGucose Test Strips was traced to a NIST, SRM 917c glucose standard
- Accurate meter display of Hi and Lo. Three meters were tested with check keys of known conductivity to elicit an out of range High and an out of range
12
Low reading, 10 replicates each. All test results passed and complied with acceptance criteria.
- Temperature and Humidity Effects. The operating temperature range and operating relative humidity (RH) were evaluated placing the iGlucose meter and strips in environmental chambers varying in temperature between 10 and 40°C and between 10 and 95% RH. Testing included 3 test strip lots and two concentrations of blood samples (normal and high), each in 5 replicates. iGlucose measurements were compared to those obtained by the YSI 2300. Results from all testing conditions met acceptance criteria and support a claim of 50 ~104 °F (10 ~ 40°C) for the operating temperature and 10 ~ 90% for the RH.
- Altitude Effects. The effect of altitudes on the performance of the iGlucose Test strip was evaluated using both control solutions and made up venous blood samples. Testing was conducted at altitudes varying between 0 and 10745 ft. At each altitude, iGlucose blood test results were compared to those of the YSI 2300. The results met acceptance criteria and support the claim that the iGlucose system can be operated at altitudes up to 10,000 ft.
- Cleaning and Disinfecting. The effect of the CaviWipes Disinfecting ● Towelettes was evaluated on five iGlucose meters following cleaning and disinfecting cycles after testing blood samples. Robustness studies were performed demonstrating that there was no change in performance or in external materials of the meter after cycles of cleaning and disinfecting with the CaviWipes Disinfecting Towelettes. The robustness studies were designed to simulate 5 years of multiple use by a single patient.
- Virucide Efficacy. The efficacy of the recommended disinfection against ● Hepatitis B virus (HBV) was evaluated on the materials of the iGlucose meter. Three lots of each material were evaluated and analytical studies included limit of detection (LoD), limit of blank, recovery, interference and, neutralization. A 2 minutes contact time of meter materials with the CaviWipe Disinfecting Towelettes produced results below the LoD and demonstrate effective inactivation of HBV when used to disinfect the iGlucose meter materials.
- Test Strip stability. Open vial and shelf life stability of the iGlucose test strip ● are assessed using real time and accelerated studies. Protocol and acceptance criteria support the claimed stability durations when stored at 4-30 ℃ and at a relative humidity of up to 90%.
13
- Readability Assessment of labeling: A Flesch-Kincaid reading level ● assessment was conducted demonstrating that the User's Manual, the Getting Started Guide and the Test Strip Insert were written at or below an 8th grade reading level.
g. Method Comparison - Lay-User Trial
A Lay-User trial was performed in a US clinic to ensure the modified device meets user requirements and to demonstrate the ability of subjects to self-train and selfmeasure their blood glucose levels using the iGlucose Blood Glucose Monitoring System. iGlucose test results were compared to those obtained by the Yellow Springs Instrument YSI 2300 Plus as the reference method, ClinicalTrials.gov ldentifier: NCT02709707.
A total of 153 subjects participated; most with a diagnosis of diabetes, either type 1 or type 2 of various duration, 23 were subjects naive to SMBG devices, 6 subjects presented with blood glucose levels