AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aequalis PerFORM Reversed & Aequalis PerFORM+ Reversed Glenoid are indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and non-repairable rotator cuff-tear with pain disabled by:

  • · Rheumatoid arthritis
  • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
  • · Correction of functional deformity
  • · Fractures of the humeral head
  • · Traumatic arthritis
  • · Revision of the devices if sufficient bone stock remains.
Device Description

The Aequalis™ PerFORM Reversed and Aequalis™ PerFORM+ Reversed Glenoid are intended to replace the glenoid part of the scapulohumeral joint as part of a reverse shoulder prosthesis. Aequalis™ PerFORM Reversed and Aequalis™ PerFORM+ Reversed glenoid component assembly must be used in association with a Tornier humeral component:

  • Humeral implants Aequalis Reversed, Aequalis Reversed Fracture or Aequalis Adjustable ● Reversed Shoulder
  • . or Humeral implants Aequalis Ascend Flex Shoulder System in reverse configuration
  • or humeral implants Aequalis Reversed FX2 ●

The glenoid component assembly includes a modular system including a baseplate with central screw or press-fit post, peripheral anchoring screws and a glenosphere. The baseplates are available in various design configurations including: standard, lateralized, half-wedge augment and full-wedge augment. The glenospheres are available in various design configurations including: standard, lateralized and eccentric. The press-fit post may be used with the standard baseplate in lieu of the central anchoring screw. The baseplates and the press-fit posts are manufactured using additive manufacturing technology.

AI/ML Overview

The document provided does not describe an acceptance criteria or study that proves the device meets specific performance metrics in the way a clinical or standalone study for an AI/algorithm-based device would.

Instead, this is an FDA 510(k) Premarket Notification for a medical device (Aequalis PerFORM Reversed and Aequalis PerFORM+ Reversed Glenoid – a shoulder prosthesis). The purpose of a 510(k) submission is to demonstrate "substantial equivalence" to a legally marketed predicate device, not necessarily to meet pre-defined performance acceptance criteria as would be typical for a diagnostic or AI device.

Here's how the information provided relates to your request, with an emphasis on what's not relevant for a typical AI/algorithm acceptance criteria study:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as pass/fail numeric targets for performance metrics like sensitivity, specificity, or AUC, as these are not relevant for a mechanical prosthesis 510(k). The "acceptance criteria" here is that the device is "substantially equivalent" to predicate devices, meaning it is as safe and effective.

  • Reported Device Performance: The document lists "Performance Data" which consists of a series of non-clinical bench tests (mechanical testing). These demonstrate the physical properties and durability of the prosthesis. Examples include:

    • Compressive Strength and Elastic Modulus Testing of Titanium Porous Metal Verification
    • Titanium Porous Structure Stereological Evaluation
    • Titanium Porous Structure Mechanical Verification
    • PerFORM Reversed Fixation Pull-out Test
    • Fatigue Test
    • Range of Motion Analysis
    • PerFORM Reversed Taper Pull-Off Test
    • Glenoid Loosening Test (100,000 cycles)
    • PerFORM+ Reversed Augment Baseplate compared to the Exactech Equinoxe Reverse Augmented Baseplate Analysis
    • PerFORM Reversed Lateralization Comparison
    • PerFORM Reversed and PerFORM+ Reversed Peripheral Screw Angulation
    • Direct Metal Laser Sintering (DMLS) Process Validation
    • Endotoxin (<20 EU / device)

    The document states that these tests were performed "to demonstrate substantial equivalence to the predicate device," implying that the results of these tests met benchmark values or comparisons to the predicate device, although specific numerical results or direct comparisons are not provided in this summary.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not applicable in the context of an AI/algorithm test set. The "samples" here refer to physical components of the prosthesis used in mechanical bench testing. The specific number of mechanical test samples for each test (e.g., how many glenoids were subjected to fatigue testing) is not provided in this summary.
  • Data Provenance: Not applicable. These are laboratory bench tests on physical devices, not clinical data from patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Experts and Qualifications: Not applicable. Ground truth, in the AI sense, relates to clinical diagnoses or interpretations, which are not part of this mechanical testing.

4. Adjudication method for the test set

  • Adjudication Method: Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This is a mechanical prosthesis, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: No. This is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Ground Truth: Not applicable in the AI/algorithm sense. The "ground truth" for these mechanical tests would be established by engineering specifications, material properties, and comparison to the predicate device's known performance characteristics.

8. The sample size for the training set

  • Training Set Sample Size: Not applicable. This device is not an AI/machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

  • Training Set Ground Truth: Not applicable.

In summary: The provided document is an FDA 510(k) summary for a physical medical device (shoulder prosthesis). It describes mechanical bench testing to demonstrate performance sufficient for "substantial equivalence" to a predicate device, rather than a clinical study or performance evaluation of an AI-based diagnostic or therapeutic algorithm, which is what your questions are primarily tailored to. Therefore, most of your questions are not directly applicable to the content of this specific document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized depiction of three human profiles facing to the right, layered on top of each other. The profiles are rendered in a simple, abstract manner.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 15, 2016

Tornier, Incorporated Ms. Laurie Lewandowski Regulatory Affairs Consultant 10801 Nesbitt Avenue South Bloomington, Minnesota 55340

Re: K161742

Trade/Device Name: Aequalis PerFORM Reversed, Aequalis PerFORM+ Reversed Glenoid Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS Dated: October 14, 2016 Received: October 17, 2016

Dear Ms. Lewandowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K161742

Device Name

Aequalis Perform Reversed and Aequalis PerFORM+ Reversed Glenoid

Indications for Use (Describe)

The Aequalis PerFORM Reversed & Aequalis PerFORM+ Reversed Glenoid are indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and non-repairable rotator cuff-tear with pain disabled by:

  • · Rheumatoid arthritis
  • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
  • · Correction of functional deformity
  • · Fractures of the humeral head
  • · Traumatic arthritis
  • · Revision of the devices if sufficient bone stock remains.

Notes:

  • · All components are single use.
  • The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "TORNIER" in a blue, sans-serif font. To the right of the word is a blue hexagon containing a stylized "T" logo. The logo consists of a "T" inside of an octagon. The overall design is clean and professional.

510(k) Summary

510(k) Number: K161742

Date Prepared: November 9, 2016

I. Administrative Information

Name:Tornier, Inc.
Address:10801 Nesbitt Avenue SouthBloomington, MN 55437United States of America
Contact Person:Laurie LewandowskiRegulatory Affairs Consultant
Phone:612-770-4038
Fax:952-426-7601

II. Device Information

Name of Device:Aequalis Perform Reversed and Aequalis PerFORM+ Reversed Glenoid
Common Name:Shoulder Prosthesis
Classification Name:21 CFR 888.3660, shoulder joint metal/polymer or semi-constrainedcemented prosthesis.
Regulatory Class:Class II
Primary Product Code:PHX
Secondary Product Code:KWS

III. Predicate Device Information

Primary Predicate:Aequalis Reversed Shoulder Prosthesis, K081059
Reference Device:Zimmer Trabecular Metal Reverse, K052906
Reference Device:Exactech Equinoxe Reversed Augmented, K110708

IV. Device Description

The Aequalis™ PerFORM Reversed and Aequalis™ PerFORM+ Reversed Glenoid are intended to replace the glenoid part of the scapulohumeral joint as part of a reverse shoulder prosthesis. Aequalis™ PerFORM Reversed and Aequalis™ PerFORM+ Reversed glenoid component assembly must be used in association with a Tornier humeral component:

  • Humeral implants Aequalis Reversed, Aequalis Reversed Fracture or Aequalis Adjustable ● Reversed Shoulder
  • . or Humeral implants Aequalis Ascend Flex Shoulder System in reverse configuration
  • or humeral implants Aequalis Reversed FX2 ●

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Tornier. The logo consists of the word "TORNIER" in blue, followed by a blue hexagon containing a stylized "T" inside. The font of the wordmark is a serif typeface, and the overall design is clean and professional.

The glenoid component assembly includes a modular system including a baseplate with central screw or press-fit post, peripheral anchoring screws and a glenosphere. The baseplates are available in various design configurations including: standard, lateralized, half-wedge augment and full-wedge augment. The glenospheres are available in various design configurations including: standard, lateralized and eccentric. The press-fit post may be used with the standard baseplate in lieu of the central anchoring screw. The baseplates and the press-fit posts are manufactured using additive manufacturing technology.

V. Intended Use

The Aequalis PerFORM Reversed & Aequalis PerFORM+ Reversed Glenoid are intended for replacement of the shoulder joint to reduce pain and improve shoulder mobility in comparison with preoperative status.

VI. Indications for Use

The Aequalis PerFORM Reversed & Aequalis PerFORM+ Reversed Glenoid are indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and nonrepairable rotator cuff-tear with pain disabled by:

  • Rheumatoid arthritis
  • Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
  • Correction of functional deformity
  • Fractures of the humeral head
  • Traumatic arthritis
  • Revision of the devices if sufficient bone stock remains. ●

Notes:

  • All components are single use.
  • The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.

VII. Comparison of Technological Characteristics with the Predicate Device

The Aequalis PerFORM Reversed has the same intended use and the same fundamental scientific technology as the primary predicate and reference devices. The augment baseplates are similar to those of the reference device, Exactech Equinoxe (K110708), both of which accommodate varying glenoid bone geometry due to wear on the articular surface. The porous structure of the baseplate is similar to that of the reference device, Zimmer Trabecular Metal Reverse, K052906 Both are designed to minimize the bone reaming needed to preserved as much of the native bone as possible. The design differences have been demonstrated to not affect safety or effectiveness or raise new issues of safety or effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "TORNIER" in blue, with a stylized blue logo to the right. The logo is a hexagon with a "T" inside of it. The font of the word "TORNIER" is a serif font.

VIII. Performance Data

Non-clinical performance bench testing (mechanical testing) was performed to demonstrate substantial equivalence to the predicate device.

  • . Compressive Strength and Elastic Modulus Testing of Titanium Porous Metal Verification
  • Titanium Porous Structure Stereological Evaluation ●
  • Titanium Porous Structure Mechanical Verification
  • PerFORM Reversed Fixation Pull-out Test
  • Fatigue Test ●
  • Range of Motion Analysis
  • PerFORM Reversed Taper Pull-Off Test ●
  • Glenoid Loosening Test (100,000 cycles)
  • . PerFORM+ Reversed Augment Baseplate compared to the Exactech Equinoxe Reverse Augmented Baseplate Analysis
  • . PerFORM Reversed Lateralization Comparison
  • PerFORM Reversed and PerFORM+ Reversed Peripheral Screw Angulation ●
  • Direct Metal Laser Sintering (DMLS) Process Validation
  • Operational Qualification - Mechanical Validation & Microstructure Validation
  • Performance Qualification - Mechanical Validation & Microstructure Validation
    • . Powder Bed Position Validation
    • . Powder Recycling Validation
    • . Powder Removal Validation
  • DMLS Device Powder Removal Cleaning ●
  • DMLS Titanium Material Validation ●
  • Endotoxin (<20 EU / device) ●

IX. Clinical Study

Clinical studies were not required to demonstrate substantial equivalence between the subject device and the predicate device.

X. Conclusions

The Aequalis PerFORM Reversed and Aequalis PerFORM+ Reversed Glenoid described in this section has the same intended use and the same fundamental scientific technology as the cleared Aequalis Shoulder Prosthesis, the porous structure is similar to the Trabecular Metal Reverse, and lateralization and augmentation are similar to Equinoxe Reversed Augmented Shoulder Prosthesis. Based on the testing presented for the design differences between the subject and predicate device, Tornier concludes that subject device is substantially equivalent to the predicate device.

§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”