K Number
K161623
Date Cleared
2016-11-10

(150 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Prontosan Wound Irrigation Solution is intended for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds and abrasions.

Device Description

Wound Irrigation Solution is a clear. clear. colorless liquid containing Prontosan undecylenamidopropyl betaine, polyaminopropyl biguanide, sodium hydroxide and purified water. The solution is aseptically filled using a blow fill seal process into low density polyethylene 40 mL ampoules; 350 mL squeeze bottles and 1000 mL bottles with screw caps. Prontosan Wound Irrigation Solution is used for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post- surgical wounds and abrasions. The mechanical action of moving across the wound provides for the mechanism of action and aids in the removal of foreign material such as dirt and debris. The Prontosan Wound Irrigation Solution Adaptor is a component that may be screwed on to the current Prontosan Wound Irrigation Solution containers (350 ml and 1000 ml bottles) and provides alternate spout geometry for which to access the solution inside the bottle. It is provided in sterile form, labeled for single patient use only.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called "Prontosan Wound Irrigation Solution." This document pertains to regulatory approval rather than a scientific study on the device's performance against detailed acceptance criteria in a clinical or standalone setting.

Therefore, the document does not contain the information requested in the prompt regarding acceptance criteria and a study proving the device meets those criteria.

Specifically, the document states:

  • "The subject device and the predicate device are exactly the same. No changes have been made to the device, manufacturing process, the packaging process or the intended use. The modification subject to this 510(k) is for labeling content only." (Page 4)
  • "No new non-clinical testing has been performed for this submission. The changes to the labeling are test results from testing previously performed on this product and submitted to the FDA in a prior submission." (Page 4)

This indicates that the current submission is not based on new performance data or a new study, but rather a labeling change for a substantially equivalent device. Consequently, none of the requested study details (sample size, ground truth, expert qualifications, MRMC, standalone performance, etc.) are present in this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 10, 2016

B. Braun Medical Inc. Nancy Skocypec Regulatory Affairs Manager 901 Marcon Blvd Allentown, Pennsylvania 18109

Re: K161623

Trade/Device Name: Prontosan Wound Irrigation Solution, 40 M1, Prontosan Wound Irrigation Solution, 350 M1, Prontosan Wound Irrigation Solution, 1000 Ml Regulatory Class: Unclassified Product Code: FRO Dated: June 6, 2016 Received: June 13, 2016

Dear Ms. Nancy Skocypec:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

51D(k) Number (if known)

K161623

Device Name Prontosan Wound Irrigation Solution

Indications for Use (Describe)

Prontosan Wound Irrigation Solution is intended for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds and abrasions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification Prontosan® Wound Irrigation Solution

510(k) Summarv

SUBMITTER:B. Braun Medical Inc.901 Marcon BoulevardAllentown, PA 18109-9341610-266-0500
CONTACT:Nancy SkocypecRegulatory Affairs Manager
DEVICE NAME:Prontosan® Wound Irrigation Solution
COMMON ORUSUAL NAME:Wound Cleanser, Wound Dressing
DEVICECLASSIFICATION:Product Code FRO, Unclassified
PREDICATEDEVICE:Prontosan® Wound Irrigation Solution, K072876B. Braun Medical Inc., General and Plastic Surgery, Unclassified

INTENDED USE

Prontosan Wound Irrigation Solution is for cleaning wounds, and for moistening and lubricating absorbent wound dressings.

INDICATIONS FOR USE:

Prontosan Wound Irrigation Solution is intended for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post-surgical wounds and abrasions.

DEVICE DESCRIPTION:

Wound Irrigation Solution is a clear. clear. colorless liquid containing Prontosan undecylenamidopropyl betaine, polyaminopropyl biguanide, sodium hydroxide and purified water. The solution is aseptically filled using a blow fill seal process into low density polyethylene 40 mL ampoules; 350 mL squeeze bottles and 1000 mL bottles with screw caps. Prontosan Wound Irrigation Solution is used for cleaning wounds and for moistening and lubricating absorbent wound dressings for ulcers, burns, post- surgical wounds and abrasions. The mechanical action of moving across the wound provides for the mechanism of action and aids in the removal of foreign material such as dirt and debris. The Prontosan Wound Irrigation Solution Adaptor is a component that may be screwed on to the current Prontosan Wound Irrigation Solution containers (350 ml and 1000 ml bottles) and provides alternate spout geometry for which to access the solution inside the bottle. It is provided in sterile form, labeled for single patient use only.

{4}------------------------------------------------

SUBSTANTIAL EQUIVALENCE:

The subject device and the predicate device are exactly the same. No changes have been made to the device, manufacturing process, the packaging process or the intended use. The modification subject to this 510(k) is for labeling content only.

NON-CLINICAL TESTING:

No new non-clinical testing has been performed for this submission. The changes to the labeling are test results from testing previously performed on this product and submitted to the FDA in a prior submission.

CONCLUSION:

Prontosan® Wound Irrigation Solution for prescription use is identical to the predicate device in formulation, manufacturing, packaging and intended use. B. Braun believes the predicate and subject devices to be substantially equivalent.

N/A