K Number
K161468
Manufacturer
Date Cleared
2017-02-08

(257 days)

Product Code
Regulation Number
880.6500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.

Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions:

OrganismNameAverage Maximum log reduction /exposure time (hours)
Test temperature
45°F72°F110°F
BacteriaStaphylococcus4.88 / 24 hours4.02 / 0.33 hours4.20 / 0.33 hours
BacteriaEscherichia coli4.31 / 24 hours4.79 / 24 hours4.40 / 0.33 hours
VirusMS2 bacteriophage4.13 / 24 hours4.25 / 24 hours5.51 / 24 hours
VirusPhi-X1744.37 / 24 hours4.37 / 24 hours4.37 / 24 hours
Mold endosporeAspergillus Niger3.91 / 72 hours3.99 / 72 hours4.22 / 72 hours
Bacterial endosporeBacillus globigii4.11 / 72 hours4.41 / 72 hours4.41 / 72 hours
Device Description

Transformair® is a patented photo-electrochemical or photo-electrocatalyst (PEC) ultraviolet air purification technology that destroys bacteria in air in medical facilities. It consists of a pre-filter, black light lamps, and a catalytic filter coated with a photocatalyst. The Transformair device has a metal housing with electronic controls and outer dimensions of 28.5 inches by 21 inches by 12.25 inches that is installed in the duct of a buildings HVAC system. The Transformair device has a pre-filter and a chamber equipped with 6 low energy ultraviolet lights (UV-A), which emit light with wavelengths between 320-400 nm, and a catalytic filter.

AI/ML Overview

The provided document describes a medical device, the Transformair Indoor Air Purifier, and its performance data to establish substantial equivalence with a predicate device. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" as a separate set of pass/fail thresholds before the study. Instead, the study results themselves demonstrate the device's efficacy, and these reported results are presented as sufficient to establish substantial equivalence with the predicate device. The performance is measured in "Average Maximum log reduction" over specific exposure times and temperatures. The FDA's acceptance of these results "proves" the device meets the implicit criteria for efficacy for its intended use.

OrganismNameTest TemperatureExposure TimeReported Average Maximum Log Reduction
BacteriaStaphylococcus epidermidis45°F24 hours4.88
72°F0.33 hours4.02
110°F0.33 hours4.20
BacteriaEscherichia coli45°F24 hours4.31
72°F24 hours4.79
110°F0.33 hours4.40
VirusMS2 bacteriophage45°F24 hours4.13
72°F24 hours4.25
110°F24 hours5.51
VirusPhi-X174 bacteriophage45°F24 hours4.37
72°F24 hours4.37
110°F24 hours4.37
Mold endosporeAspergillus Niger45°F72 hours3.91
72°F72 hours3.99
110°F72 hours4.22
Bacterial endosporeBacillus globigii45°F72 hours4.11
72°F72 hours4.41
110°F72 hours4.41

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not explicitly state the numerical sample sizes for each organism tested. It refers to "aerosolized suspensions of the claimed microorganisms" and "kill kinetics for bacteria (Staphylococcus epidermidis and Escherichia coli), viruses (MS2 and Phi-X174), and mold spores (Aspergillus Niger) and bacterial spores (Bacillus globigii)." The implication is that multiple tests were conducted for each organism at various temperatures and exposure times to derive the "Average Maximum log reduction." However, the exact number of replicates or individual samples is not provided.
  • Data Provenance: The document states, "Third party testing was performed to evaluate filter performance using aerosolized suspensions of the claimed microorganisms... This testing was done in compliance with FDA Good Laboratory Practices (GLP) as defined in 40 CFR, Part 160." This indicates the tests were conducted in a controlled laboratory environment by an independent third party, likely within the US given the FDA GLP compliance. The data is prospective as it involves active testing of the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This is not applicable as the study involves direct microbiological testing of the device's ability to destroy microorganisms, not human interpretation of data where expert consensus would establish ground truth.

4. Adjudication Method for the Test Set:

Not applicable. The study involves direct measurement of microorganism reduction, not a judgmental interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. This type of study is typically associated with diagnostic algorithms involving human readers, which is not relevant to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, a standalone performance study was done. The performance results (log reduction of microorganisms) are based solely on the efficacy of the Transformair Indoor Air Purifier device itself under specified testing conditions. There is no human intervention in the device's direct performance of destroying bacteria.

7. The Type of Ground Truth Used:

The ground truth for the device's performance is established through direct microbiological testing against known concentrations of specific organisms (Staphylococcus epidermidis, Escherichia coli, MS2 bacteriophage, Phi-X174 bacteriophage, Aspergillus Niger, and Bacillus globigii). The "log reduction" represents the quantitative measure of the device's ability to destroy these organisms, serving as the ground truth for its germicidal efficacy.

8. The Sample Size for the Training Set:

This device is not an AI/ML algorithm that requires a training set. Its performance is based on physical and chemical processes (UV-A light activating a photocatalyst), which are evaluated through direct performance testing rather than training data.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, forming a symbolic representation of the department's mission related to health and human well-being.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 21, 2019

Transformair, LLC % Dave Yungvirt CEO Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041

Re: K161468

Trade/Device Name: Transformair Indoor Air Purifier Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: January 31, 2017 Received: February 1, 2017

Dear Dave Yungvirt:

This letter corrects our substantially equivalent letter of February 8, 2017

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161468

Device Name Transformair Indoor Air Purifier

Indications for Use (Describe)

The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.

Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions:

OrganismNameAverage Maximum log reduction /exposure time (hours)
Test temperature
45°F72°F110°F
BacteriaStaphylococcus4.88 / 24 hours4.02 / 0.33 hours4.20 / 0.33 hours
BacteriaEscherichia coli4.31 / 24 hours4.79 / 24 hours4.40 / 0.33 hours
VirusMS2 bacteriophage4.13 / 24 hours4.25 / 24 hours5.51 / 24 hours
VirusPhi-X1744.37 / 24 hours4.37 / 24 hours4.37 / 24 hours
Mold endosporeAspergillus Niger3.91 / 72 hours3.99 / 72 hours4.22 / 72 hours
Bacterial endosporeBacillus globigii4.11 / 72 hours4.41 / 72 hours4.41 / 72 hours

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "transformair" in a stylized font, with a series of blue dots and a yellow arc above it. Below the word, the address "3802 Spectrum Blvd. Suite 143, Tampa, FL 33612" is printed in a smaller, simpler font. The address provides a specific location in Tampa, Florida.

510(k) SUMMARY

Transformair Air Purifier

Submitter:

Company Name:Transformair, Inc.
Company Address:3802 Spectrum Blvd.Suite 143Tampa, FL 33612

Telephone Number: 352-871-3803 Contact Person: Jaya Rao Date Prepared: 1/19/2017

  • Device:
Proprietary Name:Transformair Indoor Air Purifier
Common or Usual Name:Air purifier ultraviolet or ultraviolet air purifier
Classification Name:Medical Ultraviolet Air Purifier
Regulation Number:21 CFR 880.6500
Product Code:FRA
Device Class:Class 2
Category:General Hospital

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "transformair" in a muted green color. Above the logo is a curved line of blue dots and a yellow line. Below the logo is the address "3802 Spectrum Blvd. Suite 143, Tampa, FL 33612".

Predicate Device:

The predicate device is shown in the following table:

Table 1: Predicate Device

Predicate DeviceABRACAIR Air Cleaners QTZ300-60 and TI 100-30P
ManufacturerAbracair, Inc.
510(k) NumberK052732
Regulation Number:21 CFR 880.6500
Regulation Name:Medical Ultraviolet Air Purifier
Regulatory ClassII
Product Code :FRA
Review Panel:General Hospital
Combination ProductNo

Intended Use / Indications for Use:

The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.

Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions:

Table 2. Transformair Lethality Testing Results

OrganismNameAverage Maximum log reduction /exposure time (hours)
Test temperature
45°F72°F110°F
BacteriaStaphylococcus epidermidis4.88/ 24 hours4.02 /0.33 hours4.20 /0.33 hours
BacteriaEscherichia coli4.31/ 24 hours4.79 /24 hours4.40 /0.33 hours
VirusMS2 bacteriophage4.13/ 24 hours4.25/ 24 hours5.51/ 24 hours
VirusPhi-X174 bacteriophage4.37/ 24 hours4.37/ 24 hours4.37/ 24 hours
Mold endosporeAspergillus Niger3.91/ 72 hours3.99/ 72 hours4.22/ 72 hours
Bacterial endosporeBacillus globigii4.11/ 72 hours4.41/ 72 hours4.41/ 72 hours

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for "transformair" in a muted gray color. Above the logo is a curved line of blue dots and a yellow line. Below the logo is the address "3802 Spectrum Blvd. Suite 143, Tampa, FL 33612".

The Transformair Indoor Air Purifier is for Over-The-Counter use.

Device Description:

Transformair® is a patented photo-electrochemical or photo-electrocatalyst (PEC) ultraviolet air purification technology that destroys bacteria in air in medical facilities. It consists of a pre-filter, black light lamps, and a catalytic filter coated with a photocatalyst. The Transformair device has a metal housing with electronic controls and outer dimensions of 28.5 inches by 21 inches by 12.25 inches that is installed in the duct of a buildings HVAC system. The Transformair device has a pre-filter and a chamber equipped with 6 low energy ultraviolet lights (UV-A), which emit light with wavelengths between 320-400 nm, and a catalytic filter.

Technological Characteristics and Principles of Operation:

The Transformair Air Purifier is a medical ultraviolet air purifier unit that traps microorganisms on a filter where they are destroyed by UV photo-catalyzed generation of free radicals in the validated time periods. The air cleaning chamber consists of a metal housing that contains a non-woven polyester prefilter, glass UV-A lamps (~320-400 nm), a catalytic filter. The metal housing has electronics that plug into a standard outlet (120/220 Volt) to power the UV-A lamps.

The Transformair Air Purifier destroys microbiological contaminants in the air through a photochemical reaction that produces hydroxyl free radicals when UV-A light rays are shone on the catalytic surface.

Substantial Equivalence:

The Transformair Air Purifier is substantially equivalent to the ABRACAIR Air Cleaner (K052732) since the devices have the same general intended use and similar indications, technological characteristics, and principles of operation. A reference table is provided below comparing the Transformair Air Purifier to the predicate.

Element ofComparisonTransformairABRACAIR (K052732)
ManufacturerTransformair, Inc.Abracair, Inc.
Device typeMedical Ultraviolet Air PurifierMedical Ultraviolet Air Purifier
Element ofComparisonTransformairABRACAIR (K052732)
Regulation Number21 CFR 880.650021 CFR 880.6500
Intended Use /Indications for UseThe Transformair Indoor Air Purifier, InDuct Model 16108 is a device intendedfor medical purposes that is used todestroy bacteria in the air by exposure toultraviolet radiation.Transformair Indoor Air Purifier, In DuctModel 16108 has been demonstrated todestroy Staphylococcus epidermidis,Escherichia coli, MS2, Phi-X174,Aspergillus Niger and Bacillus globigiientrained on the filter of the subjectdevice under the exposure conditionsdescribed in Table 2.The Transformair Indoor Air Purifier isfor Over-The-Counter use.The Abracair Air Cleaner is intended forthe reduction of aerosolized mold andbacteria within hospitals, nursing homes,medical facilities. The device may beused in occupied spaces within hospitalenvironments such as baby and/orneonatal nurseries, hospital rooms,operating rooms, mortuaries, embalmingrooms.
Mechanism ofactionUV light of sufficient energy (UV-A)activates photocatalyst that destroysmicroorganisms entrained on the filterthrough a photochemical reaction thatis enhanced by a metal wire mesh.UV light of sufficient energy (UV-C)activates photocatalyst that destroysmicroorganisms through aphotochemical reaction; UV-C lightdirectly destroys microorganism.
Model(s)Model 16108QTZ300-60 and QTZ100-24
InstallationIn-DuctFree standing
Filter(s)Pre-filter made with polyester andcatalytic filter made with polyester andcoated with titanium dioxide (TiO2)Filter made with quartz fibers coated withtitanium dioxide (TiO2)
PhotocatalystCatalytic filter coated with TiO2 andmetal wire meshFilter coated with TiO2
Light sourceUV-A lamps (~320-400 nm)UV-C Xenon Flash lamps (~100-285nm)
Element ofComparisonTransformairABRACAIR (K052732)
VoltageUp to 120 Volt / 220 Volt (plugs intostandard outlet) (See Footnote1)240 Volt AC / 480 Volt AC
CurrentUp to 1.25 amps (See Footnote1)30 Amps at 240 Volts
Power ConsumptionUp to 240 Watts (See Footnote1) andFootnote2)Up to 7020 Watts
Air cleaningchamberParticulate pre-filter, UV-A lamps and acatalytic filter coated with TiO2 andmetal wire meshUV-C lamps and a catalytic filter coatedwith TiO2
ElectronicsUL compliantUL compliant
Dimensions28.5 inches by 21 inches by 12.25 inchesNot available

TABLE 3: SUBSTANTIAL EQUIVALENCE COMPARISON CHART

{6}------------------------------------------------

{7}------------------------------------------------

3802 Spectrum Blvd. Suite 143, Tampa, FL 33612

ً This does not raise any new issues of safety and efficacy since the Transformair Air Purifier complies with the applicable UL standards for an air duct mounted device.

ً Transformair Air Purifier is different from the primary predicate device since it uses upto 240 W of UV-A light (320-400 nm) with a TiO-coated catalytic filter and a metal wire mesh to enhance activation of the TiO2 photocatalyst instead of upto 7020 W of UV-C light (~100-285nm) used by the primary predicate. This does not raise any new issues of efficacy since the ability of the Transformair device to reduce the concentration of airborne microorganisms is substantiated by the performance testing.

The Transformair Air Purifier and the predicate ABRACAIR Air Cleaners use the action of UV light on a catalytic filter coated with a Titanium Oxide (TiO2) photocatalyst to destroy microbiological contaminants in the air. Both devices use conventional UV-light lamps and filter materials but, unlike the predicate, the Transformair Air Purifier uses UV-A lamps instead of UV-C lamps and has a metal wire mesh on the catalytic filter. Both devices are designed to operate in a medical facility.

The Transformair air purifier has the same intended use of reducing airborne bacteria and mold spores as the predicate device ABRACAIR Air Cleaner (K052732). In addition, the Transformair air purifier is intended to reduce viruses (MS2 and Phi-X174).

The differences in the indications for use and technological characteristics are supported by the performance testing and UL electrical safety testing and do not raise new questions of safety or efficacy.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Transformair, a company located at 3802 Spectrum Blvd. Suite 143, Tampa, FL 33612. The logo features the company name in a stylized font, with a series of blue dots and a yellow wave above the name. The address is printed in a smaller font below the logo.

Non-Clinical Performance Testing Summary:

No performance standards have been established for such devices under Sections 514 of the Federal Food, Drug and Cosmetic Act. However, the Transformair Air Purifier has been tested to demonstrate that the device does not generate Ozone. It has also been certified compliant with the following UL standards:

  • । ANSI/UL 1995-2011 & CAN-CSA C22.2 No. 236-11, Heating and Cooling Equipment and
  • -UL 2043, Heat and Visible Smoke Release For Discrete Products and Their Accessories Installed in Air-Handling Spaces.

Reference the Table below for specific UL tests conducted.

Test ConductedUL Reference StandardTest Results
Input Test (Circuits – Other than ElectricHeaters) (+)UL 1995, 4th Ed., Section 40PASS
Starting TestUL 1995, 4th ed. Section 69PASS
Leakage Current TestUL 1995, 4th ed. Section 68PASS
Temperature Operation Test (Without AnySupplementary Heating Means)(+)UL 1995, 4th Ed., Section 41PASS
Dielectric Voltage-Withstand TestUL 1995, 4th Ed., Section 54PASS
Fire Test for Heat and Visible Smoke ReleaseFor Discrete Products and Their AccessoriesInstalled in Air-Handling SpacesUL 1995, 4th Ed., Section 18 and UL2043, 3rd Ed.PASS

Table 4: Underwriters Laboratories Testing Summary

Third party testing was also performed to evaluate filter performance using aerosolized suspensions of the claimed microorganisms. This testing demonstrated retention of the test organisms on the filter. This testing was done in compliance with FDA Good Laboratory Practices (GLP) as defined in 40 CFR, Part 160.

Lastly, third party testing was performed to establish the kill kinetics for bacteria (Staphylococcus epidermidis and Escherichia coli), viruses (MS2 and Phi-X174), and mold spores (Aspergillus Niger) and bacterial spores (Bacillus globigii) over the range of temperatures expected to occur in an HVAC duct. This testing was done in compliance with FDA Good Laboratory Practices (GLP) as defined in 40 CFR, Part 160. Reference Table below for the lethality testing results.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Transformair, a company located at 3802 Spectrum Blvd. Suite 143, Tampa, FL 33612. The logo features the company name in a stylized font, with a series of blue dots and a yellow line above the name. The address is printed in a smaller font below the logo.

Table 5: Transformair Device Lethality Testing Results

CategoryAerosolizedBiologicalTest Temperature(°F)Exposure time(hours)Average Maximum logreduction
BacteriaStaphylococcusepidermidis45°F24 hours4.88
72°F0.33 hour4.02
110°F0.33 hour4.20
BacteriaEscherichiacoli45°F24 hours4.31
72°F24 hours4.79
110°F0.33 hour4.40
VirusMS2bacteriophage45°F24 hours4.13
72°F24 hours4.25
110°F24 hours5.51
VirusPhi-X174bacteriophage45°F24 hours4.37
72°F24 hours4.37
110°F24 hours4.37
MoldsporeAspergillusNiger spores45°F72 hours3.91
72°F72 hours3.99
110°F72 hours4.22
BacterialsporeBacillusglobigii spores45°F72 hours4.11
72°F72 hours4.41
110°F72 hours4.41

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for "transformair" in a stylized font. Below the logo is the address "3802 Spectrum Blvd, Suite 143, Tampa, FL 33612". The logo has a series of blue dots and a yellow swoosh above the word "transformair".

Clinical Testing Summary:

Clinical testing has not been performed for the Transformair Air Purifier. Non-clinical bench testing is sufficient to demonstrate substantial equivalence to the predicate ABRACAIR Air Cleaner.

Conclusion:

The subject device is substantially equivalent to the predicate device.

§ 880.6500 Medical ultraviolet air purifier.

(a)
Identification. A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.(b)
Classification. Class II (performance standards).