K Number
K161074
Date Cleared
2016-11-17

(213 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder-Free Nitrile Examination Glove Flock-Lined, Black Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.

Device Description

Powder Free Nitrile Examination Gloves Flock-Lined, Black Color Non-Sterile, is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. The gloves are lined with cotton flock, ambidextrous with beaded cuff, black colored, single use disposable devices that comes is six sizes (XS, S, M, L, and XL) supplied in non-sterile state. The gloves are made of synthetic rubber, designed and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical and performance characteristics meet all requirements of ASTM D6319-10. Powder was not used for this glove's manufacturing process.

AI/ML Overview

This document describes a 510(k) premarket notification for "Powder Free Nitrile Examination Gloves Flock-Lined, Black Color." The submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

CharacteristicsAcceptance Criteria (Standards)Reported Device Performance
DimensionsASTM D6319-10Meets
Physical Properties (Before/After Aging)ASTM D6319-10Meets
Freedom from pinholesASTM D6319-10 / ASTM D5151-11Meets
Powder ResidualASTM D6319-10Meets <2mg/glove
Biocompatibility: Primary Skin IrritationISO 10993-10:2010Not a primary skin irritant
Biocompatibility: Dermal SensitizationISO 10993-10:2010Not a contact sensitizer

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not specify sample sizes for the testing of the new device for dimensions, physical properties, freedom from pinholes, or powder residual. It only states "Meets" for these criteria.

For biocompatibility, the tests were conducted according to ISO 10993-10:2010. The specific number of rabbits used for primary skin irritation or the number of subjects for dermal sensitization tests is not mentioned.

The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated in relation to the specific tests performed for the new device. However, the manufacturer is stated as "ZHANGJIAGANG DAYU RUBBER PRODUCTS CO., LTD" in ZHANGJIAGANG CITY, CHINA. This suggests the testing data likely originated from China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable and not provided in the document. The device in question is a medical glove, and its performance is evaluated against established ASTM and ISO standards for physical characteristics and biocompatibility, not through human expert assessment of a "ground truth" like in imaging or diagnostic AI.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This information is not applicable and not provided. The testing involves objective physical and chemical measurements against pre-defined standards, not a subjective assessment requiring adjudication among experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic or AI-assisted devices that involve human interpretation. This document pertains to a medical glove.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable as the device is a medical glove, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context is defined by the established ASTM and ISO standards for nitrile examination gloves. These standards specify acceptable ranges and methodologies for evaluating dimensions, physical properties (tensile strength, elongation), freedom from pinholes, powder residue, and biocompatibility.

8. The sample size for the training set

This is not applicable. The device is a physical product (medical glove) and not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 17, 2016

Shen Wei (USA), Inc. % Ms. Cheryl Bailey-Kroll Compliance Manager 33278 Central Ave. Suite 102 Union City, California 94587

Re: K161074

Trade/Device Name: Powder Free Nitrile Examination Gloves Flock-Lined, Black Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: October 12, 2016 Received: October 18, 2016

Dear Cheryl Bailey-Kroll,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K161074

Device Name

Powder-Free Nitrile Examination Glove Flock-Lined, Black Color

Indications for Use (Describe)

Powder-Free Nitrile Examination Glove Flock-Lined, Black Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is labeled "Attachment One" and features a logo for "Shen Wei usa inc." The logo includes a green hand graphic above the text. Below the company name is the tagline "PROMOTING YOUR SUCCESS!" in a smaller font size.

510(k) Summary

1.0 Submitter:

Name:Shen Wei USA Inc.
Address:33278 Central Ave. Suite 102Union City, CA. 94587
FDA Est Reg#:2939388
Phone No:510-429-8692
Fax No:510-487-5347

Manufacturer:

Name:Address:ZHANGJIAGANG DAYU RUBBER PRODUCTS CO., LTDXIZHANG TOWNZHANGJIAGANG CITY, CHINA 215614
FDA Est Reg#:1000254914
Phone No:520-845-0023
Fax No:520-845-0311

Date 510k Summary was prepared: October 10, 2016

2.0 Contact Person:

Name:Cheryl Reep
Phone No:510-429-8692
Fax No:510-487-5347
Email:Creep@shenweiusa.com

3.0 Name of the device:

Trade Name:Powder Free Nitrile Examination Gloves Flock-Lined, Black Color
Device Name:Powder Free Nitrile Examination Gloves Flock-Lined, Black Color
Common Name:Patient Examination Gloves
Classification Name:Patient examination glove
Device Class:Class 1
Regulation Number:21 CFR 880.6250
Product Code:80 LZA

4.0 Predicate Device Information:

Device Name: Non Sterile, Powder Free Nitrile Examination Glove, Blue with Polymer Coating Company Name: Siam Sempermed Corp. Ltd. 510K Number: K070639

5.0 Description of the Device:

Powder Free Nitrile Examination Gloves Flock-Lined, Black Color Non-Sterile, is a disposable device intended for

{4}------------------------------------------------

medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. The gloves are lined with cotton flock, ambidextrous with beaded cuff, black colored, single use disposable devices that comes is six sizes (XS, S, M, L, and XL) supplied in non-sterile state.

The gloves are made of synthetic rubber, designed and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical and performance characteristics meet all requirements of ASTM D6319-10. Powder was not used for this glove's manufacturing process.

6.0 Labeling and Intended Use of the Device:

The Powder Free Nitrile Examination Gloves Flock-Lined, Black Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.

7.0 Shelf Life or Expiration Data: Not Applicable

8.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Examination Gloves Flock-Lined, Black Color is summarized with the following technological characteristics compared to ASTM or equivalent standards.

CharacteristicsStandardsDevice Performance
DimensionsASTM D6319-10Meets
Physical Properties(Before/After Aging)ASTM D6319-10Meets
Freedom from pinholesASTM D6319-10 / ASTMD5151-11Meets
Powder ResidualASTM D6319-10Meets <2mg/glove
BiocompatibilityISO 10993-10:2010:i. Primary Skin Irritationin Rabbitsi. Not a primary skin irritant
ii. Dermal Sensitizationii. Not a contact sensitizer

9.0 Device Interior Lining and Manufacturing Process:

The Powder Free Nitrile Examination Gloves Flock-Lined, Black Color is sprayed with a white cotton flock on the interior of the glove to help facilitate donning.

10.0 Substantial Equivalence Based On Assessment of Non-Clinical Performance Data

Comparison to Predicate- The subject device Powder Free Nitrile Examination Gloves Flock-Lined, Black Color compares favorably to the predicate device, Non Sterile, Powder Free Nitrile Examination Glove, Blue with Polymer Coating, as indicated in the tabulated summary below:

CharacteristicsPredicate DeviceProposed New DeviceComments
NameNon Sterile, Powder FreeNitrile Examination Glove,Blue with Polymer CoatingPowder Free NitrileExamination Gloves Flock-Lined, Black ColorN/A
510K NumberK070639K161074N/A
CompanySiam Sempermed Corp. Ltd.Shen Wei USA Inc.N/A
Device Description/Regulation NumberPatient examination glove /21 CFR 880.6250Patient examination glove /21 CFR 880.6250Substantially Equivalent
Product CodeLZALZASubstantially Equivalent
Intended useIntended for medicalpurpose that is worn on theexaminer's hand to preventcontamination betweenIntended for medicalpurpose that is worn on theexaminer's hand to preventcontamination betweenpatient and examiner. Thisdevice is single use only.Substantially Equivalent
Instruction for UseA disposable deviceintended for medicalpurpose that is worn on theexaminer's hand to preventcontamination betweenpatient and examiner. Thisdevice is single use only.A disposable deviceintended for medicalpurpose that is worn on theexaminer's hand to preventcontamination betweenpatient and examiner. Thisdevice is single use only.Substantially Equivalent
Non-SterileNon-SterileNon-SterileSubstantially Equivalent
MaterialsNitrileNitrileSubstantially Equivalent
ColorBlueBlackDifference
DesignAmbidextrous in differentsizes per ASTM 6319-10-dimension requirementAmbidextrous in differentsizes per ASTM 6319-10-dimension requirementSubstantially Equivalent
Lining / CoatingPolymer CoatingFlock LiningDifference
Single Patient UseSingle Patient UseSingle Patient UseSubstantially Equivalent
Performance:
I. Freedom from HolesSubstantially Equivalent
II. DimensionMeets ASTM D6319-10Meets ASTM D6319-10Substantially Equivalent
III. Physical PropertiesSubstantially Equivalent
IV. Powder ResidueSubstantially Equivalent
Single UseYesYesSubstantially Equivalent
Biocompatibility TestPasses:i. Primary Skin IrritationTest ii. Dermal SensitizationTestPasses:i. Not a primary skin irritantii. Not a contact sensitizerSubstantially Equivalent

Substantial Equivalence Comparison Table

{5}------------------------------------------------

Summary of Difference and Comparison of Safety and Effectiveness- The subject device differs from the predicate in that:

    1. The subject device is color Black while the predicate device is color Blue. Difference in color does not affect subject device's safety and effectiveness; subject device was tested and met requirements for Biocompatibility and ASTM standards.
    1. The subject device is manufactured with the inner surface coated or lined with flock to facilitate donning. The predicate device is manufactured with the inner surface coated with polymer to assist in donning. Though the materials of construction differ, the subject device's material is functionally equivalent to those of the cited predicate.
  • Summary of Clinical Testing: Not applicable. No clinical testing was performed. 11.
  1. Conclusion: The conclusion drawn from the nonclinical test demonstrate that the subject device, Powder Free Nitrile Examination Gloves Flock-Lined, Black Color, is as safe, as effective, and performs as well as the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.