AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AssureTech Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-△9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

AssureTech Methamphetamine Tests are immunochromatographic assays for the qualitative determination of d-Methamphetamine in human urine at cut-off concentration of 1000 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

AssureTech Phencyclidine Tests are immunochromatographic assays for the qualitative determination of Phencyclidine in human urine at cut-off concentration of 25 ng/mL. The tests are available in a Strip format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

Device Description

The AssureTech Methamphetamine Tests, AssureTech Phencyclidine Tests, and AssureTech Marijuana Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Methamphetamine, Phencyclidine and Marijuana (target analytes) in human urine. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.

AI/ML Overview

The provided text describes the analytical and user performance of various AssureTech drug tests (Methamphetamine, Phencyclidine, and Marijuana) across different formats (Strip, Dip Card, Cup, Turn Key Split Cup). The information outlines specific studies conducted to demonstrate the device meets acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in percentages (e.g., "must achieve 90% accuracy"). Instead, it presents results for concentrations at or around the cut-off, and above/below it, demonstrating the device's ability to correctly identify positive and negative samples.

Here's a table summarizing the reported device performance for Methamphetamine, Phencyclidine, and Marijuana tests, based on the precision and comparison studies, which effectively serve as a demonstration of meeting implied acceptance criteria for qualitative immunoassay performance.

Implicit Acceptance Criteria: The device should consistently provide correct qualitative results (positive or negative) around and beyond the defined cut-off concentrations, as confirmed by a gold standard (GC/MS). For concentrations significantly below the cut-off, the device should report negative, and for concentrations significantly above, it should report positive. At the cut-off itself, there might be some variability, but a strong trend toward expected results is desired. The lay-user study indicates a high percentage of correct results.

Reported Device Performance (Excerpted and Summarized):

TestAnalyteCut-off (ng/mL)Performance at -25% Cut-off (GC/MS)Performance at +25% Cut-off (GC/MS)Overall Expert Review Performance (Comparison Study)Lay User Performance (Overall Correctness)
AssureTech Methamphetamined-Methamphetamine1000Precision: 50-/0+ for all formats and lots (all negative as expected). Comparison Study: High number of negative reads (e.g., Viewer A Strip: 9 negative samples at or near cutoff).Precision: 50+/0- for all formats and lots (all positive as expected). Comparison Study: High number of positive reads (e.g., Viewer A Strip: 14 positive samples at or near cutoff).High agreement with GC/MS. Few discordant results mainly near cutoff, with some negative readings for samples slightly above cutoff, and some positive readings for samples slightly below cutoff. (e.g., Methamphetamine Strip Viewer A: 974 ng/mL (negative) read positive, 1045 ng/mL (positive) read negative.)Strip: 100% correct across various predefined concentrations, except for a few discordant results in the comparison studies by lab assistants. Dip Card: ~95-100% correct near cutoff for some users. Turn Key Split Cup: ~95-100% correct near cutoff for some users. Quick Cup: ~95-100% correct near cutoff for some users.
AssureTech PhencyclidinePhencyclidine25Precision: 50-/0+ for all formats and lots (all negative as expected). Comparison Study: High number of negative reads (e.g., Viewer A Strip: 9 negative samples at or near cutoff).Precision: 50+/0- for all formats and lots (all positive as expected). Comparison Study: High number of positive reads (e.g., Viewer A Strip: 14 positive samples at or near cutoff).High agreement with GC/MS. Few discordant results mainly near cutoff, with some negative readings for samples slightly above cutoff, and some positive readings for samples slightly below cutoff. (e.g., Phencyclidine Strip Viewer A: 22 ng/mL (negative) read positive, 27 ng/mL (positive) read negative.)Strip: 95-100% correct near cutoff. Dip Card: 95-100% correct near cutoff. Turn Key Split Cup: 90-100% correct near cutoff. Quick Cup: 90-100% correct near cutoff.
AssureTech Marijuana11-Nor-△9-Tetrahydrocannabinol-9-COOH50Precision: 50-/0+ for all formats and lots (all negative as expected). Comparison Study: High number of negative reads (e.g., Viewer A Strip: 9 negative samples at or near cutoff).Precision: 50+/0- for all formats and lots (all positive as expected). Comparison Study: High number of positive reads (e.g., Viewer A Strip: 14 positive samples at or near cutoff).High agreement with GC/MS. Few discordant results mainly near cutoff, with some negative readings for samples slightly above cutoff, and some positive readings for samples slightly below cutoff. (e.g., Marijuana Strip Viewer A: 46 ng/mL (negative) read positive, 52 ng/mL (positive) read negative.)Strip: 95-100% correct near cutoff. Dip Card: 90-100% correct near cutoff. Turn Key Split Cup: 95-100% correct near cutoff. Quick Cup: 95-100% correct near cutoff.

2. Sample Sizes and Data Provenance:

  • Precision Studies: Samples were prepared at various concentrations around the cut-off (-100%, -75%, -50%, -25%, cut-off, +25%, +50%, +75%, +100%). For each concentration level, 50 tests were performed (2 runs per day for 25 days). This implies 450 tests per lot per device type for each drug (9 concentration levels * 50 tests). With 3 lots per device and 4 device formats for each of the 3 drugs, the total number of precision tests is substantial (e.g., 450 * 3 lots * 4 formats * 3 drugs = 16,200).

  • Comparison Studies (Expert Review): For each drug and each device format, 80 unaltered clinical samples were used (40 negative and 40 positive). These were "blind labeled."

  • Lay-user Study: 1638 lay persons participated. They were given blind-labeled urine samples from various concentration levels (Negative, +/-75%, +/-50%, +/-25% of the cutoff). The tables show results for 21 samples per concentration level, for each drug and each device type. Considering there are 7 concentration levels listed, and assuming all 4 formats for all 3 drugs were tested, this amounts to a significant number of samples tested by lay users (21 samples/level * 7 levels * 4 formats * 3 drugs = 1,764 total sample tests, which broadly aligns with "1638 lay persons").

  • Data Provenance: The document states "Method comparison studies... were performed in-house..." and "Lay-user study was performed at three intended user sites." The country of origin is not explicitly stated for the clinical samples, but the submitting company is based in Hangzhou, China. The studies appear to be prospective as samples were prepared at specific concentrations and tested.

3. Number of Experts and Qualifications:

  • Expert Review (Comparison Study): Three laboratory assistants ("Viewer A", "Viewer B", "Viewer C") were used to read the devices. Their specific qualifications (e.g., years of experience) are not provided in this excerpt, only their role as "laboratory assistants."
  • Ground Truth Establishment: The ground truth for the test set (clinical urine samples) in the comparison studies was established using Gas Chromatography/Mass Spectrometry (GC/MS), which is stated as the "preferred confirmatory method" and considered a gold standard for drug detection. For precision studies, samples were "prepared by spiking drug in negative samples" and "Each drug concentration was confirmed by GC/MS."

4. Adjudication Method for the Test Set:

  • No explicit adjudication method (e.g., 2+1, 3+1 consensus) is described for the "Viewer" results in the comparison studies. The results for each "Viewer" are presented independently. Discordant results are specifically listed, indicating individual discrepancies rather than a pooled or adjudicated outcome from the laboratory assistants.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No MRMC comparative effectiveness study was done in the context of human readers improving with AI vs. without AI assistance. This document describes the performance of a standalone in vitro diagnostic device (immunochromatographic assay) and its comparison to a gold standard (GC/MS), along with a lay-user study. It does not involve AI assistance for human readers.

6. Standalone (Algorithm Only) Performance:

  • The devices themselves are qualitative immunochromatographic assays which provide a visual result (line appears or not). Therefore, their performance, as described in the precision and comparison studies, is essentially standalone performance. There isn't an "algorithm" in the typical AI sense; the "algorithm" is the biochemical reaction and visual interpretation of the test strips/cups. The results tabulated for the "Viewers" (laboratory assistants) and "Lay persons" represent human interpretation of these standalone devices.

7. Type of Ground Truth Used:

  • The primary ground truth used for validating the device performance is Gas Chromatography/Mass Spectrometry (GC/MS). This is a highly accurate chemical method for confirming the presence and concentration of specific substances in urine samples.
  • For the lay-user study, the ground truth was also established by GC/MS of the prepared urine samples.

8. Sample Size for the Training Set:

  • This document describes performance validation studies for a medical device (AssureTech drug tests), not an AI algorithm. Therefore, there is no "training set" in the context of machine learning. The device operates based on fixed immunochromatographic principles. The "precision" and "comparison" studies serve as validation of the device's inherent design and manufacturing consistency.

9. How the Ground Truth for the Training Set was Established:

  • As there is no AI algorithm training set, this question is not applicable. The device's operational principles are based on biochemical interactions, not learned patterns from a training dataset.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 6 2016

ASSURE TECH (HANGZHOU) CO., LTD. SHISHENG LING GENERAL MANAGER BUILDING 1, NO.10, XIYUANSAN ROAD WESTLAKE ECONOMIC ZONE HANGZHOU, ZHEJIANG 310030, CHINA

Re: K161044

Trade/Device Name: AssureTech Methamphetamine Strip Test AssureTech Methamphetamine Cup Test AssureTech Methamphetamine Dip Card Test AssureTech Methamphetamine Turn Key Split Cup Test AssureTech Phencyclidine Strip Test AssureTech Phencyclidine Cup Test AssureTech Phencyclidine Dip Card Test AssureTech Phencyclidine Turn Key Split Cup Test AssureTech Marijuana Strip Test AssureTech Marijuana Cup Test AssureTech Marijuana Dip Card Test AssureTech Marijuana Turn Key Split Cup Test Regulation Number: 21 CFR §862.3610 Regulation Name: Methamphetamine Test System Regulatory Class: II Product Code: DJC, LCM, LDJ Dated: April 09, 2016 Received: April 13, 2016

Dear Shisheng Ling:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161044

Device Name

AssureTech Marijuana Strip Test AssureTech Marijuana Dip Card Test AssureTech Marijuana Cup Test; AssureTech Marijuana Turn Key Split Cup Test

Indications for Use (Describe)

AssureTech Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-△9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K161044

Device Name

AssureTech Methamphetamine Strip Test AssureTech Methamphetamine Dip Card Test AssureTech Methamphetamine Cup Test; AssureTech Methamphetamine Turn Key Split Cup Test

Indications for Use (Describe)

AssureTech Methamphetamine Tests are immunochromatographic assays for the qualitative determination of d-Methamphetamine in human urine at cut-off concentration of 1000 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K161044

Device Name

AssureTech Phencyclidine Strip Test AssureTech Phencyclidine Dip Card Test AssureTech Phencyclidine Cup Test; AssureTech Phencyclidine Turn Key Split Cup Test

Indications for Use (Describe)

AssureTech Phencyclidine Tests are immunochromatographic assays for the qualitative determination of Phencyclidine in human urine at cut-off concentration of 25 ng/mL. The tests are available in a Strip format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) SUMMARY

1. Date:May 27, 2016
2. Submitter:Assure Tech. (Hangzhou) Co., Ltd.Building 1, No.10, Xiyuansan Road, WestlakeEconomic ZoneHangzhou, China, 310030
3. Contact person:Shisheng LingAssure Tech. (Hangzhou) Co., Ltd.Building 1, No.10, Xiyuansan Road, WestlakeEconomic ZoneHangzhou, China, 310030Telephone: +86-571-88868960Fax: +86-571-81022698Email: ling@diareagent.com
4. Device Name:AssureTech Methamphetamine Strip TestAssureTech Methamphetamine Cup TestAssureTech Methamphetamine Dip Card TestAssureTech Methamphetamine Turn Key Split Cup TestAssureTech Phencyclidine Strip TestAssureTech Phencyclidine Cup TestAssureTech Phencyclidine Dip Card TestAssureTech Phencyclidine Turn Key Split Cup TestAssureTech Marijuana Strip TestAssureTech Marijuana Cup TestAssureTech Marijuana Dip Card TestAssureTech Marijuana Turn Key Split Cup Test
Classification:Class II
Product CodeCFR #Panel
DJC21 CFR, 862.3610 Methamphetamine Test SystemToxicology
LCM21 CFR, 862.3100 Amphetamine Test SystemToxicology
LDJ21 CFR, 862.3870 Cannabinoids Test SystemToxicology
    1. Predicate Devices: K142396
      The Chemtrue® Multi-Panel Drug Screen Dip Card Tests
    1. Intended Use
      AssureTech Methamphetamine Tests are immunochromatographic assays for the qualitative determination of d-Methamphetamine in human urine at cut-off concentration of 1000 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

{6}------------------------------------------------

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

AssureTech Phencyclidine Tests are immunochromatographic assays for the qualitative determination of Phencyclidine in human urine at cut-off concentration of 25 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

AssureTech Marijuana Tests are immunochromatographic assays for the qualitative determination of 11-Nor-A9-Tetrahydrocannabinol-9-COOH in human urine at cut-off concentration of 50 ng/mL. The tests are available in a Strip format, a Cup format, a Dip Card format and a Turn Key Split Cup format.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

    1. Device Description
      The AssureTech Methamphetamine Tests, AssureTech Phencyclidine Tests, and AssureTech Marijuana Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of d-Methamphetamine, Phencyclidine and Marijuana (target analytes) in human urine. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
    1. Substantial Equivalence Information
      A summary comparison of features of the AssureTech Methamphetamine Tests, AssureTech Phencyclidine Tests and AssureTech Marijuana Tests and the predicate devices is provided in following tables.

Table 1: Features Comparison of AssureTech Methamphetamine Tests and the Predicate Devices

ItemDevicePredicate - K142396
Indication(s)for UseFor the qualitative determination of drugs ofabuse in human urine.Same (but the number ofdrugs detected is different)

{7}------------------------------------------------

Calibratord-MethamphetamineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values1000 ng/mLSame
Intended UseFor over-the-counter and prescription uses.Same
ConfigurationsStrip, Dip Card, Cup, Turn Key Split CupDip Card

Table 2: Features Comparison of AssureTech Phencyclidine Tests and the Predicate Devices

ItemDevicePredicate - K142396
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
CalibratorPhencyclidineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Cut-Off Values25 ng/mLSame
Intended UseFor over-the-counter and prescription uses.Same
ConfigurationsStrip, Dip Card, Cup, Turn Key Split CupDip Card

Table 3: Features Comparison of AssureTech Marijuana Tests and the Predicate Devices

ItemDevicePredicate - K142396
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
Calibrator11-Nor-Δ9-Tetrahydrocannabinol-9-COOHSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based on theprinciple of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame

{8}------------------------------------------------

Specimen TypeHuman UrineSame
Cut-Off Values50 ng/mLSame
Intended UseFor over-the-counter and prescription uses.Same
ConfigurationsStrip, Dip Card, Cup, Turn Key Split CupDipCard
    1. Test Principle
      The AssureTech Methamphetamine Tests, AssureTech Phencyclidine Tests, and AssureTech Marijuana Tests are rapid tests for the qualitative detection of d-Methamphetamine, Phencyclidine and Marijuana in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibodycoated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. They are 0, 255, 509, 747, 1011, 1265, 1518, 1760 and 2015 ng/mL for Methamphetamine, 0, 7, 13, 20, 27, 33, 39, 45 and 53ng/mL for Phencyclidine, and 0, 14, 28, 39, 53, 61, 73, 90 and 103ng/mL for Marijuana. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables.

AssureTech Methamphetamine Tests

Strip

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+9-/41+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+9-/41+50+/0-50+/0-50+/0-50+/0-

Dip Card

LotNumber-100%cut off-75%cut off-50%cut off-25%cut offcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-

{9}------------------------------------------------

Lot 250-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Turn Key Split Cup
Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Quick Cup
Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off

12-/38+

9-/41+

11-/39+

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50+/0-

50-/0+

50-/0+

50-/0+

AssureTech Phencyclidine Tests

50-/0+

50-10+

50-10+

50-/0+

50-/0+

50-/0+

50-/0+

50-/0+

50-/0+

Lot 1

Lot 2

Lot 3

Strip
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Dip Card
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Tum Key Split Cup
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Quick Cup
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-

AssureTech Marijuana Tests

Strip

LotNumber-100%cut off-75%cut off-50%cut off-25%cut offcut off+25%cut off+50%cut off+75%cut off+100%cut off
------------------------------------------------------------------------------------------------------------------------------------------------------------------

{10}------------------------------------------------

Lot 150-/0+50-/0+50-/0+50-/0+9-/41+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Dip Card
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+8-/42+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Tum Key Split Cup
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+11-/39+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Quick Cup
LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+12-/38+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+10-/40+50+/0-50+/0-50+/0-50+/0-

b. Linearity

Not applicable.

c. Stability

The devices are stable at 39.2°F to 86°F (4°C to 30°C) for 24 months based on the accelerated stability study at 45 ℃ and real time stability determination at both 4 ℃ and 30 °C.

d. Cut-off

A total of 150 samples of which 25 are native samples and 125 were spiked samples are used in the cut-off study. These samples are equally distributed at concentrations of -50% Cut-Off; -25% Cut-Off; +25% Cut-Off; +50% Cut-Off; +50% Cut-Off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% Cut-off and all negative at and below -25% Cut-off for Methamphetamine, Phencyclidine and Marijuana .

The following cut-off values for the candidate devices have been verified.

CalibratorCut-off (ng/mL)
d-Methamphetamine1000
Phencyclidine25
Marijuana50

e. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using

{11}------------------------------------------------

three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different devices.

Methamphetamine:
Acetamidophenol(+/-)-EphedrineOxymetazoline
AcetaminophenErythromycinPapaverine
Acetophenetidinβ-EstradiolPenicillin G
N-AcetylprocainamideEstrone-3-sulfatePentobarbital
AcetylsalicylateFenoprofenPerphenazine
AminopyrineFurosemidePhencyclidine
AmitryptylineGentisic acidPhenelzine
AmobarbitalHemoglobinPhenobarbital
AmoxicillinHydralazineL-Phenylephrine
AmpicillinHydrochlorothiazideβ-Phenylethylamine
D-AmphetamineHydrocodonePhenylpropanolamine
L-AmphetamineHydrocortisonePrednisone
L-Ascorbic Acidα-Hydroxyhippuric acidPrednisolone
Apomorphine3HydroxytyramineProcaine
AspartameIbuprofenD/L-Propanolol
AtropineImipramineD-Propoxyphene
Benzilic acid(+/-)-IsoproterenolD-Pseudoephedrine
Benzoic acidIsoxsuprineQuinidine
BenzoylecgonineKetamineQuinine
BilirubinKetoprofenRanitidine
CaffeineLabetalolSalicylic acid
CannabidiolLoperamideSecobarbital
ChloralhydrateLoxapine succinateSerotonin (5-Hydroxytyramine)
ChloramphenicolMaprotilineSulfamethazine
ChlordiazepoxideMeperidineSulindac
ChlorothiazideMeprobamateTemazepam
ChloroquineMethadoneTetracycline
ChlorpromazineMethoxyphenamineTetrahydrocortisone, 3Acetate
CholesterolMorphine-3-β-D-glucuronideTetrahydrocortisone 3 (β-D-glucuronide)
ClomipramineNalidixic acidTetrahydrozoline
ClonidineNalorphineThebaine
Cocaine hydrochlorideNaloxoneThiamine
CodeineNaltrexoneThioridazine
CortisoneNaproxenTolbutamide
CotinineNiacinamideTriamterene
CreatinineNifedipineTrifluoperazine
DeoxycorticosteroneNorcodeinTrimethobenzamide
DextromethorphanNorethindroneTrimethoprim
DiazepamNoroxymorphoneTrimipramine
DiclofenacD-NorpropoxypheneD/ L-Tryptophan
DiflunisalNoscapineTyramine
DigoxinD/L-OctopamineD/ L-Tyrosine
DiphenhydramineOxalic acidUric acid
DoxylamineOxazepamVerapamil

Methamphetamine:

{12}------------------------------------------------

EcgoninemethylesterOxycodone
Phencyclidine
Acetaminophen(-) Y EphedrineOxycodone
AcetophenetidinErythromycinOxymetazoline
N-Acetylprocainamideẞ-EstradiolPapaverine
Acetylsalicylic acidEstrone-3-sulfatePenicillin-G
AminopyrineEthyl-p-aminobenzoatePentazocine hydrochloride
AmitryptylineFenoprofenPentobarbital
AmobarbitalFurosemidePerphenazine
AmoxicillinGentisic acidPhenelzine
AmpicillinHemoglobinPhenobarbital
Ascorbic acidHydralazinePhentermine
D,L-AmphetamineHydrochlorothiazideL-Phenylephrine
Apomorphine acidHydrocodoneẞ-Phenylethylamine
AspartameHydrocortisonePhenylpropanolamine
AtropineO-HydroxyhippuricPrednisolone
Benzilic acidp-HydroxymethamphetaminePrednisone
Benzoic acid3-HydroxytyramineProcaine
BenzoylecgonineIbuprofenPromazine
BenzphetamineImipraminePromethazine
BilirubinIproniazidD,L-Propanolol
Brompheniramine(±) - IsoproterenolD-Propoxyphene
CaffeineIsoxsuprineD-Pseudoephedrine
CannabidiolKetamineQuinidine
CannabinolKetoprofenQuinine
ChloralhydrateLabetalolRanitidine
ChloramphenicolLoperamideSalicylic acid
ChlordiazepoxideMaprotilineSecobarbital
ChlorothiazideMeperidineSerotonin (5-Hydroxytyramine)
(±) ChlorpheniramineMeprobamateSulfamethazine
ChlorpromazineMethadoneSulindac
ChloroquineMethoxyphenamineTemazepam
Cholesterol(+) 3,4-Methylenedioxy-Tetracycline
Clomipramine(+)3,4-Methylenedioxy-Tetrahydrocortisone3 (β-D glucuronide)
ClonidineMorphine-3-ẞ-D glucuronideTetrahydrozoline
Cocaine hydrochlorideMorphine SulfateThiamine
CodeineNalidixic acidThioridazine
CortisoneNaloxoneD, L-Tyrosine
(-) CotinineNaltrexoneTolbutamide
CreatinineNaproxenTriamterene
DeoxycorticosteroneNiacinamideTrifluoperazine
DextromethorphanNifedipineTrimethoprim
DiazepamNorcodeinTrimipramine
DiclofenacNorethindroneTryptamine
DiflunisalD-NorpropoxypheneD, L-Tryptophan
DigoxinNoscapineTyramine
DiphenhydramineD,L-OctopamineUric acid
DoxylamineOxalic acidVerapamil

Phencyclidine

{13}------------------------------------------------

Ecgonine hydrochlorideOxazepamZomepirac
Ecgonine methylesterOxolinic acid
Marijuana
Acetamidophenolβ-EstradiolPapaverine
AcetophenetidinEstrone-3-sulfatePenicillin G
N-AcetylprocainamideEthyl-p-aminobenzoatePentazocine
Acetylsalicylic acidFenoprofenPentobarbital
AminopyrineFurosemidePerphenazine
AmitryptylineGentisic acidPhencyclidine
AmobarbitalHemoglobinPhenelzine
AmoxicillinHydralazinePhenobarbital
AmpicillinHydrochlorothiazidePhentermine
Ascorbic acidHydrocodoneL-Phenylephrine
D/L-AmphetamineHydrocortisoneβ-Phenylethlamine
L-Amphetamineα-Hydroxyhippuric acidβ-Phenyllethylamine
Apomorphine3-HydroxytyraminePhenylpropanolamine
AspartameIbuprofenPrednisolone
AtropineImipraminePrednisone
Benzilic acidIproniazidProcaine
Benzoic acid(+/-)-IsoproterenolPromazine
BenzoylecgonineIsoxsuprinePromethazine
BenzphetamineKetamineD/L-Propanolol
BilirubinLabetalolD-Propoxyphene
BrompheniramineLevorphanolD-Pseudoephedrine
CaffeineLoperamideQuinidine
ChloralhydrateMaprotilineQuinine
ChloramphenicolMeprobamateRanitidine
ChlordiazepoxideMethadoneSalicylic acid
ChlorothiazideMethoxyphenamineSecobarbital
(+/-)-Chlorpheniramine(+/-)Serotonin (5-Hydroxytyramine)
Chlorpromazine(+/-Sulfamethazine
ChloroquineMethylphenidateSulindac
CholesterolMethyprylonTemazepam
ClomipramineMorphine-3-β-D-glucuronideTetracycline
ClonidineNalorphineTetrahydrocortisone3(β-D-glucuronide)
Cocaine hydrochlorideNaloxoneTetrahydrozoline
CodeineNalidixic acidThebaine
CortisoneNaltrexoneThiamine
(-)-CotinineNaproxenThioridazine
CreatinineNiacinamideD/L-Thyroxine
DeoxycorticosteroneNifedipineTolbutamide
DextromethorphanNorcodeinTriamterene
DiazepamNorethindroneTrifluoperazine
DiclofenacD-NorpropoxypheneTrimethoprim
DiflunisalNoscapineTrimipramine
DigoxinD/L-OctopamineTryptamine
DiphenhydramineOxalic acidD/L-Tryptophan
DoxylamineOxazepamTyramine
Ecgonine hydrochlorideOxolinic acidD/L-Tyrosine

{14}------------------------------------------------

EcgoninemethylesterOxycodoneUric acid
L - Ψ-EphedrineOxymetazolineVerapamil
Erythromycinp-HydroxymethamphetamineZomepirac

f. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below. There were no differences observed for different devices.

MethamphetamineResult% Cross-Reactivity
(Cutoff=1000 ng/mL)Positive at (ng/mL)
D(+)-Methamphetamine1000100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)1000010%
D/L-Methamphetamine1000100%
p-Hydroxymethamphetamine1000010%
D-AmphetamineNegative at ≤100000<1%
L-AmphetamineNegative at ≤100000<1%
Chloroquine500002%
(+/-)-Ephedrine400025%
L-Methamphetamine1000010%
(+/-)3,4-Methylenedioxyamphetamine (MDA)Negative at ≤100000<1%
(+/-)3,4-methylenedioxymethamphetamine(MDMA)500200%
β -Phenylethylamine750013.3%
Trimethobenzamide200005%
Phencyclidine(Cut-off=25 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
Phencyclidine25100%
4-Hydroxyphencyclidine7533.3%
Marijuana(Cut-off=50 ng/mL)ResultPositive at (ng/mL)% Cross-Reactivity
11-Nor-Δ9-Tetrahydrocannabinol-9-COOH50100%
11-Hydroxy-Δ9-Tetrahydrocannabinol50100%
11-Nor-Δ8-Tetrahydrocannabinol-9-COOH50100%
Cannabinol200000.25%
Δ8-Tetrahydrocannabinol150000.33%
Δ9-Tetrahydrocannabinol150000.33%
CannabidiolNegative at ≤100000<0.05%
11-Nor-Δ9-THC-carboxy glucuronide7566.7%
(-)-11-nor-9-carboxy-Δ9-THC50100%

g. Effect of Urine Specific Gravity and pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.009 to 1.030 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. Specific gravity from 1.009 to 1.030 does not affect the test results. Similarly, pH from 4 to 9 does not affect

{15}------------------------------------------------

the test result. There were no differences observed for different format of the devices.

2. Comparison Studies

Method comparison studies for the AssureTech Methampletamine Tests, the AssureTech Phencyclidine Tests and the AssureTech Marijuana Tests were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:

Methamphetamine
StripNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cutoff)Near Cutoff Positive by GC/MS (Between the cutoff and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0011425
Negative1020910
Viewer BPositive0001425
Negative10201010
Viewer CPositive0011425
Negative1020910

Methamphetamine

Discordant Results of Methamphetamine Strip

ViewerSample NumberGC/MS Result(ng/mLof d-Methamphetamine)StripViewer Results
Viewer A50368974Positive
Viewer C50368974Positive
Viewer A963781045Negative
Viewer B796101094Negative
Viewer C963781045Negative
DipCardNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011525
Negative1020900
ViewerBPositive0011425
Negative1020910
ViewerCPositive0001525
Negative10201000

Discordant Results of Methamphetamine Dip Card

{16}------------------------------------------------

ViewerSample NumberGC/MS Result(ng/mLof d-Methamphetamine)Dip CardViewer Results
Viewer A50368974Positive
Viewer B50368974Positive
Viewer B963781045Negative
TurnKeySplitCupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative1020910
ViewerBPositive0001425
Negative10201010
ViewerCPositive0011525
Negative1020900

Discordant Results of Methamphetamine Turn Key Split Cup

ViewerSample NumberGC/MS Result(ng/mLof d-Methamphetamine)Turn Key Split CupViewer Results
Viewer A50368974Positive
Viewer C50368974Positive
Viewer A963781045Negative
Viewer B963781045Negative
QuickCupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
ViewerANegative1020910
ViewerBPositive0001525
ViewerBNegative10201000
ViewerCPositive0001425
ViewerCNegative10201010

Discordant Results of Methamphetamine Quick Cup

ViewerSample NumberGC/MS Result(ng/mLof d-Methamphetamine)Quick CupViewer Results
Viewer A50368974Positive

{17}------------------------------------------------

Viewer A963781045Negative
Viewer C796101094Negative
StripPhencyclidine
NegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerPositive0011425
ANegative1020910
ViewerPositive0011425
BNegative1020910
ViewerPositive0011525
CNegative1020900

Phencyclidine

Discordant Results of Phencyclidine Strip

ViewerSample NumberGC/MS Result(ng/mLof Phencyclidine)StripViewer Results
Viewer A6841422Positive
Viewer B4909024Positive
Viewer C4909024Positive
Viewer A3482527Negative
Viewer B4671727Negative
DipCardNegativeLow Negative byGC/MS(less than -50%)Near Cutoff Negative byGC/MS(Between -50% and cutoff)Near Cutoff Positive byGC/MS(Between the cutoff and +50%)High Positiveby GC/MS(greater than +50%)
ViewerAPositive0021425
Negative1020810
ViewerBPositive0001425
Negative10201010
ViewerCPositive0011425
Negative1020910

Discordant Results of Phencyclidine Dip Card

ViewerSample NumberGC/MS Result(ng/mLof Phencyclidine)Dip CardViewer Results
Viewer A4909024Positive
Viewer A6841422Positive
Viewer C6841422Positive

{18}------------------------------------------------

Viewer A3482527Negative
Viewer B6144327Negative
Viewer C4671727Negative
Turn KeySplit CupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
Viewer APositive0011425
Negative1020910
Viewer BPositive0011425
Negative1020910
Viewer CPositive0011525
Negative1020900

Discordant Results of Phencyclidine Turn Key Split Cup

ViewerSample NumberGC/MS Result(ng/mLof Phencyclidine)Turn Key Split CupViewer Results
Viewer A6841422Positive
Viewer B6841422Positive
Viewer C6841422Positive
Viewer A3482527Negative
Viewer B4671727Negative
QuickCupNegativeLow Negative byGC/MS(less than-50%)Near Cutoff Negative byGC/MS(Between-50% andcutoff)Near Cutoff Positive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011525
Negative1020900
ViewerBPositive0021425
Negative1020810
ViewerCPositive0001425
Negative10201010

Discordant Results of Phencyclidine Quick Cup

ViewerSample NumberGC/MS Result(ng/mLof Phencyclidine)Quick CupViewer Results
Viewer A6841422Positive
Viewer B6841422Positive
Viewer B4909024Positive

{19}------------------------------------------------

Viewer B4671727Negative
Viewer C4671727Negative
Marijuana
StripNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerPositive0011425
ANegative1020910
ViewerPositive0001325
BNegative10201020
ViewerPositive0011425
CNegative1020910

Discordant Results of Marijuana Strip

ViewerSample NumberGC/MS Result(ng/mL of11-Nor-Δ9-Tetrahydrocannabinol-9-COOH)StripViewer Result
Viewer A8718046Positive
Viewer C8718046Positive
Viewer A9789852Negative
Viewer B9789852Negative
Viewer B3299553Negative
Viewer C3299553Negative
DipCardNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative1020910
ViewerBPositive0011525
Negative1020900
ViewerCPositive0001425
Negative10201010

Discordant Results of Marijuana Dip Card

ViewerSample NumberGC/MS Result(ng/mL of11-Nor-△9-Tetrahydrocannabinol-9-COOH)Dip CardViewerResults
Viewer A3878544Positive

{20}------------------------------------------------

Viewer B8718046Positive
Viewer A9789852Negative
Viewer C3299553Negative
TurnKeySplitCupNegativeLowNegative byGC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0011425
Negative1020910
ViewerBPositive0011425
Negative1020910
ViewerCPositive0001425
Negative10201010

Discordant Results of Marijuana Turn Key Split Cup

ViewerSample NumberGC/MS Result(ng/mL of11-Nor-Δ9-Tetrahydrocannabinol-9-COOH)Turn Key SplitCupViewer Results
Viewer A3878544Positive
Viewer B8718046Positive
Viewer A9901554Negative
Viewer B9789852Negative
Viewer C3299553Negative
QuickCupNegativeLow Negative byGC/MS(less than-50%)Near Cutoff Negative byGC/MS(Between-50% andcutoff)Near Cutoff Positive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
ViewerAPositive0001425
Negative10201010
ViewerBPositive0011525
Negative1020900
ViewerCPositive0001425
Negative10201010

Discordant Results of Marijuana Quick Cup

ViewerSample NumberGC/MS Result(ng/mL of11-Nor-△9-Tetrahydrocannabinol-9-COOH)Quick CupViewer Results
Viewer B4378843Positive

{21}------------------------------------------------

Viewer A3299553Negative
Viewer C9456755Negative

Lay-user study

A lay user study was performed at three intended user sites with 1638 lay persons. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested.

Numberofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff210021100
-75% Cutoff21260021100
-50% Cutoff21489021100
-25% Cutoff21769021100
+25% Cutoff211276210100
+50% Cutoff211468210100
+75% Cutoff211795210100

Comparison between GC/MS and Lay Person Results for Methamphetamine Strip

Comparison between GC/MS and Lay Person Results for Methamphetamine Dip Card

% of CutoffNumber ofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff210021100
-75% Cutoff21260021100
-50% Cutoff21489021100
-25% Cutoff21769No. ofPositive1No. ofNegative2095
+25% Cutoff21127619290
+50% Cutoff211468210100
+75% Cutoff211795210100

Comparison between GC/MS and Lay Person Results for Methamphetamine Turn Key Split Cup

% of CutoffNumberofsamplesd-MethamphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff210021100
-75% Cutoff21260021100
-50% Cutoff21489021100

{22}------------------------------------------------

-25% Cutoff2176912095
+25% Cutoff211276210100
+50% Cutoff211468210100
+75% Cutoff211795210100

Comparison between GC/MS and Lay Person Results for Methamphetamine Quick Cup

% of CutoffNumber of samplesd-Methamphetamine Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff210021100
-75% Cutoff21260021100
-50% Cutoff21489021100
-25% Cutoff21769021100
+25% Cutoff21127620195
+50% Cutoff211468210100
+75% Cutoff211795210100

Comparison between GC/MS and Lay Person Results for Phencyclidine Strip

% of CutoffNumber of samplesPhencyclidine Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff210021100
-75% Cutoff218021100
-50% Cutoff2112021100
-25% Cutoff2120021100
+25% Cutoff213020195
+50% Cutoff2139210100
+75% Cutoff2143210100

Comparison between GC/MS and Lay Person Results for Phencyclidine Dip Card

% of CutoffNumber of samplesPhencyclidineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff210No. ofPositiveNo. ofNegative100
-75% Cutoff218021100
-50% Cutoff2112021100
-25% Cutoff212012095
+25% Cutoff213020195
+50% Cutoff2139210100
+75% Cutoff2143210100
Comparison between GC/MS and Lay Person Results for Phencyclidine Turn Key Split Cup

{23}------------------------------------------------

% of CutoffNumberofsamplesPhencyclidineConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff210021100
-75% Cutoff218021100
-50% Cutoff2112021100
-25% Cutoff2120021100
+25% Cutoff213019290
+50% Cutoff2139210100
+75% Cutoff2143210100

Comparison between GC/MS and Lay Person Results for Phencyclidine Quick Cup

% of CutoffNumber of samplesPhencyclidine Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff210021100
-75% Cutoff218021100
-50% Cutoff2112021100
-25% Cutoff2120021100
+25% Cutoff213019290
+50% Cutoff2139210100
+75% Cutoff2143210100

Comparison between GC/MS and Lay Person Results for Marijuana Strip

% of CutoffNumber of samplesMarijuana Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff210021100
-75% Cutoff2113021100
-50% Cutoff2126021100
-25% Cutoff213912095
+25% Cutoff216120195
+50% Cutoff2176210100
+75% Cutoff2186210100

Comparison between GC/MS and Lay Person Results for Marijuana Dip Card

% of CutoffNumberofsamplesMarijuanaConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff210021100
-75% Cutoff2113021100
-50% Cutoff2126021100
-25% Cutoff213921990
+25% Cutoff2161210100

{24}------------------------------------------------

+50% Cutoff2176210100
+75% Cutoff2186210100

Comparison between GC/MS and Lay Person Results for Marijuana Turn Key Split Cup

% of CutoffNumber of samplesMarijuana Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff210021100
-75% Cutoff2113021100
-50% Cutoff2126021100
-25% Cutoff213912095
+25% Cutoff216120195
+50% Cutoff2176210100
+75% Cutoff2186210100

Comparison between GC/MS and Lay Person Results for Marijuana Quick Cup

% of CutoffNumberofsamplesMarijuanaConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff210021100
-75% Cutoff2113021100
-50% Cutoff2126021100
-25% Cutoff213912095
+25% Cutoff2161210100
+50% Cutoff2176210100
+75% Cutoff2186210100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the AssureTech Methamphetamine Tests, AssureTech Phencyclidine Tests and AssureTech Marijuana Tests are substantially equivalent to the predicate.

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).