K Number
K161043
Device Name
Profound System
Date Cleared
2016-09-12

(152 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Profound System is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. Specifically, the 25º Dermal handpiece and cartidge are used for percutaneous treatment of facial wrinkles, and the 75° SubQ handpiece and cartridge are used to improve the appearance of cellulite in patients with Fitzpatrick skin types I-III as supported by long-term clinical data (6 months).

Device Description

The Profound System is comprised of the following components: a re-usable console containing a radiofrequency (RF) generator and graphical user interface; two re-usable treatment applicators; and two disposable, single use, sterile electrode cartridges. Bipolar RF energy is delivered from the RF generator through the electrodes into the dermal layers beneath the surface of the skin. The volume of the treated area is defined by the geometry of the individual micro-electrode needle pairs, which are all electrically isolated from each other and controlled independently by separated RF channels within the console. In addition, temperature sensors provide real-time feedback of tissue temperature. The treatment dose is controlled by the physician.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Primary Endpoint)Reported Device Performance
Primary: 80% of treated thighs show at least 1 point level of improvement in the appearance of cellulite at 3 months post-treatment relative to baseline photos, as assessed by blinded review (considering improvement in dimples and/or undulation irregularities).Met: 94% (79/84) of the treated thighs, based on observed data, demonstrated an improvement in cellulite appearance per blinded review at 3 months. 88% (37/42) of the treated subjects showed improvement in both thighs in dimples and/or undulation at 3 months follow-up. (Specifically: 86% of treated thighs with dimples at baseline showed improvement in dimples; 76% of treated thighs with undulation at baseline showed improvement in undulation irregularities.)
Secondary (Effectiveness): Cellulite appearance improvement by blinded evaluators at 6 months post-treatment visit.Met: Improvement in cellulite appearance was maintained at 6 months follow-up; 93% (78/84) of the treated thighs (per-subject success of 86%) showed improvement in cellulite appearance at 6 months post-treatment compared to baseline based on blinded review.
Secondary (Effectiveness): Cellulite appearance improvement by study investigators and subject improvement/satisfaction assessments.Met: Investigator cellulite reduction assessment results demonstrated progressively improved treatment effects for the large majority of thighs at 3 and 6 months. Cellulite improvement per Nurnberger-Muller scale showed increasing levels of improvement throughout the study. Investigator satisfaction increased gradually and was demonstrated for the majority of subjects at 3 and 6 months. Subject improvement results showed moderate to excellent reduction for most thighs at 3 and 6 months. Subject satisfaction was reported by a majority of subjects at 3 and 6 months.
Safety: No device-related serious adverse events. Anticipated treatment responses resolve without medical intervention.Met: Of 50 subjects, only 3 reported a total of 3 adverse events, none serious, and each resolved during the study. Anticipated treatment-associated responses (e.g., erythema, edema) resolved completely without medical intervention.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): 50 treated subjects (100 thighs).
  • Data Provenance: The study was conducted across 4 U.S. sites. The study was prospective, single-arm, and self-controlled.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states "blinded review" but does not specify the number or qualifications of the experts who performed this blinded review for establishing the ground truth.

4. Adjudication Method for the Test Set

The document states "blinded review" for assessing improvement in cellulite. However, it does not specify a formal adjudication method like 2+1 or 3+1. It implies direct assessment by the blinded reviewers against baseline photos.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study was a single-arm, self-controlled study focused on the Profound System's performance rather than comparing it against human readers or other devices in a multi-reader context. Therefore, there is no effect size reported for human readers improving with AI vs without AI assistance, as AI assistance in image interpretation is not the function of this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device (Profound System) is an electrocoagulation and hemostasis device used for percutaneous treatment of facial wrinkles and cellulite. It is a physical medical device that delivers energy, not an AI algorithm for diagnostic interpretation. Therefore, a standalone algorithm-only performance study is not applicable and was not performed. The performance evaluated is that of the device itself on patients, not an AI interpreting data.

7. The Type of Ground Truth Used

The ground truth for the effectiveness endpoints was established through blinded expert review of clinical photographs at various follow-up intervals compared to baseline photos. Additionally, investigator assessments and patient self-assessments contributed to the understanding of effectiveness.

8. The Sample Size for the Training Set

The document describes a single clinical study of the device for its new indication. It does not refer to a "training set" in the context of an AI model. The 50 subjects (100 thighs) are the test set for evaluating the device's clinical efficacy for its intended use. If this device had AI components, its training set would be mentioned separately. As it is a physical device, this concept does not apply in the same way.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" for an AI model is not applicable in this context. The clinical study described served as the primary evidence to demonstrate the device's performance for regulatory clearance.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three overlapping human profiles facing to the right, with flowing lines beneath them, resembling a bird or a wave.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 12, 2016

Syneron Candela Corporation % Janice Hogan Regulatory Counsel Hogan Lovells Us Llp 1835 Market Street 29th Floor Philadelphia, Pennsylvania 19103

Re: K161043

Trade/Device Name: Profound System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: August 15, 2016 Received: August 15, 2016

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Binita S.Ashar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last paqe

510(k) Number (if known)

K161043

Device Name

Profound System

Indications for Use (Describe)

The Profound System is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. Specifically, the 25º Dermal handpiece and cartidge are used for percutaneous treatment of facial wrinkles, and the 75° SubQ handpiece and cartridge are used to improve the appearance of cellulite in patients with Fitzpatrick skin types I-III as supported by long-term clinical data (6 months).

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Profound System

Submitted by:Syneron Candela Corporation530 Boston Post RoadWayland, MA 01778-1886
Contact Person:Ruthie AmirGlobal Vice President of Clinical, Regulatory, and EducationTel: 508-358-7400 x330Fax: 508-358-5602
Submission Correspondent:Janice HoganHogan Lovells US LLP1835 Market St., 29th FloorPhiladelphia, PA 19103Tel: (267) 675-4600Fax: (267) 675-4601janice.hogan@hoganlovells.com
Date prepared:September 12, 2016
Trade Name:Profound System
Common Name:Electrosurgical unit and accessories
Classification:Class II per 21 C.F.R. § 878.4400 – Electrosurgical cutting andcoagulation device and accessoriesProduct Code PBX, GEI
Predicate Devices:Primary Predicate DevicesSyneron-Candela's Profound System (K082391, K080145)Viora Ltd.'s V10 System (K150035)Reference Predicate DeviceCynosure's Cellulaze Laser (K123407)

Intended Use / Indications for Use:

The Profound System is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. Specifically, the 25° Dermal handpiece and cartridge are used for percutaneous treatment of facial wrinkles, and the 75° SubQ handpiece and cartridge are used to improve the appearance of cellulite in patients with Fitzpatrick skin types I-III as supported by longterm clinical data (6 months).

Device Description:

The Profound System is comprised of the following components: a re-usable console containing a

{4}------------------------------------------------

radiofrequency (RF) generator and graphical user interface; two re-usable treatment applicators; and two disposable, single use, sterile electrode cartridges. Bipolar RF energy is delivered from the RF generator through the electrodes into the dermal layers beneath the surface of the skin. The volume of the treated area is defined by the geometry of the individual micro-electrode needle pairs, which are all electrically isolated from each other and controlled independently by separated RF channels within the console. In addition, temperature sensors provide real-time feedback of tissue temperature. The treatment dose is controlled by the physician.

Technoloqical Characteristics:

The Profound System has similar technological characteristics as its predicates. The primary purpose of this submission is to expand the Profound's indications for use to include long-term improvement in the appearance of cellulite.

The Profound design and components are very similar to those previously cleared. Specifically, the cleared 25° Dermal handpiece and associated electrode cartridge have remained essentially the same, with minor hardware and software changes. The company has also added an option for the physician to use a 75° SubQ handpiece and associated electrode cartridge. The modified device is based on the same principles of operation, i.e., thermal heating. The RF frequency and maximum output voltage per channel are the same as for the cleared Profound System. Same as the previously cleared Profound, the system has an electrically powered console with a graphical user interface through which the physician pre-sets treatment parameters (as appropriate for each patient) and obtains real-time feedback regarding ongoing treatment. Minor software changes have been made and the updated software has been subjected to verification and validation. The technological differences compared to the predicates do not raise any different types of safety or effectiveness questions, because the modified device operates in the same manner and the treatment parameters are the same as or within the range of the predicates.

Summary of Substantial Equivalence:

The subject Profound System has the same intended use and similar indications for use as its predicates. The indications of the subject device are the same as those of the predicate. K082391. except for the proposed additional indication for use: "...improve the appearance of cellulite in patients with Fitzpatrick skin types I-III as supported by long-term clinical data (6 months)." The newly added indications for use, while not the same, is similar to the Viora V10 System predicate which is cleared for "...Temporary reduction in the appearance of cellulite." Moreover, the newly added indications for use is similar to the reference predicate device, Cellulaze (K123407), indications for use, which is as follows: "Cynosure Cellulaze ™ laser is intended for the improvement in the appearance of cellulite as supported by long-term clinical data (at least 6 months with no observed reduction in treatment benefits up to 9 months of observation)."

The subiect device and the primary predicate devices are radio frequency (RF) based electrosurgical devices. The subject device has similar components and the same energy output as the Profound primary predicate, and minor design changes to the previously cleared console and 25° handpiece do not impact device performance. The added 75º cartridge and associated handpiece have similar technological characteristics as the previously cleared cartridge and handpiece.

The similarities and differences in terms of indications for use and technological characteristics between the subject and predicate devices are summarized in the below side comparison table. To address the differences in the indications for use and technological parameters, a clinical study was performed, as described below in the Performance Data section of this 510(k) Summary.

{5}------------------------------------------------

The clinical data support substantial equivalence of the Profound compared to its predicate devices. The technological differences have also been evaluated in bench testing. Minor software changes to accommodate the expanded technological features have been verified and validated.

ProductSyneron-Candela Corp.'sProfound System(K161043)Primaeva Medical, Inc.'sMiratone System(K082391, K080145)Viora Ltd.'s V10 System(K150035)
Intended Use/Indications for UseThe Profound System isindicated for use indermatologic and generalsurgical procedures forelectrocoagulation andhemostasis. Specifically,the 25° Dermal handpieceand cartridge are used forpercutaneous treatment offacial wrinkles, and the 75°SubQ handpiece andcartridge are used toimprove the appearance ofcellulite in patients withFitzpatrick skin types I-III assupported by long-termclinical data (6 months).The Primaeva MedicalMiratone System is indicatedfor use in dermatologic andgeneral surgical proceduresfor electrocoagulation andhemostasis, and thepercutaneous treatment offacial wrinkles.The Viora V10 system isintended for dermatologicalprocedures. The V-FormHandpiece, with Body-Contour (BC) and Facial-Contour (FC) applicators,is indicated for deliveringnon-thermal RF combinedwith massage:• relief of minor muscleaches and pain, relief ofmuscle spasm,temporary improvementof local bloodcirculation and• Temporary reduction inthe appearance ofcellulite.
Product CodePBX, GEIGEIPBX, ISA
Energy TypeBipolar fractionated RFBipolar fractionated RFBipolar RF, Vacuum
RF Frequency460 +/- 5kHz460 +/- 5kHz0.8, 1.7, or 2.45 MHz, orcombination of all 3frequencies
Components• RF generator with userinterface• Reusable electrodeinsertiondevice/applicators (25°with cooling plate and 75°)• Single patient use,disposable, 25° and 75°electrode cartridges• RF generator with userinterface• Cooler controller andhandpiece/applicator• Reusable electrodeinsertion device/applicator• Single patient usedisposable, 25° electrodecartridge.• Console (RF generator,user interface, etc.)• Applicators (with cable,connector to console,and vacuum pump; 2available for celluliteindication)
Maximum OutputVoltage84 VRMS84 VRMSNot specified
Channels25° applicator: 5independent channels(electrode pairs)75° applicator: 7independent channels25° applicator: 5independent channels(electrode pairs)Multi-channel;6 electrodes (V-Formhandpiece)
Target TemperatureRange65-75°C +/- 1°C increments65-75°C +/- 1°C incrementsNot specified
Treatment AreaWidth14 mm for the 25° mode (5pairs)20 mm for the 75° mode (7pairs)14 mm (5 pairs)BC applicator: 97 x 83 mmFC applicator: 0-24 mm
User InterfaceGUI - color LCD displayGUI - color LCD displayDisplay unit
SystemDimensions125 x 46.5 x 44.5 cm /29.5 x 18.5 x 17.5 inches125 x 46.5 x 44.5 cm /29.5 x 18.5 x 17.5 inches45 x 35 x 40 cm
ProductSyneron-Candela Corp.'sProfound System(K161043)Primaeva Medical, Inc.'sMiratone System(K082391, K080145)Viora Ltd.'s V10 System(K150035)
ElectricalRequirements100-240 VAC; 2.5 A; 50-60 Hz; single phase100-240 VAC; 2.5 A; 50-60 Hz; single phase90-264 VAC; 50-60 Hz;single phase
SterilizationCartridges are sterilizedCartridges are sterilizedNot specified

{6}------------------------------------------------

In summary, minor differences between the subject Profound and the predicates do not present different types of safety or effectiveness questions, since the Profound's energy parameters are similar to or within the range of the predicates and the device operates in the same manner and for the same general intended use as the primary predicates. Further, the device's safety and performance have been confirmed by results of both clinical and non-clinical investigations. Therefore, the subject Profound System has the same intended use and similar indications for use. technological characteristics, and principles of operation as the predicate devices.

Performance Data:

Electrical Safety and Electromagnetic Compatibility: Electrical safety and electromagnetic compatibility (EMC) testing for the modified Profound System was conducted by an independent test laboratory in accordance with the applicable standards (IEC 60601-1, Medical electrical equipment, Part 1: General requirements for basic safety and essential performance. EN 60601-1-2, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests, and IEC 60601-2-2. Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories). The Profound System was determined to be in conformance with these standards.

Biocompatibility: The biocompatibility of the Profound device has been established per ISO 10993 guidelines as well as comparison to previously cleared Syneron applicators using the same chrome plating for the same type and duration of patient contact.

Software: Software verification and validation testing was conducted and results were found acceptable for software release.

Bench Testing: Performance testing on the RF generator, 25º Dermal handpiece, and 25º Dermal cartridge was previously conducted and confirmed device performance. Bench testing performed on the new 75º SubQ handpiece and cartridge demonstrated that the modified device performs as intended and device outputs met specifications.

Clinical Data: In support of this 510(k), a single arm, self-controlled, multi-center study was conducted to evaluate the safety and effectiveness of the Profound for improving cellulite appearance. The clinical study evaluated 50 treated subjects (100 thighs), all females, across all skin types. Mean age was 44 years and the majority of the subjects were Caucasian. Data demonstrated that the Profound performs as intended when used for this indication. Treatment with the Profound did not lead to any device-related serious adverse events. The mean pain level reported (3.74/10) reflects patient comfort when the procedure is performed as indicated, i.e., with tumescent anesthesia and cold air. Anticipated treatment responses reported immediately after the procedure (e.g., erythema, edema) all completely resolved without medical intervention. The study met the primary endpoint, where 94% of treated thighs demonstrated improvement in cellulite

{7}------------------------------------------------

appearance as assessed by blinded review at 3 months following treatment when considering improvement in dimples and/or undulation irregularities. When considering improvement in each cellulite feature separately, 86% of the treated thighs with dimples at baseline) showed improvement in dimples, and 76% of the thighs had undulation at baseline) showed improvement in undulation irregularities. Per subject results were 88% (37/42) of the treated subjects showed improvement in both thighs in dimples and/or undulation at 3 months follow up. When considering each cellulite feature separately, 71% of the treated subjects showed improvement in dimples in both thighs, and 64% showed improvement in undulation irreqularities in both thighs.

Further, additional assessments of treatment effectiveness were also favorable. These included blinded review at 6 months post-treatment, where improvement in cellulite appearance was maintained at 6 months follow up; 93% (78/84) of the treated thighs (per-subject success (both thighs) of 86%) showed improvement in cellulite appearance at 6 months post-treatment compared to baseline based on blinded review. In addition, assessments also included, openlabel assessments by investigators and patients of the severity of cellulitic features, overall improvement in cellulite appearance, and satisfaction with treatment at 1, 3, and 6 months following treatment. Therefore, the study results for the subject device did not present any different types of safety or effectiveness questions as compared to the predicates. The study design and results are summarized in the table below.

StudyDesignProspective, single-arm, self-controlled, multicenter clinical study
Sample size50 patients (100 thighs) were treated across 4 U.S. sites.
PrincipalEligibilityCriteriaSeeking treatment of cellulite in the upper thighs areas Presenting with cellulite stage II or III as graded using the Nurnberger-Muller scale classification Healthy females ages 25 to 60 years of age Willing to receive the proposed Profound treatment and comply with thefollow-up protocol
Follow upintervals4 follow-up visits: 1 week, 1 month, 3 months and 6 months following treatment.
EndpointsPrimary: The study primary endpoint was achieved when 80% of thighs showedat least 1 point level of improvement in the appearance of cellulite at 3 monthfollow-up visit relative to baseline photos based on blinded review.Secondary: Cellulite appearance improvement by blinded evaluators at 6 monthspost treatment visit. Cellulite appearance improvement by study investigators. Investigator improvement and satisfaction assessments. Subject improvement and satisfaction assessments.

{8}------------------------------------------------

EffectivenessResultsPrimary: Met endpoint; 94% (79/84) of the treated thighs based on observeddata had an improvement in cellulite appearance per blinded review. 88%(37/42) of the treated subjects showed improvement in both thighs in dimplesand/or undulation at 3 months follow up.
Secondary:Improvement in cellulite appearance was maintained at 6 months followup; 93% (78/84) of the treated thighs (per-subject success (both thighs)of 86%) showed improvement in cellulite appearance at 6 months post-treatment compared to baseline based on blinded review.Investigator cellulite reduction assessment results demonstratedprogressively improved treatment effects for the large majority of thighsat 3 and 6 months following treatment.Cellulite improvement per the Nurnberger-Muller scale alsodemonstrated increasing levels of improvement throughout the study(majority of the thighs showed improvement at 3 and 6 months followingtreatment).Nearly all of the thighs showed improvement at 1 month followingtreatment and all thighs showed improvement at both 3 and 6 monthsfollowing treatment.Investigator satisfaction increased gradually along the study course andwas demonstrated for the majority of the subjects at 3 and 6 monthfollow up.Subject improvement results showed moderate to excellent reduction formost of the thighs at 3 and 6 month follow up.Subject satisfaction was reported by a majority of subjects at 3 and 6month follow-up visits.
SafetyResultsOf the 50 total subjects, only 3 subjects reported a total of 3 adverse events.None were serious and each event resolved during the study. Anticipatedtreatment-associated responses experienced during the first week followingtreatment were also reported. None of the immediate responses were observedat the next visit (i.e., 1 month follow-up visit).

The distribution of numbers of insertions in the study suggested that most treatments were covered by a range of up to 300 insertions per thigh, or up to a maximum treatment density of 1.333. No clear dose-response relationship was observed, and treatment beyond these levels has not been shown to improve outcomes and is not recommended.

Conclusions:

Testing of the Profound System demonstrated that the device performs as intended with a favorable safety profile. The non-clinical data further support device safety. Verification and validation testing of the modified software demonstrates that the device performs as intended. Thus, the Profound System is substantially equivalent to the predicate devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.