(97 days)
Not Found
No
The device description and performance studies focus on the material properties and biocompatibility of the porous polyethylene implants, with no mention of AI or ML.
No
The device, "Biopor Porous Polyethylene Implants," is intended for augmentation or reconstruction of the maxillofacial skeleton, which places it in the category of a surgical implant rather than a therapeutic device that treats or cures a disease.
No
Explanation: The device description clearly states that Biopor Porous Polyethylene Implants are intended for "augmentation or reconstruction of the maxillofacial skeleton." This indicates a therapeutic or reconstructive purpose rather than a diagnostic one. The performance studies also focus on biocompatibility and implantation, not on identifying diseases or conditions.
No
The device description clearly states it is a physical implant made of porous high-density polyethylene, with options for coatings and embedded titanium mesh. This is a hardware device, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is for "augmentation or reconstruction of the maxillofacial skeleton." This describes a surgical implant used directly on the patient's body.
- Device Description: The description details a physical implant made of porous polyethylene, designed for surgical implantation and tissue ingrowth.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue samples) to provide information for diagnosis, monitoring, or screening.
IVD devices are used outside the body to analyze samples and provide diagnostic information. This device is a surgical implant used inside the body for structural support and reconstruction.
N/A
Intended Use / Indications for Use
Biopor Porous Polyethylene Implants in block, sheet, and anatomical shapes are intended for the augmentation or reconstruction of the maxillofacial skeleton.
Product codes
KKY
Device Description
Biopor Porous Polyethylene Implants are manufactured of porous high-density polyethylene, a biomaterial that is easily contoured or carved to suit the anatomical and functional requirements of the patient. The interconnecting pores of the Porous HDPE material permits fibrovascular ingrowth into the implant. The implants are manufactured with the option of a coating with a water-soluble alkylene oxide copolymer blend and/or the option of embedded titanium mesh. AOC™ Porous Polyethylene Surgical Implants are provided STERILE and should not be resterilized.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
maxillofacial skeleton
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Performance testing of the Biopor® Porous Polyethylene Implants was completed. Biocompatibility and Implantation studies were also conducted in accordance with ISO 10993. The test data reports are listed below:
Cytotoxicity Study Using the End-Point Titration Murine Local Lymph Node Assay (LLNA) ISO Modified Intracutaneous Study Solution USP and ISO Modified Systemic Toxicity Study Solution ISO Muscle Implantation Study Genotoxicity Mouse Bone Marrow Micronucleus Study Genotoxicity: Bacterial Reverse Mutation Study Genotoxicity: In Vitro Chromosomal Aberration Study Bone Implantation Study in the Femur of the Rabbit AOC Polymer Hemolysis Porous Polyethylene Post-irradiation Cytotoxicity ISO Intramuscular Implantation AOC Coated & Uncoated PPE Pyrogen Test in NZW Rabbits Suture Pullout from PPE
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 878.3500 Polytetrafluoroethylene with carbon fibers composite implant material.
(a)
Identification. A polytetrafluoroethylene with carbon fibers composite implant material is a porous device material intended to be implanted during surgery of the chin, jaw, nose, or bones or tissue near the eye or ear. The device material serves as a space-occupying substance and is shaped and formed by the surgeon to conform to the patient's need.(b)
Classification. Class II.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and support. The profiles are connected to three flowing ribbons. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2016
Ceremed Incorporated Ms. Chelsea Mitchell Vice President of Regulatory Affairs 3643 Lenawee Avenue Los Angeles. California 90016
Re: K160988
Trade/Device Name: Biopor®, AOC™ Porous Polyethylene, Cerepor™ Regulation Number: 21 CFR 878.3500 Regulation Name: Polytetrafluoroethylene with carbon fibers composite implant material Regulatory Class: Class II Product Code: KKY Dated: June 8, 2016 Received: June 23, 2016
Dear Ms. Mitchell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K160988
Device Name Biopor Porous Polyethylene Implants
Indications for Use (Describe)
Biopor Porous Polyethylene Implants in block, sheet, and anatomical shapes are intended for the augmentation or reconstruction of the maxillofacial skeleton.
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
Exemption Under 21 CFR 58.1 Subpart D | For EPA-Specific Use (21 CFR 58.1(a)(2)) |
---|---|
------------------------------------------------------------------------------------------ | --------------------------------------------------------------------------------------------- |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary of Safety and Effectiveness
Submitted by:
Chelsea Mitchell Ceremed, Inc. 3643 Lenawee Ave. Los Angeles, California 90016 Tel: (424) 258-1888 Fax: (310) 815-2130
Contact Person: | Chelsea Mitchell |
---|---|
Date Prepared: | March 31, 2016 |
Common/Usual Name: | Porous High Density Polyethylene |
(HDPE) Surgical Implant | |
Proprietary Name: | Biopor®, AOC™ Porous |
Polyethylene, Cereport™ | |
Classification Name: | Polytetrafluoroethylene with carbon |
fibers composite implant material | |
Product Code: | KKY |
Predicate Devices:
-
- Ceremed. Inc. Biopor Porous Polyethylene Surgical Implants (K141880)
-
- Poriferous, LLC Su-Por® Surgical Implants (K140437)
-
- Stryker® (Formerly Porex) Medpor® Surgical Implant Material (K922489)
Description of the device:
Biopor Porous Polyethylene Implants are manufactured of porous high-density polyethylene, a biomaterial that is easily contoured or carved to suit the anatomical and functional requirements of the patient. The interconnecting pores of the Porous HDPE material permits fibrovascular ingrowth into the implant. The implants are manufactured with the option of a coating with a water-soluble alkylene oxide copolymer blend and/or the option of embedded titanium mesh. AOC™ Porous Polyethylene Surgical Implants are provided STERILE and should not be resterilized.
Intended use (Indications):
Biopor Porous Polyethylene Implants in block, sheet, and anatomical shapes are intended for the augmentation or reconstruction of the maxillofacial skeleton.
4
Substantial equivalence:
Biopor Porous Polyethylene Implants in this application represent a line extension of additional shapes and sizes, and have the same intended use and indications for use as the predicate Biopor® Porous Polyethylene Implants (K141880). The additional shapes and sizes are equivalent to those of the predicates Medpor® Surgical Implant Material (K922489) and Su-Por® Surgical Implants (K140437).
The biocompatibility of the alkylene oxide copolymer blend is in accordance with the standards set forth in ISO 10993 Biological Testing of Medical and Dental Materials and Devices.
The mechanical properties of AOC™ Porous Polyethylene Surgical Implants are substantially equivalent to the corresponding properties of the predicate devices made of porous polyethylene, and any minor differences raise no new issues of safety and efficacy.
Biocompatibility and Performance Testing:
Performance testing of the Biopor® Porous Polyethylene Implants was completed. Biocompatibility and Implantation studies were also conducted in accordance with ISO 10993. The test data reports are listed below:
Cytotoxicity Study Using the End-Point Titration Murine Local Lymph Node Assay (LLNA) ISO Modified Intracutaneous Study Solution USP and ISO Modified Systemic Toxicity Study Solution ISO Muscle Implantation Study Genotoxicity Mouse Bone Marrow Micronucleus Study Genotoxicity: Bacterial Reverse Mutation Study Genotoxicity: In Vitro Chromosomal Aberration Study Bone Implantation Study in the Femur of the Rabbit AOC Polymer Hemolysis Porous Polyethylene Post-irradiation Cytotoxicity ISO Intramuscular Implantation AOC Coated & Uncoated PPE Pyrogen Test in NZW Rabbits Suture Pullout from PPE