K Number
K160809
Device Name
Reveal LINQ
Manufacturer
Date Cleared
2016-05-26

(63 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reveal LINQ ICM is an insertable automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:

  • patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
  • patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia
    The device has not been tested specifically for pediatric use.
Device Description

The Reveal® LINQ™ Insertable Cardiac Monitor (ICM) Model LNQ11 is designed to record the occurrence of arrhythmias in a patient. Arrhythmia may be classified as atrial tachyarrhythmia/atrial fibrillation (AT/AF), bradyarrhythmia, pause, or (fast) ventricular tachyarrhythmia. The Reveal LINQ ICM provides storage of ECG and Marker Channel during patient-activated and automatically-detected (auto-activated) events. Auto-activation may help to detect abnormal heart rhythms in patients who may not activate/trigger the ICM. The Reveal LINQ Model LNQ11 is a small, leadless implantable device that is typically implanted under the skin, in the chest. Two electrodes on the body of the implantable device continuously monitor the patient's subcutaneous ECG.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Medtronic Reveal LINQ Insertable Cardiac Monitor (ICM), Model LNQ11. However, the submission does not contain specific acceptance criteria or an explicit study describing device performance against those criteria.

The focus of this 510(k) is to demonstrate substantial equivalence of the Reveal LINQ ICM (Model LNQ11) when used with a new mobile application (Reveal LINQ Mobile Manager App Model MSW002 for iOS) compared to its use with an existing mobile application (Reveal LINQ Mobile Manager App Model MSW001 for Android).

The document explicitly states:
"The Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 have no impact on the physical characteristics, materials, mechanical, electrical, hardware, components, firmware, operation, performance, and telemetry protocol of the Reveal LINQ ICM."
And, "There are no differences in the performance, features, materials, hardware, intended use or programmer-to-implantable device interface between the Reveal LINQ ICM used with the Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 versus the existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965."

Therefore, the provided document does not include detailed acceptance criteria or a study proving the device (the ICM itself) meets specific performance metrics in the way a de novo or more comprehensive submission might. It relies on the assertion that the core device (ICM Model LNQ11) and its performance remain unchanged from its previous clearance (K150614) and that the new mobile app does not alter this performance.

Given this, I cannot directly extract the specific information requested in the prompt based on the provided text for acceptance criteria and a study demonstrating performance. The document focuses on demonstrating that the addition of a new mobile app platform does not negatively impact the already-cleared device.

However, I can provide a summary of what the document does include regarding testing and equivalence:

1. A table of acceptance criteria and the reported device performance:

  • Acceptance Criteria: Not explicitly stated as pass/fail thresholds for specific performance metrics of the ICM. The acceptance hinges on demonstrating that the new mobile app (MSW002) does not alter the performance or intended use of the already-cleared ICM (LNQ11).
  • Reported Device Performance: The document asserts that the performance, features, materials, hardware, intended use, and programmer-to-implantable device interface are the same when using the new iOS app (MSW002) as with the existing Android app (MSW001). The inherent performance of the LNQ11 ICM for "recording the occurrence of arrhythmias" is considered equivalent to its predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Sample Size: Not specified for any particular test set. The document refers to "system verification and system validation" for the ICM model LNQ11 itself, and for the new mobile app and patient connector. These are typically internal engineering and software tests rather than clinical studies with patient data.
  • Data Provenance: Not applicable in the context of this submission, as it focuses on software and system equivalence rather than clinical data from patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not applicable. Ground truth as typically defined in clinical performance studies (e.g., expert consensus on diagnostic accuracy) is not a component of this submission, which focuses on device and software functional equivalence.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods are relevant for clinical studies involving human interpretation or challenging diagnoses, which is not the subject of this 510(k).

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is an Insertable Cardiac Monitor, which automatically detects arrhythmias. It is not an AI-assisted diagnostic imaging device that involves human readers interpreting output.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The Reveal LINQ ICM is described as an "automatically-activated monitoring system that records subcutaneous ECG." This indicates its primary function is standalone detection and recording of arrhythmias. The 510(k) implies that the core detection algorithms of the ICM (e.g., "Full View + P-wave presence filter") remain unchanged from the predicate device. However, the document does not detail a standalone performance study in this specific submission. It relies on the fact that the ICM Model LNQ11 already had standalone performance established in its prior clearance (K150614).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not applicable to this specific 510(k) submission for mobile app equivalence. For the original ICM clearance (K150614), the ground truth for arrhythmia detection would typically involve a comparison of the device's automatic detections against expert-annotated ECGs or other clinical evidence. This detail, however, is not in the provided text.

8. The sample size for the training set:

  • Not applicable. This submission is for a device and an associated app, not a new algorithm developed via machine learning requiring an explicit training set described within this document. The detection algorithms are stated to be "Same" as the predicate device.

9. How the ground truth for the training set was established:

  • Not applicable for the same reasons as #8.

In summary: The provided document is a 510(k) submission focused on demonstrating substantial equivalence for a software update (new mobile app) for an already cleared device. It asserts that the performance of the core device remains unchanged, rather than providing new performance criteria or studies for the device itself. Therefore, the specific details regarding acceptance criteria and performance studies for the ICM are not contained within this particular document. They would have been part of the original K150614 submission for the Reveal LINQ ICM (Model LNQ11).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 26, 2016

Medtronic, Inc. Dianna Johannson Sr. Principal Regulatory Affairs Specialist 8200 Coral Sea St Nw Mounds View, Minnesota 55112

Re: K160809

Trade/Device Name: Reveal LINO Insertable Cardiac Monitor, Model LNO11 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: DSI Dated: April 26, 2016 Received: April 27, 2016

Dear Dianna Johannson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160809

Device Name

Reveal LINQ Insertable Cardiac Monitor (Model LNQ11)

Indications for Use (Describe)

The Reveal LINQ ICM is an insertable automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:

· patients with clinical syndromes or situations at increased risk of cardiac arrhythmias

· patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia

The device has not been tested specifically for pediatric use.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo consists of a blue circular icon on the left and the word "Medtronic" in blue bold font on the right. The icon appears to be a stylized representation of a person in motion.

Date Prepared:18 March 2016
510(k) Owner / Address:Medtronic, Inc.Cardiac Rhythm and Heart Failure8200 Coral Sea StreetMounds View, MN 55112
Contact:Dianna L. JohannsonSr. Pr. Regulatory Affairs Specialist
Telephone:(763) 526-2376
Fax:(651) 367-0603
E-mail:dianna.johannson@medtronic.com
Trade / Proprietary Name:Reveal® LINQTM Insertable Cardiac Monitor,Model LNQ11
Common Name:Insertable Cardiac Monitor
Classification /Classification Name:Class II (special controls)Arrhythmia detector and alarm(21 CFR 870.1025)
Product Code:DSI

510(k) Summary

Summary of Substantial Equivalence

The intended use, design, materials and performance of the Reveal LINQ ICM (Model LNQ11) in association with the Reveal LINQ Mobile Manager App (Model MSW002) and patient connector (Model 24965) (referred to as the Reveal LINQ Mobile Manager) are substantially equivalent to the following predicate device:

  • Reveal LINO ICM (Model LNO11) cleared via K150614 on 06AUG2015. ●

Device Description

The Reveal® LINQ™ Insertable Cardiac Monitor (ICM) Model LNQ11 is designed to record the occurrence of arrhythmias in a patient. Arrhythmia may be classified as atrial tachyarrhythmia/atrial fibrillation (AT/AF), bradyarrhythmia, pause, or (fast) ventricular tachyarrhythmia. The Reveal LINQ ICM provides storage of ECG and Marker Channel during patient-activated and automatically-detected (auto-activated) events. Auto-activation may help to detect abnormal heart rhythms in patients who may not activate/trigger the ICM. The Reveal LINQ Model LNQ11 is a small, leadless implantable device that is typically implanted under the skin, in the chest. Two electrodes on the body of the implantable device continuously monitor the patient's subcutaneous ECG.

{4}------------------------------------------------

Medtronic Reveal LINQ (Model LNQ11) ICM

The Reveal LINQ ICM is a small, leadless implantable device that is implanted under the skin, in the chest. The implantable device uses two electrodes on its surface to monitor the patient's subcutaneous ECG continuously. The implantable device memory can store up to 27 min of ECG recordings from automatically detected arrhythmias and up to 30 min of ECG recordings from patient-activated episodes. The system provides 3 options for segmenting the patientactivated episode storage: up to four 7.5 min recordings, up to three 10 min recordings, or up to two 15 min recordings. Arrhythmia detection parameters are set to pending automatically, based on patient information entered on the programmer during pre-insertion implantable device setup: the patient's Date of Birth and the clinician's Reason for Monitoring the patient. Arrhythmia detection parameters can also be programmed manually by the clinician. The following table describes the Reveal LINQ ICM Model LNQ11 when used with the new Reveal LINQ Mobile Manager App Model MSW002 for iOS and patient connector as compared to the Reveal LINQ ICM Model LNQ11 when used with the existing Reveal LINQ Mobile Manager App Model MSW001 for Android and patient connector.

ParameterReveal LINQ ICM ModelLNQ11 (with existing RevealLINQ Mobile Manager AppModel MSW001 for Androidand patient connector Model24965)Reveal LINQ ICM ModelLNQ11 (with new RevealLINQ Mobile Manager AppModel MSW002 for iOS andpatient connector Model24965)
Longevity3 yearsSame
Electrode Spacing (inside-to-inside)37.7 mmSame
Volume1.2 ccSame
Mass2.4 gSame
Episode Storage57 minSame
Patient Symptom MarkPatient AssistantSame
Cardiac CompassYesSame
MRI CompatibilityMR ConditionalSame
Clinician NotificationNightly Transmission /NotificationsSame
Bi-Directional TelemetryBSame
Detection AlgorithmsFull View + P-wave presencefilterSame
CareLinkYesSame
Wireless Telemetry1-Way, Transmit OnlySame
Patient's CareLink ClinicName and IDYesSame

{5}------------------------------------------------

The Reveal LINQ ICM Model LNQ11 with existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965 was initially cleared via a 510(k) application K150614 on 06AUG2015. This submission is to add an app (Model MSW002) to the Reveal LINQ Mobile Manager for use on iOS mobile devices. The Reveal LINQ Mobile Manager for iOS devices is comprised of the Reveal LINO Mobile Manager App Model MSW002 installed on a clinician's off-the-shelf non-medical iOS mobile device and patient connector Model 24965 (also referred to as a telemetry head) used in conjunction with the CareLink Network. The Reveal LINO ICM continues to have the same intended use when used with the Reveal LINO Mobile Manager App Model MSW002 and patient connector Model 24965. The Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 provide the essential capabilities for programming, interrogating and managing the Reveal LINQ ICM in a clinical or hospital environment.

The Reveal LINO Mobile Manager App Model MSW002 and patient connector Model 24965 have no impact on the physical characteristics, materials, mechanical, electrical, hardware, components, firmware, operation, performance, and telemetry protocol of the Reveal LINO ICM. There are no differences in the performance, features, materials, hardware, intended use or programmer-to- implantable device interface between the Reveal LINQ ICM used with the Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 versus the existing Reveal LINO Mobile Manager App Model MSW001 and patient connector Model 24965. The Reveal LINO Mobile Manager App Model MSW002 and patient connector Model 24965 are capable of interrogating and programming the Reveal LINQ ICM like the existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965.

Indications for Use

There are no changes to the cleared indications for use for the Reveal LINO ICM Model LNO 1 when used with the Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 as compared to the Reveal LINQ ICM Model LNQ11 indications when used with the existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965.

The indication statement is as follows:

The Reveal LINO ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases:

  • patients with clinical syndromes or situations at increased risk of cardiac arrhythmias ●
  • . patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia

The device has not been tested specifically for pediatric use.

{6}------------------------------------------------

Technological Characteristics

The Reveal LINQ ICM consists of three major subassemblies which include the hermetically enclosed battery, hermetically enclosed electronics module, and molded header assembly. The battery is a custom D-shaped cell based on LiCFx chemistry and supports a 12-month shelf life and 3-year usable capacity. The Electronics Module contains the hybrid comprised of a printed circuit board with surface mount components on one side and an over molded stack on the other side which contains analog, digital and memory IC's along with capacitor arrays. The molded Header Assembly contains a Titanium Nitride coated sensing electrode, an embedded miniature RF antenna, suture hole, and a mounting bracket.

The Reveal LINO ICM will continue to use the same technology. It is designed to automatically record the occurrence of an arrhythmia in a patient, continuously senses the patient's subcutaneous ECG, and analyzes the timing of ventricular events to detect possible episodes of arrhythmia. The Reveal LINQ ICM has a small form factor, and uses Titanium, Parylene, Urethane, and Titanium Nitride coating on the sensing electrodes as body contacting materials.

The use of the Reveal LINO Mobile Manager App Model MSW002 and patient connector Model 24965 to support the Reveal LINQ ICM does not alter the existing technology of the Reveal LINQ ICM. Additionally, from the system perspective, the Reveal LINQ Mobile Manager App still communicates via an RF head (the patient connector) which in turn communicates with the Reveal LINQ ICM to program diagnostic settings and to extract data from the implantable device for viewing on either the Reveal LINQ Mobile Manager App or the Medtronic CareLink Network. The patient connector utilizes the same telemetry protocol (Telemetry B) to communicate with the Reveal LINQ ICM; however, it utilizes Bluetooth Low Energy technology to communicate with the mobile device, and has rechargeable battery-powered operation. The Reveal LINQ Mobile Manager App utilizes existing infrastructure of the clinician's mobile device to run the programming application, facilitating portability and mobility, direct connectivity with CareLink via cellular and Wi-Fi, Bluetooth Low Energy connectivity to communicate with the patient connector, software updates via publicly accessible channels (i.e. Apple App Store), and a user interface with workflows, streaming ECG waveform and marker display.

Summary of Testing

Testing was performed to demonstrate equivalency of the Reveal LINQ Model LNQ11 Insertable Cardiac Monitor with Reveal LINO Mobile Manager App Model MSW002 and patient connector Model 24965 to the Reveal LINQ ICM with the existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965. The Reveal LINQ ICM with Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 were subjected to the same use scenarios during testing as the Reveal LINO ICM with the existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965, thus supporting substantial equivalency. Testing included:

Reveal LINQ ICM Model LNQ11 Testing

System verification and system validation were completed with the Reveal LINO Model LNQ11 Insertable Cardiac Monitor.

{7}------------------------------------------------

Reveal LINO Mobile Manager App Model MSW002 Testing

Performance verification as well as software verification, system verification and system validation were completed for the Reveal LINQ Mobile Manager App Model MSW002.

Patient Connector Model 24965 Testing

Electromagnetic compatibility (EMC), RF compliance, and wireless coexistence testing were completed on and with the patient connector. Additionally, firmware verification, system verification and system validation were completed with the patient connector. Electrical safety, mechanical, packaging, biocompatibility and performance verification testing were performed on the patient connector as part of K150614 but did not warrant repeating given the scope of this submission, that the Model 24965 electrical and mechanical design is unchanged, and that the Model 24965 hardware / materials and packaging are unchanged. The patient connector is provided non-sterile.

The results of the testing indicate that the Reveal LINQ Insertable Cardiac Monitor Model LNQ11 with the Reveal LINO Mobile Manager App Model MSW002 and patient connector Model 24965 perform as intended, and are safe for their intended use.

The FDA recognized standards that the Reyeal LINO Insertable Cardiac Monitor Model LNQ11, Reveal LINQ Mobile Manager App (Model MSW002), and patient connector (Model 24965) complies are identified in the following tables.

Standard
StandardsOrganization /NumberStandards TitleDate /Version
AAMI / ANSI /IEC 62366Medical devices - Application of usability engineering to medicaldevices2007
ISO 10993-1Biological evaluation of medical devices - Part 1: Evaluation andtesting within a risk management process2010 (ForthEdition)
ISO 10993-7Biological evaluation of medical devices - Part 7: Ethylene oxidesterilization residuals2008(SecondEdition)
ISO 11607-1Packaging for terminally sterilized medical devices - Part 1:Requirements for materials, sterile barrier systems and packagingsystems2009
ISO 14971Medical devices - Applications of risk management to medicaldevices2007(SecondEdition)
IEC 62304Medical device software - Software life-cycle processes2006

Standards Referenced for the Reveal LINQ ICM

{8}------------------------------------------------

Standard
StandardsOrganization /NumberStandards TitleDate /Version
IEC 62304Medical device software – Software life cycle processes2006
AAMI/ANSI/IEC 62366Medical devices – Application of usability engineering to medicaldevices2007/(R)2013
ISO 14971Medical devices – Application of risk management to medicaldevices2007
ISO 15223-1Medical devices – Symbols to be used with medical device labels,labeling, and information to be supplied — Part 1: Generalrequirements2012

Standards Referenced for the Reveal LINQ Mobile Manager App

Standards Referenced for the Patient Connector

Standard
StandardsOrganization /NumberStandards TitleDate / Version
AAMI/ANSI/ISO 10993-1Biological evaluation of medical devices. Evaluation and testing2009/(R)2013
IEC 60601-1-2Medical electrical equipment - Part 1: General requirements forbasic safety and essential Performance2007
IEC 60601-1-6Medical electrical equipment - Part 1-6: General requirements forbasic safety and essential Performance2010
IEC 62133Secondary cells and batteries containing alkaline or other non-acidelectrolytes - Safety requirements for portable sealed secondarycells, and for batteries made from them, for use in portableapplications.2012
IEC 62304Medical device software – Software life cycle processes2006
AAMI/ANSI/IEC 62366Medical devices – Application of usability engineering to medicaldevices2007/(R)2013
ISO 14971Medical devices – Application of risk management to medicaldevices2007 / 2012
ISO 15223-1Medical devices - Symbols to be used with medical device labels,labeling, and information to be supplied – Part 1: Generalrequirements2012

{9}------------------------------------------------

Conclusion

Medtronic has demonstrated that the Reveal LINQ ICM (Model LNQ11) with the Reveal LINQ Mobile Manager App Model MSW002 and patient connector Model 24965 described in this submission result in a substantially equivalent device because the fundamental scientific principle, operating principle, design features and intended use are unchanged from the predicate Reveal LINQ ICM with the existing Reveal LINQ Mobile Manager App Model MSW001 and patient connector Model 24965, and it continues to be safe, effective, and performs as intended.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.