(13 days)
The Acclarix LX8 Diagnostic Ultrasound System is intended for use by a qualified physician or sonographer for ultrasound evaluation. Clinical applications include: Abdominal, Gynecology (including endovaginal), Obstetic, Cardiac, Small parts (Breast, Testes, Thyroid, etc.), Urology, Musculoskeletal, and Peripheral vascular, and Intra-operative.
The Edan Acclarix LX8 Ultrasound system consists of a main system along with associated transducers. The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The waves are then reflected within the body and detected by the transducer, which then converts back to an electrical signal. The Acclarix LX8 system then analyzes the returned signal to generate an image or conduct Doppler processing. The Acclarix LX8 system gives the operator the ability to measure anatomical structures, and offers analysis packages that provide information used by competent health care professionals to make a diagnosis. The system provides both touch screen and hard buttons for the User Interface.
The provided document is a 510(k) summary for the EDAN INSTRUMENTS, INC. Acclarix LX8 Diagnostic Ultrasound System. It outlines the device's indications for use and compares it to predicate devices to establish substantial equivalence. However, the document explicitly states that clinical testing was not required for this submission (Page 15, "Clinical test: Clinical testing is not required."). Therefore, the detailed information requested regarding acceptance criteria and studies proving the device meets those criteria for clinical performance is not available within this document.
The document focuses on non-clinical testing and benchmarking against predicate devices to demonstrate safety and effectiveness.
Here's a breakdown of the available information based on your request:
1. A table of acceptance criteria and the reported device performance:
Since no clinical studies were performed, there are no specific clinical performance acceptance criteria or reported clinical performance metrics in this document. The "acceptance criteria" appear to be compliance with relevant safety and performance standards for ultrasound devices, which are met through non-clinical testing.
| Acceptance Criteria (Non-clinical) | Reported Device Performance |
|---|---|
| IEC 60601-1 Electrical Safety | Complies |
| IEC 60601-1-2 Electromagnetic Compatibility | Complies |
| IEC 60601-2-37 Basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment | Complies |
| NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment | Complies |
| NEMA UD 3 Standard for real-time display of thermal and mechanical acoustic output indices | Complies |
| Acoustic output testing (per FDA guideline) | Complies |
| ISO 10993-1, -5, -10, -12 Biocompatibility | Complies |
2. Sample size used for the test set and the data provenance:
- Test set sample size: Not applicable, as no clinical test set was used.
- Data provenance: Not applicable, as no clinical data was generated for the purpose of this submission. The testing was non-clinical (electrical safety, electromagnetic compatibility, acoustic output, biocompatibility) and based on engineering validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable, as no clinical test set with human-established ground truth was used.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable, as no clinical test set requiring adjudication was used.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. The device is a diagnostic ultrasound system, not an AI-powered image interpretation device, and no MRMC study was performed or required for this 510(k) submission.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is a diagnostic imaging system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable. For the non-clinical tests conducted, the "ground truth" would be established by reference standards or engineering specifications.
8. The sample size for the training set:
Not applicable, as this is not an AI/Machine Learning device requiring a training set.
9. How the ground truth for the training set was established:
Not applicable, as this is not an AI/Machine Learning device requiring a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, with flowing lines representing movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2016
EDAN INSTRUMENTS, INC. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K160790
Trade/Device Name: Acclarix LX8 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: March 21, 2016 Received: March 22, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K160790
Device Name Acclarix LX8 Diagnostic Ultrasound System
Indications for Use (Describe)
The Acclarix LX8 Diagnostic Ultrasound System is intended for use by a qualified physician or sonographer for ultrasound evaluation. Clinical applications include: Abdominal, Gynecology (including endovaginal), Obstetic, Cardiac, Small parts (Breast, Testes, Thyroid, etc.), Urology, Musculoskeletal, and Peripheral vascular, and Intra-operative.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Acclarix LX8 Diagnostic Ultrasound System
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify)[2]3 | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal / Obstetrics | N | N | N | N | N | N | |||
| Abdominal | N | N | N | N | N | N | N | ||
| Intra-operative (Specify) | N | N | N | N | N | N | |||
| Intra-operative (Neuro logical) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify) * | N | N | N | N | N | N | |||
| FetalImaging& Other | Neonatal Cephalic | ||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | N | N | N | N | N | N | |||
| Trans-vaginal | N | N | N | N | N | N | |||
| Trans-urethral | |||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | |||
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | |||
| Intravascular | |||||||||
| Other (Specify) ** | N | N | N | N | N | N | |||
| Adult Cardiac | N | N | N | N | N | N | N | ||
| Pediatric Cardiac | N | N | N | N | N | N | N | ||
| Cardiac | Intravascular(Cardiac) | ||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra- cardiac | |||||||||
| Peripheralvascular | Peripheral vascular | N | N | N | N | N | N | ||
| vascular | Other (Specify) |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication: P = previously cleared by FDA: E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW, B+CW, B+Color+CW, B+PDI/DPDI+CW,
Note * Small Organ includes Thyroid, Testes, Breast
| [1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging | ||||||||
|---|---|---|---|---|---|---|---|---|
[2]: Biopsy Guidance .
[3]: Harmonic Imaging, This feature does not use contrast agent.
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Prescription Use (Per 21 CFR 801.109)
{4}------------------------------------------------
ACCLARIX LX8 with L10-4D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify) [2][3] | |
| Ophthalmic | Ophthalmic | Ophthalmic | |||||||
| Fetal / Obstetrics | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro logical) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| FetalImaging& Other | Small Organ (Specify) * | N | N | N | N | N | N | N | |
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | N | ||
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | N | ||
| Intravascular | |||||||||
| Other (Specify) ** | |||||||||
| Cardiac | Adult Cardiac | ||||||||
| Pediatric Cardiac | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra- cardiac | |||||||||
| Peripheralvascular | Peripheral vascular | N | N | N | N | N | N | N | |
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW Note * Small Organ includes Thyroid, Testes, Breast.
** Other use includes Urology, Gynecology.
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{5}------------------------------------------------
ACCLARIX LX8 with E8-4D Transducer
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | ||
|---|---|---|
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify) [2][3] |
| Ophthalmic | Ophthalmic | |||||||
| Fetal / Obstetrics | N | N | N | N | N | N | ||
| Abdominal | ||||||||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro logical) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| FetalImaging& Other | Small Organ (Specify) * | |||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | N | N | N | N | N | N | ||
| Trans-vaginal | N | N | N | N | N | N | ||
| Trans-urethral | ||||||||
| Musculo-skeletal(Conventional) | ||||||||
| Musculo-skeletal (Superficial) | ||||||||
| Intravascular | ||||||||
| Other (Specify) ** | ||||||||
| Adult Cardiac | ||||||||
| Pediatric Cardiac | ||||||||
| Cardiac | Intravascular(Cardiac) | |||||||
| Trans-esoph.(Cardiac) | ||||||||
| Intra- cardiac | ||||||||
| Peripheralvascular | Peripheral vascular | |||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW * Small Organ includes Thyroid, Testes, Breast Note
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging ,DPDI: Directional Power Doppler Imaging
「2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{6}------------------------------------------------
ACCLARIX LX8 with L17-7HD Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify) [2][3] | |
| Ophthalmic | Ophthalmic | ||||||||
| Fetal / Obstetrics | |||||||||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro logical) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| FetalImaging& Other | Small Organ (Specify) * | N | N | N | N | N | N | ||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | |||
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | |||
| Intravascular | |||||||||
| Other (Specify) ** | |||||||||
| Cardiac | Adult Cardiac | ||||||||
| Pediatric Cardiac | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra- cardiac | |||||||||
| Peripheralvascular | Peripheral vascular | N | N | N | N | N | N | ||
| Other (Specify) |
N = new indication: P = previously cleared by FDA: E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW Note * Small Organ includes Thyroid, Testes, Breast
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{7}------------------------------------------------
ACCLARIX LX8 with C5-2XD Transducer
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | |||
|---|---|---|---|
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify) [2][3] |
| Ophthalmic | Ophthalmic | |||||||
| Fetal / Obstetrics | N | N | N | N | N | N | ||
| Abdominal | N | N | N | N | N | N | ||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro logical) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| FetalImaging& Other | Small Organ (Specify) * | |||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | ||
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | ||
| Intravascular | ||||||||
| Other (Specify) ** | N | N | N | N | N | N | ||
| Adult Cardiac | ||||||||
| Pediatric Cardiac | ||||||||
| Cardiac | Intravascular(Cardiac) | |||||||
| Trans-esoph.(Cardiac) | ||||||||
| Intra- cardiac | ||||||||
| Peripheralvascular | Peripheral vascular | |||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW * Small Organ includes Thyroid, Testes, Breast Note
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
「2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Prescription Use (Per 21 CFR 801.109)
{8}------------------------------------------------
ACCLARIX LX8 with P5-1XD Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined (Specify) [1] | Other (Specify) [2][3] |
| Ophthalmic | Ophthalmic | |||||||
| Fetal / Obstetrics | ||||||||
| Abdominal | N | N | N | N | N | N | N | |
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro logical) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| FetalImaging& Other | Small Organ (Specify) * | |||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Musculo-skeletal(Conventional) | ||||||||
| Musculo-skeletal (Superficial) | ||||||||
| Intravascular | ||||||||
| Other (Specify) ** | ||||||||
| Adult Cardiac | N | N | N | N | N | N | N | |
| Pediatric Cardiac | N | N | N | N | N | N | N | |
| Cardiac | Intravascular(Cardiac) | |||||||
| Trans-esoph.(Cardiac) | ||||||||
| Intra- cardiac | ||||||||
| Peripheralvascular | Peripheral vascular | |||||||
| Other (Specify) |
N = new indication: P = previously cleared by FDA: E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+CW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW, B+Color+CW, B+PDI/DPDI+CW
Note * Small Organ includes Thyroid, Testes, Breast
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{9}------------------------------------------------
ACCLARIX LX8 with L17-7SD Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify) [2][3] |
| Ophthalmic | Ophthalmic | |||||||
| Fetal / Obstetrics | ||||||||
| Abdominal | ||||||||
| Intra-operative (Specify) | N | N | N | N | N | N | ||
| Intra-operative (Neuro logical) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (Specify) * | ||||||||
| FetalImaging& Other | Neonatal Cephalic | |||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Musculo-skeletal(Conventional) | N | N | N | N | N | N | ||
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | ||
| Intravascular | ||||||||
| Other (Specify) ** | ||||||||
| Adult Cardiac | ||||||||
| Pediatric Cardiac | ||||||||
| Cardiac | Intravascular(Cardiac) | |||||||
| Trans-esoph.(Cardiac) | ||||||||
| Intra- cardiac | ||||||||
| Peripheralvascular | Peripheral vascular | |||||||
| Other (Specify) |
N = new indication: P = previously cleared by FDA: E = added under this appendix PDI=Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW Note * Small Organ includes Thyroid, Testes, Breast
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{10}------------------------------------------------
ACCLARIX LX8 with C5-2MD Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PW | CW | Color | Combined(Specify) [1] | Other(Specify)[2]3 | |
| Ophthalmic | Ophthalmic | ||||||||
| FetalImaging& Other | Fetal / Obstetrics | N | N | N | N | N | N | ||
| Abdominal | |||||||||
| Intra-operative (Specify) | |||||||||
| Intra-operative (Neuro logical) | |||||||||
| Laparoscopic | |||||||||
| Pediatric | |||||||||
| Small Organ (Specify) * | |||||||||
| Neonatal Cephalic | |||||||||
| Adult Cephalic | |||||||||
| Trans-rectal | |||||||||
| Trans-vaginal | |||||||||
| Trans-urethral | |||||||||
| Musculo-skeletal(Conventional) | |||||||||
| Musculo-skeletal (Superficial) | |||||||||
| Intravascular | |||||||||
| Other (Specify) ** | |||||||||
| Cardiac | Adult Cardiac | ||||||||
| Pediatric Cardiac | |||||||||
| Intravascular(Cardiac) | |||||||||
| Trans-esoph.(Cardiac) | |||||||||
| Intra- cardiac | |||||||||
| Peripheralvascular | Peripheral vascular | ||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix PDI= Power Doppler Imaging Additional comments: Combined mode: B+M, B+PW, B+Color, B+PDI/DPDI, B+Color+PW, B+PDI/DPDI +PW Note * Small Organ includes Thyroid, Testes, Breast
** Other use includes Urology, Gynecology
[1]: PDI: Power Doppler Imaging, DPDI: Directional Power Doppler Imaging
[2]: Biopsy Guidance
[3]: Harmonic Imaging, This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
{11}------------------------------------------------
510(k) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc. | ||
|---|---|---|---|
| 3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#,Shekou, Nanshan Shenzhen, 518067 P.R. China | |||
| Tel.: (0755) 26858739 | |||
| Fax: +1 (408) 418-4059 | |||
| Contact Person: | Queena Chen | ||
| Date prepared: | March 28, 2016 | ||
| 2. Device nameand classification: | Device Name: Diagnostic Ultrasound SystemModel: Acclarix LX8 | ||
| Classification Name: | |||
| 892.1550 | System, Imaging, Pulsed Doppler, UltrasonicProduct code: IYN | ||
| 892.1560 | Ultrasonic, Pulsed echo, ImagingProduct code: IYO | ||
| 892.1570 | Transducer, Ultrasonic, DiagnosticProduct code: ITX | ||
| Regulatory Class: Class II | |||
| 3.PremarketNotification ClassIII Certificationand Summary | Not applicable, the subject device is Class II. | ||
| 4. PredicateDevice(s): | Acclarix AX8 Diagnostic Ultrasound System/ K150999/Shenzhen EDAN Instruments, Inc. | ||
| ZS3 Ultrasound System / K141641 (Reference) / ZONAREMedical Systems, Inc | |||
| 5. Reason forSubmission | Acclarix LX8 is a new device. | ||
| 6. Pre-Submission,IDE | Not applicable, there is no prior submission. |
{12}------------------------------------------------
| 7. DeviceDescription: | The Edan Acclarix LX8 Ultrasound system consists of a mainsystem along with associated transducers. |
|---|---|
| The system circuitry generates an electronic voltage pulse,which is transmitted to the transducer. In the transducer, apiezo electric array converts the electronic pulse into anultrasonic pressure wave. When coupled to the body, thepressure wave transmits through body tissues. The waves arethen reflected within the body and detected by the transducer,which then converts back to an electrical signal. The AcclarixLX8 system then analyzes the returned signal to generate animage or conduct Doppler processing. | |
| The Acclarix LX8 system gives the operator the ability tomeasure anatomical structures, and offers analysis packagesthat provide information used by competent health careprofessionals to make a diagnosis. | |
| The system provides both touch screen and hard buttons for theUser Interface. | |
| 8. Intended Use: | The ACCLARIX LX8 Diagnostic Ultrasound System isintended for use by a qualified physician or sonographer forultrasound evaluation. Clinical applications include:Abdominal, Gynecology (including endovaginal), Obstetric,Cardiac, Small parts (Breast, Testes, Thyroid, etc.), Urology,Musculoskeletal, and Peripheral vascular, and Intra-operative. |
9. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table:
| Item | Acclarix LX8 DiagnosticUltrasound System(Edan Instruments) | Acclarix AX8 DiagnosticUltrasound System (EdanInstruments) | ZS3 Ultrasound Systems(ZONARE MedicalSystems) |
|---|---|---|---|
| 510(k) Number | Current Submission | K150999 | K141641 |
| Intended Use | Diagnostic ultrasoundimaging or fluid flowanalysis of the human body | Same | Same |
| Indications forUse | The Acclarix LX8Diagnostic UltrasoundSystem is intended for useby a qualified physician orsonographer for ultrasoundevaluation. Clinicalapplications include: | Same | The ZS3 Ultrasound Systemis intended for use by aqualified physician forultrasound evaluation ofOphthalmic; Fetal/obstetric,gynecological; Abdominal(renal, GYN/Pelvic); |
| Abdominal, Gynecology(including endovaginal),Obstetric, Cardiac, Smallparts (Breast, Testes,Thyroid, etc.), Urology,Musculoskeletal, andPeripheral vascular, andIntra-operative. | Intra-operative (abdominal,thoracic, and vascular),Intra-operative neurological;Pediatric: small organ(thyroid, breast, testes, etc.),Adult & Neonatal Cephalic;Trans-rectal, Trans-vaginal,Trans-cranial,Trans-esophageal(non-cardiac and cardiac);Musculoskeletal(conventional &superficial); 3D/4D; Cardiac- Adult/ Pediatric/ Fetal;Echo, Intra-Cardiac; Pelvic;Peripheral vascular;harmonic tissue and contrastimaging and Tissueelasticity. | ||
| Installation andUse | Trolley Mobile Equipment | Portable (laptop) MobileEquipment | Same |
| SafetyStandards | IEC 60601-1IEC 60601-1-2IEC 60601-2-37ISO 10993-1, -5, -10, -12AIUM, NEMA UD 2, UD3 | Same | Same |
| Patient ContactMaterials | Complies with ISO 10993 | Same | Same |
| Mode ofOperations | B-Mode, M-Mode, Color,PDI/DPDI, PW, CW, 3D(Note 1) | Same except 3D | B-Mode, M-Mode, PWDMode, CWD, CD Mode,Elastorgraphy, ContrastEnhanced, 3D/4D, ECG(forcardiac cycle referencedtiming only)Combined Modes includeB+CD, B+PW, B+CD+PW,B+M, M+CM,B+CD+M+CM,B+Elastorgraphy, B+CEUS,and + ECG Trace |
| Measurements | B-Mode: Distance,Circ/Area, Angle, Volume,Stenosis ratioM-Mode: Distance, Time,Slope and Heart Rate | Same | B-mode (2D): Depth,Distance, Cire/Area/VolumeM-Mode: Depth, Distance,HR |
| D-Mode: Velocity, RI,Time, PI, Heart Rate, AutoTrace, PG, S/D, ΔV,Acceleration, PHT, VTI | PWD(Manual): Velocity,Velocity Pairs, RI, Accl,S/D, A/B, PI, HR/PWD(AutoTrace: RI, PI,Accl, S/D, HR, AT, TAMX) | ||
| and TAMN) | |||
| Principle ofOperation | Applying high voltage burstto the Piezoelectric materialin the transducer and detectreflected echo to constructdiagnostic image | Same | Same |
| AcousticOutput | Track 3: MI, TIS, TIC,TIB(TI Range 0-6.0)Derated ISPTA: 720 W/cm²maximum, Mechanic Index<1.9 maximum or DeratedISPPA 190 W/cm² max | Same | Same |
| TransducerTypes | Convex ArrayLinear ArrayEndocavity-Micro ConvexArrayPhased Array | Same | Linear ArrayCurved Linear ArrayPhased ArrayTrans-esophagealPencil ProbeIntracavitary |
| TransducerFrequency | 2.5-15.0 MHz | Same | 1.0 - 20.0 MHz |
| PrimaryDisplay | Primary Screen: 21.5 inch(1920 x 1080)(Note 2) | Primary Screen: 15 inch(1920x1080) | Color 19" Liquid CrystalDisplay (LCD) |
| TransducerPorts | Multi-Transducer Port(Four)(Note 3) | One | Multi-Transducer Port(Three) |
| Dimensions/Weight | Height: 53 - 70 in(1355-1780mm)Width: 23 in (585mm)Depth: 36.6 in (930mm)Weight: 111 Kg(noperipherals)(Note 4) | Height: 3 in (77mm)Width: 16 in (407mm)Depth: 15.3 in (388mm)Weight: 9.1Kg(withrechargeable battery,without power adaptor ortransducers.) | Height. max (in operationaluse) 157.5cm (62in)Height. min (in operationaluse) 128cm (50.5in)Height min (displayed lowerfor transport) 104cm (41in)Width: 51cm (21.1in)Depth: 72cm(28.2)Weight: 65.3kg (144lb) |
| PowerRequirements | 100-240V, 50/60Hz, 2.5Amax | Same | 100-240V options,~50-60Hz, 6A max |
{13}------------------------------------------------
{14}------------------------------------------------
The subject device has same intended use, similar product design, same performance effectiveness, and performance safety as the predicate device.
The differences between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
10. Effectiveness and Safety Considerations:
Clinical test:
Clinical testing is not required.
{15}------------------------------------------------
Non-clinical test:
The Acclarix LX8 Ultrasound Imaging System complies with:
(1) IEC 60601-1 Electrical Safety
(2) IEC 60601-1-2 Electromagnetic Compatibility
(3) IEC 60601-2-37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
(4) NEMA UD 2 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment (5) NEMA UD 3 Standard for real-time display of thermal and mechanical acoustic output indies on diagnostic ultrasound equipment.
(6) Acoustic output testing as per the guideline "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008.
The following biocompatibility standards are conducted on the subject device: (1) ISO 10993-1, ISO 10993-5, ISO 10993-10 and ISO 10993-12
The tests were selected to show substantial equivalence between the subject device and the predicate.
11. Substantially Equivalent Determination
Verification and validation testing has been conducted on the Acclarix LX8 Ultrasound Imaging System. This premarket notification submission demonstrates that Acclarix LX8 Ultrasound Imaging System is substantially equivalent to the predicate devices.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.