K Number
K160481
Device Name
Vivo 60
Manufacturer
Date Cleared
2016-11-09

(261 days)

Product Code
Regulation Number
868.5895
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Vivo 60 ventilator (with or without the iOxy and CO2 sensor) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.) The Vivo 60 with the iOxy is intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. The Vivo 60 with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas. The device is intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 60 is not intended to be used as a transport or critical care ventilator.
Device Description
The Vivo 60 Ventilator is a portable, microprocessor controlled turbine based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation. Flow and pressure are read through flow and pressure sensors. Essential parameters such as pressure, flow and volume are presented on the ventilator screen, both as graphs and numbers. Operator actions are performed via the front panel where the buttons and an LCD screen are located. There are dedicated LEDs and buttons for managing alarm conditions and an Information button which provides integrated user support. The Vivo 60 can be operated by external AC or DC power supply and contains an integrated battery as well as an additional click on battery. The Vivo 60 can be used with three types of patient circuits: single limb patient circuits including an active exhalation valve, single limb patient circuits including a passive leakage port, and dual limb patient circuits.
More Information

Not Found

No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the device description focuses on microprocessor control and sensor-based measurements without mentioning AI/ML capabilities.

Yes
The device is a ventilator, which provides mechanical ventilation, directly treating individuals who require respiratory support. This aligns with the definition of a therapeutic device designed to treat or alleviate a medical condition.

Yes

The device, when equipped with the iOxy or CO2 sensor, is intended to measure functional oxygen saturation (%SpO2), pulse rate, and CO2 in inspiratory and expiratory gas, which are diagnostic functions.

No

The device description clearly outlines hardware components such as a microprocessor, turbine, sensors (flow, pressure, iOxy, CO2), screen, buttons, LEDs, and power sources (AC, DC, integrated battery, click-on battery). The performance studies also include testing related to hardware aspects like electrical safety, biocompatibility, and battery operation.

Based on the provided information, the Vivo 60 ventilator is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The primary intended use of the Vivo 60 is to provide mechanical ventilatory support to patients. While it can measure physiological parameters like SpO2 and CO2, these measurements are used to monitor the patient's respiratory status and adjust the ventilation, not to diagnose a condition based on in vitro examination of specimens.
  • Device Description: The description focuses on the mechanical and electronic aspects of the ventilator, including sensors for flow and pressure, and controls for ventilation parameters.
  • Lack of IVD Characteristics: The description does not mention the analysis of biological specimens (like blood, urine, or tissue) or the use of reagents, which are hallmarks of IVD devices.
  • Performance Studies: The performance studies described focus on the functionality and safety of the ventilator itself, as well as the accuracy of the integrated sensors for SpO2 and CO2 measurement in the context of ventilation. They do not involve studies related to the diagnostic accuracy of analyzing biological samples.

In summary, the Vivo 60 is a life-support device that assists with breathing. While it incorporates sensors to monitor physiological parameters, these are used for patient management during ventilation, not for in vitro diagnostic purposes.

N/A

Intended Use / Indications for Use

The Vivo 60 ventilator (with or without the iOxy and CO2 sensor) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.)

The Vivo 60 with the iOxy is intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate.

The Vivo 60 with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas.

The device is intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 60 is not intended to be used as a transport or critical care ventilator.

Product codes (comma separated list FDA assigned to the subject device)

NOU, CBK, DQA, CCK

Device Description

The Vivo 60 Ventilator is a portable, microprocessor controlled turbine based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation.

Flow and pressure are read through flow and pressure sensors. Essential parameters such as pressure, flow and volume are presented on the ventilator screen, both as graphs and numbers.

Operator actions are performed via the front panel where the buttons and an LCD screen are located. There are dedicated LEDs and buttons for managing alarm conditions and an Information button which provides integrated user support.

The Vivo 60 can be operated by external AC or DC power supply and contains an integrated battery as well as an additional click on battery.

The Vivo 60 can be used with three types of patient circuits: single limb patient circuits including an active exhalation valve, single limb patient circuits including a passive leakage port, and dual limb patient circuits.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

pediatric through adult patients weighing more than 5 kg (11 lbs.)

Intended User / Care Setting

intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance Testing:
Waveform performance testing was conducted comparing the Vivo 60 to Vivo 50 and the Vivo 60 to the Astral. Characteristics tested included flow, pressure and volume waveforms. The comparison of the recorded waveforms supports the claim that Vivo 60 is substantially equivalent to the predicate devices.

Triggering testing of Vivo 60 was performed which showed that the Vivo 60 performed as intended, detecting each patient effort within the permissible trigger delay without false-triggers.

Testing of the Vivo 60 was performed to confirm accuracy of controls and monitored values. The testing confirmed that the Vivo 60 meets its accuracy specifications.

Alarms testing of the Vivo 60 was performed which confirmed proper operation of physiologic and technical alarms.

Power management testing confirmed proper operation of the Vivo 60 power management system including transitioning between the different internal and external power sources, power source alarms, and battery operating time.

Treatment and alarm settings testing confirmed the range and operation of settings for all treatment and alarm parameters conform to specifications.

Cybersecurity testing conformance with all cybersecurity specifications.

Software verification and validation were performed at the unit, integration, and system level according to plans and protocols with predetermined pass/fail criteria. All tests passed.

Summative usability / human factors testing was performed including critical tasks associated with the changes from the Vivo 50 to the Vivo 60.

Testing for immunity to emissions from RFID sources was conducted at 125kHz, 134.2kHz, 13.56MHz, 902MHz and using an Electronic Article Surveillance system.

Cleaning validation was performed to ensure no physical or performance degradation occurred.

Performance testing conducted with the Vivo 60 demonstrates compliance with recognized standards and product specifications. This testing including comparative testing with the predicate devices supports use of the device for its intended use and in its intended environments.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K123144, K133868

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three heads, representing the department's mission to protect the health of all Americans. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 9, 2016

Breas Medical AB % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, Massachusetts 01864

Re: K160481

Trade/Device Name: Vivo 60 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: NOU, CBK, DQA, CCK Dated: October 12, 2016 Received: October 13, 2016

Dear Ms. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

1

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K160481

Device Name Vivo 60

Indications for Use (Describe)

The Vivo 60 ventilator (with or without the iOxy and CO2 sensor) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.)

The Vivo 60 with the iOxy is intended to measure function of arterial hemoglobin (%SpO2) and pulse rate.

The Vivo 60 with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas.

The device is intended to be used in home, institution, hospitals applications such as wheelchairs and gurneys. It may be used for both invasive ventilation. The Vivo 60 is not intended to be used as a transport or critical care ventilator.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Date Summary Prepared:November 8, 2016
Submitter Information:
Company Name:Breas Medical AB
Company Address:Företagsvagen 1
Mölnlycke Vastra Gotalands Lan (Se-14)
SWEDEN 435 33
Contact Person:Maureen O'Connell
O'Connell Regulatory Consultants, Inc.
5 Timber Lane
North Reading, MA 01864
Telephone: 978-207-1245
Device Information:
Trade Name:Vivo 60
Common Name:Portable Ventilator
Classification Name:Continuous Ventilator, Home Use
Product code: NOU
21 CFR §868.5895
Additional product codes:
Continuous Ventilators, Facility Use
Product code: CBK
21 C.F.R. §868.5895.
Oximeters
Product code: DQA
21 C.F.R. §870.2700.
Carbon Dioxide Gas Analyzer
Product code CCK
21 C.F.R. §868.1400.
Device Class:Class II
Predicate Devices:
Device: Vivo 50
510(k) Number: K123144
Manufacturer: Breas Medical
Device: Astral 110/150
510(k) Number: K133868
Manufacturer: ResMed Corporation

4

Intended Use:

To provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Indications for Use:

The Vivo 60 ventilator (with or without the iOxy and CO2 sensor) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.)

The Vivo 60 with the iOxy is intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate.

The Vivo 60 with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas.

The device is intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 60 is not intended to be used as a transport or critical care ventilator.

Device Description:

The Vivo 60 Ventilator is a portable, microprocessor controlled turbine based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation.

Flow and pressure are read through flow and pressure sensors. Essential parameters such as pressure, flow and volume are presented on the ventilator screen, both as graphs and numbers.

Operator actions are performed via the front panel where the buttons and an LCD screen are located. There are dedicated LEDs and buttons for managing alarm conditions and an Information button which provides integrated user support.

The Vivo 60 can be operated by external AC or DC power supply and contains an integrated battery as well as an additional click on battery.

The Vivo 60 can be used with three types of patient circuits: single limb patient circuits including an active exhalation valve, single limb patient circuits including a passive leakage port, and dual limb patient circuits.

Technological Characteristics Compared to Predicate:

The indications for use for the Vivo 60 and the Vivo 50 are identical with the exception that the Vivo 60 can be used in pediatric patients with a weight as low as 5 kg while the Vivo 50 is used in patients beginning at 10 kg. The Astral 100/150 has similar indications for use and can also be used in patients to 5 kg. All three of the devices are used in home,

5

institution and portable applications. Both the Vivo 60 and Vivo 50 are not to be used as a transport or critical care ventilator.

The Vivo 60, Vivo 50 and Astral ventilators have technological characteristics. All three are software controlled devices that receive ambient air through an inlet, utilize a turbine to generate the required pressures and flows, and pass the air to the patient via an outlet. All three devices provide a bleed-in connection for a low pressure, low flow supplemental oxygen supply. There are minor differences between the Vivo 60 and the Vivo 50 including: addition of compatibility with dual limb patient circuits, separate adult and pediatric patient types, addition of Synchronized Intermittent Mandatory Ventilation breath-mode ("SIMV"), and minor changes to performance specifications.

The Vivo 50. Vivo 60 and Astral ventilators can all be used with single limb with leak patient circuits, and single limb with active exhalation valve patient circuits. Additionally, the Vivo 60 and Astral can be used with dual limb patient circuits. The Vivo 60, Vivo 50, and Astral ventilators are for use by pediatric through adult patients. The Vivo 60 and the Astral have selectable "Pediatric" and "Adult" patient type settings. The Vivo 50 has a single fixed patient type setting to be used for both pediatric and adult patients. The selected patient type only determines the range of available settings (but does not alter the ventilation algorithms). The Vivo 60, Vivo 50 and Astral have nine ventilation modes in common, including pressure support ventilation (with and without target volume), pressure controlled ventilation (with and without target volume and/or assist), volume controlled ventilation (with or without assist), and CPAP. The ventilation algorithms utilized within these nine ventilation modes are identical between the Vivo 60 and Vivo 50. The Vivo 60 and Astral have two additional ventilation modes in common: pressure controlled ventilation with SIMV, and volume controlled ventilation with SIMV. These SIMV modes in the Vivo 60 utilize the same algorithms for delivering individual breaths as in the Vivo 50. In both the Vivo 60 and Astral, the SIMV modes are mixed ventilation modes delivering pressure-controlled or volume-controlled mandatory breaths, and pressure-supported spontaneous breaths.

The user interface, indicators and physical specifications are identical between the Vivo 60 and Vivo 50 as are power management and environmental characteristics.

Although there are differences between the Vivo 60 and the predicate devices, they do not impact safety and effectiveness. Therefore, the Vivo 60 is substantially equivalent to the Vivo 50 and Astral predicate devices.

Performance Testing:

The Vivo 60 was subjected to performance testing which verified conformance with all requirements specifications and applicable standards, and which included comparative testing with the Vivo 50 and Astral predicate devices which supported substantial equivalence. Performance testing included testing to the standards and procedures listed below:

6

Performance Testing to Standards
Electrical SafetyIEC 60601-1: 2005+CORR. 1:2006+CORR.2:
2007+AM1:2012 Medical Electrical Equipment: General
requirements for basic safety and essential performance
Electromagnetic
compatibilityIEC 60601-1-2: 2014 (Ed. 4) Medical electrical equipment -
Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic
compatibility
Alarms systemsIEC 60601-1-8: 2006 General requirements for basic safety
and essential performance – Collateral Standard: General
requirements, tests and guidance for alarm systems in
medical electrical equipment and medical electrical systems
Medical equipment used
in home healthcare
environmentIEC 60601-1-11: 2010 General requirements for basic safety
and essential performance-Collateral Standard:
Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
Respiratory gas monitorsISO 80601-2-55: 2011 (1st Ed) Medical electrical
equipment-Part 2-55: Particular requirements for the basic
safety and essential performance of respiratory gas monitors
Pulse oximeter
equipmentISO 80601-2-61:2011 (1st Ed) Medical electrical equipment-
Part 2-61: Particular requirements for basic safety and
essential performance of pulse oximeter equipment
Home healthcare
environment ventilators
for ventilator-dependent
patientsISO 80601-2-72:2015 (1st Ed) Medical electrical equipment-
Part 2-72: Particular requirements for basic safety and
essential performance of home healthcare environment
ventilators for ventilator-dependent patients
Rough handling shocksIEC 60068-2-31:2008 Environmental testing. Part 2-31:
Tests - Rough handling shocks, primarily for equipment-
type specimens. Drop and Topple test and Free Fall –
procedure 1.
Battery testingIEC 62133: 2012 (2nd Ed) Secondary cells and batteries
containing alkaline or other non-acid electrolytes – Safety
requirements for portable sealed secondary cells, and for
batteries made from them, for use in portable applications
BiocompatibilityISO 10993-1:2009 Biological evaluation of medical
devices-Part 1: Evaluation and testing within a risk
management process.

Biocompatibility testing performed:
• Cytotoxicity
• Sensitization
• Irritation/intracutaneous reactivity
• Acute systemic toxicity
• Pyrogenicity |
| | |
| | Subacute/subchronic toxicity Genotoxicity Implantation Hemocompatibility |
| VOC testing | Compendium Method TO-15 "Determination Of Volatile
Organic Compounds (VOCs) In Air Collected In Specially-
Prepared Canisters And Analyzed By Gas
Chromatography/Mass Spectrometry (GC/MS)", U.S. EPA,
and
ASTM D5466-01 Standard test method for determination of
volatile organic compounds in atmospheres (canister
sampling methodology) |
| Particulates testing | EPA PM2.5 standard |

7

Performance Testing

Waveform performance testing was conducted comparing the Vivo 60 to Vivo 50 and the Vivo 60 to the Astral. Characteristics tested included flow, pressure and volume waveforms. The comparison of the recorded waveforms supports the claim that Vivo 60 is substantially equivalent to the predicate devices.

Triggering testing of Vivo 60 was performed which showed that the Vivo 60 performed as intended, detecting each patient effort within the permissible trigger delay without false-triggers.

Testing of the Vivo 60 was performed to confirm accuracy of controls and monitored values. The testing confirmed that the Vivo 60 meets its accuracy specifications.

Alarms testing of the Vivo 60 was performed which confirmed proper operation of physiologic and technical alarms.

Power management testing confirmed proper operation of the Vivo 60 power management system including transitioning between the different internal and external power sources, power source alarms, and battery operating time.

Treatment and alarm settings testing confirmed the range and operation of settings for all treatment and alarm parameters conform to specifications.

Cybersecurity testing conformance with all cybersecurity specifications.

Software verification and validation were performed at the unit, integration, and system level according to plans and protocols with predetermined pass/fail criteria. All tests passed.

Summative usability / human factors testing was performed including critical tasks associated with the changes from the Vivo 50 to the Vivo 60.

Testing for immunity to emissions from RFID sources was conducted at 125kHz, 134.2kHz, 13.56MHz, 902MHz and using an Electronic Article Surveillance system.

Cleaning validation was performed to ensure no physical or performance degradation occurred.

Performance testing conducted with the Vivo 60 demonstrates compliance with recognized standards and product specifications. This testing including comparative

8

testing with the predicate devices supports use of the device for its intended use and in its intended environments.

Conclusion:

The Vivo 60 is substantially equivalent to the predicate devices, as the devices share a common intended use, and technological differences between the new devices and the predicates do not raise new questions of safety and effectiveness.