K Number
K160412
Date Cleared
2016-12-14

(306 days)

Product Code
Regulation Number
862.1120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GEM Premier 5000 is a portable critical care system for use by health care professionals to rapidly analyze heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH and pO2 from venous, arterial and capillary heparinized whole blood, as well as quantitative measurements of pCO2 from venous and arterial heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status.

pH, pCO2 and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Device Description

The GEM Premier 5000 system provides health care professionals in central laboratory or point-of-care clinical settings with fast, accurate, quantitative measurements of pH and pQ2 from venous, arterial and capillary heparinized whole blood, as well as quantitative measurements of pCO₂ from venous and arterial heparinized whole blood.

Key Components:
Analyzer: Employs a unique color touch screen and a simple set of menus and buttons for user interaction. The analyzer guides operators through the sampling process with simple, clear messages and prompts.
GEM Premier 5000 PAK (disposable, multi-use GEM PAK): Houses all required components necessary to operate the instrument once the cartridge is validated. These components include the sensors, CO-Ox optical cell, Process Control (PC) Solutions, sampler, pump tubing, distribution valve and waste bag. The GEM PAK has flexible menus and test volume options to assist facilities in maximizing efficiency. NOTE: The EEPROM on the GEM PAK includes all solution values and controls the analyte menu and number of tests.
Step 1: After inserting the GEM PAK, the instrument will perform an automated PAK warm-up during which the sensors are hydrated and a variety of checks occur, all of which take about 40 minutes. During warm-up, the instrument requires no user intervention.
Step 2: After GEM PAK warmup, Auto PAK Validation (APV) process is automatically completed: two completely independent solutions (PC Solution D and E) that are traceable to NIST standards, CLSI procedures or internal standards, containing two levels of concentration for each analyte, are run by the analyzer to validate the integrity of the PC Solutions and the overall performance of the analytical system.
Step 3: After successful performance of APV, iQM2 manages the quality control process, replacing external quality controls.
Intelligent Quality Management 2 (iQM2): iQM2 is an active quality process control program designed to provide continuous monitoring of the analytical process before, during and after sample measurement with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions. iQM2 is a statistical process control system that performs 5 types of continuous, quality checks to monitor the performance of the GEM PAK, sensors, CO-Ox, and reagents. These checks include System, Sensor, IntraSpect, Pattern Recognition and Stability Checks.

AI/ML Overview

The GEM Premier 5000 is a medical device designed for the rapid analysis of heparinized whole blood samples for pH, pCO2, and pO2 measurements. The provided document is a 510(k) summary, which demonstrates substantial equivalence to a predicate device (GEM Premier 4000) for these parameters.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct numerical targets in the document. Instead, the studies aim to demonstrate that the device performs "within specification" or shows "substantial equivalency" to the predicate device and recognized standards (CLSI guidelines). The reported performance is summarized below, with an implied acceptance that these values fall within expected clinical or analytical ranges and compare favorably to the predicate.

The reported performance values are presented as part of precision studies, linearity studies, analytical specificity (absence of interference), and method comparison against predicate devices or reference methods. The "acceptance criteria" are implicitly met if the reported values demonstrate acceptable precision, linearity, and agreement with reference methods/predicate devices.

Metric (Analyte)Acceptance Criteria (Implied / Contextual)Reported Device Performance (GEM Premier 5000)Study Reference
pH Precision (Aqueous Controls)Low SD/CV (within specification)Level 1: Mean 7.14, Total SD 0.008, Total %CV 0.1%Level 2: Mean 7.38, Total SD 0.006, Total %CV 0.1%Level 3: Mean 7.57, Total SD 0.003, Total %CV 0.0%Internal Precision Study - Aqueous Controls
pCO2 Precision (Aqueous Controls)Low SD/CV (within specification)Level 1: Mean 87, Total SD 2.3, Total %CV 2.7%Level 2: Mean 35, Total SD 0.8, Total %CV 2.3%Level 3: Mean 14, Total SD 0.3, Total %CV 2.3%Internal Precision Study - Aqueous Controls
pO2 Precision (Aqueous Controls)Low SD/CV (within specification)Level 1: Mean 31, Total SD 2.4, Total %CV 7.9%Level 2: Mean 88, Total SD 2.2, Total %CV 2.4%Level 3: Mean 370, Total SD 5.9, Total %CV 1.6%Internal Precision Study - Aqueous Controls
pH Precision (Whole Blood)Low SD/CV (within specification)Normal Mode: Total SD 0.005-0.014, Total %CV 0.1-0.2%Micro Mode: Total SD 0.007-0.015, Total %CV 0.1-0.2%Internal Precision Study - Whole Blood
pCO2 Precision (Whole Blood)Low SD/CV (within specification)Normal Mode: Total SD 0.8-2.9, Total %CV 1.5-7.7%Internal Precision Study - Whole Blood
pO2 Precision (Whole Blood)Low SD/CV (within specification)Normal Mode: Total SD 0.9-32.7, Total %CV 1.2-4.3%Micro Mode: Total SD 0.9-34.0, Total %CV 1.4-4.8%Internal Precision Study - Whole Blood
pH Reproducibility (POC)Low SD/CV (within specification)GSE 1: Reproducibility SD 0.002, %CV 0.0%GSE 2: Reproducibility SD 0.008, %CV 0.1%GSE 3: Reproducibility SD 0.003, %CV 0.0%Reproducibility Study - POC Setting
pCO2 Reproducibility (POC)Low SD/CV (within specification)GSE 1: Reproducibility SD 1.1, %CV 1.2%GSE 2: Reproducibility SD 0.9, %CV 2.7%GSE 3: Reproducibility SD 0.5, %CV 3.5%Reproducibility Study - POC Setting
pO2 Reproducibility (POC)Low SD/CV (within specification)GSE 1: Reproducibility SD 1.9, %CV 6.7%GSE 2: Reproducibility SD 1.9, %CV 2.2%GSE 3: Reproducibility SD 9.3, %CV 2.6%Reproducibility Study - POC Setting
Linearity (pH)R² close to 1, slope close to 1, small interceptSlope: 0.972, Intercept: 0.191, R²: 0.998Linearity Study
Linearity (pCO2)R² close to 1, slope close to 1, small interceptSlope: 1.045, Intercept: -2.027, R²: 0.998Linearity Study
Linearity (pO2)R² close to 1, slope close to 1, small interceptSlope: 1.028, Intercept: -4.069, R²: 0.995Linearity Study
Analytical SpecificityNo observed interferenceNo interference with listed substances at specified concentrations for pH, pCO2, pO2.Analytical Specificity
Method Comparison (pH vs. Predicate)Good correlation with predicate (GEM Premier 4000) or reference (Tonometry)Slope: 0.953, Intercept: 0.344, R²: 0.993Internal Method Comparison
Method Comparison (pCO2 vs. Predicate)Good correlation with predicate (GEM Premier 4000) or reference (Tonometry)Slope: 1.026, Intercept: -0.991, R²: 0.997Internal Method Comparison
Method Comparison (pO2 vs. Predicate)Good correlation with predicate (GEM Premier 4000) or reference (Tonometry)Slope: 1.027, Intercept: -1.266, R²: 0.999Internal Method Comparison
Clinical Testing (POC vs. Predicate) - Normal ModeHigh correlation (r close to 1)pH: r=0.991; pCO2: r=0.991; pO2: r=0.996Clinical Testing
Clinical Testing (POC capillary) - Total ErrorBias within 95% CI of Bias, Total Error (TEa) within specificationpH: MDL 7.30 Bias 0.002, 95% CI (-0.015 to 0.020), TEa $\pm$0.04pCO2: Not provided for capillarypO2: MDL 30 Bias 6.1, 95% CI (1.3 to 11.6), TEa $\pm$9.0Clinical Testing

The total error calculations are compared against "GEM Premier 5000 Total Error Specifications," which are not explicitly provided in the document but are indicated as having been met.

2. Sample Sizes and Data Provenance

The document provides sample sizes for various tests:

  • Internal Precision Study - Aqueous Controls: n=120 per analyte per level (3 analyzers, 20 days, 2 runs/day, 1 replicate/run). Data provenance is internal (Instrumentation Laboratory Co.).
  • Internal Precision Study - GEM PAK Process Control Solutions D and E: n=120 per analyte per level (3 analyzers, 20 days, 2 runs/day, 1 replicate/run). Data provenance is internal (Instrumentation Laboratory Co.).
  • Internal Precision Study - Whole Blood: n=120 per analyte per sample mode per level (3 analyzers, 5 days, 1 run/day, 8 replicates/run). Data provenance is internal (Instrumentation Laboratory Co.).
  • Reproducibility Study with Aqueous Controls – POC Setting: n=90 per material/level (3 external POC sites, 9 operators, 3 instruments, 5 days, 2 runs/day, 3 replicates/run). Data provenance is external Point-of-Care sites within the USA (implied as no other country explicitly mentioned). This was a prospective study.
  • External Precision – Whole Blood:
    • POC Setting: n=30 to 54 per site (various N given per analyte/mode/site) for pH, pCO2, pO2. Pooled POC data: pH (N=102-126), pCO2 (N=117), pO2 (N=45-51). Performed by 11 operators on 3 instruments. Data provenance is 3 external POC locations, prospective.
    • Central Lab Setting: n=24 to 30 per site. Pooled lab data: pH (N=90), pCO2 (N=78), pO2 (N=66-72). Performed by 3 operators on 3 instruments. Data provenance is 2 external central laboratories and 1 internal Customer Simulation Laboratory (CSL). It includes native and contrived specimens. This was a prospective study.
  • LoB, LoD, and LoQ: LoB, LoD, LoQ were established using three (3) lots of GEM Premier 5000 PAKs (cartridges). Specific sample sizes for individual measurements within the establishment process are not detailed, but the method used is CLSI EP17-A2. Data provenance is internal (Instrumentation Laboratory Co.).
  • Linearity: 8 or 9 levels per analyte, 9 replicates per level (on 3 test analyzers). Total N per analyte: 72-81. Data provenance is internal (Instrumentation Laboratory Co.).
  • Analytical Specificity: Interference study with various substances. Specific N for each substance not detailed, but it involves screening tests at two different analyte concentrations. Data provenance is internal (Instrumentation Laboratory Co.).
  • Internal Method Comparison:
    • pH: N=373 clinical samples.
    • pCO2: N=150 clinical samples.
    • pO2: N=148 clinical samples.
      Data provenance is internal, using clinical samples. These were retrospective or prospective, but likely prospective as samples were "altered as needed to cover the medical decision levels."
  • Clinical Testing (Method Comparison):
    • Normal Mode (Syringe Samples): pH: N=479; pCO2: N=492; pO2: N=506. Pooled data from 3 external POC sites and 1 internal Customer Simulation Laboratory (CSL). Data includes patient samples. This was a prospective study.
    • Native Capillary Samples: N=171 for pH, N=167 for pO2. Collected at an external POC site (N=65) and an internal CSL (N=106). Data provenance is external POC site and internal CSL. This was a prospective study.
    • Pooled Capillary Results (Native + Contrived): pH: N=189; pO2: N=218. The contrived samples were prepared internally. Data provenance is external POC site, internal CSL, and internal contrived samples. This was a prospective study.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of human "experts" in the traditional sense (e.g., radiologists, pathologists) to establish ground truth for the device's performance. Instead, ground truth for quantitative measurements of blood gases and pH is established using:

  • Reference materials/standards: NIST standards, CLSI procedures, internal standards (for calibration and control solutions).
  • Predicate devices: The GEM Premier 4000 serves as the predicate for method comparison studies.
  • Reference methods: Tonometry is used as a reference for pO2 and pCO2 in Linearity and Internal Method Comparison studies due to instability of blood gases.

The quality control process (iQM2) also uses "Process Control (PC) Solutions" whose values are traceable to NIST standards, CLSI procedures, or internal standards.

4. Adjudication Method

No adjudication method (e.g., 2+1, 3+1 consensus) is described because the ground truth is established through quantitative measurements against reference materials, reference methods, or comparison to a predicate device, rather than subjective expert interpretation requiring consensus.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The device measures quantitative laboratory parameters (pH, pCO2, pO2) and does not involve human readers interpreting images or data where "AI assistance" would be relevant in the context of diagnostic performance improvement. The studies focus on analytical performance metrics like precision, accuracy, and linearity.

6. Standalone Performance Study

Yes, standalone performance was done. All the precision, linearity, LoB/LoD/LoQ, analytical specificity, and method comparison studies described are standalone performance studies. These studies evaluate the device's ability to accurately and precisely measure pH, pCO2, and pO2 in various conditions (aqueous controls, whole blood, different modes) independently. The "method comparison" studies compare the device's readings directly against a predicate device or a reference method.

7. Type of Ground Truth Used

The ground truth used for the studies is based on:

  • Calibration and Control Solutions: Values traceable to NIST standards, CLSI procedures, or internal standards.
  • Reference Methods: Tonometry for pCO2 and pO2 in linearity and method comparison studies.
  • Predicate Device: The GEM Premier 4000 for method comparison studies for pH.
  • Clinical Samples: Patient samples (heparinized whole blood, native capillary samples) are used for precision and method comparison studies, with their "true" values established by the reference methods or predicate device for comparison. For some internal precision studies and clinical testing, contrived samples are also used to extend the tested range.

8. Sample Size for the Training Set

The document describes premarket validation studies for a medical device that performs quantitative measurements, not a machine learning or AI model that requires a "training set." Therefore, there is no "training set" in the context of AI model development mentioned or implied in this document. The studies are validation studies demonstrating the device's performance against established analytical standards.

9. How Ground Truth for the Training Set Was Established

As noted above, this device does not involve a "training set" in the context of AI/ML. The measurements and comparisons are against established analytical methodologies and reference values.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine and healing. The caduceus is depicted as a staff with two snakes coiled around it, and a pair of wings at the top.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 14, 2016

INSTRUMENTATION LABORATORY CO. CAROL MARBLE REGULATORY AFFAIRS DIRECTOR 180 HARTWELL ROAD BEDFORD MA 01730

Re: K160412

Trade/Device Name: GEM Premier 5000 (Measured Parameters: pH. pCO2, pO2) Regulation Number: 21 CFR 862. 1120 Regulation Name: Blood gases (pCO2 and pO2) and blood pH test system Regulatory Class: II Product Code: CHL Dated: December 7, 2016 Received: December 8, 2016

Dear Ms. Marble:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160412

Device Name

GEM Premier 5000 (Measured Parameters: pH, pCO2, pO2)

Indications for Use (Describe)

The GEM Premier 5000 is a portable critical care system for use by health care professionals to rapidly analyze heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH and pO2 from venous, arterial and capillary heparinized whole blood, as well as quantitative measurements of pCO2 from venous and arterial heparinized whole blood. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status.

pH, pCO2 and pO2 measurements in whole blood are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Instrumentation Laboratory, a Werfen Company. The logo consists of two overlapping shapes in green and blue, followed by the company name in gray. Below the company name, it states "A Werfen Company" in a smaller font, also in gray.

510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

Submitter's InformationInstrumentation Laboratory (IL) Co.180 Hartwell RoadBedford, MA 01730, USA
-------------------------------------------------------------------------------------------------------------
Contact PersonCarol Marble, Regulatory Affairs DirectorPhone: 781-861-4467Fax: 781-861-4207Email: cmarble@ilww.com
----------------------------------------------------------------------------------------------------------------------------------
Preparation DateDecember 7, 2016
------------------------------------
Device Trade NameGEM Premier 5000(Measured Parameters: pH, pCO2 and pO2)
--------------------------------------------------------------------------------
Predicate DeviceGEM Premier 4000K133407
---------------------------------------------
Regulatory Information
AnalyteRegulation SectionRegulatory DescriptionClassProduct CodePanel
pH, pCO₂, pO₂,862.1120Blood Gases (pCO₂, pO₂) and Blood pH systemIICHL75

{4}------------------------------------------------

Device Description

The GEM Premier 5000 system provides health care professionals in central laboratory or point-of-care clinical settings with fast, accurate, quantitative measurements of pH and pQ2 from venous, arterial and capillary heparinized whole blood, as well as quantitative measurements of pCO₂ from venous and arterial heparinized whole blood.

Key ComponentsDescription
AnalyzerEmploys a unique color touch screen and a simple set of menus andbuttons for user interaction. The analyzer guides operators through thesampling process with simple, clear messages and prompts.
GEM Premier 5000 PAK(disposable, multi-use GEM PAK)Houses all required components necessary to operate the instrumentonce the cartridge is validated. These components include the sensors,CO-Ox optical cell, Process Control (PC) Solutions, sampler, pumptubing, distribution valve and waste bag. The GEM PAK has flexiblemenus and test volume options to assist facilities in maximizingefficiency.NOTE: The EEPROM on the GEM PAK includes all solution values andcontrols the analyte menu and number of tests.
Step 1:After inserting the GEM PAK, the instrument will perform anautomated PAK warm-up during which the sensors arehydrated and a variety of checks occur, all of which takeabout 40 minutes. During warm-up, the instrument requiresno user intervention.
Step 2:After GEM PAK warmup, Auto PAK Validation (APV) process isautomatically completed: two completely independentsolutions (PC Solution D and E) that are traceable to NISTstandards, CLSI procedures or internal standards, containingtwo levels of concentration for each analyte, are run by theanalyzer to validate the integrity of the PC Solutions and theoverall performance of the analytical system.
Step 3:After successful performance of APV, iQM2 manages thequality control process, replacing external quality controls.
Intelligent Quality Management 2(iQM2)iQM2 is an active quality process control program designed to providecontinuous monitoring of the analytical process before, during andafter sample measurement with real-time, automatic error detection,automatic correction of the system and automatic documentation of allcorrective actions.iQM2 is a statistical process control system that performs 5 types ofcontinuous, quality checks to monitor the performance of the GEMPAK, sensors, CO-Ox, and reagents. These checks include System,Sensor, IntraSpect, Pattern Recognition and Stability Checks.

{5}------------------------------------------------

Indications for Use / Intended Use
GEM Premier 5000The GEM Premier 5000 is a portable critical care system for use byhealth care professionals to rapidly analyze heparinized wholeblood samples at the point of health care delivery in a clinicalsetting and in a central laboratory. The instrument providesquantitative measurements of pH and pO2 from venous, arterialand capillary heparinized whole blood, as well as quantitativemeasurements of pCO2 from venous and arterial heparinized wholeblood. These parameters, along with derived parameters, aid in thediagnosis of a patient's acid/base status.pH, pCO2, pO2 measurements in whole blood are used in thediagnosis and treatment of life-threatening acid-base disturbances.
Substantial Equivalency
The GEM Premier 5000 system is substantially equivalent in function and intended use to the following predicate device for pH, pCO2, pO2:
ItemPredicate DeviceNew Device
Trade NameGEM Premier 4000GEM Premier 5000
510(k) No.K133407K160412
ManufacturerInstrumentation Laboratory Co.Instrumentation Laboratory Co.
Indicationsfor UseThe GEM Premier 4000 is a portable critical care system for use byhealth care professionals to rapidly analyze whole blood samples atthe point of health care delivery in a clinical setting and in a centrallaboratory. The instrument provides quantitative measurements ofpH, pCO2, pO2, sodium, potassium, chloride, ionized calcium,glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb,O₂Hb, COHb, MetHb, HHb) parameters. Total bilirubin can also bequantitated from heparinized plasma samples when analyzed in thetBili/CO-Ox mode. These parameters, along with derivedparameters, aid in the diagnosis of a patient's acid/base status,electrolyte and metabolite balance and oxygen delivery capacity.Total bilirubin measurements are used in the diagnosis andmanagement of biliary tract obstructions, liver disease and varioushemolytic diseases and disorders involving the metabolism ofbilirubin. In neonates, the level of total bilirubin is used to aid inassessing the risk of kernicterus.The GEM Premier 5000 is a portable critical care system for useby health care professionals to rapidly analyze heparinizedwhole blood samples at the point of health care delivery in aclinical setting and in a central laboratory. The instrumentprovides quantitative measurements of pH and pO2 fromvenous, arterial and capillary heparinized whole blood, as wellas quantitative measurements of pCO2 from venous and arterialheparinized whole blood. These parameters, along with derivedparameters, aid in the diagnosis of a patient's acid/base status.pH, pCO2, pO2 measurements in whole blood are used in thediagnosis and treatment of life-threatening acid-basedisturbances.
Substantial Equivalency (Cont.)
ItemPredicate DeviceNew Device
Trade NamesGEM Premier 4000K133407GEM Premier 5000K160412
Intended UserCentral Laboratory and Point-of-CareSame
Sample TypeHeparinized whole blood$pH, pO2$ Arterial, venous or capillary heparinized whole blood$pCO2$ Arterial or venous heparinized whole blood
Blood Gas MeasurementPotentiometry: $pH$ and $pCO2$Amperometry: $pO2$Same
Sample IntroductionAspirationSame
PAK Shelf-Life StabilityUp to 180 daysSame
PAK Storage Temperature15-25°CSame
System Operating Temperature12-32°CSame
Operating System SoftwareLinux-basedSame
Calibration2-point calibrationSame
QC MaterialCVP 1 and 2PC Solution D and E (PAK Internal)
CVP 3 and 4PC Solution D and E (PAK Internal)
CVP 5 tBiliSame; No Formulation Change

Special Conditions for Use Statement

  • . For prescription use only.
  • For clinical laboratory and point-of-care use. .

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Substantial Equivalency (Cont.)
ItemPredicate DeviceNew Device
Trade NameGEM Premier 4000K133407GEM Premier 5000K160412
Instrument DimensionsGEM Premier 4000 Instrument:• Height: 18 inches• Width: 12 inches• Depth: 15 inches• Weight: 44 poundsGEM Premier 5000 Instrument:• Height: 18.6 inches• Width: 13.0 inches• Depth: 16.4 inches• Weight: 45.4 pounds
Cartridge (PAK) DimensionsGEM Premier 4000 Cartridge (PAK):• Height: 6.75 inches• Width: 10 inches• Depth: 8 inches• Weight: 8 poundsGEM Premier 5000 Cartridge (PAK):• Height: 6.75 inches• Width: 10 inches• Depth: 8 inches• Weight: 8.1 pounds
Reportable RangeAnalyteGEM Premier 4000GEM Premier 5000
pH7.00 to 8.007.00 to 7.92
pCO26 to 125 mmHg6 to 125 mmHg
pO25 to 690 mmHg6 to 690 mmHg

{9}------------------------------------------------

Performance Summary

Internal Precision Study - Aqueous Controls

In accordance with CLSI EP05-A3, an internal 20-day precision study was performed on the GEM Premier 5000, with GEM System Evaluator. Each of the control levels was run on three (3) GEM Premier 5000 analyzers for twenty (20) days, with two (2) runs per day and one (1) replicate measured per run per level (n=120). All results were within specification.

MaterialAnalyteLevelMeanNWithinAnalyzerSDWithinAnalyzer%CVTotal SDTotal%CV
GEM SystemEvaluatorpHLevel 17.141200.0080.1%0.0080.1%
Level 27.381200.0040.1%0.0060.1%
Level 37.571200.0030.0%0.0030.0%
pCO2(mmHg)Level 1871202.32.7%2.32.7%
Level 2351200.71.9%0.82.3%
Level 3141200.32.2%0.32.3%
pO2(mmHg)Level 1311201.96.1%2.47.9%
Level 2881201.21.4%2.22.4%
Level 33701204.81.3%5.91.6%

Internal Precision Study – GEM PAK (Cartridge) Process Control Solutions D and E

In accordance with CLSI EP05-A3, an internal 20-day precision study was performed with the GEM PAK (cartridge) Process Control Solutions (PCS) D and E run on three (3) GEM Premier 5000 analyzers for twenty (20) days, with two (2) runs per day and one (1) replicate measured per run per level (N=120 per analyte/per level). All results were within specification.

MaterialAnalyteMeanNWithin AnalyzerSDWithin Analyzer%CV
PCS DpH7.351200.0010.0%
PCS EpH7.211200.0010.0%
PCS D$pCO_2$ (mmHg)251200.10.4%
PCS E$pCO_2$ (mmHg)681200.30.5%
PCS D$pO_2$ (mmHg)551201.11.9%
PCS E$pO_2$ (mmHg)981201.11.1%

{10}------------------------------------------------

Internal Precision Study – Whole Blood

In accordance with CLSI EP05-A3, an internal precision study was performed using five (5) different concentrations of whole blood per analyte, each run on three (3) GEM Premier 5000 analyzers per sample mode for five (5) days, with one (1) run per day and eight (8) replicates measured per run per level (N=120 per analyte/per sample mode). All results were within specification.

Sample Modes and Volumes:

  • . Normal Mode 150 µL
  • . Micro Mode 65 µL
AnalyteModeLevelMeanNWithin RunSDWithin Run%CVTotal SDTotal %CV
pHNormalMode17.111200.0040.1%0.0050.1%
27.331200.0070.1%0.0070.1%
37.351200.0040.1%0.0050.1%
47.421200.0050.1%0.0060.1%
57.681200.0120.1%0.0140.2%
MicroMode17.101200.0060.1%0.0070.1%
27.321200.0040.1%0.0060.1%
37.351200.0040.1%0.0050.1%
47.411200.0050.1%0.0060.1%
57.671200.0130.2%0.0150.2%
pCO2(mmHg)NormalMode11121202.72.4%2.92.3%
2701201.11.5%1.52.0%
3501200.61.2%0.81.5%
4361200.51.3%0.81.9%
5101200.54.6%0.87.7%

{11}------------------------------------------------

AnalyteModeLevelMeanNWithin Run SDWithin Run %CVTotal SDTotal %CV
pO2(mmHg)NormalMode1321200.41.2%0.92.7%
2621200.71.1%0.91.2%
32041202.31.1%3.11.4%
44151208.62.1%13.02.8%
572212018.62.6%32.74.3%
MicroMode1311200.93.0%1.65.3%
2621200.71.1%0.91.4%
32041204.32.1%5.52.5%
440212015.33.8%16.43.8%
569312026.53.8%34.04.8%

Internal Precision Study – Whole Blood (Cont.)

{12}------------------------------------------------

Reproducibility Study with Aqueous Controls – Point-of-Care (POC) Setting

In accordance with CLSI EP05-A3, a reproducibility study was performed at three (3) external clinical point-of-care (PCC) sites. The studies were run by a total of nine (9) different operators on three (3) different GEM Premier 5000 instruments, using a single lot of GEM Premier 5000 PAKs (cartridges). Each site used the same lots of GEM System Evaluator (GSE), running each control level in triplicate, twice a day for 5 days, for a total of 30 poled). All results at all sites were within specification.

Pooled Multi-Site POC Data
AnalyteMaterial/LevelNInsertRangeTargetSD/CVSpecMeanRepeatabilityBetween-RunBetween-DayBetween-SiteReproducibility
SD%CVSD%CVSD%CVSD%CVSD%CV
pHGSE 1907.10-7.187.140.027.140.0020.0%0.0010.0%0.0000.0%0.0000.0%0.0020.0%
GSE 2907.35-7.437.390.027.390.0070.1%0.0030.0%0.0030.0%0.0020.0%0.0080.1%
GSE 3907.53-7.617.570.027.570.0020.0%0.0000.0%0.0030.0%0.0000.0%0.0030.0%
pCO2(mmHg)GSE 19078-96874%870.40.5%0.70.9%0.60.7%0.00.0%1.11.2%
GSE 29029-39342.5340.92.5%0.00.0%0.30.9%0.00.0%0.92.7%
GSE 39010-16132.5130.43.3%0.00.0%0.21.2%0.00.0%0.53.5%
pO2(mmHg)GSE 19021-41315280.72.3%0.41.4%0.62.2%1.65.8%1.96.7%
GSE 29077-97875871.41.7%0.00.0%0.60.7%1.01.2%1.92.2%
GSE 390325-3893575%3577.92.2%0.00.0%0.00.0%4.81.3%9.32.6%

{13}------------------------------------------------

External Precision – Whole Blood

To evaluate whole blood precision on the GEM Premier 5000 system in the central laboratory and point-ofcare (POC) settings, whole blood patient samples were tested at 2 external central laboratories and 1 internal Customer Simulation Laboratory (CSL), as well as at 3 external POC locations. For the central laboratory setting, the studies were performed by 3 operators on 3 GEM Premier 5000 instruments using a single lot of GEM Premier 5000 PAK (cartridge). For the POC setting, the studies were performed by 11 operators on 3 GEM Premier 5000 instruments, using a single lot of GEM Premier 5000 PAK (cartridge). At least two whole blood specimens were analyzed in triplicate daily for 5 days in both normal mode (150 µL) for pH, pCO2 and pO2, and micro capillary (65 µL) mode for pH and pO2. At the internal Customer Simulation Laboratory (CSL), contrived whole blood specimens were analyzed in addition to native specimens in order to cover the low and high medical decision levels of each analyte.

The precision results are summarized below:

AnalyteModeSiteNMeanWithin SampleSD
pHNormalModePOC1547.360.008
POC2427.330.009
POC3307.350.008
POC-All1267.350.008
CSL307.320.009
Lab1307.360.009
Lab2307.320.007
Lab-All907.330.008
MicroModePOC1307.290.009
POC2367.320.012
POC3367.280.012
POC-All1027.300.011
CSL307.300.008
Lab1307.330.007
Lab2307.260.009
Lab-All907.300.008
AnalyteModeSiteNMeanWithin SampleSD
pCO2(mmHg)NormalModePOC148431.1
POC239420.8
POC330491.7
POC-All117441.2
NormalModeCSL24531.3
Lab130461.3
Lab224460.8
Lab-All78481.2
pO2(mmHg)NormalModePOC127521.5
POC212630.4
POC312712.0
POC-All51591.5
CSL30520.6
MicroModeLab121550.8
Lab215503.9
Lab-All66530.8
POC121531.5
MicroModePOC26451.2
POC318581.1
POC-All45541.3
CSL30540.8
MicroModeLab124601.0
Lab218521.1
Lab-All72561.0

{14}------------------------------------------------

External Precision – Whole Blood (Cont.)

{15}------------------------------------------------

LoB, LoD and LoQ

In accordance with CLSI EP17-A2, LoB, LoD and LoQ were established for pH, pCO2 and pO2 using three (3) lots of GEM Premier 5000 PAKs (cartridges).

AnalyteLoBLoDLoQ
pH*8.908.878.04
pCO2 (mmHg)014
pO2 (mmHg)NANA2

Following are the combined data results for LoB, LoD and LoQ:

  • *Note: By definition, LoQ must be ≥ LoD except for pH. Since pH is measured on a negative log scale, larger pH values represent lower analyte levels. Thus, for pH, LoQ must be ≤ LoD.

Linearity

In accordance with CLSI EP06-A, eight (8) or nine (9) levels per analyte were prepared by tonometering whole blood to challenge the claimed reportable range for each parameter. Each blood level was analyzed in triplicate on three (3) GEM Premier 5000 test analyzers and results compared to standard reference procedures (i.e. tonometry for gases).

Combined data from limit of quantitation (LOQ) and linearity were used to support the lower limits of the claimed reportable ranges.

Analyte# ofLevelsN perLevelSlopeInterceptR2TestedRangeReportableRange
pH890.9720.1910.9986.67 to 7.977.00 to 7.92
pCO2(mmHg)891.045-2.0270.9981 to 1496 to 125
pO2(mmHg)991.028-4.0690.9955 to 7276 to 690

{16}------------------------------------------------

Analytical Specificity

In accordance with EP07-A2, an interference study was conducted on the GEM Premier 5000. A screening test was used to identify substances that produce a clinically significant interference based on measurement at two different analyte concentrations. The table below lists substances that were screen tested with no observed interference on pH, pCO₂ and pO₂ results:

No Observed Interference with pH, pCO2 and pO2
SubstanceConcentration
Acetaminophen1324 µmol/L
Albumin (Human)60 g/L
Amoxicillin206 µmol/L
Aprotinin50 mg/L
Atracurium50 mg/L
Benzalkonium (Chloride)5 mg/L
Bilirubin20 mg/dL
Ceftriaxone1460 µmol/L
Ciprofloxin30.2 µmol/L
Diazepam18 µmol/L
Epinephrine0.5 µmol/L
Ethanol86.8 mmol/L
Etomidate50 mg/L
Fentanyl0.02 µg/ml
Gentamycin21 µmol/L
Halothane759 µmol/L
Hematocrit25%75%
Hemoglobin (Hemolysis)2 g/dL (20%)
Lithium (Chloride)3.2 mmol/L
Methadone6.46 µmol/L
Midazolam0.5 µg/mL
Morphine1.75 µmol/L
Omeprazole17.4 µmol/L
Phenobarbital431 µmol/L
Propofol0.05 mg/mL
Suxamethonium68 µmol/L
Thiopental248 µmol/L
Thyroxine1.29 µmol/L
Triglycerides (Intralipids)2% or 4012 mg/dL

{17}------------------------------------------------

Internal Method Comparison

In accordance with EP09-A3, an internal method comparison study was conducted using clinical samples to compare the GEM Premier 5000 to the following predicate devices:

  • Tonometry: pO2, pCO2 (tonometry used due to instability of blood gases)
  • . GEM Premier 4000: pH

Samples were altered as needed to cover the medical decision levels. All parameter levels passed specification for all sample modes.

AnalyteNSlopeInterceptMedical DecisionLevelBias at MedicalDecision Level
pH3730.9530.3440.9937.300.001
7.35-0.001
7.45-0.006
pCO2(mmHg)1501.026-0.9910.99735-0.1
500.3
701.2%
pO2(mmHg)1481.027-1.2660.99930-0.5
450.0
600.4

{18}------------------------------------------------

Whole Blood Performance at Medical Decision Levels

The data from the internal method comparison and precision studies were combined to assess the performance at medical decision levels.

Total Error was computed based on the following equation and the results were compared to the GEM Premier 5000 Total Error Specifications:

Total Error Observed = Bias + 2 * SD (or %CV)

Note: Previously shown bias and precision data were used in Total Error computations below.

AnalyteMedicalDecisionLevelAbsolute Value ofBias at MedicalDecision Level2*(SD or %CV)Total Error ObservedBias + 2*(SD or %CV)
pH7.300.0010.0130.014
7.350.0010.0080.009
7.450.0060.0100.016
pCO2(mmHg)350.10.91.0
500.31.21.5
701.2%3.0%4.2%
pO2(mmHg)300.50.71.2
450.01.41.4
600.41.41.8

{19}------------------------------------------------

Reference Ranges

AnalyteReference RangeUnit
pH7.35 to 7.45pH
cH35.5 to 44.7nmol/L
cH35.5 to 44.7nEq/L
pH7.32 to 7.43 (venous)pH
pCO₂Arterial blood: 35 to 48 (male) and 32 to 35 (female)Arterial blood: 4.6 to 6.4 (male) and 4.3 to 6.0 (female)Venous blood (right atrium) - 6-7 mmHg (0.80-0.93 kPa) higher than arterial pCO₂mmHgkPa
pO₂83 to 10811.0 to 14.4mmHgkPa
TCO₂19.0 to 24.022.0 to 26.0 (venous)mmol/L
BE-2.0 to 3.0mmol/L
HCO₃⁻21 to 28mmol/L
21 to 28mEq/L
22 to 29 (venous)mmol/L
22 to 29 (venous)mEq/L

Sources: Burtis, Carl and David Bruns, Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, Elsevier Saunders, 7th edition, 2015, pp 952-982

Wu, A., Tietz Clinical Guide to Laboratory Tests, W.B. Saunders Co., St. Louis MO, 4th Edition, 2006: 951-982

TCO₂, HCO₃ and BE (Base Excess) are derived parameters. Note:

{20}------------------------------------------------

Clinical Testing

In accordance with EPO9-A3, a method comparison study was conducted on the GEM Premier 5000 compared to the predicate device, the GEM Premier 4000, in the point-of-care (POC) setting using heparinized whole blood patient samples from the intended use population.

  • Study Design:
    • Three (3) external point-of-care (POC) sites.
    • One (1) internal Customer Simulation Laboratory (CSL) at IL, where three (3) intended POC users were brought on site to run the samples, allowing spiking to cover the reportable ranges.

The pooled results from the POC sites and the IL internal Customer Simulation Laboratory (CSL) for the Normal Mode (with samples collected in syringes) are presented below.

Pooled Point-of-Care Site and CSL Data - Normal Mode (with Syringe Samples)
AnalyteNSlopeInterceptrSample Range
pH4790.9410.4270.9917.01 to 7.92
pCO2 (mmHg)4920.9553.5450.99111 to 117
pO2 (mmHg)5060.9925.0930.9966 to 685

To support capillary claims, finger-stick samples were collected at an external POC site (N=65 native samples) and the IL internal Customer Simulation Laboratory (CSL) (N=106 native samples) with POC operators. The observed total error at the medical decision levels is shown below:

Pooled Point-of-Care Site and CSL Data with Native Capillary Samples Only
AnalyteNRangeMinRangeMaxMDLBias at MDL95% CI of Bias at MDLTEa
pH1717.367.597.300.002-0.015 to 0.020$\pm$ 0.04
7.350.005-0.006 to 0.016$\pm$ 0.04
7.450.0100.005 to 0.014$\pm$ 0.04
pO2(mmHg)16752105306.11.3 to 11.6$\pm$ 9.0
455.12.0 to 8.6$\pm$ 9.0
604.12.0 to 5.7$\pm$ 9.0

{21}------------------------------------------------

Clinical Testing

In addition, the data from the native capillary samples (finger-stick samples) previously presented were pooled with contrived capillary samples prepared internally. The regression analysis is shown below:

Pooled Point-of-Care Site and CSL Data with Additional Contrived Capillary Results
AnalyteNSlopeInterceptrSample Range
pH1890.9350.4940.9757.07 to 7.89
pO2 (mmHg)2181.0082.5450.9966 to 676
Conclusion
The technological and functional characteristics of the new GEM Premier5000 as described above are substantially equivalent to that of thepredicate device (GEM Premier 4000) for pH, pCO2, pO2.The analytical and clinical study results demonstrate that the GEMPremier 5000 is safe and effective for its intended purpose and equivalentin performance to the predicate device.

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.