K Number
K160278
Device Name
VSTAAR AdjusteR
Date Cleared
2016-11-04

(276 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VSTAAR is a mechanical device intended to be used by chiropractors, medical doctors and other licensed health care professions for the external analysis and adjustment of the human spinal column, and for soft tissue musculoskeletal mobilization.

Device Description

The VSTAAR AdjusteR is a mechanical device used by chiropractors, medical and other licensed health care professionals for the external analysis and adjustment of the human spinal column, and for soft tissue and musculoskeletal mobilization. The graphical user interface is mouse driven. All of the screens are easy to find and easy to understand. Navigating from screen to screen is quick and intuitive.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the VSTAAR AdjusteR device, seeking substantial equivalence to existing predicate devices. The submission focuses on non-clinical performance data and electrical safety testing.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" for the VSTAAR AdjusteR in terms of performance metrics like sensitivity, specificity, or accuracy, which are common for diagnostic or AI-driven devices. Instead, it relies on demonstrating substantial equivalence to predicate devices through comparisons of technological characteristics and non-clinical testing.

The performance reported is primarily confirmation of meeting design specifications and intended uses, with electrical safety and electromagnetic compatibility being verified against international standards.

Acceptance Criterion (Implicit)Reported Device Performance
Conformance to device specificationsThe performance data records documents that the VSTAAR AdjusteR device met its stated requirements and design specifications as intended.
Conformance to intended usesValidation testing... to confirm the device met its intended uses and requirements.
Electrical Safety (IEC 60601-1)Tested in conformance with IEC 60601-1 / EN 60601-1 Medical electrical equipment - Part 1: General requirements for safety (IEC 60601-1:1988 + A1:1991 + A2:1995), by Intertek, Twinsburg, OH.
Electromagnetic Compatibility (IEC 60601-1-2)Tested in conformance with IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests, by a Keystone Compliance, New Castle, PA.
Similar Technological Characteristics to Predicates (Implicit)As demonstrated by the "PRODUCT COMPARISON TABLE" on page 8, the VSTAAR AdjusteR has similar IFU, power source, force settings, applicator tip velocity, pulse rate, range of displacement, duration of force, max count of impulses to predicate devices, with differences in method of delivering thrust force, patient-contacting material, and sensor within handheld device noted to not adversely affect safety and effectiveness.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "significant verification and validation testing" and "performance data records". However, it does not specify any sample sizes for a test set, nor does it refer to patient data or its provenance (country of origin, retrospective/prospective). The testing described is non-clinical, focusing on device performance specifications and safety standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and not provided in the document. The testing described is non-clinical laboratory testing, not human subject or diagnostic performance studies that would require expert-established ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and not provided. As above, the testing is non-clinical and does not involve subjective assessment requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. The device, VSTAAR AdjusteR, is a mechanical adjusting instrument, not an AI or imaging device that would typically leverage such studies to evaluate human reader improvement with AI assistance. The document is entirely focused on the mechanical and electrical safety/performance of the device itself, not its impact on human diagnostic performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is not applicable to the VSTAAR AdjusteR as described. The device is a mechanical instrument used by healthcare professionals. It is not an algorithm, nor an AI system capable of "standalone" performance in the sense of a diagnostic or assistive software.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical performance and safety testing, the "ground truth" would be the established engineering specifications, test procedures, and international standards (e.g., IEC 60601-1, IEC 60601-1-2) that the device was tested against. The document does not describe any studies involving expert consensus, pathology, or outcomes data, as these are typically associated with clinical efficacy or diagnostic accuracy studies, which are not detailed here.

8. The sample size for the training set

This information is not applicable and not provided. The VSTAAR AdjusteR is a mechanical device, not a machine learning or AI model, thus it does not have a "training set" in that context.

9. How the ground truth for the training set was established

This information is not applicable and not provided for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 4, 2016

Spinal Acoustics, LLC Albert A. Torrence President and CEO Spinal Acoustics, LLC 640 Fourth Street Beaver, Pennsylvania 15009

Re: K160278 Trade/Device Name: VSTAAR AdjusteR Regulatory Class: Unclassified Product Code: LXM Dated: October 1, 2016 Received: October 4, 2016

Dear Albert Torrence:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

MichaelJ. Hoffmann -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160278

Device Name VSTAAR AdjusteR

Indications for Use (Describe)

The VSTAAR is a mechanical device intended to be used by chiropractors, medical doctors and other licensed health care professions for the external analysis and adjustment of the human spinal column, and for soft tissue musculoskeletal mobilization.

Type of Use (Select one or both, as applicable)

☑ Exportation (Part 21 CFR 600.81, Subpart E)☐ Own Use Consignment (21 CFR 600.80, Subpart D)
--------------------------------------------------------------------------------------------------------------------------------------------------

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

510(k) Number:K160278
Company Name:Company Address:Spinal Acoustics LLCFourth Street Professional Building640 Fourth StreetBeaver, PA 15009
Contact:Phone/FAX Numbers:Albert Torrence, President and CEO(724) 513-0354 (mobile)(724) 775-5200
Email Address:torrencelaw@comcast.net
Trade Name:VSTAAR AdjusteR
Common/Classification Name:Chiropractic Adjusting Instrument/Manipulator, Plunger-like Joint
Classification Regulation:Unclassified
Device Class:Class II
Product Code (Procode):LXM
Preparation Date:June 29, 2012

LEGALLY MARKETED PREDICATE DEVICE - as required by A. 807.92(a)(3)

One of the identified legally marketed predicate devices identified by the submitter is Smart Adjuster, K962239, submitted by Sigma Company [now known as Sigma Instruments]. A second legally marketed predicate device is PulStarFRAS, K973914, submitted by Sense Technology. A third predicate device is the legally marketed Impulse IQ Adjusting Instrument, K080261, submitted by Neuromechanical Innovations, LLC.

SUMMARY DEVICE DESCRIPTION - as required by 807.92(a)(4) B.

The VSTAAR AdjusteR is a mechanical device used by chiropractors, medical and other licensed health care professionals for the external

{4}------------------------------------------------

analysis and adjustment of the human spinal column, and for soft tissue and musculoskeletal mobilization.

The graphical user interface is mouse driven. All of the screens are easy to find and easy to understand. Navigating from screen to screen is quick and intuitive.

The main features of the device include:

Patient Data Screen - for patient demographic entry.

Single Evaluation test mode - to determine the articular and spinal tissue's relative compliance with a single impact on controlled preload instructions.

Lateral Evaluation - allows the user to evaluate the motion from side to side of the vertebrae, measuring the relative differences between two directions, identifying subluxation in the lateral axis.

Treatment - information from the Evaluation test mode allows the user to set a forced between 5 and 60 lbs and an impulse rate between 1 and 25 cycles per second.

Preload Bar Graph - allows the user to set a range that will be used to keep the device at a specific preload during all of the patients' treatments and assessments, typically based on body type for each patient - user feedback is provided graphically showing the pliability increase during the treating segment.

Real time Feedback - assists the user's treatment decisions. The LVDT position sensor measures depressions in the tissue made by the percussive force. The user can observe when there is a greater excursion of the LVDT denoting an increase in pliability or motion of the segment and tissue, documenting a change in the tissue consistency.

Treatment Focus – the user can focus on one region of concern at a time – cervical, thoracic or lumbar/sacral – allowing for a comparison between more like segments for bone structure.

Test "Pulse" - on seqments of the spine for comparison of relative compliance of those segments. Any vertebrae that will be considered a potential problem may or may not be selected by the practitioner for an adjustment.

{5}------------------------------------------------

Impulse Analysis - measuring depth of penetration and acceleration of penetration from an impact of a consistent force and the force has been shown to be essentially constant for specific substrates in both the evaluation of our instrument. The analysis assesses a region determining areas of low compliance (Minimal Pliability) and high compliance (Relatively High pliability compared to areas of Minimal Pliability). Once areas greater than 10% from one segment to the next are identified, the instrument is then used in a percussive mode (treatment) to break up muscle tension and loosen soft or articular tissue increasing mobility of the treated segments or regions. Bv repeated percussion affected areas are loosened, increasing the pliability of a muscle or segmental level thus increasing an area of low compliance or minimal pliability to a relatively higher level of compliance after mechanically palpating the tissue. This decreases muscle tension through mechanical compression and by simply mobilizing the affected areas allowing for more function and ROM in a segment or muscle.

PRODUCT AND TECHNICAL SPECIFICATIONS

Control Unit:

Power120VAC/60Hz
Single-phase2A
DimensionsW 16.75"x D 16.75"x H 6.7"
Weight15 lbs
CabinetBrushed Aluminum & Steel

Accessories:

DisplayPL 1900 LCD monitor or AOC e1649FwuUSB LCD monitor
ApplicatorSingle Tip/Dual Tip Applicators

Adiustment Device:

DimensionDiameter 2.5″ x Length 8.75′
Weight1.88 lbs
CoverPolished Aluminum
Cord8 ft

Operating Conditions:

Temperature+18°C - +30°C
Humidity30% - 75%
Atmospheric Pressure500 - 1060 hPa

Shipping and Storage Conditions:

{6}------------------------------------------------

Temperature................................................................................................................................................................... Humidity ..................................................................................................................................................................... Atmospheric Pressure .......... 500 - 1060 hPa

Operating System Software

Microsoft Windows XP and Windows 7

Application Software - Designed with National Instruments' LabView development software

Provides preloaded settings user selects of 1 – 60 settings
Feedbackpreload feedback provided
Force settinguser definable default
Pulsesmultiple pulses provided
Pulse rateuser selects from 1 to 25 Hz
User interfacetouch screen or mouse driven

C. INTENDED USE - as required by 807.92(a)(5)

The VSTAAR AdjusteR device is a tool designed by Chiropractors (DC), medical doctors (M.D. and D.O.), Physical Therapists (PT), Physicians Assistant (PA), Nurse Practitioner (NP), engineers and researchers to analyze and administer soft tissue, articular, spinal and extremity manipulations, including adjustment, and mobilization.

The graphical user interface is mouse driven. All of the screens are easy to find and easy to understand. Navigating from screen to screen is quick and intuitive.

INDICATIONS FOR USE - as required by 807.92(a)(5) D.

The VSTAAR is a mechanical device used by chiropractors, medical and other licensed health care professionals for the external analysis and adjustment of the human spinal column, and for soft tissue and musculoskeletal mobilization.

TECHNOLOGICAL CHARACTERISTICS SUMMARY - as required E. by 807.92(a)(6)

Like the predicate devices, the VSTAAR AdjusteR device operates on 120v 50/60 Hz power and is designed for a typical professional medical clinic environment. Within that environment, the VSTAAR device, like its predicates, is easily transported.

{7}------------------------------------------------

Like the predicate devices, the VSTAAR AdjusteR operates within a computer hardware/computer operating system/submitter developed application software environment.

The VSTAAR AdjusteR device, like its predicates, delivers a light mechanical impulse through the hand held treatment device with user selected force, pulse rate The VSTAAR AdjusteR device, like its predicates, has more than one tip for the adjusting or hand held treatment device. Like the predicates, the VSTAAR AdjusteR device has a sensor within the adjusting or hand held treatment device. Like the predicates, the VSTAAR AdjusteR device provides the professional user with an electromagnetic coil, user definable default settings and pulse rate settings.

Like the predicate devices, the VSTAAR AdjusteR device provides the user with a software interface for patient information input, operational setting selections, and storage of system generated data. Like the predicates, there is a choice of preload settings for the VSTAAR AdjusteR device. The VSTAAR AdjusteR device, like the predicates, provides multiple pulses and pulse rates.

Unlike some of the predicate devices, the VSTAAR AdjusteR device provides "preload feedback" to show the professional user if/how much the preload changes during treatment.

The submitted VSTAAR AdjusteR device has the same technological characteristics as the predicate devices. The Substantial Equivalence Table provides a detailed COMPARISON MATRIX of the submitted device and the identified predicate devices.

{8}------------------------------------------------

PRODUCT COMPARISON TABLE

510(k) #Subject Device (K122038Predicate Device (K962239)Predicate Device (K973914)Predicate Device (K080261)Justification
Device NameVSTAAR AdjusteRSmart AdjusterPulStarFrasImpulse IQ Adjusting InstrumentN/A
IFUThe VSTAAR AdjusteR is a mechanical device intended to be used by chiropractors, medical doctors and other licensed health care professions for the external analysis and adjustment of the human spinal column, and for soft tissue musculoskeletal mobilization.The smart adjuster is intended to be used to do both analysis and adjustments to the human spinal column in accordance with standard chiropractic procedures. These procedures include the location and return to normal compliance of joint fixations and adhesions.Indications for use of the pulstarfras include: musculoskeletal pain due to joint subluxation, restricted joint mobility, myofacial spasm, ligamentous strain.The impluse IQ adjusting instrument is intended for chiropractic adjustment, mobilization, or manipulation of the musculoskeletal joints of the spine and/or extremities or for soft-tissue musculoskeletal mobilization by a licensed health care professional only. For external use onlySimilar
Power sourceAC PoweredAC PoweredAC PoweredAC PoweredIdentical
Method of delivering thrust forceSpringSolenoidSolenoidSolenoidDifferent but does not adversely affect the safety and effectiveness of the subject device
Force settings5 - 60lbs10 - 35lbs10 - 35lbs22.5 - 90lbsSimilar
Applicatortip velocity5.65 ft/sUnknown3.0 ft/s6.29 ft/sSimilar
Pulse Rate1 - 12 HzUnknown2 -12 Hz4 - 12 HzSimilar
Range ofdisplacement (inches)Max 0.210Max 0.260Max 0.12Max 0.145Similar
Duration ofForce2-4 msUnknownUnknown2-4 msSimilar
Patient-contactingmaterial60 DurometerSiliconeRubberTippedProbeUnknownRubber contactendDifferentbut doesnotadverselyaffect thesafety andeffectiveness of thesubjectdevice
SensorwithinhandhelddeviceLinearVariableDifferentialTransformer(LVDT)ForcesensorForcesensorAccelerometerDifferentbut doesnotadverselyaffect thesafety andeffectiveness of thesubjectdevice
Max countofimpulses60Unknown1 - 20012Similar

{9}------------------------------------------------

NON-CLINICAL PERFORMANCE DATA TESTING AND REVIEW -F. as required by 807.92(b)(1)

Non-Clinical Testing

The submitted VSTAAR AdjusteR device has undergone significant verification and validation testing. Validation testing included verification testing of major components of the device to document conformance to device specifications and validation testing to confirm the device met its intended uses and requirements.

{10}------------------------------------------------

The performance data records documents that the VSTAAR AdjusteR device met its stated requirements and design specifications as intended.

Electrical Safety Testing - Electromagnetic Compatibility & Electrical Safety

The submitted device has been tested in conformance with IEC 60601-1 / EN 60601-1 Medical electrical equipment - Part 1: General requirements for safety (IEC 60601-1:1988 + A1:1991 + A2:1995), by Intertek, Twinsburg, OH, and IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests, by a Keystone Compliance, New Castle, PA, both qualified testing laboratories.

SUMMARY STATEMENT

Spinal Acoustics, LLC believes the VSTAAR AdjusteR device's summary of non-clinical testing and electrical safety testing data further documents the submitter's claim of substantial equivalence.

CONCLUSION

The performance testing and testing data document that the submitted VSTAAR AdjusteR device is substantially equivalent to the Sigma Company's [now known as Sigma Instruments] predicate Smart Adjuster, K962239, Sense Technology Inc.'s predicate PulseStarFRAS, K973914, and Neuromechanical Innovations, LLC's predicate Impulse IQ Adjusting Instrument, K080281.

N/A