K Number
K160252
Date Cleared
2016-04-27

(86 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.

The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared.

Abbott FreeStyle test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.

Device Description

The proposed device provides for the management of insulin therapy and blood glucose monitoring by patients with diabetes mellitus. It is comprised of two primary components: the disposable insulin infusion pump (Pod) and an associated wireless remote controller referred to as the Personal Diabetes Manager (PDM) with an embedded blood glucose meter.

AI/ML Overview

This document describes the 510(k) premarket notification for the Insulet OmniPod Insulin Management System (K160252). The notification details modifications made to the device and provides evidence of performance testing to support its substantial equivalence to a predicate device (K122953).

Here's the breakdown of the acceptance criteria and supporting studies based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria / Performance AspectReported Device Performance
Risk ManagementPredetermined acceptance criteria were met, and the device is safe for use, as demonstrated by verification activities conducted in accordance with ISO 14971:2007.
Soft Cannula Depth and AngleAll Pod samples tested met the specified requirements for soft cannula insertion depth and angle.
Cam Finger DesignThe cam finger design has been verified to perform as defined by its requirements.
Insertion Mechanism ReleaseTesting confirmed that the Release Bar design intent was met and it is capable of releasing the insertion mechanism.
Cannula Insertion RetentionTesting confirmed that the Catch Beam design intent was met and it is capable of retaining the Soft Cannula at the specified depth.
Cannula Insertion IndicatorAn indicator has been provided and tested to confirm when the cannula is fully inserted.
Biocompatibility (Silicone Lubricant)An assessment rationalized the additional silicone lubricant for human use as acceptable. The added silicone is not in contact with the fluid path. (Note: Predicate device testing per ISO 10993: Part 3, Part 5, Part 6, Part 10 is referenced for the same silicone at a different location on the predicate).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the specific sample sizes for each test mentioned (e.g., number of Pods for cannula depth testing). It generally refers to "all Pod samples tested" or "testing was conducted."

The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated for these performance tests. The manufacturing site is listed as Flextronics Industrial (Shenzhen) Co., LTD. in Shenzhen, Guangdong, China. It is highly probable the performance testing was conducted at or in conjunction with the manufacturing site or the submitting company (Insulet Corporation, Billerica, MA, USA).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to the provided document. The performance data presented refers to engineering and bench testing of mechanical and material properties of the device, not clinical studies requiring expert interpretation of diagnostic output.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. As stated above, these are engineering and bench tests, not clinical studies requiring human adjudication of ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device described is an insulin management system with a built-in glucose meter, not an AI-based diagnostic imaging or interpretation device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical medical device (infusion pump and glucose meter), not a standalone algorithm. The performance tests are for the physical components and their functionality.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the performance tests described, the "ground truth" or reference standards are the device specifications and requirements established by the manufacturer and relevant industry standards (e.g., ISO 14971, ISO 10993). For example, for cannula depth, the ground truth is the specified depth. For biocompatibility, it's the acceptance criteria defined by ISO 10993.

8. The sample size for the training set

This information is not applicable. The device is a physical medical device, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a wing-like shape above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 27, 2016

Insulet Corporation Mr. Matthew King Director, Regulatory Affairs and Design Ouality Assurance 600 Technology Park Drive Suite 200 Billerica, Massachusetts 01821

Re: K160252

Trade/Device Name: Omnipod Insulin Management System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG, NBW Dated: January 30, 2016 Received: February 1, 2016

Dear Mr. King:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -

for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160252

Device Name OmniPod Insulin Management System

Indications for Use (Describe)

The OmniPod Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.

The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared.

Abbott FreeStyle test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Complying with 21 CFR 807.92

510(k)K160252
Date prepared:April 26, 2016
Submitter Name:Insulet Corporation
Submitter Address:600 Technology Park Drive, Suite 200Billerica, MA 01821
FDA Establishment Owner/OperatorNumber:9056196
FDA Establishment Registration Number:3004464228
Contact Person:Matthew KingDirector of Regulatory Affairs and Design Quality Assurance
Phone:(978) 600-7427 (office)(603)459-9755 (mobile)
Fax:(978) 600-0120
Manufacturing Site:Flextronics Industrial (Shenzhen) Co., LTD.Building 2-3, Yusheng Industrial Park, 467 Xixiang SectionNational Highway 107, Xixiang, Baoan DistrictShenzhen Guangdong, 518126, ChinaFDA Establishment Registration Number 3307710663
Device Trade/Proprietary Name:OmniPod® Insulin Management System
Device Common Name:Pump, Infusion, Insulin
Regulation Description:Infusion pump
Requlation Medical Specialty:General Hospital
Review Panel:General Hospital
Product Code:LZGNBW
Submission Type:Traditional 510(k) Corrective Actions Being Effected
Regulation Number:21 CFR 880.5725
Device Class:2
Model number (Pod/10-pack):ZXP425
Model number (PDM):14500-5A (UST400)
Device predicate:K122953 OmniPod® Insulin Management System

{4}------------------------------------------------

Insulet Corporation

Device Description:

The proposed device provides for the management of insulin therapy and blood glucose monitoring by patients with diabetes mellitus. It is comprised of two primary components: the disposable insulin infusion pump (Pod) and an associated wireless remote controller referred to as the Personal Diabetes Manager (PDM) with an embedded blood glucose meter.

Indications for use:

The following Indications for Use are identical to the predicate device:

The OmniPod® Insulin Management System is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin and for the quantitative measurement of glucose in fresh whole capillary blood (in vitro) from the finger.

The glucose measurements should not be used for the diagnosis of or screening for diabetes. The PDM glucose meter is intended for single-patient use and should not be shared.

Abbott FreeStyle® test strips are used with the built-in FreeStyle meter for the quantitative measurement of blood glucose in fresh whole capillary blood from the finger, upper arm and palm. Abbott Freestyle Control Solutions are used to verify that the meter and test strips are working together properly and that the test is performed correctly.

Technological Characteristics:

The technological characteristics, design, material composition, energy source, principal of operation, and configuration of the OmniPod® Insulin Management System are not changed from those cleared in K122953 as a result of the modifications described in this pre-market notification.

Summary of Device Modifications

The following three device modifications were submitted in accordance with K95-1 Memorandum "510(k) Requirements for Proposes Fixes for Devices Undergoing Recall."

    1. Modify the component responsible for preventing the cannula from retracting after insertion into the skin tissue (e.g., remove catch wings from the insertion mechanism and replace them with a catch beam).
    1. Modify the dimensions of the insertion mechanism release bar to improve performance.
    1. Add medical grade silicone lubricant to the modified release bar to improve performance and correct an issue found through an increase in needle mechanism field complaints.

The three modifications were made to address the needle deployment mechanism recall Z-0393-2016 and Z-0394-2016. These modifications are described in detail, analyzed for their risk, and results of verification testing are summarized in this submission. The sum of these changes do not raise different questions of safety or effectiveness of the OmniPod® Insulin Management System.

{5}------------------------------------------------

Insulet Corporation

Performance Data

Insulet completed the appropriate validation and verification activities recommended by the Guidance for Industry and FDA Staff – Total Product Life Cycle: Infusion Pump – Premarket Notification [510(k)] Submissions Guidance and other guidance, as applicable. The following performance testing has confirmed the OmniPod Insulin Management System to be substantial equivalent to the predicate device.

  • Risk Management: was completed in accordance with /SO 14971:2007 Medical devices -. Application of risk management to medical devices. Verification activities, as required by the risk analysis, demonstrated that the predetermined acceptance criteria were met and the device is safe for use.
  • . Soft Cannula Depth and Angle Studies; testing was conducted to verify that the soft cannula insertion depth and angle of the Pod meets the specification. All Pod samples tested met the requirements.
  • . Cam finger; the cam finger design has been verified to perform as defined by its requirements.
  • Insertion mechanism release; testing was conducted to demonstrate that the Release Bar meets its design intent and releases the insertion mechanism. Testing confirmed that design intent was met and the Release Bar is capable of releasing the insertion mechanism.
  • . Cannula insertion retention; testing was conducted to demonstrate that the Catch Beam meets its design intent and retains the Soft Cannula at the specified depth. Testing confirmed that design intent was met and the Catch Beam is capable of retaining the Soft Cannula at the specified depth.
  • . Cannula insertion indicator; an indicator has been provided and tested to confirm when the cannula is fully inserted.
  • . Biocompatibility; an assessment of the additional silicone lubricant has been completed which rationalizes the additional silicone for human use as acceptable. The added silicone is not in contact with the fluid path. The same silicone is cleared with the predicate device for use at the needle tip/ needle cap interface which is in the fluid path. The predicate submission includes verification testing per ISO 10993. Part 3: Tests for genotoxicity. carcinogenicity, and reproductive toxicity, Part 5: Tests for in vitro cytotoxicity, Part 6: Tests for local effects after implantation, and Part 10: Tests for irritation and skin sensitization.

Substantial Equivalence Conclusion

The modified OmniPod® Insulin Management System uses the same technology, indications for use, materials, and modes of operation as the device cleared as K122953. The proposed modifications do not affect the intended use or technological characteristics of the predicate device. Performance bench testing demonstrated that the subject device met all device specifications. Therefore, the subject device is substantially equivalent to the predicate device.

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).