(136 days)
The Omron Heat Pain Pro is intended for:
The relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities.
Environments of Use: Clinics, hospital and home environments
Patient Population: Adult
The Heat Pain Pro is a small battery operated OTC device that provides a combination of transcutaneous electrical nerve stimulation (TENS) for pain relief and superficial heat for a warming sensation. It delivers TENS only or alternating combinations of TENS and heat. TENS and heat are never applied at the same time. The device can connect to a specified external IEC 60601-1 compliant power supply for charging of the internal lithium ion battery. The device complies with AAMI/ ANSI/ES60601-1, IEC 60601-1-2 and IEC 60601-2-10.
Stimulations from the device are intended for application to the following areas: lower back, arms, legs, shoulder or foot. Software controls all controls and indicators. Software controls waveform characteristics.
The accessories include: An external power adaptor, an electrode cord / cable attached to electrodes pads and a Pad holder for storage. The electrode pads Omron "HV-PAD with PMGEL" allow for stimulation and have resistive elements to provide heat which is powered by the device.
The device is battery powered and can be connected to an external power supply for charging the internal battery. The battery is lithium ion and is not user serviceable or accessible. The only external connections on the device are the power input and the electrode connector there is no connection to any other device.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Heat Pain Pro device:
The provided document is a 510(k) summary for the Omron Heat Pain Pro, a Transcutaneous Electrical Nerve Stimulator (TENS) device with heat for pain relief. The purpose of a 510(k) submission is to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving a new device's absolute safety and effectiveness through extensive clinical trials. Therefore, the "acceptance criteria" discussed here primarily relate to demonstrating equivalence in performance and safety to the predicate device, not necessarily a pre-defined clinical efficacy target like a reduction in pain scores.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by similarity to the predicate device (Omron PM3032, K141978) and compliance with specific electrical and medical device safety standards. The reported device performance is primarily a comparison of its electrical output parameters to those of the predicate.
| Acceptance Criteria (Implicit) | Reported Device Performance (Heat Pain Pro) |
|---|---|
| Functional Equivalence | |
| Indications for Use equivalent to predicate. | The Omron Heat Pain Pro is intended for: The relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities. (Virtually identical to predicate.) |
| OTC device. | Yes |
| Design and Technology equivalent to predicate. | Equivalent design and features compared to the predicate (PM3032). Combines TENS and heat through the same basic technology as reference device (ezFit, K070299). |
| Patient Population equivalent to predicate. | Adult (Identical to predicate PM3032) |
| Environment of Use equivalent to predicate. | Clinics, hospital and home environments (Identical to predicate PM3032) |
| Electrical Performance Equivalence (compared to PM3032) | Parameters are generally very close or within comparable ranges to the PM3032 across various modes (TAP, KNEAD, RUB, ARM, LBACK, LEG, FOOT, JOINT, SHLDR). See detailed tables {10-11} for specific values, including: - Waveform: Biphasic, Rectangular (matches predicate) - Max Output Voltage, Max Output Current across various loads - Duration of primary phase, Pulse Duration, Frequency - Net Charge, Max Phase Charge, Max Current Density, Max Average Power Density - Burst Mode parameters (Pulses per burst, Bursts per second, Burst duration, Duty cycle) - ON Time, OFF Time |
| Material Equivalence | Patient contacting materials (PM-GELs) constructed of the same materials as the patient contacting portion of the Long Life Pads cleared in K120516. |
| Safety and Regulatory Compliance | |
| Compliance with AAMI / ANSI ES60601-1 | Complies (ES60601-1:2005 + A1: 2012) |
| Compliance with IEC 60601-1-2 (EMC) | Complies (IEC 60601-1-2: 2007) |
| Compliance with IEC 60601-1-11 (Home Healthcare) | Complies (IEC 60601-1-11: 2015) |
| Compliance with IEC 60601-2-10 (Nerve/Muscle Stimulators) | Complies (IEC 60601-2-10: 2012) |
| Compliance with 21 CFR 898 | Yes |
| Patient Leakage Current within limits. | Normal Condition: 1 uA (matches predicate). Single Fault Condition: 5.52 uA max (predicate: 8.9 uA max - acceptable). |
| Average DC current through electrodes (no pulse) at 0 uA. | 0 uA (matches predicate). |
| Automatic No-Load Trip feature. | Yes (matches predicate). |
| Automatic Shut Off feature. | Yes (matches predicate). |
| Low Battery and Voltage/Current Level Indications. | Yes (matches predicate). |
| Timer Range (30 mins vs 15 mins for predicate is a difference, but considered insignificant). | 30 minutes (Predicate: 15 minutes). The document implicitly accepts this difference as insignificant in terms of safety or effectiveness, likely due to patient safety features like automatic shut-off and user override. |
| Drop, Vibration, and Environmental Temperature/Humidity testing. | Performed. |
| Usability testing performed. | Performed (to address OTC use and differences vs. Rx reference device). |
2. Sample Size Used for the Test Set and Data Provenance
The submission does not describe a clinical test set from a clinical study. Instead, the "test set" for the non-clinical performance evaluation appears to be the device itself subjected to a series of engineering and electrical tests. The provenance is the manufacturer's internal testing.
- Sample Size for Test Set: Not applicable in the context of a clinical test set. Performance was evaluated on the device itself against established electrical and safety standards.
- Data Provenance: The data is from non-clinical testing performed by the manufacturer, verifying the device's electrical output characteristics and compliance with standards. It is not patient data or human subject data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. There was no clinical ground truth established for a test set of patient data, as no clinical studies were performed. The "ground truth" for the non-clinical testing was defined by industry-recognized electrical and medical device safety standards and the performance parameters of the predicate device.
4. Adjudication Method for the Test Set
Not applicable, as there was no clinical test set requiring expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The document explicitly states: "No clinical testing was performed." Therefore, no MRMC study was done, and no effect size for human readers with/without AI assistance is provided or applicable.
6. Standalone Performance Study
Yes, in a non-clinical sense. The device's performance was evaluated in a "standalone" fashion by subjecting it to various engineering and electrical tests as described in the "Non-Clinical Testing Summary" (e.g., testing controls, indicators, battery state, waveforms, compliance with standards like ES60601-1). This is performance of the algorithm/device itself in a lab setting, not a standalone clinical performance study on patients.
7. Type of Ground Truth Used
The "ground truth" for the substantial equivalence claim is:
- Predicate Device Performance: The electrical output and functional specifications of the legally marketed predicate device (Omron PM3032).
- Regulatory Standards: Compliance with recognized medical device safety and electrical performance standards (e.g., AAMI / ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10).
- Reference Device Performance: The basic technology and combined TENS+heat capabilities of the ezFit Digital Heating TENS (K070299) as a reference for integrating heat functionality.
8. Sample Size for the Training Set
Not applicable. There is no mention of an algorithm based on machine learning or AI that would require a training set. The device's operation is based on predefined electrical stimulation parameters and heating control.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there was no training set for an AI/ML algorithm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, with three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2016
Omron Healthcare, Inc. % Paul Dryden Consultant Promedic. Inc. 24301 Woodsage Dr. Bonita Springs, Florida 34134
Re: K160115
Trade/Device Name: Heat Pain Pro Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: May 2, 2016 Received: May 3, 2016
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael J.Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
Heat Pain Pro
Indications for Use (Describe)
The Omron Heat Pain Pro is intended for:
The relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities.
Environments of Use: Clinics, hospital and home environments
Patient Population: Adult
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14
{3}------------------------------------------------
510(k) Summary Page 1 of 9 6/3/2016
| Omron Healthcare, Inc. | |
|---|---|
| 1925 West Field Court | Tel - 847-247-5626 |
| Lake Forest, IL 60045 USA | Fax - 847-680-6269 |
| Official Contact: | Renee Thornborough – Director QA/RA |
| Proprietary or Trade Name: | Heat Pain Pro |
| Common/Usual Name: | Transcutaneous electrical nerve stimulator for pain relief |
| Classification Name/Code: | 21CFR 882.5890 NUH - stimulator, nerve, transcutaneous, over-the-counter |
| Device Name: | Heat Pain Pro |
| Predicate Device:Reference Device: | K141978 - Omron - PM3032K070299 – ezFit Technology – ezFit Digital Heating TENS |
Device Description:
The Heat Pain Pro is a small battery operated OTC device that provides a combination of transcutaneous electrical nerve stimulation (TENS) for pain relief and superficial heat for a warming sensation. It delivers TENS only or alternating combinations of TENS and heat. TENS and heat are never applied at the same time. The device can connect to a specified external IEC 60601-1 compliant power supply for charging of the internal lithium ion battery. The device complies with AAMI/ ANSI/ES60601-1, IEC 60601-1-2 and IEC 60601-2-10.
Stimulations from the device are intended for application to the following areas: lower back, arms, legs, shoulder or foot. Software controls all controls and indicators. Software controls waveform characteristics.
The accessories include: An external power adaptor, an electrode cord / cable attached to electrodes pads and a Pad holder for storage. The electrode pads Omron "HV-PAD with PMGEL" allow for stimulation and have resistive elements to provide heat which is powered by the device.
The device is battery powered and can be connected to an external power supply for charging the internal battery. The battery is lithium ion and is not user serviceable or accessible. The only external connections on the device are the power input and the electrode connector there is no connection to any other device.
Intended User OTC
Indications for Use:
The Heat Pain Pro is intended for:
The relief of pain associated with sore or aching, muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household work activities.
{4}------------------------------------------------
510(k) Summary Page 2 of 9 6/3/2016
Environments of Use: Clinics, hospital and home environments Patient Population: Adult
Contraindications:
Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death.
Predicate Device Comparison:
The Omron Heat Pain Pro is compared to the predicates in the device comparison tables below.
{5}------------------------------------------------
510(k) Summary Page 3 of 9
6/3/2016
| Table 1 Device Comparison | ||
|---|---|---|
| ProposedOmron Heat Pain Pro | PredicateOmronPM 3032K141978 | ||
|---|---|---|---|
| Feature | |||
| Indications for Use | The Omron Heat Pain Pro isintended for:The relief of pain associatedwith sore or aching, muscles ofthe lower back, arms, legs,shoulder, or foot due to strainfrom exercise or normalhousehold work activities.Environments of Use: Clinics,hospital and home environmentsPatient Population: Adult | This device is intended forthe relief of pain associatedwith sore or aching, musclesof thelower back, arms, legs,shoulder or foot due tostrain from exercise ornormal household and workactivities. | |
| Over-the-Counter (OTC) | Yes | Yes | |
| Power Source(s) | External supply Lithium-ionbattery | AAA alkaline batteries x 2 | |
| - Method of Line Current Isolation | N.A.(internal power source) | N.A.(internal power source) | |
| - Patient Leakage Current | --- | --- | |
| - Normal Condition (uA) | 1 | 1 | |
| - Single Fault Condition (uA) | 5.52 max | 8.9 max | |
| Average DC current through electrodeswhen device is on but no pulse arebeing applied (uA) | 0 (uA) | 0 (uA) | |
| Number of output Modes | -3 Therapies(2 TENS with HEATcombination therapies and 1TENS therapy)-9 TENS Stimulation Modes-2 Heat Level Settings | 9 TENS output Modes | |
| Number of outputchannels | Synchronousor Alternating | 1 ch | 1 ch |
| Method ofChannelIsolation | None | None | |
| Regulated Current or Regulated Voltage | Regulated Current | Regulated Current | |
| Software/Firmware/MicroprocessorControl? | Microprocessor | Microprocessor | |
| Automatic Overload Trip? | No | No | |
| Automatic No-Load Trip? | Yes | Yes | |
| Automatic shut Off? | Yes | Yes | |
| User over ride control? | User activated On/Off | User activated On/Off | |
| Indication display | Yes | Yes | |
| Feature | ProposedOmron Heat Pain Pro | PredicateOmronPM 3032K141978 | |
| Low Battery? | Yes | Yes | |
| Voltage /CurrentLevel? | Yes | Yes | |
| Timer Range (minutes) | 30 | 15 | |
| Compliance with Voluntary standards? | ES60601-1, IEC60601-1-2,IEC60601-2-10, IEC 60601-1-11 - same as PM3032 –K141978 | ES60601-1, IEC60601-1-2,IEC60601-2-10, IEC 60601-1-11 | |
| Compliance with 21 CFR 898? | Yes | Yes | |
| Weight | Approx. 200g (incl. batteries) | Approx. 100g (incl.batteries) | |
| Dimensions (W x H x D) | 71(W)x165(H)x30.5(D)mm | 52(W)x112(H)x25(D)mm | |
| Operating and Storage Temperature,Humidity | Operating Temperature,10 to 40°C30 to 80 %RH700 to 1060 hPaStorage Temperature,0 to 40°C30 to 80 %RH700 to 1060 hPa | 10 to 40°C30 to 80 % RH | |
| Transportation Temperature,Humidity, Air Pressure | -20 to 60°C10 to 95% RH,700 to 1060 hPa | -20 to 60°C10 to 95% RH,700 to 1060 hPa | |
| Electrode style | PMGELReusable | Long Life PadsReusable | |
| Patient contacting accessory | Electrode | Electrode |
{6}------------------------------------------------
510(k) Summary Page 4 of 9 6/3/2016
{7}------------------------------------------------
510(k) Summary Page 5 of 9 6/3/2016
Table 2 Comparison to Reference Device - K070299 ezFit
| Feature | ProposedOmron Heat Pain Pro | ReferenceezFit TechnologyezFit - K070299 |
|---|---|---|
| Indications for Use | The Omron Heat Pain Pro is intendedfor:The relief of pain associated with sore oraching, muscles of the lower back, arms,legs, shoulder, or foot due to strain fromexercise or normal household workactivities.Environments of Use: Clinics, hospitaland home environmentsPatient Population: Adult | For Transcutaneous Electrical NerveStimulation, ezFit Digital Heating TENS(Model No.: HR-661/ UC-101) is intended for* Symptomatic relief and management ofchronic intractable pain.For powered heating therapy, ezFit DigitalHeating TENS (Model No .: HR-661/UC-101) is intended forTemporary relief of minor aches and pains andmuscle spasms |
| Compliance with Voluntary standards? | ES60601-1, IEC60601-1-2, IEC60601-2-10, IEC 60601-1-11 | IEC 60601-1, IEC 60601-1-2, ISO 14081 |
| Power Source | AC adaptor / Rechargeable battery(Lithium Ion) | AC adaptor / Rechargeable battery (Ni-H) |
| Functions | TENS and electrical heating viaelectrode pad | TENS and electrical heating via electrode pad |
| TENS Modes | Nine Modes | Not specified |
| Heating Setting | High and low | Adjustable |
| Heating temperature | High: 43°C maximumLow: 42°C maximum | 36-42°C |
{8}------------------------------------------------
510(k) Summary Page 6 of 9 6/3/2016
Discussion of Substantial Equivalence - Differences Between A Legally Marketed Predicate Device:
The Omron Heat Pain Pro is viewed as substantially equivalent to the predicate device because: The Heat Pain Pro uses the exact same technology and has substantially equivalent indications for use. The differences that exist between the devices are insignificant in the terms of safety or effectiveness.
Indications - These indications are virtually identical to the predicate.
Prescriptive - The Heat Pain Pro and predicate are all OTC.
Design and Technology - The Heat Pain Pro has equivalent design and features when compared to the predicate.
Performance and Specifications – The Heat Pain Pro has equivalent specifications of performance when compared to the predicate.
Compliance with standards -The Heat Pain Pro and predicate comply with the same standards: AAMI ANSI ES60601-1, IEC 60601-1-2. IEC 60601-2-10 and IEC 60601-1-11 for home healthcare.
Materials - The patient contacting materials of the Heat Pain Pro are the PM-GELs. The PM-GELs are constructed of the same materials as the patient contacting portion of the Long Life Pads as cleared in K120516.
Patient Population – The Heat Pain Pro and predicate are indicated for adults
Environment of Use - Clinics, hospital and home environments. Identical to the predicate PM3032
Rationale for Reference Device:
The Omron Heat Pain Pro is combines the ability to provide some surface heat as well as TENS. We selected a reference, ezFit Technology ezFit - K070299, as a reference device as it incorporates heat and TENS through the same basic technology. The differences that exist between the devices are related to heat therapy provided by the reference device vs. heat to provide a warming sensation.
Indications - The indications are similar except the reference device provides heat therapy vs. the subject device provide a warming sensation.
Prescriptive - The Heat Pain Pro is OTC vs. the reference is Rx.
Design and Technology - The basic design of providing heat and TENS via a single electrode is similar.
Performance and Specifications - The Heat Pain Pro has similar maximum delivered heat specifications and TENS stimulation.
{9}------------------------------------------------
510(k) Summary Page 7 of 9 6/3/2016
As indicated in Table 2, one can see that the subject device and reference device are similar and the difference between heat therapy of the reference device and providing a warming sensation for the subject does not raise any new safety concerns. The difference of OTC vs. prescriptive is has been addressed through our risk and hazard analysis and labeling as supported by the usability tested for the lay user population as an OTC device.
Non-Clinical Testing Summary:
The device has been tested to insure that all requirements have been met, this includes:
- Testing of all controls ●
- Testing of all indicators
- Testing of battery state indicators
- . Testing of waveforms
The device has also been tested to the requirements of the following standards:
- AAMI / ANSI ES60601-1:2005 + A1: 2012 Medical electrical equipment part 1: ● general requirements for basic safety and essential performance
- . IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 60601-1-11: 2015, medical electrical equipment part 1-11: general requirements ● for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-2-10: 2012 Medical electrical equipment - part 2-10: particular requirements for the basic safety and essential performance of nerve and muscle stimulators
The device has also been tested for drop, vibration and environmental temperature and humidity.
Usability testing was performed.
Clinical Testing Summary:
No clinical testing was performed
Substantial Equivalence Conclusion
Omron maintains that the Heat Pain Pro is substantially equivalent to the predicate device in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards.
{10}------------------------------------------------
510(k) Summary
Page 8 of 9
6/3/2016
| Parameter | Subject | PM3032 | Subject | PM3032 | Subject | PM3032 | Subject | PM3032 | |
|---|---|---|---|---|---|---|---|---|---|
| Mode of Program Name | TAP | TAP | KNEAD | KNEAD | RUB | RUB | ARM | ARM | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum Output Voltage [V] | @500Ω | 67.2 | 66.3 | 40.7 | 51.1 | 37.3 | 42.3 | 67.1 | 66.3 |
| @2KΩ | 85.6 | 87.6 | 56.6 | 67.9 | 51.2 | 55.9 | 86.0 | 87.9 | |
| @10kΩ | 95.9 | 95.9 | 62.4 | 73.9 | 56.7 | 59.9 | 95.7 | 95.9 | |
| Maximum Output Current [mA] | @500Ω | 134.4 | 132.6 | 81.4 | 102.2 | 74.6 | 84.6 | 134.2 | 132.6 |
| @2KΩ | 42.8 | 43.8 | 28.3 | 34.0 | 25.6 | 28.0 | 45.0 | 44.0 | |
| @10KΩ | 9.6 | 9.6 | 6.2 | 7.4 | 5.7 | 6.0 | 9.6 | 9.6 | |
| Duration of primary phase [usec] | 32 | 32 | 32 | 32 | 32 | 32 | 32 | 32 | |
| Pulse Duration [usec] | 96 | 100 | 96 | 150 | 96 | 150 | 96 | 100 | |
| Frequency [Hz] | 1 - 20.13 | 1 - 15.43 | 51.65 | 25.28 - 79.22 | 99 | 84.86 - 132 | 2 - 51.65 | 1 - 132 | |
| For multiphasic waveforms | Symmetrical phases | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| only: | Phase Duration | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| Net Charge(µC per pulse) (@500Ω) [uC] * | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (@500Ω) [µC] | 4.30 | 4.24 | 2.60 | 3.27 | 2.39 | 2.71 | 4.29 | 4.24 | |
| Maximum Current Density (@500Ω) [mA/cm²] | 8.43 | 3.03 | 5.10 | 2.33 | 4.68 | 1.93 | 8.41 | 3.03 | |
| Maximum Current Density (@500Ω) [mA/cm²] r.m.s. | 0.08 | 0.03 | 0.36 | 0.14 | 0.46 | 0.22 | 0.12 | 0.03 | |
| Maximum Average Power Density (@500Ω) [W/cm²] | 5.219E-09 | 0.00017 | 5.107E-06 | 0.00020 | 1.576E-05 | 0.00039 | 2.081E-08 | 0.00060 | |
| Burst Mode | (a) Pulses per burst | 60.4 | 46 | 222.1 | 71 | 544.5 | 118 | 60.4 | 23 |
| (b) Bursts per second | 0.32 | 0.33 | 0.12 | 0.59 | 0.09 | 0.59 | 0.32 | 1.00 | |
| **(c) Burst duration | 3.10 | 3.00 | 8.60 | 8.50 | 11.00 | 17.00 | 3.10 | 8.00 | |
| (d) Duty cycle: Line(b)xLine(c) | 1.00 | 1.00 | 1.00 | 5.00 | 1.00 | 10.00 | 1.00 | 8.00 | |
| ON Time (seconds) | 3.00 | 3.00 | 7.80 | 0.90 | 10.50 | 0.90 | 3.00 | 0.30 | |
| OFF Time (seconds) | 0.10 | 0.00 | 0.80 | 0.80 | 0.50 | 0.80 | 0.10 | 0.70 | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
Subject device is Heat Pain Pro
{11}------------------------------------------------
510(k) Summary
Page 9 of 9
6/3/2016
| Parameter | Subject | PM3032 | Subject | PM3032 | Subject | PM3032 | Subject | PM3032 | Subject | PM3032 | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mode of Program Name | LBACK | LBACK | LEG | LEG | FOOT | FOOT | JOINT | JOINT | SHLDR | SHLDR | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum Output Voltage [V] | @500Ω | 67.1 | 66.3 | 66.4 | 66.3 | 66.9 | 66.3 | 40.3 | 66.3 | 67.1 | 66.3 |
| @2kΩ | 84.0 | 87.9 | 85.5 | 87.9 | 85.9 | 87.9 | 56.8 | 87.9 | 84.8 | 87.9 | |
| @10kΩ | 94.0 | 95.9 | 95.5 | 95.9 | 95.8 | 95.9 | 63.1 | 95.9 | 95.4 | 95.9 | |
| Maximum Output Current [mA] | @500Ω | 134.2 | 132.6 | 132.8 | 132.6 | 133.8 | 132.6 | 80.6 | 132.6 | 134.2 | 132.6 |
| @2kΩ | 42.0 | 44.0 | 42.8 | 44.0 | 43.0 | 44.0 | 28.4 | 44.0 | 42.4 | 44.0 | |
| @10kΩ | 9.4 | 9.6 | 9.6 | 9.6 | 9.6 | 9.6 | 6.3 | 9.6 | 9.5 | 9.6 | |
| Duration of primary phase [usec] | 32 | 32 | 32 | 32 | 32 | 32 | 32 | 32 | 32 | 32 | |
| Pulse Duration [usec] | 96 | 150 | 96 | 100 | 96 | 100 | 96 | 100 | 96 | 100 | |
| Frequency [Hz] | 2 - 108 | 1 - 237.6 | 2 - 51.65 | 1 - 19.16 | 2 -11 | 1 - 19.16 | 42.43 - 108 | 42.43 - 237.6 | 1 - 19 | 1 - 19.16 | |
| For multiphasic waveformsonly: | Symmetrical phases | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| Phase Duration | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
| Net Charge(µC per pulse) (@500Ω) [µC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | |
| Maximum Phase Charge (@500Ω) [µC] | 4.29 | 4.24 | 4.25 | 4.24 | 4.28 | 4.24 | 2.58 | 4.24 | 4.29 | 4.24 | |
| Maximum Current Density (@500Ω) [mA/cm2] | 8.41 | 3.03 | 8.33 | 3.03 | 8.39 | 3.03 | 5.05 | 3.03 | 8.41 | 3.03 | |
| Maximum Current Density (@500Ω) [mA/cm2] r.m.s. | 0.12 | 0.04 | 0.12 | 0.03 | 0.12 | 0.03 | 0.32 | 0.20 | 0.08 | 0.03 | |
| Maximum Average Power Density (@500Ω) [W/cm2] | 2.081E-08 | 0.00015 | 2.038E-08 | 0.00040 | 2.069E-08 | 0.00016 | 3.379E-06 | 0.00015 | 5.203E-09 | 0.00082 | |
| Burst Mode | (a) Pulses per burst | 540.0 | 1425 | 258.3 | 57 | 55.0 | 57 | 324.0 | 1425 | 76.0 | 76 |
| (b) Bursts per second | 0.10 | 0.14 | 0.10 | 0.25 | 0.19 | 0.25 | 0.25 | 0.14 | 0.20 | 0.20 | |
| **(c) Burst duration | 20.00 | 14.00 | 10.00 | 8.00 | 5.20 | 8.00 | 16.00 | 14.00 | 10.00 | 10.00 | |
| (d) Duty cycle: Line(b)xLine(c) | 2.00 | 2.00 | 1.00 | 2.00 | 1.01 | 2.00 | 4.00 | 2.00 | 2.00 | 2.00 | |
| ON Time (seconds) | 20.00 | 6.00 | 9.00 | 3.00 | 5.00 | 3.00 | 11.00 | 6.00 | 8.00 | 4.00 | |
| OFF Time (seconds) | 0.00 | 1.00 | 1.00 | 1.00 | 0.15 | 1.00 | 4.00 | 1.00 | 2.00 | 1.00 | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
Subject device is the Heat Pain Pro
§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.
(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).