K Number
K160050

Validate with FDA (Live)

Date Cleared
2016-05-03

(113 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker SERFAS 90-S Electrosurgical Probe is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip, ankle, and wrist.

Device Description

The Stryker SERFAS 90-S Electrosurgical Probe (hereafter referred to as "Proposed device") is an accessory to the SERFAS Energy System, marketed through K041810, which is intended for resection, ablation, and coagulation of soft tissue via radiofrequency (RF) ablation. RF ablation probes are the main tool used in most arthroscopic procedures for the removal of tissue and the coagulation of bleeding vessels. The Proposed device is a disposable single-use electrosurgical device provided sterile via Ethylene Oxide sterilization.

AI/ML Overview

This document describes the Stryker SERFAS 90-S Electrosurgical Probe and its substantial equivalence to a predicate device. The information provided is primarily focused on demonstrating that the new device does not introduce new safety or effectiveness concerns compared to its predecessor.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly "Pass" for all tests, indicating the device met the internal design specifications and applicable performance standards.

Test NameDescriptionAcceptance Criteria (Implicit)Reported Device Performance
Aggressive UseDetermines if a probe can last its full lifetime without failure while being subjected to an extended activation period on tissue at the highest power setting.PassPass
Tip CantileverDetermines if the probe tip can withstand the normal, side, and back force required to permanently deform the lumen.PassPass
Electrode BendVerifies the electrode can withstand a 7lb force for a worst-case scenario.PassPass
Thermal Expansion (Heat) TestVerifies that the SERFAS probe tip assembly does not experience thermal expansion capable of causing catastrophic failure when subject to excessive heat.PassPass
LeakVerifies the condition of the ceramic-outer lumen assembly and the condition of the suction clamp of all styles of Stryker and non-Stryker RF ablation probes.PassPass
System CompatibilityVerifies that the electrosurgical probe is compatible with Stryker's legally marketed consoles.PassPass
BiocompatibilityAssessed in accordance with ISO 10993-1:2009 and related collateral standards for patient contacting materials. (Implicitly "Pass" as it contributes to substantial equivalence)PassPass
SterilizationAssessed in accordance with ISO 11135:2014 and ISO 10993-7 for sterilization residuals. (Implicitly "Pass" as it contributes to substantial equivalence)PassPass

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample size used for each "Test Name" listed (e.g., how many probes were subjected to the Aggressive Use test). It also does not provide details on data provenance (e.g., country of origin, retrospective or prospective data) as these are non-clinical (bench) tests rather than clinical studies with human subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. These were non-clinical bench tests performed against design specifications and industry standards, not assessments requiring expert interpretation of clinical ground truth.

4. Adjudication Method for the Test Set

Not applicable. The tests are objective measurements against defined criteria (e.g., pass/fail for mechanical stress, leak detection). There is no mention of an adjudication method typical for subjective clinical assessments.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document describes a medical device (electrosurgical probe), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC comparative effectiveness study involving human readers and AI is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for these tests are the internal design specifications of the device and applicable performance standards (e.g., ISO standards for biocompatibility and sterilization). The goal was to demonstrate that the device met these engineering and regulatory requirements.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not a machine learning or AI device that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 3, 2016

Stryker Corporation Somi Ekwealor Regulatory Affairs Analyst 5900 Optical Court San Jose, California 95138

Re: K160050

Trade/Device Name: SERFAS 90-S Electrosurgical Probe Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: April 1, 2016 Received: April 4, 2016

Dear Somi Ekwealor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K160050DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationPage 1 of 1Page 19 of 681
Indications for UseForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.
510(k) Number (if known)K160050
Device NameStryker® SERFAS 90-S Electrosurgical Probe
Indications for Use (Describe)The Stryker SERFAS 90-S Electrosurgical Probe is indicated for resection, ablation, and coagulation of soft tissue andhemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip,ankle, and wrist.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)Page 1 of 1PSC Publishing Services (301) 443-6740EF
-------------------------------------------------------------------------------

189 ງຸງ 10 6 ໂ ລອງປີ 1

{3}------------------------------------------------

1. General Information

510(k) SponsorStryker Corporation
Address5900 Optical CourtSan Jose, CA 95138
FDA Registration Number2936485
Correspondence PersonMr. Somi Ekwealor, MSRS, RACRegulatory Affairs AnalystStryker Corporation
Contact InformationEmail: somi.ekwealor@stryker.comPhone: (408) 754-2356Fax: (408) 754-2507
Date Prepared08 Jan 2016

2. Device Identification

Proposed Device:

Proprietary NameStryker® SERFAS 90-S Electrosurgical Probe
Common NameRF Probe
Classification NameElectrosurgical Cutting and Coagulation Deviceand Accessories
Regulation Number21 CFR 878.4400
Product CodeGEI
Regulatory ClassII

Predicate Device:

Proprietary NameSERFAS Energy System (probe only*)
Common NameRF System, RF Generator, RF Probe, RF Footswitch
Premarket NotificationK041810
Classification NameElectrosurgical Cutting and Coagulation Device and Accessories
Regulation Number21 CFR 878.4400
Product CodeGEI
Regulatory ClassII
  • The predicate device includes the entire energy system and accessories while the Proposed device includes the electrosurgical probe (90-S Redesign) only. This premarket notification refers to the SERFAS 90-S Redesign unless Original SERFAS 90-S is stated.

{4}------------------------------------------------

Device Description

The Stryker SERFAS 90-S Electrosurgical Probe (hereafter referred to as "Proposed device") is an accessory to the SERFAS Energy System, marketed through K041810, which is intended for resection, ablation, and coagulation of soft tissue via radiofrequency (RF) ablation. RF ablation probes are the main tool used in most arthroscopic procedures for the removal of tissue and the coagulation of bleeding vessels. The Proposed device is a disposable single-use electrosurgical device provided sterile via Ethylene Oxide sterilization.

3. Indication for Use

The Stryker SERFAS 90-S electrosurgical probe is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip, ankle, and wrist.

4. Comparison of Technological Characteristics with the Predicate Device

The Stryker SERFAS 90-S electrosurgical probe and predicate device are similar in technological characteristics and design. The predicate device includes the entire SERFAS Energy system while the Proposed device includes the probe only.

The differences between the Proposed device and the predicate are minor and raise no new questions of safety and effectiveness, therefore supporting that the SERFAS 90-S electrosurgical probe is substantially equivalent to the predicate device currently on the market. The following differences exist between the Proposed and predicate device:

  • 트 Ceramic and Electrode Key Feature Design
  • Electrode Face Suction Path
  • 트 Handle Design

Ceramic and Electrode Key Feature Design

Both the Proposed device and predicate device ceramic raw materials are identical but the devices have slightly different dimensional specifications. The intent of this change is to increase the mechanical robustness of the probe's tip and thereby address the failure mode of tip breakage. This difference is not critical to the intended surgical use of the device and does not impact the effectiveness of the device when used as labeled, however the safety of this product will be improved as a result of this change given the reduction in occurrence of tip breakage.

{5}------------------------------------------------

This minor difference in the Proposed device does not introduce additional risks and raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.

Electrode Face Suction Path

The Proposed device contains minor electrode face suction path design differences compared to the predicate device. Both the Proposed device and predicate device electrode raw materials are identical but the devices have slightly different dimensional specifications.

This minor difference in the Proposed device does not introduce additional risks and raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.

Handle Design

The patient contacting raw materials of the Proposed device and predicate device handle are identical. The Proposed device has a slightly different design specification for improved safety. This difference is not critical to the intended surgical use of the device and does not affect the effectiveness of the device when used as labeled. however the safety of this product will be improved as a result of this change because the occurrence of unintentional activation will be decreased.

This minor difference in the Proposed device does not introduce additional risks and raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.

5. Performance Testing

The Stryker SERFAS 90-S Electrosurgical probe was tested for performance in accordance with internal design specification and with the applicable performance standards, which established the substantial equivalence determination. The following non-clinical tests were conducted and are summarized in this premarket notification:

Test NameDescriptionResults
Aggressive UseDetermines if a probe can last its full lifetime without failurewhile being subjected to an extended activation period ontissue at the highest power setting.Pass
Tip CantileverDetermines if the probe tip can withstand the normal, side,and back force required to permanently deform the lumenPass

{6}------------------------------------------------

Test NameDescriptionResults
Electrode BendVerifies the electrode can withstand a 7lb force for a worstcase scenario.Pass
ThermalExpansion (Heat)TestVerifies that the SERFAS probe tip assembly does notexperience thermal expansion capable of causing catastrophicfailure when subject to excessive heatPass
LeakVerifies the condition of the ceramic-outer lumen assemblyand the condition of the suction clamp of all styles of Strykerand non-Stryker RF ablation probes.Pass
SystemCompatibilityVerifies that the electrosurgical probe is compatible withStryker's legally marketed consolesPass

Biocompatibility was assessed in accordance with ISO 10993-1:2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" and related collateral standards for patient contacting materials (see Section 15 -Biocompatibility).

Sterilization was assessed in accordance with ISO 11135:2014 - Sterilization of health-care products - Ethylene Oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. Sterilization residuals were assessed in accordance with ISO 10993-7 - Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (see Section 14 - Sterilization and Shelf Life).

Bench performance testing was conducted to ensure that the device functioned as intended and met design specifications and acceptance criteria. Test results obtained verified the safety and effectiveness of the device per design specifications and applicable standards (see Section 18 - Performance Testing - Bench).

6. Conclusion

Based on the intended use, technological characteristics, performance testing, and nonclinical testing in comparison to the predicate device, the Stryker SERFAS 90-S Electrosurgical Probe does not introduce additional risks and raises no new questions of safety and effectiveness, thus demonstrating substantial equivalence.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.